Exp 3 emg sim and stim
-
Upload
bvm-engineering-college -
Category
Engineering
-
view
58 -
download
1
Transcript of Exp 3 emg sim and stim
PRACTICAL NO: 3
Aim: To study EMG SIMULATOR & STIMULATOR.
Material required:
[1] Matrix trainer kit-EMG SIMULATOR & STIMULATOR.
[2] DIGITAL STORAGE CRO.
Procedure:
Setup-
[1] Keep the POWER switch in off condition.
[2] Connect power chord to mains (230 volts A.C).
[3] Connect CRO at the output at T1 w.r.t Ground T2.
[4] Set the required rate by P1.
Experiment-
[1] Put on the trainer & CRO.
[2] Note the output waveform.
[3] Select X10 MODE on CRO note voltage waveform across T4 w.r.t T3
(Jumper J1 at B)
[1]Note waveform amplitude across T4 w.r.t T3.
Observation:
ENG SIMULATOR & STIMULATOR
INTRODUCTION:
A system is able to produce signals, as some other system is known as simulator.
Biomedical simulator is those are producing artificial signal, like nature part produce
in live body. The Contraction of skeletal muscle results in the generation of potentials
in individual muscle. The action potential lasting only for few milliseconds. The
electrical activity of the under lying muscle mass can be observed by means of surface
electrodes on the skin. In voluntary contraction of skeletal muscle, the muscle
potential range from 50µV To 5mV and duration 2 to 15mS.
This signal known as the Electromyogram. This signal is artificially generated by
means of electronics components by EMG simulator.
STIMULATOR: stimulator used to stimulate muscle under test. Stimulator provide
single or double or train of pulse. Stimulator amplitude, duration, repetition & delay
are adjustable. The o/p is either constant voltage & constant current type. The
constant voltage type stimulator provides rectangular pulse with amplitude 0-500V.
CIRCUIT DESCRIPTION:
Fig3.2 show circuit of EMG STIMULATOR & SIMULATOR. It consists of a pulse
generator based on time: IC LM555. The frequency of oscillation is variable through
P1.
Jumper J1 disable stimulator o/p when it is on a position. This jumper breaks power
supply to switching transistor through Q1.
When jumper J1 on a position circuit works only as EMG stimulator. Combination
differentiator integrator circuit by C4,R3 & C5, does this. The signal looks like as
below.
TYPICAL EMG SIGNAL:
Fig.3.1: EMG signal
When juper J1 on B position transistor Q1 coupler power supply to transistor Q2
through pulse transformer Tx1. Each pulse differentiates by C2 and D1 and switch on
q1. Moment Switching ON of Q2 produce High voltage Stimulator pulse across
secondary of TX1 (TX1 is a step-up transformer with 1:30 ratio).
This output is used to stimulate Nerve at the time of EMG.
Figure: 3.2: EMG SIMULATOR AND STIMULATOR
Conclusion: