Example Cert

3
CERTIFICATE OF CALIBRATION Sig 1 Sig 2 Sig 3 Time Electronics Unit 11 Sovereign Way Tonbridge, Kent. TN9 1RH email: [email protected] Tel: 01732 355993 Fax: 01732 770312 Approved Signatory Date of Issue : 20 May 2009 DM10005 Issued By : Time Electronics Certificate Number Page 1 of 3 Relative Humidity: 50% ±10% Temperature: 22.0°C ±4.0°C Enviromental Conditions Date Received: 20 May 2009 FLUKE-725-AUTO Procedure Comments Instrument /Device Supply Voltage: 230V ±2V @ 50Hz ±5Hz Failed on 3.2kOhm Measure EC4 419456 725 Fluke Process Calibrator Serial Number Model Number ID Number Manufacturer Type Customer /Owner Cal Systems Last Cert: 10034 Cal House Unit 34 Lock Drive Sevenoaks Kent Tel: 01732 355559 Fax: Email: [email protected] Traceability Information Cal Due Cal Date Cert No Serial No Instrument Description 5025 Multifunction Calibrator 1089G08 UKAS-90897 08 May 2009 07 May 2010 AC Current: 0pA-200uA=0.07%+300nA|200uA-2mA=0.05%+300nA|2mA-20mA=0.05%+3uA|20mA-200mA=0.05%+30uA|200mA-2A=0.1%+0.5mA|2A-20A=0.2%+5mA| AC Voltage: 1mV-20mV=0.05%+250uV|20mV-200mV=0.04%+150uV (45Hz-10kHz)|200mV-2V=0.03%+250uV (45Hz-10kHz)|2V-20V=0.03%+3mV (45Hz-10kHz)|20V-200V=0.06%+20mV|200V- 1.05kV=0.08%+90mV| DC Current: 0pA-200uA=150ppm+15nA|200uA-2mA=100ppm+40nA|2mA-20mA=80ppm+200nA|20mA-200mA=80ppm+3uA|200mA-2A=250ppm+40uA|2A-20A=600ppm+2mA| DC Voltage: 0pV-20mV=100ppm+4uV|20mV-200mV=30ppm+6uV|200mV-2V=15ppm+20uV|2V-20V=15ppm+150uV|20V-200V=30ppm+6mV|200V-1.05kV=50ppm+30mV| Resistance: 0pR-100Ω=0.01%+7m|100Ω-1kΩ=0.01%+5m|1kΩ-10kΩ=0.02%+20m|10kΩ-100kΩ=0.01%+1|100kΩ-1MΩ=0.01%+10|1MΩ-10MΩ=0.02%+100m|10MΩ-120MΩ=0.1%+1k| 5075 Digital Multimeter 1099C4 39803 19 May 2009 18 May 2010 AC Current: 0pA-200uA=0.07%+300nA|200uA-2mA=0.05%+300nA|2mA-20mA=0.05%+3uA|20mA-200mA=0.05%+30uA|200mA-2A=0.1%+0.5mA|2A-20A=0.2%+5mA| AC Voltage: 1mV-20mV=0.05%+250uV|20mV-200mV=0.04%+150uV (45Hz-10kHz)|200mV-2V=0.03%+250uV (45Hz-10kHz)|2V-20V=0.03%+3mV (45Hz-10kHz)|20V-200V=0.06%+20mV|200V- 1.05kV=0.08%+90mV| DC Current: 0pA-200uA=150ppm+15nA|200uA-2mA=100ppm+40nA|2mA-20mA=80ppm+200nA|20mA-200mA=80ppm+3uA|200mA-2A=250ppm+40uA|2A-20A=600ppm+2mA| DC Voltage: 0pV-20mV=100ppm+4uV|20mV-200mV=30ppm+6uV|200mV-2V=15ppm+20uV|2V-20V=15ppm+150uV|20V-200V=30ppm+6mV|200V-1.05kV=50ppm+30mV| Resistance: 0pR-100Ω=0.01%+7m|100Ω-1kΩ=0.01%+5m|1kΩ-10kΩ=0.02%+20m|10kΩ-100kΩ=0.01%+1|100kΩ-1MΩ=0.01%+10|1MΩ-10MΩ=0.02%+100m|10MΩ-120MΩ=0.1%+1k| Calibration Due: 19 May 2010 Date of Calibration: 20 May 2009 Calibrated by: Robert Martins This certificate has been produced by EasyCal Calibration Software from Time Electronics Ltd Cert.rpt v8.0

description

etete

Transcript of Example Cert

  • CERTIFICATE OF CALIBRATION

    Sig 1Sig 2Sig 3

    Time ElectronicsUnit 11Sovereign WayTonbridge, Kent. TN9 1RH

    email: [email protected]

