DTR_FCR-15A_FCD-15A

73
ACSE Sp. z o.o. ZESTAW INSTRUKCJI DO MIKROPROCESOROWYCH REGULATORÓW TEMPERATUYRY I PROCESÓW Z SERII FCR-15A I FCD-15A ORAZ INTERFEJS KOMUNIKACYJNY RS-232 I RS485 PROTOKÓŁ KOMUNIKACYJNY SHINKO MODBUS ASCII

Transcript of DTR_FCR-15A_FCD-15A

Page 1: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

ZESTAW INSTRUKCJI

DO MIKROPROCESOROWYCH

REGULATORÓW TEMPERATUYRY

I PROCESÓW

Z SERII

FCR-15A I FCD-15A

ORAZ

INTERFEJS KOMUNIKACYJNY RS-232 I RS485

PROTOKÓŁ KOMUNIKACYJNY

SHINKO

MODBUS ASCII

Page 2: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

ON/OFF SERVO

DIGITAL INDICATING CONTROLLER

FCD-15A, FCR-15AINSTRUCTION MANUAL

Page 3: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

2

PrefaceThank you for purchasing our ON/OFF SERVO digital indicating controller FCD-15A, FCR-15A.This manual contains instructions for the mounting, functions, operations and notes whenoperating the FCD-15A, FCR-15A. To prevent accidents arising from the misuse of thiscontroller, please ensure the operator receives this manual.Notes• This instrument should be used in accordance with the specifications described in the manual.

If it is not used according to the specifications, it may malfunction or cause a fire.• Be sure to follow the warnings, cautions and notices. If they are not observed, serious injury or

malfunction may occur.• Specifications, external appearance of the FCD-15A, FCR-15A and the contents of this

instruction manual are subject to change without notice.• Care has been taken to ensure that the contents of this instruction manual are correct, but if

there are any doubts, mistakes or questions, please inform our sales department.• Measures must be taken to ensure that the operator cannot touch power terminals or other

high voltage sections.• Any unauthorized transfer or copying of this document, in part or in whole, is prohibited.• Shinko Technos CO., LTD. is not liable for any damage or secondary damage(s) incurred as a

result of using this product, including any indirect damage.SAFETY PRECAUTIONS (Be sure to read these precautions before using our products.)The safety precautions are classified into 2 categories: “Warning” and “Caution”.Depending on the circumstances, procedures indicated by Caution may cause seriousresults, so be sure to follow the directions for usage.

Warning

Warning• To prevent an electric shock or fire, only Shinko or qualified service personnel may handle

the inner assembly.• To prevent an electric shock, fire or damage to the instrument, parts replacement may

only be undertaken by Shinko or other qualified service personnel.

SAFETY PRECAUTIONS• To ensure safe and correct use, thoroughly read and understand this manual before using

this instrument.• This instrument is intended to be used for industrial machinery, machine tools and measuring

equipment. Verify correct usage after consulting purpose of use with our agency or main office.(Never use this instrument for medical purposes with which human lives are involved.)

• External protection devices such as protection equipment against excessive temperature rise,etc. must be installed, as malfunction of this product could result in serious damage to thesystem or injury to personnel. Also proper periodic maintenance is required.

• This instrument must be used under the conditions and environment described in this manual.Shinko Technos Co., Ltd. does not accept liability for any injury, loss of life or damage occurringdue to the instrument being used under conditions not otherwise stated in this manual.

Caution with respect to Export Trade Control OrdinanceTo avoid this instrument from being used as a component in, or utilization in the manufacture ofweapons of mass destruction (i.e. military applications, military equipment, etc.), pleaseinvestigate the end users and the final use of this instrument.In the case of resale, ensure that this instrument is not illegally exported.

Caution

Procedures which may lead to dangerous conditions and causedeath or serious injury, if not carried out properly.

Procedures which may lead to dangerous conditions and causesuperficial to medium injury or physical damage or may degradeor damage the product, if not carried out properly.

Page 4: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

3

11.. IInnssttaallllaattiioonn pprreeccaauuttiioonnss

CautionThis instrument is intended to be used under the following environmentalconditions (IEC61010-1): Overvoltage category , Pollution degree 2Ensure the mounting location corresponds to the following conditions:• A minimum of dust, and an absence of corrosive gases• No flammable, explosive gases• No mechanical vibrations or shocks• No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 )that does not change rapidly, and without icing• An ambient non-condensing humidity of 35 to 85%RH• No large capacity electromagnetic switches or cables through which large current is flowing.• No water, oil or chemicals or where the vapors of these substances can come into

direct contact with the unit• Take note that the ambient temperature of this unit as well as the control panel must

not exceed 50 (122 ) if mounted through the control panel. Otherwise the life ofelectronic components (especially electrolytic capacitors) may be shortened.

Note: Avoid setting this instrument directly on or near flammable material even thoughthe case of this instrument is made of flame-resistant resin.

22.. WWiirriinngg pprreeccaauuttiioonnss

Caution• Do not leave bits of wire in the instrument, because they could cause fire or

malfunction.• Use the solderless terminal with an insulation sleeve in which an M3 screw fits when

wiring the FCD-15A, FCR-15A.• The terminal block of this instrument is designed to be wired from the left side.

The lead wire must be inserted from the left side of the terminal, and fastened withthe terminal screw.

• Tighten the terminal screw within the specified torque. If excessive force is applied tothe screw when tightening, the screw or case may be damaged.

• Do not apply a commercial power source to the sensor connected to the inputterminal nor allow the power source to come into contact with the sensor, as the inputcircuit may be burnt out.

• This controller has no built-in power switch, circuit breaker or fuse. It is necessary toinstall them near the controller.(Recommended fuse: Time-lag fuse, rated voltage 250V AC, rated current 2A)

• For a 24V AC/DC power source, do not confuse polarity when using direct current (DC).

33.. RRuunnnniinngg aanndd mmaaiinntteennaannccee pprreeccaauuttiioonnss

Caution• It is recommended that PID auto-tuning be performed on the trial run.• Do not touch live terminals. This may cause electric shock or operation problems.• Turn the power supply OFF before retightening the terminal or cleaning.

Working or touching the terminal with the power switched ON may result in severeinjury or death due to Electric Shock.

• Use a soft, dry cloth when cleaning the instrument.(Alcohol based substances may tarnish or deface the unit.)

• As the display section is vulnerable, do not strike or scratch it with a hard objector press hard on it.

Page 5: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

4

--- CONTENTS ---1 Model

1.1 Model …………………………………………………………… 61.2 How to read the model label ………………………………………….. 7

2 Name and functions of the sections2.1 Name and displays …………………………………………………….. 8

3 Mounting to the control panel3.1 Site selection …………………………………………………… 103.2 External dimensions (FCD-15A) (Scale: mm)………………………. 103.3 Panel cutout (FCD-15A) (Scale: mm) ………………………………. 103.4 External dimensions (FCR-15A) (Scale: mm)………………………. 113.5 Panel cutout (FCR-15A) (Scale: mm) ……………………..………… 113.6 Mounting ………………………………………………………………. 12

4 Wiring4.1 Terminal arrangement ……………….………………………………. 144.2 Lead wire solderless terminal ……………………………………….. 164.3 Wiring examples ……………………………………………………… 16

5. Setup5.1 Taking the inner assembly out ……………………………………….. 185.2 Switch setting (multi-function) ………………………………….. 185.3 Insertion of the internal assembly ………………………………. 20

6. Operation6.1 Operation flowchart ……………………………………………… 226.2 Main setting mode ……………………………………………… 24

SV …………………………………………………………….. 246.3 Sub setting mode

Set value memory number selection …………………………… 24AT (Auto-tuning) Perform/Cancel ……………………………… 24Proportional band ……………………………………………... 24Integral time …………………………………………………… 24Derivative time ………………………………………………… 24Open/Closed output dead band ………………………………… 24A1 value……………………………………………………….. 25A3 value……………………………………………………….. 25A4 value ………………………………………………………. 25LA (Loop break alarm) time……………………………………. 25LA (Loop break alarm) span …………………………………… 25

6.4 Auxiliary function setting mode 1Set value lock …………………………………………………. 26SV high limit …………………………………………………… 26SV low limit ……………………………………………………. 27Sensor correction ……………………………………………… 27Remote/Local setting ………………………………………….. 27Instrument number …………………………………………….. 27Communication speed ………………………………………… 27

6.5 Auxiliary function setting mode 2Scaling high limit …………………………………………………….. 28Scaling low limit ……………………………………………………… 28Decimal point place………………………………………………….. 28PV filter time constant……………………………………………….. 28A3 type …………………………………………………….…………. 28A4 type .…………………………………………………….………… 29A1 hysteresis ………………………………………………………… 29

Page 6: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

5

A3 hysteresis …………………………………………….………….. 29A4 hysteresis ……………………………………………..………….. 29A1 action delayed timer ……………………………………………. 29A3 action delayed timer ………………………………….…………. 29A4 action delayed timer ………………………………….…………. 29External setting input high limit …………………………………….. 29External setting input low limit ..………………………………........ 30Transmission output …………………………………………………. 30Transmission output high limit ……………………………………… 30Transmission output low limit ………………………………………. 30Indication selection when control output is OFF …………………. 30MV computation cycle ………………………………....................... 30SV rise rate …………………………………………………………… 31SV fall rate ……………………………………………………………. 31Overshoot suppression function selection ………………………… 31

6.6 Open/Closed output time setting modeOpen output time …………………………………………………….. 32Closed output time …………………………………………………… 32

6.7 Program mode ………………………………………………………… 33Fixed value control/Program control ……………………………….. 34Steps 1 to 7 time ……………………………………………………… 34

6.8 Auto/Manual control …………………………………………………… 356.9 Control output OFF function …………………………………………. 356.10 Open/Closed output MV, Remaining step time indication ………. 35

7 Set value memory function (SM option)…………………… 368 Running

8.1 When using the instrument as a Temperature controller …………. 378.2 When using the instrument as a Simplified program controller ….. 38

9 Action explanations9.1 Standard action ……………………………………………………….. 399.2 Open/Closed output dead band action ……………………………… 399.3 Pattern end action …………………………………………………….. 409.4 Alarm 1 (A1), Alarm 3 (A3) and Alarm 4 (A4) action ………………. 41

10. Control actions10.1 Fuzzy self-tuning ………..…………………………………………… 4310.2 Explanations of PID …………………………………………………. 4410.3 PID auto-tuning of this controller …………………………………… 45

11. Other functions …………………………………………………….. 4612. Specifications

12.1 Standard specifications ……………………………………………… 4712.2 Optional specifications ………………………………………………. 50

13. Troubleshooting ……………………………………………………. 5214. Character table ……………………………………………………… 54

Characters used in this manual

Indication

Number, / -1 0 1 2 3 4 5 6 7 8 9

Indication

Alphabet A B C D E F G H I J K L M

Indication

Alphabet N O P Q R S T U V W X Y Z

Page 7: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

6

1 Model1.1 Model

Standard specifications

F C - 1 5 A - R / MD FCD-15A: 96 x 96 x 110mm (W x H x D)SeriesR FCR-15A: 48 x 96 x 110mm (W x H x D)

Control action (OUT1) 5 ON/OFF servo *1Alarm A Alarm action *2Output R Relay contact outputInput M Multi-range input *3

*1: Fuzzy self-tuning PID and PID action are selectable by internal DIP switch.*2: 12 types of alarm and no alarm action are selectable by internal rotary switch

and DIP switch.*3: 15 types of input (Thermocouple, RTD, DC Current and Voltage) are selectable

by internal rotary switch and DIP switch.

Alphanumeric characters to represent functions or types are entered in the .(Example)F C D -1 5 A - R / M, SA , TV

Transmission output (DC voltage)Alarm 3, 4 outputMulti-range inputRelay contact output

Optional specifications

Code SpecificationsSA Alarm 3 (A3), Alarm 4 (A4) output (only for FCD-15A)TATV

Transmission output: PV, SV, Open/Closed output MVTA: DC current (4 to 20mA DC)TV: DC voltage (0 to 1V DC)

C5C

C5: Serial communication (EIA RS-485)C: Serial communication (EIA RS-232C)

SM Set value memory number external selectionEAEV

EA: External setting; DC current (0 to 20mA DC, 4 to 20mA DC)EV: External setting; DC voltage (0 to 1V DC, 1 to 5V DC)

LA Loop break alarmBK Color: BlackBL Screw type mounting brackets (only for FCR-15A)IP Dust-proof/Drip-proof (IP54)TC Terminal cover

(See page 50 for the options in detail.)

Page 8: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

7

1.2 How to read the model label

WarningDo not take the inner assembly out nor touch the terminals with the power

supply ON.Touching the terminals with the power switched ON may result in severe

injury or death due to Electric Shock.

Model labels are attached to the case and the left side of the inner assembly.

(Model label) (Example)

FCD–15A–R/MSA, TV

No.

(1): Model(2): Option codes(3): Serial number (Indicated only on inner assembly)

(1)

(2)

(3)

Relay contact output/Multi-range inputAlarm 3 (A3), Alarm 4 (A4) output,Transmission output (voltage)

Page 9: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

8

2 Name and functions of the sections2.1 Name and displays

[FCD-15A]

(11)

(4) (12)(5) (13)(6)(7) (14)(8) (15)(9) (16)

(10) (17)

[FCR-15A]

(1)

(2)

(3)

(7)

(8)

(9)

(10)

(4)

(5)

(6)

(11)

(14)

(15)

(16)

(17)

(3) (2) (1)

Page 10: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

9

(1) PV display

Indicates the PV (process variable) with a red LED.(2) SV display

Indicates the SV (desired value), Open/Closed output MV, TIME or set values (in thesetting modes) with a green LED.

(3) MEMO display

Indicates the Set value memory number with a yellow LED.(4) SV indicator

The green LED lights when the SV is displayed on the SV display.(5) MV indicator

The red LED lights when the Open/Closed output MV is displayed on the SV display.(6) TIME indicator

The yellow LED lights when Time is displayed on the SV display.(7) MANUAL control indicator

The red LED lights for Manual control.(8) Control motor OPEN output ON indicator

The green LED lights when the Control motor OPEN output is ON.(9) Control motor CLOSED output ON indicator

The yellow LED lights when the Control motor CLOSED output is ON.(10) Serial communication indicator (Option)

The green LED lights during serial communication (TX).(11) A1 indicator (including Pattern end output)

The red LED lights when the Alarm 1 (A1) output or Pattern end output is ON.(12) A3 indicator (for FCD-15A) (Option)

The red LED lights when the Alarm 3 (A3) output is ON.(13) A4 indicator (for FCD-15A) (Option)

Then red LED lights when the Alarm 4 (A4) output is ON.(14) REMOTE indicator (Option)

The red LED lights during Remote action.(15) SB (Sensor burnout) indicator

The red LED lights when the Sensor is burnt out.(16) LA (Loop break alarm) indicator (Option)

The red LED lights if any problems occur such as a Sensor burnout or Control motortrouble has been found on the loop.

(17) AT (Auto-tuning) indicator

The yellow LED flashes during auto-tuning.

KeysMain functions are described below, however, the key has other functions in each mode.Refer to Section 6.1 Operation flowchart.

Increase key: Increases the numeric value on the SV display during settingmode. If pressed continuously, it makes the numerical value change faster.

Decrease key: Decreases the numeric value on the SV display during settingmode. If pressed continuously, it makes the numerical value change faster.

