EndoSoft DICOM Conformance Statement EndoSoft © Version 6.0 EndoVault © Version 1.3 Revision 1.10

Transcript of EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from...

Page 1: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

 

 

EndoSoft DICOM Conformance Statement   

  

EndoSoft© Version 6.0 

EndoVault©  Version 1.3 

 

Revision 1.10  

 

 

 

 

 

 

 

 

 

 

 

Page 2: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Revision History  

Date      Notes 

01/25/2007    Version 1.3 

02/25/2008    Version 1.4 

03/06/2008    Version 1.5 

05/07/2008    Version 1.6 

09/15/2008    Version 1.7 

11/23/2008    Version 1.8 

03/10/2009    Version 1.9 

06/22/2010    Version 1.10 (Added DICOM Encapsulated PDF information) 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 3: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

 

Table of Contents  

1.    Introduction              1 

1.1.    Intended Audience            3 

1.2.    Scope                3 

1.3.    Important Remarks            3 

1.4.    References              4 

1.5.    Definitions              4 

1.6.    Modalities supported           5 

 

2.     Network Conformance Statement        5 

2.1.    Implementation Model          5 

2.2.    Application Data Flow Diagram        6 

2.3.    Functional Definitions of AE’s         7 

2.4.    Sequencing of Real‐World Activities         

 

3.     AE Specifications            8 

3.1.    EndoSoft systems Specification        8   

3.1.1.    Association Establishment Policies      8 

3.1.1.1    General          8 

3.1.1.2    Number of Associations      8 

3.1.1.3   Asynchronous Nature       9 

3.1.1.4   Implementation Identifying Information  9 

 

3.1.2.    Association Initiation Policy        9 

3.1.2.1   Associated Real‐World Activity    9 

3.1.2.2   Associated Real‐World Activity     

Query Modality Worklist Items from     

a remote node        9 

3.1.2.3   Associated Real‐World Activity 

Store images in a Destination Node    10 

 

3.1.3.     Proposed Presentation Contexts      11 

3.1.3.2   Association Acceptance Policy    11 

 

3.2.     SOP Specific Conformance          11 

Page 4: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

 

Table of Contents Continued 

 

3.2.1     SOP Specifics Conformance for Storage SOP Class  11 

 

3.2.1.1   “Secondary Capture” Image Storage  

Attributes          11 

3.2.1.2   Error Handling         18 

 

3.2.2       SOP Specific Conformance for Modality  

Worklist          18 

3.2.2.1   Error Handling         19 

     

4.  Communication Profiles            19 

4.1.  Supported Communications Stacks (Part 8)        19 

4.2.  TCP/IP Stack                19 

4.3.  Physical Media Support            19 

 

5.  Extensions/Specializations/Privatizations        19 

 

6.  Configuration                19 

 

7.  Support of Extended Character Sets         19 

 

8.  Codes and Controlled Terminology          19 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 5: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

1. Introduction  

Overview 

This DICOM Conformance statement covers the following EndoSoft systems: EndoSoft GI, EndoSoft Urology, 

EndoSoft Pulmonary, EndoSoft Pain Management, EndoSoft Orthopaedic, EndoSoft ENT, EndoSoft Cardiology, 

EndoSoft Nuclear Medicine, EndoSoft Ophthalmology, EndoSoft Pathology, EndoSoft Dermatology, EndoSoft 

General Surgery and EndoSwift© Image management system.  Throughout the document these systems will be 

collectively referred to as EndoSoft systems. 

 

This DICOM Conformance Statement is divided into sections as described on the next page. 

1   

Section 1 (Introduction) begins with an overview and then describes the overall structure, intent, and 

references for this Conformance Statement. 

 

Section 2 (Network Conformance Statement) begins with an Implementation Model; which contains an 

Application Dataflow Diagram. This section also contains the definition of AE’S and sequencing of real world 

application. 

 

Section 3 (AE Specifications) describes the overall structure, systems specification, intent, and references for 

this Conformance Statement.  It also describes the association Policies and SOP specification. 

       

Section 4 (Communication Profiles) specifies the EndoSoft systems compliance to the DICOM requirements 

for the implementation of Communications and Networking features. 

 

This document is considered essential to provide interoperability between imaging equipment and modalities 

and EndoSoft systems which is necessary for efficient patient care. 

 

EndoSoft systems utilizes a DICOM module in order to communicate and export visible light as well as 

ultrasound images to other DICOM compliant systems by which it can fully participate in a DICOM 

environment. 

