EMMI LAB MANUAL_correct.pdf

download EMMI LAB MANUAL_correct.pdf

of 41

Transcript of EMMI LAB MANUAL_correct.pdf

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    1/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Exp. No:

    Date:

    CALIBRATION OF THREE PHASE ENERGYMETER

    AIM

    To calibrate the given three phase static energy meter at upf

    (i)  By direct loading.

    (ii)  By phantom loading

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Energy meter 1no.

    2. Wattmeter 2no.

    3. Voltmeter 1no.

    4. Ammeter 1no.

    5. 3φ autotransformer   1no.

    6. 3φ r esistive load 1no.

    7. Rheostat 1no.

    8. Stop Watch 1no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    2/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    CIRCUIT DIAGRAM

    (i)  Direct Loading

    Fig. 1

    (ii)  Phantom Loading

    Fig. 2

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    3/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    PRINCIPLE

    In order to check the calibration of a three phase energy meter, reading of energy meter has

    to be compared with that of standard instrument. For determining the true energy consumption a

    standard wattmeter and an accurate stop watch is used. From the calculated true energy, the error and

     percentage error in the energy meter is determined.

    (i)  Direct Loading

    In direct loading, current coil of energy meter and wattmeter are connected to a three phase

    supply in series with a loading device, where as in each phase pressure coils of energy meter are

    connected altogether to neutral wire. The pressure coils of two wattmeters are connected to Y phase.

    By adjusting the auto transformer take the reading of voltmeter which connected in between two phases

    as 415V. Supply is given to the circuit. Then by adjusting the loading device required current had got

    on the ammeters. Then energy consumption (measured power) is got by observing the time taken for

    3 revolutions of energy meter. True energy is calculated from wattmeter reading and time indicated inthe stop watch.

    Measured Power = ∗ ∗1000  Watts

    K = Energy meter constant in rev/kWh

     N = Number of revolutions made by energy meter disc

    t = Time for N revolutions of energy meter disc

    % error =( . . . ∗ 100)  , where M.P is measured power and T.P is true power.(ii)  Phantom loading

    Calibration is done by phantom loading of three phase energy meter. In phantom loading

    using standard wattmeter, the current coil is fed from a low voltage supply and pressure coil

     by rated voltage. Hence total power required for conducting the test is small.

    Measured Power = ∗ ∗1000  Watts

    K = Energy meter constant in rev/kWh

     N = Number of revolutions made by energy meter disc

    t = Time for N revolutions of energy meter disc

    % error =( . . . ∗ 100)  , where M.P is measured power and T.P is true power.

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    4/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    PROCEDURE

    (i)  Direct Loading

    Connections are made as shown in the fig (1). Autotransformer is kept at minimum position

    and the supply is switched on. Autotransformer is adjusted to apply rated voltage in the circuit. Then by adjusting the loading devices the required current readings are made on ammeters. Different

    Wattmeter readings and the time taken for 3 revolutions of the energy meter using a stopwatch are

    noted. The true reading is obtained from the wattmeter reading.

    (ii) Phantom Loading

    Connections are made as shown in fig (2). Autotransformer is kept at minimum position and

    the supply is switched on. Autotransformer is then varied and different currents are passed through

    the current coil of the energy meter. All the meter readings, wattmeter reading for all currents and

    time taken for three revolutions of the disc are noted.

    OBSERVATION

    (i)  Direct Loading

    Sl

    no.

    Voltmeter

    Reading

    (V)

    Ammeter

    Reading

    (A)

    W1 

    (W)

    W2 

    (W)

    T.P=

    W1+W2(W)

    Time for 3

    revolutions

    (sec)

    M.P

    (W)

    Error=M.P - T.P

    (W)

    % error=.−.

    . ×1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    5/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    (ii)  Phantom Loading

    l

    o.

    Vol

    tme

    ter

    Readin

    g

    (V)

    Am

    mete

    r

    Reading

    (A)

    W1 

    (W)

    W2 

    (W)

    T.P=

    W1+W2

    (W)

    Time

    for 3

    revol

    utions

    (sec)

    M.P

    (W)

    Error 

    =

    M.P-

    T.P(W)

    % error=

    .−.

