Easychem Coli on-line
Transcript of Easychem Coli on-line
Technical Data
MEASURING PRINCIPLE Colorimetric (total & faecal coliforms) and fluorimetric (E. coli)
MEASUREMENT TYPE Fully automated, direct read, on-line discrete analyzer
MEASURING
FREQUENCY Typically one sample / day
MEASURING TIME Up to 15-18 hours for coliforms presence/absence detection
MEASURING POINTS One
OPERATOR INTERFACE Colour graphic touch-screen 15” display
OUTPUT SIGNAL AND
COMMUNICATIONS PORTS
Digital outputs for alarm; RS232 output; RS485 optional
INPUT SIGNALS Analysis: 1 digital contact with photocoupler, galvanically isolated
ALARM CONTACTS
Coliform presence: 1 potential free switch SPDT, max load 24 AC
DC 0.5 A General: 1 potential free switch SPDT, max load 24 AC DC 0.5 A
ALARM MESSAGES On operator interface
SAMPLE DELIVERY
Pressure: atmospheric
Temperature: 10° - 35 °C Sample volume: 100 mL per analysis Sample connection: Standard silicone 2x4, other on request
REAGENTS IDEXX Colilert®-18
NUMBER OF SAMPLES 2 x 20, on automatic rotating circular trays
SAMPLES TO BE PROCESSED IN
PARALLEL
2: Total coliforms + E. Coli and Faecal coliforms
REAGENTS
REPLACEMENT Typically every 20 days, up to 40 days
ENVIRONMENTAL TEMP. 10-35°C (Reagents up to 25°C)
MOUNTING Stand-alone industrial cabinet
STANDARD PROTECTION
IP 55
HARDWARE Industrial microcontroller PC panel with touch screen graphic display
POWER SUPPLY 12/24 Vdc; external power supply from 110/220 Vac to 12/24 Vdc included in the scope of delivery
POWER ABSORPTION 100 W
WEIGHT 100 Kg without reagents
DIMENSIONS 190 (H) x 80 (W) x 50 (D) cm
Subject to change without notice
SYSTEA SpA
Facilities : Via Paduni, 2/A - 03012 ANAGNI (FR) -
Tel. (+39) 0775.776058 - Fax (+39) 0775.772204
e mail: [email protected] Web site: http://www.systea.it Document ID:Easychem Coli on-line.doc
EASYCHEM COLI on-line
DIRECT READING DISCRETE
ANALYZER FOR ON LINE DETECTION
OF TOTAL COLIFORMS, FAECAL
COLIFORMS AND E. COLI FOR
POTABLE, SURFACE, WASTE
AND BATHING WATER
MONITORING
COLILERT®-18 TEST
Simultaneously detects both total coliforms and Escherichia coli in water, or faecal
coliforms in wastewater. Results in 18 hours.
IDEXX Colilert®-18 has become the new
International Organization for Standardization (ISO) standard 9308-2:2012. Colilert-18 is also U.S. EPA-
approved and included in Standard Methods for Examination of Water and Wastewater.
BATCH MEASUREMENT SESSION PREPARATION
Open the tap of each 100 ml bottle provided with COLILERT®-18 test kit
Open the blister and pour the reaction
powder inside the bottle
Close the bottle using a Parafilm and the special tap with hole provided as
standard accessory with the on-line analyser and place it inside one of the sample carousels of the on-line analyser.
At the end of each batch measurement session the 20/40 bottles have to be manually removed from the analyser and properly disposed.
COMBINED COLORIMETRIC AND
FLUORIMETRIC DETECTORS
Presence of Total and Faecal coliforms is detected by variation of the colour of the sample solution, while E. coli by its
fluorimetric emission by the light excitation at specific UV wavelength.
EASYCHEM COLI on-line ANALYZER
TOUCH SCREEN DISPLAY
EASYCHEM COLI on-line
Microprocessor controlled on-line analyzer automating COLILERT®-18
test to detect coliforms in drinking, surface, waste and bathing water.
BACTERIA DETECTABLE:
• TOTAL COLI - COLORIMETRIC • E. COLI - FLUORIMETRIC
• FAECAL COLI - COLORIMETRIC
Total + E. coli incubated at 35±0.5 °C
Faecal coli incubated at 44±0.5 °C
ROBUST AND RELIABLE
Designed for industrial and environmental on-line applications, ensures the highest level of robustness in the electronics, mechanics and hydraulics components.
EASY TO INSTALL The analyzer is delivered only after a successful final tests. It is provided ready for installation, with a spares set for start-up operations.
MAIN APPLICATION FOR BACTERIA DETECTION
• Drinking water
• Bathing water
• Surface (lakes & rivers)
• Waste water treatment plants outlet.
STATISTICAL EVALUATION OF CONTAMINATION
The on-line analyser allows to estimate the level of contamination degree by correlating the detection times of colour (Total or Faecal
coliforms depending on temperature) and fluorescence (E. coli) development with the number of bacterial individuals.
FEATURES/BENEFITS
� Fully automatic operation
� 20 days autonomy, one sample per day, with automatic detection of Total + E. coli and Faecal coliforms or
� 40 samples autonomy with automatic detection of Total + E.
coli or, alternatively, Faecal coliforms
� Use of disposable 100 ml bottles ensures water sample
representativeness and avoids any cross contamination
� No need of long hydraulics washing using special cleaning solutions
� Automatic generation of local and remote alarms through
Ethernet or optional GSM/GPRS remote connection.
� Yearly maintenance.