DSA DEEPAK GUPTA

92
DSA ITS APPLICATION AND ADVANCEMENT Moderator: MR. S.R.CHOWDHURY TUTOR, DEPT.OF RADIO-DIAGNOSIS & IMAGING, PGIMER CHANDIGARH Presented by: DEEPAK GUPTA BSc final year Student , DEPT.OF RADIO-DIAGNOSIS & IMAGING PGIMER CHANDIGARH

description

more informative to ask any question related seminar ask me on [email protected]

Transcript of DSA DEEPAK GUPTA

Page 1: DSA  DEEPAK GUPTA

DSA ITS APPLICATION AND ADVANCEMENT

Moderator:MR. S.R.CHOWDHURYTUTOR,DEPT.OF RADIO-DIAGNOSIS & IMAGING,PGIMER CHANDIGARH

Presented by:DEEPAK GUPTABSc final year Student ,DEPT.OF RADIO-DIAGNOSIS & IMAGINGPGIMER CHANDIGARH

Page 2: DSA  DEEPAK GUPTA

04/11/2023 2

ANGIO?

ANGIO means blood vessel

And angiography is the radiological study of blood vessel in the body after the introduction of iodinated contrast media.

SUBTRACTION?

It is simply a technique by which bone structures images are subtracted or canceled out from a film of bones plus opacified vessels, leaving an unobscured image of the vessels.

Page 3: DSA  DEEPAK GUPTA

04/11/2023 3

"DSA" means "Digital Subtraction Angiography"

WHAT DO YOU MEAN BY DSA

DSA-The acquisition of digital fluoroscopic images combined with injection of contrast material and real-time subtraction of pre- and post contrast images to perform angiography is referred to as digital subtraction angiography

Page 4: DSA  DEEPAK GUPTA

04/11/2023 4

The Portuguese neurologist Egas Moniz,( Nobel Prize

winner 1949), in 1927developed the technique of

contrast x-ray cerebral angiography to diagnose

diseases, such as tumors and arteriovenous

malformations.

HISTORY

The idea of subtraction images was first proposed by the Dutch radiologist Ziedses des Plantes in the 1935, when he was able to produce subtracted images using plain films.With the introduction of the Seldinger

technique in 1953, the procedure became

safer as no sharp devices need to remain

inside the vascular lumen.

Page 5: DSA  DEEPAK GUPTA

04/11/2023 5

HISTORICAL DEVELOPMENT

CONVENTIONAL SUBTRACTION TECHNIQUE

Photographic method used to eliminate unwanted images.No addition of information;only purpose to make diagnostically important information to see.First described by a Dutch radiologist, Zeides des Plantes, 1935.3 conditions:SCOUT FILM ANGIOGRAM FILM-CONTRAST NO MOTION OF HEAD

Page 6: DSA  DEEPAK GUPTA

04/11/2023 6

The principles of subtraction are based on the following:The scout film shows the structural details of the skull and the adjacent soft tissue.Angiogram film shows exactly the same anatomic details, if the patient does not move, plus the opacified blood vessels.If all the information in the scout film could be subtracted from the angiogram film, only the opacified vessel pattern would remain visible.

CONTD.

Page 7: DSA  DEEPAK GUPTA

04/11/2023 7

EQUIPMENTTHE AOT CHANGER:Can change films –6/s or 1/5s Two of these changers can be coupled electronically and mechanically and will operate out of phase or synchronously with each other.

Radiographs in 2 planes simultaneously-one mounted vertically and other horizontal.

Page 8: DSA  DEEPAK GUPTA

04/11/2023 8

LOADING MAGZINE Large steel container.

Inside the container are arranged a no. of strong wire separators.

30 sheets of film.Precautions :each film should slide into position

by it’s own wt.

RECIEVING / LOADING MAGZINE

AOT CONTD.

Page 9: DSA  DEEPAK GUPTA

04/11/2023 9

30 SEPERATING WIRES

EXPOSURE AREA :Rectangular, slightly recessed area at the top of the changer.Defined by heavy metal frame.Within this is a secondary grid.Supported on 4 small helical springs placed cornerwise and to consist of a stout Al plate, to the lower side of which is attached an intensifying screen.

