ESPECIFICACIONES CLIA

download ESPECIFICACIONES CLIA

of 7

Transcript of ESPECIFICACIONES CLIA

  • 8/22/2019 ESPECIFICACIONES CLIA

    1/7

    Has compliance with CLIA requirements really improved

    quality in US clinical laboratories?

    Sharon S. Ehrmeyera,*, Ronald H. Laessiga,b

    aDepartments of Pathology and Laboratory Medicine, University of Wisconsin Medical School, Room 6175,

    1300 University Avenue, Madison, WI 53706 USAbPopulation Health Sciences, University of Wisconsin Medical School, Madison, WI 53706, USA

    Received 18 December 2003; accepted 23 December 2003

    Abstract

    Background: The Clinical Laboratory Improvement Amendments of 1988 (CLIA88) mandate universal requirements for all

    U.S. clinical laboratory-testing sites. The intent of CLIA88 is to ensure quality testing through a combination of minimum

    quality practices that incorporate total quality management concepts. These regulations do not contain established, objective

    indicators or measures to assess quality. However, there is an implicit assumption that compliance with traditionally accepted

    good laboratory practicesfollowing manufacturers directions, routinely analysing quality control materials, applying quality

    assurance principles, employing and assessing competent testing personnel, and participating in external quality assessment or

    proficiency testing (PT)will result in improved test quality. Methods: The CLIA88 regulations do include PT performancestandards, which intentionally or unintentionally, define intra-laboratory performance. Passing PT has become a prime

    motivation for improving laboratory performance; it can also be used as an objective indicator to assess whether compliance to

    CLIA has improved intra-laboratory quality. Results: Data from 1994 through 2002 indicate that the percentage of laboratories

    passing PT has increased. In addition to PT performance, subjective indicators of improved qualityfrequency of inspection

    deficiencies, the number of government sanctions for non-compliance, and customer satisfactionwere evaluated.

    Conclusions: The results from these subjective indicators are more difficult to interpret but also seem to show improved

    quality in US clinical laboratories eleven years post-CLIA88.

    D 2004 Elsevier B.V. All rights reserved.

    Keywords: Clinical Laboratory Improvement Amendments (CLIA88); Laboratory compliance; Quality laboratory results; Testing deficiencies;

    Indicators of laboratory quality; Good laboratory practices

    1. Introduction

    To assess improvements resulting from the Clin-

    ical Laboratory Improvement Amendments of 1988

    and the most recent revisions in 2003, one must

    first understand their underlying philosophy [13].

    The CLIA88 regulations like the ISO 9000 docu-

    ments codify, for all testing sites, a collection of

    good laboratory practices associated with quality

    laboratories [4]. By incorporating a total quality

    management-based approach into the regulatory

    process, the laboratory director is made responsible

    for guaranteeing that appropriate quality practices

    0009-8981/$ - see front matterD 2004 Elsevier B.V. All rights reserved.doi:10.1016/j.cccn.2003.12.033

    * Corresponding author. Tel.: +1-608-262-0859; fax: +1-608-

    262-9520.

    E-mail address: [email protected] (S.S. Ehrmeyer).

    www.elsevier.com/locate/clinchim

    Clinica Chimica Acta 346 (2004) 3743

  • 8/22/2019 ESPECIFICACIONES CLIA

    2/7

    are followed. CLIA88 specifically states that the

    laboratory director must ensure that the . . . testing

    systems. . .used. . .provide quality laboratory services

    [covering] all aspects of test performance, whichincludes the pre-analytic, analytic, and post-analytic

    phases of testing [2]. To achieve this goal, CLIA88

    specifies a universal (applicable to all test sites) set

    of minimum, good laboratory practices: (1) having

    trained and competent testing personnel; (2) follow-

    ing manufacturers procedural directions (calibration,

    maintenance, etc.); (3) routinely performing and

    evaluating daily quality control, responding to out-

    of-control results and correcting problems; (4) ap-

    plying total quality management and continuous

    quality improvement principles and practices; (5)

    participating in external quality assessments; (6)

    and documenting all activities.

    The only quantitative performance requirement in

    CLIA88 is to successfully analyze three sets of PT

    challenges per year. The mandated, minimum good

    laboratory practices provide no absolute perfor-

    mance standards. Instead, they take a quality manage-

    ment approach requiring the director to implement

    and set limits for the quality control parameters

    appropriate for the intended use of the laboratory

    results. To assess compliance with these practices,

    CLIA88 requires that testing sites be inspected bi-annually. Test sites found out of compliance (either

    with PT or inspection requirements) are subject to a

    series of progressive sanctions ranging from requiring

    plans of correction, to monetary fines, to testing

    limitations, to revocation of the sites CLIA certifi-

    cate, i.e., prohibiting testing entirely.