Mode key: Selects a setting mode.OUT/OFF key: Turns the control output ON or OFF.

Starts or stops the program control.Auto/Manual key: Switches Automatic or Manual control.Fast key: This makes the numerical value change faster by pressing the

key and the (or ) key simultaneously.

Page 11: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

10

96 10010106 (for TC option)

9110

6.2

3. Mounting to the control panel3.1 Site selection

This instrument is intended to be used under the following conditions (IEC61010-1):Overvoltage category , Pollution degree 2Ensure the mounting location corresponds to the following conditions:(1) A minimum of dust, and an absence of corrosive gases(2) No flammable, explosive gases(3) No mechanical vibrations or shocks(4) No exposure to direct sunlight, an ambient temperature of 0 to 50 (32 to 122 )

that does not change rapidly, and without icing(5) An ambient non-condensing humidity of 35 to 85%RH(6) No large capacity electromagnetic switches or cables through which large current

is flowing(7) No water, oil or chemicals or where the vapors of these substances can come into

direct contact with the unit(8) Take note that the ambient temperature of this unit as well as the control panel must

not exceed 50 (122 ) if mounted through the control panel. Otherwise the life ofelectronic components (especially electrolytic capacitor) may be shortened.

3.2 External dimensions (FCD-15A) (Scale: mm)

(Fig. 3.2-1)3.3 Panel cutout (FCD-15A) (Scale: mm)

Lateral close mountingn: Number of units mounted

(Fig. 3.3-1) □92 +0.80

92+0

.80 13

0

n x 96-3 +0.50

Page 12: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

11

48

96

10 100106 (For TC option)

91 109

48

96

10 100106 (For TC option)

9110

6.2

(For

BLop

tion)

3.4 External dimensions (FCR-15A) (Scale: mm)

One-touch type mounting brackets

(Fig. 3.4-1)Screw type mounting brackets

(Fig. 3.4-2)3.5 Panel cutout (FCR-15A) (Scale: mm)

Lateral close mountingn: Number of units mounted

(Fig. 3.5-1)

n x 48-3 +0.50

92+0

.8 0

130

92+0

.8 0

45 +0.50

Page 13: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

12

3.6 Mounting

NoticeAs the case is made of resin, do not use excessive force while screwing in themounting bracket. The torque should be approximately 0.12N•m.

[FCD-15A]

Insert the FCD-15A from the front of the panel. Slot the mounting bracketinto the holes at the top and bottom of the case, and screw in place.Mounting panel thickness: 1 to 8mm

(Fig. 3.6-1)

[FCR-15A]

Catch the mounting bracket 1 to thetop and bottom of the instrument first.Then, insert the FCR-15A 2 from thefront of the mounting panel.Mounting panel thickness: 1 to 3mm.When the Soft-type Front cover (FC-R)is used, mounting panel thickness is1 to 2.5mm.

(Fig. 3.6-2)

UP

UP

Page 14: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

13

When using screw type mounting brackets (BL option)

NoticeAs the case is made of resin, do not use excessive force while screwing inthe mounting bracket. The torque should be approximately 0.12N•m.

Insert the FCx-15A from the front of the panel. Slot the mounting bracketinto the holes at the top and bottom of the case, and screw in place.Mounting panel thickness: 1 to 8mm

(Fig. 3.6-3)

UPUP

Page 15: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

14

22

28

29

30

31

34

35

32

33

38

39

36

37

1

4

COM

7

8

9

10

100 to 240V ACor 24V AC/DC

2

18

19

18

19

11

12

14

15

16

TC

RTD

Ground

5

A1/Pattern end

25

26

27

21

23

24

R/L

A/M

COM

EA or EV

TA or TV

+

+

RS-232C[RS-485]

TX [YA(-)]

RX [YB(+)]

b0

b1

b2

COM

B

B

A

SM

A4

A3

+18

19

+V DC

18

19

+

mA DC

17

3 13

6

17

20

OPEN

CLOSE

OpenoutputClosedoutput

+

4. Wiring

WarningTurn the power supply to the instrument off before wiring or checking.Working or touching the terminal with the power switched on may result insevere injury or death due to Electric Shock.Moreover, the instrument must be grounded before the power supply to theinstrument is turned on.

Caution• The terminal block of this instrument is designed to be wired from the left side.

The lead wire must be inserted from the left side of the terminal, and fastened withthe terminal screw.

• Use a thermocouple and compensating lead wire according to the sensor inputspecifications of this controller.

• Use a 3-wire RTD according to the sensor input specifications of this controller.• This controller has no built-in power switch or fuse. It is necessary to install them in

the circuit near the external controller.(Recommended fuse: Time-lag fuse, rated voltage 250V AC, rated current 2A)

• For a 24VAC/DC of power source, do not confuse polarity when using a direct current (DC).• When wiring, keep input wires (thermocouple, RTD, etc.) away from AC sources or

load wires to avoid external interference.• Use a thick wire (1.25 to 2.0mm2) for grounding.• Use an auxiliary electromagnetic switch between the FCx-15A and the control motor.• Dotted lines show options. Option terminals are equipped only when they are added.

4.1 Terminal arrangement

[FCD-15A]

(Fig. 4.1-1)

Page 16: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

15

Ground 1

4

7

8

9

10

100 to 240V ACor24V AC/DC

2

18

19

18

19

11

12

14

15

16

TC

RTD

5

A1/Pattern end

RS-232C[RS-485]

TX [YA(-)]

RX [YB(+)]

b0

b1

b2

COM

B

B

A

SM

+18

19

+V DC

18

19

+

mA DC

17

3 13

6

17

20

+

COM

OPEN

CLOSE

EA or EV

+

+

TA or TV

Open output

Closed output

[FCR-15A]

(Fig. 4.1-2)A1 : Alarm 1 (A1) outputPattern end: Pattern end outputTA, TV : Transmission outputEA, EV : External setting inputR/L : Remote/Local setting (External operation input) (for FCD-15A)A/M : Auto/Manual control (External operation input) (for FCD-15A)A3 : Alarm 3 (A3) output (for FCD-15A) (option)A4 : Alarm 4 (A4) output (for FCD-15A) (option)RS-232C : Serial communication EIA RS-232CRS-485 : Serial communication EIA RS-485SM : Set value memory external selectionTC : Thermocouple inputRTD : RTD inputV DC : DC voltage inputmA DC : DC current input

Page 17: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

16

4.2 Lead wire solderless terminal

Use a solderless terminal with an insulation sleeve in which an M3 screw fits as shownbelow. The torque should be approximately 0.63N・m.Solderless terminal Manufacturer Model Tightening torque

Nichifu Terminal Industries CO.,LTD. TMEV1.25Y-3Y typeJapan Solderless Terminal MFG CO.,LTD. VD1.25-B3ANichifu Terminal Industries CO.,LTD. TMEV1.25-3Round typeJapan Solderless Terminal MFG CO.,LTD. V1.25-3

0.63N•m

(Fig. 4.2-1)

4.3 Wiring examplesWiring example of OPEN/CLOSED output

(Fig. 4.3-1)

* To prevent the unit from harmful effects of the unexpected level noise,it is recommended that a surge absorber be installed between the coilsof the external relay.

5

7

6

19

18

OPEN output

CLOSED output *

Power supply for motor drive

*

Electromagneticswitch

Valve

Input

Control motor

5.8m

mor

less

3.2mm3.2mm

5.8m

mor

less

Furnace

+

-

COM

COMOPEN

CLOSED

Electromagneticswitch

Page 18: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

17

Wiring example of External operation input (Only for FCD-15A)

Contact input252627

(Fig 4.3-2)

Remote/Local, Auto/Manual switching

Setting item External operation Key operationRemote setting 25 - 27 Closed Not required (invalid)Local setting 25 - 27 Open See p.27.Manual control 26 - 27 Closed Not required (invalid)Automatic control 26 - 27 Open See p.35.

External operation has priority.

Wiring example of External setting (EA, EV), Transmission output (TA, TV)(only for FCR-15A)

(Fig. 4.3-3)

R/LA/MCOM

Page 19: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

18

5. Setup5.1 Taking the inner assembly out

Before the power supply to this instrument is turned on, take the internal assemblyout from the case by pushing the hook (bottom of the instrument) in the directionas shown by the arrow and by holding the latches.

(Fig. 5.1-1) (Fig. 5.1-2)

5.2 Switch setting (multi-function)

Using a small flat-head screwdriver or tweezers, set the Input type, Alarm 1 type,

Control action, Heating (reverse)/Cooling (direct) control action, Alarm 1 standby

function, / and Program start Auto/Manual by the rotary switch and DIP switch asfollows.

Control action Alarm 1 typeHeating/Cooling control action Input typeAlarm 1 standby action

/Input type

Program start Auto/Manual

(Fig. 5.2-1)

DIP switch

Rotary switch

Page 20: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

19

The following items can be selected by the DIP switch (SW304).Default value: All switches OFF [ ].(Table 5.2-1)

ItemDIP SW304

No.Selection Status

Fuzzy self-tuning PID action OFFControl action 1 PID action ONHeating (reverse) control action OFFHeating/Cooling

control action 3 Cooling (direct) control action ONStandby Disabled OFFAlarm 1 (A1)

standby action 4 Standby Enabled ONOFF/ 6 ON

K, J, R, B, N, PL- , Pt100, JPt100 OFFInput type *1 7 S, E, T, C, 4 to 20mA, 0 to 20mA, 0 to 1V ONManual start OFFProgram start

Auto/Manual *2 8 Automatic start ON

Note: Switches No. 2 and 5 are not used. They will not function even if they areset to ON side.

*1: Use the rotary switch (SW303) as well.*2: If Automatic start is selected, the unit proceeds to warm-up status for approx. 2

seconds after power-on, then program control starts automatically from step 1.If Manual start is selected, the unit proceeds to warm-up status for approx. 2

seconds after power-on, then proceeds to standby status. In this status,if the key is pressed, program control starts from step 1.

Select the input type with the rotary switch (SW303) and DIP switch (SW304).Default: K (-200 to 1370 ) [ ].

(Table 5.2-2)RotarySW303.

No.

DIPSW304No. 7

Input type Scale range

0 OFF K -200 to 1370 -320 to 25001 OFF J -200 to 1000 -320 to 18002 OFF R 0 to 1760 0 to 32003 OFF B 0 to 1820 0 to 33004 OFF PL- 0 to 1390 0 to 25005 OFF N 0 to 1300 0 to 23006 OFF Pt100 -199.9 to 850.0 -199.9 to 999.97 OFF JPt100 -199.9 to 500.0 -199.9 to 900.00 ON S 0 to 1760 0 to 32001 ON E 0 to 1000 0 to 18002 ON T -199.9 to 400.0 -199.9 to 750.03 ON C (W/Re5-26) 0 to 2315 0 to 42004 ON 4 to 20mA DC -1999 to 99995 ON 0 to 20mA DC -1999 to 99996 ON 0 to 1V DC -1999 to 99997 ON Pt100 -200 to 850 -320 to 1560

Page 21: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

20

Alarm type and the pattern end output can be selected by the rotary switch (SW302).

When using the FCx-15A as a Simplified program controller, the Pattern end outputcan be selected instead of Alarm 1 (A1) type.Default value: No alarm action, [ ].

(Table 5.2-3)

Alarm 1 (A1) typeRotary

SW302 No.

No alarm action 0High limit alarm 1Low limit alarm 2High/Low limits alarm 3High/Low limit range alarm 4Process high alarm 5Process low alarm 6Pattern end output 7

5.3 Insertion of the internal assembly

When the setup is completed, insert the internal assembly into the case.Surely insert the assembly until it is locked by the hook at the bottom of the instrument.(Click sound audible)

CautionDo not confuse the top and bottom of the internal assembly.If inserting the assembly into the case forcibly in the wrong direction, the PCBmay be damaged.

Page 22: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

21

Note before operating keys

• Pressing the key for approx. 1 second from any mode turns the control output OFF.• Once the control output OFF function is enabled, the function cannot be released

even if the power to the instrument is turned OFF and ON again.To cancel the function, press the key again for approx. 1 second.

• Pressing the key registers the set value and proceeds to the next setting item.If the key is pressed for approx. 3 seconds from any mode, the unit reverts to thePV/SV display mode.

Turn the Power ON

The PV display indicates the input type, and the SV display indicates input range highlimit value selected during Input type selection for approx. 2 seconds after the power isturned on. See (Table 6.1-1) below.During this time, all outputs and LED indicators are in OFF status.Control will then start indicating the PV (process variable) on the PV display and SV(desired value) on the SV display.

(Table 6.1-1)

Input PV display SV display PV display SV displayKJRB

PL-NSETC

Pt100JPt100

4 to 20mA DC0 to 20mA DC0 to 1V DC

Scaling highlimit value

Scaling highlimit value

Page 23: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

22

6. Operation6.1 Operation flowchart

(Approx.1sec) Control outputOFF function

Manual controlPV/SV display

Open/Closed Output MVindication

Remaining steptime indication *1

[Main setting mode] [Sub setting mode] [Auxiliary function setting mode 1]

SV[ ]

Set value memory number[ ]

Set value lock[ ]

AT Perform/Cancel[ ]

SV high limit[ ]

Proportional band[ ]

SV low limit[ ]

Integral time[ ]

Sensor correction[ ]

Derivative time[ ]

Remote/Local[ ]

Open/Closed outputdead band [ ]

Instrument number[ ]

A1 value[ ]

Communication speed[ ]

A3 value[ ]

A4 value[ ]

Loop break alarm time[ ]

Loop break alarm span[ ]

Warm-up status lasts 2sec after power-on.*1: Remaining step time can be indicated

during program control.*2: If AT is performed, the unit reverts to the

PV/SV display mode.If AT is cancelled, the unit proceeds tothe “Proportional band setting”.

+ + 3sec

3sec

• : This means that if the keyis pressed, the unit proceeds to thenext setting item mode.

• Setting items with dotted lines areoptional, and they appear onlywhen the options are added.

• Alarm 3 (A3) and 4(A4) setting itemsare available for the FCD-15A.

*2

Page 24: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

23

*3: If (Fixed value control) is selected during “Fixed value control/Program control selection” andthe key is pressed, the controller reverts to the PV/SV display mode.

+ : Press the key while holding down the key.+ 3sec: Press the key for approx. 3 seconds while holding down the key.+ + 3sec: Press the key for approx. 3sec while holding down the and keys.+ + 3sec: Press the key for approx. 3sec while holding down the and keys.+ 3sec: Pressing the key, press the key for approx. 3 seconds.

[Auxiliary function [Open/Closed output [Program mode]

setting mode 2] setting mode]Scaling high limit[ ]

Transmission outputhigh limit [ ]

Open outputtime [ ]

Fixed value control/Program control[ ] *3

Scaling low limit[ ]

Transmission outputlow limit [ ]

Closed outputtime [ ] Step 1 time

[ ]Decimal point place[ ]

Indication selection whencontrol output is off[ ]

Step 2 time[ ]

PV filter time constant[ ] MV computation cycle

[ ]Step 3 time[ ]

A3 type[ ] SV rise rate

[ ]Step 4 time[ ]

A4 type[ ] SV fall rate

[ ]Step 5 time[ ]

A1 hysteresis[ ] Step 6 time

[ ]A3 hysteresis[ ] Step 7 time

[ ]A4 hysteresis[ ]

A1 action delayedtimer [ ]

A3 action delayedtimer [ ]

A4 action delayedtimer [ ]

External setting inputhigh limit [ ]

External setting inputlow limit [ ]

Transmission output[ ]

+ + 3sec + 3sec+ + 3sec

Page 25: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

24

6.2 Main setting mode

In the PV/SV display mode, if the key is pressed, the mode proceeds to theMain setting mode.The SV can be increased or decreased normally by pressing the , , andfaster with the key.If the key is pressed, the SV will be registered and the controller will revert tothe PV/SV display mode.

