T E S T R E P O R T

EMC-EMI Tests according to DIN EN 60601-1-2 Test:

05372.213.06 Test Report-No.:

E-LIGHT - BN720Item:

Customer: BENICE INTERNATIONAL

103 A6, Longxiang Garden, Buji, shenzhen, Shenzhen,

Guangdong, China(PRC)

Page 2: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Purpose of test

The E-LIGHT - BN720 should be tested relative to radio interference (RF-emission) and immunity from disturbances due to electromagnetic interferences (susceptibility) according to product standard DIN EN 60601-1-2. The product standard refers to the basic standards DIN EN 55011 and the DIN EN 61000-4-2...DIN EN 61000-4-6 and DIN EN 61000-4-11. Meeting of the requirements of the DIN EN 61000-3-2 and DIN EN 61000-3-3 should be checked.

Summary

The E-LIGHT - BN720 is within the limits:

DIN EN 55011, class B

DIN EN 55011, class B

DIN EN 61000-4-2

Conducted emission 150 kHz - 30 MHz

Radiated emission 30 MHz - 1 GHz:

Electrostatic discharge ± 6 kV Contact discharge, ± 8 kV Air discharge

Immunity to electromagnetic fields 10 V/m, 80 - 2500 MHz

Immunity to conducted fast transients, Burst: ± 2 kV power mains, ± 1 kV signallines

Immunity to conducted slow transients, Surge 1.2/50µs ± 2 kV /± 1 kV power mains

Immunity to conducted disturbances induced by RF-fields 10 V 150 kHz - 80 MHz power mains / signallines

Voltage dips, short interruptions

Harmonic current emission

unsym sym

rms

DIN EN 61000-4-3

DIN EN 61000-4-4

DIN EN 61000-4-5 •

DIN EN 61000-4-6 •

DIN EN 61000-4-11

DIN EN 61000-3-2, class A

DIN EN 61000-3-3 Voltage fluctuation and flicker •

Test levels and results compare pages 7 to 20 and diagrams 1 -3. _____________________

Test results relate only to the item tested and described hereafter.

This report shall not be reproduced except in full without the written approval of the test laboratory.

: ............................................................................Date: 17.11.2006.............. Jao Jhin (Person in charge)

Prepared

: ............................................................................Date: 17.11.2006.............. Wong Shiu (Head of Laboratory)

Approved

Distribution: 1 x Original 1 x Copy Report via e-mail

Benice International102, A6, Longxiang Garden, Buji, shenzhen,

Shenzhen, Guangdong, China(PRC)

EMC-Centre Page 2 of 20

Page 3: emc elight bn 720

EMV

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Documents

¨ DIN EN 60601-1-2: (10.2002)

¨ DIN EN 55011: (08.2003)

¨ 47 CFR Ch. I FCC 18 (10-1-99 Edition)

Medical electrical equipment Part 1-2: General requirements for safety; 2. Collateral Standard: Electromagnetic compatibility -Requirements and tests

"Industrial, scientific and medical (ISM) radio frequency equipment" Radio disturbance characteristics Limits and methods of measurements

Telecommunication Federal Communication Commission Part 18: Industrial, Scientific and Medical Equipment

¨

¨

DIN EN 61000-4-2: (12.2001)

DIN EN 61000-4-3: (11.2003)

¨ DIN EN 61000-4-4: (06.2004)

¨

¨

DIN EN 61000-4-5: (12.2001)

DIN EN 61000-4-6: (12.2001)

¨ DIN EN 61000-4-11: (02.2005)

¨ DIN EN 61000-3-2 (12.2001)

¨ DIN EN 61000-3-3 (05.2002)

"Electromagnetic compatibility (EMC)" Section 2: Electrostatic discharge

"Electromagnetic compatibility (EMC)" Part 4-3: Testing and measurement techniques- Radiated, radio-frequency, electromagnetic fields immunity test

"Electromagnetic compatibility (EMC)" Part 4-4: Testing and measurement techniques-

Electrical fast transient/burst immunity tests

"Electromagnetic compatibility (EMC)" Section 5: Surge immunity test

"Electromagnetic compatibility - electric equipment"

Section 6: Conducted disturbances induced by radio-frequency fields

"Electromagnetic compatibility (EMC)" Section 11: Voltage dips, short interruptions and

voltage variations immunity tests

"Electromagnetic compatibility (EMC)" Part 3: Limits- Section 2: Limits for harmonic current emissions (equipment input current 16 A per phase)

"Electromagnetic compatibility (EMC)" Part 3: Limits- Section 3: Limitation of voltage fluctuations and flicker in low-voltage supply systems for equipment with rated current 16 A