RECEIVING CASSETTE:Shallow,polished metal container.Lid is a sliding section in a side which can be pushed down by firm finger pressure and is retained by central spring loaded catch.

Page 10: DSA  DEEPAK GUPTA

04/11/2023 10

THE ROLL FILM CHANGER:

Mechanically easier to wind roll films.12/s.

Page 11: DSA  DEEPAK GUPTA

04/11/2023 11

DIGITAL SUBTRACTION

ANGIOGRAPHY

HISTORY:Developed in 1970s.University of Wisconsin, University of Arizona,University of Kiel.Commercial systems introduced in 1980.

Page 12: DSA  DEEPAK GUPTA

04/11/2023 12

Angiography theory ages

Single shot angiographyManual changerAOT PUCKDSA

Page 13: DSA  DEEPAK GUPTA

04/11/2023 13

EQUIPMENT AND APPARATUS

Both film and digital capabilities are usually present and the fluoroscopy equipment operates in the conventional way.Brief review :

Output phosphorLight intensity

Electrical signal

Input surface Visible light Light detectedamplified

Page 14: DSA  DEEPAK GUPTA

04/11/2023 14

DIGITAL IMAGE PROCESSOR

DIGITIZER DIGITAL ARITHEMATIC UNIT

DIGITAL IMAGE MEMORY

DISPLAY SECTION

COMPUTER MONITOR

DIGITALDISK

STORAGE

MULTIFORMAT CAMERA

X-RAY

CONTROL

SIGNALS

Page 15: DSA  DEEPAK GUPTA

04/11/2023 15

X-Ray Unit

Should be capable of rapid serial exp. @2-8

frames/sec for up to 20 sec.

Near monochromatic beam consistent output.

Isocentre motorized table ,tilt 90-0-90 , up-down,

sliding ,floating capable

Ceiling mounted C-arm with versatile movements

II system with multi field ( 6”,9”,12”,16”)

High resolution TV /CCD camera

Computer with high definition monitor system

Page 16: DSA  DEEPAK GUPTA

04/11/2023 16

High output preferably high frequency

generator

High heat loading capacity ,fast heat

dissipation, high speed anode

Target material

tungsten,rhenium,graphite.

Oil-water close loop cooling

Contd.

Page 17: DSA  DEEPAK GUPTA

04/11/2023 17

In our dept.Siemens polydoros SX 65-80Polystar

Page 18: DSA  DEEPAK GUPTA

04/11/2023 18

GAS , AIR & VACUUM PORTS

Boyle apparatus

NECESSARY EQUIPMENT USED WITH ANGIO(DSA)

Page 19: DSA  DEEPAK GUPTA

04/11/2023 19

BED SIDE MONITOR

INJECTOR

Page 20: DSA  DEEPAK GUPTA

04/11/2023 20

Page 22: DSA  DEEPAK GUPTA

04/11/2023 22

WORKING

Page 23: DSA  DEEPAK GUPTA

04/11/2023 23

Modern DSA systems are based on digital fluoroscopy/fluorography systems, which are equipped with special software and display facilities. Digital subtraction angiography (DSA) was developed to improve vessel contrast. This is a technique that uses a computer to subtract two images, obtained before and after contrast media is injected into the vessels of interest. The anatomical structures that are the same in the two images can be removed and the resulting image shows the vessels only.

Page 24: DSA  DEEPAK GUPTA

04/11/2023 24

 The image before the contrast agent is administered is called the mask image. Once the contrast is administered, a sequence of images are taken by a television camera in analog form, which is then digitised by computer.

 The DSA processor has two separate image memories, one for the mask and the other for the images with contrast medium. These two image memories are subtracted from one another arithmetically, and the result goes to an image processing and display unit.

Page 25: DSA  DEEPAK GUPTA

04/11/2023 25

Adjust contrast & brightness.Remasking – correcting misregistered images.Pixel shifting.Edge enhancement – edges of the vessels can

be enhanced so that small details can be made more obvious.

Image zoom.land marking – a small amount of original

image is added into the subtracted image.Noise smoothing – operates by reducing the

statistical fluctuations in each pixel by averaging the pixel with it's closest neighbors. The visual prominence of noise has been suppressed by averaging, but resolution is decreased.

