Evaluation Of a Fully Automated GSP Neonatal G6PD Kit For ......CV% 7.0 3.6 2.9 3.3 4.1 3.6 6.4...
Transcript of Evaluation Of a Fully Automated GSP Neonatal G6PD Kit For ......CV% 7.0 3.6 2.9 3.3 4.1 3.6 6.4...
GSP® Neonatal G6PD kit (3310- 0010) is a new, fully automated G6PD (glucose-6-phos-phate dehy drogenase) enzyme activity assay to be used with the GSP® system. Optimized for the quantitative determination of G6PD activity in blood specimens dried on filter paper, the kit is an effective aid in screening newborns for G6PD deficiency (Figure 1.). To determine analytical
performance of the kit, verification studies were performed at PerkinElmer, Turku, Finland and an evaluation study was conducted at a customer site to ensure that the new product meets the user requirements and performs in its intended use.
GSP Neonatal G6PD kit 3310-0010
T E C H N I C A L N O T E
Authors
Hanna Polari, Teemu Korpimäki, Pauliina Mäkinen, Tuukka Pölönen, Terhi Kolari, Hanne Lindroos and Liisa Meriö
PerkinElmerMustionkatu, Turku, FINLAND
Evaluation Of a Fully Automated GSP Neonatal G6PD Kit For Newborn Screening
Key advantages over manual methods:
• Fully automated assay
• Reliable sample results even with floating disks
• 24 hrs valid calibration curve
• Elution control for missing or poorly eluted samples
• All reagents and QC material are bar-coded
• On-board stability 14 days
• Simultaneous screening of any other GSP analytes in any preferred order
Figure 1. Workflow comparison of manual G6PD assay and automated GSP Neonatal G6PD kit
Punch
Punch Result
Load Result
Manual
GSP
Reagent preparation
Reagent preparation
Retesting of floating disksMeasurementPipetting Incubation
with shaking Pipetting
32
Complete selection of reagents
One kit contains reagents for running 12 plates (Figure 2.)
• GSP G6PD Calibrators – 7 dried blood spot sets
• GSP G6PD Controls – 4 dried blood spot sets
• GSP G6PD Assay Buffer – 3 bottles, ready-for-use
• GSP G6PD Substrate 1 (NADP) – 3 vials, lyophilized
• GSP G6PD Substrate 2 (G6P) – 3 vials, ready-for-use
Figure 3. Run scenario for 24 plates
Fully Automated GSP Neonatal G6PD Kit Assay Protocol
After punching of the samples and reconstitution of the NADP substrate, the assay is fully automated from plate loading to completion. The assay time for one plate is 1 hour 22 minutes and a batch of 24 plates can be run in 10 hrs 12 min (Figure 4.).
Validated assay method
The GSP Neonatal G6PD assay is based on the same enzymatic reaction as the manual, PerkinElmer Neonatal G6PD (ND-1000) assay (Figure 3.). G6PD enzyme present in the sample catalyses the conversion of glucose-6-phosphate to 6-phosphogluconate while reducing NADP+ to NADPH. To determine G6PD activity levels, the fluorescence of NADPH is measured using an excitation wavelength of 340 nm and emission wavelength of 460 nm.
NADP+
G-6-P
6-PG
H+
NADPH+G6PD
Methods and instrumentation
Novel technology to measure floating disks
With the GSP Neonatal G6PD assay, a novel measurement technology enabled the measurement of wells with floating disks. The accuracy of GSP Neonatal G6PD kit to measure floating disks was verified by repeated measurements of both floating and non-floating dried blood samples of varying G6PD activity level.
Specimen stability
The influence of storage time, temperature, and humidity on G6PD activity was studied using several dried whole blood spot samples.