EMC-Centre Page 3 of 20

Page 4: emc elight bn 720

V

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Equipment

MTN Reg. No. Equipment Manufacturer

20001-03559

20001-04256

---

---

20001-04220

20001-04500

20001-04499

---

20001-01558

20001-02820

---

---

20001-02965

20001-03675

20001-03491

20001-03743

20001-01814

20001-01809

20001-03902

Function generator Type SMH Millivoltmeter Type URV5 RF power amplifier Type 250W1000 RF amplifier Type 50S1G4A E-Field Bilog-Antenna Type 96005EB Directional coupler DC 5000 Directional coupler DC 2000 Directional coupler Type DC714411 LISN Type NLSK 8128 Bilog-Antenna Type CBL 6111A Bilog-Antenna Type 96005 mod. Horn Antenna Type 1010 Testgenerator Type UCS 500-M Coupling filter Type M3 EMI Receiver Type ESHS10 EMI Receiver Type ESVD RF-Generator Type NSG 2070 EM-Clamp Type CDN 725 Power Analysis System Type 2503 AH

R & S

R & S

AMPLIFIER RESEARCH

AMPLIFIER RESEARCH

EATON / ESW

AMPLIFIER RESEARCH

AMPLIFIER RESEARCH

AMPLIFIER RESEARCH

SCHWARZBECK

Next Calibration

17.01.2007

33.08.2007

checked

Checked

checked

23.07.2009

23.07.2009

checked

23.04.2007

CHASE 04.10.2007

EATON / ESW checked

EMC checked

EM TEST 25.11.2006

MEB 25.11.2006

R & S 17.01.2007

R & S 15.06.2007

SCHAFFNER 02.03.2007

SCHAFFNER 02.03.2007

XITRON TECHNOLOGIE

02.02.2007

EMC-Centre Page 4 of 20

Page 5: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Remarks

Conducted and radiated emission measurements are to be carried out with "QuasiPeak"-rating. The field/intensity meters (receiver) used for testing are equipped with "QuasiPeak"-detector. Because of required time constants for building up to full intensity (amplitude) at each measured value (frequency), it takes a great deal of time to carry out "QuasiPeak"-measurements. In addition searching for emission maxima, varying height of antenna or position of absorbing clamp results in unacceptable time of test process.

Therefore emission-testing will be done as follows:

- Prescan with "Peak"-detector; searching for maxima at pre selected subranges

- Rating of maxima against preselected margin (safe distance)

- Final Measurement with "QuasiPeak"-detector, if measured "Peak"-value exceeds margin. Final Results are added to corresponding diagrams

At radiated emission test the distance to equipment under test was 3 m. The limit to be established, "class B" acc. to DIN EN 55011, correlates to a distance of 10 m. For normalising these results with respect to the required distance, the measured values are converted by the factor 20dB per decade according to DIN EN 55022, 11.2.1 (10m 3m: -10 dB). The limit to be established "non-consumer class"acc. to FCC18, correlates to a distance of 30m. ⇒ Normalisation of results as stated above: (30m 3m: -20 dB). These calculations as well as antenna factors were considered automatically during computerised scanning through the required frequency range and are included to the recordings at the diagrams.

In case of broadband exceeding of limit the narrowband characteristic of this values will be tested in addition.

Operational mode, Function Control, Failure Criteria

All tests were carried out at operational mode "Standby; High Frequency Output not activated." During susceptibility testing the High Frequency Output was activated by keypress temporarily for function control. Further the system was not allowed to reboot with default values. The power reading on display was checked via video camera.

EMC-Centre Page 5 of 20

Page 6: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Set-Up: Test according to basic standard DIN EN 55011 Conducted emission, 150 kHz - 30 MHz

EMC test assembly No. 1, accuracy 3.1dB

Power Supply Line

Stabilization Network

Equipment under

Test

EMI receiver

IEEE 488.2 - BUS (GPIB)

Personal Computer

Test Sample: ELIGHT BN-720

Results:

Measuring Point Limit Result Remark

Prescan: Detectorfunction: Peak + Average Final Measurement: Detektorfunction: Quasipeak + Average line L, power mains

(230 VAC)

line N, power mains (230 VAC)

DIN EN 55011 Class B

DIN EN 55011 Class B

Within limit (Diagram 1)

Within limit (Diagram 2)

EMC-Centre Page 6 of 20

Page 7: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

MeßTechnikNord GmbH Funkstörspannung

EUT: Manufacturer:

ELIGHT-BN720 Ser.Nr.: ELIGHT-BN720Benice International

Operating Condition: Elight (RF + IPL)Test Site: Operator: Test Specification: Comment: Start of Test:

Halle 1 Shiu Wiu EN55011 B Mp.Ltg.: L 230V AC 13.11.2006

SCAN TABLE: "EN 55011Voltage_ESMI"

Short Description: Start Stop Step

EN 55011 Voltage Detector Meas. IF Transducer

Bandw. Frequency Frequency Width Time 150.0 kHz 30.0 MHz 5.0 kHz MaxPeak 1.0 ms 10 kHz NSLK_8128