[ ] SV• Sets SV.• Setting range: SV low limit to SV high limit• Default: 0

6.3 Sub setting mode

In the PV/SV display mode, if the key is pressed while holding down thekey, the unit proceeds to the Sub setting mode.The set value can be increased or decreased normally by pressing the , andfaster with the key.Pressing the key registers the set value, and proceeds to the next setting item.

[ ] Set value memory number• Selects the memory number to be set or to be retrieved.• Selection range: 1 to 7• Default: Memory number 1[ ] AT (Auto-tuning) Perform/Cancel• Sets AT (Auto-tuning) Perform or Cancel.

If AT Perform is selected and the key is pressed, the controller reverts tothe PV/SV display mode.

• Selection item: : AT Cancel: AT Perform

• Default: AT Cancel[Auto-tuning]• Refer to Section 10.3.• During auto-tuning, none of the settings can be performed.• If the auto-tuning is cancelled during the process, P, I, D values revert to their

previous value.[ ] Proportional band• Sets the proportional band.• Setting range: 0.1 to 999.9%• Default: 2.5%[ ] Integral time• Sets the integral time.• Setting range: 1 to 3600 seconds• Default: 200 seconds[ ] Derivative time• Sets the derivative time.Setting the value to 0 disables the function. (PI action)• Setting range: 0 to 3600seconds• Default: 50 seconds[ ] Open/Closed output dead band• Sets the dead band value of the Open output and Closed output• Setting range: 0.1 to 100.0% of the proportional band• Default: 3.0%

Page 26: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

25

[ ] A1 value• Sets Alarm 1 (A1) value.

Setting the value to 0 or 0.0 disables the function.(Excluding process high alarm and process low alarm)

• Not available if Rotary switch (SW302) is set to No.0 or No.7.• Setting range: Refer to (Table 6.3-1).• Default: 0[ ] A3 value (for FCD-15A)• Sets Alarm 3 (A3) value.

Setting the value to 0 or 0.0 disables the function.(Excluding process high alarm and process low alarm)

• Not available if Alarm 3 (SA option) is not added or if [ ] is selected duringAlarm 3 (A3) type selection.

• Setting range: Refer to (Table 6.3-1).• Default: 0[ ] A4 value (for FCD-15A)• Sets Alarm 4 (A4) value.

Setting the value to 0 or 0.0 disables the function.(Excluding process high alarm and process low alarm)

• Not available if Alarm 4 (SA option) is not added or if [ ] is selected duringAlarm 4 (A4) type selection.

• Setting range: Refer to (Table 6.3-1).• Default: 0[ ] LA (Loop break alarm) time• Sets the time to assess the Loop break alarm.• Available only when Loop break alarm (LA option) is added• Setting range: 0 to 200 minutes• Default: 0 minutes[ ] LA (Loop break alarm) span• Sets the span to assess the Loop break alarm.• Available only when Loop break alarm (LA option) is added• Setting range:

Thermocouple except T: 0 to 100T input, RTD input: 0.0 to 100.0DC input: 0 to 1000 (The placement of the decimal point follows the selection.)

• Default: 0

[Setting range of A1, A3 and A4] (including standby function)(Table 6.3-1)

Alarm type Setting rangeHigh limit alarm –Input span to Input span *1Low limit alarm –Input span to Input span *1High/Low limits alarm 0 to Input span *1High/Low limit range alarm 0 to Input span *1Process high alarm Input range low limit to Input range high limit *2Process low alarm Input range low limit to Input range high limit *2

When input has a decimal point, the negative low limit value is -199.9.For DC input, the negative low limit value is –1999. (The placement of the decimalpoint follows the selection.)*1: For DC input, the Input span is the same as the scaling span.*2: For DC input, Input range low (or high) limit value is the same as the Scaling low

(or high) limit value.

Page 27: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

26

6.4 Auxiliary function setting mode 1

In the PV/SV display mode, if the key is pressed for approx. 3 seconds whileholding down the key, the unit proceeds to Auxiliary function setting mode 1.The set value can be increased or decreased normally by pressing the , ,and faster with the key.Pressing the key registers the set value, and proceeds to the next setting item.

[ ] Set value lock• Locks the set value to prevent setting errors.The setting item to be locked differs depending on the selection.• When selecting Lock, set the necessary items in the status Unlock,then select Lock 1, Lock 2 or Lock 3.• Fuzzy self-tuning PID or Auto-tuning (AT) will not function if Lock 1 or Lock 2is selected.[About Lock 3]When using the unit as a Fixed value controllerThe set values of the selected Set value memory number can be changedtemporarily. However, if the memory number is changed, the changed valuesof the previous number are cancelled. (Returns to the values before setting.)When using the unit as a Program controllerThe set values of the currently performing step number can temporarily bechanged. However, if the step number is changed, the changed set values arecancelled. (Returns to the values before setting.)Set value change during running standby status will be invalidated.(Running starts at the values memorized in the memory.)

• Be sure to select Lock 3 when changing the set value frequently viacommunication function considering the life of non-volatile memory.

• Selection item:(Unlock): All set values can be changed.(Lock 1): None of the set values can be changed.(Lock 2): Only SV can be changed.(Lock 3): All set values can be changed temporarily. However, changedvalues revert to their previous value after the power is turned off becausethey are not saved in the non-volatile memory.This function is suitable when changing value temporarily.Since this function has no relation to the memory life, it is well suited whenchanging set values frequently using Shinko programmable controller(PC-900, PCD-33A with SVTC option).Do not change any setting item in Auxiliary function setting mode 2. If anyitem in the mode is changed, it will affect other setting items such as theSV and Alarm value.

• Default: Unlock[ ] SV high limit• Sets SV high limit value.• Setting range: SV low limit to Input range high limit value(For DC input, the placement of the decimal point follows the selection.)DC input: SV low limit to Scaling high limit value

• Default: 400

Page 28: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

27

[ ] SV low limit• Sets SV low limit value.• Setting range: Input range low limit value to SV high limit

(The placement of the decimal point follows the selection.)For DC input: Scaling low limit to SV high limit(The placement of the decimal point follows the selection.)

• Default: 0[ ] Sensor correction• Sets the sensor correction value.• Setting range: –100.0 to 100.0 ( )• Default: 0.0Sensor correction function

This corrects the input value from the sensor. When a sensor cannot be set atthe exact location where control is desired, the sensor measured temperaturemay deviate from the temperature in the controlled location.When controlling with plural controllers, sometimes the measured temperatures(input value) do not concur due to differences in sensor accuracy or dispersion ofload capacities. In such a case, the control can be set at the desired temperatureby adjusting the input value of sensors.However, it is effective within the input rated range regardless of the sensorcorrection value.PV after sensor correction= Current PV+ (Sensor correction value)

[ ] Remote/Local• SV can be set with either Remote or Local method.• Available only when External setting (EA, EV option) is added.• Selection item:

(Local setting) : The SV can be set by the front keypad operation.(Remote setting) : The SV can be set in analog by the remote operation

externally.• Default: Local setting[ ] Instrument number• Sets the instrument number of this unit. (The instrument number should be set

individually when communicating by connecting plural instruments in serialcommunication. Otherwise communication is impossible.)

• Available only when the Serial communication (C, C5 option) is applied• Setting range: 0 to 95• Default: 0[ ] Communication speed• Selects the communication speed of this instrument.(The communication speed of this instrument must be equal to that of the hostcomputer, otherwise communication is impossible.)

• Available only when the Serial communication (C, C5 option) is applied• Selection item:

(2400bps)(4800bps)(9600bps)(19200bps)

• Default: 9600bps

Page 29: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

28

6.5 Auxiliary function setting mode 2

In the PV/SV display mode, if the key is pressed for approx. 3 seconds while holdingdown the and keys, the unit proceeds to Auxiliary function setting mode 2.The set value can be increased or decreased normally by pressing the , ,and faster with the key.Pressing the key registers the set value, and proceeds to the next setting item.

[ ] Scaling high limit• Sets scaling high limit value.Not available if thermocouple or RTD is selected during Input type selection.

• Setting range: Scaling low limit value to Input range high limit value• Default: 9999[ ] Scaling low limit• Sets scaling low limit value.Not available if thermocouple or RTD is selected during Input type selection.

• Setting range: Input range low limit value to scaling high limit value• Default: –1999[ ] Decimal point place

• Selects the decimal point place.• Not available if thermocouple or RTD is selected during Input type selection• Selection item:

(No decimal point)(1 digit after the decimal point)(2 digits after the decimal point)(3 digits after the decimal point)

• Default: No decimal point[ ] PV filter time constant• Sets PV filter time constant.

When PV fluctuates due to disturbance, find a value which does not fluctuate bygradually increasing the value. However, if the value is set too large, it affectscontrol result due to the delay of response.

• Setting range: 0.0 to 10.0 seconds• Default: 0.0 seconds[ ] A3 type (for FCD-15A)• Selects the Alarm 3 (A3) type.• Available only when Alarm 3 (SA option) is added• Default: No alarm action• Selection item

: No alarm action: High limit alarm: High limit alarm with standby: Low limit alarm: Low limit alarm with standby: High/Low limits alarm: High/Low limits alarm with standby: High/Low limit range alarm: High/Low limit range alarm with standby: Process high alarm: Process high alarm with standby: Process low alarm: Process low alarm with standby

Page 30: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

29

[ ] A4 type (for FCD-15A)

• Selects the Alarm 4 (A4) type.• Available only when Alarm 4 (SA option) is added• Selection items are the same as those of Alarm 3 (A3) type selection.• Default: No alarm action[ ] A1 hysteresis• Sets Alarm 1 (A1) hysteresis.• Not available if Rotary switch (SW302) is set to No.0 or No.7 during A1 type selection.• Setting range: Thermocouple, RTD input: 0.1 to 100.0

DC input: 1 to 1000 (The placement of the decimal point followsthe selection.)

• Default: 1.0[ ] A3 hysteresis (for FCD-15A)• Sets Alarm 3 (A3) hysteresis.• Not available if Alarm 3 (SA option) is not added, or if [ ] is selected during

Alarm 3 type selection even if SA option is added• Setting range and default value are the same as those of A1 hysteresis setting.[ ] A4 hysteresis (for FCD-15A)• Sets Alarm 4 (A4) hysteresis.• Not available if Alarm 4 (SA option) is not added, or if [ ] is selected during

Alarm 4 type selection even if SA option is added• Setting range and default value are the same as those of A1 hysteresis setting.[ ] A1 action delayed timer setting• Sets A1 action delayed timer.

When setting time has elapsed after the input enters the alarm output range, thealarm is activated.

• Not available if the Rotary switch (SW302) is set to No.0 or No.7 during A1 typeselection.

• Setting range: 0 to 9999 seconds• Default: 0 seconds[ ] A3 action delayed timer (for FCD-15A)• Sets A3 action delayed timer.When setting time has elapsed after the input enters the alarm output range, thealarm is activated.

• Not available if Alarm 3 (SA option) is not applied, or if [ ] is selected duringAlarm 3 type selection even if SA option is added

• Setting range and default value are the same as those of A1 action delayed timersetting.

[ ] A4 action delayed timer (for FCD-15A)• Sets A4 action delayed timer.When setting time has elapsed after the input enters the alarm output range, thealarm is activated.

• Not available if Alarm 4 (SA option) is not applied, or if [ ] is selected duringAlarm 4 type selection even if SA option is added

• Setting range and default value are the same as those of A1 action delayed timersetting.

[ ] External setting input high limit• Sets the high limit value for external setting input.For EA option (4 to 20mA), the value corresponds to 20mA input.

• Available only when External setting (EA or EV option) is added• Setting range: External setting input low limit value to Input range high limit value• Default: 400

Page 31: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

30

[ ] External setting input low limit

• Sets the low limit value for external setting input.For EA option (4 to 20mA), the value corresponds to 4mA input.

• Available only when External setting (EA or EV option) is added• Setting range: Input range low limit value to External setting input high limit value• Default: 0[ ] Transmission output

• Selects the Transmission output type.• Available only when Transmission output (TA or TV option) is added• Selection item:

: PV (process variable) transmission: SV (desired value) transmission: Open/Closed output MV (manipulated variable) transmission

• Default: PV transmission[ ] Transmission output high limit• Sets the Transmission output high limit value.

For TA option, the value correponds to 20mA output.• Available only when Transmission output (TA or TV option) is added• Transmission output during control standby:

PV transmission: PVSV transmission: 0Open/Closed output MV transmission: 0

• Setting range: Transmission output low limit value to Input range high limit value• Default: 400[ ] Transmission output low limit

• Sets the Transmission output low limit value.For TA option, the value correponds to 4mA output.

• Available only when Transmission output (TA or TV option) is applied• Transmission output during control standby:

PV transmission: PVSV transmission: 0Open/Closed output MV transmission: 0

• Setting range: Input range low limit value to Transmission output high limit value• Default: 0[ ] Indication selection when control output is OFF

• Selects the indication when control output is off.• Selection item:

: OFF is indicated on the PV display.: No indication (all unlit): Only PV is indicated.

• Default:[ ] MV computation cycle

• Sets MV computation cycle for Open/Closed output.• Setting range: 1 to 120sec• Default: 3sec

Page 32: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

31

[ ] SV rise rate

• Sets the SV rise rate (Rising value per minute).Setting the value to 0 or 0.0 disables the function.

• Setting range:Thermocouple except T, RTD without a decimal point : 0 to 9999 /min ( /min)Thermocouple T, RTD with a decimal point : 0.0 to 999.9 /min ( /min)DC input: 0 to 9999 (The placement of the decimal point follows the selection.)

• Default: 0 /minute ( /min)[ ] SV fall rate

• Sets the SV fall rate (Falling value per minute).Setting the value to 0 or 0.0 disables the function.

• Setting range and default value are the same as those of SV rise rate setting.

Page 33: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

32

6.6 Open/Closed output time setting mode

In the PV/SV display mode, if the key is pressed for approx. 3 seconds whileholding down the and keys, the unit proceeds to Open/Closed outputtime setting mode.The set value can be increased or decreased normally by pressing the ,and faster with the key.Pressing the key registers the set value, and proceeds to the next setting item.

[ ] Open output time• Sets time from the valve fully closed to the valve fully open.• Setting range: 0.1 to 999.9sec• Default: 30.0secAutomatic measuring method

(1) Check that the valve is at the fully closed position, and press the andkeys simultaneously.

Time count starts from 0 seconds to measure open output time.Control motor OPEN output will turn ON after count starts, and makes thevalve open.

(2) Press the key when the valve is fully open. Time count stops, open outputtime is registered, and the control motor OPEN output turns OFF.

[ ] Closed output time setting• Sets time from the valve fully open to the valve fully closed.• Setting range: 0.1 to 999.9sec• Default: 30.0secAutomatic measuring method

(1) Check that the valve is at the fully open position, and press the andkeys simultaneously.Time count starts from 0 seconds to measure the closed output time.Control motor CLOSED output will turn ON after count starts, and makes thevalve closed.