IMAGE PROCESSING

Page 26: DSA  DEEPAK GUPTA

04/11/2023 26

Road mapping is useful for the placement of catheters and wires in complex and small vasculature. A DSA sequence is performed, and the frame with maximum vessel opacification is identified;this frame becomes the road map mask. The road map mask is subtracted from subsequent live fluoroscopic images to produce real-time subtracted fluoroscopic images overlaid on a static image of the vasculature

Road Mapping

Page 27: DSA  DEEPAK GUPTA

04/11/2023 27

When a digital subtraction technique is used, patient motion that occurs between acquisition of the precontrast images and acquisition of the postcontrast images will result in artifacts due to misregistration of the two images. If these arti-facts are observed, it is possible to reregister the pre- and postcontrast images by shifting the subtraction mask (precontrast image) with respect to the postcontrast image and resubtracting the two images

Page 28: DSA  DEEPAK GUPTA

04/11/2023 28

VIDICON CAMERA TUBE

• Diameter 1 inch

• Length 6 inch

• Main parts -

• Target section

• Electron gun

• Electromagnetic focusing coil

• Two pair of electrostatic deflecting coil

• A scanning section

Page 29: DSA  DEEPAK GUPTA

04/11/2023 29

Working principle of T.V monitor

Control Grid

No. of e-ns & the brightness of individual

dots

Produce bright area in the TV Picture

Dark Area

Strikes the fluorescent Screen

Emits large no. of light photons

VISIBLE TV IMAGE

Receive Video signal from ccu & regulate

Grid cuts off the e-ns flow almost completely

E-ns is Accelerated

Page 30: DSA  DEEPAK GUPTA

04/11/2023 30

• In the scanning section of tube, externally mounted coil (emitting focusing coil and electrostatic deflecting coil) produce an axial electromagnetic field by means of which-

• The electrons are focused on the target of tube and

• The beam is moved over the target area in an orderly scanning scan

• Two types scanning system

• Sequential scanning pattern and

• Interlaced scanning pattern

SCANNING SECTION/ SYSTEMS

Page 31: DSA  DEEPAK GUPTA

04/11/2023 31

SEQUENTIAL SCANNING PATTERN

• Also called progressive scanning

Disadvantages • Technical• Requires a wide frequency band during

transmission

Page 32: DSA  DEEPAK GUPTA

04/11/2023 32

Interlaced scanning

• Instead of scanning all 625 line

consequently, only the even

no. of lines are scanned the

first half of the frame and only

the odd no. lines are scanned

during second half.

• In this pattern the transmission

is easier and cheaper because

the actual picture frequency is

lower

Page 33: DSA  DEEPAK GUPTA

04/11/2023 33

TELEVISION IMAGE QUALITY

The assessment of the quality of TV images is a complex subject but there are a no. of image characteristics relating specifically to TV system

• Resolution • Contrast • Brightness • Image lag• Distortion

Page 34: DSA  DEEPAK GUPTA

04/11/2023 34

Resolution is the process by which something is separated into its component part

The resolution of imaging system refers to the amount of detail which is observable

It may be formulated as the number of pairs of black and white lines which an image forming or image recording device can demonstrate in a length of 1 mm

Vertical resolution is determined by no. of vertical scan lines (e.g. 625).

In recent years the manufactures of radio diagnostic imaging equipment have developed his definition TV system which employ 1249 or even more lines.

Resolution

Page 35: DSA  DEEPAK GUPTA

04/11/2023 35

The contrast of the image on a TV screen clearly depends on the

contrast of the original image focused onto the signal plate of the

pick up tube.

Both camera and monitor affect the contrast of a TV image.

A vidicon camera reduces contrast by a factor of approx. 0.8 and the

monitor enhances by a factor of 2.