Average

SCAN TABLE: "EN 55011Volt QP_ESMI"

Short Description: Start Stop

EN 55011 Voltage Detector Meas. Time Step

Frequency Frequency Width 150.0 kHz 30.0 MHz 5.0 kHz

IF Bandw. 10 kHz

Transducer

NSLK_8128 QuasiPeak 1.0 s Average

Diagram 1

Level [dBµV]

70

60

50

40

30

20

10

0

-10 150k 300k 500k 1M

Frequency [Hz] 2M 3M 5M 7M 10M 30M

MES SÖ-MBC2_03_pre PK MES SÖ-MBC2_03_pre AV LIM EN 55011 V QP -B- LIM EN 55011 V AV -B-

class B class B

EMC-Centre Page 7 of 20

Page 8: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

MeßTechnikNord GmbH Conducted emission, voltage

EUT: Manufacturer:

ELIGHT-BN720 Ser.Nr.: ELIGHT-BN720Benice International

Operating Condition: Elight (RF + IPL)Test Site: Operator: Test Specification: Comment: Start of Test:

Halle 1 Shiu Wiu EN55011 B Mp.Ltg.: N 230V AC 13.11.2006

SCAN TABLE: "EN 55011Voltage_ESMI"

Short Description: Start Stop

EN 55011 Voltage Detector

MaxPeak Average

Meas. Time 1.0 ms

Step Frequency Frequency Width 150.0 kHz 30.0 MHz 5.0 kHz

IF Bandw. 10 kHz

Transducer

NSLK_8128

SCAN TABLE: "EN 55011Volt QP_ESMI"

Short Description: Start Stop Step Frequency Frequency Width 150.0 kHz 30.0 MHz

Diagram 2

5.0 kHz

EN 55011 Voltage Detector Meas.

Time QuasiPeak 1.0 s Average

IF Bandw. 10 kHz

Transducer

NSLK_8128

Level [dBµV]

70

60

50

40

30

20

10

0

-10 150k 300k 500k 1M

Frequency [Hz] 2M 3M 5M 7M 10M 30M

MES SÖ-MBC2_02_pre PK MES SÖ-MBC2_02_pre AV LIM EN 55011 V QP -B- LIM EN 55011 V AV -B-

class B class B

EMC-Centre Page 8 of 20

Page 9: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Set-Up: Test according to basic standard DIN EN 55011 Radiated emission, 30-1000MHz

EMC test assembly No. 3: accuracy 4.5dB

Receiving antenna 3m

Power Supply Line Stabilization

Network

Equipment under

Test

EMI receiver

IEEE 488.2 - BUS (GPIB)

Personal Computer

Test Sample: ELIGHT-BN720

Results:

Measuring Point Limit Result Remark

Prescan: Detectorfunction: Peak Final Measurement: Detektorfunction:

Quasipeak Antenna horiz. + vert. oriented

3 m distance DIN EN 55011 Class B

Within limit (Diagram 3)

EMC-Centre Page 9 of 20

Page 10: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

MeßTechnikNord GmbH Funkstörfeldstärke

EUT: Manufacturer:

ELIGHT-BN720 Ser.Nr.: ELIGHT-BN720Benice International

Operating Condition: Elight (RF + IPL)Test Site: Operator: Test Specification: Comment: Start of Test:

Halle 1 Shiu Wiu EN55011 B 3m Antennenabstand 13.11.2006

SCAN TABLE: "EN 55011 Field ESMI"

Short Description: Start Stop

EN 55011 Field Strength 3m Detector

MaxPeak MaxPeak

Meas. Time 10.0 ms 10.0 ms

IF Bandw.

Step Frequency Frequency Width 30.0 MHz 1.0 GHz

1.0 GHz 2.5 GHz

50.0 kHz 50.0 kHz

Transducer

120 kHz CBL 6111A 1832 120 kHz EATON 96001

SCAN TABLE: "EN 55011 Field QP ES"

Short Description: Start Stop

EN 55011 Field Strength 3m Detector Meas.

Time IF Bandw.