 

The Introduction serves to familiarize the reader with DICOM terminology and general concepts.  It should be 

read prior to reading the Conformance Statements for the individual products.  

 

In addition, the EndoSoft systems Conformance Statement, specifies the Lower Layer communications that it 

supports (e.g., TCP/IP).  However, the Technical Specifications are defined in the DICOM Part 8 standard.  For 

more information including Network Architecture and basic DICOM concepts, please refer to the Introduction.  

 

For more information regarding DICOM, copies of the Standard may be obtained on the Internet at 

http://www.medical.nema.org. 

Page 6: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Comments on the Standard may be addressed to:  

 

    DICOM Secretariat NEMA  

    1300 N. 17th Street, Suite 1847  

    Rosslyn, VA 22209 USA  

    Phone: +1.703.841.3200  

 

1.1  Intended Audience 

 

This Conformance statement is intended for existing or potential users of EndoSoft systems, system 

administrators of institutions using EndoSoft systems, as well as developers of systems wishing to 

communicate with EndoSoft systems using the DICOM protocol. 

   

The reader of this document is concerned with software design and/or system integration issues. It is assumed 

that the reader of this document is familiar with the DICOM Standard and with the terminology and concepts 

which are used in that Standard.  

 

If readers are unfamiliar with DICOM terminology they should first refer to the http://www.medical.nema.org 

then read the DICOM Standard itself, prior to reading this DICOM Conformance Statement document. 

 

1.2   Scope 

 

It is the intent of this document to provide an unambiguous specification for DICOM implementations of 

EndoSoft systems. This specification, called a Conformance Statement, includes a DICOM Conformance  

 

Statement and is necessary to ensure proper processing and interpretation of EndoSoft systems medical data 

exchanged using DICOM. The EndoSoft systems DICOM Conformance Statements are available to the public. 

 

1.3  Important Remarks 

 

  The use of these DICOM Conformance Statements, by itself, it is not sufficient to ensure that inter‐

operation will be successful. The user (or user's agent) needs to proceed with caution and address at least the 

following issues:  

     

• Integration ‐ The integration of any system into an integrated practice, department or enterprise 

requires an analysis of the applications requirements. The responsibility to analyze the applications 

requirements and to design a solution that integrates EndoSoft systems with the user is strongly 

advised to ensure that such an integration analysis is correctly performed. 

     

Page 7: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

• Validation ‐ Testing the complete range of possible interactions between any EndoSoft Systems and 

non–EndoSoft Systems, before the connection is declared operational, should not be overlooked. 

Therefore, the user should ensure that any non–EndoSoft Systems provider accepts full responsibility 

for all validation required for their connection with EndoSoft Systems.  

 

This includes the accuracy of the data across the interface, and the suitability of data for the intended 

applications. Such a validation is required before any clinical use (diagnosis and/or treatment) is performed. 

 

• Future Evolution ‐ EndoSoft understands that the DICOM Standard will evolve to meet the   user's 

growing requirements.  

 

• EndoSoft reserves the right to discontinue or make changes to the systems features including those 

reflected by this DICOM   Conformance statement. 

 

• Interaction – It is the sole responsibility of the Non‐EndoSoft systems providers to ensure that 

communication with the interfaced equipment does not cause degradation of EndoSoft systems 

performance and/or function. 

 

1.4   References 

 

Digital Imaging and Communications in Medicine (DICOM) standard by the National Electrical Manufacturers 

Association (NEMA). 

 

  DICOM Standard is available on internet: 

  http://www.medical.nema.org 

 

1.5  Definitions 

 

Table 1: Definitions and Abbreviations 

 

AE  DICOM Application Entity 

AET  Application Entity Title 

DICOM  Digital Imaging and Communications in Medicine 

FSC  File Set Creator 

IOD  DICOM Information Object Definition 

PDU  Protocol Data Unit 

SCP  Service Class Provider 

SCU  Service Class User 

SOP  Service Object Pair 

TCP/IP  Transmission Control Protocol/Internet Protocol 

UID  Unique Identifier 

 

Page 8: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

 

1.6  Modalities supported 

   

CP  =  Colposcopy  

CS   =   Cystoscopy  

CT   =   Computed Tomography  

DM   =   Digital Microscopy  

ES   =   Endoscopy  

LP   =   Laparoscopy  

NM   =   Nuclear Medicine  

RF   =   Radio Fluoroscopy  

US   =   Ultrasound  

 

 2.   Network Conformance Statement This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance 

requirements for the relevant Networking features on this EndoSoft Systems product. The EndoSoft Systems 

utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a 

network.  