(2) Press the key when the valve is fully closed. Time count stops, closedoutput time is registered, and the control motor CLOSED output turns OFF.

Page 34: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

33

6.7 Program mode

In PV/SV display mode, if the key is pressed for approximately 3 seconds whileholding down the key, the unit enters Program mode.The set value can be increased or decreased normally by pressing the ,and faster with the key.Pressing the key registers the set value, and proceeds to the next setting item.

• Each value of Set value memory numbers (1 to 7) set during the Fixed value controlare assigned to the value for Steps 1 to 7.For example, values of Set value memory number 1 become Step 1 values,and values of Set value memory number 2 become Step 2 values.

• If the Pattern end output is selected and program control is performed,the Pattern end output is turned on when the program is completed.

• If the key is pressed while the Pattern end output is on, the Pattern end outputis turned off. If the key is pressed again, the program will be performed.

• Set the step time to 00.00 for the unnecessary steps.

Program Example

Step number (Setvalue memory number) 1 2 3 4 5 6 7

SV 500 500 1000 1000 200 200 100Proportional band 2.0 1.8 2.0 1.8 2.5 1.8 2.0Integral time 180 80 180 80 200 80 200Derivative time 30 20 30 20 50 20 50Open/Closed dead band 1.0 1.0 1.0 1.0 1.0 1.0 1.0Alarm 1 value 2 5 2 5 10 0 5Alarm 3 value 2 5 2 5 10 0 5Alarm 4 value 2 5 2 5 10 0 5Step time 00:30 01:00 00:40 01:00 02:00 00:30 01:00

0

500

1000

Page 35: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

34

[ ] Fixed value control/Program control

• Selects fixed value control or program control.• If the key is pressed in the fixed value control, the controller reverts to thePV/SV display mode.(If the controller is not set to program control mode, Step time setting (from Step 1time) is impossible.• Selection item: (Fixed value control)

(Program control)• Default: Fixed value control

If “ (program control)” is selected during Fixed value control/Program controlselection mode and the key is pressed, Steps 1 to 7 time can be set.

Steps 1 to 7 time setting (indicated only during program control)

[ ] Step 1 time

• Sets Step 1 time. (Available only for program control)• 00.00

MinuteHour

(e.g.) When setting 1 hour 58 minutes, set as [ ].• Setting range: 00.00 to 99.59• Default: 00.00

[ ] Step 2 time• Sets Step 2 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

[ ] Step 3 time• Sets Step 3 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

[ ] Step 4 time• Sets Step 4 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

[ ] Step 5 time• Sets Step 5 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

[ ] Step 6 time• Sets Step 6 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

[ ] Step 7 time• Sets Step 7 time. (Available only for program control)• Setting range and default value are the same as those of Step 1 time setting.

Page 36: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

35

6.8 Auto/Manual control

Manually output MV (manipulated variable) can be changed.As the proportional action, the MV is automatically outputted in proportion to thedeviation. This is an automoatic control.As compared with the above, in the manual control, the value set by front keypadcan be outputted regardless of the deviation.The balance/bumpless function is provided to prevent rapid output change when thecontrol mode is changed from Automatic to Manual and vice versa.Each time the key is pressed, the control changes from Automatic to Manualand vice versa.

Manual control

MV 0.0% : Valve fully ClosedMV 100.0% : Valve fully OpenFor the MV (0.1 to 99.9%) other than 0.0% and 100.0%, Open output ON time andClosed output ON time will be determined by the relation between the MV at thegiven time and set values (Open/Closed output dead band, MV computation cycle,Open output time, Closed output time, etc.).• Setting range: 0.0 to 100.0%

6.9 Control output OFF function

• A function to pause the control action or turn the control output of the unusedinstrument of the plural units OFF even if the power to the instrument is supplied.“ ” is indicated on the PV display while the function is working.

• Pressing the key for approx. 1 second from any mode enables the controloutput OFF function. PV display indicates , No indication or PV previouslyselected during “Indication selection when control output is OFF” mode.To cancel the function, press the key again for approx. 1 second.

• Once the control output OFF function is enabled, the function cannot be releasedeven if the power to the instrument is turned OFF and ON again.To cancel the function, press the key again for approx. 1 second.

• During program control, the key becomes the Program Start/Stop key, sothe control output OFF function is disabled.

6.10 Open/Closed output MV, Remaining step time indication

In the PV/SV display mode, press the key for approx. 3 seconds.Keep pressing the key until the Open/Closed output MV appears, thoughthe main setting mode appears temporarily during the process.The MV indicator lights, and the SV display indicates Open/Closed output MV.In the fixed value control, when the key is pressed again, the mode revertsto the PV/SV display.

In the program control, if the key is pressed while an output MV(manipulatedvariable) is indicated, the unit changes to the Remaining step time indication mode.The TIME indicator lights, and the SV display indicates remaining step time.If the key is pressed again, the mode reverts to the PV/SV display mode.

Page 37: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

36

7. Set value memory function (SM option)If the SM option is applied, Set value memory number can be selected externally.A maximum of 7 files with 8 pieces of data (6 pieces for FCR-15A) can be memorized.Control can be performed with the desired file.In one file, 8 kinds of set value (SV, PID values, Open/Closed output dead band, A1,A3, A4 value) are included.(A3, A4 values are applicable only to the FCD-15A.)To select the set value memory number, connect terminals 13 to 16 as shown below.Up to 50 units can be connected in parallel.

Terminal connection for Set value memory numberSet value

memory No.Connectingterminals

1 2 3 4 5 6 7

13 - 16 (b0-COM) O O O14 - 16 (b1-COM) O O O15 - 16 (b2-COM) O O O

: Closed O : Open• If the Set value memory number is selected externally, the number cannot be selected

by front keypad.• When selecting the memory number by front keypad, all terminals (b0, b1, b2 and

COM) should be open.• Set value memory number cannot be changed during setting mode or AT (auto-tuning).

Operation procedures

(1) In the PV/SV display mode, select a set value memory number by connectingterminals.

(2) In the setting mode, set each value: SV, PID values, Open/Closed output dead band,A1, A3, A4 values

(3) After setting, press the key to revert to the PV/SV display mode.

Registration• Each set value is registered in the file, the number of which is indicated on the MEMO

display.

• When any number is retrieved by connecting terminals, the selected number isindicated, and control is performed using the data (set values) of the indicatedfile number.

• To change set values, repeat the Operation procedures above.

Page 38: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

37

8. RunningAfter the unit is mounted to the control panel and wiring is completed, operate the unitfollowing the procedures below.

8.1 When using the instrument as a Temperature controller

(1) Turn the power supply to the instrument ON.

For approx. 2sec after power-on, the input type and temperature unit selected duringthe Input type selection are indicated on PV display, and the input range high limitvalue is indicated on the SV display. See (Table 8.1-1).During this time, all outputs and LED indicators are in OFF status.After that, the instrument enters the PV/SV display mode.While the Control output OFF function is working, the item selected during“Indication selection when control output is OFF” is displayed on the PV display.

(Table 8.1-1)

Input PV display SV display PV display SV displayKJRB

PL-NSET

C(W/Re5-26)

Pt100

JPt1004 to 20mADC0 to 20mADC0 to 1V DC

Scaling highlimit value

Scaling highlimit value

(2) Input the set value, referring to Chapter 6. Operation.

When controlling with Fuzzy self-tuning PID action, select “Perform” during Auto-tuningPerform/Cancel to start control in optimum conditions.

(3) Turn the load circuit power ON.

Control action starts so as to keep the control target at the SV (desired value).

Page 39: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

38

8.2 When using the instrument as a Simplified program controller

(1) Turn the power supply to the instrument ON.For approx. 2sec after power-ON, input characters and temperature unitselected during Input type selection are indicated on PV display, and the inputrange high limit value is indicated on the SV display (Table 8.1-1).

During this time, all outputs and LED indicators are in OFF status.After that, the instrument enters PV/SV display mode.

While Control output OFF function is working, the item selected during “Indicationselection when control output is OFF” is displayed on the PV display.

(2) Input SV and step time, referring to Chapter 6. Operation.The PV display indicates actual temperature and the unit enters standby status.

(3) Turn the load circuit power ON.

(4) Program control start

If the Automatic start is selected during Program start Auto/Manual selection,the unit proceeds to warm-up status for approx. 2 seconds after power-on,then program control starts automatically from step 1.

If Manual start is selected during Program start Auto/Manual selection, the unitproceeds to warm-up status for approx. 2 seconds after power-on, then proceedsto standby status. In this status, if the key is pressed, program control startsfrom step 1.

While program control is performing, Step number change (Set value memorynumber) is invalidated.

To make the step time progress faster.

The step time progress becomes 60 times faster than usual while the keyis being pressed.

To end the program control in process.

The program control ends if the key is pressed for approx. 1 second or longer.

PV/SV display, Open/Closed output MV, Remaining step time indication

In the PV/SV display, if the key is pressed for approximately 3 seconds,the unit indicates Open/Closed output MV (manipulated variable), and if the

key is pressed again, remaining step time will be indicated.

Instrument status when power is restored.

When power is restored during program control, the instrument resumes the programcontinuously.The PV display flashes until the step at which the power failure occurred is finished.

Page 40: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

39

Heating (reverse) action

Controlaction

Cooling (direct) action

ON

OFF

Proportional band

SV

ON

OFF

Proportional band

SV

Open output5

6

Lit Unlit LitUnlit

Closed output5 7

Cycle action is performedaccording to deviation.

Cycle action is performedaccording to deviation.

5 6

7

5

6

7

5

6

7

5

6

7

5

6

7

5

6

7

Indicator

Yellow LitUnlit Lit Unlit

IndicatorOpen outputGreen

Closed output

9. Action explanations9.1 Standard action

9.2 Open/Closed output dead band action

Heating (reverse) action Cooling (direct) action

ON

OFF

SV

Closedoutput

ON

OFF

Dead band

Openoutput

SV

Dead band

ON

OFF

SV

Dead band

SV

Dead bandON

OFF

Page 41: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

40

Pattern end output iscancelled by pressingthe key.

ON

OFFPattern end

action

Time

9

10

9

10

9

10

Unlit UnlitLit

Pattern endoutput

IndicatorRed (A1)

(Program control ends.)OUT

OFF

9.3 Pattern end action

Pattern end output is turned ON after program control ends, and it isnot cancelled until the key is pressed.

Page 42: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

41

High limit alarm with standby Low limit alarm with standby

OFF

ON

SV SV

OFF

ON

OFF

ON

Process high alarm Process low alarm

High limit alarm Low limit alarm

High/Low limit range alarmHigh/Low limits alarm

hysteresis

hysteresis

Alarmaction

Alarmoutput

Alarmaction

Alarmoutput

Alarmaction

Alarmoutput

Alarmaction

Alarmoutput

-A1 value +A1 value valuevalue

A1 value A1 valueSV

+side

- side

+side

- side

A1 hysteresis

A1

OFF

ON

A1 hysteresis

value SV valueA1 A1

valueOFF

ON

hysteresisA1

A1OFF

ON

hysteresis

value

A1

A1

OFF

ON

SV-A1 value +A1

+side

- side

A1 hysteresis

SV

OFF

ON

hysteresis

value-A1 value

+side

- side

A1

+A1value

A1

-A1 +A1

9.4 Alarm 1 (A1), Alarm 3 (A3) and Alarm 4 (A4) action

Page 43: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

42

Alarmaction

Alarmoutput

Alarmaction

Alarmoutput

ON

High/Low limits alarm with standby

A1 hysteresis

A1 value A1 valueSV

High/Low limit range with standby

OFF

ON

A1 hysteresis

value SV valueA1 A1

Process high alarm with standby Process low alarm with standby

valueOFF

ON

hysteresisA1

A1OFF

ON

hysteresis

value

A1

A1

OFF

For A3 (A4) output, read “A3 (A4)” for “A1”.

For A3 output, use terminals 29 and 30.For A4 output, use terminals 28 and 30.

A1, A3 and A4 indicators light when their output terminals are connected,and go off when their output terminals are disconnected.(A3 and A4 are applicable to the FCD-15A.)

: A1 output terminals 9 and 10 are connected.

: A1 output terminals 9 and 10 are connected or disconnected.

: A1 output terminals 9 and 10 are disconnected.

: Standby functions.

Page 44: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

43

10. Control actions10.1 Fuzzy self-tuning

Fuzzy self-tuning is a function to perform a fine adjustment of PID values automatically.The stable control can be carried out even if the conditions of the production process arechanged due to things like variation and variability.

If Lock 1 or Lock 2 is selected, Fuzzy self-tuning or Auto-tuning does not work.In the control system in which load fluctuation periodically occurs, the Fuzzyself-tuning may malfunction.In such cases, use the controller with the PID auto-tuning mode.

(1) When the control initiates, the unit performs this function by the PID valuespreviously adjusted.

(2) When the control result is disordered by disturbance or a change in theprocess, the controller checks the converging status, and performs a fineadjustment of PID values if necessary.

(a) If the convergence is performed smoothly, the PID values are not changed.(b) If the convergent speed is slow, the controller corrects the PID values

to accelerate the convergence.(c) When overshoot is generated during the convergence, the controller changes

the PID values to correct this.(d) When hunting occurs, the controller checks its waveform and performs a

fine adjustment of PID values.

The instrument is always in self-tuning status, and when deviation occurs, the tuningstarts. Even in Fuzzy self-tuning status, when very large hunting occurs and thecontrol is not stabilized, auto-tuning starts automatically.When the auto-tuning “Perform” is selected by the keypad, auto-tuning initiates.When the control is stabilized, the auto-tuning is released and the controllerreturns to self-tuning status.

The controller can be used without Fuzzy self-tuning by selecting the internal switch.When using the controller without Fuzzy self-tuning, perform the PID auto-tuning orset the proper PID values by key operation.

Page 45: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

44

10.2 Explanations of PID

(1) Proportional band (P)

Proportional action is the action which the control output varies in proportion to thedeviation between the SV (desired value) and the PV (process variable).If the proportional band is narrowed, even if the output changes by a slight variationof the PV (process variable), better control results can be obtained as the offsetdecreases.However, if the proportional band is narrowed too much, even slight disturbancesmay cause variation in the PV (process variable), control action changes to ON/OFFaction and the so-called hunting phenomenon occurs.Therefore, when the PV (process variable) comes to the balanced position nearthe SV and a constant temperature is maintained, the most suitable value isselected by gradually narrowing the proportional band while observing the controlresults.

(2) Integral time (I)

Integral action is used to eliminate offset. When the integral time is shortened, thereturning speed to the setting point is accelerated. However, the cycle of oscillationis also accelerated and the control becomes unstable.

(3) Derivative time (D)

Derivative action is used to restore the change in the PV (process variable) accordingto the rate of change. It reduces the amplitude of overshoot and undershoot width.If derivative time is shortened, restoring value becomes small, and if derivative time isextended, an excessive returning phenomenon may occur and the control system mayoscillate.

Page 46: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

45

10.3 PID auto-tuning of this controllerIn order to set each value of P, I, D and ARW automatically, the auto-tuning processshould be made to fluctuate to obtain an optimal value.Sometimes the auto-tuning process will not fluctuate if auto-tuning is performedat or near room temperature. Therefore auto-tuning might not finish normally.

(A) In the case of large difference between the SV and PV (process variable)as the temperature is rising.The AT process will fluctuate at a temperature of 1.5% input span lower than the SV.

