Calibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina) Quantity Instrument or artifact Instrument Type or Method Minimum value Maximum value units Parameter Specifications Value Units Coverage Factor Level of Confidence Is the expanded uncertainty a relative one? Uncertainty matrix NMI Service Identifier Comments DC voltage sources: single values Solid state voltage standard Direct comparison with a bank of solid state standards 10 10 V 1 μV/V 2 95% Yes 1 Approved on 11 May 2017 DC voltage sources: low values DC voltage source, multifunction calibrator Comparison with reference digital multimeter 0.01 10 V 4 to 17 μV/V 2 95% Yes DCV_M1.1.2 2 Approved on 11 May 2017 DC voltage sources: intermediate values DC voltage source, multifunction calibrator Comparison with reference digital multimeter 10 1000 V 4 to 8.5 μV/V 2 95% Yes DCV_M1.1.3 3 Approved on 11 May 2017 DC voltage meters: intermediate values DC voltmeter, multimeter Comparison with reference digital multimeter 0.01 1000 V 4 to 17 μV/V 2 95% Yes DCV_M1.2.2-1 4 Approved on 11 May 2017 DC voltage meters: intermediate values DC voltmeter, multimeter Direct comparison with multifunction calibrator 0.01 1000 V 5 to 45.5 μV/V 2 95% Yes DCV_M1.2.2-2 4 Approved on 11 May 2017 DC resistance standards and sources: intermediate values Fixed resistor DC current comparator 0.1 1 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.1 Approved on 11 May 2017 DC resistance standards and sources: intermediate values Fixed resistor DC current comparator 1 1.00E+05 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.2 Approved on 11 May 2017 DC resistance standards and sources: Multiple ranges Multifunction calibrator Comparison with reference digital multimeter 1 1.00E+08 Ω 8 to 135 µΩ/Ω 2 95% Yes DCR_M2.1.5 Approved on 11 May 2017 DC resistance meters: intermediate values Ohmeter, multimeter Comparison to standard resistor 1 1.00E+09 Ω 1 to 19 µΩ/Ω 2 95% Yes DCR_M2.2.2 Approved on 11 May 2017 Calibration or Measurement Services Measurand Level or Range Measurement Conditions/Independent variables Expanded Uncertainty The BIPM key comparison database, June 2017 1/31

Transcript of Electricity and Magnetism, Electricity and Magnetism...

Page 1: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

QuantityInstrument or

artifact

Instrument Type

or Method

Minimum

value

Maximum

valueunits Parameter Specifications Value Units

Coverage

Factor

Level of

Confidence

Is the expanded

uncertainty a

relative one?

Uncertainty

matrix

NMI Service

IdentifierComments

DC voltage sources:

single values

Solid state voltage

standard

Direct comparison

with a bank of solid

state standards

10 10 V 1 µV/V 2 95% Yes 1Approved on

11 May 2017

DC voltage sources:

low values

DC voltage source,

multifunction

calibrator

Comparison with

reference digital

multimeter

0.01 10 V 4 to 17 µV/V 2 95% Yes DCV_M1.1.2 2Approved on

11 May 2017

DC voltage sources:

intermediate values

DC voltage source,

multifunction

calibrator

Comparison with

reference digital

multimeter

10 1000 V 4 to 8.5 µV/V 2 95% Yes DCV_M1.1.3 3Approved on

11 May 2017

DC voltage meters:

intermediate values

DC voltmeter,

multimeter

Comparison with

reference digital

multimeter

0.01 1000 V 4 to 17 µV/V 2 95% Yes DCV_M1.2.2-1 4Approved on

11 May 2017

DC voltage meters:

intermediate values

DC voltmeter,

multimeter

Direct comparison

with multifunction

calibrator

0.01 1000 V 5 to 45.5 µV/V 2 95% Yes DCV_M1.2.2-2 4Approved on

11 May 2017

DC resistance

standards and

sources:

intermediate values

Fixed resistorDC current

comparator0.1 1 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.1

Approved on

11 May 2017

DC resistance

standards and

sources:

