Expanded Pupil Eyepieces user guide.pdf · the left and right eyepieces, to obtain optimum focus of...
Transcript of Expanded Pupil Eyepieces user guide.pdf · the left and right eyepieces, to obtain optimum focus of...
. . . Expanded Pupil Eyepieces
USER GUIDE
CONTENTS
Unpacking Page 1
Assembly & Attachment Page 1
Set-up Microscopes with Zoom Page 2
Set-up Microscopes without Zoom Page 3
Graticule Fitting Page 3
Routine Maintenance Page 4
Catalogue Numbers Inside back page
UNPACKING
The ISIS and all its accessories areshipped in a single customised pack.
Remove the ISIS system from thepacking case and check that all therelevant parts are present (seeFigure 1).
O The items shown in Figure 1 will
only be present if they are
necessary for the ISIS system
ordered (see Figure 2 and
Table 1).
STAGE 1:ASSEMBLY & ATTACHMENT
1. Refer to Figure 2 to identify the
parts and their locations.
2. Refer to the Stereo Intermediate
Adapter column (SIA) of Table 1.
If no adapter is required for the
host microscope then continue
from step 3. If an adapter is
required then the binocular head
of the host microscope should be
removed and the Stereo
Intermediate Adapter fitted
between the microscope and the
binocular head (see Figure 2).
3. Refer to the Extending
Lens(es) column of Table 1 for
selection of the appropriate
lens(es) to suit the host
microscope.
O The X value shown is the equivalent of eyepiece magnification.
4. To fit the Extending Lenses, first remove the two Protective Caps then screw the extending lenses into the exposed
threads in the centre of the parfocal control (see Figure 3). Taking care not to touch the optical disc and not to
over-tighten the Extending Lenses.
5. Refer to the Parfocal Spacer column (PSP) of Table 1. The spacers (if required) slide over the extending lenses.
6. Set the Parfocal Control to mid travel and slightly slacken the Link Knob (see Figure 3).
1
Graticule
ISIS
ExtendingLenses
StereoIntermediateAdaptor
UserGuide
DC PowerSupply
ParfocalSpacers
ExtendingLenses
Figure 1 Packing Contents
Host MicroscopeHost Microscope
Binocular HeadBinocular Head
00++
ISIS UniversalBinocular Head
Graticule
ExtendingLens Pairs
Parfocal Spacer Pairs (PSP)
Stereo Intermediate Adaptors (SIA)
Binocular TubeInterface (BTI)
Figure 2 ISIS Parts Identification & Location
7. Remove the standard eyepieces from the microscope tubes. If the microscope has diopter adjustment on one or
both of its eyepiece tubes these will need to be set to the nominal zero position. If the microscope has thumb
screws to retain the eyepieces which appear to interfere with the ISIS these should be removed.
8. Slide the ISIS into the microscope tubes ensuring the assembly fits against the end of the microscope tubes.
9. Set the Eyepiece Focus Control (see Figure 3) on each eyepiece to the nominal position indicated by the dotted
line.
10. The method of initial setting of the ISIS differs slightly between microscopes with or without a zoom system. The
two methods are described in stage 2.
O If you normally wear spectacles for close work then for maximum comfort and image clarity they should also be worn
when using and setting up the ISIS.
STAGE 2:SET-UP MICROSCOPES WITH ZOOM (Refer to Figure 3)
1. Switch the microscope illumination on, view through the ISIS and if necessary adjust the Eyepiece Focus Control on
the left and right eyepieces, to obtain optimum focus of the granular optical surface of the ISIS disk.
O At this stage the ISIS power is not connected and the image from the microscope may be out of focus.
2. Connect the ISIS DC Power Supply to a suitable mains supply and connect its two jack plugs as shown in Figure 3.
Check the red power indicators are illuminated and the granular optical surface can no longer be seen when viewing
through the eyepieces.
3. View through the right hand eyepiece and focus the microscope with the zoom set at maximum magnification. Then
set the zoom to minimum magnification (do not readjust the focus control) and set the axial position of the ISIS in
the microscope tube by adjustment of the right hand Parfocal Control until the image is clear.
4. Set the zoom to maximum magnification and if the image focus is not clear, repeat step 3.
2
II SSII
SS
--
PPAA
RRFF
OOCC
AALL
SS
PPAA
CCEE
RR
Eyepiece FocusControl
Link Knob
Power Indicators
Parfocal Control
Extending Lens
Parfocal Spacer
Protective Cap
12V DC Jack Plug
12V DC Jack Plug
ParfocalShim
Figure 3 Fitting the Extending Lens & Parfocal Spacer (if necessary)
5. Without refocusing the microscope, set the zoom to minimum magnification and view through the left hand
eyepiece and adjust the left hand Parfocal Control until the image is in focus.
