APPLICATION NOTE 21778 Keywords NIBRT, biopharmaceutical, bioproduction, QA/QC, biotherapeutic, IgG, monoclonal antibody (mAb), critical quality attribute, intact protein analysis, charge variant analysis, cetuximab, trastuzumab, bevacizumab, infliximab, innovator, ion exchange chromatography, MAbPac SCX-10, pH gradient buffer, CX-1 pH buffer, Vanquish Flex UHPLC Authors Amy Farrell, Craig Jakes, and Jonathan Bones Characterization and Comparability Laboratory, NIBRT – The National Institute for Bioprocessing Research and Training, Dublin, Ireland. Application benefits Simplified ion exchange analysis (IEX) of charge variants using a pH gradient Easy and fast method development for pH gradient based IEX compared to salt gradients Reduced risk of method variability of pH gradients compared to salt gradients for IEX Goal To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority of therapeutic monoclonal antibody species. To show the ease of optimization and improved reproducibility of the CX-1 pH gradient buffer system when compared to salt-based gradient systems used for charge variant analysis. Introduction Monoclonal antibodies (mAbs) are well-established pharmacological therapeutics with high specificity to target antigens, long serum half-life in humans, and capabilities for the treatment of a wide range of ailments, such as inflammatory diseases and cancer. Recombinant mAbs are highly heterogeneous biomolecules that may undergo a wide range of enzymatic and chemical modifications during bioprocessing and storage, which have Evaluation and application of salt- and pH-based ion-exchange chromatography gradients for analysis of therapeutic monoclonal antibodies

Transcript of Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared...

Page 1: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

APPLICATION NOTE 21778

KeywordsNIBRT, biopharmaceutical, bioproduction, QA/QC, biotherapeutic, IgG, monoclonal antibody (mAb), critical quality attribute, intact protein analysis, charge variant analysis, cetuximab, trastuzumab, bevacizumab, infliximab, innovator, ion exchange chromatography, MAbPac SCX-10, pH gradient buffer, CX-1 pH buffer, Vanquish Flex UHPLC

AuthorsAmy Farrell, Craig Jakes, and Jonathan Bones Characterization and Comparability Laboratory, NIBRT – The National Institute for Bioprocessing Research and Training, Dublin, Ireland.

Application benefits • Simplified ion exchange analysis (IEX) of charge variants using a pH gradient

• Easy and fast method development for pH gradient based IEX compared to salt gradients

• Reduced risk of method variability of pH gradients compared to salt gradients for IEX

Goal To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority of therapeutic monoclonal antibody species. To show the ease of optimization and improved reproducibility of the CX-1 pH gradient buffer system when compared to salt-based gradient systems used for charge variant analysis.

IntroductionMonoclonal antibodies (mAbs) are well-established pharmacological therapeutics with high specificity to target antigens, long serum half-life in humans, and capabilities for the treatment of a wide range of ailments, such as inflammatory diseases and cancer. Recombinant mAbs are highly heterogeneous biomolecules that may undergo a wide range of enzymatic and chemical modifications during bioprocessing and storage, which have

Evaluation and application of salt- and pH-based ion-exchange chromatography gradients for analysis of therapeutic monoclonal antibodies

Page 2: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

2

the potential to alter the safety and efficacy profile of mAbs by impacting protein binding, reducing drug activity, and giving rise to adverse toxicological and immunological responses in vivo.1 Hence, it is essential to detect, characterize, and quantify mAb variants and modifications (i.e. their critical quality attributes (CQAs)), to ensure the safety, quality, and efficacy of therapeutic biomolecules and thereby enable regulatory approval of mAbs for clinical use. Charge-related microheterogeneity of mAbs is frequently observed following bioprocessing due to both enzymatic processes and non-enzymatic degradation of therapeutic proteins.2

Determination of a mAb charge variant profile provides information regarding commonly encountered product CQAs.3 Bioprocess-related protein modifications may impart a positive or negative charge on mAbs, resulting in the generation of modified mAb variants with different isoelectric point (pI) values. Chemical modifications, such as deamidation, may give rise to a negative charge on mAbs, thereby decreasing their pI value and resulting in the formation of acidic mAb variants.4,5 Similarly, C-terminal cleavage of lysine residues results in the loss of positively charged amino acids on the mAb also creating acidic protein variants.6 Other modifications that may affect the charge of mAbs include cyclisation of N-terminal glutamine to form pyroglutamate and peptide bond cleavage.7