(1) Calculating PIDconstant

(2) PID constantcalculated

(3) Controlled by thePID constant setby auto-tuning.

(4) 1.5% of input span

(Fig. 10.3-1)

(B) When the control is stable or when control temperature is within 1.5% of inputspan of the SVThe AT process will fluctuate around the SV.

(1) Calculating PIDconstant

(2) PID constantcalculated

(3) Controlled by thePID constant setby auto-tuning.

(Fig. 10.3-2)

(C) In the case of a large difference between the SV and PV (process variable)as the temperature is fallingThe AT process will fluctuate at a temperature of 1.5% input span higher than the SV.

(1) Calculating PIDconstant

(2) PID constantcalculated

(3) Controlled by thePID constant setby auto-tuning.

(4) 1.5% of input span

(Fig. 10.3-3)

1.5% of input span lower than the SV

(1) (2) (3)

(4)

AT starting point

SV

Temperature

Time

AT starting point

(1) (2) (3)

Temperature

SV

Time

AT starting point

(1) (2) (3)

(4)

Temperature

SV

Time

1.5% of input span higher than the SV

Page 47: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

46

11. Other functions(1) Burnout

Overscale

When the thermocouple or RTD is burnt out, or if the input value exceeds [Input rangehigh limit value +1% of input span], Open output is turned OFF (Closed output ON),and the PV display flashes “ ”.With the RTD Pt100 ( ) (with a decimal point), if the input value exceeds 999.9, the PVdisplay flashes “ ”, however, the control will perform to [Input range high limitvalue +1% of input span].For DC input, if the input value exceeds [Scaling high limit value+10% of the scaling span],Open output is turned OFF (Closed output ON), and the PV display flashes “ ”.If the input exceeds 9999, the PV display flashes “ ”, however, the control performsto [scaling high limit value +10% of scaling span].

Underscale

When the input value drops below [Input range low limit value -1% of input span], Openoutput is turned OFF (Closed output ON), and PV display flashes “ ”.With the Thermocouple T and RTD input, if the input value drops below -199.9, and thePV display flashes “ ”. However, the control will perform to [Input range low limitvalue -1% of input span].For DC input, if the input value drops below [Scaling low limit value -1% of the scalingspan], Open output is turned OFF (Closed output ON), and the PV display flashes“ ”.If the input value drops below -199.9, the PV display flashes “ ”. However,the control will perform to [Scaling low limit value -1% of scaling span].

(2) Self-diagnosis

The CPU is monitored by a watchdog timer, and when an abnormal status is found onthe CPU, the controller is switched to warm-up status.

(3) Automatic cold junction temperature compensation (Thermocouple input type)This detects the temperature at the connecting terminal between thermocouple andthe instrument, and always keeps it at the same status as if the reference junctionis located at 0 (32 ).

(4) Warm-up indication

For approximately 2 seconds after the power supply to the instrument is turned on, theinput type and the temperature unit are indicated on the PV display, and the input rangehigh limit value (for DC input, scaling high limit value) on the SV display.

(5) SV (desired value) ramp function

When the SV is adjusted, it approaches the new SV by the preset rate of change.When the power is turned on, the control starts from the PV (process variable)and approaches the SV by the rate of change.

Page 48: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

47

12. Specifications12.1 Standard specifications (Common to FCD-15A, FCR-15A)

Mounting : FlushSetting : Input system using membrane sheet keyDisplay

[FCD-15A]

PV display : Red LED display 4 digits, size, 14.3(H) x 8(W)mmSV/MV/TIME display: Green LED display 4 digits, size, 10(H) x 5.5(W)mmMEMO display : Yellow LED display 1 digit, size, 8(H) x 4(W)mm[FCR-15A]

PV display : Red LED display 4 digits, size, 8(H) x 4(W)mmSV/MV/TIME display: Green LED display 4 digits, size, 8(H) x 4(W)mmMEMO display : Yellow LED display 1 digit, size, 8(H) x 4(W)mm

Accuracy (Setting, Indication)

Thermocouple : Within 0.2% of input span 1digitHowever,K, J, T input, range less than 0 (32 ): Within 0.4% of

input span 1digitR, S input, range 0 to 200 (0 to 400 ): Within 4 (8 )(Cold junction temperature compensating accuracy:

1 , at 0 to 50 )RTD : Within 0.2% of input span 1digitDC current : Within 0.2% of input span 1digitDC voltage : Within 0.2% of input span 1digit

Input sampling period : 0.125 seconds (0.25 seconds for EA, EV option)Input

Thermocouple : K, J, R, S, B, E, T, N, PL- , C (W/Re5-26),External resistance, 100 or lessFor input burnout, Overscale

RTD : Pt100, JPt100, 3-wire systemAllowable input lead wire resistance, 10 or less per wireFor input burnout, Overscale

DC current : 0 to 20mA DC, 4 to 20mA DCInput impedance, 50Allowable input current, 100mA or lessFor input burnout, 0 to 20mA: the same as 0mA

4 to 20mA: UnderscaleDC voltage : 0 to 1V DC

Input impedance, 1M or moreAllowable input voltage, 5V or lessAllowable signal source resistance, 2k or lessFor input burnout, Overscale

Control output

Relay contact : 1a x 2Control capacity, 3A 250V AC (resistive load)

1A 250V AC (inductive load cos =0.4)

Page 49: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

48

Alarm 1 output

The alarm action point is set by deviation from the SV (desired value) (exceptProcess alarm).When the input is out of the range, the output turns ON or OFF (in the case ofHigh/Low limit range alarm).One alarm is selectable by the rotary switch and DIP switch from the choice of;High limit alarm, Low limit alarm, High/Low limits alarm, High/Low limit rangealarm, Process high alarm and Process low alarm, and standby function isapplied to them respectively, as well as No alarm action and Pattern end output.Setting accuracy: Within 0.2% of input span 1digitAction : ON/OFF action

Hysteresis:Thermocouple, RTD input: 0.1 to 100.0 ( )DC input: 1 to 1000 (The placement of the decimal point place

follows the selection.)Output: Relay contact, 1a

Control capacity, 3A 250V AC (resistive load)1A 250V AC (inductive load cos =0.4)

Controlling action

Fuzzy self-tuning PID or PID action is selectable by the DIP switch.Fuzzy self-tuning PID action

Proportional band (P) : AutomaticIntegral time (I) : AutomaticDerivative time (D) : AutomaticMV computation cycle : 1 to 120secAnti-reset windup (ARW) : AutomaticDead band : 0.1 to 100.0% of proportional bandOpen output time setting : 0.1 to 999.9secClosed output time setting: 0.1 to 999.9sec

PID action (with auto-tuning function)Proportional band (P) : 0.1 to 999.9%Integral time (I) : 1 to 3600secDerivative time (D) : 0 to 3600sec (off when set to 0)MV computation cycle : 1 to 120secAnti-reset windup (ARW) : AutomaticDead band : 0.1 to 100.0% of proportional bandOpen output time : 0.1 to 999.9secClosed output time : 0.1 to 999.9sec

Supply voltage : 100 to 240V AC, 50/60Hz, 24V AC/DC, 50/60HzAllowable voltage fluctuation : 100 to 240V AC: 85 to 264V AC

24V AC/DC : 20 to 28V AC/DCAmbient temperature : 0 to 50 (32 to 122 )Ambient humidity : 35 to 85%RH (non-condensing)Power consumption : Approx. 15VA

Page 50: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

49

Ground terminal Power supply

CPU Communication

Non-insulated

Externalsettinginput

Alarm 3 (A3) output

Insulated

AB

Insulated Insulated Insulated

Open/Closedoutput

Insulated

Insulated

Input

Insulated

Non-insulated

Alarm 1 (A1) output

Alarm 4 (A4) output

InsulatedNon-insulated

Set value memorynumber external

selection

D

Transmissionoutput

C

Insulated

Outputs

Non-insulated

Externaloperation

InsulatedInsulated

Insulated Insulated Insulated

Open/Closedoutput

Insulated

Insulated

CPU Communication

InsulatedNon-insulated Non-insulated

Ground terminal

Alarm 1 (A1) output Transmissionoutput

Power supply

ExternalsettinginputA

B

Input

Set value memorynumber external

selectionD

C

Circuit insulation configuration

[FCD-15A]

[FCR-15A]

Insulation resistance10M or more, at 500V DCInsulation test must not be carried out between A-B and C-D of Circuitinsulation configuration, because they are not insulated from each other.

Dielectric strengthBetween input terminal and ground terminal, 1.5kV AC for 1 minuteBetween input terminal and power terminal, 1.5kV AC for 1 minuteBetween power terminal and ground terminal, 1.5kV AC for 1 minuteOpen/Closed output terminal and ground terminal, 1.5kV AC for 1 minuteOpen/Closed output terminal and power terminal, 1.5kV AC for 1 minute

Weight : FCD-15A: Approx. 500g

FCR-15A: Approx. 320g

External dimensions: FCD-15A: 96 x 96 x 110mm (W x H x D)FCR-15A: 48 x 96 x 110mm (W x H x D)

Material : Base and Case: Flame-resistant resinColor : Base and Case: Light gray

Page 51: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

50

Attached functions : Control output OFF, Set value lock,Sensor correction, Multi-range input,Alarm action delayed timer, Multi-function,Simplified program controller, Set value memory,Power failure countermeasure, Self-diagnosis,SV ramp, Burnout function,Warm-up display function,Loop break alarm indicating function,Automatic cold junction temperature compensation

Accessories : Mounting brackets 1 setInstruction manual 1 copyUnit label 1Terminal cover: FCD-15A: 2 pieces (When TC option is added)

FCR-15A: 1 piece (When TC option is added)Wire harness: 3m 1 length (For FCR-15A with TA, TV, EA, EV option)

12.2 Optional specifications

Alarm 3, 4 output (Option code: SA) (for FCD-15A)

The alarm action point is set by deviation from the SV (except Process alarm), andwhen the input exceeds the range, the output turns ON or OFF (in the case ofHigh/Low limit range alarm).Alarm 3 and 4 are selectable by key operation from 13 types of alarm;High limit alarm, Low limit alarm, High/Low limits alarm, High/Low limit rangealarm, Process high alarm and Process low alarm, and standby function isapplied to them respectively, as well as No alarm action.Setting accuracy: Within 0.2% of input span 1digitAction : ON/OFF action

Hysteresis:Thermocouple, RTD input: 0.1 to 100.0 ( )DC input: 1 to 1000 (The placement of the decimal

point follows the selection.)Output : Relay contact 1a x 2

Control capacity 3A 250V AC (resistive load)1A 250V AC (inductive load cos =0.4)

(However, terminals common to A3 and A4: maximum 3A)Transmission output (Option code: TA, TV)

Converting the value (PV transmission, SV transmission, Open/Closed output MVtransmission) to analog signal every 0.125 seconds, outputs the value in current orvoltage.Transmission output (PV, SV, Open/Closed output MV) can be selected by keypad.Resolution 1/10000Current (TA) 4 to 20mA DC (load resistance, maximum 550 )Voltage (TV) 0 to 1V DC (load resistance, minimum 100k )Output accuracy Within 0.3% of input span

Serial communication (Option code: C5, C)Following operations can be executed from the external computer.(1) Reading and setting of the SV, PID values and various set values(2) Reading of the PV and the action status (3) Change of the functionsCommunication line EIA RS-485 (C5) or

EIA RS-232C (C)Communication method Half-duplex communicationSynchronization method Start-stop synchronization

Page 52: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

51

Communication speed 2400, 4800, 9600, 19200bps (Selectable by key operation)Data format Start bit : 1

Data bit : 7Parity : Even parityStop bit : 1

Digital external setting:Receives digital set value from Shinko programmable controllers such asPCD-33A (with SVTC option) or PC-900 (with SVTC option).Be sure to select Lock 3 during the Set value lock selection for the FCx-15A.

Set value memory number external selection (Option code: SM)Selects the set value memory number from 7 files (mentioned below is 1 file) byexternal terminals.(SV, PID values, Open/Closed dead band, Alarm 1, 3 and 4 values)Memory numbers: 1 to 7 (7 files)Data : FCD-15A: 8 pieces of data

FCR-15A: 6 pieces of dataExternal setting (Option code: EA, EV)

The SV (desired value) can be set using the external analog signal.Setting signal : DC current (EA)

0 to 20mA DC, 4 to 20mA DCInput impedance, 50 (Not insulated from the input)

: DC voltage (EV)0 to 1V DC, 1 to 5V DCAllowable input voltage, 5V or less for 0 to 1V DC

10V or less for 1 to 5V DCInput impedance, 100k (Not insulated from the input)

Setting signal sampling period: 0.25 seconds (If this function is applied, the inputsampling period will also change to 0.25 seconds.)

External operation:If External setting option is applied, Remote/Local setting and Auto/Manual controlcan be switched externally.

Terminals 25 - 27 Closed : Remote settingOpen : Local setting

Terminals 26 - 27 Closed : Manual controlOpen : Automatic control

Color black (Option code: BK)Front panel : Dark grayCase : Black

Screw type mounting bracket (Option code: BL)Mounting panel thickness: 1 to 8mm

Terminal cover (Option code: TC)Electrical shock protection terminal cover

Dust-proof/Drip-proof (IP)

Dust-proof and Drip-proof specification (IP54)• Effective for only front panel surface, case section is excluded.• Effective only when the FCx-15A is mounted vertically.

User specifiedInput, Scale range : Shipped at specified rangeAlarm action : Shipped at specified alarm action (A1, A3, A4)Control action : Shipped at specified control action (PID)Transmission output : Shipped at specified output

(e.g.) SV transmission 0-20mA DC (Transmission output option)External setting : Shipped at specified input.

(e.g.) 1-5V DC (External setting option)

Page 53: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

52

13. TroubleshootingIf any malfunctions occur, refer to the following after checking the power supply to the controller.

WarningTurn the power supply to the instrument off before wiring or checking.Working or touching the terminal with the power switched ON may result insevere injury or death due to Electric Shock. Moreover, the instrument mustbe grounded before the power supply to the instrument is turned on.

<Indication>

Problem Presumed cause and solution[ ] is indicated onthe PV display.

• Control output OFF function is working.Press the key for approx. 1sec to release the function.

[ ] is flashingon the PV display.

• Burnout of thermocouple, RTD or disconnection of DCvoltage (0 to 1V DC). Change each sensor.

How to check whether the sensor is burnt out[Thermocouple]If the input terminals of the instrument are shorted, and if a valuearound room temperature is indicated, the instrument is likely tobe operating normally, however, the sensor may be burnt out.[RTD]If approx. 100 of resistance is connected to the inputterminals between A-B of the instrument and between B-Bis shorted, and if a value around 0 (32 ) is indicated, theinstrument is likely to be operating normally, however, thesensor may be burnt out.[DC voltage (0 to 1V DC)]If the input terminals of the instrument are shorted, and ifa scaling low limit value is indicated, the instrument is likelyto be operating normally, however, the signal wire may bedisconnected.• Check whether the input terminals of thermocouple, RTD or

DC voltage (0 to 1V DC) are securely mounted to theinstrument input terminals.

[ ] is flashingon the PV display.

• Check if polarity of thermocouple or compensating leadwire is correct.

• Check whether codes (A, B, B) of RTD agree with theinstrument terminals.