    The regulation of US laboratories has a brief

    history. From 1967 to 1988, 12,000 large clinical

    and reference laboratories were required to follow

    the quality standards embodied in two federal regu-

    lations, the Clinical Laboratory Improvement Act of1967 (CLIA67) and Medicare 1968 [5,6]. The pur-

    pose of these regulations was to improve the quality

    of laboratory tests in federally funded programs and to

    ensure that federal healthcare dollars were spent on

    high quality laboratory tests. In the mid-1980s, a

    series of newspaper articles highlighted laboratory

    errors associated with Pap smear testing. The U.S.

    congress generalized these errors and enacted the

    CLIA88 amendments designed to ensure quality in

    all laboratories including physicians offices. For the

    first time, all clinical laboratory-testing sites were

    required to meet uniform standards.

    CLIA88 as implemented by the February 28, 1992

    and subsequent regulations recognize three majorcategories of testing: waived, moderate complexity

    and high complexity. Only laboratories registering as

    moderate and high complexity are required to partic-

    ipate in PT and to undergo bi-annual inspections for

    the mandated, good laboratory practices. Waived tests

    were originally limited to a total of nine analytes using

    methodologies described as so simple to perform

    that there is essentially no possibility of a wrong test

    result or, if erroneous, pose no risk to the patient.

    Most of these were already available, over-the-count-

    er, without prescription to the general public.

    2. Indicators of quality

    Before determining whether the quality of test

    results has improved as a result of compliance with

    CLIA88, one must first attempt to do what CLIA

    does notthat is to define quality and, to a lesser

    degree, what level of quality is needed to meet good

    laboratory practice standards. While adherence to

    government-mandated good laboratory practices is

    intuitively assumed to improve the quality of testresults, linking quality to compliance, as measured

    by inspection, is difficult because of the lack of a

    direct relationship. The only objective measurement

    of quality in CLIA88 regulations is performance in

    external quality assessment or PT programs. The

    Centers for Medicare and Medicaid Services (CMS)

    website claims that PT is a measure of laboratory

    quality and that PT performance is an indicator of a

    laboratorys ability to provide quality patient test

    results [7]. This official, government-sponsored, web

    site also declares that test accuracy, as assessed by PT:(1) reduces repetitive testing, (2) aids in rapid and

    appropriate patient diagnoses, (3) contributes to ef-

    fective treatment and (4) reduces overall costs of

    medical care. The CMS concludes that the review of

    a laboratorys PT performance data educates the

    laboratory staff. This educational purpose closely

    parallels and further validates the premise first dis-

    cussed by Belk and Sunderman, who beginning in

    1946, demonstrated that mere PT participation moti-

    vated laboratories to improve performance [8]. Most

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 374338

  • 8/22/2019 ESPECIFICACIONES CLIA

    3/7

    importantly, now as then, a laboratory, which fails PT

    by reporting wrong results, is better able to recog-

    nize and presumably correct a performance problem.

    A fundamental assumption, then and embodied inCLIA88, is that a performance problem identified

    by PT is also present in patient test results. Belk and

    Sundermans findings motivated the College of Amer-

    ican Pathologists to offer, for educational purposes,

    PT surveys beginning in 1947. Twenty years later, the

    US government mandated PT participation as part of

    the CLIA67 requirements governing 12,000 large

    hospital and reference laboratories [5,9]. CLIA88

    extended mandatory PT to all non-waived testing

    sites.

    The CLIA 88 mandated, analyte specific, PT

    tolerance limits for selected chemistry analytes are

    shown in Table 1 along with examples of acceptable

    performance ranges for stated target values. These

    tolerance limits are total error standards that incorpo-

    rate both inaccuracy and imprecision and are defined

    as either fixed limits (target valueF an absolute per-

    centage of the target value) or target valueF a multi-

    ple (usually 3.0) of the PT group standard deviation.

    For example if the target value for sodium is 140

    mmol/l, using the CLIA88 tolerance limit of 4.0

    mmol/l, a laboratorys PT result to be acceptable

    must fall within the range of 136144 mmol/l. EachPT event includes five samples for an analyte. Ac-

    ceptable performance in a single event is achieving at

    least four out of five correct results (>80% correct

    performance). More than one incorrect result ( < 80%

    correct) constitutes a failed analyte in that PT event.

