DIFFERENTIAL DIFFERENTIAL THERMAL THERMAL
ANALYSIS (DTA)ANALYSIS (DTA)
Prepared by:- Guided by:-Mr. Jiten B. Patel Mrs. Henal Chorawala M.pharm(1st sem) Pro. M.pharm
B.M.C.P.
![Page 2: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/2.jpg)
Thermal Analysis TechniquesThermal Analysis Techniques
• When a material is heated its structural and chemical composition can undergo changes such as fusion, melting, crystallization, oxidation, decomposition, transition, expansion and sintering.
• Using Thermal Analysis such changes can be monitored in every atmosphere of interest. The obtained information is very useful in both quality control and problem solving.
![Page 3: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/3.jpg)
Types of thermal analysis Types of thermal analysis
• TG (Thermogravimetric) analysis: weight
• DTA (Differential Thermal Analysis): temperature
• DSC (Differential ScanningCalorimetry): temperature
![Page 4: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/4.jpg)
• In Differential Thermal Analysis, the temperature difference that develops
between a sample and an inert reference material is measured, when both are subjected
to identical heat - treatments.• The related technique of Differential
Scanning Calorimetry relies on differences in energy required to maintain the sample and reference at an identical temperature.
![Page 5: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/5.jpg)
Thermogravimetry (TG)Thermogravimetry (TG)
• Thermogravimetry is the measurement of the mass of a sample as the temperature increases. This method is useful for determining sample purity and water, carbonate, and organic content; and for studying decomposition reactions.
![Page 6: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/6.jpg)
INTRODUCTIONINTRODUCTION
• This is a comparison method
• Analytical method for recording the difference in temperature (∆T) b/w a substance and an inert reference material as a function of temperature or time
• Any transformation – change in specific heat or an enthaply of transition can be detected by DTA
![Page 7: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/7.jpg)
• In DTA both test sample & an inert reference material (alumina) – controlled heating or cooling programming
• If zero temperature difference b/w sample & reference material – sample does not undergo any chemical or physical change.
• If any reaction takes place temperature difference (∆T) will occur b/w sample & reference material
![Page 8: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/8.jpg)
• A DTA curve can be used as a finger print for identification purposes, for example,
• in the study of clays where the structural similarity of different forms renders diffraction experiments difficult to interpret.
![Page 9: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/9.jpg)
∆T VS Temp.
Sharp Endothermic – changes in crystallanity or fusionBroad endotherms - dehydration reactionPhysical changes usually result in endothermic curvesChemical reactions are exothermic
![Page 10: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/10.jpg)
ApparatusApparatus
• The key features of a differential thermal analysis kit are as follows
1. Sample holder comprising thermocouples, sample containers and a ceramic or metallic
block. 2. Furnace. 3. Temperature programmer. 4. Recording system.
![Page 11: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/11.jpg)
![Page 12: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/12.jpg)
![Page 13: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/13.jpg)
![Page 14: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/14.jpg)
• Heart of the analysis – heating block• Identical pair of cavities for the sample, ref.
material• Whole unit is set in an oven- control
pressure• Thermocouple is place directly in contact
with the sample and another in contact with the reference
• Temp.of the block is raised, the temperature of the sample & reference follow
• Zero temp. difference – no physical or chemical change
• If any reaction – difference in ∆T
![Page 15: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/15.jpg)
![Page 16: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/16.jpg)
![Page 17: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/17.jpg)
Differential Thermal AnalysisAdvantages:
• instruments can be used at very high temperatures
• instruments are highly sensitive
• characteristic transition or reaction temperatures can be accurately determined
Disadvantages:
• uncertainty of heats of fusion, transition, or reaction estimations is 20-50%
DTA
![Page 18: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/18.jpg)
Factors affect results in DTAFactors affect results in DTA
• Sample weight• Particle size• Heating rate• Atmospheric conditions• Conditions of sample packing into
dishes
![Page 19: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/19.jpg)
![Page 20: Dta](https://reader035.fdocuments.in/reader035/viewer/2022070322/5593258e1a28ab14208b45cc/html5/thumbnails/20.jpg)
ApplicationsApplications
• Quantitative identification and purity assessment of materials are accomplished by comparing the DTA curve of sample to that of a reference curve
• Impurities may be detected by depression of the M.P