• For 4-20mA DC input, if the input to the input terminals ofthe instrument is 4mA DC and if a scaling low limit value isindicated, the instrument is likely to be operating normally,however, the signal wire may be disconnected.• For 0-20mA DC input, check whether the signal wire is

securely connected to the instrument input terminals.The indication of thePV display is abnormalor unstable.

• Designation of the sensor input is not proper.Set the input type properly with the Rotary switch (SW303)and DIP switch (SW304).

• Check whether the temperature unit ( or ) is correct.Set the unit properly with the DIP switch (SW304).

• Sensor correction value is not suitable.Set it to a suitable value.

• Check whether the specification of the sensor is correct.• AC is leaking into the thermocouple or RTD circuit.Use an ungrounded type sensor.

• There may be equipment that interferes with or makesnoise near the controller. Keep equipment that interfereswith or makes noise away from the controller

Page 54: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

53

<Key operation>

Problem Presumed cause and solutionSetting mode cannotbe selected.

• Manual control is selected.Change the mode to Automatic control.

It is impossible tochange the modefrom Remote settingto Local setting bykey operation.

• Remote setting has been selected using the Externalsetting (EA or EV option).External operation has priority.

The mode cannot bechanged from Manualcontrol to Automaticcontrol by key operation.

• Manual control has been selected using an Externalsetting (EA or EV option).External operation has priority.

Settings are impossible.The value does notchange by the ,

keys.

• Lock 1 or Lock 2 is selected during Set value lock selection.Release the lock.

• During PID auto-tuning.Cancel auto-tuning if necessary.

The setting indicationdoes not change in theinput range even if the

, keys arepressed, and new valuesare unable to be set.

• SV high limit or low limit may be set at the point wherethe value does not change.Set it to a suitable value while in Auxiliary functionsetting mode 1.

Program control doesnot start even if the

key is pressedin the Program mode.

• Step time has not been set.Set the step time.

<Control>

Problem Presumed cause and solutionPV (Temperature)does not rise.

• Sensor is out of order. Replace the sensor.• Check whether the Sensor is securely mounted to the

instrument input terminals.Ensure that the sensor is securely mounted to theinstrument input terminals.

• Check whether the wiring of sensor is correct.Control does notstart.(The PV display isindicated)

• Program mode is selected.If running by Program mode, press the key.When performing by Fixed value control, press the

key for approx. 3sec while holding down thekey to select the Fixed value control mode.

The value is fullyopen or fully closedregardless of thecontrolling temperaturehigh or low.

• Check whether Open/Closed output time is set suitably.Set the time to a suitable value.

For all other malfunctions, please contact our main office or dealers.

Page 55: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

54

14. Character tablePhotocopiable material

Alarm 3 and Alarm 4 (value, type, hysteresis, delayed timer) are applicable to theFCD-15A.

<Main setting mode>Character Item Default Data

SV 0

<Sub setting mode>Character Item Default Data

Set value memory number 1AT (Auto-tuning) Perform/Cancel CancelProportional band 2.5%Integral time 200secDerivative time 50secOpen/Closed output dead band 3.0%Alarm 1 value 0Alarm 3 value 0Alarm 4 value 0Loop break alarm time 0 minLoop break alarm span 0

<Auxiliary function setting mode 1>Character Item Default Data

Set value lock CancelSV high limit 400SV low limit 0Sensor correction 0.0Remote/Local LocalInstrument number 0Communication speed 9600bps

<Auxiliary function setting mode 2>Character Item Default Data

Scaling high limit 9999Scaling low limit −1999Decimal point place No decimal point

PV filter time constant 0.0secAlarm 3 type No alarmAlarm 4 type No alarmAlarm 1 hysteresis 1.0Alarm 3 hysteresis 1.0Alarm 4 hysteresis 1.0Alarm 1 delayed timer 0secAlarm 3 delayed timer 0secAlarm 4 delayed timer 0secExternal setting high limit 400External setting low limit 0

Page 56: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

55

Transmission output (PV)Transmission output high limit 400Transmission output low limit 0Indication selection when controloutput is OFFMV computation cycle 3secSV rise rate 0 /minuteSV fall rate 0 /minute

<Open/Closed output time setting mode>Character Item Default Data

Open output time 30.0secClosed output time 30.0sec

<Program mode>Character Item Default Data

Fixed value control/Program control Fixed valueStep 1 time 00.00Step 2 time 00.00Step 3 time 00.00Step 4 time 00.00Step 5 time 00.00Step 6 time 00.00Step 7 time 00.00

Page 57: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

SHINKO TECHNOS CO.,LTD.OVERSEAS DIVISION

:::

Reg. OfficeURLE-mail

2-5-1, Senbahigashi, Minoo, Osaka, Japanhttp://[email protected]

Tel :Fax:

81-72-727-610081-72-727-7006

***** Inquiry *****

For any inquiries about this controller, after checking the following as to thecontroller, please contact the shop where you purchased the unit or our agency.

(Example)Model -------------------- FCD-15A-R/MInput type -------------- KOption ------------------- SA, TVSerial number --------- No. xxxxxx

In addition to the above, please let us know the details of malfunction, if any,and the operating conditions.

FCDR12E7 2008.08

Page 58: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

1

RS-485

COMMUNICATION INSTRUCTION MANUAL FC SERIES (C, C5)

No.FC1CE11 2006.12 To prevent accidents arising from the use of this controller, please ensure the operator receives this manual. Models for Shinko protocol : FCS-23A, FCR-13A, FCR-15A, FCR-23A, FCD-13A, FCD-15A Models for Modbus protocol : FCS-23A, FCR-13A, FCR-23A, FCD-13A

Warning Turn the power supply to the instrument OFF before wiring or checking. Working or touching the terminal with the power switched ON may result in severe injury or death due to Electric Shock.

1. System configuration RS-232C (C option)

(Fig. 1-1)

RS-485 Multi-drop connection (C5 option)

(Fig. 1-2) (Fig. 1-3)

2. Wiring RS-232C (C option)

(Fig. 2-1)

Host computer

RS-485 RS-485

FC series No. 0

FC series No. 1

FC series No. 2

FC series No. 30

Communication converter

IF-400 232C 485

FC series No. 0

FC series No. 1

FC series No. 30

Host computer FC series

TXD

RXD

SG

FG

RTS

CTS

DSR

DTR

CD

2

3

7

1

4

5

6

20

8

Host computer

TX

RX

COM

Shielded wire

D-sub 25-pin connector

FCD-13A

11

12

16

RS-232C

Host computer RS-232C

Page 59: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

2

D-sub 25-pin connector

Host computer

D-sub 9-pin connector

Host computer

RXD

TXD

DCD

DTR

DSR

RTS

CTS

2

3

5

1

4

6

7

8

GND

RI 9

RS-232C RS-485

IF-400

CDM

FG FG

FG

FG

TXD

RXD

SG

FG

RTS

CTS

DSR

DTR

CD

2

3

7

1

4

5

6

20

8

RS-232C RS-485

IF-400

CDM

FG

FG

FG

When using communication converter IF-400 D-sub 9-pin connector

(Fig. 2-2)

D-sub 25-pin connector

(Fig. 2-3)

Shielded wire Connect only one side of the shielded wire to the FG or GND terminal so that current cannot flow to the shielded wire. (If both sides of the shielded wire are connected to the FG or GND terminal, the circuit will be closed between the shielded wire and the ground. As a result, current will run through the shielded wire and this may cause noise.) Be sure to ground FG and GND terminals. Recommended cable: OTSC-VB 2PX0.5SQ (made by Onamba Co., Ltd.) or equivalent (Use a twisted pair cable.)

Terminator (Terminal resistor) The terminator prevents signal reflection and disturbance. Connect a terminator at the end of the communication cable. The IF-400 (sold separately) is available as a communication converter.

Shielded wire Shielded

wire

FCD-13A

11 YA(-) 12 YB(+) 16 COM 1 GND

Shieldedwire

Shieldedwire

11 YA(-) 12 YB(+) 16 COM 1 GND

11 YA(-) 12 YB(+) 16 COM 1 GND

FCD-13A

11 YA(-) 12 YB(+) 16 COM 1 GND

Shieldedwire

Shieldedwire

11 YA(-) 12 YB(+) 16 COM 1 GND

11 YA(-) 12 YB(+) 16 COM 1 GND

43 1 6

43 1 6

Page 60: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

3

3. Communication parameters setting Set Communication parameters as follows. (Refer to the Instruction manual.)

(1) Instrument number setting Set an instrument number to each of the FC series individually when communicating by connecting plural units. (Default: 0)

(2) Communication speed selection Select a communication speed for the FC series according to that of the host computer. (Default: 9600bps)

(3) Communication protocol selection (For the FCD-13A, FCR-13A, FCR-23A, FCS-23A) Select a communication protocol. (Default: Shinko protocol)

4. Communication procedure Communication starts with command transmission from the host computer (hereafter Master) and ends with the response of the FC series (hereafter Slave).

• Response with data When the master sends the reading command, the slave responds with the corresponding set value or current status. • Acknowledgement When the master sends the setting command, the slave responds by sending the acknowledgement after the processing is terminated. • Negative acknowledgement When the master sends a non-existent command or value out of the setting range, the slave returns a negative acknowledgement as a response. • No response

The slave will not respond to the master in the following cases: • Global address (Shinko protocol) is set. • Communication error (framing error, parity error) (Fig.4-1) • Checksum error (Shinko protocol), LRC discrepancy (Modbus)

Communication timing between master and slave Master side (Notice on programming)

Set the program so that the master can disconnect the transmitter from the communication line within a 1 character transmission period after sending the command in preparation for reception of the response from the slave. To avoid the collision of transmissions between the master and the slave, send the next command after carefully checking that the master received the response.

Slave side When the slave starts transmission through the RS-485 communication line, the slave is arranged so as to provide an idle status (mark status) transmission period of 1 or more characters before sending the response to ensure the synchronization on the receiving side. The slave is arranged so as to disconnect the transmitter from the communication line within a 1 character transmission period after sending the response.

5. Shinko protocol

5.1 Transmission mode Shinko protocol is composed of ASCII codes. Hexadecimal (0 to 9, A to F), which is divided into high order (4-bit) and low order (4-bit) out of 8-bit binary data in command is transmitted as ASCII characters. Data format Start bit: 1 bit Data bit: 7 bits Parity : Even Stop bit: 1 bit Error detection: Checksum

5.2 Command configuration All commands are composed of ASCII. The data (set value, decimal number) is represented by hexadecimal figures, and ASCII codes are used for the command. The negative numbers are represented by 2's complement.

CommandData

CommandAcknowledgement

CommandNegative

acknowledgement

Command

No response

Master Slave

Page 61: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

4

(1) Setting command Header (02H) Address Memory

number Commandtype (50H) Data item Data Check-

sum Delimiter

(03H) 1 1 1 1 4 4 2 1 (Fig. 5.2-1) Number of characters

(2) Reading command Header (02H) Address Memory

number Commandtype (20H) Data item Check-

sum Delimiter

(03H) 1 1 1 1 4 2 1 Number of characters (Fig. 5.2-2)

(3) Response with data Header (06H) Address Memory

number Commandtype (20H) Data item Data Check-

sum Delimiter

(03H) 1 1 1 1 4 4 2 1 (Fig. 5.2-3) Number of characters

(4) Acknowledgement Header (06H) Address Check-

sum Delimiter

(03H) 1 1 2 1 Number of characters (Fig. 5.2-4)

(5) Negative acknowledgement Header (15H) Address Error

code Check-

sum Delimiter

(03H) 1 1 1 2 1 Number of characters (Fig. 5.2-5)

Header : Control code to represent the beginning of the command or the response. ASCII codes are used. Setting command, Reading command : STX (02H) fixed Response with data, Acknowledgement : ACK (06H) fixed Negative acknowledgement : NAK (15H) fixed

Address (Instrument number): Numbers by which the master discerns each slave. Instrument number 0 to 94 and Global address 95. ASCII codes (20H to 7FH) are used by adding 20H to instrument numbers 0 to 95 (00H to 5FH). 95 (7FH) is called Global address, which is used when the same command is sent to all the slaves connected. However, a response is not returned.

Memory number: Set value memory number (1 to 7) Zero (0) is set to the data item which has no relation to the Set value memory. The numbers (20H to 27H) are used by adding 20H of bias.

Command type : Code to discern Setting command (50H) and Reading command (20H) Data item : Data classification of the command object Composed of hexadecimal 4 digits (Refer to the Communication command table.) Data : The contents of data (set value) differ depending on the setting command. Composed of hexadecimal 4 digits (Refer to the Communication command table.) Checksum : 2-character data to detect communication errors (Refer to 5.3 Checksum calculation.) Delimiter : Control code to represent the end of command 03H fixed Error code : Represents an error type. Composed of hexadecimal 1 digit. 0 (30H)---- Unknown error 1 (31H)-----Non-existent command

2 (32H)-----Not used 3 (33H)-----Set value out of the setting range 4 (34H)-----Unsettable status (e.g. AT is performing) 5 (35H)-----During setting mode by keypad operation

5.3 Checksum calculation Checksum is used to detect receiving errors in the command or data. Set the program for the master side as well to calculate the checksum of the response data from the slaves so that the communication errors can be checked. The ASCII code (hexadecimal) corresponding to the characters which range from the address to that before the checksum is converted to binary notation, and the total value is calculated. The lower 2-digits of the total value are converted to 2’s complements and then to hexadecimal figures, that is, ASCII code for the checksum.

Page 62: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

5

STX ETXP 0 0 0 1 0 2 5 8

02H 21H 21H 50H 30H 30H 30H 31H 30H 32H 35H 38H 03H

[Characters above are represented by ASCII]

Checksum

Checksum calculation range[e.g.]

D

44H 45H

E

21H21H50H30H30H30H31H30H32H35H38H

0010 00010010 00010101 00000011 00000011 00000011 00000011 00010011 00000011 00100011 01010011 1000+

10 0010 0010

[Hexadecimal] [Binary]

1101 11011+

1101 1110

D E

44H 45H

[1's complement]

[2's complement]

[Hexadecimal]

[ASCII]

Checksum

! !

Checksum calculation example Address (instrument number): 1 (21H) Memory number: 1 (21H) SV: 600 (0258H)

• 1’s complement: Reverse each binary bit. 0 will become 1 and vice versa. • 2’s complement: Add 1 to 1’s complements.

5.4 Command table (Shinko protocol) When the data (set value) has a decimal point, remove the decimal point and represent it as a whole number, then express it in hexadecimal figures.