    Tel: 01732 355993 Fax: 01732 770312

    Approved SignatoryDate of Issue : 20 May 2009

    DM10005

    Issued By : Time Electronics

    Certificate Number

    Page 1 of 3

    Relative Humidity: 50% 10%Temperature: 22.0C 4.0CEnviromental Conditions

    Date Received: 20 May 2009

    FLUKE-725-AUTOProcedure

    Comments

    Instrument/Device

    Supply Voltage: 230V 2V @ 50Hz 5Hz

    Failed on 3.2kOhm Measure

    EC4

    419456

    725

    Fluke

    Process Calibrator

    Serial Number

    Model Number

    ID Number

    Manufacturer

    Type

    Customer /Owner

    Cal Systems

    Last Cert: 10034

    Cal HouseUnit 34 Lock DriveSevenoaksKent

    Tel: 01732 355559Fax:Email: [email protected]

    Traceability InformationCal DueCal DateCert NoSerial NoInstrument Description

    5025 Multifunction Calibrator 1089G08 UKAS-90897 08 May 2009 07 May 2010 AC Current: 0pA-200uA=0.07%+300nA|200uA-2mA=0.05%+300nA|2mA-20mA=0.05%+3uA|20mA-200mA=0.05%+30uA|200mA-2A=0.1%+0.5mA|2A-20A=0.2%+5mA|AC Voltage: 1mV-20mV=0.05%+250uV|20mV-200mV=0.04%+150uV (45Hz-10kHz)|200mV-2V=0.03%+250uV (45Hz-10kHz)|2V-20V=0.03%+3mV (45Hz-10kHz)|20V-200V=0.06%+20mV|200V-1.05kV=0.08%+90mV|DC Current: 0pA-200uA=150ppm+15nA|200uA-2mA=100ppm+40nA|2mA-20mA=80ppm+200nA|20mA-200mA=80ppm+3uA|200mA-2A=250ppm+40uA|2A-20A=600ppm+2mA|DC Voltage: 0pV-20mV=100ppm+4uV|20mV-200mV=30ppm+6uV|200mV-2V=15ppm+20uV|2V-20V=15ppm+150uV|20V-200V=30ppm+6mV|200V-1.05kV=50ppm+30mV|Resistance: 0pR-100=0.01%+7m|100-1k=0.01%+5m|1k-10k=0.02%+20m|10k-100k=0.01%+1|100k-1M=0.01%+10|1M-10M=0.02%+100m|10M-120M=0.1%+1k|

    5075 Digital Multimeter 1099C4 39803 19 May 2009 18 May 2010 AC Current: 0pA-200uA=0.07%+300nA|200uA-2mA=0.05%+300nA|2mA-20mA=0.05%+3uA|20mA-200mA=0.05%+30uA|200mA-2A=0.1%+0.5mA|2A-20A=0.2%+5mA|AC Voltage: 1mV-20mV=0.05%+250uV|20mV-200mV=0.04%+150uV (45Hz-10kHz)|200mV-2V=0.03%+250uV (45Hz-10kHz)|2V-20V=0.03%+3mV (45Hz-10kHz)|20V-200V=0.06%+20mV|200V-1.05kV=0.08%+90mV|DC Current: 0pA-200uA=150ppm+15nA|200uA-2mA=100ppm+40nA|2mA-20mA=80ppm+200nA|20mA-200mA=80ppm+3uA|200mA-2A=250ppm+40uA|2A-20A=600ppm+2mA|DC Voltage: 0pV-20mV=100ppm+4uV|20mV-200mV=30ppm+6uV|200mV-2V=15ppm+20uV|2V-20V=15ppm+150uV|20V-200V=30ppm+6mV|200V-1.05kV=50ppm+30mV|Resistance: 0pR-100=0.01%+7m|100-1k=0.01%+5m|1k-10k=0.02%+20m|10k-100k=0.01%+1|100k-1M=0.01%+10|1M-10M=0.02%+100m|10M-120M=0.1%+1k|

