Eddy Current Testing
description
Transcript of Eddy Current Testing
EDDY CURRENT TESTING
By: Jabran Younas
A NEW BEGINNING
Applications
Metal thickness
Alloy sorting
Heat treat condition
Heat damage
Plating thicknessCladding Thickness
Insulation Thickness
Nonmetallic coating thicknessvibrationDiameter of tubing or bar stock
CracksSeamsPorosityCorrosionErosionSegregationInclusion
R2 XL2 Z2
Resistance
Inductive Reactance
Impedance
Ohm's Ohm's LawLaw
R=ρL/A
Resistivity (Ohm-cm)
Conductivity (mho or % IACS)
Faraday Faraday LawLaw
Ohm's Ohm's LawLaw
Faraday Faraday LawLaw
Ohm's Ohm's LawLaw
Єemf=NΔφΔt
Єemf=Induced emfN= Number of TurnsΔφ= Change in Flux
Faraday Faraday LawLaw
Ohm's Ohm's LawLaw
Lenz’s Lenz’s LawLaw
The direction of induced current is such as to oppose the cause producing it.
Faraday Faraday LawLaw
Ohm's Ohm's LawLaw
Lenz’s Lenz’s LawLawBiot Biot
Savart’s Savart’s LawLaw
B=μo/2π x I/r
Coils Magnetic Field
Coils Magnetic Field
Coils Magnetic Field
Coils
CoilsAC Coils
CoilsCoils
Conducting MaterialConducting Material
Eddy Currents
Eddy Currents Magnetic Field
Magnetic FieldFrom Test Coil
Magnetic Field From
Eddy Currents
Eddy Currents
Crack
Crack DetectionCrack Detection
Depth of Penetration & Skin Effect
• Eddy current density is greatest at surface
• Reduces exponentially with depth
• At standard D of P = 1/e (37%) of surface value
• δ = 50 ……ρ…. μr f
Effect Of Frequency
Effect of conductivity
Effect of Permeability
Dep
th
Eddy Current Density
Dep
th
Eddy Current DensityLow FrequencyLow ConductivityLow Permeability
High FrequencyHigh ConductivityHigh Permeability
Standard Depthof
Penetration(Skin Depth)
1/e or 37 % of surface density
Crack parallel to eddy
Currents - not detected
Crack interrupts eddy
Currents - detected
Discontinuities
• Edges produce signals just like cracks and mislead inspector
• Different techniques can be used to avoid edge effect
Edge effect
Inductive Coupling
Lift Off and Fill Factor
Usually, 70-90% "fill-factor" is targeted for reliable inspection
Phase Lag
Eddy Current Variables Comparison
A Typical ECT Instrument
Types of Probes
Surface
Probe
External Probe
Internal Probe
Mode of 0perationAbsolute Differential
Sensitive to both sudden and gradual changes in properties.
Not Sensitive gradual changes in properties
Easy to interpret Difficult to interpret
Show total length of long flaws Detect only ends of long flaws
Sensitive to drift due to temperature changes
Less Sensitive to drift due to temperature changes
Sensitive to probe wobble Less Sensitive to probe wobble
Single Coil Pair of coils
Absolute value of impedance and induced voltage is measured
Changes in impedance or induced voltage is mesured
Eddy Current Probes
Eddy Current Test Equipments
Digital Meter Analogue Meter
Multi-Freq Equipments
Calibration Standards
Testing Procedure• Balance on sound portion
• Set Sensitivity (Drive Level & Gain)
• Set Frequencies
• Draw Calibration Curve
• Take signal with constant speed
• Note Phase lag and amplitude for each indication
Effect of frequency on signal
Effect of Frequency
44khz 22khz
5khz
Wobble
Dent
High Speed
Correct Speed
Signal Filtering
Signal+ Noise Correct Filter is used
Incorrect Filter is used
Stripper
Thickness Monitoring Methodology
Stripper Tube Signal
Selection of Frequency
Selection of Probe
10% OD shown
36% OD
Defect Sizing
Defect Sizing
60% OD
80% OD
Support
60% changed to 78% near
support
80% OD
60% OD
Defect Sizing
Wrong Indications
RFT Theory
RFT Zones
RFT Signals
Effect of Support plate
Tube Inspection
Depth Curve
Magnet
Card Board
Non Magnetic Strip
Magnetic Strip
Shielding Effect