    .×100

    Power

    consumed by

    Wattmeter

    Power

    consumed byEnergymeter

    Total

    Power

    =

    T.P+C.C+

    P.C(W)

    C.C (W)

    P.C (W)

    C.C (W)

    P.C (W)

    SAMPLE CALCULATION

    (i)  Direct Loading

    Time for 3 revolutions of energy meter disc = ------------------sec

    Energy meter constant, K= ---------------------rev/kWh

    Measured Power = ∗ ∗1000  Watts = -------------------W

    True Power = W1+W2 = ---------------W

    % error = ( . . . ∗ 100) 

    (ii) 

    Phantom Loading

    Time for 3 revolutions of energy meter disc = ------------------sec

    Energy meter constant, K= ---------------------rev/kWh

    Measured Power = ∗ ∗1000  Watts = -------------------W

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    6/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    True Power = W1+W2 = ---------------W

    % error = ( . . . ∗ 100)  = ----------------Power loss in wattmeter current coil = 2*I2R = --------------W

    Power loss in wattmeter pressure coil = 2*V2/R= ----------------W

    Power loss in energy meter current coil = 2*I2R = --------------W

    Power loss in energy meter pressure coil = 2*V2/R= ----------------W

    Total Power = T.P +C.C+P.C = --------------W

    Power Savings = True power obtained from direct loading- Total power consumed in

     phantom Loading

    = ------------W

    RESULT

    The given three phase energy meter is calibrated at upf by direct loading and phantom loading.

    INFERENCE

    Exp No:

    Date :

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    7/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    CALIBRATION OF SINGLE PHASE ENERGYMETER

    AIM

    To calibrate the given single phase static energy meter at upf by direct loading and phantom

    loading using standard wattmeter and draw the error and calibration curve.

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Energy meter 1no.

    2. Wattmeter 1no.

    3. Voltmeter 1no.

    4. Ammeter 1no.

    5. 1φ autotransformer   1no.

    6. Rheostat 1no.

    7. Resistive load 1no.

    8. Stop Watch 1no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    (i)  Direct Loading

    Fig. 1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    8/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    (ii) 

    Phantom Loading

     

    Fig. 2

    PRINCIPLE

    In order to check the calibration, single phase energy meter is compared with that of standard

    instrument. For determining the true energy consumption a standard wattmeter and an accurate stop

    watch is used. From the calculated true energy the error and percentage error in the energy meter is

    determined. The given energy meter can be calibrated for different loads by comparing the measured

    value of power obtained from it and the wattmeter reading. The wattmeter gives the true power.

    In direct loading, the load voltage appears across the pressure coil. Hence the energy meter

    and wattmeter reads the actual value of energy and power dissipated in the load.

    Phantom loading is employed for testing energy meter of high capacity. The phantom loading

    method is usually used for calibration of single phase or three phase energy meter. Here external loadis not connected and the current and pressure coils are supplied separately so that it will consume

    only less power. In this connection the voltage across pressure coil will be supply voltage even if the

    autotransformer is in minimum position and current coil is supplied from a separate low voltage

    supply.

    For checking the calibration of energy meter at upf, the current coils of energy meter and

    wattmeter are connected in series to the supply(R phase) through an autotransformer, which reduces

    the voltage to a low value. The pressure coils are connected directly to the supply to the same phase(R

     phase).

    Energy meter constant, K= 900 rev/KWh which means

    900 revolutions = 1 kWh

    = 1*1000*3600

    ie, one revolution = 1000∗3600 900   = 4000 W

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    9/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    For ‘N’ revolutions, power = 4000 N W 

    Time ‘t’ in second is required for N revolutions. 

    Power expected in one second = 4000  WTrue power = Wattmeter reading

    Measured Power = 4000 W% error =( . . . ∗ 100) 

    PROCEDURE

    (1) 

    Direct loading of energy meter at upf condition.

    Made the connections as shown in the fig(1). Supply is switched on and apply full voltage (230 V)

     by using autotransformer while keeping the load resistance in the circuit as minimum as possible. Switch

    on the load and the voltmeter, ammeter, wattmeter readings and time taken for 3 revolutions of the

    energy meter are noted. Experiment is repeated for various load currents up to rated value.

    (2)  Phantom loading of energy meter at upf condition.

    Keep the autotransformer in minimum position and switch on the supply. Keep the rheostat at

    constant value throughout the experiment. Give a voltage so that the ammeter readings do not exceed

    the rated current. Take readings for different current by adjusting the autotransformer. 

