ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of...

53
ENTC 4350 Modern Ventilators

Transcript of ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of...

Page 1: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

ENTC 4350

Modern Ventilators

Page 2: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Modern Ventillators

Ventilation assistance is provided under either of two conditions:

(1) breathing initiated by a timing mechanism or

(2) patient-initiated breathing.

Page 3: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Automatically timed breathing is usually provided for patients who cannot breathe on their own. • It provides inspiration and expiration at fixed

rates and durations except for periodic sigh; a sigh is a rest period for the patient.

Page 4: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Patient-initiated breathing may be given to one who has difficulty breathing due to high airway resistance. • The patient’s effort to inhale triggers the

respirator unit to deliver air at the positive pressure prescribed.

Page 5: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Ventilator Modes of Operation

The following definitions are commonly used to describe respirator/ventilator operation:• CMV Continuous mandatory ventilation:

Once initiated by either the ventilator operator or the patient, the breath is driven to the patient.

• CPAP Continuous positive airway pressure: Breaths are spontaneous, unless the operator

intervenes. The spontaneous breaths are determined entirely by patient effort. However, the air/oxygen mixture is set by the ventilator.

Page 6: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

SIMV Synchronized intermittent mandatory ventilation: These breaths are initiated by either the machine, the operator, or the patient. The breaths may be either

spontaneous or mandatory. That is, if the patient does not breathe within a preset time period, the ventilator will deliver a breath.

PEEP Positive end-expiratory pressure: The pressure maintained by the ventilator that

the patient must exhale against.

Page 7: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

• Apena The patient has stopped breathing.

• Sigh A breath delivered by the ventilator that differs in

duration and pressure from a nominal breath.

• Nebulizer A device for producing a fine spray of liquid or medication into the patient’s air.

Page 8: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The block diagram shows the external flexible tubing and the ventilator unit.

Page 9: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Air from the ventilator during patient inspiration passes through a bacterial filter and humidifier. • A nebulizer may spray medication into the air.

This air then forces valve 1 up to close off the spirometer and deliver air to the patient.

Page 10: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

After the inspiration air is turned off by the ventilator, valve 1 drops and the patient exhales into the bellows, which has its outlet valve held closed pneumatically by the ventilator unit. • During the subsequent patient inspiration

cycle, that valve will open, causing the bellows to fall and empty.

Page 11: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

During patient expiration, the direction of the air in the pneumatic system is determined by the main solenoid, which is switched appropriately by the system electronics.• Room air is drawn from the air inlet filter by

the main compressor and is directed through the main solenoid to hold closed the upper outlet valve of the bellows located inside the unit.

Page 12: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Next, the weight of the bellows causes the bottom bellows chamber outlet valve to open, as the main solenoid directs air to close the inlet bellows chamber valve. • The weight of the falling bellows draws

oxygen-enriched air into it in preparation for the patient-inspiration part of the cycle.

Page 13: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The oxygen content of the air flowing into the bellows is controlled by a percentage control valve, which regulates the resistance to room air and oxygen appropriately.

Page 14: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

At the end of patient expiration, the system electronics trip the main solenoid, thereby initiating the patient-inspiration part of he cycle.

Page 15: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

During patient inspiration, the compressor draws room air through an air filter and then through the main solenoid. • It forces the bottom inlet valve of the internal

bellows chamber open and forces the bottom bellows chamber outlet valve closed.

Page 16: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The high pressure in the bellows chamber compresses the bellows, forcing open the upper outlet valve set free by the main solenoid. • This allows the oxygen-enriched air to pass

through the main bacteria filter into the external tubes and then to the patient lungs.

Page 17: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

A sensitivity control monitors the negative pressure necessary to initiate inspiration when the respirator is used in the patient-initiated breathing mode called the assist mode. • A nebulizer compressor may draw air from the

bellows and force it through an aspirator to mix medication into the patient-inspired air.

Page 18: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

• When inspiration is complete, the main solenoid switches the direction of the pneumatic air to repeat the expiration cycle, and so on.

Page 19: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The black bag at the end of the tubing simulates a compliant lung. • This respirator may be

operated using compressed air from the hospital air supply.

Page 20: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

In that mode, the ventilator can be removed from its internal compressor, thereby decreasing it in size; turning off the compressor also reduces problems in instrument noise control.

Page 21: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

To aid patient respiration, hospital air and oxygen enter the pneumatic compartment, where it is filtered. • A check valve reduces the pressure to a

nominal 10 psi.

Page 22: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

A proportional solenoid valve assembly allows the air/oxygen mix to be controlled by the system electronics.

• A check valve directs air to the patient during the inspiration cycle.

Page 23: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

During the subsequent expiration phase, the system electronics opens the check valve, CV5, to provide a vent for the patient exhalation air. • In this case, the pneumatically operated

valves of older ventilators have been replaced by valves controlled with microprocessor-based electronics.

Page 24: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

A small positive-pressure ventilator is illustrated with both front and back views.

Page 25: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

PNEUMOTACHOGRAPH AIRFLOW MEASUREMENT

Patient airflow may be measured by changes in resistance of a thermistor in the airstream due to the cooling effect of flowing air. • But it must be calibrated to compensate for

changing ambient temperature.