Precision, lower limits of detection and linearity
The precision of the assay was determined in accordance with CLSI document EP05-A2 [1] using human whole blood spot samples, 3 kit lots and 3 GSP instruments. The study was performed with 54 plates measured over 20 working days, each plate had 4 replicates per sample. The total number of measurements was 216 per sample. An analysis of variance approach was used to calculate the variance components. The analytical limits were determined in accordance with CLSI document EP17-A2 [2] and linearity was determined in accordance with CLSI document EP06-A [3].
Independent evaluation study
An evaluation study was conducted at the Newborn Screening Center, National Institutes of Health of the Philippines, Manila. The study aimed:
• To determine the usability of the GSP Neonatal G6PD kit
• To produce normal newborn population distribution data for the GSP Neonatal G6PD kit
• Dispensing Assay Buffer (100 µL/well)
• Dispensing Substrate 1 (5 µL/well)
• Shaking/Incubation (40 min)
• Dispensing Assay Buffer (100 µL/well)
• Dispensing Substrate 2 (5 µL/well)
• Shaking/Incubation (10 min)
• NADPH Measurement (G6PD activity)
• Measurement (Elution Control)
The study included 2,075 routine newborn screening specimens and 38 archived confirmed G6PD deficient specimens. G6PD activity levels were measured using GSP Neonatal G6PD kit (3310-0010) and a manual method (ND-1000). Patient mean, median, minimum, maximum and cut-off values corresponding to 10th, 5th and 3.25th percentile were calculated for the GSP Neonatal G6PD kit.
Figure 4. Run scenario for 24 plates with GSP Neonatal G6PD assay protocol
Figure 2. Content of the GSP Neonatal G6PD kit for 12 plates
Figure 3. The assay principle of the GSP Neonatal G6PD kit
54
Assay Performance
Measurement of floating disks
The ability of the GSP Neonatal G6PD kit to measure wells containing floating disks was verified by repeated measurements of both floating and non-floating samples with varying G6PD
PARAMETER VALUE
LoB 1.8 U/dL
LoD 2.4 U/dL
LoQ 2.4 U/dL
Measuring range from 2.4 to 130 U/dL
Linearity from 2.5 to 228 U/dL
Precision
The within-run, within-lot and total variation results for the GSP Neonatal G6PD kit are shown in Table 1.
GSP Neonatal G6PD kit calibration curve
The calibration unit was defined as units per decilitre (U/dL) of whole blood. One unit of G6PD enzyme catalyses the formation of 1 µmol of NADPH in 1 minute at 25 °C and pH 8.1. The kit calibrators cover the clinically relevant activity range and the
Figure 5. Effect of floating disks on GSP G6PD activity measurements
Figure 6. A typical calibration curve for GSP Neonatal G6PD assay
Table 1. Variation of GSP Neonatal G6PD kit using one calibration curve valid for 24 hours.
Table 2. Analytical limits of detection, measuring range and linearity
Results
reagents include controls for low (15 U/dL) and high (50 U/dL) G6PD activity. The calibration curve is valid for 24 hours. A typical calibration curve is shown in Figure 6.
activity levels. The measured sample activity was not significantly affected by floating disks (Figure 5.).
Analytical limits, measuring range and linearity
Analytical limits, measuring range and linearity of the GSP Neo-natal G6PD kit are summarized in Table 2. The Limit of Blank (LoB) is defined as the 95th percentile of a distribution of blank samples (n=150), the Limit of Detection (LoD) is based on 863 determinations of low level samples and the Limit of Quantita-tion (LoQ) is defined as the lowest activity with a total CV equal to or less than 20%.
Figure 7. The G6PD activity (% of reference without storage, 52 U/dL) during storage at different temperatures and humidity conditions.
Specimen stability
Storage of specimens at elevated temperature and humidity conditions significantly decreases the observed G6PD activity. At +35 °C with high humidity, 50 % of the G6PD activity may
be lost in just 1 day (Figure 7.).