    OBSERVATIONS

    (1) Direct loading

    Sl

     No.

    Voltmeter

    Reading

    (V)

    Ammeter

    Reading

    (A)

    True

    Power

    (W)

    Time for 3revolutions

    (Sec)

    Measured

    Power

    (W)

    Error=

    M.P-T.P

    (W)

    % error=

    .−.. ×100

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    10/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    (2) Phantom loading

    Sl

     No.

    Volt

    meter

    Read

    ing(V)

    Ammet

    er

    Readin

    g(A)

    True

    Power 

    (W)

    Time

    for 3revoluti

    ons(Sec)

    M.P

    (W)

    Error=

    M.P-

    T.P

    (W)

    % error=.−.

    . ×100Power

    consumed by

    wattmeter

    Power

    consumed by

    energymeter

    Total

     power (W)=

    T.P+C.C+P.C

    C.C

    (W)

    P.

    C

    (W)

    C.C

    (W)

    P.C

    (W)

    SAMPLE CALCULATION

    (i) 

    Direct Loading

    True Power = Wattmeter reading = ---------------------W

    K= 900 rev/KWh

     Number of revolution of energy meter disc, N = 3

    Time for 3 revolutions of energy meter disc, t = --------------sec

    Measured power =∗∗

    ∗  = -------------------W

    % error = ( . . . ∗ 100) =  (ii)  Phantom Loading

    True Power = Wattmeter reading = ---------------------W

    K= 900 rev/KWh

     Number of revolution of energy meter disc, N = 3

    Time for 3 revolutions of energy meter disc, t = --------------sec

    Measured power =∗∗

    ∗  = -------------------W

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    11/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    % error = ( . . . ∗ 100) =  Power loss in wattmeter current coil = I2R = --------------W

    Power loss in wattmeter pressure coil = V2/R= ----------------W

    Power loss in energy meter current coil = I2R = --------------W

    Power loss in energy meter pressure coil = V2/R= ----------------W

    Total Power = T.P +C.C+P.C = --------------W

    Power Savings = True power obtained from direct loading- Total power consumed in

     phantom Loading

    = ------------W

    RESULT

    The single phase energy meter is calibrated at upf by direct and phantom loading.

    INFERENCE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    12/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Exp No:

    Date :

    CALIBRATION OF SINGLE PHASE ENERGYMETER AT 0.5 pf

    AIM

    To calibrate the given single phase static energy meter at 0.5 pf lag and lead by phantom

    loading.

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Energy meter 1no.

    2. Wattmeter 1no.

    3. Voltmeter 1no.

    4. Ammeter 1no.

    5. 1φ autotransformer   1no.

    6. Rheostat 1no.

    7 Stop Watch 1no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM0.5 pf lag

     

    Fig. 1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    13/41

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    14/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    (i) 0.5 pf lag

    Connections are made as shown in the fig (1). Supply is switched on and apply full voltage (230 V)

     by using autotransformer while keeping the load resistance in the circuit as minimum as possible. Switch

    on the load and the voltmeter, ammeter, wattmeter readings and time taken for 3 revolutions of the

    energy meter are noted. Experiment is repeated for various load currents up to rated value.

    (iii)  0.5 pf lead

    Connections are made as shown in the fig (2). Supply is switched on and apply full voltage (230

    V) by using autotransformer while keeping the load resistance in the circuit as minimum as possible.

    Switch on the load and the voltmeter, ammeter, wattmeter readings and time taken for 3 revolutions

    of the energy meter are noted. Experiment is repeated for various load currents up to rated value.

    OBSERVATIONS

    (i) 0.5 pf lag

    Sl

     No.

    Voltmeter

    Reading

    (V)

    Ammeter

    Reading

    (A)

    True

    Power

    (W)

    Time for 3

    revolutions

    (Sec)

    Measured

    Power

    (W)

    Error=

    M.P-T.P

    (W)

    % error=

    .−.. ×100

    (ii) 0.5 pf lead

    Sl

     No.