Ion-exchange chromatography (IEX), using either a salt or pH separation gradient, is the gold standard for determination of the charge variant profile of therapeutic mAbs.8 Traditionally, salt-based gradients, wherein the pH is maintained at a constant value while the ionic strength is increased over the duration of the gradient, were most frequently used for IEX. However, optimization of a salt gradient for charge variant analysis of individual mAbs often requires significant effort. Salt gradient buffer optimization may include the choice of salt reagent, salt concentration, and the pH value of the buffers. Use of pH gradients for IEX separation has been demonstrated as an excellent alternative to salt-based gradients for IEX of mAbs. pH-gradient-based ion-exchange chromatographic methods are reported to have greater resolving power and peak capacities when compared to ionic strength elution ion-exchange chromatography.9

A protein’s pI value corresponds to the pH value at which the protein has no net charge. In pH gradient SCX, the charge on the protein is reduced as the pH increases;

elution occurs when the protein has no electrostatic interaction with the SCX stationary phase. Unlike a salt gradient, optimization of a pH gradient for IEX does not require alteration of the mobile phase composition and hence facilitates more rapid method optimization. Furthermore, reproducible batch-to-batch preparation of salt-based buffers may be challenging when using certain buffer reagents. In this application note, the reproducibility of different salt buffers prepared across multiple preparation dates by different analysts is evaluated for both salt- and pH-based gradient buffers. In addition, two approaches to the generation of a pH gradient were investigated. One approach consisted of quaternary pump mixing of an acid, a base, a salt solution, and water to produce a pH separation gradient. The second approach was generation of a pH gradient using a Thermo Scientific™ cation-exchange pH gradient buffer platform consisting of a low pH buffer at pH 5.6 (Thermo Scientific™ CX-1 pH Gradient Buffer A) and a high pH buffer at pH 10.2 (Thermo Scientific™ CX-1 pH Gradient Buffer B). Using a gradient from 100% of CX-1 pH Gradient Buffer A to 100% of CX-1 pH Gradient Buffer B, a linear pH gradient from pH 5.6 to pH 10.2 may be generated. Since the majority of mAbs have pI values in the range of pH 6 to 10, pH-gradient-based separation methods based on the Thermo Scientific pH buffer platform and Thermo Scientific™ MAbPac™ SCX-10 columns may serve as a generic platform for mAb charge variant analysis. Furthermore, the described platform enables rapid, simple method optimization by merely using a shallower pH gradient over a narrower pH range following initial separations in the typical mAb pI range of pH 6 to 10.

Currently, six of the top 10 selling biopharmaceutical products are therapeutic mAbs, all of which will have come off patent by 2019.10 As a consequence, many biosimilars corresponding to these innovator mAbs are currently being marketed or are under development. Thus, the availability of a simple, readily optimized, widely applicable platform for charge variant analysis is of great benefit to producers of innovator therapeutic mAbs and biosimilars. The charge variant profile of the commercial mAbs bevacizumab, cetuximab, infliximab, and trastuzumab were also determined in this application note using a MAbPac SCX-10 column on a Thermo Scientific™ Vanquish™ Flex UHPLC with CX-1 pH gradient buffers. Minimal method optimization in <1.5 hours facilitated excellent separation of mAb variants. This approach could be easily applied to multiple mAbs and mAb biosimilar products.