Memory Number

Command type Data item Data

1 to 7 20H/50H 0001H SV or step SV

Set value (Decimal point ignored)

0 20H/50H 0002H Set value memory number or step number

Selected value

0 20H/50H 0003H AT Perform/Cancel

0000H: Cancel 0001H: Perform

1 to 7 20H/50H 0004H OUT1 proportional band Set value (Decimal point ignored)

1 to 7 20H/50H 0005H OUT2 proportional band (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

1 to 7 20H/50H 0006H Integral time Set value 1 to 7 20H/50H 0007H Derivative time Set value

0 20H/50H 0008H OUT1 proportional cycle (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value

0 20H/50H 0009H OUT2 proportional cycle (For FCD-13A, FCR-13A, FCR-23A)

Set value

0 20H/50H 000AH Manual reset (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

1 to 7 20H/50H 000BH A1 value Set value (Decimal point ignored)

1 to 7 20H/50H 000CH A2 value (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

1 to 7 20H/50H 000DH A3 value (For FCD-13A, FCD-15A)

Set value (Decimal point ignored)

1 to 7 20H/50H 000EH A4 value (For FCD-13A, FCD-15A)

Set value (Decimal point ignored)

Page 63: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

6

0 20H/50H 000FH Heater burnout alarm (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

0 20H/50H 0010H Loop break alarm time Set value 0 20H/50H 0011H Loop break alarm span Set value

(Decimal point ignored.) 0 20H/50H 0012H Set value lock

(If Lock 3 is selected, the set data will not saved in the memory. By turning the power off, the set values return to the previous one)

0000H: Unlock 0001H: Lock 1 0002H: Lock 2 0003H: Lock 3

0 20H/50H 0013H SV high limit Set value (Decimal point ignored)

0 20H/50H 0014H SV low limit Set value (Decimal point ignored)

0 20H/50H 0015H Sensor correction value Set value (Decimal point ignored)

1 to 7 20H/50H 0016H Overlap band/Dead band (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

0 20H/50H 0017H Remote/Local (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

0000H: Local 0001H: Remote

0 20H/50H 0018H Scaling high limit

Set value (Decimal point ignored)

0 20H/50H 0019H

Scaling low limit

Set value (Decimal point ignored)

0 20H/50H 001AH Decimal point place (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

0000H: XXXX (No decimal point)

0001H: XXX.X (1 digit after the decimal point)

0002H: XX.XX (2 digits after the decimal point)

0003H: X.XXX (3 digits after the decimal point)

0 20H/50H 001BH PV filter time constant Set value (Decimal point ignored)

1 to 7 20H/50H 001CH OUT1 high limit (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

1 to 7 20H/50H 001DH OUT1 low limit (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

0 20H/50H 001EH OUT1 ON/OFF action hysteresis (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

0 20H/50H 001FH OUT2 action mode (For FCD-13A, FCR-13A, FCR-23A)

0000H: Air cooling (Linear characteristic)

0001H: Oil cooling (1.5th power of the linear)

0002H: Water cooling (2nd power of the linear)

1 to 7 20H/50H 0020H OUT2 high limit (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

1 to 7 20H/50H 0021H OUT2 low limit (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

0 20H/50H 0022H OUT2 ON/OFF action hysteresis (For FCD-13A, FCR-13A, FCR-23A)

Set value (Decimal point ignored)

0 20H/50H 0023H 0024H

A3 type (For FCD-13A, FCD-15A) A4 type (For FCD-13A, FCD-15A)

0000H: No alarm 0001H: High limit alarm 0002H: High limit alarm with standby 0003H: Low limit alarm 0004H: Low limit alarm with standby 0005H: H/L limits alarm 0006H: H/L limits alarm with standby 0007H: H/L limit range alarm

Page 64: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

7

0008H: H/L limit range alarm

with standby 0009H: Process high alarm000AH: Process high alarm

with standby 000BH: Process low alarm 000CH: Process low alarm

with standby 0 20H/50H 0025H A1 hysteresis Set value(Decimal point ignored)0 20H/50H 0026H A2 hysteresis

(For FCD-13A, FCR-13A, FCR-23A, FCS-23A)Set value (Decimal point ignored)

0 20H/50H 0027H A3 hysteresis (For FCD-13A, FCD-15A)

Set value (Decimal point ignored)

0 20H/50H 0028H A4 hysteresis (For FCD-13A, FCD-15A)

Set value (Decimal point ignored)

0 20H/50H 0029H A1 action delayed timer Set value 0 20H/50H 002AH A2 action delayed timer

(For FCD-13A, FCR-13A, FCR-23A, FCS-23A)Set value

0 20H/50H 002BH

A3 action delayed timer (For FCD-13A, FCD-15A)

Set value

0 20H/50H 002CH A4 action delayed timer (For FCD-13A, FCD-15A)

Set value

0 20H/50H 002DH External setting input high limit (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

Set value (Decimal point ignored)

0 20H/50H 002EH External setting input low limit (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

Set value (Decimal point ignored)

0 20H/50H 002FH Transmission output mode (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

0000H: PV transmission 0001H: SV transmission 0002H: MV transmission

0 20H/50H 0030H Transmission output high limit (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

Set value (Decimal point ignored)

0 20H/50H 0031H Transmission output low limit (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

Set value (Decimal point ignored)

0 20H/50H 0032H Indication selection when control output is OFF

0000H: OFF indication 0001H: No indication 0002H: PV indication

0 20H/50H 0033H SV rise rate Set value (Decimal point ignored)

0 20H/50H 0034H SV fall rate Set value (Decimal point ignored)

0 20H/50H 0035H Fixed value control/Program control

0000H: Fixed value control 0001H: Program control

1 to 7 20H/50H 0036H Step (1 to 7) time Set value (Decimal point ignored) Converted to minutes, thento hexadecimal figures. (See “Data” on p.14)

0 20H/50H 0037H Control output OFF function For fixed value control: 0000H: ON 0001H: OFF

For program control: 0000H: STOP 0001H: RUN

0 20H/50H 0038H Auto/Manual control (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

0000H: Automatic control 0001H: Manual control

0 20H/50H 0039H Manual MV (manipulated variable) (For FCD-13A, FCD-15A, FCR-13A, FCR-23A, FCR-15A)

• Settable only in the Manual mode.

• The setting range depends on OUT1(OUT2)high and low limit values.

• Set value(Decimal point ignored)

Page 65: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

8

1 to 7 20H/50H 003AH Open/Closed dead band

(For FCD-15A, FCR-15A) Set value

0 20H/50H 003BH Open output time (For FCD-15A, FCR-15A)

Set value

0 20H/50H 003CH Closed output time (For FCD-15A, FCR-15A)

Set value

0 20H/50H 003DH MV computation cycle setting (For FCD-15A, FCR-15A)

Set value

0 20H/50H 003EH Infrared emissivity (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

Set value (Decimal point ignored)

0 20H/50H 003FH Control output OFF function selection for excess input (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

0000H: Disabled 0001H: Enabled

0 20H/50H 0040H A1 action Energized/Deenergized (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

0000H: Energized 0001H: Deenergized

0 20H/50H 0041H A2 action Energized/Deenergized (For FCD-13A, FCR-13A, FCR-23A, FCS-23A)

0000H: Energized 0001H: Deenergized

0 20H/50H 0042H A3 action Energized/Deenergized (For FCD-13A, FCD-15A)

0000H: Energized 0001H: Deenergized

0 20H/50H 0043H A4 action Energized/Deenergized (For FCD-13A, FCD-15A)

0000H: Energized 0001H: Deenergized

0 20H 0080H PV PV (Decimal point ignored)0 20H 0081H OUT1 MV (manipulated variable) OUT1 MV

(Decimal point ignored) 0 20H 0082H OUT2 MV (manipulated variable)

(For FCD-13A, FCR-13A, FCR-23A) OUT2 MV (Decimal point ignored)

0 20H 0083H Current SV during program control

Current SV (Decimal point ignored)

0 20H 0084H Remaining time during program control

Remaining time (Decimal point ignored) Converted to minutes, thento hexadecimal figures. (See “Data” on p.14)

0 20H 0085H Status flag

0000 0000 0000 0000 215 to 20

20 digit: OUT1 (Control output 1)

0: OFF, 1: ON 21 digit: OUT2 (Control

output 2) 0: OFF, 1: ON 22 digit: A1 output 0: OFF, 1: ON 23 digit: A2 output 0: OFF, 1: ON 24 digit: A3 output 0: OFF, 1: ON 25 digit: A4 output 0: OFF, 1: ON 26 digit: Heater burnout

alarm output 0: OFF, 1: ON 27 digit: Loop break alarm

output 0: OFF, 1: ON 28 digit: Overscale 0: OFF, 1: ON 29 digit: Underscale 0: OFF, 1: ON 210 to 215 digit: Not used.

Always 0 0 20H 0086H Selected Set value memory number

(running step number) Selected value

Page 66: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

9

6. Modbus protocol (Not available for FCR-15A, FCD-15A) 6.1 Modbus message framing

Modbus protocol has 2 transmission modes (ASCII and RTU), and all commands are made up of ASCII codes since the FC series (slave) runs on the ASCII mode. Data (set value, decimal number) is converted to hexadecimal figures, and ASCII codes are used. Negative numbers are represented by 2's complement. In ASCII mode, messages start with a colon (: 3AH) character, and end with a carriage return-line feed (CRLF) pair (0DH and 0AH). Intervals of up to one second can elapse between characters within the message. If a greater interval occurs, the receiving device assumes an error has occurred. A typical message frame of the slave is shown in (Table 6.1-1). (Table 6.1-1)

Header ( : ) Slave address Function code

Address Data Number of data

or data Error check (LRC) Delimiter (CR•LF)

6.2 Slave address

Slave addresses are represented by two ASCII characters. Slave address (instrument number): 0 to 95 (00H to 5FH) Slave address (instrument number) is set to the FC series in the range of 0 to 95 (00H to 5FH) beforehand by the key operation. A master addresses a slave by placing the slave address in the address field of the message. When the slave sends its response, it places its own slave address in this address field of the response to let the master know which slave is responding. Address 0 slave receives a message and returns the response message ignoring the broadcast address [0 (00H)] since slaves are not supported by the broadcast address.

6.3 Function code Function code is represented by two ASCII characters. When the message is transmitted from the master to the slaves, function code field tells the slave what kind of action to perform.

Function code Contents 03 (03H) Reading the set value and information from slaves

(Only one piece of data can be read.) 06 (06H) Setting to slaves (Only one piece of data can be written.)

When the slave responds to the master, function code field is used to indicate if it is a normal response or if an error has occurred. For a normal response, the slave simply echoes the original function code. For an exception response, the slave returns the value by adding 1 to the most significant bit of the original function code. This tells the master what kind of error occurred by adding the exception code to the response.

Exception code Contents 0 (00H) Reserve (Normal or undefined) 1 (01H) Illegal function (Non-existent function) 2 (02H) Illegal data address (Non-existent data address) 3 (03H) Illegal data value (Set value out of the range)

6.4 Data field

Data field consists of the address and number of data, and is represented by four ASCII characters. A request message from the master is composed of address, number of data or data. A response message from the slave is composed of number of bytes, data and exception code in negative acknowledgement. Effective range of data is –32768 to 32767 (8000H to 7FFFH).

Page 67: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

10

6.5 Error check field After calculating LRC (Longitudinal Redundancy Check) from the slave address to the end of data, the calculated 8-bit data is converted to two ASCII characters and are appended to the end of the message. [Error checking methods] The LRC is applied to the entire message. Both the character check and message frame check are calculated in the master device, and applied to the message contents before transmission. (The slave device checks each character and the entire message frame during reception.) The master is configured by the user to wait for a predetermined timeout interval before aborting the transaction. This interval is set to be long enough for any slave to respond normally. If the slave detects a transmission error, the message will not be valid. The slave will not construct a response to the master. Thus the timeout will expire and allow the master’s program to handle the error. The message addressed to a nonexistent slave device will also cause a timeout. [LRC checking] The LRC field checks the contents of the message, excluding the beginning colon and ending CR•LF pair. It is applied regardless of any parity check method used for the individual characters of the message. The LRC field is one byte, containing an eight-bit binary value. The LRC value is calculated by the transmitting device, which appends the LRC to the message. The receiving device calculates an LRC during reception of the message, and compares the calculated value to the actual value it received in the LRC field. If the two values are not equal, an error message is returned. How to calculate LRC The LRC is calculated by adding together successive eight-bit bytes of the message (excluding the beginning colon and ending CR•LF), discarding any carries, and then is calculated by two’s compliment of the result.

6.6 Example of transmitting message (1) Reading of SV and status (Address 1, Memory number 1, SV)

• A request message from the master Header

(3AH )

Slave address

(30H 31H)

Function code

(30H 33H)

Address

(30H 30H 30H 30H)

Number of data

(30H 30H 30H 31H)

Error check LRC

(46H 42H)

Delimiter

(0DH 0AH)1 2 2 4 4 2 2

(Fig. 6.6-1) Number of characters

• Response message from the slave in normal status (When SV is 600 ) Slave address, function code, number of data, etc are represented by hexadecimal value, and replaced to the messages character by character. There are two response bytes since it is 16-bit data per channel. For the message, it becomes 4 times the number of characters since one character is composed of 4 bits. (*1: “30H 34H” for the actual instrument) Header

(3AH)

Slave address

(30H 31H)

Function code

(30H 33H)

Number of response byte *1

(30H 34H)

Data

(30H 32H 35H 38H)

Error check LRC

(39H 45H)

Delimiter

(0DH 0AH)1 2 2 2 4 2 2

(Fig. 6.6-2) Number of characters (2) Reading of set value and status (Address 1, PV)

• A request message from the master Header (3AH )

Slave address (30H 31H)

Function code (30H 33H)

Address (30H 30H 39H 39H)

Number of data (30H 30H 30H 31H)

Error check LRC

(36H 32H)

Delimiter(0DH 0AH)

1 2 2 4 4 2 2 (Fig. 6.6-3) Number of characters

• Response message from the slave in normal status (When PV is 600 ) Slave address, function code, number of data, etc are represented by hexadecimal value, and replaced to the messages character by character. There are two response bytes since it is 16-bit data per channel. For the message, it becomes 4 times the number of characters since one character is composed of 4 bits. (*1: “30H 34H” for the actual instrument) Header

(3AH)

Slave address

(30H 31H)

Function code

(30H 33H)

Number of response byte *1

(30H 34H)

Data

(30H 32H 35H 38H)

Error check LRC

(39H 45H)

Delimiter

(0DH 0AH)1 2 2 2 4 2 2

(Fig. 6.6-4) Number of characters

Page 68: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

11

• Response message from the slave in exception (error) status (data item mistaken) Header

(3AH)

Slave address

(30H 31H)

Function code

(38H 33H)

Exception code [02H](30H 32H)

Error checkLRC

(37H 41H)

Delimiter

(0DH 0AH) 1 2 2 2 2 2 (Fig. 6.6-5) Number of characters The slave returns the error contents to the master by adding 1 (binary) to the most significant bit of function code.

(3) Setting to the slave (Address 1, Memory number 1, SV is set to 600 )

• A request message from the master Header

(3AH )

Slave address

(30H 31H)

Function code

(30H 36H)

Data item

(30H 30H 30H 30H)

Data

(30H32H35H 38H)

Error check LRC

(39H 46H)

Delimiter

(0DH 0AH) 1 2 2 4 4 2 2 (Fig. 6.6-6) Number of characters

• Response message from the slave in normal status Slave address, function code, number of data, etc are represented by hexadecimal value, and replaced to the messages character by character. Header

(3AH )

Slave address

(30H 31H)

Function code

(30H 36H)

Data item

(30H 30H 30H 30H)

Data

(30H32H 35H 38H)

Error check LRC

(39H 46H)

Delimiter

(0DH 0AH) 1 2 2 4 4 2 2

(Fig. 6.6-7) Number of characters • Response message from the slave in exception (error) status (When “set value out of the range” is set.)