    Because CLIA88 only identifies one objective

    measurement of quality, subjective indicatorsthe

    frequency of inspection deficiencies and government

    sanctionsalso were evaluated. CLIA88 requireslaboratories to be inspected every 2 years. The

    inspectors look for compliance with the mandated,

    good laboratory practices. Inspectors also determine

    the degree to which total quality management con-

    cepts, such as regular assessment of the total testing

    process by the laboratory director, are incorporated

    into the laboratory operation. Conceptually, the in-

    spection process is viewed as a quality improvement

    opportunity. Laboratories cited for deficiencies during

    inspection must correct them. Most laboratories read-

    ily comply, but a small percentage chooses not to

    rectify deficiencies. These laboratories are subject to

    sanctions on re-inspection. The most severe sanction

    is revocation of the CLIA-certificate.

    One of the most subjective of the CLIA88 require-

    ments is that the laboratory director must assure that

    the quality (accuracy, precision, long term stability,

    sensitivity, specificity, etc.) of test results meets the

    physicians needs. This makes the laboratory an active

    participant in patient care. A recent study suggests that

    US medical care, including the laboratory compo-

    nents, may be falling short of its quality goal.

    3. Results and discussion

    Tables 2a 2c summarize PT participant perfor-

    mance data for three events in 1994, one event in

    1997 through 1999, and three PT events in 2002. The

    Wisconsin State Laboratory of Hygiene PT Program is

    one of 14 CLIA-approved programs [7]. Participants

    are primarily from smaller hospital and physician

    office laboratories. Data for six chemistry analytes

    were selected because of the highest incidence of PTfailures. The PT data, spanning 8 years, show that

    laboratory performance, evidenced by a decreased

    percentage of failures, has improved and indicates

    that laboratories can and do correct problems. The

    mean failure rate for the six analytes over this 8-year

    period has decreased roughly by a factor of 3.

    Our studies provide some insight into intra-labora-

    tory performance requirements necessary to pass PT

    [1012]. Through computer modeling techniques and

    direct statistical calculations, we demonstrated that the

    Table 1

    CLIA88 proficiency testing, tolerance limits and acceptable

    performance (AP) ranges

    Test or analyte CLIA88 tolerance limit (TL) AP= VFTL

    Albumin TL = 10%, If TV = 40 g/l;

    AP=3644 g/l

    Calcium, total TL= 0.25 mmol/l, If TV= 2.20 mmol/l;

    AP= 1.95 2.45 mmol/l

    Cholesterol, total TL= 10%, If TV= 5.0 mmol/l;

    AP= 4.5 5.5 mmol/l

    Glucose TL = 10%, If TV = 6.7 mmol/l;

    AP= 6.03 7.37 mmol/l

    Sodium TL = 4 .0 mmol/l, If TV = 140.0 mmol/l;

    AP= 136 144 mmol/l

    Urea nitrogen TL=9%, If TV= 6.0 mmol Urea/l;

    AP= 5.46 5.54 mmol Urea/l

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 3743 39

  • 8/22/2019 ESPECIFICACIONES CLIA

    4/7

    PT criteria could be translated directly into total

    allowable error, or intra-laboratory performance

    specifications. These consist of maximum allowablecoefficient of variation (CV) and bias combinations.

    Statistically, a laboratory with no bias (inaccuracy)

    and a day-to-day CV (imprecision) equal to one-third

    of the specified PT criterion has virtually a 100%

    chance of passing a 5-sample PT event for that

    analyte. For example, the PT criterion for both glucose

    and cholesterol is target valueF 10%. A laboratory

    having a methodology with no bias and a CV that is

    100% of the specified PT limit (i.e., an internal CV of

    10% of the target value for either glucose or choles-

    terol) would fail PT (achieve < 80% correct results in

    one PT event) 51% of the time. This is tantamount tofailing every other PT event. If the laboratory reduces

    its internal CV for glucose or cholesterol to 5% of the

    target value, or to one-half of the PT limit, the chance

    of a single event failure is dramatically reduced to

    about 2%, or one in 50 PT events. Most importantly, if

    the CV is reduced to one-third of the PT limit, the

    chance of achieving < 80% correct results in one PT

    event is virtually zero. This assumes that the labora-

    tory is in statistical control, i.e., the method is func-

    tioning properly. It clearly delineates the apparent

    CLIA88 mandated view of what constitutes adequate

    performance.