 

Note that the format of this section strictly follows the format defined in DICOM  

Standard PS 3.2 (Conformance). Please refer to that part of the standard while reading this section. 

 

2.1  Implementation Model 

 

EndoSoft systems support the DICOM Modality Worklist as well as Storage. It provides following DICOM 

Service class. 

 

DICOM Modality Worklist SCU (C‐FIND). 

DICOM Storage SCU (C‐STORE). 

DICOM Second Capture Image Storage SCU (C‐STORE). 

DICOM Encapsulated Document Image Storage SCU (C‐STORE) 

Verification SOP Class 

 

 

 

 

 

 

 

Page 9: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

2.2  Application Data Flow Diagram 

 

This section contains the Application Data Flow model for the EndoSoft systems Application Entity (AE) and 

how it works with remote DICOM devices. There is a functional definition of the AE and its Real‐World 

Activities, and a description of the Real‐World sequence of events of the AE. 

 

The network application model for the EndoSoft systems is shown in the following Illustration:  

 

  

The EndoSoft DICOM SERVER Application Entity (AE) is an application that handles DICOM protocol 

communication. The DICOM SERVER AE is automatically brought up when the EndoSoft Systems is powered on 

and the main application is started. 

 

All remote DICOM entities must be manually entered on the EndoSoft Systems by a system admin or an 

EndoSoft field engineer. 

 

 

Page 10: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

The DICOM SERVER AE is invoked by the following: 

 

Manual Transmit from the EndoSoft Systems to a Remote Host 

For this operation, the user selects a series from the EndoSoft patient database window and then sends the 

selected series to a remote DICOM AE.  Application selects the host name from a configuration file.  To change 

the host name, User has to access the menu item on patient data screen under tools by the name “Node” and 

changing the Node name as the data destination. 

 

The declaration of the remote host is done through the Node Configuration window that can be accessed 

through the System Administration menu.  The Node Configuration window is used to configure the local host 

and define remote entities. 

 

The status of the transfer is displayed on the status bar of the patient database window. Any error in transfer 

is communicated to the user through a pop‐up message indicating the nature of the error. 

 

Only one series at a time may be selected for network transfer, but additional transfers can be initiated while 

another data set is being transferred. The series information will be held in a buffer and the additional transfer 

requests will be executed in the order that they were initiated as each previous transfer is completed. 

 

2.3  Functional Definitions of AE’s 

 

When EndoSoft systems are invoked, it shall have “Delphi” as it’s calling  AE by default. It is configurable and 

can be defined at the time of installation. It is capable of utilizing the full 16‐bit range (1‐65535) of port 

numbers. EndoSoft systems sends images to remote DICOM AE’s after first converting them to DICOM from 

their locally stored file format.  

 

When the modality worklist data is requested, the application initiates the DICOM association request for the 

modality worklist service from the remote DICOM Modality Worklist SCP. When the association is established, 

the application further issues one or more C‐FIND requests to the remote SCP.    

 

Ones the set of modality items are received from the remote Modality Worklist SCP, the application inserts 

these items into the EndoSoft database. 

 

Under EndoSoft DICOM Storage Module, EndoSoft Systems capture Images in jpegs and then converts it into 

DICOM images.  

 

Attributes are automatically taken from the modality Worklist item, but the user may enter the performed 

procedure data, and certain attributes are pre‐configured.  The captures object will be sent to the remote 

archive server or image viewer with DICOM C‐STORE service. 

 

Page 11: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

The modality Worklist, and/or storage commitment transactions may be disabled according to the site. 

 

2.4  Sequencing of Real‐World Activities 

 

EndoSoft systems DICOM AE waits for user requests providing a user interface selection of files to be sent to 

remote DICOM entities. After files have been selected and the procedure has been signed off, the DICOM 

server running on the server will pick those images and will be send to the selected DICOM Entity. 

 

Likewise for WORKLIST, EndoSoft systems can query for Patient and study information (i.e., issue “C‐FIND” 

request). EndoSoft systems can request the list of patients that are scheduled based on certain search criteria.  

 

3  AE Specifications    This section describes EndoSoft systems Application Entity. 

 

3.1  EndoSoft Systems Specification 

 

  EndoSoft systems provide Standard Conformance to the following DICOM V3.0 SOP Classes as an SCU. 