    Calibration Due: 19 May 2010Date of Calibration: 20 May 2009Calibrated by: Robert Martins

    This certificate has been produced by EasyCal Calibration Software from Time Electronics LtdCert.rpt v8.0

  • CERTIFICATE OF CALIBRATIONDM10005Certificate Number

    Date of Issue : 20 May 2009

    Issued By : Time ElectronicsPage 2 of 3

    Pass/FailRqd Value Actual Value % of SpecTest Name Allowed Error

    UPPER DISPLAY

    Voltage Measure0V DC 0.000V0.000V 0.002V Pass0%15V DC 14.997V15.000V 0.005V Pass-60%20V DC 19.997V20.000V 0.006V Pass-50%30V DC 29.997V30.000V 0.008V Pass-38%

    mA Measure4mA 4.000mA4.000mA 0.003mA Pass0%12mA 11.999mA12.000mA 0.005mA Pass-20%24mA 23.994mA24.000mA 0.007mA Pass-86%

    LOWER DISPLAY

    mV/TC Measure0mV DC 0.01mV0.00mV 0.02mV Pass50%45mV DC 45.00mV45.00mV 0.03mV Pass0%90mV DC 89.99mV90.00mV 0.04mV Pass-25%

    Voltage Measure0V DC 0.000V0.000V 0.002V Pass0%10V DC 9.999V10.000V 0.004V Pass-25%20V DC 19.998V20.000V 0.006V Pass-33%

    Freqency Measure10kHz 10.00kHz10.00kHz 0.02kHz Pass0%

    mA Measure4mA 4.000mA4.000mA 0.003mA Pass0%12mA 12.000mA12.000mA 0.005mA Pass0%24mA 23.997mA24.000mA 0.007mA Pass-43%

    Thermocouple MeasureCJC Value 23.40C25.00C 5C Pass-32%0C -0.20C0.00C 0.7C Pass-29%

    Resistance 4 Wire Measure15 15.0015.00 0.1 Pass0%350 349.90350.00 0.1 Marginal-100%500 500.0500.0 0.5 Pass0%1500 1499.81500.0 0.5 Pass-40%3200 3198.73200.0 1 Fail-130%

    Resistance/RTD 3 Wire Measure350 350.02350.00 0.1 Pass20%

    Cert.rpt v8.1 This certificate has been produced by EasyCal Calibration Software from Time Electronics Ltd

  • CERTIFICATE OF CALIBRATIONDM10005Certificate Number

    Date of Issue : 20 May 2009

    Issued By : Time ElectronicsPage 3 of 3

    Pass/FailRqd Value Actual Value % of SpecTest Name Allowed Error

    Voltage Source0V 0.0000V0.0000V 0.002V Pass1%5V 5.0001V5.0000V 0.003V Pass5%10V 10.0005V10.0000V 0.004V Pass12%

    mV Source0mV 0.002mV0.000mV 0.02mV Pass10%45mV 44.996mV45.000mV 0.03mV Pass-14%100mV 99.997mV100.000mV 0.04mV Pass-8%

    Frequency Source10kHz 10.0000kHz10.0000kHz 0.025kHz Pass0%

    mA Source4mA 4.0004mA4.0000mA 0.0028mA Pass13%12mA 11.9996mA12.0000mA 0.0044mA Pass-9%24mA 23.9989mA24.0000mA 0.0068mA Pass-16%

    Ohms Source15 15.015.0 0.1 Pass15%360 360.0360.0 0.1 Pass23%500 500.2500.0 0.5 Pass30%

    Thermocouple SourceCold Junction Value -23.83C-25.00C 5C Pass-23%0C 0.12C0.00C 0.7C Pass17%180C 180.19C180.00C 0.7C Pass28%-180C -179.46C-180.00C 0.7C Pass-77%

    Comments

    Cert.rpt v8.1 This certificate has been produced by EasyCal Calibration Software from Time Electronics Ltd