Contrast

Page 36: DSA  DEEPAK GUPTA

04/11/2023 36

The term luminance and brightness are often used when discussing the aspect of tv imageLuminance can be defined as light emitted per unit area from a surface. It is measured in candela per square meter (cd/m2)

Brightness

Automatic brightness control (ABC)

• ABC helps to maintain the image intensifier exposure rate based on the subject’s thickness

• It is critical to the patient dose and image quality

Page 37: DSA  DEEPAK GUPTA

04/11/2023 37

The ABC monitors the light output from an area of the face of the image intensifier and it tries to maintain the signal travel within an approximate range of that output and adjust tube potential (kvp) and tube current (mA) a/c to predefined algorithm

It refers to control of x-ray exposure levels

Automatic gain control (AGC)If brightness is controlled by varying sensitivity of

the TV system the term AGC is used.It is fairly simple and inexpensive way to control

image brightness It does not change the x-ray dose rate to the patient

37

Page 38: DSA  DEEPAK GUPTA

04/11/2023 38

Image lag Lag is the term used to describe inability of an

imaging system to follow rapid changes in its input image

Distortion Distortion of the image occur in the image

intensifier or in the optical coupling system which links its output phosphor to the signal plate of the TV pick up tube

38

Page 39: DSA  DEEPAK GUPTA

04/11/2023 39

ArtifactsLagVignetting,Veiling glare Pincushion distortionS distortion.

39

Page 40: DSA  DEEPAK GUPTA

04/11/2023 40

Lag Lag is the persistence of luminescence after x-

ray stimulation has been terminated. Lag degrades the temporal resolution of the dynamic image. Older image intensifier tubes had phosphors with lag times on the order of 30–40 msec. Current image intensifier tubes have lag times of approximately 1 msec.

40

Vignetting

A fall-off in brightness at the periphery of an image is called vignetting. Vignetting is caused by the unequal collection of light at the center of the image intensifier compared with the light at its periphery. As a result, the center of an image intensifier has better resolution, increased brightness, and less distortion.

Page 41: DSA  DEEPAK GUPTA

04/11/2023 41

Veiling GlareScattering of light and the defocusing of photoelectrons within the image intensifier are called veiling glare. Veiling glare degrades object contrast at the output phosphor of the image intensifier. As mentioned, the contrast ratio is a good measure of determining the veiling glare of an image intensifier. X-ray, electron, and light scatter all contribute to veiling glare.

41

Pincushion Distortion• Pincushion distortion is a geometric, nonlinear

magnification across the image. • The magnification difference at the periphery of the

image results from the projection of the x-ray beam onto a curved input surface.

• The distortion is easily visualized by imaging a rectangular grid with the fluoroscope.

Page 42: DSA  DEEPAK GUPTA

04/11/2023 42

S DISTORTION

Electrons within the image intensifier move in paths along designated lines of flux. External electromagnetic sources affect electron paths at the perimeter of the image intensifier more so than those nearer the center. This characteristic causes the image in a fluoroscopic system to distort with an S shape .Manufacturers include a highly conductive mu-metal shield that lines the canister in which the vacuum bottle is positioned to reduce the effect of S distortion.

42

Page 43: DSA  DEEPAK GUPTA

04/11/2023 43

INDICATIONSNon traumatic Subarachnoid Hemorrhage

(SAH).Arterial dissection or lacerationAneurysmPseudoaneurysmThrombosis.Arterio-venous malformation (AVM). Arterio-venous fistula (AVF).Tumor vascularity.Gastrointestinal bleed/Uterine artery bleedTherapeutic.

43

Page 44: DSA  DEEPAK GUPTA

04/11/2023 44

CONTRAINDICATIONSNo absolute contraindication.

Poor renal reserve.

Deranged coagulogram.

Allergic to contrast media.

Recent barium examination –abd.angio

44

Page 45: DSA  DEEPAK GUPTA

04/11/2023 45

Contrast Media

Blood vessels are not normally seen in an x-ray image, because of low tissue contrast.

To increase image contrast, contrast agents, which are dense fluids with elements of high atomic numbers, such as iodine, are injected into a blood vessel during angiography. Because of its higher density and high atomic number, iodine absorbs photons more than blood and tissue.

This creates detailed images of the blood vessels in real time.

The first contrast media used for intravascular injection were called high-osmolar contrast media (HOCM). (osmolality is the measure of the particle concentration in a solution.)

45

Page 46: DSA  DEEPAK GUPTA

04/11/2023 4646

 HOCM had osmolarity seven to eight times

higher than plasma. This high osmolarity

caused adverse effects such as pain,

endothelial damage, thrombosis, and

increased pressure in the pulmonary

circulation.