Page 3: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

3

Experimental Recommended consumables • CX-1 pH Gradient Buffer A, pH 5.6 (P/N 085346)

• CX-1 pH Gradient Buffer B, pH 10.2 (P/N 085348)

• Deionized (DI) water, 18.2 MΩ∙cm resistivity

• MabPac SCX-10, 4 × 250 mm, 10 µm column(P/N 074625)

• Thermo Scientific™ Virtuoso™ Vial, clear 2 mL kit withsepta and cap (P/N 60180-VT402)

• Thermo Scientific™ Virtuoso™ Identification System(P/N 60180-VT100)

• Carboxypeptidase B (150 units/mg) (RocheDiagnostics™) (P/N 10103233001)

• Sodium phosphate monobasic (Acros Organics™)(P/N 10226410)

• Sodium phosphate dibasic (Acros Organics)(P/N 10036020)

• Sodium chloride (Fisher Chemicals™) (P/N 10428420)

• MES monohydrate (Acros Organics) (P/N 10517241)

Sample preparationAll untreated mAb samples were injected directly in formulation buffer.

To identify peaks corresponding to mAb variants containing C-terminal lysine, samples were digested with carboxypeptidease B as follows: 200 µg of mAb and 15 units of carboxypeptidase B were placed in a microcentrifuge tube, prepared to 60 µL with DI water, and mixed by pipette action. Samples were incubated for 2 hours at 37 oC.

Buffer preparationPhosphate buffersEluent A: 20 mM sodium phosphate,

pH 6.2Eluent B: 20 mM sodium phosphate,

pH 6.2, 300 mM NaClMES buffersEluent A: 20 mM MES, pH 5.6, 60 mM

NaClEluent B: 20 mM MES, pH 5.6, 300 mM

NaCl

pH gradient buffers: Four solution blending gradientEluent A: 100 mM MES monohydrateEluent B: 100 mM sodium phosphate

dibasicEluent C: 1 M sodium chlorideEluent D: Deionized water

pH gradient buffers: Two solution linear gradientEluent A: 10-fold dilution of CX-1 buffer A

(pH 5.6) with DI waterEluent B: 10-fold dilution of CX-1 buffer B

(pH 10.2) with DI water. Mobile phase B was stored in an amber borosilicate glass bottle.

Separation conditions Instrumentation Vanquish Flex Quaternary UHPLC system equipped with: • Quaternary Pump F (P/N VF-P20-A)• Column Compartment H (P/N VH-C10-A)• Split Sampler FT (P/N VF-A10-A) with a 25 µL Sample

Loop• Diode Array Detector HT (P/N VH-D10-A) with a

Thermo Scientific™ LightPipe™ 10 mm Standard FlowCell (P/N 6083.0100)

Instrument settings Flow rate: 1.0 mL/min Column temperature: 30 °C Injection details: 50 µg of mAb

(2 µL of 25 mg/mL bevacizumab, 10 µL of 5 mg/mL cetuximab, 5 µL of 10 mg/mL infliximab, 2.4 µL of 21 mg/mL trastuzumab)

DAD detector settings: 280 nm Gradient: Tables 1–3

Page 4: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

4

Time (minutes)

Optimized Gradient Conditions (% Eluent B)

Phosphate Buffers MES Buffers pH Gradient Buffers (Two Solution Gradient)

0.0 10 15 10

30.0 50 45 50

30.1 10 15 10

40.0 10 15 10

Table 1. Gradient conditions for mAb charge variant separations by IEX using either a salt gradient (phosphate- and MES-based buffers) or pH gradient (CX-1 buffer kit)

Table 2. Gradient conditions for mAb charge variant separations using a pH gradient of pH 6.5 to 7.2, generated using quaternary pump mixing of four eluents: 100 mM MES monohydrate (A), 100 mM sodium phosphate dibasic (B), 1 M sodium chloride (C), and water (D)

Time (minutes)

% Eluent A

% Eluent B

% Eluent C

% Eluent D

pH Salt Concentration

(mmol/L)

0.00 13.3 6.7 4 76 6.5 40

15.00 7.7 12.3 4 76 7.2 40

15.01 7.7 12.3 50 30 7.2 500

20.01 7.7 12.3 50 30 7.2 500

20.02 13.3 6.7 4 76 6.5 40

35.02 13.3 6.7 4 76 6.5 40

Table 3. pH gradient conditions used for mAb charge variant separations by IEX using a Thermo Scientific CX-1 pH gradient buffer kit

Time (minutes)

% Eluent B

Initial Gradient

Conditions

Optimized Gradient Conditions

Bevacizumab Cetuximab Infliximab Trastuzumab

-10.0 0 20 10 15 20

0.0 0 20 10 15 20

30.0 100 50 40 45 60

30.1 0 20 10 15 20

40.0 0 20 10 15 20

Data processing Thermo Scientific™ Chromeleon™ Chromatography Data System (CDS) 7.2.4 was used for data acquisition and analysis.