Step Transducer Frequency Frequency Width 30.0 MHz 1.0 GHz

1.0 GHz 2.5 GHz

60.0 kHz 60.0 kHz

QuasiPeak 1.0 s QuasiPeak 1.0 s

120 kHz CBL 6111A 1832 120 kHz EATON 96001

Diagram 3

Level [dBµV/m]

40

30

x x

x

x 20

10

0

-5 30M 50M 70M 100M 200M

Frequency [Hz] 300M 500M 1G 2G

x MES SÖ_MBC2_01_fin QP MES SÖ_MBC2_01_pre PK LIM EN 55022 F QP B MES SÖ_MBC2_01a_pre PK

EMC-Centre Page 10 of 20

Page 11: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Set-Up: Test according to basic standard DIN EN 61000- 4-3 Immunity to Electromagnetic Fields, 80MHz -2.5GHz

EMC test assembly No. 6: accuracy 29.8 %

2.0

Power Supply Line

Stabilization Network

Equipment under Test

Transmitting antenna

Directional Coupler

Millivolt- Meter

Signal- generator

Broadband amplifier

IEEE 488.2 - BUS (GPIB)

Personal Computer

Test Sample: ELIGHT-BN720Frequency steps: 1% / 1000 ms

Results:

Frequency Range Test Level Result Remarks

80 MHz -2.5 GHz 10 V/m Crit. A

No malfunction Criteria A fulfilled

Testsignal modulated: AM, 1kHz/80% Antenna horizontal + vertical

EMC-Centre Page 11 of 20

Page 12: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Set-Up: Test according to basic standard DIN EN 61000-4-5 Immunity to Conducted Slow Transients, Surge

EMC test assembly No. 10: accuracy 11.5 %

Power Supply Coupling

filter

Equipment under Test

Hybrid- generator

Test Sample: ELIGHT-BN720

Results:

Measuring point Testlevel Result Remarks

5 pulses per polarity, 1ppm Pulseform 1.2/50µs power mains

L PE ± 2 kV unsym.

Critt. B

± 2 kV N PE unsym.

Critt. B

± 1 kV L N sym.

No malfunction Criteria A fulfilled

No malfunction Criteria A fulfilled

No malfunction Criteria A fulfilled Critt. B

EMC-Centre Page 12 of 20

Page 13: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Test Set-Up: Test according to basic standard DIN EN 61000-4-8 Immunity to Power frequency magnetic field immunity test, 3 A/m, 50 Hz

EMC test assembly No. 11: accuracy ±19 %

USB Kabel (geschirmt)

Notebook

Equipment Under

Test Isolation (10cm)

Groundplane (min. 1 x 1m)

AC Power Source

Result:

Test point Test Level Result Remark

EUT in center of loop 3 A/m no malfunction; criteria A fulfilled

Tested with 50 Hz/ 60 Hz Tested at EUT axis X , Y und Z

EMC-Centre Page 13 of 20

Page 14: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

Fig. 5: Set up for magnetic field immunity test, x-axis

Fig. 7: Set up for magnetic field immunity test, z-axis

EMC-Centre Page 14 of 20

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

TEST SET-UP: Test according to basic standard DIN EN 61000-4-11 Voltage dips, short interruptions

EMC test assembly No. 13, accuracy 16.3 %

Power Supply Variable

Power

Supply

Equipment

under Test

RESULTS:

Level Result Remark

0 V for 10ms

161 V for 500 ms

92 V for 100 ms

no malfunction criteria C fulfilled

no malfunction criteria A fulfilled

no malfunction criteria A fulfilled

Interruption Rebooting with default values

Voltage dip 30%

- 1 x

- 3 x

Voltage dip 60%, - 3 x

EMC-Centre

Page 15: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

TEST SET-UP: Test acc. to DIN EN 61000-3-2, Limitation of harmonic current emissions

EMC test assembly No. 27: Power 0,2 % Voltage 0,1 % Current 0,1 %

Power factor 0,002 L 10 % R 20 %

Power Supply Line

Stabilization

Network EN 61000-3-2

Equipment

under

Test

Power

Analysis

System

IEEE 488.2 - BUS (GPIB)

Personal

Computer Printer

Test Sample: ELIGHT-BN720

RESULTS:

Measuring Point Result Remark

power mains (230 V ) AC

Pass; within limit, class A compare protocol, page 22, 23

EMC-Centre Page 15 of 20

Page 16: emc elight bn 720

Benice International

E-light BN720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

EMC-Centre Page 16 of 20

Page 17: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

EMC-Centre Page 17 of 20

Page 18: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

TEST SET-UP: Test acc. to DIN EN 61000-3-3, Limitation of voltage fluctuations and flicker

EMC test assembly No.. 27: Power 0,2 % Voltage 0,1 % Current 0,1 %

Power factor 0,002 L 10 % R 20 %

Power Supply Line

Stabilization

Network

EN 61000-3-3

Equipment

under

Test

Power

Analysis

System

IEEE 488.2 - BUS (GPIB)

Personal

Computer Printer

Test Sample: ELIGHT-BN720

RESULTS:

Measuring Point Result Remark

Power supply pass , within limit compare protocol, page 25, 26

EMC-Centre Page 18 of 20

Page 19: emc elight bn 720

Benice International

BN720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

EMC-Centre Page 19 of 20

Page 20: emc elight bn 720

Doc.-No. Date

: :

05372.213.06 17.11.2006

EMC-Test E-LIGHT - BN720

EMC-Centre Page 20 of 20