 

Table 2: SOP Class 

SOP Class Name  SOP Class UID  

Verification SOP class  1.2.840.10008.1.1 

Secondary Capture Image Storage  1.2.840.10008.5.1.4.1.1.7 

Modality Worklist ‐ Find   1.2.840.10008.5.1.4.31 

Encapsulated PDF Storage  1.2.840.10008.5.1.4.1.1.104.1 

 

 

3.1.1  Association Establishment Policies 

 

3.1.1.1  General 

 

EndoSoft systems will attempt to establish associations whenever it is invoked with appropriate parameters.  

The maximum PDU size which EndoSoft systems will use is configurable, with a default of 64K. 

 

3.1.1.2  Number of Associations 

 

There are no inherent limitations on the total number of simultaneous associations which the Application 

Entity represented by EndoSoft systems can maintain. 

Page 12: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

3.1.1.3  Asynchronous Nature 

 

EndoSoft systems will not perform asynchronous operations window negotiation. All operations will be 

performed synchronously. 

 

3.1.1.4  Implementation Identifying Information 

 

The software version number of the systems using this DICOM software is used as the Implementation Version 

Number and a component of the Implementation UID.   

   

This is a 5 digit number that will be updated whenever EndoSoft updates the software version. 

 

•  Implementation Class UID:  1.2.826.0.1.3680043.1.1.X…, 

•  Implementation Version Name : 6.1.5.02 

 

3.1.2  Association Initiation Policy 

 

When the DICOM SERVER Application Entity initiates an Association for  any Real‐World Activity, it will 

propose the Presentation Contexts for all Real‐World Activities; i.e., there is only a single, comprehensive 

Presentation Context Negotiation proposed for the AE. 

 

3.1.2.1 Associated Real‐World Activity 

 

When a DICOM association is requested, an association is initiated so as  to check for Validity and verification 

of the remote AE’s ability to handle the requested association from EndoSoft systems. Once acceptance of the 

association request has been verified, the local DICOM image files stored at the local DICOM AE will be sent to 

the remote DICOM AE until finished. If more than a single file is to be sent, each file is first stored Confirmation 

of remote storage success received, and then the next file sent until all files are successfully processed. 

 

3.1.2.2 Associated Real‐World Activity ‐ Query Modality Worklist Items from a remote Node 

 

The associated real world activity is a C‐FIND request invoked by EndoSoft Systems. Once the association is 

successfully established, EndoSoft sends the C‐FIND request along with Modality Worklist query criteria to the 

remote node. When the remote node returns the matched data items with its response EndoSoft systems will 

add these data items in the Modality Task Worklist. 

 

 

 

 

 

Page 13: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

3.1.2.3 Associated Real‐World Activity ‐ Store images in a Destination Node 

 

The associated real world activity is a C‐Store request invoked by EndoSoft Systems. Once the association is 

successfully established, EndoSoft sends the acquired image, which is already processed, to the destination 

DICOM node that is assigned to store the image. 

 

3.1.3  Proposed Presentation Contexts 

 

EndoSoft systems will only propose a single Presentation Context in each association. This presentation 

context will use the SOP Class UID of the instance to be transmitted as the proposed Abstract Syntax, and the 

proposed Transfer Syntaxes will be as shown in Table 3. 

 

Table 3: Proposed Presentation Contexts for EndoSoft systems 

 

Abstract Syntax    Transfer Syntax    Role  Extended Negotiation 

Name  UID  Name List  UID List  Role   

Secondary Capture  1.2.840.10008.5.1.4.1.1.7  Implicit VR Little Endian 

1.2.840.10008.1.2 

SCU  None 

MWL  1.2.840.10008.5.1.4.31  Implicit VR Little Endian 

1.2.840.10008.1.2.1 

SCU  None 

Verification  1.2.840.10008.1.1  Implicit VR Little Endian 

1.2.840.10008.1.2 

SCU  None 

Encapsulated PDF  Storage 

1.2.840.10008.5.1.4.1.1.104.1  Implicit VR Little Endian 

1.2.840.10008.1.2.1 

SCU  None 

    Implicit VR Little Endian 

1.2.840.1000.8.1.2 

SCU  None 

 

3.2  SOP Specific Conformance 

 

3.2.1 “SOP Specifics Conformance for Storage SOP Class” 

   

EndoSoft systems do not check, nor modify any elements of the image files it transfers, though it may 

compress or decompress the pixel data according to the negotiated transfer syntax. 

