Technical Data MEASURING PRINCIPLE Colorimetric (total & faecal coliforms) and fluorimetric (E. coli) MEASUREMENT TYPE Fully automated, direct read, on-line discrete analyzer MEASURING FREQUENCY Typically one sample / day MEASURING TIME Up to 15-18 hours for coliforms presence/absence detection MEASURING POINTS One OPERATOR INTERFACE Colour graphic touch-screen 15” display OUTPUT SIGNAL AND COMMUNICATIONS PORTS Digital outputs for alarm; RS232 output; RS485 optional INPUT SIGNALS Analysis: 1 digital contact with photocoupler, galvanically isolated ALARM CONTACTS Coliform presence: 1 potential free switch SPDT, max load 24 AC DC 0.5 A General: 1 potential free switch SPDT, max load 24 AC DC 0.5 A ALARM MESSAGES On operator interface SAMPLE DELIVERY Pressure: atmospheric Temperature: 10° - 35 °C Sample volume: 100 mL per analysis Sample connection: Standard silicone 2x4, other on request REAGENTS IDEXX Colilert ® -18 NUMBER OF SAMPLES 2 x 20, on automatic rotating circular trays SAMPLES TO BE PROCESSED IN PARALLEL 2: Total coliforms + E. Coli and Faecal coliforms REAGENTS REPLACEMENT Typically every 20 days, up to 40 days ENVIRONMENTAL TEMP. 10-35°C (Reagents up to 25°C) MOUNTING Stand-alone industrial cabinet STANDARD PROTECTION IP 55 HARDWARE Industrial microcontroller PC panel with touch screen graphic display POWER SUPPLY 12/24 Vdc; external power supply from 110/220 Vac to 12/24 Vdc included in the scope of delivery POWER ABSORPTION 100 W WEIGHT 100 Kg without reagents DIMENSIONS 190 (H) x 80 (W) x 50 (D) cm Subject to change without notice SYSTEA SpA Facilities : Via Paduni, 2/A - 03012 ANAGNI (FR) - Tel. (+39) 0775.776058 - Fax (+39) 0775.772204 e mail: [email protected] Web site: http://www.systea.it Document ID:Easychem Coli on-line.doc EASYCHEM COLI on-line DIRECT READING DISCRETE ANALYZER FOR ON LINE DETECTION OF TOTAL COLIFORMS, FAECAL COLIFORMS AND E. COLI FOR POTABLE, SURFACE, WASTE AND BATHING WATER MONITORING

Transcript of Easychem Coli on-line

Page 1: Easychem Coli on-line

Technical Data

MEASURING PRINCIPLE Colorimetric (total & faecal coliforms) and fluorimetric (E. coli)

MEASUREMENT TYPE Fully automated, direct read, on-line discrete analyzer

MEASURING

FREQUENCY Typically one sample / day

MEASURING TIME Up to 15-18 hours for coliforms presence/absence detection

MEASURING POINTS One

OPERATOR INTERFACE Colour graphic touch-screen 15” display

OUTPUT SIGNAL AND

COMMUNICATIONS PORTS

Digital outputs for alarm; RS232 output; RS485 optional

INPUT SIGNALS Analysis: 1 digital contact with photocoupler, galvanically isolated

ALARM CONTACTS

Coliform presence: 1 potential free switch SPDT, max load 24 AC

DC 0.5 A General: 1 potential free switch SPDT, max load 24 AC DC 0.5 A

ALARM MESSAGES On operator interface

SAMPLE DELIVERY

Pressure: atmospheric

Temperature: 10° - 35 °C Sample volume: 100 mL per analysis Sample connection: Standard silicone 2x4, other on request

REAGENTS IDEXX Colilert®-18

NUMBER OF SAMPLES 2 x 20, on automatic rotating circular trays

SAMPLES TO BE PROCESSED IN

PARALLEL

2: Total coliforms + E. Coli and Faecal coliforms

REAGENTS

REPLACEMENT Typically every 20 days, up to 40 days

ENVIRONMENTAL TEMP. 10-35°C (Reagents up to 25°C)