intermediate values

Fixed resistorDC current

comparator1 1.00E+05 Ω 0.65 µΩ/Ω 2 95% Yes DCR_M2.1.2

Approved on

11 May 2017

DC resistance

standards and

sources: Multiple

ranges

Multifunction

calibrator

Comparison with

reference digital

multimeter

1 1.00E+08 Ω 8 to 135 µΩ/Ω 2 95% Yes DCR_M2.1.5Approved on

11 May 2017

DC resistance

meters: intermediate

values

Ohmeter, multimeterComparison to

standard resistor1 1.00E+09 Ω 1 to 19 µΩ/Ω 2 95% Yes DCR_M2.2.2

Approved on

11 May 2017

Calibration or Measurement Services Measurand Level or RangeMeasurement

Conditions/Independent variablesExpanded Uncertainty

The BIPM key comparison database, June 2017 1/31

Page 2: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

QuantityInstrument or

artifact

Instrument Type

or Method

Minimum

value

Maximum

valueunits Parameter Specifications Value Units

Coverage

Factor

Level of

Confidence

Is the expanded

uncertainty a

relative one?

Uncertainty

matrix

NMI Service

IdentifierComments

Calibration or Measurement Services Measurand Level or RangeMeasurement

Conditions/Independent variablesExpanded Uncertainty

DC resistance

meters: intermediate

values

Ohmeter, multimeter

Direct comparison

with reference

digital multimeter

1 1 MΩ 5.2 µΩ/Ω 2 95% YesApproved on

11 May 2017

DC current sources:

low values

DC current generator,

multifunction

calibrator

Drop voltage

across the

standard resistor

1 100 μA 6.5 to 20 μA/A 2 95% Yes DCI_M3.1.1-1Approved on

11 May 2017

DC current sources:

low values

DC current generator,

multifunction

calibrator

Direct comparison

with reference

digital multimeter

10 100 μA 13.5 to 50 μA/A 2 95% Yes DCI_M3.1.1-2Approved on

11 May 2017

DC current sources:

intermediate values

DC current generator,

multifunction

calibrator

Drop voltage

across the

standard resistor

0.0001 0.2 A 6.5 μA/A 2 95% Yes DCI_M3.1.2-1Approved on

11 May 2017

DC current sources:

intermediate values

DC current generator,

multifunction

calibrator

Direct comparison

with reference

digital multimeter

0.0001 1.E+01 A 11 to 620 μA/A 2 95% Yes DCI_M3.1.2-2Approved on

11 May 2017

DC current meters:

low values

DC current meter,

multimeter

Drop voltage

across the

standard resistor

1 100 μA 6.5 to 25 μA/A 2 95% Yes DCI_M3.2.1-1Approved on

11 May 2017

DC current meters:

low values

DC current meter,

multimeter

Direct comparison

with multifunction

calibrator

10 100 μA 41 to 190 μA/A 2 95% Yes DCI_M3.2.1-2Approved on

11 May 2017

DC current meters:

low values

DC current meter,

multimeter

Direct comparison

with reference

digital multimeter

10 100 μA 13.5 to 50 μA/A 2 95% Yes DCI_M3.2.1-3Approved on

11 May 2017

The BIPM key comparison database, June 2017 2/31

Page 3: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

QuantityInstrument or

artifact

Instrument Type

or Method

Minimum

value

Maximum

valueunits Parameter Specifications Value Units

Coverage

Factor

Level of

Confidence

Is the expanded

uncertainty a

relative one?

Uncertainty

matrix

NMI Service

IdentifierComments

Calibration or Measurement Services Measurand Level or RangeMeasurement

Conditions/Independent variablesExpanded Uncertainty

DC current meters:

intermediate values

DC current meter,

multimeter

Drop voltage

across the

standard resistor

0.1 200 mA 6.5 μA/A 2 95% Yes DCI_M3.2.2-1Approved on

11 May 2017

DC current meters:

intermediate values

DC current meter,

multimeter

Direct comparison

with multifunction

calibrator

0.0001 10 A 27 to 625 μA/A 2 95% Yes DCI_M3.2.2-2Approved on

11 May 2017

DC current meters:

intermediate values

DC current meter,

multimeter

Direct comparison

with reference

digital multimeter

0.0001 10 A 11 to 620 μA/A 2 95% Yes DCI_M3.2.2-3Approved on

11 May 2017

AC voltage up to

1100 V: sources

Multifunction

calibrator

Direct comparison

with reference

multimeter

0.01 1000 V Frequency 50 Hz to 1 MHz 0.08 to 31 mV/V 2 95% Yes ACV_M5.2.1Approved on

11 May 2017

AC voltage up to

1100 V: meters

AC voltmeter,

multimeter

Direct comparison

with reference

multimeter

0.01 1000 V Frequency 50 Hz to 1 MHz 0.08 to 31 mV/V 2 95% Yes ACV_M5.2.2-1Approved on

