EyeSuite Biometry - Holladay IOL Consultant Software HICSOAP PRO LenStar... · 2019-09-11 ·...

8
EyeSuite Biometry V 1.1.1 Extract of Software Instruction Manual: - Connection to Holladay IOL Consultant Instruction Manual

Transcript of EyeSuite Biometry - Holladay IOL Consultant Software HICSOAP PRO LenStar... · 2019-09-11 ·...

EyeSuite Biometry V 1.1.1

Extract of Software Instruction Manual: - Connection to Holladay IOL Consultant

Instruction Manual

How to connect to 3rd party IOL calculation softwareEyeSuite is able to connect to third party IOL calculation software. In this case EyeSuite is acting as a pureprovider of measurement data to the third party software. Haag-Streit is no liable for any output of the thirdparty software.

Connection to Holladay IOL Consultant

The Holladay IOL Consultant & Surgical Outcomes Assessment is a "smart" database program thatautomatically minimizes a surgeon's prediction error for intraocular lens power calculations every timepostoperative results are entered. The program can handle an unlimited number of IOL's, surgeons andpatients. (Source: www.hicsoap.com/)

EyeSuite settings for Holladay IOL Consultant

To connect to Holladay IOL Consultant, EyeSuite should be set-up as follows:

ModeTo connect to Holladay IOL Consultant this should be set to Automatic

FormatTo connect to Holladay IOL Consultant this should be set to EyeSuite Template Format

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :42/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

TemplateTo connect to Holladay IOL Consultant the template Holladay_Template.xml has to be selected.

The template can be found in the folder templates of the installation path of EyeSuite (standard path for 32 bitWindows systems: C:\Program Files\Haag-Streit\EyeSuite\templates; standard path for 64 bit Windowssystems:C:\Program Files (x86)\Haag-Streit\EyeSuite\templates) or can be downloaded form the Haag-Streithomepage in the product section LENSTAR (www. haag-streit.com).

Output FolderTo connect to Holladay IOL Consultant this should be set e.g. to C:\HolladayII, or any other system folderEyeSuite and Holladay IOL Consultant can access.

FilenameTo connect to Holladay IOL Consultatn this should be set to:{TIMESTAMPSTD}_{ID}_{NAME}_{FIRSTNAME}_{BIRTHDAYDAY}_{BIRTHDAYMONTH}_{BIRTHDAYYEAR}.xml

ExecuteTo connect to Holladay IOL Consultant this should be empty

Execute inTo connect to Holladay IOL Consultant this should be empty

Generate PDFsTo connect to Holladay IOL Consultant no PDF are needed and therefore no PDF output should be selected

Settings in the Holladay IOL Consultant

To connect to Holladay IOL Consultant, an import directory has to be specified in the Holladay IOL Consultant:

Open Holladay IOL Consultant

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :43/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

Click on Locate Patient Record

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :44/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

Click on Import Patient(s) from IOL Master or Lenstar...

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :45/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

Click on Add Lenstar...

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :46/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

Select the folder specified in the settings of EyeSuite as the export forlder for the Holladay IOL Consultatn, e.g.C:\Holladay_II

Instruction Manual EyeSuite Biometry"6. General program functions"Doc. Nbr. : 1500 7220056 06011Page :47/84

Wiki RevisionID: 48916 created: 07.01.2009 IW / changed: 04.06.2010 TBe / reviewed:04.06.10 JBr approved: 04.06.10 TBe

The system is now set to import any exam taken by the LENSTAR into the Holladay IOL Consultant.