MOUNTING Stand-alone industrial cabinet

STANDARD PROTECTION

IP 55

HARDWARE Industrial microcontroller PC panel with touch screen graphic display

POWER SUPPLY 12/24 Vdc; external power supply from 110/220 Vac to 12/24 Vdc included in the scope of delivery

POWER ABSORPTION 100 W

WEIGHT 100 Kg without reagents

DIMENSIONS 190 (H) x 80 (W) x 50 (D) cm

Subject to change without notice

SYSTEA SpA

Facilities : Via Paduni, 2/A - 03012 ANAGNI (FR) -

Tel. (+39) 0775.776058 - Fax (+39) 0775.772204

e mail: [email protected] Web site: http://www.systea.it Document ID:Easychem Coli on-line.doc

EASYCHEM COLI on-line

DIRECT READING DISCRETE

ANALYZER FOR ON LINE DETECTION

OF TOTAL COLIFORMS, FAECAL

COLIFORMS AND E. COLI FOR

POTABLE, SURFACE, WASTE

AND BATHING WATER

MONITORING

Page 2: Easychem Coli on-line

COLILERT®-18 TEST

Simultaneously detects both total coliforms and Escherichia coli in water, or faecal

coliforms in wastewater. Results in 18 hours.

IDEXX Colilert®-18 has become the new

International Organization for Standardization (ISO) standard 9308-2:2012. Colilert-18 is also U.S. EPA-

approved and included in Standard Methods for Examination of Water and Wastewater.

BATCH MEASUREMENT SESSION PREPARATION

Open the tap of each 100 ml bottle provided with COLILERT®-18 test kit

Open the blister and pour the reaction

powder inside the bottle

Close the bottle using a Parafilm and the special tap with hole provided as

standard accessory with the on-line analyser and place it inside one of the sample carousels of the on-line analyser.

At the end of each batch measurement session the 20/40 bottles have to be manually removed from the analyser and properly disposed.

COMBINED COLORIMETRIC AND

FLUORIMETRIC DETECTORS

Presence of Total and Faecal coliforms is detected by variation of the colour of the sample solution, while E. coli by its

fluorimetric emission by the light excitation at specific UV wavelength.

EASYCHEM COLI on-line ANALYZER

TOUCH SCREEN DISPLAY

EASYCHEM COLI on-line

Microprocessor controlled on-line analyzer automating COLILERT®-18

test to detect coliforms in drinking, surface, waste and bathing water.

BACTERIA DETECTABLE:

• TOTAL COLI - COLORIMETRIC • E. COLI - FLUORIMETRIC

• FAECAL COLI - COLORIMETRIC

Total + E. coli incubated at 35±0.5 °C

Faecal coli incubated at 44±0.5 °C

ROBUST AND RELIABLE

Designed for industrial and environmental on-line applications, ensures the highest level of robustness in the electronics, mechanics and hydraulics components.

EASY TO INSTALL The analyzer is delivered only after a successful final tests. It is provided ready for installation, with a spares set for start-up operations.

MAIN APPLICATION FOR BACTERIA DETECTION

• Drinking water

• Bathing water

• Surface (lakes & rivers)

• Waste water treatment plants outlet.

STATISTICAL EVALUATION OF CONTAMINATION

The on-line analyser allows to estimate the level of contamination degree by correlating the detection times of colour (Total or Faecal

coliforms depending on temperature) and fluorescence (E. coli) development with the number of bacterial individuals.

FEATURES/BENEFITS

� Fully automatic operation

� 20 days autonomy, one sample per day, with automatic detection of Total + E. coli and Faecal coliforms or

� 40 samples autonomy with automatic detection of Total + E.

coli or, alternatively, Faecal coliforms

� Use of disposable 100 ml bottles ensures water sample

representativeness and avoids any cross contamination

� No need of long hydraulics washing using special cleaning solutions

� Automatic generation of local and remote alarms through

Ethernet or optional GSM/GPRS remote connection.

� Yearly maintenance.