11 May 2017

AC voltage up to

1100 V: meters

AC voltmeter,

multimeter

Direct comparison

with multifunction

calibrator

0.01 1000 V Frequency 50 Hz to 1 MHz 0.15 to 31 mV/V 2 95% Yes ACV_M5.2.2-2Approved on

11 May 2017

AC current up to

100 A: sources

AC current generator,

multifunction

calibrator

Direct comparison

with reference

multimeter

1.00E-04 10 A Frequency 50 Hz to 10 kHz 0.36 to 3.5 mA/A 2 95% Yes ACI_M6.2.1Approved on

11 May 2017

AC current up to

100 A: meters

AC ammeter,

multimeter

Drop voltage

across the

standard current

shunt

10 10 mA Frequency 300 Hz, 1 kHz 90 μA/A 2 95% YesApproved on

11 May 2017

AC current up to

100 A: meters

AC ammeter,

multimeter

Direct comparison

with a reference

standard

1.00E-04 10 A Frequency 50 Hz to 10 kHz 0.36 to 3.5 mA/A 2 95% Yes ACI_M6.2.2 15Approved on

11 May 2017

The BIPM key comparison database, June 2017 3/31

Page 4: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCV_M1.1.2

Expanded uncertainty (µV/V)

10 mV 17

20 mV 10

50 mV 6.5

100 mV 4.5

190 mV 4.5

0.2 V 4.5

0.5 V 4.5

1 V 4

1.9 V 4

5 V 4

10 V 4

The BIPM key comparison database, June 2017 4/26

Page 5: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCV_DCV_M1.1.3

Expanded uncertainty (µV/V)

15 V 4

19 V 4

20 V 6.5

50 V 5.5

100 V 5.5

190 V 6.5

200 V 8.5

500 V 6.5

900 V 5.5

1000 V 5.5

The BIPM key comparison database, June 2017 5/26

Page 6: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCV_M1.2.2-1

Expanded uncertainty (µV/V)

10 mV 17

20 mV 10

50 mV 6.5

100 mV 4.5

190 mV 4.5

0.2 V 4.5

0.5 V 4.5

1 V 4

1.9 V 4

5 V 4

10 V 4

15 V 4

19 V 4

20 V 6.5

50 V 5.5

100 V 5.5

190 V 6.5

200 V 8.5

500 V 6.5

900 V 5.5

1000 V 5.5

The BIPM key comparison database, June 2017 6/26

Page 7: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCV_M1.2.2-2

Expanded uncertainty (µV/V)

10 mV 45.5

20 mV 25

50 mV 12

100 mV 8

190 mV 6.5

0.2 V 6.5

0.5 V 6

1 V 5

1.9 V 5

5 V 5

10 V 5

15 V 5

19 V 5

20 V 7

50 V 6.5

100 V 6.5

190 V 7

200 V 9

500 V 7.5

900 V 6.5

1000 V 6.5

The BIPM key comparison database, June 2017 7/26

Page 8: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCR_M2.1.1

Expanded uncertainty (µΩ/Ω)

0.1 Ω to 1 Ω 0.65

The BIPM key comparison database, June 2017 8/26

Page 9: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCR_M2.1.2

Value Expanded uncertainty (µΩ/Ω)

1 Ω to 10 Ω 0.65

10 Ω to 10 kΩ 0.65

10 kΩ to 100 kΩ 0.65

The BIPM key comparison database, June 2017 9/26

Page 10: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCR_M2.1.5

Expanded uncertainty (µΩ/Ω)

1 Ω 16.0

10 Ω 9.0

100 Ω 8.0

1 kΩ 8.0

10 kΩ 8.0

100 kΩ 8.0

1 MΩ 8.5

10 MΩ 20

100 MΩ 135

The BIPM key comparison database, June 2017 10/26

Page 11: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCR_Μ2.2.2

Expanded uncertainty (µΩ/Ω)

1 Ω 1

1.9 Ω 1

10 Ω 1

100 Ω 1

1 kΩ 1

10 kΩ 1

19 kΩ 1

100 kΩ 1

1 MΩ 4

10 MΩ 4

100 MΩ 6.50

1 GΩ 19

The BIPM key comparison database, June 2017 11/26

Page 12: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.1.1-1

Expanded uncertainty (µA/A)

1 μA 20

2 μA 15

5 μA 9

10 μA 7

20 μA 6.5

50 μA 6.5

100 μA 6.5

The BIPM key comparison database, June 2017 12/26

Page 13: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.1.1-2

Expanded uncertainty (µA/A)