 Low-osmolar contrast media (LOCM) were

first developed in the 1970's and these

helped to reduce these side effects.

 One of the major risks of modern iodine

contrast media is an allergic reaction to

iodine.

Page 47: DSA  DEEPAK GUPTA

04/11/2023 47

Non ionic Iso-osmolar contrast media.

30-40% dilution with normal saline.

50 ml of diluted contrast media is enough to

do a standard cerebral angiogram with total 8

projections.

Approx. 5-8 ml diluted contrast / injection.

47

Page 48: DSA  DEEPAK GUPTA

04/11/2023 48

Materials used

Catheters

Arterial sheath

Medicut

Guidewires

Contrast

Connector/100

cm.tubing

Surgical blade48

Saline

Disposable

syringes

Local anesthesia

Heparin

Surgical gloves

Elastoplast

Page 49: DSA  DEEPAK GUPTA

04/11/2023 4949

Picard

Pigtail

Sim-1

Cathetars

Page 50: DSA  DEEPAK GUPTA

04/11/2023 5050

VASCULAR SHEATH

MEDICUT

GUIDE WIRES

ELASTOPLAST STRIPS

Page 51: DSA  DEEPAK GUPTA

04/11/2023 51

CONTRAST ANESTHESIA

ANTICOAGULANT

PVA-PARTICLESGEL FOAM

Page 52: DSA  DEEPAK GUPTA

04/11/2023 52

PREPARATION

Appointment time

Nil orally 4-6 hrs.

On trolley

In hospital gown

Groin shave

Records

PTI

52

Should be well

hydrated.

Should void before

procedure.

Peripheral pulses

marked.

I.V line in place.

INFORMED CONSENT

MUST

Page 53: DSA  DEEPAK GUPTA

04/11/2023 5353

DSA

DIAGNOSTIC• Cerebral- 4 vessel• Upper & lower limb • Renal • IVC graphy • Pulmonary • Abdominal - coelic

& branches• -sup.

mesentric A.• - inf.

mesentric A.• Transjugular liver

biopsy• Balloon occlusion test.

THERAPEUTIC

• Embolization • - coil • - gel• - particle• - ballon• Stenting

• TACE

• Vertebroplasty

Page 54: DSA  DEEPAK GUPTA

04/11/2023 54

PROCEDUREGaining arterial access.

Selective arterial catheterization.

Image acquisition.

Closure of arterial access.

Post processing

Hard copy

54

Page 55: DSA  DEEPAK GUPTA

04/11/2023 5555

Page 56: DSA  DEEPAK GUPTA

04/11/2023 56

During angiography, patients may be sedated to reduce anxiety.

Their heart rate and rhythm, breathing, and oxygen saturation are monitored throughout the procedure.

Patient clean draped . A local anesthetic is usually used in the area

where the catheter is to be inserted, most commonly the femoral artery.

First, a small incision given, medicut is inserted into the artery. fluoroscopy is used to guide the needle to the proper position .

The needle is then removed after placing guide wire in the artery and vascular sheath is inserted over the guide wire . The catheter is then inserted along the guide wire through the sheath.

56

Page 57: DSA  DEEPAK GUPTA

04/11/2023 57

When the catheter is in the correct position, the wire is pulled out and dye is injected through the catheter.

Images are acquired during contrast injection. Injections can be made directly into the artery of interest (selective arteriography)

Complications from an arteriogram are very rare, but there is some risk. Most problems that occur can be detected at the time of the procedure or immediately after the procedure. The artery may be injured at the puncture site or along the artery where the catheter is passed.

57

Page 58: DSA  DEEPAK GUPTA

04/11/2023 58

COMPLICATIONS0.16% major complication rate.Local complications: hematoma,

vessel laceration, dissection, peudoaneurysm, AV fistula.

Systemic complications: contrast reactions, fever, sepsis, dehydration, death.

CNS complication: aggravation of preexisting complaints, neurological deficit.

