(English) - bsuh.nhs.uk · logic which indicate the year of manufacture and a . sequential unit...
Transcript of (English) - bsuh.nhs.uk · logic which indicate the year of manufacture and a . sequential unit...
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
Operation and Maintenance Manual(English)
User ResponsibilityThis Product will perform in conformity with the description contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked prior to use following the Pre-Use Checkout procedure described in section 3. A defective Product should not be used. Parts that are broken, missing, visibly worn, distorted or contaminated should be replaced immediately.
Should such repair or replacement become necessary, the manufacturer recommends that a telephone request for service advice be made to the local distributor. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by the manufacturer or local distributor. The Product must not be altered. The user of this Product shall have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Ikaria.
Outside the U.S.A. and Canada, check local laws for any restrictions that may apply. Caution: U.S. Federal and Canadian law restrict this device to sale by or on the order of a licensed medical practitioner.
Inhaled Nitric Oxide mixtures must be handled and stored in compliance with federal, state and local regulations.
Ikaria products have unit serial numbers with coded logic which indicate the year of manufacture and a sequential unit number for identification.
©2012 Ikaria, Inc.INOMAX®, INOmax DSIR
®, and INOblender® are registered trademarks of INO Therapeutics LLC and their respective owners. INO Therapeutics LLC is a wholly-owned subsidiary of Ikaria, Inc.
BL 20051234 The first two digits indicate the product, the next four numeric digits indicate the year of manufacture, and the next four digits are the sequential unit number produced that year.
10026 INOblender (DISS, 800 ppm, English)10004 INOblender (DISS, 800 ppm, English)10021 INOblender (NIST, 400 ppm, English)
Part No. 20181 Rev-022012-06
Part No. 20181 Rev-022012-06
1
1. General Information .............................................................................................................3 Introduction .......................................................................................................................4. DefinitionsandAbbreviations ...........................................................................................4. Theory of Operation ..........................................................................................................5 EnvironmentalEffects .......................................................................................................6
2. Setup ...................................................................................................................................7
3. Pre-UseCheckout ...............................................................................................................9
4.. Operation ...........................................................................................................................12
5. Maintenance ......................................................................................................................15 ReplacingRegulatorTips................................................................................................16 CylinderLeakCheck.......................................................................................................18 PartsandAccessories ....................................................................................................19
6. Specifications.....................................................................................................................21
Contents
Part No. 20181 Rev-022012-06
2
(Intentionallyleftblank)
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
1/ General Information
General
Information
Part No. 20181 Rev-022012-06
1/ General Information
Gen
eral
Info
rmat
ion
Part No. 20181 Rev-022012-06
3
CautionsandWarningsWARNINGS:•ThepurgeproceduremustbefollowedtohelpensureNO2ispurgedfromthepressureregulator,INOblenderandhosesbeforethemanualresuscitatorbagornasalcannulaisconnectedtothepatient.ThemanualbagshouldbesqueezedcontinuouslyduringusetoavoidNO2buildingupinthebag.IfthebagisnotsqueezedcontinuouslywhiledeliveringINOMAXthebagshouldberemovedfromthepatientandthepurgeprocedureperformedbeforecontinuing.
• TheINOblendershouldonlybeusedinaccordancewiththeindications,usage,contraindications,warningsandprecautionsdescribedintheINOMAXdrugpackageinsertandlabeling.Refertothismaterialpriortouse.
• PersonsusingthisdeviceshouldbetrainedonandexperiencedintheuseofthisdevicetoassureeffectiveadministrationofINOMAXandtoavoidinjurytothepatientorothersresultingfrominhalationofexcessINOMAX,nitrogendioxideorotherreactionproducts.
• TheNationalInstituteforOccupationalSafetyandHealth(NIOSH)intheU.S.hasarecommendedtime-weighted(8hours)averageconcentrationof25ppmforNOandalimitof1ppmforNO2.PersonsusingtheINOblenderwhomaybeparticularlysensitivetoNOorNO2,orwhomaybeexposedtotheseagentsforprolongedperiodsasaresultoftheuseofthisdevice,shouldbeawarethattheINOblenderdoesnotscavengetheexhaustdrug,andthatthisisventedattheresuscitatorbag.AmbientconcentrationsofNOorNO2areexpectedtobelessthan0.2ppmwhenusingtheINOblender,seeSection1EnvironmentalEffects.