10 μA 50

20 μA 29

50 μA 18.5

100 μA 13.5

The BIPM key comparison database, June 2017 13/26

Page 14: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.1.2-1

Expanded uncertainty (µA/A)

200 μA 6.5

500 μA 6.5

1 mA 6.5

2 mA 6.5

5 mA 6.5

10 mA 6.5

20 mA 6.5

50 mA 6.5

100 mA 6.5

200 mA 6.5

The BIPM key comparison database, June 2017 14/26

Page 15: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.1.2-2

Value Expanded uncertainty (µA/A)

190 μA 12

0.2 mA 27.5

0.5 mA 16

1 mA 12.5

1.9 mA 11

2 mA 29

5 mA 17

10 mA 13.5

19 mA 12

20 mA 75

50 mA 50

100 mA 45

190 mA 40

0.2 A 330

0.5 A 275

1 A 210

1.9 A 200

2 A 620

5 A 500

10 A 430

The BIPM key comparison database, June 2017 15/26

Page 16: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.1-1

Value Expanded uncertainty (µA/A)

1 μA 25

2 μA 15

5 μA 9

10 μA 7

20 μA 6.5

50 μA 6.5

100 μA 6.5

The BIPM key comparison database, June 2017 16/26

Page 17: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.1-2

Value Eexpanded uncertainty (µA/A)

10 μA 190

20 μA 110

50 μA 58

100 μA 41

The BIPM key comparison database, June 2017 17/26

Page 18: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.1-3

Value Expanded uncertainty (µA/A)

10 μA 50

20 μA 29.5

50 μA 18.5

100 μA 13.5

The BIPM key comparison database, June 2017 18/26

Page 19: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.2-1

Value Expanded uncertainty (µA/A)

200 μA 6.5

500 μA 6.5

1 mA 6.5

2 mA 6.5

5 mA 6.5

10 mA 6.5

20 mA 6.5

50 mA 6.5

100 mA 6.5

200 mA 6.5

The BIPM key comparison database, June 2017 19/26

Page 20: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.2-2

Value Expanded uncertainty (µA/A)

190 μA 33

0.2 mA 41

0.5 mA 35

1 mA 29

1.9 mA 27

2 mA 38

5 mA 35

10 mA 29

19 mA 27

20 mA 79

50 mA 60

100 mA 53

190 mA 48

0.2 A 335

0.5 A 280

1 A 220

1.9 A 205

2 A 625

5 A 520

10 A 450

The BIPM key comparison database, June 2017 20/26

Page 21: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix DCI_M3.2.2-3

Value Expanded uncertainty (µA/A)