6. Hold the microscope binocular head and adjust the eyepiece centre distance for comfortable viewing. Lightly
tighten the Link Knob and apply gentle downward pressure on top of each ISIS unit to ensure image alignment.
Finally adjust the Eyepiece Focus Control on each side for image clarity, and the ISIS is ready for use.
STAGE 2:SET-UP MICROSCOPES WITHOUT ZOOM (Refer to Figure 3)
1. Switch the microscope illumination on, view through the ISIS and if necessary adjust the Eyepiece Focus Control on
the left and right eyepieces, to obtain optimum focus of the granular optical surface of the ISIS disk.
O At this stage the ISIS power is not connected and the image from the microscope may be out of focus.
2. Connect the ISIS DC Power Supply to a suitable mains supply and connect its two jack plugs as shown in Figure 3.
Check the red power indicators are illuminated and the granular optical surface can no longer be seen when viewing
through the eyepieces.
3. View through the right hand eyepiece and adjust the microscope focus to obtain a clear image.
4. Without adjusting the microscope focus controls check the image through the left hand eyepiece and if necessary
adjust the Parfocal Control to obtain optimum focus.
5. Hold the microscope binocular head and adjust the eyepiece centre distance for comfortable viewing. Lightly
tighten the Link Knob and apply gentle downward pressure on top of each ISIS unit to ensure image alignment.
Finally adjust the Eyepiece Focus Control on each side for image clarity, and the ISIS is ready for use.
6. To maintain optimum focus with the ISIS, POS shims should be fitted between the objective and the microscope
nosepiece for all objectives with a magnification of x20 or less (see inset of Figure 3 and Table 1).
GRATICULE FITTING (refer to Figure 4)
The ISIS has the facility for a reference graticule to be fitted to either the left or the right optical path. Select in whichside you wish to fit or exchange a graticule and proceed as follows, referring to Figure 4:
1. Disconnect the two DC Jack Plugs and withdraw the ISIS from the microscope.
2. Remove the three Case Securing Screws and carefully lift the rear cover away from the assembly.
3. Remove the three Adapter Screws and remove the Adapter.
4. Remove the Field Stop or the previously fitted Graticule Cell and store safely for future use.
5. Fit the Graticule Cell into the main casting and ensure that the gear teeth engage.
O Take care not to touch the Optical Disc or the Graticule surface.
6. Replace the Adapter and its 3 screws, ensuring the Graticule rotates freely after finally tightening the screws.
7. Replace the rear cover and the three Case Securing Screws.
8. Return the ISIS to the microscope ready for use or set up as previously described.
3
ROUTINE MAINTENANCE
Periodically clean the ISIS eyepieces with the cloth supplied. Do not use any form of cleaning liquid. No other routinemaintenance of the ISIS is required.
4
Extending LensAdapter
Case SecuringScrews
AdapterScrews
Field Stop or RotatingGraticule Cell if fitted
Rotating GraticuleControl
Optical Disc
Rotating Graticule Controls(only 1 in use)
12V DC Jack Plug
12V DC Jack PlugRear Cover
Figure 4 Fitting the Graticule (right hand side shown)
WARRANTY
This product is warranted to be free from defects in material and workmanship for a period of one year from the date ofinvoice to the original purchaser.
If, during the warranty period the product is found to be defective, it will be repaired or replaced at facilities of VisionEngineering or elsewhere, all at the option of Vision Engineering. However, Vision Engineering reserves the right torefund the purchase price if it is unable to provide replacement, and repair is not commercially practicable or cannot betimely made. Parts not of Vision Engineering manufacture carry only the warranty of their manufacturer. Expendablecomponents such as lamps and fuses carry no warranty.
This warranty does not cover damage caused in transit, damage caused by misuse, neglect, or carelessness, or damageresulting from either improper servicing or modification by other than Vision Engineering approved service personnel.Further, this warranty does not cover any routine maintenance work on the product described in the instruction manual orany other minor maintenance work which is reasonably expected to be performed by the purchaser.
No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage, or other conditions beyond the control ofVision Engineering.
Except as stated herein, Vision Engineering MAKES NO OTHER WARRANTIES, EXPRESS OR IMPLIED BY LAW, WHETHER OR MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHERWISE. Further, Vision Engineering shall notunder any circumstances be liable for incidental, consequential or other damages.