    Voltmeter

    Reading

    (V)

    Ammeter

    Reading

    (A)

    True

    Power

    (W)

    Time for 3

    revolutions(Sec)

    Measured

    Power

    (W)

    Error=

    M.P-T.P

    (W)

    % error=

    .−.. ×100

    SAMPLE CALCULATION

    (i)  0.5 pf lag

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    15/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    True Power = Wattmeter reading = ---------------------W

    K= 900 rev/KWh

     Number of revolution of energy meter disc, N = 3

    Time for 3 revolutions of energy meter disc, t = --------------sec

    Measured power =∗∗

    ∗  = -------------------W

    % error = ( . . . ∗ 100) =  (ii) 0.5 pf lead

    True Power = Wattmeter reading = ---------------------W

    K= 900 rev/KWh

     Number of revolution of energy meter disc, N = 3

    Time for 3 revolutions of energy meter disc, t = --------------sec

    Measured power =∗∗

    ∗  = -------------------W

    % error = ( . . . ∗ 100) =  RESULT

    The single phase energy meter is calibrated at 0.5 pf lag and lead by phantom loading.

    INFERENCE

    Exp No:

    Date :

    CALIBRATION OF SINGLE PHASE ENERGYMETER AT 0.866 pf

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    16/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    AIM

    To calibrate the given single phase static energy meter at 0.866 pf lag and lead using phase

    shifting transformer.

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Energy meter 1no.

    2. Wattmeter 1no.

    3. Voltmeter 1no.

    4. Ammeter 1no.

    5. 1φ autotransformer   1no.

    6. Rheostat 1no.

    7 Stop Watch 1no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    (i)  0.866 pf lag

    ` Fig.1

    (i)  0.866 pf lead

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    17/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Fig. 2

    Exp. No: 

    Date:

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    18/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    MEASUREMENT OF SELF AND MUTUAL INDUCTANCE

    AIM:-

    To determine self inductance, mutual inductance and coefficient of coupling of coupled coil.

    APPARATUS REQUIRED

    SL NO APPARATUS SPECIFICATION QUANTITY

    1. Iron cored coils (Transformer 1φ)  1no.

    Voltmeter 1no.

    3. Voltmeter 1no.

    4. Voltmeter 1no.

    5. Ammeter 1no.

    6. Ammeter 1no.

    7. Autotransformer 1no.

    8. DC regulated power supply 1no.

    9. Rheostat 1 no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    Additive Polarity

    Fig. 1

    Subtractive Polarity

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    19/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Fig. 2

    To measure resistance of Primary coil

    Fig. 3

    To measure resistance of Secondary coil

    Fig. 4 

    PRINCIPLE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    20/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Inductance is the property of a circuit element by which energy is capable of being stored in

    a magnetic flux field and any circuit element exhibit the property of inductance is called an inductor.

    Self Inductance of a coil is the property by which it opposes any flux through it. Mutual

    inductance of a coil is the ability to produce an EMF in the neighboring coil by induction, when the

    current in the first coil changes.

    Consider two magnetically coupled coils of self inductance  and . Let M be the mutualinductance of the coils connected in series so that flux is produced by current I through the coils are

    in the same direction, then the effective inductance

     =  +  + 2MIf coils are connected such that the flux produced by the current in opposite direction, then

    effective inductance

     =  + - 2MTherefore mutual inductance M= (  - ) / 4Coupling Coefficient k = M / √    In the first case, if and  are the applied voltage and current, then =  /   ,   = √    , =  2⁄ , where  is the DC resistance of primarycoil.

    Similarly for the second case

    =  /   ,   =     , =  2⁄   , where  is the DC resistance of secondarycoil.

     /  =  / )2 From the above equations  ,  , M and k can be found out. The experimental determinations ofthe above parameters are carried out for a pair of transformer winding.

    PROCEDURE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    21/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    For determining the inductance of transformer coil, connections are made as shown in figure

    (1) when the windings are connected in series for additive polarity. Supply is switched on and the

    rated voltage is applied in the circuit by adjusting the auto transformer. The Voltmeter and Ammeter

    readings are noted. Then the terminals of the coils are interchanged as shown in the figure (2) for

    subtractive polarity. Auto transformer is adjusted to give a voltage to the primary and all other

    readings are taken.(It is desirable to keep V2 constant for both aiding and opposing).

    To measure the resistance of the coils, connections are made as shown as figure (3) and (4)

    and dc supply is switched on. The rheostat is adjusted for different values of Ammeter and Voltmeter

    reading. R = ⁄  After obtaining the voltages and currents, inductance  , , M and coupling coefficient k

    for a pair of transformer windings are determined.