Page 5: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

5

Results and discussion Evaluation of buffer preparation reproducibility for both salt and pH-based IEX eluents Preparation of IEX buffers from salts followed by manual adjustment of pH has been reported to result in variations in chromatography from preparation to preparation of eluents. To evaluate the buffer preparation reproducibility for salt- and pH-based IEX eluents, phosphate salt buffers, MES salt buffers, and CX-1 pH gradient buffers were prepared by two different analysts on three different days. These buffers were then applied for charge variant analysis using a Vanquish Flex Quaternary UHPLC system. For each buffer system, triplicate injections of infliximab were performed. Excellent reproducibility was observed for replicate mAb injections analyzed using the same buffer preparation, as shown in the example in Figure 1, reflecting the excellent instrument precision achievable with a the Vanquish Flex UHPLC.

Figure 1. Overlay of chromatograms generated from triplicate injections of infliximab with a salt gradient using the same phosphate buffer preparation

Charge variant analysis performed using phosphate buffers resulted in slight retention time shifts for different buffers prepared by different analysts on different days despite using the same procedure, reagent, and equipment (Figure 2). However, the charge variant profile was unchanged following analysis with different phosphate buffer preparations. MES-based salt buffers were found to result in increased reproducibility when compared to phosphate buffers with very minor variations in retention time observed (Figure 3). Charge variant profiles generated using pH gradient buffers were found to be the most reproducible of the eluents evaluated when comparing different buffer preparations (Figure 4). Furthermore, the pH gradient yielded the best separation of peaks, resulting in the resolution of 12 mAb variants compared to 10 mAb species separated using salt gradients, with minimal buffer optimization (<1.5 hours).

Abso

rban

ce (m

AU)

Time (min)

-10

0

-10

20

30

40

50

60

70

80

90

100

111

31.62.00.1 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0-20

50.000

10.000

Page 6: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

6

Figure 2. Overlay of chromatograms generated from IEX analysis of infliximab with a salt gradient using different phosphate buffer eluents, prepared by two different analysts on three different days

Abso

rban

ce (m

AU)

Time (min)

-10

0

-10

20

30

40

50

60

70

80

90

100

111

31.62.00.1 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0-20

50.000

10.000

Figure 3. Overlay of chromatograms generated from IEX analysis of infliximab with a salt gradient using different MES buffer eluents, prepared by two different analysts on three different days

Abso

rban

ce (m

AU)

Time (min)

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

40.0

45.0

50.0

55.0

59.5

2.00.2 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0 32.0-4.8

45.000

15.000

Page 7: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

7

Figure 4. Overlay of chromatograms generated from IEX analysis of infliximab using three different pH buffer eluents, prepared by two different analysts on three different days. pH gradient buffers were prepared using a Thermo Scientific CX-1 buffer kit (pH range 6.1 to 7.9).

Abso

rban

ce (m

AU)

Time (min)

05

101520253035404550556065707580859095

100

2.00.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0 32.3-4

50.000

10.000

Comparison of CX-1 buffer kits with other pH gradient generation strategy. The Vanquish Flex UHPLC is equipped with a biocompatible, low-pressure mixing quaternary pump. This quaternary pump enables the generation of a pH gradient using a four-solution blending system comprising a weak acid and cognate base for pH buffer adjustment, and sodium chloride and water to alter the ionic strength of the buffer. Similar to the CX-1 pH gradient buffer kit, a pH gradient generated using a four-solution mixing gradient resulted in excellent separation

of infliximab charge variants (Figure 5). However, this pH gradient required considerable optimization to achieve an optimized separation gradient, increased buffer preparation times, and is only possible with a high-performing, inert quaternary pump capable of efficient, reproducibly mixing of eluents. The use of MES/phosphate buffer is restricted to pH 5.5 to 8 as it does not perform with decent buffering capacity outside these pH values. mAb samples with IP values outside this region would require method optimization with a different buffer system.