Page 26: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

To eliminate this disadvantage, a strain-gauge wire mesh is often used.

Page 27: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The airflow in either direction puts a strain on the screen and changes the resistance of its strain gauge.

Page 28: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The strain gauge is a component of a Wheatstone bridge.

Page 29: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Here the change in resistance, R, is proportional to the airflow, F, past the wire mesh.

RkF

Page 30: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

BBAAAB VR

R

RRR

RVVV

2

RRR

RVV BBA

R

RVV BBB 2

Voltage divider rule

Page 31: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

BBAB VR

R

RR

RR

RR

R

R

RV

22

2

22

2

Getting a common denominator

BBAB VRRR

RRR

RRR

RV

24

2

24

22

2

2

2

BBAB VRR

RV

24

Page 32: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

BBAB VR

RV

4

If R » R

For the diff amplifier with a gain of AD ,

k

FRfor

kR

FVA

R

RVAV BBDBBD

F

44

Page 33: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Integrator Circuit

For the integrator circuit,

dtR

v

Cdti

CV

t

t

in

t

t

Cout 00

11

dtR

RVA

RCV

t

t

BBDout

0

4

1

Page 34: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Note that the term inside the integral is a constant,

kttCR

RVAdt

CR

RVAV BBD

t

t

BBDout

2

0

2 44

Page 35: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Pneumotachograph Volume Measurement

A volume exhaled by a patient is measured with the pneumotachograph by first closing and then opening the reset switch.• This sets the initial charge on the capacitor to

zero and fixes Vout at zero.

Page 36: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The patient is then asked to exhale through the pneumotach mouthpiece.• The resulting change in R

creates a voltage VF as a function of time in proportion to flow.

Page 37: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The air volume expired by the patient, beginning at t = 0, when the reset is activated, equals the area under the flow vs. time curve.• Mathematically, this area is computed by

integration.

Page 38: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The output voltage, Vout ,is proportional to the volume of air expired from time t = 0 to the time, t, desired.• The flow, F, is a function of time that may

increase, decrease, or stay constant, so long as it goes in one direction.

dtkR

FVA

RCV

t

t

BBDout

0

4

1

Page 39: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

dtFCkRR

VAV

t

t

BBDout

014

VOLvolumelungTotaldtF

t

t

0

VOLkVOLCkRR

VAV BBD

out 14

This implies that Vout is proportional to the total volume of air that is passed through the pneumotach for the time of observation.

Page 40: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

THE PLETHYSMOGRAPH

The pneumotachograph can be used to measure the rate of airflow during respiration and the vital air capacity of the lung VC. • It cannot, however, measure the total lung

capacity, TLC.

Page 41: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The reason for this is that the pneumotachograph can only measure the amount of air a person can exchange in respiration, and cannot detect the residual volume of air, RV, left in the lung after a forced exhaling. • To measure the TLC, a body plethysmograph

may be used.

Page 42: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The plethysmograph consists of an airtight chamber the patient can enter and sit in.

Page 43: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

The principle of operation of the plethysmograph depends directly on the gas law for an ideal gas of volume V and pressure P, namely Boyle’s Law

• where k1 is a constant and T is the absolute temperature (K). In the chamber, the temperature remains constant.

TkPV 1

Page 44: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

To measure TLC, the patient enters the chamber.• The door is sealed, and the valve on the

mouthpiece is closed.

Page 45: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Since the patient cannot breathe with the valve closed, the air pressure in the mouthpiece equals that in the lung, PT.

• That is, when the flow of air is zero, the pressure drop from mouthpiece to lung is also zero.

Page 46: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

With the valve closed, a formula for the thoracic volume is derived as follows.• The gas equation for constant temperature

holds inside the lung

V

P

dV

dP

Page 47: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Using the previous equation, the TLC can be written as:

TT P

TLC

dP

dTLC

Page 48: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Boyle’s Law also holds in the chamber, so that

• Here VOLC is the chamber volume and PC is the chamber pressure.

C

C

C

C

P

Vol

dP

dVol

Page 49: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Because the chamber is closed, any increase in the thoracic volume introduced by breathing motions causes a decrease in the chamber volume of air. • That is

dTLCdVol C

Page 50: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

Combining the equations

• with the previous fact:

TT P

TLC

dP

dTLC

C

C

C

C

P

Vol

dP

dVol

dTLCdVol C

Page 51: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

This yields:

TT

CC

CC dP

P

TLCdTLCdP

P

VoldVol

CC

CT

T

dPP

VoldP

P

TLC

Page 52: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

During the test PT = PC approximately, since the changes in pressure induced by breathing motions are small when the patient is resting.• Thus,

T

CC dP

dPVolTLC

Page 53: ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.

This equation gives the means for measuring the lung volume, TLC, by the following steps:1.Close the mouthpiece valve on the patient

sealed in the chamber.

2.Ask the patient to make breathing motions.

3.Read the change in pressure dPT on meter 1.

4.Read the change in pressure dPC in the chamber on meter 2.

5.Since the chamber volume VOLC is a known specification of the plethysmograph, use the result in steps 3 and 4 to compute TLC.