Cautions:•OutsidetheU.S.andCanada,checklocallawsforanyrestrictionsthatmayapply. U.S.federallawandCanadianlawrestrictthisdevicetosalebyorontheorderofalicensedmedical practitioner.
•Properuseofthisproductdependsoncarefulreadingandunderstandingoflabelingandinstructions.
•Tohelpensureproperoperationofthisproduct,completethepre-usecheckoutpriortoeachuse.
•Whennotinuse,theO2flowmeterandtheINOMAXcylindervalveshouldbeturnedoff.
•Whencleaning,donotsubmergeinliquid.
•Donotautoclave.
•Refertothemanufacturer’sproceduresforusingtheresuscitationbag.
WARNING:Warningstellyouaboutdangerousconditionsthatcancauseinjurytotheoperatororthepatientifyoudonotobeyalloftheinstructionsinthismanual.
Caution:Cautionstellyouabouthowtoproperlyusetheequipmentandconditionsthatcouldcausedamagetotheequipment.
Readandobeyallwarningsandcautions.
Note:Notesprovideclarificationorsupplementalinformation.
Bluearrowdenotesrequireduseraction.
Part No. 20181 Rev-022012-06
4.
Pneumatic Inlet
Pneumatic Outlet
RxONLY Prescriptionuseonly
Pressure
Productpackagingisabletobe recycled
Protectfromheatandradioactivesources
SN SerialNumber
StorageTemperatureLimitation
StorageHumidityLimitation
ThisproductisincompliancewiththeEuropeanCouncilDirective93/42/EECanditsamendmentsconcerningmedicaldevices
Attention,consultaccompanyingdocuments!
AuthorizedRepresentativeintheEuropeanCommunity
CatalogNumber
Date of Manufacture
Donotuseifpackageisdamaged
Fragile,handlewithcare
Keep Dry
Manufacturer
NO or O2GasInlet
NOGasOutlet
IntroductionTheINOblenderallowsuserstoselectaconcentrationofINOMAX,(nitricoxideinabalanceofnitrogen),tobemixedintoausersetflowofoxygenwhichisdeliveredtoapatient.TheintendedusefortheINOblenderisasabackuptoaprimaryINOMAXdeliverysystem;orforshorttermattendedusewhenaprimarydeliverydevicecannotpracticablybeused.Thisintendeduseincludesapplicationswithinamedicalfacilityandtransportoutsideofamedicalfacility.TheINOblenderisnotintendedforuseasaprimaryINOMAXdeliverysystemforlongtermuse.
DefinitionsandabbreviationsN2 NitrogenNO INOMAX(nitricoxide)forinhalationNO2 NitrogendioxideNO/N2 Nitricoxide(NO)andnitrogen(N2)gasmixtureO2 Oxygenppm PartsPerMillionSet NO ThedoseofINOMAXsetbytheuserv/v Volume to volume
Note:ThismanualshowstheSetNOdisplaysassociatedwiththe0-80ppmrange.
SymbolsusedinthismanualoronthesystemSymbolsreplacewordsontheequipmentand/orinthismanual.Thesesymbolsinclude:
Part No. 20181 Rev-022012-06
5
INOblenderSchematicDiagram
Theory of Operation1. TheINOblenderconnectstoa3-5bar(45-75psig)oxygensupply.Theoxygenflowisthen
controlledbyanintegratedoxygenflowmeter(2-14lpm).Thereisafixedrestrictorintheoxygenflowpathwhichcreatesapressureproportionaltotheoxygenflowrate.
2. TheINOblenderisalsoconnectedtoa2bar(30psig)INOMAX(800ppmor400ppm)supply.
•TheINOMAXflowiscontrolledbyabalanceregulatorinproportiontotheoxygenpressurecausedbythefixedrestrictorandtheINOblendersetting.
•ThebalanceregulatormaintainsaconstantINOMAXdoseforvaryingoxygenflowrates.