    For regulatory purposes, CLIA88 defines passing

    PT as being successful in two out of three consecutive

    PT events. Our studies show that a laboratory having

    no bias will pass PT, long term, with a CV of

    approximately 40% of the specified CLIA88 limits.

    However, passing PT does not necessarily indicate

    that a laboratorys results are good enough to meet

    the needs of the physician. There are no data to

    support the assumption that a laboratory successfully

    passing PT is providing physicians with clinically

    relevant data. The PT criteria defining successfulperformance (Table 1) are open to question, since

    they are primarily based on 19851990 state-of-the-

    art interlaboratory performance data.

    A the critical level of 5 mmol/l for cholesterol, we

    can assume that among laboratories passing PT, some

    are doing so with day-to-day CVs at or near 40% of

    the CLIA88 tolerance limit or 0.2 mmol/l. This also

    means that a cholesterol result of 5.0 mmol/l would be

    reported, 5% of the time, as >5.4 or < 4.6 mmol/l. The

    question is will data with this amount of inherent

    variation permit adequate diagnosis and treatmentsuch as prescribing a cholesterol-lowering drug.

    As seen in Tables 2a 2c, the number of partic-

    ipants from 1994 to 2002 has significantly decreased.

    This reflects the fact that some test sites have

    changed PT providers while others have shifted from

    moderate and high complexity test methods to

    waived testing methodologies, which do not re-

    quire participation in PT. Likewise, some poorer

    performing laboratories, those failing PT, have dis-

    continued laboratory testing entirely. The latter case

    Table 2a

    Participant failures (%) in Wisconsin PT program1994

    PT event #1 PT event #2 PT event #3

    % N % N % NAlbumin 9 223 7 223 3 232

    Calcium 10 314 8 314 5 323

    Cholesterol 15 1094 12 1083 11 1057

    Glucose 12 1099 7 1080 8 1060

    Sodium 17 479 18 476 16 481

    Urea nitrogen 14 851 4 768 5 738

    Mean failure rate for these six analytes over 3 PT even-

    ts= 10.1F 4.6.

    Table 2b

    Participant failures (%) in Wisconsin PT program1997 1999

    PT event

    #1 (97)

    PT event

    #1 (98)

    PT event

    #1 (99)

    % N % N % N

    Albumin 3 232 2 219 0 224

    Calcium 5 260 4 245 1 244

    Cholesterol 6 926 5 690 2 548

    Glucose 6 827 4 679 3 602

    Sodium 7 374 8 356 4 352

    Urea nitrogen 9 508 7 469 1 428

    Mean failure rate for these six analytes over 3 PT events= 4.3F 2.6.

    Table 2c

    Participant failures (%) in Wisconsin PT program2002

    PT event #1 PT event #2 PT event #3

    % N % N % NAlbumin 2 301 1 302 5 304

    Calcium 1 333 1 222 2 337

    Cholesterol 4 436 4 428 3 417

    Glucose 8 502 5 500 4 497

    Sodium 3 375 4 372 5 374

    Urea nitrogen 3 415 3 413 4 413

    Mean failure rate for these six analytes over 3 PT events = 3.4F 1.8.

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 374340

  • 8/22/2019 ESPECIFICACIONES CLIA

    5/7

    could be regarded as a CLIA-induced improvement

    in the quality of testing.

    The percentages of test sites with CLIA88

    inspection deficiencies are shown in Table 3. Thereis a decline in the percentage of laboratories cited

    for deficiencies over the last four inspection cycles

    (approximately 8 years). Before declaring the

    CLIA88 approach to ensuring quality through

    enforced compliance to be evidence of success, it

    must be noted that the number of laboratories

    inspected for CLIA88 compliance has dropped over

    this period from 37,000 in 1995 to 22,000 in 2002.

    The reason for the decline of 15,000 laboratories is

    uncertain. Some test sites undoubtedly stopped

    testing entirely while others have chosen to be

    inspected by other government-approved profession-

    al programs such as the Joint Commission on

    Accreditation of Healthcare Organizations, the Col-

    lege of American Pathologists, or COLA (formerly

    the Commission of Office Laboratory Testing). Still

    others, particularly physician office laboratories,

    have abandoned CLIA-classified moderate and high

    complexity test methodologies and now perform

    only CLIA-waived testing that does not require

    CLIA88 inspections. Today, the waived list of tests

    includes 72 analytes and 1410 test systems, which

    provide testing in most disciplines of laboratorymedicine [7,13]. As a result, analytes such as

    hemoglobin Alc, HDL cholesterol and H. pylori

    are available to test sites that want to provide

    testing services, but do not wish to be CLIA

    inspected or participate in mandatory PT.

