Example- FDA Proposal Final Junior Year

download Example- FDA Proposal Final Junior Year

of 16

Transcript of Example- FDA Proposal Final Junior Year

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    1/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 1

    Design Verification ProposalFood and Drug Administration

    Submitted under guidelines established by 21 CFR 820.30

    Effectiveness of Polylactic Acid Polymers in Biodegradable Stents

    for Use in Coronary Arteries

    Team Woosa

    1/28/11

    David Baruela

    Samyukta Gade

    Jennifer Go

    Alex Hussinger

    Austin Schader

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    2/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 2

    Report Title:Effectiveness of Polylactic Acid Polymers in Biodegradable Stents for

    Use in Coronary Arteries

    Product: Biodegradable StentsSubject: Biodegradation, tensile strength and pH

    1. EXECUTIVE SUMMARY

    Stents have evolved from the simple bare metal design to the more complex, drug

    eluting design. However, there are a few drawbacks to each of these products, such as

    encouraging restenosis (re-narrowing) to occur or even thrombus formation; these serious

    issues have prompted groups such as ours to find a solution to the problems presented.

    The patient is at his or her greatest risk for restenosis in the first 6 months after the

    angioplasty, but by leaving a stent in too long we would run the risk of dangerous blood clot

    formation. We think that by creating completely biodegradable stents, we will be able to

    reduce the risk of clot formation in the blood vessels by both removing the surfaces the

    body perceives as foreign, and by eventually having the stents degrade away.

    Our design will follow the ASTM standards and FDA regulations for bare metal stents in

    order to compare the effectiveness of our product to stents currently used in industry. Our

    stent will undergo tensile testing and degradation tests to simulate the environment found in

    coronary arteries. Our samples will be tested for strength after being degraded in order to

    verify that they maintain structural support to the artery after a minimum of six months.

    Afterwards, we will analyze and extrapolate the data to simulate the ideal lifetime of the

    stent, six to twelve months.

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    3/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 3

    2. DESIGN AND DEVELOPMENT PLANNING

    2.1. Introduction

    2.1.1. Medical Condition

    81.1 million people were diagnosed with heart disease in 20061

    . Of those, one inone hundred of them died. The main cause of heart disease is stenosis, or closing

    of one or more of your arteries, due to buildup of plaque along the endothelial lining.

    Stenosis has become one of the most common and dangerous internal injuries. As

    a result of this problem doctors used angioplasty to unclog the blockages, although,

    restenosis occurred in 40% of the cases within one year.

    2.1.2. Current Technology

    Together with doctors engineers created stents a small wire mesh frame

    cylinders in order to reduce the chance of restenosis. Since their inception stents

    have evolved from bare metal platforms, to drug eluding stents that further reduce

    chances of restenosis. Eventually stents will transition to fully biodegradable stents

    made from biodegradable polymers. The implantation of stents has grown to

    become the most common surgical procedure and the industry has grown to multi-

    billion dollar proportions.

    2.1.3. Current Limitations

    Bare metal stents and drug-eluding stents have greatly decreased the chance of

    restenosis; however, they are not without their drawbacks. Bare metal stents trigger

    and auto-immune response that can lead to restenosis as the body tries to destroy

    the foreign object (stent). Drug-eluding stents emit drugs to prevent the auto-

    immune response; however, this can lead to blood clots that may travel to the heart

    or brain causing heart attack or stroke.

    2.2. Problem Statement

    2.2.1. Problem Statement

    There currently is a necessity for a geometrically similar device that does notcause blood clots while maintaining the necessary strength to keep the artery open

    and minimize the auto-immune response.

    1Cardiovascular Disease Statistics. American Heart Association.

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    4/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 4

    2.3. Project Goals, Solution and Objectives

    2.3.1. Team Goals

    Polymers chosen for biodegradable stents must exhibit strength near those of the

    metal stents and biodegrade with six months to one year (the ideal time frame to

    prevent restenosis or collapse of the artery and allow the body to repair the

    damaged artery).