3. TheflowsofO2 and INOMAXaremixedpriortoexitingtheINOblenderoutlet.
INOMAXCylinder
INOMAXRegulator
OxygenSource
BalanceRegulator
DoseKnob
INOMAXandO2Outlet
FixedLaminarRestrictor
INOMAXOxygenOxygen/INOMAX
Part No. 20181 Rev-022012-06
6
EnvironmentalEffectsTheNationalInstituteforOccupationalSafetyandHealth(NIOSH)hasrecommendedexposurelimitsasfollows(ref1).
NO time-weighted(8hours)average concentration limit of 25 ppm
NO2 ceiling limit of 1 ppm.
Theenvironmentalbuildupofnitricoxide in a well ventilatedICUroomcanbeevaluatedusingthefollowingcalculation:
Roomsize 10ftsquare(approx.3meters)
Roomvolume 27,000L of air
Roomventilation 2,700L/min at 6 room changes/hr
NOflowintoroom 14L/minat80ppm
Averageroom 80x14/2,700=0.4ppmofNO concentration of NO
ThistheoreticcalculationcanbesupplementedbymeasurementsasperformedbyHessetal.TheNO and NO2concentrationsweremeasuredusingachemiluminescenceanalyzerwhen100ppmofNOat8L/minwasdeliveredintoaroomwithnoscavengingbeingused.ThemaximumNO and NO2 concentrationsmeasuredoveraone-hourperiodwere 0.12 ppm of NO and 0.03 ppm of NO2.
Bothmethodsshowthattheexposurelevelsaresignificantlylessthanthelevelsrecommended byNIOSH.
IfthelocationforusingNOhasuncertainventilation,thenthelocationshouldbeevaluatedforNO and NO2builduppriortouse.
Ref. 1 Centers for Disease Control, Atlanta, GA 30333 USA. NIOSH Recommendations for Occupational Safety and Health Standards 1988. August 26, 1988 / vol. 37 /No. 9.
Ref. 2 Hess et al, Use of Inhaled Nitric Oxide in patients with Acute Respiratory Distress Syndrome, Respiratory Care, 1996, vol. 41, No 5, pg. 424-446.
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
2/ Setup
Setup
Part No. 20181 Rev-022012-06
2/ Setup
Setu
p
Part No. 20181 Rev-022012-06
7
SetupChecktheINOMAXgascylinderforthecorrectproductidentitylabels,cylinderconcentration(800or400ppm)andexpirationdate.EnsuretheINOMAXgascylinderhas>14bar(200psig).
1.Connectahigh-pressureregulatortoan INOMAXcylinderandhandtightenthefitting totheINOMAXcylinder.
2.ConnecttheINOMAXregulatorhosetothe INOblenderinlethoseusingthequick connectfittings.
3.ConnecttheoxygensupplytotheO2 inlet fitting.(seepartlistforcorrectoxygenhose) Note:3.5bar(50psig)nominal
1.MakesuretheNOdosesettingdialis turnedtozero.
2.Attachamanualresuscitationbagtothe INOblenderoutletfitting.
Tooxygensource
TheINOblenderisnowsetup.ProceedtoSection3/Pre-useCheckout.
1
23
1
2
Note: •FortheCGA-typeINOMAXregulatorconnector,ensurethewhiteplastictipisnotchippedorcracked.Removeandreplaceasnecessary.(seeReplacingthetipontheINOMAXregulator,Page16).
•FortheISO-typeregulatorconnector,checkthattheO-ringispresentandisnotdamaged(seeReplacingthetipontheINOMAXregulator,Page17).
Part No. 20181 Rev-022012-06
8
(Intentionallyleftblank)
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
3/ Pre-Use Checkout
Pre-U
se Checkout
Part No. 20181 Rev-022012-06
3/ Pre-Use Checkout
Pre
-Use
Che
ckou
t
Part No. 20181 Rev-022012-06
9
Pre-useCheckout
High-PressureLeakTest
1.MakesureNOdosesettingdialisturnedtozeroandflowmeterisOFF.
2.Openandthenclosethecylindervalve.3.Pressuregaugeshouldindicate>14bar (200psig).
4.Monitorpressuregaugefor30secondsforanysignsofpressuredecrease.Ifnopressuredecreaseisobserved,high-pressureleaktestsuccessful,proceedtoDeliveryConfirmationand Purge.