    OBSERVATION

    Aiding Circuit

     Z  A(Ω) 

    Opposing Circuit

     Z  B(Ω) 

    V(V)  V 1(V) V 2(V) I(A) V(V) V 1(V) V 2(V) I(A)

    Resistance Measurement

    Transformer PrimaryWinding

    Transformer SecondaryWinding

    V(V)   I(A) R P (Ω)  V(V) I(A) RS (Ω) 

    Sample Calculation

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    22/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    =  /   = -------------   = √    = ------------- =  2⁄ = ------------- =  /  = -------------

      =     = ------------- =  2⁄  = -------------M = (  - ) / 4 = -------------

     =  +  + 2M = -------------

     =  + - 2M = ------------- + = 2 ( +  = -------------( +  = (  +  ) / 2 = -------------

     /  =  / )2 =  / )2 From the above equations the values of  ,  , M and k are calculated.

    RESULT

    Self inductance, Mutual inductance and Coefficient of coupling for a pair of transformer

    windings are determined.

    Self inductance of coil 1,   = -------------Self inductance of coil 2 ,  = -------------Mutual inductance , M = -------------

    Coefficient of Coupling, k = ---------------

    INFERENCE

    Expt No:

    Date:

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    23/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    MEASUREMENT OF B-H CURVE USING CRO

    AIM 

    To study the hysteresis loop of a given specimen

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Hysteresis loop module ITB-026A 1

    2. CRO 1

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    PRINCIPLE

    Magnetic field is a phenomenon where under certain conditions, energy or force transfer canoccur through space. it can be established only by its effective which is used to determine the

    magnetic property of the materials.

    Hysteresis loop is nothing but a plot of flux density ‘B’ versus magnetizing for ‘H’. Many

    other parameters can be determined from this loop. The hysteresis in any process is the

    nonconformity of the loading and unloading curve of the process. The reason for occurrence of

    hysteresis is that of all energy that has been pumped in to the system during the loading period, is

    not being recovered completely due to losses in the system.

    The a-b-c-d-e-f-a curve is called hysteresis curve for the magnetic material.

    BR  is the residual flux density. This is what enables the creation of permanent magnets. The

    magnetic force HD is required coerce the material to reduce its flux density level to zero is called

    coercive force. The unit for magnetic force ‘H’ is ampere turn per meter At/m. flux density B is

    called Tesla (Wb/m2) or gauss. One gauss is 10,000 gauss

    Thus the hysteresis loop is often called B-H curve. The understanding of B-H curve is

    extremely importance in design of transformer, chokes, coils and inductors

    PROCEDURE

    (iii) 

    Connect the variable power supply to the input terminals.

    (iv)  Connect the X input of the CRO to the terminal T3

    (v)  Connect the Y input of the CRO to the terminal T6

    (vi)  Keep the CRO in XY mode

    (vii)  Hysteresis loop appears as in figure

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    24/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    (viii)  Vary the input AC voltage and calculate VX , VY and tabulate the readings.

    (ix)  Tabulate the VX  and VY from CRO

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    25/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    OBSERVATION

    SL.NO. VX VY I1 H B

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    26/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    FORMULAE

    VX = R 1 * I1 VX=0.22 * I1

    I1= VX/0.22

    H=N1I1 / L 

    VY= (1/C) (N2AS/R 2) B

    B= VYCR 2/ N2AS

    Where

    R 1= 0.22 ohms (from manufacturer)

    L= length of magnetic path= 0.1139m

     N1= no. of turns in primary=2475

     N2= no. of turns in secondary+102

    As= area of the specimen+0.00045m2

    R 2= 4700Ω (from manufacturer)

    C= 10μF (from manufacturer)

    RESULT

    Thus the graph is plotted between B and H

    INFERENCE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    27/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    CIRCUIT DIAGRAM

    a) 

    Calibration of voltmeter

    Fig (1)

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    28/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Expt No:

    Date:

    VERNIER POTENTIOMETER

    AIM 

    To calibrate the given voltmeter, ammeter and wattmeter using Vernier potentiometer and

    hence draw the calibration curve.