Abso

rban

ce (m

AU)

Time (min)

-5.0

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

40.0

45.0

50.0

55.0

60.0

65.068.8

2.00.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0 32.0 34.0 36.0 38.4-8.2

%D: 76.0%

30.0

76.0

%C: 4.0%

50.0

4.0

B: 4.800%

12.300

4.800

Figure 5. Overlay of chromatograms generated from IEX analysis of infliximab using a pH gradient from pH 6.5 to 7.2, generated from quaternary pump mixing of an acid, a base, a salt solution, and water

Page 8: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

8

Application of CX-1 pH buffer kits for IEX analysis of therapeutic monoclonal antibodies. Charge variant analysis of four therapeutic mAbs was performed with a pH gradient mode, produced using CX-1 pH gradient buffers on a MAbPac SCX-10 column (4.0 x 250 mm, 10 µm) and a Vanquish Flex UHPLC. Use of a linear gradient from 100% of CX-1 Gradient Buffer A to 100% of CX-1 Gradient Buffer B results in the generation of a pH gradient from pH 5.6 to pH 10.2. As the majority of therapeutic mAbs have pI values that fall within this pH range, CX-1 pH gradient buffers may be applied for the charge variant analyses of most therapeutic mAbs. Bevacizumab (Figure 6-A), cetuximab (Figure 7-A), infliximab (Figure 8-A), and trastuzumab (Figure 9-A) were analyzed using the described platform using the full pH gradient method from 100% of CX-1 buffer A to 100% of CX-1 buffer B, showing good

separation of mAb variants for each therapeutic protein. Following initial survey runs, the method was optimized for each therapeutic mAb by decreasing the pH range and hence gradient slope of the method (Figures 6-B to 9-B). Ease of optimization was facilitated by the use of a linear pH gradient method in combination with a MAbPac SCX-10 column. For each therapeutic mAb analyzed the pH gradients, as outlined in Table 3, were fully optimized in <1.5 hours. Simple optimization of the method enabled the separation of mAb variants, which were initially undetected for all mAbs determined. This straightforward, easily optimized, widely applicable approach to charge variant analysis of therapeutic mAbs has the potential to greatly expedite the characterisation of mAb charge variants during drug development and routine quality control analyses.

Abso

rban

ce (m

AU)

8.78 9.50

Time (min)

10.00 10.50 11.00 11.50 12.00 12.50 13.00 13.50 14.00 14.45

607

-4

50

100

150

200

250

300

350

400

500

550

450

CX-1 Gradient Buffer B: 29.264%48.177

1 2 3 4 5 679 10 11 12

12

Abso

rban

ce (m

AU)

6.3 7.0

Time (min)

8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.3

226

-1

20

40

60

80

100

120

140

160

180

200

1 2 3 4 5 6 7 9 10 11 12

8

CX-1 Gradient Buffer B: 26.3%

37.3

Figure 6. Chromatographic separation of bevacizumab charge variants using pH gradient (A) 0–100% over 30 min and (B) 20–50% B over 30 min, on a MAbPac SCX-10, 10 µm, 4 × 250 mm column

CX-1 Gradient Buffer B: 28.7%44.3

1 23

4Abso

rban

ce (m

AU)

8.62 9.00

Time (min)

9.50 10.00 10.50 11.00 11.50 12.00 12.50 13.30

123

-1

102030405060708090

100110

5

6

7

89

10

Abso

rban

ce (m

AU)

9.9 12.0

Time (min)

14.0

12

3

16.0 18.0 20.0 22.0 24.0 26.5

34.0

-0.8

4.0

8.0

12.0

16.0

20.0

24.0

28.0

CX-1 Gradient Buffer B: 19.9%

36.54

5

67

89

10

11

Figure 7. Chromatographic separation of cetuximab charge variants using pH gradient (A) 0–100% over 30 min and (B) 10–40% B over 30 min, on a MAbPac SCX-10, 10 µm, 4 × 250 mm column

A

B

A

B

Page 9: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

9

Abso

rban

ce (m

AU)

8.65 9.50

Time (min)