    2.3.2. Proposed Design Solution

    The proposed design solution is a collapsible, biodegradable Polylactic acid

    based stent that degrades fully within one year and maintains structural strength for

    at least six months after the angioplasty.

    2.3.3. Objectives

    Our project seeks to test biodegradable polymers (Polylactic acid based) for the

    rate of degradation and for the fall in strength vs. % degradation. These two

    characteristics represent the most important factors for a biodegradable stent.

    2.4. Testing Justification and Overview

    2.4.1. Brief description

    During our project will test seven PLA polymer strips approximately 4 inches in

    length and 3/4 inch in width and 1/8 inch. These strips will be fabricated from plastic

    biodegradable spoons and melted down in a furnace at 250 C then poured into

    Styrofoam molds lined in aluminum foil, and will then be sanded into the ideal shape

    as specified above.

    After forming, the strips will be placed into a saline solution with a measured pH

    using either pH probe or pH strips. This solution will be held at 37 2C. These

    strips will then be tested one per day for seven days in the Instron Tensile Tester

    and the hardness tester using the Vickers scale. Additionally they will be measured

    using calipers and weighed on a balance.

    2.4.2. Verification

    These tests will be conducted to monitor the strength and amount of material lost

    over time and to confirm the strength is adequate for the first six months of

    implantation. Additionally, the amount of sample lost will be used to verify that the

    sample will be completely biodegraded within the one year timeframe.

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    5/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 5

    2.4.3. FDA Approval

    The FDA should approve PLA for use in stents if it meets or exceeds the tests

    below (section 5) as it has demonstrated adequate strength and biodegradability for

    the human body. Additionally, drug coatings may reduce chances of forming a

    thrombus and further research need to be conducted into combinations of drug-

    eluding biodegradable polymers.

    2.5. Important Terminology

    2.5.1. Technical or Medical Terms

    Artery-blood vessels that carry blood away from the heart

    Catheter-tube inserted into body cavity

    Coronary-having to do with the heart

    Endothelial-cells that line that inner walls of blood vessels

    Endothelialization-when the endothelial cells grow over the implant and incorporate

    it into the body

    Macrophage-white blood cells that engulf and digest pathogens

    Mitotic inhibitors-a drug that inhibits cell division

    Plaque-in blood vessels consists of accumulation of ruptured macrophages,

    calcium and cholesterol

    Restenosis-formation of blockage after the artery has been treated with angioplasty

    Stenosis-narrowing of blood vessel due to formation of plaque

    Thrombus-blot clot formed within a blood vessel

    2.6. Organizational Responsibilities

    2.6.1. Team Member Responsibilities

    The main project responsibilities were divided into four roles: the delegator,

    budget analyst, Gantt chartist, and Computer Aided Designers. The delegator,

    Jennifer Go, will be the liaison between the team and our coaches, Dr. Harding and

    Dr. Vanasupa. She will ask questions on behalf of the team, and check in with the

    coaches periodically as the project moves forward. David Baruela will do the

    budget analysis and make sure that the team is within budget and that the rest of

    the project can be completed with the remaining resources. He will keep an up to

    date spreadsheet of bill of materials. Samyukta Gade will create a Gantt chart

    detailing individual tasks for each team member, critical milestones, and associated

    completion dates. She will also make sure all deadlines are met and update the

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    6/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 6

    Gantt chart if any obstacles arise. Austin Schader and Alex Hussinger are in charge

    of computer aided design of stent.

    In addition to the roles listed above, each team member will have an assigned

    task to be completed every week. These tasks and their due dates are listed in

    section 2.7.3.

    2.7. Document Tracking, Deliverables and Timeline

    2.7.1. Tracking and Maintaining Important Documents

    A Google Groups page has been created to keep track of all documents related

    to this project. Documents include the team contract, related articles, list of sources,

    related terminology, and deliverables. The list of sources will be updated whenever

    a new source is used in the project. Related terminology will be continuously

    updated every time a team member comes across medical term needed forfundamental knowledge of stents. These documents can be read and altered by any

    member of the team.