190 μA 12

0.2 mA 27.5

0.5 mA 16

1 mA 12.5

1.9 mA 11

2 mA 29

5 mA 17

10 mA 13.5

19 mA 12

20 mA 75

50 mA 50

100 mA 45

190 mA 40

0.2 A 330

0.5 A 275

1 A 210

1.9 A 200

2 A 620

5 A 500

10 A 430

The BIPM key comparison database, June 2017 21/26

Page 22: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix ACV_M5.2.1

50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz

10 mV 0.72 0.50 1.20 3.00 3.00 - -

20 mV 0.40 0.30 0.75 1.80 1.80 - -

100 mV 0.16 0.15 0.40 1.00 1.00 - -

190 mV 0.15 0.15 0.35 0.85 0.85 - -

1 V 0.11 0.09 0.25 0.72 0.72 - -

1.9 V 0.10 0.08 0.25 0.62 0.62 - -

10 V 0.11 0.09 0.25 0.72 0.72 5.10 31.00

19 V 0.10 0.08 0.25 0.62 0.62 4.10 21.00

100 V 0.11 0.10 0.25 0.72 0.72 - -

190 V 0.10 0.08 0.25 0.62 0.62 - -

500 V 0.13 0.13 0.30 - - - -

700 V 0.13 0.13 0.35 - - - -

1000 V 0.11 0.11 0.29 - - - -

The expanded uncertainties are expressed in mV/V

The BIPM key comparison database, June 2017 22/26

Page 23: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix ACV_M5.2.2-1

50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz

10 mV 0.72 0.51 1.20 2.85 3.00 - -

20 mV 0.40 0.30 0.75 1.80 1.80 - -

100 mV 0.16 0.15 0.40 1.00 1.00 - -

190 mV 0.15 0.15 0.35 0.85 0.85 - -

1 V 0.11 0.10 0.25 0.72 0.72 - -

1.9 V 0.10 0.08 0.25 0.62 0.62 - -

10 V 0.11 0.09 0.25 0.72 0.72 5.10 31.00

19 V 0.10 0.08 0.25 0.62 0.62 4.10 21.00

100 V 0.11 0.10 0.25 0.72 0.72 - -

190 V 0.10 0.08 0.25 0.62 0.62 - -

500 V 0.13 0.13 0.29 - - - -

700 V 0.13 0.13 0.35 - - - -

1000 V 0.11 0.11 0.29 - - - -

The expanded uncertainties are expressed in mV/V

The BIPM key comparison database, June 2017 23/26

Page 24: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix ACV_M5.2.2-2

50 Hz 1 kHz 20 kHz 50 kHz 100 kHz 300 kHz 1 MHz

10 mV 0.90 0.70 1.30 3.00 3.20 - -

20 mV 0.50 0.50 0.85 2.00 2.00 - -

100 mV 0.25 0.25 0.50 1.10 1.20 - -

190 mV 0.20 0.20 0.40 0.90 1.10 - -

1 V 0.15 0.15 0.30 0.80 0.80 - -

1.9 V 0.15 0.15 0.30 0.70 0.70 - -

10 V 0.15 0.15 0.30 0.80 0.80 5.50 31.00

19 V 0.15 0.15 0.30 0.70 0.70 4.50 21.00

100 V 0.15 0.15 0.30 0.75 0.75 - -

190 V 0.15 0.15 0.30 0.70 0.70 - -

500 V 0.20 0.20 0.40 - - - -

700 V 0.20 0.20 0.40 - - - -

1000 V 0.15 0.15 0.40 - - - -

The expanded uncertainties are expressed in mV/V

The BIPM key comparison database, June 2017 24/26

Page 25: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix ACI_M6.2.1

Value 50 Hz 1 kHz 5 kHz 10 kHz

100 μA 0.50 0.50 0.50 0.55

190 μA 0.45 0.45 0.45 0.45

0.2 mA 1.30 1.30 1.30 1.30

1 mA 0.45 0.45 0.45 0.45

1.9 mA 0.36 0.36 0.36 0.36

2 mA 1.30 1.30 1.30 1.30

10 mA 0.45 0.45 0.45 0.45

19 mA 0.36 0.36 0.36 0.36

20 mA 1.30 1.30 1.30 1.30

100 mA 0.45 0.45 0.45 0.45

190 mA 0.36 0.36 0.36 0.36

0.2 A 1.60 1.60 1.70 1.70

1 A 0.80 0.80 0.90 0.90

1.9 A 0.71 0.71 0.82 0.82

2 A 1.80 1.8 3.50 3.50

10 A 1.00 1.00 2.70 2.70

The expanded uncertain The expanded uncertainties in the table are expressed in mA/A

The BIPM key comparison database, June 2017 25/26

Page 26: Electricity and Magnetism, Electricity and Magnetism ...kcdb.bipm.org/appendixC/EM/BA/EM_BA.pdfCalibration and Measurement Capabilities Electricity and Magnetism, Electricity and Magnetism,

Calibration and Measurement Capabilities

Electricity and Magnetism, Bosnia and Herzegovina, IMBiH (Institute of Metrology of Bosnia and Herzegovina)

Matrix ACI_M6.2.2

50 Hz 1 kHz 5 kHz 10 kHz

100 μA 0.50 0.50 0.50 0.55

190 μA 0.45 0.45 0.45 0.45

0.2 mA 1.30 1.30 1.30 1.30

1 mA 0.45 0.45 0.45 0.45

1.9 mA 0.36 0.36 0.36 0.36

2 mA 1.30 1.30 1.30 1.30

10 mA 0.45 0.45 0.45 0.45

19 mA 0.36 0.36 0.36 0.36

20 mA 1.30 1.30 1.30 1.30

100 mA 0.45 0.45 0.45 0.45

190 mA 0.36 0.36 0.36 0.36

0.2 A 1.60 1.60 1.70 1.70

1 A 0.80 0.80 0.90 0.90

1.9 A 0.71 0.71 0.82 0.82

2 A 1.80 1.80 3.50 3.50

10 A 1.00 1.00 2.70 2.70

The expanded uncertain The expanded uncertainties in the table are expressed in mA/A

The BIPM key comparison database, June 2017 26/26