The slave returns the error contents to the master by adding 1 (binary) to the most significant bit of function code. Header

(3AH)

Slave address

(30H 31H)

Function code

(38H 36H)

Exception code [03H](30H 33H)

Error checkLRC

(37H 36H)

Delimiter

(0DH 0AH) 1 2 2 2 2 2

(Fig. 6.6-8) Number of characters

6.7 Command table (Modbus protocol) R: Reading, W: Writing (setting) Address Contents R/W Data

0000H to 0006H SV or step SV Corresponds to memory 1 to 7 R/W Set value Decimal point ignored

0007H to 000DH OUT1 proportional band Corresponds to memory 1 to 7 R/W Set value Decimal point ignored

000EH to 0014H OUT2 proportional band (For FCD-13A, FCR-13A, FCR-23A)

Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

0015H to 001BH Integral time Corresponds to memory 1 to 7 R/W Set value 001CH to 0022H Derivative time Corresponds to memory 1 to 7 R/W Set value 0023H to 0029H Alarm 1 (A1) value Corresponds to memory 1 to 7 R/W Set value

Decimal point ignored002AH to 0030H Alarm 2 (A2) value Corresponds to memory 1 to 7 R/W Set value

Decimal point ignored0031H to 0037H Alarm 3 (A3) value

(For FCD-13A) Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

0038H to 003EH Alarm 4 (A4) value (For FCD-13A)

Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

003FH to 0045H Overlap/Dead band (For FCD-13A, FCR-13A, FCR-23A)

Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

0046H to 004CH OUT1 high limit Corresponds to memory 1 to 7 R/W Set value Decimal point ignored

004DH to 0053H OUT1 low limit Corresponds to memory 1 to 7 R/W Set value Decimal point ignored

0054H to 005AH OUT2 high limit (For FCD-13A, FCR-13A, FCR-23A)

Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

005BH to 0061H OUT2 low limit (For FCD-13A, FCR-13A, FCR-23A)

Corresponds to memory 1 to 7

R/W Set value Decimal point ignored

0062H to 0068H Step (1 to 7) time Corresponds to step 1 to 7 R/W Set value, Decimal point ignored Converted to minutes, then to hexadecimal figures. (See “Data” on p.14)

Page 69: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

12

0069H Set value memory number or Step number R/W Selection 006AH AT Perform/Cancel R/W 0000H: Cancel

0001H: Perform 006BH OUT1 proportional cycle R/W Set value 006CH OUT2 proportional cycle (For FCD-13A,FCR-13A,FCR-23A) R/W Set value 006DH Manual reset R/W Set value

Decimal point ignored006EH Heater burnout alarm value

(For FCD-13A, FCR-13A, FCR-23A) R/W Set value

Decimal point ignored006FH Loop break alarm action time R/W Set value 0070H Loop break alarm action span R/W Set value

Decimal point ignored0071H Set value lock

(If Lock 3 is selected, the set data will not saved in the memory. By turning the power off, the set values return to the previous one)

R/W 0000H:Unlock 0001H: Lock 1 0002H: Lock 2 0003H: Lock 3

0072H SV high limit R/W Set value Decimal point ignored

0073H SV low limit R/W Set value Decimal point ignored

0074H Sensor correction value R/W Set value Decimal point ignored

0075H Remote/Local (For FCD-13A, FCR-13A, FCR-23A)

R/W 0000H: Local 0001H: Remote

0076H Scaling high limit R/W Set value Decimal point ignored

0077H Scaling low limit R/W Set value Decimal point ignored

0078H Decimal point place R/W 0000H: XXXX 0001H: XXX.X 0002H: XX.XX 0003H: X.XXX

0079H PV filter time constant R/W Set value Decimal point ignored

007AH OUT1 ON/OFF action hysteresis R/W Set value Decimal point ignored

007BH OUT2 action mode (For FCD-13A, FCR-13A, FCR-23A)

R/W 0000H: Air cooling(Linear)0001H: Oil cooling (1.5th power of the linear)

0002H: Water cooling(2ndpower of the linear)

007CH OUT2 ON/OFF action hysteresis (For FCD-13A, FCR-13A, FCR-23A)

R/W Set value Decimal point ignored

007DH A3 type (For FCD-13A)

R/W 0000H: No alarm 0001H: High limit alarm0002H: High limit alarm with standby 0003H: Low limit alarm0004H: Low limit alarm with standby 0005H: H/L limits alarm0006H: H/L limits alarm

with standby 0007H: H/L limit range

alarm 0008H: H/L limit range

alarm with standby 0009H:Process high

alarm 000AH: Process high

alarm with standby000BH:Process low

alarm 000CH:Process low alarm with standby

007EH A4 type (For FCD-13A) R/W The same as A3 type

Page 70: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

13

007FH A1 hysteresis R/W Set value Decimal point ignored

0080H A2 hysteresis R/W Set value Decimal point ignored

0081H A3 hysteresis (For FCD-13A)

R/W Set value Decimal point ignored

0082H A4 hysteresis (For FCD-13A)

R/W Set value Decimal point ignored

0083H A1 action delayed timer R/W Set value 0084H A2 action delayed timer R/W Set value 0085H A3 action delayed timer (For FCD-13A) R/W Set value 0086H A4 action delayed timer (For FCD-13A) R/W Set value 0087H External setting input high limit

(For FCD-13A, FCR-13A, FCR-23A) R/W Set value

Decimal point ignored 0088H External setting input low limit

(For FCD-13A, FCR-13A, FCR-23A) R/W Set value

Decimal point ignored 0089H Transmission output mode

(For FCD-13A, FCR-13A, FCR-23A) R/W 0000H: PV transmission

0001H: SV transmission0002H: MV transmission

008AH Transmission output high limit (For FCD-13A, FCR-13A, FCR-23A)

R/W Set value Decimal point ignored

008BH Transmission output low limit (For FCD-13A, FCR-13A, FCR-23A)

R/W Set value Decimal point ignored

008CH Indication selection when control output is OFF R/W 0000H: OFF indication0001H: No indication 0002H: PV indication

008DH SV rise rate R/W Set value Decimal point ignored

008EH SV fall rate R/W Set value Decimal point ignored

008FH Fixed value control/Program control R/W 0000H: Fixed value control0001H: Program control

0090H Control output OFF function R/W For fixed value control0000H: ON 0001H: OFF

For program control 0000H: STOP 0001H: RUN

0091H Auto/Manual control (For FCD-13A, FCR-13A, FCR-23A)

R/W 0000H: Automatic 0001H: Manual

0092H Manual MV (manipulated variable) (For FCD-13A, FCR-13A, FCR-23A)

R/W • Settable only in the Manual mode.

• The setting range differs depending on OUT1 (OUT2) high and low limit values.

• Set value Decimal point ignored

0093H Infrared emissivity R/W Set value Decimal point ignored

0094H Control output OFF function selection for excess input R/W 0000H: Prohibited 0001H: Allowed

0095H A1 action Energized/Deenergized R/W 0000H: Energized 0001H: Deenergized

0096H A2 action Energized/Deenergized R/W 0000H: Energized 0001H: Deenergized

0097H A3 action Energized/Deenergized (For FCD-13A)

R/W 0000H: Energized 0001H: Deenergized

0098H A4 action Energized/Deenergized (For FCD-13A)

R/W 0000H: Energized 0001H: Deenergized

0099H PV R PV Decimal point ignored

Page 71: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

14

009AH OUT1 MV (manipulated variable) R OUT1 MV Decimal point ignored

009BH OUT2 MV (manipulated variable) (For FCD-13A, FCR-13A, FCR-23A)

R OUT2 MV Decimal point ignored

009CH Current SV during program control R Current SV Decimal point ignored

009DH Remaining time during program control R Remaining time Decimal point ignored Converted to minutes, thento hexadecimal figures. (See “Data” on p.14)

009EH Status flag R 0000 0000 0000 0000215 to 20

20 digit: OUT1 (Control output 1)

0: OFF, 1: ON 21 digit: OUT2 (Control

output 2) 0: OFF, 1: ON 22 digit: A1 output 0: OFF, 1: ON 23 digit: A2 output 0: OFF, 1: ON 24 digit: A3 output 0: OFF, 1: ON 25 digit: A4 output 0: OFF, 1: ON 26 digit: Heater burnout

alarm output 0: OFF, 1: ON 27 digit: Loop break

alarm output 0: OFF, 1: ON 28 digit: Overscale 0: OFF, 1: ON 29 digit: Underscale 0: OFF, 1: ON 210 to 215 digit:

Not used. Always 0009FH Selected Set value memory number (running step) R Selected number Data

Note on setting, reading command • The data (set value, decimal) is converted to hexadecimal figures. A negative number is represented

by 2's complement. • When connecting plural slaves, the address (instrument number) must not be duplicated. • For Step (1 to 7) time (0036H, 0062H to 0068H for Modbus), and Remaining time during program control

(0084H, 009DH for Modbus), time is converted to minutes, then to hexadecimal figures, and ASCII is used. (e.g.) 1 hour 30 minutes 90minutes 005AH (hexadecimal)

99 hours 59 minutes 5999minutes 176FH (hexadecimal) Setting command

• Setting range of each item is the same as that of keypad operation. • When the data (set value) has a decimal point, a whole number (hexadecimal) without a decimal point is used. • If the alarm type is changed during Alarm 3 type selection (0023H, 007DH for Modbus) and Alarm 4 type

selection (0024H, 007EH for Modbus), the alarm value will revert to “0”. Also alarm output status will be initialized.

• It is possible to set the set value with the setting command of the communication function even when the set value is locked.

• Although the options are not applied, setting the items for the options is possible using the setting command. However, they will not function.

• Instrument numbers, communication speed and protocol of the slave cannot be set by communication function. • When sending a command by Global address (Shinko protocol) or Broadcast address (Modbus protocol),

the same command is sent to all the slaves connected. However, the response is not returned. • The memory can store up to 1,000,000 (one million) entries.

If the number of settings exceeds the limit, the data will not be saved. So frequent transmission via communication is not recommended.

Reading command • When the data (set value) has a decimal point, a whole number (hexadecimal) without a decimal point

is used for a response.

Page 72: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

15

Negative acknowledgement The slave will return Error code 1 (31H) (Shinko protocol) or Exception code 1 (01H) (Modbus protocol) in the following cases.

• When AT Perform/Cancel (0003H, 006AH for Modbus protocol) is selected during PI action or ON/OFF action. The slave will return Error code 4 (34H) (Shinko protocol) or Exception code 17 (11H) (Modbus protocol) in the following cases.

• If “Cancel (0000H)” of AT Perform/Cancel (0003H, 006AH for Modbus protocol) is selected while Auto- tuning is being cancelled.

• When “Perform (0001H)” of AT Perform/Cancel (0003H, 006AH for Modbus protocol) is selected while Auto-tuning is performing

Notes on programming monitoring software How to speed up the scan time

When monitoring plural units of FC series, set the program so that requisite minimum pieces of data such as PV (0080H, 0099H for Modbus protocol), OUT1 MV (0081H, 009AH for Modbus), Status flag (0085H, 009EH for Modbus), etc. can be read, and for other data, set the program so that they can be read only when their set value has changed. This will speed up the scan time.

Note when sending all set values at one time • If alarm type is changed during Alarm 3 type selection (0023H, 007DH for Modbus) or Alarm 4 type

selection (0024H, 007EH for Modbus), the alarm value will revert to “0”. First, send the selected alarm type, then send the alarm value.

When communicating with a PLC Command example (Shinko protocol) when communicating with a Mitsubishi PLC (FX series, etc.) • Reading (Top D register: D100)

Address 1, PV reading Sending Data (STX)(!)( )( )(0)(0)(8)(0)(D)(7)(ETX)

Command Register Code Header (STX) 02H D100(LSB) 02H Address 1 D100(MSB) 21H Sub address 20H D101(LSB) 20H Command type 20H D101(MSB) 20H

D102(LSB) 30H D102(MSB) 30H D103(LSB) 38H Data item &H80

D103(MSB) 30H D104(LSB) 44H Checksum D104(MSB) 37H

Delimiter (ETX) 03H D105(LSB) 03H Communication setting [MOVP H0C86 D8120] Reading + Setting [RS D100 K11 D108 K26] Reading only [RS D100 K11 D106 K26]

1 [MOV H2102 D100] 2 [MOV H2020 D101] 3 [MOV H3030 D102] 4 [MOV H3038 D103] 5 [MOV H3744 D104] 6 [MOV H03 D105]

• Setting (Top D register: D120) Address 1, Memory No. 1, SV setting [When setting SV to 600 (0258H)]

Sending Data (STX)(!)(!)(P)(0)(0)(0)(1)(0)(2)(5)(8)(D)(E)(ETX) Command Register Code

Header (STX) 02H D120(LSB) 02H Address 1 D120(MSB) 21H Sub address 1 D121(LSB) 21H Command type P D121(MSB) 50H

D122(LSB) 30H D122(MSB) 30H D123(LSB) 30H Data item &H1

D123(MSB) 31H D124(LSB) 30H D124(MSB) 32H D125(LSB) 35H Data item 600

D125(MSB) 38H D126(LSB) 44H Checksum D126(MSB) 45H

Delimiter (ETX) 03H D127(LSB) 03H

Page 73: DTR_FCR-15A_FCD-15A

ACSE Sp.

z o.o.

16

SHINKO TECHNOS CO.,LTD.OVERSEAS DIVISION

:::

Reg. OfficeURLE-mail

2-5-1, Senbahigashi, Minoo, Osaka, Japanhttp://[email protected]

Tel :Fax:

81-72-727-610081-72-727-7006

Reading + Setting [RS D120 K15 D128 K26] Reading only [RS D120 K15 D128 K22]

1 [MOV H2102 D120] 2 [MOV H5021 D121] 3 [MOV H3030 D122] 4 [MOV H3130 D123] 5 [MOV H3230 D124] 6 [MOV H3835 D125] 7 [MOV H4544 D126] 8 [MOV H03 D127]

7. Specifications Cable length : RS-232C: 10m (Max), RS-485: 1km (Max) Communication method: Half-duplex communication Communication speed : 9600bps (2400, 4800, 9600, 19200bps) Selectable by keypad Synchronization : Start-stop synchronization Code form : ASCII Error correction : Command request repeat system Error detection : Parity check, Checksum (Shinko protocol), LRC (Modbus protocol) Communication protocol : Shinko protocol/Modbus ASCII (Selectable by keypad) Data format : Start bit: 1, Data bit: 7, Parity: Even, Stop bit: 1

8. Troubleshooting If any malfunctions occur, refer to the following items after checking the power supply of the host computer and the FC series.

Problem: Communication failure Check the following

The connection or wiring of communication is not secure. A burnout or imperfect contact on the communication cable and the connector Communication speed of the FC series does not coincide with that of the host computer. The data bit, parity and stop bit of the host computer do not accord with those of the FC series. The instrument number (address) of the FC series does not coincide with that of the command. The instrument numbers (addresses) are duplicated in multiple FC series. When communicating by RS-485 (C5 option), make sure that the program is appropriate for the transmission timing. For the C5 option, when communicating with the host computer using RS-485 board (or customers’own board) instead of Shinko IF-400, communications sometimes cannot be carried out successfully. In this case, connect 2 resistors of 680 (1/4W) as shown on the right.

Problem: Although communication is occurring, the response is 'NAK'.

Check the following Check whether or not a non-existent command code has been sent. The setting command data exceeds the setting range of the FC series.The controller cannot be set when functions such as AT are performing.The operation mode is under the front keypad operation setting mode.

For further inquiries, please consult our agency or the shop where you purchased the unit.

Added resistors680 (1/4W)

+5V line

COM

YB

YA

RS-485 board(Customers’own board)

(Fig. 8-1)