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Vernier potentiometer

    2. STD cell

    3. Fixed 2V supply

    4. DC regulated power supply

    5. Voltmeter

    6. Galvanometer

    7. Volt ratio box

    8. Standard Resistance

    9. Rheostat

    10. Ammeter

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    PRINCIPLE

    The potentiometer is an instrument used for measurement of an unknown EMF or potential difference

     by balancing it, wholly or partially by a known potential difference produced by the flow of a known

    current in a network or circuit of known characteristics. Potentiometers are extensively used in

    measurements where the precision required is higher than that can be obtained by ordinary deflection

    instruments. EMFS are measured directly with a potentiometer in terms of the EMF of a standard

    cell. By using, in addition, a standard resistor, current can also be measured. From the current and

    voltage measurements, power can be calculated. The 

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    29/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    b) Calibration of ammeter

    c) 

    Calibration of Wattmeter

    Fig (2)

    Fig (3)

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    30/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

     potentiometer is widely employed for calibration of voltmeters, ammeters and wattmeters.

    The potentiometer works on the principle of opposing the unknown EMF by a known EMF with the

     – VE terminals of two EMFS connected together and also the +VE terminals connected together

    through a galvanometer. Galvanometer gives no deflection if two EMFS are equal.

    PROCEDURE

    a)  Standardization

    The potentiometer is energized by giving a 2V dc supply to the terminals given in fig. The

    selector switch is thrown to position ‘STD’ to include standard cell in the circuit. Range selector key

    should be in position X0.1. In position X0.1, all dial readings should be multiplied by a factor 0.1 in

    addition to factors shown near dials. The dials are adjusted to read standard cell voltage i.e. if standard

    cell voltage is 1.0186, the first dial should read 10; the second dial 18 and third dial 60. The

    galvanometer is connected to the terminal pair marked as ‘GALV’. The galvanometer key is pressed

    and observed the galvanometer. The coarse, medium and fine rheostats are adjusted until the

    galvanometer gets null deflection.

     b) 

    Calibration of voltmeter

    The voltmeter to be calibrated is connected across the variable dc source. The voltmeter is

    connected to the high voltage side of volt ratio box. The low voltage side of volt ratio box is connected

    to the potentiometer at appropriate terminals. The selector switch is thr own to ‘TEST’ position and

    voltage at output of volt ratio box is measured using potentiometer. Then the true value of voltage

    across voltmeter can be determined by multiplying the voltage at low voltage side of the volt ratio

     box by multiplication factor. The voltmeter reading is noted. The ratio of difference between

    measured value and true value is the percentage error. The errors at various voltmeter readings are

    calculated.

    % Error = −   × 100

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    31/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    OBSERVATIONS

    a)  Calibration of voltmeter

    Sl

     No.

    Volts 

    X1 ×0.1 X2 ×0.001 X3 ×0.00001 V(V) −   × 100

     b) 

    Calibration of ammeter

    Sl

     No.

    Current I  X1×0.1 X2×0.001 X3×0.00001 V I =/  −   × 100

    c)  Calibration of Wattmeter

    Sl

    No  V Wind  X1 

    ×0.1

    X2 

    ×0.001

    X3 

    ×0.00001E1 

    X1 

    ×0.1

    X2 

    ×0.001

    X3 

    ×0.00001E2

    I= 2/ 

    Wact 

    W−WW× 100

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    32/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    a) 

    Calibration of ammeter

    An Ammeter to be calibrated is connected in series with a standard resistance R s of suitable value.

    The current supplied by a D.C. supply passes through ammeter as well as standard resistance R s.

    using D.C. potentiometer; a voltage across standard resistance can be measured. Thus, current

    flowing through R s (and ammeter also) is given by,

    I =R  ; where Vs = voltage across Rs measured using potentiometer

    Rs = resistance of standard resistor  

    Since the resistance of standard resistance is known accurately and also the voltage across Rs is

    measured using standardized potentiometer, the method of calibrating ammeter is very accurate. If

    the actual current Iact  and the current indicat5ed by ammeter Iind  are not matching, error is

    indicated. The percentage error is given by

    % error =−

      × 100 ; where Iind  = current indicated by ammeter and Iact  = I =R 

    The calibration curve of ammeter can be obtained by plotting % error against the reading of

    ammeter i.e. Iind.