    Government sanctions imposed on laboratories not

    correcting identified CLIA88 deficiencies are sum-

    marized in Table 4. These data are difficult to interpret

    because of the very small percentage of laboratories

    sanctioned for repeated violations. The 1996 2002

    increase in the frequency of citations is possibly due

    to fewer laboratories surveyed (37,000 versus 22,000)

    and to inspectors becoming more willing to issue

    sanctions. In the inspectors view most test sites have

    11 years experience with the CLIA88 requirements

    and should be in compliance.

    The last indicator considered was determiningwhether the quality of test results, as a result of

    universal adoption of CLIA88 requirements, better

    meet customers (physicians and patients) needs.

    According to the 2000 U.S. Institute of Medicine

    (IOM) report, To Err is Human: Building a Safer

    Health System, between 44,000 and 98,000 hospi-

    talized Americans die each year from a variety of

    preventable medical errors [14]. Some portion of

    these serious medical errors must be attributed, in

    full or in part, to failures in the laboratory testing

    process. Beginning with Belk and Sunderman, thereis overwhelming data to support the idea that

    problems in laboratory testing are not due to a

    failure to appreciate the need for quality test results.

    Rather, the critical concern is how to better manage

    the testing process to ensure that the appropriate

    level of quality is achieved. Simply mandating, by

    regulation, specific quality practices such as those in

    CLIA88 does not appear to be the entire answer.

    The IOM study recommends several general solu-

    tions for building a better and safer health care

    Table 3

    Most frequently cited CLIA inspection deficiencies (%) [7]

    Deficiencies f 1995

    1996af 1997

    1998

    f 1999

    2000

    f 2001

    2002b

    Failure to

    run QC

    30% 27% 24% 18%

    Failure to

    follow

    manufacturers

    directions

    13% 12% 15% 6%

    No quality

    assurance

    program

    7% 7% 9% 3%

    Failure of

    laboratory

    director to

    meet

    responsibilities

    6% 8% 9% 2%

    a Approximately 37,000 CLIA inspected test sites.b Approximately 22,000 CLIA inspected test sites.

    Table 4

    Percent (%) of CLIA-inspected laboratories issued sanctions in

    1996 and 2002 for failure to correct deficiencies [7]

    Sanctions 1996a 2002b

    CLIA certificate

    suspensions, limitations

    or revocations

    0.19% 0.36%

    Test sites fined or

    given directed plans

    of corrections

    0.01% 0.13%

    a Approximately 37,000 CLIA inspected test sites.b Approximately 22,000 CLIA inspected test sites.

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 3743 41

  • 8/22/2019 ESPECIFICACIONES CLIA

    6/7

    system. It specifically recommends incorporating

    safeguards that mandate the use of quality processes

    to ensure quality and reduce the likelihood of

    human err. This can apply fully to the testingprocess governed by the CLIA88 regulations. Con-

    ceptually, CLIA88 envisioned a unique, total qual-

    ity management partnership among regulators,

    manufacturers, and laboratorians to achieve quality

    laboratory testing [15]. In the 11-years post CLIA,

    we have witnessed the evolution of test systems by

    which, to a lesser or greater extent, manufacturers

    have taken the testing process out of the hands of

    testing personnel and have produced instruments

    capable of delivering assured quality. Examples

    exist of systems with internal liquid calibrators and

    controls that use sophisticated analytical algorithms.

    These systems perform the calibration, but also run

    the controls, assess performance, rectify problems

    and, then and only then, release patient test results

    [16]. Certainly this new generation of instruments is

    in part due to the manufacturers desire to sell more

    testing systems by helping customers meet CLIA88

    quality requirements.

    4. Conclusions

    In the last analysis, the fundamental question

    remains, Do fewer PT failures, fewer inspection

    deficiencies, and a very limited number of govern-

    ment sanctions mean that the quality of patient test

    results has improved? Intuitively we all believe

    that adhering to the CLIA88-mandated, good labo-

    ratory practices and passing PT improves laboratory

    test quality. The indicators assessed in this study

    generally support this conclusion. However, the

    direct link between regulatory compliance and qual-

    ity test results is difficult to prove. Clearly, thedecision to make PT the cornerstone of CLIA88

    compliance has been a good one. The success of PT

    as a quality tool has been demonstrated by over 50

    years of experience.