    2.7.2. Publishing Test Results

    Test results will be published as a newsletter and sent to all team members at

    the end of the testing period. This is an effective way to distribute a summarized

    report that the team members can review as well as the FDA.

    2.7.3. Timeline

    The project was broken

    down into three parts: FDA

    proposal, testing, analysis, and

    presentation. We allocated one

    week for gathering materials

    and one week for testing. An

    extra was set aside just in

    case unforeseen delays or

    obstacles occur during the

    ordering or testing of material.

    The list of tasks along with

    tentative completion dates are

    in figure 1 and a Gantt chart

    depicting critical milestones is

    shown in figure 2.

    Figure 1: List of tasks along with tentative completion dates

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    7/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 7

    Figure 2: Gantt chart depicting several milestones and a critical path

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    8/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 8

    3. DESIGN INPUTS

    3.1. User and Patient Needs

    The surgeon needs a stent that is small enough to fit inside an artery, but will be

    able to hold open the artery for six to twelve months. The patient needs a coronarystent that is strong enough to hold open an artery and stiff enough to keep its shape

    for 6 to 12 months. The stent will need to fully degrade in 6 to 12 months.

    3.2. Performance Requirements

    3.2.1. Functional Requirements

    The stent must be able to fit inside the coronary artery. The patient needs a stent

    that is able to hold open the artery for 6 to 12 months. The stent must fully degrade

    after 6 to 12 months. The stent must not be toxic to humans. The stent must be

    able to keep its strength while undergoing degradation in the artery. However, the

    stent must not degrade until it is fully deployed within the artery.

    3.2.2. Service Environment

    The stent will be inside a coronary artery. The stent will be exposed to blood,

    oxygen, and plaque. The stent will not be exposed to a lot of motion, since it will be

    placed in the heart. However, the stent will be subject to the cyclic pressure

    changes normally seen in the heart.

    3.2.3. Regulatory Requirements

    There are currently mandatory FDA regulations and voluntary ASTM standards in

    place for coronary stents. However, FDA regulations are not as strict as the ASTM

    standards. The stent diameter must be measured to the nearest 0.25 or 0.5 mm. 2

    The stent length must be measured to the nearest 1mm. Standard sizes for stents

    undergoing tests are 8 or 24 mm long and 2.5 or 4 mm in diameter3. The stent

    must not undergo a length change greater than 1% from the undeployed state to

    the deployed state. However, stents that have been deemed clinically useful can

    have a shortening of up to 20%. The strut thickness should be measured to the

    nearest 0.013 mm. The average strut thickness is between 0.025 0.177 mm. The

    2F2081 Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular Stents

    3 Guidance for Industry and FDA Staff Non-Clinical Engineering Tests and Recommended Labeling for

    Intravascular Stents and Associated Delivery Systems

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    9/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 9

    stent must not be toxic to humans4. When undergoing a three-point bending test, a

    stent that is 20-24 mm long cannot deflect more than 3.2 mm5.

    3.3. Design Input Assessment Plan

    Our design will be measured following the ASTM instructions for vascular stents,

    and will follow the FDA regulations for stents. The design will undergo tests similar to

    those found in ASTM standards for testing balloon expanded stents, bare metal

    stents, and drug-eluting stents.

    4 Guidance for Industry Coronary Drug-Eluting Stents Nonclinical and Clinical Studies5 F2606 Standard for Three-Point Bending of Balloon Expandable Vascular Stents and Stent Systems

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    10/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 10

    4. PROPOSED SOLUTION

    4.1. Conceptual Design Description

    Biodegradable stents are an attractive alternative to bare metal and drug eluting

    stents which are often rejected by the users body or cause a thrombus which could

    have a cascade of negative effects, respectively. Currently, there are many

    biodegradable materials out on the market however the material which is most

    appealing for use in our project is Polylactic acid (PLA) based polymers. As stated

    previously, PLA polymers are completely biodegradable if given enough time and are

    contained in an oxygen-rich environment. They are very biocompatible with the

    human body (see section 4.3 below), and can serve as vectors for drugs to provide

    the same function and benefits that a drug eluting stent does without the drawbacks.