    Calibration of Wattmeter

    For the calibration of a Wattmeter, a low voltage supply is given to the current coil (CC)

    whose current can be adjusted by using a rheostat R h in series with low voltage supply and a high

    voltage supply is given across the potential coil (PC). The voltage is stepped down by volt-ratio box.

    A voltmeter measures voltage V and ammeter measures current I which gives true power as,

    Wind = V I

    This value can be compared with a value indicated by watt meter. If two values are not matching, a

     positive or negative error is indicated which is given by,

    % error =W−W

    W  × 100

    RESULT

    The given voltmeter, ammeter and wattmeter were calibrated the using vernier potentiometer

    and hence the calibration curve were drawn.

    INFERENCE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    33/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Expt.No.

    Date:

    WHEATSTONE BRIDGE AND KELVIN’S DOUBLE BRIDGE 

    AIM

    a)  To use wheatstone bridge for the measurement of medium resistance

     b)  To use Kelvin Double bridge for the measurement of low resistance.

    APPARATUS REQUIRED

    SL No APPARATUS SPECIFICATION QUANTITY

    1. Wheatstone Bridge 1 no.

    2. Voltmeter 1 no.

    3. Voltmeter 1 no.

    4. Rheostat 1 no.

    5. Kelvin Double Bridge 1 no.

    6. Galvanometer 1 no.

    7. DC regulated power supply 1 no.

    8. Ammeter 1 no.

    9. Ammeter 1 no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    Wheatstone Bridge

    Fig.1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    34/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Fig.2

    PRINCIPLE

    a)  Wheatstone Bridge

    This is the  best and common method of measuring medium resistances (1Ω to 0.1MΩ).Majority of the electrical apparatus and equipments have resistances within these limits and hence

    the wheat stone bridge is a very useful instrument for resistance measurement.

    The General circuit arrangement is given in the figure (1), where P and Q are two known fixed

    resistances, S being a known variable resistance and R be the unknown resistance (i.e. voltmeter

    or rheostat). At balanced condition, no current flows through the galvanometer, and the unknown

    resistance R is given by R =  .The arms P and Q are the ratio arms of decade dial is provided in the portable bridge

    for this purpose. S is the known standard resistance whose values may be varied. Decade dials in

    different ranges are provided in the portable bridges. The product of the range selector (   ) andthe total value of variable resistance (S) gives the unknown resistance R.

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    35/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Kelvin Double bridge

    Fig .4

    Fig .3

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    36/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Kelvin Double Bridge

    This method is the best available for the precise measurement of low resistances (less than

    1Ω). It is a development of the wheat stone bridge by which the errors due to contact and lead

    resistance are eliminated. Kelvin Double Bridge is suitable for measuring resistance fitted withfour terminals i.e. two current terminals C1 and C2 and two potential terminals P1 and P2. This is

    to reduce contact and lead resistances completely. It incorporates the idea of a second set of ratio

    arms. Hence the name double bridge and the use of four terminals.

    Figure (2) shows the schematic diagram of Kelvin’s double bridge. The first set of ratio arms

    is Q and M. The second set of ratio arms q and m is used to connect the galvanometer to a point

    at the appropriate potential between Q and M to eliminate the effect of connecting lead resistance

     between the known resistance R and standard resistance S. In the figure (2), R is the low resistance

    to be measured (i.e. resistance of ammeter, given length of wire etc.), S is the standard variable

    resistance and Q, M, q and m are four non inductive resistances, one pair of each (M and m Or

    Q and q) are variable. These are connected to two sets of ratio arms which are used for range

    selection. The ratio ⁄  is kept same as  ⁄  under balanced condition. These ratios along withS being varied until zero deflection of the galvanometer are obtained. Then ⁄  = ⁄  = ⁄  or R = ⁄  × S. So the product of ratio arms ⁄  and variable resistance S gives the unknownresistances.

    PROCEDURE

    a)  Wheatstone bridge circuit is shown in figure ( 1 ). Initially, set the four knobs of the decade

    resistance box at zero position. Then adjust the galvanometer pointer to zero by using ‘SET

    ZERO’ knob after checking the position of key 10, which should be in the normal raised position

    and connect the given voltmeter to the external terminal ( 3 and 4) of the bridge.[Ensure that the

    terminals 1 and 2 and also 7 and 8 are shorted.]. The range selector (  ) is properly selected andresistance S (i.e. by varying the four decade resistances) is varied and the galvanometer is

    checked. This is continued until the balanced condition of the bridge is obtained. The readings

    of the range selector and the four dials of the variable resistance S  are noted. The experiment is

    repeated for different values of range selector.