    A laboratorys motivation may not reflect a true

    quality goal in the total quality management or

    ISO spirit though the outcome is desirable. Adop-

    tion of good laboratory practices and successful PT

    performance, if only to achieve and retain CLIA-

    certification, most certainly improves test quality.

    Other laboratories do benefit from the guidance pro-

    vided by the CLIA88 regulations. At a minimum,

    there is strong suggestive evidence that bad laborato-

    ries are selectively pushed out of the testing businessor, as an alternate hypothesis, have become better

    performers. The small numbers of government im-

    posed sanctions, one-tenth of one percent; speak to the

    laboratories overwhelming commitment to do good

    quality laboratory work.

    While our overall assessment is that compliance to

    CLIA88, has improved US clinical laboratory perfor-

    mance, we as scientists must distinguish what we

    know from what we think (assume) we know.

    References

    [1] Public Law 100 578, Section 353 Public Health Service Act

    (42 U.S.C. 263a) October 31, 1988. U.S. Department of

    Health and Human Services.

    [2] Medicare. Medicaid and CLIA programs: regulations imple-

    menting the Clinical Laboratory Improvement Amendments

    of 1988 (CLIA). Final rule. Fed Reg 1992;57:7002186.

    [3] U.S. Department of health and Human Services. Medicare,

    medicaid and CLIA programs; laboratory requirements relat-

    ing to quality systems and certain personnel qualifications;

    final rule. Fed Reg 2003;68:3640714.

    [4] URL: http://www.iso.ch/.

    [5] U.S. Department of Health. Education and Welfare, Public

    Health Service, Center for Disease Control, 1967: Clinical

    Laboratories Improvement Act of 1967. Part F of Title III of

    the Public Health Service Act. Sec. 353. 68. Atlanta: CDC;

    1967.

    [6] U.S. Department of Health, Education and Welfare, Social

    Security Administration: Federal Health Insurance for the

    Aged Regulations: Conditions for Coverage of Service of

    Independent Laboratories. Publ. No. HIR-13. 68. Washington:

    DHEW; 1968.

    [7] http://www.cms.gov/clia/.

    [8] Belk W, Sunderman W. A survey of the accuracy of chemical

    analyses in clinical laboratories. Am J Clin Pathol 1947;17:

    85361.

    [9] Ross J. Evolution of evaluation criteria in the College of Amer-ican Pathologists Surveys. Arch Pathol Lab Med 1988;112:

    3349.

    [10] Laessig R, Ehrmeyer S. Use of computer modeling to predict

    the magnitude of intralaboratory error tolerated by proposed

    CDC interlaboratory proficiency testing performance criteria.

    Clin Chem 1988;34:184953.

    [11] Ehrmeyer S, Laessig R, Leinweber J, Oryall J. 1990 Medicare/

    CLIA final rules for proficiency testing: minimum intralabor-

    atory performance characteristics (CV and bias) needed to

    pass. Clin Chem 1990;36:1736 40.

    [12] Laessig R, Ehrmeyer S, Leinweber J. Health care financing

    administrations new proficiency testing rules: characterizing

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 374342

    http://%20http//www.iso.ch/http://%20http//www.iso.ch/http://%20http//www.iso.ch/http://%20http//www.cms.gov/clia/http://%20http//www.cms.gov/clia/http://%20http//www.cms.gov/clia/http://%20http//www.cms.gov/clia/http://%20http//www.iso.ch/
  • 8/22/2019 ESPECIFICACIONES CLIA

    7/7

    long term performance as successful, probation and

    suspended. Arch Pathol Lab Med 1992;116:770 6.

    [13] Pontius C. Committee tackles agenda. Med Lab Obs 2003;

    35:26.

    [14] Committee on Quality of Health Care in America. Institute ofMedicine. In: Kohn L, Corrigan J, Donaldson M, editors. To

    Err is Human: Building a Safer Health system. Washington,

    DC: National Academy Press; 2000, p. 1287.

    [15] Laessig R, Ehrmeyer S. Quality: the next six months. Clin

    Chem 1997;43(5):9037.

    [16] Fallon K, Ehrmeyer S, Laessig R, Mansouri S, Ancy J. From

    quality control and quality assurance to assured quality. Point

    Care 2003;2(3):18894.

    S.S. Ehrmeyer, R.H. Laessig / Clinica Chimica Acta 346 (2004) 3743 43