58

Page 59: DSA  DEEPAK GUPTA

04/11/2023 59

POSTPROCEDURAL CARE

After the catheter is removed compression is applied to the puncture site.The patient is asked for bed rest for a minimum of 4 hoursDuring rest patient is monitored and vital sign like peripheral pulse like distal to Puncture are regularly The extremity is also checked for warmth, color, numbness to ensure circulation has not been disrupted.

Page 60: DSA  DEEPAK GUPTA

04/11/2023 60

Oral fluid is given and analgesics are given if required.Special care should be given in case of children and geriatric patient since additional Patience is required. They have given warming blanket to keep the body temp. Normal. Geriatric patient also frequently feel nervous and afraid of falling off the Table ,reassurance and additional care from the technologist will enable the Patient to feel secure and comfortable.

Page 61: DSA  DEEPAK GUPTA

04/11/2023 61

Digital subtraction techniques

DSA has many techniques

Mask subtraction

Dual energy subtraction

• Time interval differencing

Temporal filtering

Page 62: DSA  DEEPAK GUPTA

04/11/2023 62

Mask subtractionThe mask image is

subtracted from the

image with dye to

reveal just the vessels

uncontaminated by

overlying bone

structures.

PRE CONTRAST POST CONTRAST

SUBTRACTED ENHANCED

Page 63: DSA  DEEPAK GUPTA

04/11/2023 63

Dual energy subtraction

2 images are taken within a very short period (no pt. movement ).

One at low KVp & other at high KVp.

K EDGE HYBRID

Page 64: DSA  DEEPAK GUPTA

04/11/2023 64

K edge subtractionIodine attenuates diagnostic x-rays almost

entirely by P.E.E.One method for achieving high iodine

contrast is removal of all x-rays from the beam except those that lie below the k-shell B.E of ‘I’.

An x-ray filter made of rare earth named cerium can do this well, since B.E = 40 KeV.

Thick filter will remove most of the x-rays above 40 KeV.

The image to be subtracted should contain few

x-rays.

Page 65: DSA  DEEPAK GUPTA

04/11/2023 65

Hybrid subtraction

High KVp & low KVp image pair is collected.

No pt. movement – soft tissues cancel properly. but bone

edges cause severe artifact.

Consider the subtracted images to consist of only 2 atomic

no. materials (iodine & bone).

The hybrid subtraction produces 2 sets of subtracted images.

The same bone & iodine structures are present on both sets.

D.E.S can now be used to eliminate bone leaving only iodine.

The final image – more noise.

Mask subtraction + dual energy subtraction

Page 66: DSA  DEEPAK GUPTA

04/11/2023 66

Time interval differencing

In simple mask subtraction, an early image is

used as a mask.

TID – a new mask for each subtraction.

Each subtracted image is the difference b/w the

images separated by some fixed interval of time.

Page 67: DSA  DEEPAK GUPTA

04/11/2023 67

Temporal filtering

Temporal filters are used.

This filter generates one final image by

adding & subtracting some of the

original images together, & then shifting

& repeating to form the next image.

Page 68: DSA  DEEPAK GUPTA

04/11/2023 68

ADVANCEMENTSIN OUR DEPARTMENTPhilips Allura Xper FD 20/10 and interventional tools

The Allura Xper FD20/10 is a biplane flat

detector system with integrated 3D for

intricate neurovascular procedures.

Combining the proven stable geometry with

'real-time' Allura

3D- reconstructions,

Xper CT,

SPECTRA BEAM

3D Roadmapping,

and multi-modality information integration,

Page 69: DSA  DEEPAK GUPTA

04/11/2023 6969 69

Page 70: DSA  DEEPAK GUPTA

04/11/2023 70

the Allura Xper FD20/10 balances

superb image quality with excellent

dose safety.  The flat detector's

unique distortion-free imaging chain

redefines image clarity and captures

information at a resolution four

times greater than conventional X-

ray systems.

Page 71: DSA  DEEPAK GUPTA

04/11/2023 71

Xper ctWith XperCT clinicians can access CT-like imaging right on the angio system so that they can assess soft tissue, bone structure and other body structures before, during or after an interventional procedure. The XperCT reconstruction is created from a rotational acquisition performed on the Allura Xper system. This reconstruction can be overlaid with the 3D vascular image.