5.Ifobservedpressuredecreasecontinues,seeSection5/Maintenance;CylinderLeakCheck.
6.Ifleakcannotbetraced,replacetheINOblender.
DeliveryConfirmationandPurge
1.SettheINOblenderto40ppmwhenusingan 800ppmcylinder(20ppmwhenusinga400 ppmcylinder).
2.SettheoxygenflowontheINOblenderflow meterto10L/mintobeginpurge.
3.Ensurethepressuregaugedecreases approximately14bar(200psig)in10seconds (+2sec.).
4.Continuepurginguntilpressuregauge readszero.
Caution: Tohelpensureproperoperation,completethepre-usecheckoutpriortoeachuse.
Set-uptheINOblenderasdescribedinSection2/Setup
TheINOblenderisnowreadyforuse.ProceedtoSection4/Operation.
1
2
3
1
2
4. Note: Ifthepressuredoesnotdecrease,then
theINOblenderisnotdeliveringNOandtheINOblendershouldbereplaced.
Part No. 20181 Rev-022012-06
10
(Intentionallyleftblank)
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
4/ Operation
Operation
Part No. 20181 Rev-022012-06
4/ Operation
Ope
ration
Part No. 20181 Rev-022012-06
11
(Intentionallyleftblank)
Part No. 20181 Rev-022012-06
12
Operation
Cautions:• Whennotinuse,theoxygenflowmeterandtheINOMAXcylindervalveshouldbeturnedoff.
• WhentheINOblenderisusedwithanoxygen/airblender:
- ThespecificationforINOMAXdeliverywhenusingtheINOblenderwith100%oxygenis+/-20%ofsettingor2ppm(whicheverisgreater).Theuseof100%oxygenat3.4bar(50psig)isthelabeledspecificationfortheINOblender.
- Ausermaydeterminethatsomeclinicalconditionsmaynecessitatetheuseofanoxygen/airblenderwiththeINOblendertoachieveFiO2levelslessthan100%.
- Usingoxygen/airmixtures(21%to95%v/v)willreducethedeliveredNOconcentrationbyupto10%ofsettingor1ppm(whicheverisgreater)comparedtousing100%oxygenalone,resultinginacumulativeerrorupto+/-30%ofsettingor3ppm(whicheverisgreater).
• Refertothemanufacturer’sprocedureswhenusingthemanualresuscitatorbag.• Whenfinished,turntheNOcylindervalveOFFandallowtheoxygentoflowuntiltheNOpressuregaugereadszero,thenturntheoxygenflowOFF.
WARNINGS:• ThepurgeproceduremustbefollowedtohelpensureNO2ispurgedfromthesystembeforethemanualresuscitatorbagisconnectedtothepatient.
• ThemanualbagshouldbesqueezedrepeatedlyduringusetoavoidNO2frombuildingupinthebag.
• IfthebagisnotsqueezedrepeatedlywhiledeliveringINOMAX,thebagshouldberemovedfromthepatientandthebagpurgeprocedureperformedbeforecontinuing.
• TheINOblendershouldbeuprightwhensettingtheoxygenflowrateforaccuratesetting.
• DonotusepneumaticallypowerednebulizerswiththeINOblender.ThiswillresultinsignificantoverdeliveryofINOMAX(nitricoxide)forinhalationinexcessof80ppm(40ppmwitha400ppmcylinder).
- TheINOblenderoutletpressurehasbeenvalidatedforuseupto400millibar(5.8psig)pressure.Theamountofback-pressuregeneratedbypneumaticnebulizersissignificantlygreater[1.4-2.0bar(20-30psig)]andwillresultinoverdeliveryofINOMAXinexcessof80ppmwithan800ppmcylinder(40ppmwitha400ppmcylinder).TheuseradjusteddosesettingontheINOblenderwillnotcorrelatewith,orhaveaneffecton,theactualdelivereddose.
- Inaddition,theINOblenderflowmeterisnotback-pressurecompensatedandwilldisplayalowerthanactualflowratewhenpressureisappliedtotheoutlet.
Part No. 20181 Rev-022012-06
13
ThepurgeprocedurecontainedinthePre-useCheckouthelpsensureanyNO2ispurgedfromthepressureregulator,INOblenderandhosesbeforethemanualresuscitatorbagisconnectedtothepatient.