    Also, there are various forms of PLA polymers which give our experiment flexibility

    when choosing our specific polymer. Overall, PLA polymers have strength

    comparable to similar polymer, 6-8% ductility, and have a low modulus of elasticity,

    which would provide enough support for a stent without being brittle. They also can

    theoretically support a blood vessel which is undergoing constant pressure and

    sometimes load deformation.

    Based on the limited flexibility of PLA polymers, the traditional diamond shaped

    strut stent design cannot be used. PLA is not suited to be put into place by plastic

    deformation. Instead, a new stent design will be used in tandem with the old method

    of stent implantation to get the stent into the stenotic artery. The PLA stent will be

    made of square shaped struts which are overlapped onto one another when the stent

    is collapsed. If the stent needs to be placed in a stenotic artery the collapsed stent

    will be threaded onto a catheter and guided to the blockage, once the stent/catheter

    has reached the blockage the balloon at the end of the catheter will expand, pushing

    the rings out on the stent. The rings of the stent will lock into place with a latch

    mechanism that catches once the stent struts fall into place, causing the stent to

    maintain its expanded shape.

    A PLA polymer stent will rectify the problems encountered with traditional stents

    because over time the polymer and thus stent will biodegrade. The rate at which PLA

    degrades is slow enough so that it is able to undergo endothelialization and be

    integrated into the vessel wall to provide support for the surrounding vessel; but once

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    11/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 11

    the stents job is complete, it will degrade and not cause any long term lasting effects.

    PLA polymers are biocompatible which means that for the majority of cases, unlike

    bare-metal stents, the body will not reject the stent. Additionally, since the PLA stent

    can be incorporated with a drug of some-sort, the rate of restenosis (re-narrowing),

    which is a common occurrence in a bare metal stent, of the vessel in question can be

    decreased dramatically. Also, unlike drug-eluting stents, thrombus formation with

    PLA stents is minimized or completely absent 6.

    As shown, a PLA stent would provide many, if not all, of the benefits that

    traditional stents provide without all the negative drawbacks. Manufacturing a stent of

    pure PLA polymer is still in the research stages but due to its beneficial properties,

    PLA is an attractive option of stent use.

    4.2. Design Performance

    Our completely PLA polymer stent will be used in the same way traditional stents

    are, although their design and method of expansion are novel. The completed stent

    will be threaded onto the end of a catheter guide wire and led to the location of

    implantation, typically where the blockage is located in one of the coronary arteries.

    The balloon will be expanded and in doing so flatten the components of the blockage

    against the sides of the vessel wall as well as expand the PLA stent and hold it into

    place. Once the PLA stent is in place, it will start to release the drug which has been

    integrated with the stent. These drugs can range from mitotic inhibitors which inhibit

    macrophage multiplication, to immunosuppressants which prevent the body from

    attacking and rejecting the stent. With the stent in place, the vessel wall will be less

    likely to collapse and the vessel will be clear for blood to flow normally. Over the

    course of the next two to three months, the stent will start to become endothelialized

    and incorporated into the vessel. The stent will also start to degrade once implanted

    which prevents thrombus formation. It will take the stent anywhere from six months,

    to one year to fully biodegrade; this time frame is long enough to provide the

    necessary support as well as short enough to not cause any long term effects from a

    foreign object located within the body.

    6Zamiri, P. Kuang, Y. Sharma, U. Ng, T.F. Busold, R.H. Rago, A.P. Core, L.A. Palasis, M. The biocompatibility of

    rapidly degrading polymeric stents in porcine carotid arteries.