IT Bring 3D soft tissue imaging to your angio suite

3D soft tissue imaging supports diagnosis, planning, interventions and treatment follow-up, XperCT can be combined with Allura 3D-RA images to visualize soft tissue and vascular anatomy on one image. Using CT-like imaging for a post-procedural check can help reduce delays and improve patient comfort.

Page 72: DSA  DEEPAK GUPTA

04/11/2023 72

XperCT overlay on both the internal carotid and vertebral arteries

Page 73: DSA  DEEPAK GUPTA

04/11/2023 7373 73

Page 74: DSA  DEEPAK GUPTA

04/11/2023 74

Spectra Beam

It is basically a selectable copper beam filtration

The combination of SpectraBeam with the MRC- tube allows increased X-ray output with better filtration of soft radiation. SpectraBeam offers filters of 200, 500 and 1000 microns thickness Cu equivalent to provide a high level of radiation protection regardless of the projection or patient absorption. This reduces patient X-ray dose for cardiac and vascular applications, while maintaining the same image quality.

Page 75: DSA  DEEPAK GUPTA

04/11/2023 7575 75

When the acquisition sequence is initiated, the C-arm rotates from left

to right for 220o at 40o per second but

dataacquisition is accomplished at the

range Of 200o, acquiring at 8.8 frames per

second

for 5 seconds resulting in 44mask images

The gantry then rotates back to the starting position. When the C-arm returns to the starting position, the Power injector is initiated according to the injector delay protocol, which is Mask-to-Image delay of 1.1 seconds and Injector-to-Image delay of 1.1 seconds.The Power injector injects 18

milliliterof Visipaque 280 at the rate of 3.6ml/second for Internal Carotid Artery(ICA). The gantry repeats the maskacquisition path and acquires

another44 images with contrast. Real timesubtraction angiographs are

displayedon post processing screen.

Page 76: DSA  DEEPAK GUPTA

04/11/2023 7676

System components

GeneratorsX-Ray TubesII/Flat detectorImaging ChainPatient Table

Page 77: DSA  DEEPAK GUPTA

04/11/2023 7777

Generator characteristicsX-ray generators High powered x-ray generators to provide

almost constant potential power supply to the tube.

Two Velara CVFD microprocessor-controlled 100 kW high-

frequency converter generators.

Minimum exposure time of 1 ms.

Voltage range: 40 kV to 125 kV.

Max current:

- MRC-GS 04-07: 802 mA at 80 kV.

- MRC-GS 05-08: 1062.5 mA at 80 kV.

Automatic kV and mA control for optimal image quality prior

to run to reduce dose.

77

Page 78: DSA  DEEPAK GUPTA

04/11/2023 7878

Generator characteristics cont. Max continuous power for fluoroscopy: 2.4 kW for 0.5 hour

and 2 kW for 8 hours.Nominal power (highest electrical power): 100 kW (1000

mA at 100 kV)Reference loading conditions: 110 kV, 18 mA continuous. Pulsed X-ray of 3.75, 7.5, 15 and 30 frames/sec. in

monoplane and biplane mode for pulsed fluoroscopy. Pulsed X-ray of 0.5 to 6 frames/sec. for digital subtracted

acquisition in monoplane and biplane mode.Pulsed X-ray up to 6 frames/sec. for digital acquisition in

monoplane and biplane mode. 15 and 30 frames/sec. Noise < 55 dB(A).

78

Page 79: DSA  DEEPAK GUPTA

04/11/2023 7979

X-Ray Tubes

79

The high performance metal ceramic tubes .Most relevant feature is support of the rotor anode in a spiral groove bearingThe spiral groove bearing glides on a thin film of liquid metal to maximize heat dissipation allowing virtually unlimited X-ray sessions without forced cool down delays.• The MRC tube runs silently offering a more comfortable patient environment.• Another secret of the MRC tube is the enormous heat storage capacity of its 200 mm diameter anode, eliminating waiting times.• And, the MRC X-ray tube can image the heaviest patients even at steep angles, while maintaining image quality.