1.OpentheINOMAXcylindervalve.
2.Checkforsupplypressureof>14bar (200psig).
1
2
AdjustSettings
1.TurntheINOblendersettingdialtothe desiredconcentration(5to80ppmforan 800 ppm cylinder and 2.5 to 4.0 ppm for a 400ppmcylinder).
2. Turn the O2flowmetertothedesiredflowrate (5to14L/min).
3.Squeezethemanualresuscitator3-4times to purge the NO2fromthesystem.
TheINOblenderisnowreadyforpatientuse.
1
2
3
Priortooperation,completethePre-UseCheckoutinSection3.
Note: DepressurizetheINOMAXregulatorbyusingthepurgeportonthebackoftheINOmaxDSIR prior to removing from the cylinder valve.
Part No. 20181 Rev-022012-06
14.
(Intentionallyleftblank)
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
5/ Maintenance
Maintenance
Part No. 20181 Rev-022012-06
5/ Maintenance
Mai
nten
ance
Part No. 20181 Rev-022012-06
15
Cautions: • Whencleaning,donotsubmergeinliquid.• Donotautoclave.
CleaningProcedure
Maintenance
CleaningAgent ActiveIngredientsPreciseHospitalFoamCleanerDisinfectantbyCaltechIndustries
o-Phenylphenol<0.37%Otheringredients99.63%
PureGreen24byPureGreen,LLC SDC–silverions0.003%Citricacid4.84%Otheringredients95.157%
PDISuperSaniClothbyPDI n-Alkyl(68%C12,32%C14)dimethylethylbenzylammoniumchlorides0.25%n-Alkyl(60%C14,30%C16,5%C12,5%C18)dimethylbenzylammoniumchlorides0.25%Isopropylalcohol55%Inertingredients44.50%
SaniClothHBbyPDI n-Alkyl(60%C14,30%C16,5%C12,5%C18)dimethylbenzylammoniumchlorides0.07%n-Alkyl(68%C12,32%C14)dimethylethylbenzylammoniumchlorides0.07%Inertingredients99.86%
Asepti-HBbyEcolabInc. n-Alkyl(60%C14,30%C16,5%C12,5%C18)dimethylbenzylammoniumchlorides0.07%n-Alkyl(68%C12,32%C14)dimethylethylbenzylammoniumchlorides0.07%Inertingredients99.86%
CavicideandCaviWipesbyMetrex Diisobutylphenooxyethoxyethyldimethylbenzylammoniumchloride0.28%Isopropylalcohol17.2%Inertingredients82.52%
Caution:• Applycleaningagenttoaclothbeforeapplication;donotspraycleaningagentdirectlyonthedevice. Itisimportanttopreventpoolinganddirectcontactwithopenconnectionswhichcancausedamage over time.- CleantheoutersurfaceoftheINOblenderwithasoftclothdampenedinamildsoapandwatersolution(Example:PalmoliveUltraantibacterial,4dropsper50mLofwarmwater),isopropylalcohol(70%),orwithoneofthefollowingcleaningagentswhilefollowingthemanufacturer’srecommendations.
Note: TheINOblenderdoesnotcontainanyuserrepairableparts.
BioquellHydrogenPeroxideSterilantBioquellHydrogenPeroxideSterilantandhydrogenperoxidevaporgeneratorsareregulatedbytheUSEnvironmentalProtectionAgency(EPA)aspesticidechemicalsinaccordancewithFederalInsecticide,Fungicide,andRodenticideAct(FIFRA).Useoftheseproductsincleaning/disinfectionprocesseshavenotbeenvalidatedwiththeINOblender.DonotusetheseproductstodecontaminatetheINOblenderoranyancillaryproductsusedwiththeINOblender.
Part No. 20181 Rev-022012-06
16
ReplacingtheCGA626tipontheINOMAXregulator
YearlyCalibrationTheINOblendercalibrationshouldbecheckedevery12monthsusingtheNOmonitoroftheprimaryNOdeliverysystem.Thetestpointsshouldincludeallthecontrolsettingsatoxygengasflowsof5and14L/min.IftheINOblenderisoutsideoftheaccuracyspecification,itshouldnotbeused.