9.00 10.00 10.50 11.00 11.50 12.00 12.50 13.00 13.50 14.37

140

-5

102030405060708090

100110120130

CX-1 Gradient Buffer B: 28.8%47.9

810

1 2 3 45

6

7

9

13

1211

Abso

rban

ce (m

AU)

10.9

Time (min)

12.0

1

13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.6

68.9

-0.9

10.0

20.0

30.0

40.0

50.0

60.0

CX-1 Gradient Buffer B: 24.6%

38.8

2 34

5

6

8

7

13

9101112

Figure 8. Chromatographic separation of infliximab charge variants using pH gradient (A) 0–100% over 30 min and (B) 10–40% B over 30 min, on a MAbPac SCX-10, 10 µm, 4 × 250 mm column

3

4

5 6 78

9

Abso

rban

ce (m

AU)

12.39

Time (min)

13.00 13.50 14.00 14.50 15.00 15.50 16.00 16.50 17.00 17.46

277

-19

406080

100120140160180200220240260

200

CX-1 Gradient Buffer B: 41.3%58.2

1 210

11

3

4

5 6Ab

sorb

ance

(mAU

)

12.7

Time (min)

14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 22.8

116

-2

10

20

30

40

50

60

70

80

90

100

CX-1 Gradient Buffer B: 36.9%

50.4

78

9

1 2 1011

12 13

Figure 9. Chromatographic separation of trastuzumab charge variants using pH gradient (A) 0–100% over 30 min and (B) 10–40% B over 30 min, on a MAbPac SCX-10, 10 µm, 4 × 250 mm column

Both enzymatic and non-enzymatic modifications result in charge heterogeneity in therapeutic mAbs, e.g. glycosylation, deamidation, and lysine truncation. Two common mAb variants include moieties incorporating either one or two C-terminal lysine residues, which are reported to produce distinctive peaks eluting after the main mAb peak upon SCX analysis.11 To determine which chromatographic peaks detected correspond to mAb variants containing zero, one, or two C-terminal lysine residues, each of the mAb samples under determination were processed with carboxypeptidase B, an endopeptidase that cleaves C-terminal lysines with high specificity. Treatment of the therapeutic mAbs with carboxypeptidase B resulted in the removal of peaks in the basic region of the chromatograms acquired for charge variant analysis for the majority of the mAbs. A loss of peaks 7 and 8 was observed for bevacizumab

(Figure 10), while multiple peaks were eliminated from the chromatogram showing the charge variant profile of cetuximab, namely peaks 6 to 10 (Figure 11). Digestion of infliximab with carboxypeptidase B resulted in the loss of peaks 7 to 13 from the chromatogram and a notable increase in peak 6 which is the non-truncated form, as shown in Figure 12. Notably, digestion of trastuzumab drug product with carboxypeptidase B did not appear to result in the removal of peaks from the chromatograms suggesting the absence of C-terminal lysine variants in the drug substance (Figure 13).

A

B

A

B

Page 10: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

10

Abso

rban

ce (m

AU)

Time (min)

0.0

50.0

75.0

100.0

125.0

150.0

175.0

200.0

225.0

250.0

25.0

254.3

19.92.05.1 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0-19.0

CX-1 Gradient Buffer B: 24.8%

39.6

12 3 4 5

7 8

12 3 4

5

6

6

9

910

11

Figure 10. Characterization of bevacizumab drug product C-terminal lysine truncation variants using MAbPac SCX-10 column (4.0 × 250 mm, 10 µm) and pH gradient mode

Abso

rban

ce (m

AU)

Time (min)

0.0

5.0

10.0

15.0

20.0

25.0

30.0

35.0

40.0

45.0

49.4

37.2812.509.12 15.00 17.50 20.00 22.50 25.00 27.50 30.00 32.50 35.00-19.0

1

1

CX-1 Gradient Buffer B: 19.1%

40.0

10.0

2

2

3

3 4

6

8 9

4

5

5

7

10

Figure 11. Characterization of cetuximab drug product C-terminal lysine truncation variants using MAbPac SCX-10 column (4.0 × 250 mm, 10 µm) and pH gradient mode