    Biomaterials 8 August 2010

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    12/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 12

    4.3. Patient Safety Assurance

    Polylactic acid polymers are biocompatible and are bioabsorbable under aerobic

    conditions. Since we intend on placing the PLA stent in the coronary artery, which is

    an oxygen-rich environment, we can expect the polymer to be able to degrade

    naturally.

    In studies performed on stenotic porcine carotid arteries, metal stents laced with

    different forms of PLA polymer to observe the effects of this material in vivo, returned

    very positive results. There was minimal inflammation at the site of implementation,

    successful widening of the carotid artery, as well as complete endothelialization and

    beginning stages of absorption by the time the study was complete7.

    According to ASTM standards for PLA polymers, there is a potential for increase

    in local acidity as the polymer degrades depending on implant form. However, since

    the application is stents, which have a very low volume, and the material is not being

    used for larger solid devices, the residual monomer may be acceptable8. To ensure

    the device will not drastically increase the localized pH, we will monitor the pH of the

    solution during testing.

    7Zamiri, P. Kuang, Y. Sharma, U. Ng, T.F. Busold, R.H. Rago, A.P. Core, L.A. Palasis, M. The biocompatibility of

    rapidly degrading polymeric stents in porcine carotid arteries.

    Biomaterials 8 August 2010

    8ASTM standard: F1925-09 Standard specification for Semi-crystalline Poly(lactide) Polymer and Copolymer

    Resins for surgical Implants

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    13/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 13

    5. DESIGN VERIFICATION PLAN

    5.1 Design of Experiment

    5.1.1 DOEOur experiment will involve placing different samples of PLA into phosphorus-

    buffered saline solution (PBS) designed to simulate the ph level of human blood. We

    will measure both the strength and reduced dimensions of the PLA samples for one

    week in order to identify whether a stent made completely out of PLA, that

    biodegrades between 6-12 months in the bloodstream, will have enough strength to

    function as a coronary stent. We will be using a pH level of 7.4, the same as human

    blood, and subject PLA samples to it for up to one week, but we will be extrapolating

    our data linearly up to a year.

    5.1.2 Control variables

    Our main control variables will be the ph of the PBS (7.4) and the temperature of

    the mixture (37 Degrees Celsius). Other factors like air pressure, humidity, and

    possible impurities may affect the experiment, but not enough to warrant direct

    controls given our limited resources.

    5.1.3 Response Variable

    The Variables we will be measuring are mass of the part in grams (calculating

    the change in dimensions), yield strength of the part in KPA (expected to be around

    54 MPA), and tensile strength of the part in KPA (Expected to be around 97.5 MPA)

    5.1.4 Number of Replications and Justification

    We intend to only repeat the experiment one time, as we feel that our resources

    and time is limited and that by the nature of the experiment, we wont need to average

    multiple data sets to acquire a meaningful value. We intend to use at least 7 samples

    in the one test however.

    5.2 Sample Description

    5.2.1 Materials

    We intend to use Polylactic acid, to be supplied from either an independent

    supplier (Ecoproducts) or from Splash in the form of the biodegradable plastic

    utensils. The Phosphorus-buffered Saline solution will be created by simply adding

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    14/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 14

    purchased table salt (NaCl) to De-Ionized water along with phosphates acquired from

    (the chemistry department), and will be measured by a Ph probe.

    5.2.2 Sample Preparation

    We intend to prepare the samples by mixing the salt and buffer with de-ionized

    water gradually, measuring the ph until it reaches the levels we need (approximately

    .1M of each). We also intend to melt the polylactic acid (which is a thermoplastic) in a

    furnace and re-cast it using a Styrofoam mold into at least 7 rectangular blocks, 4

    inches long by 3/4th inch wide by 1/8th inch deep (approximately 50 grams each) that

    fits easily into the Instron tensile tester and into the beakers we will be using.