ContactCustomerSupportatthenumberonthebackcoverofthismanualforrepair.
Figure 5-1
1.DisconnecttheregulatorfromtheINOMAX gascylinder.
2.RemovetheoldCGA626tip(forISOsee Figure5-2)bypullingonthetip(a)andturning itcounterclockwise(b)(seeFigure5-1).
3.Ensurethethreadsarecleanonthe regulatortip(ifrequired,usealintfreecloth).
4.Installthenewtip: Flexthefourprongsbysqueezingtwoprongs atatimeusingonlyyourfingers.Thiswillhelp startthenewtipintothethreads.Turnthetip clockwisewhenthreadingthetip.Whenthe tipisfullyinserted,itshouldturnfreely.
a
b
Note: •DepressurizetheINOMAXregulatorbyusingthepurgeportonthebackoftheINOmaxDSIR prior to removing from the cylinder valve.
•ChecktheINOblenderandconnectionhosesforsignsofwearanddamage. •FortheCGA-typeregulatorconnector,ensurethewhiteplastictipisnotchippedorcracked.Remove
andreplaceasnecessary.(seeFigure5-1). •FortheISO-typeregulatorconnector,checkthattheO-ringispresentandisnotdamaged
(seeFigure5-2).
Part No. 20181 Rev-022012-06
17
Figure 5-2
1.DisconnecttheregulatorfromtheINOMAXtherapygascylinder.
2.RemovetheoldISOO-ringbyrollingitoffitsgroove(seeFigure5-2).
3.Cleantheconnectortip(ifrequired,usealintfreecloth).
4.RollthenewO-ringintoitsgroove.Whencorrectlyinstalled,itshouldnotberemovablebyturningit.
Caution: DonotusehardobjectstoremovetheO-ringastheymaydamagethemetalglandand causealeak.
O-ring
Gland
ReplacingtheO-ringontheISO5145INOMAXregulatorfitting
Note: DepressurizetheINOMAXregulatorbyusingthepurgeportonthebackoftheINOmaxDSIR prior to removing from the cylinder valve.
Part No. 20181 Rev-022012-06
18
CylinderLeakCheckIfaleakissuspectedduringthehigh-pressureleaktest(seeSection3/Pre-UseCheckout;HighPressureLeakTest),thefollowingstepscanbetakentocheckforleaks(seeFigure5-3forpossiblecylindergasleaklocations)intheINOMAXregulatororINOMAXcylinder.
Figure 5-3
1. CylinderValveRegulatorConnection2. INOMAXRegulatorHandWheelConnection3. RegulatorEndCapConnection4.. Tamper Evident Tape5. Valve Nut6. SafetyPressureReleaseDevice
1. EnsurethatINOMAXregulatorisconnectedtocylindervalveoutlet(hand-tightenonly),thatthecylindervalveisopen,andthatthecylinderhasmorethan14bar(200psig).
2. Applysoapywatertopoints1,2,3,5and6(seeFigure5-3);ifbubblesform,thereisaleak.
3. Iftherearenobubbles,leakmaybeinsideINOblenderorINOmaxDSIRandcannotberepaired.ReplaceINOblenderorINOmaxDSIR andcontactCustomerSupport.
Recommendedactionsshouldaleakbedetected:
1. Aleakdetectedatpoints1and2(seeFigure5-3)maybecorrectedbytighteningtheINOMAXregulator hand wheel.a. Ifcylindervalveisopen,closecylindervalve
andtightenINOMAXregulatorhandwheel.b.Opencylindervalveandreapplysoapywater
topoints1and2.c.Ifbubblesform,thereisaleak.d.RemoveINOMAXregulatorandcheckwhite
plastictiponINOMAXregulatorforchipsorcracks.Replaceifnecessary(seereplacingthetipontheINOMAXregulator,page16).FortheISOtyperegulatorconnector,checkthattheO-ringispresentandisnotdamaged.Replaceifnecessary(seereplacingtheO-ringontheINOMAXregulator,page17).Repeatstepb(note:Ifleakremains,replaceINOMAXregulator).
2. Ifaleakisdetectedbetweentheregulatorbodyandregulatorendcap(seepoint3),replaceINOMAXregulatorandcontactCustomerSupport.