  

The content of the images sent/stored therefore depends entirely on the contents of the files provided in the 

source directory. 

 

 

 

 

 

 

10 

Page 14: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

3.2.1.1 “Secondary Capture” Image Storage Attributes. 

 

USER – this attribute value source is from User input. 

AUTO – this attribute value is generated automatically  

IMG File – this attribute value is generated via vendor  

MWL – this attribute value is selected from Modality Worklist when enabled 

CONFIG – this attribute value source is a configurable parameter 

 

For the Presence of Value Column 

 

ALWAYS – this attribute is always in the SOP Instance and always with a  valid value 

VNAP – this attribute is always in the SOP Instance but may be "set to" 

Zero length (meaning unknown) 

EMPTY – this attribute value is always “set to” zero length 

ANAP – this attribute is not always present 

 

Table 4:  

Patient Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Patient’s name  0010,0010  PN  USER/ 

Img 

File/M

WL 

ALWAYS Patient Full Name(Table Patients)

Patient ID  0010,0020  LO   USER/I

mg File/ 

MWL 

ALWAYS Patient MRN (Table Patients) 

Patient’s Birth 

Date 

0010,0030  DA  USER/I

mg File/ 

MWL 

ALWAYS Patient  DOB (Table Patients) 

Patient’s Sex  0010,0040  CS  USER/I

mg File/ 

MWL 

ALWAYS Patient Sex (Table Patients) 

Ethnic Group  0010,2160  SH  USER/I

mg File/ 

MWL 

ALWAYS Patient Race (Table Patients) 

Page 15: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Other Patient 

IDs 

0010,1000    USER/m

wl 

VNAP Account Number field in Patient 

Table 

Patient 

Comments 

0010,4000  LT  USER VNAP Comments Field (Table Patients) 

 

General Study Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Study Instance 

UID 

0020,000D  UI  MWL/

Auto 

ALWAYS Identifier Unique to Study – From 

Modality or Generated by Device. 

Study ID  0020,0010  SH  MWL/

Auto 

ALWAYS Procedure Number  (Procedure 

Table) 

Study Date  0008,0020  DA  USER/I

mg File/ 

MWL 

ALWAYS Date of Procedure (Procedure Table)

Study Time  0008,0030  TM  USER/I

mg File/ 

MWL 

ALWAYS Time of Procedure (Procedure Table)

Referring 

Physician’s 

Name 

0008,0090  PN  User VNAP Referring Physician (Procedure 

Table).Physician making referral 

Accession 

Number 

0008,0050  SH  MWL/

User 

ANAP From Modality Worklist or user 

input 

Study 

Description 

0008,1030  LO  User ALWAYS ER_ProcedurePerformed (Procedure 

Table) 

Reference 

study 

sequence 

0008,1110  SQ  MWL ANAP Study Scope sequence 

From Modality Worklist 

SOP Instance 

UID 

0008,0018  UI  AUTO ALWAYS Identifies the SOP instance uniquely. 

Generated by Device.  

>Referenced 

SOP Class UID 

0008,1150  UI  MWL VNAP Uniquely identifies the referenced 

SOP class. From Modality Worklist 

Page 16: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

>Referenced 

SOP Instance 

UID 

0008,1155  UI  MWL VNAP From Modality Worklist 

Physician(s) of 

Record 

0008,1048    User ALWAYS Physician (Procedure Table). 

Responsible for the patient’s overall 

care during study. 

 

Patient Study Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Additional 

Patient History 

0010,21B0    USER/I

mg File/ 

MWL 

ANAP History Field (Table Patients) 

 

General Series Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Modality  0008,0060  CS  Auto ALWAYS ES

Series Instance 

UID 

0020,000E  UI  Auto ALWAYS Identifier unique to series. 

Generated by device 

Series Number  0020,0011  IS  Auto ALWAYS Series ID number. Generated by 

device 

Laterality  0020,0062      VNAP

Series Date  0008,0021  DA  Auto ALWAYS DateTime series began. 

<yyyymmdd> 

Series Time  0008,0031  TM  Auto ALWAYS Time series began.  

<hhmmss> 

Series 

Description 

0008,103E  LO  User VNAP Description provided by series user. 

Maximum 512 characters. 