SAMPLE
1 2 3 4 5 6 7
MEAN G6PD ACTIVITY (U/dL)
3.0 9.7 12.9 18.0 24.9 37.6 94.1
WITHIN RUN VARIATION
CV% 7.0 3.6 2.9 3.3 4.1 3.6 6.4
WITHIN LOT VARIATION
CV% 10.9 4.7 3.3 3.9 4.8 4.7 7.8
TOTAL VARIATION
CV% 12.7 5.3 4.2 4.8 4.9 4.8 8.1
N
216 216 216 216 216 216 216
76
Evaluation Study
Frequency distribution for newborn screening specimens
The frequency distributions of the newborn screening specimens for G6PD activity are shown in Figure 8. Confirmed G6PD deficient specimens are shown with red bars and light blue bars indicate presumed normal specimens, which include non-confirmed G6PD deficient specimens.
Screening performance
The new GSP Neonatal G6PD assay was compared to the manual Neonatal G6PD kit (ND-1000) by measuring G6PD activity of neonatal samples with both methods. The distribution is visualized in Figure 9. Red circles represent confirmed G6PD
specimens and light blue circles presumed normal specimens (which include non-confirmed G6PD deficient specimens). By using lower 5th percentile cut-off value, 20.5 U/dL, GSP Neonatal G6PD kit found all the confirmed G6PD deficient specimens and missed none.
Figure 8. Frequency distributions of the newborn screening specimens for G6PD activity. Red bars represent confirmed G6PD deficient specimens (including both retrospective and confirmed G6PD deficient routine specimens) and light blue bars presumed normal (include non-confirmed G6PD deficient specimens).
Descriptive statistics of the routine newborn screening specimens results for G6PD activity are shown in Table 3 and the results for the retrospective G6PD deficient specimens in Table 4.
SAMPLE TYPE
Routine*
SAMPLE SIZE 2075
MEAN (U/dL) 46.4
MEDIAN (U/dL)
47.7
5.00%
20.5
MIN (U/dL)
0.1
MAX (U/dL)
141.7
3.25%
16.1
10.00%
26.4
* Include also confirmed and non-confirmed G6PD deficient specimens.
SAMPLE TYPE
Retrospective confirmed G6PD deficient
SAMPLE SIZE 38
MEAN (U/dL) 7.7
MEDIAN (U/dL)
7.8
MIN (U/dL)
0.3
MAX (U/dL)
17.9
Table 3. Range, mean and median G6PD activity values with lower percentiles for the routine specimens measured with GSP Neonatal G6PD kit
Table 4. Range, mean and median G6PD activity values for the retrospective, confirmed G6PD deficient specimens measured with GSP Neonatal G6PD kit
Figure 9. Scatter plot of GSP Neonatal G6PD results as a function of manual G6PD activity. Red circles represent confirmed G6PD deficient samples.
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2014, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 1599-9790 TCH GSP Neonatal G6PD kit 3310-0010
Conclusions
• Reliable sample results obtainable even with floating disks
• The GSP Neonatal G6PD kit demonstrates good analytical performance
• Storage of specimens with elevated temperature and humidity increases the risk of false positive results
• All retrospective confirmed G6PD deficient samples were identified by the GSP Neonatal G6PD kit
• The fully automated GSP Neonatal G6PD assay saves time and money in a newborn screening laboratory.
Acknowledgements
Dr. Carmencita D. Padilla, Newborn Screening Center, Institute of Human Genetics, NIH University of the Philippines Manila
References
[1] CLSI (2004): Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition; CLSI Document EP05-A2. Wayne, PA: Clinical and Laboratory Standards Institute.
[2] CLSI (2012): Evaluation of Detection Capability for Clinical Laboratory Measurement procedures; Approved Guideline – Second Edition; CLSI Document EP17-A2. Wayne, PA: Clinical and Laboratory Standards Institute.
[3] CLSI (2003): Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI document EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute.
The product is not available in the USA, Canada, China, Taiwan, Brazil, Mexico and some other Asian and Latin American countries. Please contact your local sales representative for availability.