    5.3 Test Procedures

    5.3.1 ASTM or ISO standards

    F1635 In vitro degradation of poly (L-Lactic acid) is the standard is being used as

    it allows us to measure exactly what we need (both the rate of decomposition and

    the resulting mechanical properties of the degraded material), and seems completely

    feasible with the resources available to us9. We intend to change only three details

    from the ASTM standard. We will only be using one sample per testing time (as

    using three would be too burdensome and costly for our scale), and no microbial

    agent and a lower solution to sample ratio (as we are only going to be testing over

    one week, so the odds that significant changes in the pH of the solution or large-

    scale microbial growth occurring and interfering with the data are minimal)

    5.3.2 Procedure

    For each test, we will mix the saline solutions and Phosphate buffer together

    ahead of time to the desired ph level. We will then place the samples into a water

    bath at 37 degrees Celsius (+/- 2 degrees Celsius), and once thermal equilibrium is

    attained, we will insert the samples into separate beakers, making sure none of the

    samples touch. Each day, we will remove one sample and measure the yield

    strength and tensile strength via the Instron tensile tester while the part is still wet.

    We will then dry it off and measure its mass by electric scale.

    9 F1635 Standard test method for in vitro degradation testing of hydrolytically degradable polymer

    resins and fabricated forms for surgical implants

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    15/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 15

    5.3.3 Assumptions

    We are assuming that the material we are using is pure PLA (L-type) when in fact

    it might be only mostly PLA. However, the cups are marketed as being completely

    biodegradable, so even if the material is not pure PLA, it should be able to

    biodegrade at roughly the same rate. We are assuming that the PLA will biodegrade

    solely based on the pH of the solution. We are assuming that simply leaving the

    sample wet will adequately simulate loading while still submerged in the blood, and

    that mechanical loading of the actual stent does not greatly affect the rate of

    degradation inside the bloodstream.

    5.4 Equipment

    5.4.1 Equipment

    The equipment we intend to use is a hot plate supplied from the MATEdepartment, a water bath (2L Beaker and water), a furnace that can go up to 300C,

    a pH meter, an electric scale that measures down to hundredths of a gram, and an

    Instron tensile tester.

    5.4.2 Expected data and level of accuracy

    The data we expect to get from these tests are both the mechanical strength of

    the PLA after exposure to the PBS, and the rate of degradation of the PLA as a

    function of time. The accuracy for the former will be relatively high, as the Instron

    tensile tester machine is able to very precisely calculate the force needed to both

    break the part and cause the initial yield (% error of about 1%). The accuracy of the

    latter however will be lower, as we will be calculating minute changes in the size of

    the part using both mass and measured dimensions (%error of about 3%)

    5.5 Analytical Techniques

    5.5.1 Significance Level

    For our linear regression test the p-value used will be

  • 8/3/2019 Example- FDA Proposal Final Junior Year

    16/16

    Document No.: Revised: 05/01/12 File Name: PLA based stents

    Page 16

    5.5.3 Description of Data

    Additionally, all strength measurements will be converted into percents of the

    original strength and compared to data from CES for common Nickel-Titanium

    Stents to see if the PLA based polymer will be adequate in maintaining the artery

    from collapsing. The loss in mass of the PLA sample will also be regression tested

    in Minitab to see if we can extrapolate, using the linear trend formula, whether or not

    the polymer will biodegrade within one year completely.

    5.6 Contingency Plan

    If we cannot get the saline solution to the proper pH is that well use baking soda

    (Sodium Bicarbonate) instead. If the samples do not all fit inside the beaker, we will

    use a larger beaker. If the samples do not cast properly, we will either re-cast them. If

    we think it was a simple mis-cast, and sand the part down to an acceptable alternative

    size if not. If our testing results show no significant degradation after one week (less

    than .01 g mass loss), then we will simply sand down the samples to different cross-

    sectional areas. If our experiment is unsuccessful for any other reason, we will

    attempt to find additional recorded values for the data we need and extrapolate a

    conclusion based on that data. If we cant get phosphates we will attempt to run the

    experiment without it.