3. Aleakdetectedatcylindervalvenutconnection(seepoint5)maynotberepaired. ReplaceINOMAXcylinderandcontactCustomerSupport.
4. Aleakdetectedatsafetypressurereleasedevice(seepoint6)maynotberepaired.ReplaceINOMAXcylinderandcontactCustomerSupport.
6
5
1
2 3
4
Note: Refertohospitalpoliciesandprocedures fordealingwithleakinggascylinders. Additionalinformationregarding environmentaleffectscanbefoundin thesection1/GeneralInformation.
Part No. 20181 Rev-022012-06
19
PartsandAccessories
WARNING: Useonlyparts/accessoriesdesignatedforusewiththissystem.
Parts/Accessories PartNumberClampAssembly 10008HoseAssembly,OxygenGreen,DISStoDISS,15Feet 8034.4.HoseAssembly,OxygenWhite,DISStoDISS,15Feet 8034.5HoseAssembly,OxygenWhite,DISStoSIS,15Feet 80353HoseAssembly,OxygenWhite,NISTtoNIST,15Feet 804.81INOMAXRegulator,CGA626 10006INOMAXRegulator,ISO5145 10020INOMAXRegulatorExtensionHose 10014.Operation and Maintenance Manual 20181O-ring,ISO5145INOMAXRegulator 804.70Tip,CGA626INOMAXRegulator 1605-314.9-000TransportMountingBracketAssembly/Spacer 5004.1
OperationandMaintenanceManual PartNumberEnglish 20181Czech 20308Danish 20309Dutch 20310Finnish 20311French 20312German 20313Hungarian 20314.Italian 20315Korean 20316Norwegian 20317Polish 20318Portuguese 20319Spanish 20305Swedish 20320
Part No. 20181 Rev-022012-06
20
(Intentionallyleftblank)
Part No. 20003 Rev - 01 Part No. 20181 Rev-022012-06
6/ Product Specifications
Part No. 20181 Rev-022012-06
6/ Product Specifications
Part No. 20181 Rev-022012-06
21
Specifications
INOblender
Ü NOSetRange 5to80ppm(800ppmcylinder) 2.5to40ppm(400ppmcylinder) Settingaccuracy +/-20%indicatedor2ppm,whicheverisgreater Gasflowrange 5to14L/minofO2 O2inletconnection DISS(male),3.4bar(50psig,nominal) NO/N2inletpressure 1.7to2.4bar(25to35psig) NOinlet Quickdisconnectwithshut-off(Female) NO/O2outlet Barbfittingforoxygensupplytubing Max.Dimensions 200mm(W)X120mm(H)X110mm(D)DISS 200mm(W)X120mm(H)X131mm(D)NIST Max.weight 1.5kg
OperatingandStorageConditions
OperatingTemperature 10to40°C OperatingHumidity 15%to95%non-condensing OperatingBarometricPressure 57to110kPa StorageTemperature -40to+70°C StorageHumidity 15%to98%noncondensing StorageBarometricPressure 57to110kPa
INOMAXPressureRegulator
CylinderValveConnector CGA626orISO5145valve InletPressure 14to155bar(200to2,248psig) NOOutletPressure 1.7to2.4bar(25to35psig)
ManualResuscitatorsThe INOblender is compatible with most types of manual resuscitation bags when used within the specifications of this device. The INOblender has been validated against the following manual resuscitation bags.*
*AirlifeAdultManualResuscitatorwithreservoirtubing *Allegiance½LPediatricManualResuscitatorwithreservoirtubing
Part No. 20181 Rev-022012-06
22
(Intentionallyleftblank)
17Part No. 20181 Rev-02
2012-06
23
(Intentionally left blank)
EuropeEuropean Authorized Representative
MDSS GmbHSchiffgraben 41
30175 Hannover, Germany
AustraliaAustralia RepresentativeIkaria Australia Pty Ltd.
+61 3 9882 4855
Europe
Part No. 20181 Rev-022012-06
ManufacturerNorth America
INO Therapeutics LLC d/b/a Ikaria, Inc.Perryville III Corporate Park
53 Frontage Road, Third Floor Hampton, NJ 08827-9001, USA
1-877-566-9466www.ikaria.com