Content Date  0008,0023  DA  User ANAP <yyyymmdd>

Content Time  0008,0033  TM  User ANAP <hhmmss>

Patient 

Position 

0018,5100    User ANAP

Page 17: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Patient 

Orientation 

0020,0020  CS  Auto Empty

Abstract 

Syntax 

(Secondary 

Capture Image 

Storage) 

0008,0016    Auto ALWAYS 1.2.840.10008.5.1.4.1.1.7 

Referenced 

Study 

Component 

Sequence 

0008,1111    Auto ANAP Uniquely identifies series referenced 

study component SOP instance. 

>Referenced 

SOP Class UID 

0008,1150  UI  MWL VNAP Uniquely identifies the referenced 

SOP class. From Modality Worklist 

>Referenced 

SOP Instance 

UID 

0008,1155  UI  MWL VNAP Uniquely identifies the referenced 

SOP instance. From Modality 

Worklist 

>Requested 

ProcedureID 

0040,1001    Auto Identifier used for identification of 

requested procedure included in the 

Image service request. 

>Scheduled 

Procedure 

Step ID 

0040,0009  SH  MWL VNAP Classification of Scheduled 

Procedure Step Described. From 

Modality Worklist 

Performed 

Procedure 

Step Start 

Date 

0040,0244  DA  User ALWAYS Date When the Performed 

Procedure Step began. Date of 

Procedure (Procedure TABLE) 

Performed 

Procedure 

Step Start 

Time 

0040,0245  TM  User ALWAYS Time When the Performed 

Procedure Step began. 

TimeofProcedure (Procedure TABLE) 

Operator 

Name 

0008,1070  PN  User VNAP Operator field in Study list. 

Maximum 64 characters. 

Page 18: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Performing 

Physician 

Name 

0008,1050  PN  User VNAP Physician field in Study list. 

Maximum 64 characters. 

Scheduled 

Protocol Code 

Sequence 

0040,0008  SQ  MWL VNAP From Modality Worklist 

Instance 

Number 

0008,0013  IS  Auto ALWAYS Generated by device 

 

General Equipment Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Manufacturer  0008,0070  LO  Auto Always Endosoft systems

Institution 

Name 

0008,0080  LO  Auto VNAP Name of the Hospital. From 

Configuration 

Station Name  0008, 1010  SH  Auto Always Capture Station. From Configuration

Institutional 

Department 

Name 

0008,1040    Auto Always Name of the department at the 

Institution where the Equipment 

was installed. 

Manufacturer 

Model Name 

0008,1090  LO  Auto ANAP Model name given to Equipment by 

the manufacturer. From 

Configuration 

Software 

Version(s) 

0018,1020    Auto ANAP Software version of the Equipment 

that Generated the Digital Image. 

 

General Image Module 

Attribute 

Name 

Tag  VR  Source Presence of 

Value 

Default Value and Notes 

Instance 

Number 

0020,0013  IS  Auto Always Generated by device 

Patient 

Orientation 

0020,0020  CS  Auto EMPTY Set Length=0 if no value can be 

specified when an image is created. 

Page 19: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Image Date  0008,0023  DA  Auto Always Sets the date of start of image 

generation. 

Image Time  0008,0033  TM  Auto Always Sets the Time of start of image 

generation. 

Conversion 

Type 

0008,0064    Auto Always WSD

       

Image Type  0008,0008  CS  Auto VNAP

Acquisition 

Number 

0020,0012  IS  Auto Always Issues numbers sequentially in units 

of IPs read at the Image Reader. 

Generated by device 

Acquisition 

Date 

0008,0022  DA  Auto VNAP Sets date of start of IP reading. 

Acquisition 

Time 

0008,0032  TM  Auto VNAP Sets Time of start of IP reading. 

>Referenced 

SOP Class UID 

0008,1150  UI  Auto Always From referenced image, 

>Referenced 

SOP Instance 

UID 

0008,1155  UI  Auto Always From referenced image, 

Lossy Image 

Compression 

0028.2110      ANAP Clarifies whether image has been 

subjected to irreversible 

compression 

Pixel Spacing  0028,0030  DS  Auto ANAP Physical distance, within the patient, 

between the center’s of each pixel. 

Image 

Comment 

0020,4000  LT  User ANAP User‐defined comment on the 

image. From user input. Maximum 

1024 characters. 

