Scale Terminal Calibration
Mettler-Toledo IND560
Training Purposes Only2
How To : Traditional Calibration
Scale System Calibration
Main area of proficiency:- Mettler-Toledo IND560 Scale Terminal
Training Purposes Only3
How To : Traditional Calibration
Entering Setup Mode
Press the Down Arrow to viewthe SETUP softkey.
The configuration of the IND560 terminal is accessed through the SETUP softkey
Softkeys
Training Purposes Only4
How To : Traditional Calibration
With the Down Arrow navigate to Scale Sub-branch
Use the UP and DOWN navigation keys to move focus through the branches of the setup menu tree.
Press the RIGHT navigation key to expand a branch and the LEFT navigation key to collapse a branch. When the focus is on a sub-branch, focus can quickly be moved back to the main branch by pressing the LEFT navigation key.
When a branch such as Calibration is in focus, press the ENTER key to display the setup screen for that function.
Training Purposes Only5
How To : Traditional Calibration
Navigate to Calibration- Press the Down Arrow twice.
When at Calibration-Press Enter
Training Purposes Only6
How To : Traditional Calibration
Navigate to Calibration- Press the Down Arrow four (4) times.
Geo Code Base Serial Number
Calibration Units Linearity Adjust
Training Purposes Only7
How To : Traditional Calibration
Press Zero softkey - Remove all weight from scale
- Press Start
Training Purposes Only8
How To : Traditional Calibration
Press Span softkey
- Last test weight value of 100 Lb. is displayed
- Press Enter twice
- Add test weight either manually or via Automated Test Weight System
Training Purposes Only9
How To : Traditional Calibration
Span- Press Start
Training Purposes Only10
How To : Traditional Calibration
- Press Exit twice when done
- Unit will return to normal weighing mode
-Weigh Hopper Discharge Feeder needs to be placed back into Automatic.