Page 11: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

11

Abso

rban

ce (m

AU)

Time (min)

-20

-15

-10

-5

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

78

22.28.06.6 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.0-25

CX-1 Gradient Buffer B: 21.6%

37.2

12 3

4

51 2 34

5 7

8

6

9 10 1112

6

13

Figure 12. Characterization of infliximab drug product C-terminal lysine truncation variants using MAbPac SCX-10 column (4.0 × 250 mm, 10 µm) and pH gradient mode

Abso

rban

ce (m

AU)

Time (min)

-30

-20

-10

0

10

20

30

40

50

60

70

80

90

100

110

119

28.213.0

1 2 3

4

5

11.4 14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 23.0 24.0 25.0 26.0 27.0-38

CX-1 Gradient Buffer B: 35.2%

57.6

6 78 10

11 12

1 2 3

4

5 6 78 10

11 12

9

9

Figure 13. Characterization of trastuzumab drug product C-terminal lysine truncation variants using MAbPac SCX-10 column (4.0 × 250 mm, 10 µm) and pH gradient mode

Page 12: Evaluation and Application of Salt- and pH-Based Ion ......To demonstrate a simple, easily prepared pH buffer system that is capable of charge variant determination for the majority

Conclusions• Thermo Scientific CX-1 pH gradient buffers in

combination with Thermo Scientific MAbPac SCX-10analytical columns enable rapid method developmentfor charge variants analysis of most mAbbiotherapeutics.

• Use of a pH gradient for mAb charge variantanalysis by IEX enables increased reproducibility ofchromatography across different buffer preparations.

• This straightforward, easily optimized, widely applicableapproach to charge variant analysis of therapeuticmAbs generated using Thermo Scientific CX-1 pHgradient buffers has the potential to greatly expeditethe characterization of mAb charge variants duringdrug development and routine quality control analyses.

References1. Berkowitz, S.A., et al., Analytical tools for characterizing biopharmaceuticals and the

implications for biosimilars. Nat. Rev. Drug Discov., 2012, 11(7), 527-40.

2. Khawli, L.A., et al., Charge variants in IgG1: Isolation, characterization, in vitro binding properties and pharmacokinetics in rats. MAbs, 2010, 2(6), 613-24.

3. Rathore, A.S., Roadmap for implementation of quality by design (QbD) for biotechnology products. Trends Biotechnol., 2009, 27(9), 546-53.

4. Zheng, J.Y. and L.J. Janis, Influence of pH, buffer species, and storage temperature on physicochemical stability of a humanized monoclonal antibody LA298. Int. J. Pharm., 006, 308(1-2), 46-51.

5. Huang, L., et al., In vivo deamidation characterization of monoclonal antibody by LC/ MS/MS. Anal. Chem., 2005, 77(5), 1432-9.

6. Antes, B., et al., Analysis of lysine clipping of a humanized Lewis-Y specific IgG antibody and its relation to Fc-mediated effector function. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci., 2007, 852(1-2), 250-6.

7. Liu, H., et al., Heterogeneity of monoclonal antibodies. J. Pharm. Sci., 2008, 97(7), 2426-47.

8. Parr, M.K., O. Montacir, and H. Montacir, Physicochemical characterization of biopharmaceuticals. J. Pharm. Biomed. Anal., 2016.

9. Farnan, D. and G.T. Moreno, Multiproduct high-resolution monoclonal antibody charge variant separations by pH gradient ion-exchange chromatography. Anal. Chem, 2009, 81(21), 8846-57.

10. Walsh, G., Biopharmaceutical benchmarks 2014. Nat. Biotechnol., 2014, 32(10), 992-1000.

11. Dick, L.W., Jr., et al., C-terminal lysine variants in fully human monoclonal antibodies: investigation of test methods and possible causes. Biotechnol. Bioeng., 2008, 100(6), 1132-43.

Find out more at thermofisher.com/chargevariants

For Research Use Only. Not for use in diagnostic procedures. ©2018 Thermo Fisher Scientific Inc. All rights reserved. Roche Diagnostics is a trademark of Hoffmann-LaRoche Corporation. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN21778-EN 0318S