 

 

 

 

16

Page 20: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

 

Image Pixel Module 

Attribute Name  Tag  VR Source Presence of 

Value 

Default Value and Notes 

Samples per Pixel  0028, 0002  US Auto Always 3 (Number of sample surfaces image 

has) 

Pixel Representation  0028, 0003  US Auto Always 0,1

Photometric 

Interpretation 

0028, 0004  CS  Auto Always RGB

Planar Configuration  0028, 0006    0

Rows  0028,0010  US  Auto Always Number of Rows in an image

Column  0028,0011  US  Auto Always Number of Column in an image, 

Bits allocated  0028, 0100  US  Auto Always 8

Bits stored  0028, 0101  US  Auto Always 8

High bits  0028, 0102  US  Auto Always 7

 

Encapsulated DICOM Module 

Attribute Name  Tag  VR Source Presence of 

Value 

Default Value and Notes 

Burned In Annotation  0028, 0301  CS  Auto Always NO

Document Title  0042, 0010  ST  Auto Always Report Title 

Concept Name Code 

Sequence 

0040, A043  SQ Auto Always Null

Encapsulated Document stream  

0042, 0011  OB Auto Always containing a document encoded                                                      according to the MIME Type  

MIME Type Encapsulated Document  

0042, 0012 

 

LO  Auto Always application/pdf  

 

 

 

 

 

17

Page 21: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Copyright © 1997‐2011 EndoSoft, LLC 

 

  

3.2.1.2 Error Handling 

EndoSoft systems are able to abort the association when a time out or an error occurs. 

 

If the C‐STORE operation is not successful, the image(s) are set to “Not Send” status in the database. A 

user‐configured number of additional attempt are made to store the image(s). If these attempts fail the 

user must select the job and press “Retry Job” to complete the C‐STORE operation. A complete log file 

gets created which a support team can analyze. 

 

3.2.2 SOP Specific Conformance for Modality Worklist 

 

Modality Worklist enables EndoSoft systems to query (using C‐FIND) for  Patient Demographics and Study 

Details from the MWL SCP, normally part of the PACS. The modality asks for a list of patients with chosen 

criteria via Standard C‐FIND operation and Modality Worklist responds accordingly.  

 

EndoSoft Systems Encodes the key listed in Table 5 in the query criteria. 

 

Table 5: Query Key Attributes in Modality Worklist Query 

Attribute Name  Tag  Matching Type

Scheduled Procedure 

Step 

Schedule Procedure Step 

Sequence 

(0040,0100)

  > Schedule Procedure Step 

Start Time 

(0040,0002)

  > Schedule Procedure Step 

Stop Time 

(0040,0003)

  > Modality (0008,0060)

Patient Identification  Patient’s Name (0010,0010)

  Patient ID  (0010,0020)

  Other Patient Ids (0010,1040)

Patient Demographics  Patient Birth Date (0010,0030)

  Patient Sex (0010,0040)

Imaging Service request  Accession Number (0008,0050)

  Referring Physicians Name (0008,0090)

Requested Procedure  Requesting Physician (0032,1032)

 

18 

         

Page 22: EndoSoft DICOM Conformance 2012utilize DICOM services to both receive and transfer DICOM data from other compliant DICOM entities over a network. Note that the format of this section

 

Copyright © 1997‐2011 EndoSoft, LLC 

3.2.2.1 Error Handling 

 

If an error occur when processing a Worklist response message, this message is skipped and EndoSoft 

system proceeds to the next message. 

 

4  Communication Profiles  

4.1  Supported Communications Stacks 

 

EndoSoft systems provide DICOM V3.0 TCP/IP Network Communication Support as defined in PS 3.8. 

 

4.2  TCP/IP Stack 

 

EndoSoft systems inherit its TCP/IP stack from the Windows system upon which it executes. 

 

4.3  Physical Media Support 

 

EndoSoft systems are indifferent to the physical medium over which TCP/IP executes; it inherits this from 

the Windows system upon which it executes. 

 

5  Extensions/Specializations/Privatizations  

Not Applicable. 

 

6  Configuration  

EndoSoft systems obtains configuration information from an ini  file located in its root directory name 

EndoLocl.ini as well as from the Incident table in the database and only system admin can configure it. 

 

7  Support of Extended Character Sets  

EndoSoft systems does not, itself, generate or use extended character sets, but if present in any DICOM 

instances received are preserved. 

 

8  Codes and Controlled Terminology  

No codes or controlled terminology are used by EndoSoft systems itself,  though such codes present in any 

DICOM instances received are preserved. 

19