    Kelvin Double bridge circuit is shown in figure ( 3 ). Provision has been given in

    the bridge for current terminals (C and C1) and potential terminals (P and P1) separately. But

    when connecting the unknown resistance having two terminals (like ammeters), they are

    connected to only C and C1 while C to P and C1 to P1 are short circuited. An external dc power

    supply (at ‘CURRENT INPUT’  terminal) and a galvanometer (between G1 and G2) are

    connected. After setting the CURRENT SWITCH to NOMAL and adjusting the current to the

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    37/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    required value, select a proper value for range multiplier. Then vary the resistance S by adjusting

    the decade dials. This is continued until the balanced condition of the bridge is obtained. The

    readings of the range selector and the four dials of the variable resistance S are noted. The

    experiment is repeated for the REVERSE CURRENT SWITCH condition also. The mean of the

    two readings should be taken as the correct value.

    OBSERVATIONS

    a)  For Wheatstone bridge

    Sl

     No.

      S1 × 1000 S2 × 100 S3 ×10 S4 × 1 R3 = S1+S2+S3+S4 R =  

     

    b) For Kelvin Double bridge

    Sl No.

    Remarks ⁄   S1 S2 S = S1 + S2 R = ⁄  × S

    RESULT

    1.  Resistance of Voltmeter =

    2.  Resistance of Ammeter =

    INFERENCE

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    38/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Expt.No.

    Date:

    B - H CURVE USING SINGLE PHASE TRANSFORMER

    AIM

    To plot the B- H characteristics of a transformer core.

    APPARATUS REQUIRED

    SL No APPARATUS SPECIFICATION QUANTITY

    1. Autotransformer 1 no.

    2. Voltmeter 1 no.

    3. Ammeter 1 no.

    4. Wattmeter 1 no.

    5. Transfomer 1 no.

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    Fig. 1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    39/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    Expt.No.

    Date:

    DETERMINATION OF HYSTERISIS LOOP

    AIM

    To trace the hysteresis loop using CRO.

    APPARATUS REQUIRED

    SL NO. APPARATUS SPECIFICATION QUANTITY

    1. Hysteresis loop module ITB-026A 1

    2. CRO 1

    < Students are expected to calculate instrument ranges based on the machine ratings before start of

    experiment and get it approved before connections are made>

    CIRCUIT DIAGRAM

    Fig. 1

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    40/41

     

    Dept. of Electrical & Electronics Engg. College of Engineeri ng, Kidangoor

    PRINCIPLE

    Magnetic field is a phenomenon where under certain conditions, energy or force transfer can

    occur through space. it can be established only by its effective which is used to determine the

    magnetic property of the materials.

    Hysteresis loop is nothing but a plot of flux density ‘B’ versus magnetizing for ‘H’. Many

    other parameters can be determined from this loop. The hysteresis in any process is the

    nonconformity of the loading and unloading curve of the process. The reason for occurrence of

    hysteresis is that of all energy that has been pumped in to the system during the loading period, is

    not being recovered completely due to losses in the system.

    The a-b-c-d-e-f-a curve is called hysteresis curve for the magnetic material.

    BR  is the residual flux density. This is what enables the creation of permanent magnets. The

    magnetic force HD is required coerce the material to reduce its flux density level to zero is called

    coercive force. The unit for magnetic force ‘H’ is ampere turn per meter At/m. flux density B is

    called Tesla (Wb/m2) or gauss. One gauss is 10,000 gauss

    Thus the hysteresis loop is often called B-H curve. The understanding of B-H curve is

    extremely importance in design of transformer, chokes, coils and inductors

    PROCEDURE

    (i)  Connect the variable power supply to the input terminals.

    (ii)  Connect the X input of the CRO to the terminal T3

    (iii) 

    Connect the Y input of the CRO to the terminal T6(iv)  Keep the CRO in XY mode

    (v)  Hysteresis loop appears as in figure

    (vi)  Vary the input AC voltage and calculate VX , VY and tabulate the readings.

    (vii)  Tabulate the VX  and VY from CRO

  • 8/18/2019 EMMI LAB MANUAL_correct.pdf

    41/41