Page 80: DSA  DEEPAK GUPTA

04/11/2023 8080 80

Tube Specifications

Page 81: DSA  DEEPAK GUPTA

04/11/2023 8181 81

Tube Specifications …

Page 82: DSA  DEEPAK GUPTA

04/11/2023 8282

DetectorsMore recently, however,

the image intensifier/CCD combination has been replaced by an advanced flat detector imaging device.

The flat detector converts the light generated by the absorbed X-ray energy in the scintillation layer directly into a digital electrical signal.

82

Page 83: DSA  DEEPAK GUPTA

04/11/2023 8383

Flat detector frontalSize of housing 42x52 cmMaximum field view 30x38cmImage matrix

2480x1920 pixel at 14 bit heightDetector zoom fields

30x30cm,22x22cm, 16x16cm,13.5x13.5cm,

11x11cm formats Pixel size 154x154 μm Detective Quantum Efficiency

(DQE): >73% at low spatial frequencies

83

Page 84: DSA  DEEPAK GUPTA

04/11/2023 8484

Flat detector lateral

• Size of housing 28x28 cm

• Maximum field view 18x18cm

• Image matrix 1024x1024 pixel at 14 bit height

• Detector zoom fields 14.5x14.5cm, 11x11cm formats

Pixel size 184x184 μm Detective Quantum Efficiency

(DQE): 75%84

Page 85: DSA  DEEPAK GUPTA

04/11/2023 8585

Patient supportAngioDIAGNOST 5

Length: 303 cm (119.3 inch). Metal-free overhang: 125 cm

(49.2 inch). Width: 50 cm (19.7 inch).Longitudinal float: 100 cm (39.4

inch). Transverse float: 36 cm (14.2

inch). Height adjustment (without

pivot option): 76 to 104 cm (29.9 to 40.9

inch). Maximum levelling speed (height

adjustment):- 2 cm/s (0.8 inch/s).Table tilt:L-arc position Angulation

(degrees) Rotation (degrees)Head-end of table 45 Caudal to 45

cranial 27 RAO to 115 RAO85

Page 86: DSA  DEEPAK GUPTA

04/11/2023 8686

Examination lightLight intensity: 30,000 Lux.

Colour temperature: 4300° K.

Focusable light field size: 14 - 25 cm (5.5

- 9.8 inch).

Working distance: 70 -140 cm (27.6 -

55.1 inch).

Light intensity at 30,000 Lux: 114 W/m2.

Lamp type: halogen 22.8 / 24 V 50 W.

Mains power: 220/240 V.

86

Page 87: DSA  DEEPAK GUPTA

04/11/2023 87

Quality Assurance

Weekly 3D calibration is carried out. The calibration data is valid for only 14 days (7 days pre- and post- acquisition).

The calibration is to rectify S-distortion as well as pin cushion distortion.

S-distortion is caused by geo- magnetic field s on moving electrons within the Image Intensifier and pin cushion distortion is caused by the curvature of the input phosphor of the Image Intensifier.

Page 88: DSA  DEEPAK GUPTA

04/11/2023 8888

Radiation Protection- Dose reductionDoseWise is active throughout at every level of system

- from spectra beam, pulsed fluoroscopy to syncratouch. It ensures every possible opportunity to

reduce dose, while maintaining image quality.

DoseWise includes three highly effective strategies for dose management:

Spectra Beam technology to remove unwanted ‘soft’ radiation and minimize scatter radiation, automatically. Spectra Beam filters during fluoro and exposures remove unwanted soft radiation, i.e. those X-rays that hit the patient but do not have enough energy to reach the image detector. In this way, filtering significantly reduces patient dose and scattered radiation for staff while maintaining a high image quality.

 

88

Page 89: DSA  DEEPAK GUPTA

04/11/2023 8989

Advantages Dose reduction3-D road mappingXper CTBolus ChaseQuantitative analysis packages

89

Page 90: DSA  DEEPAK GUPTA

04/11/2023 90

DIS-ADVANTAGES OF DSA

Mis-registration .

Poor resolution compared to conventional

angiography. User dependent success rate. Risk of emboli may reaching to healthy

tissue. Not suitable for everyone.

Page 91: DSA  DEEPAK GUPTA

04/11/2023 9191

ConclusionDespite recent advances in CT angiography and MR

angiography, digital subtraction angiography (DSA) remains the standard imaging technique for evaluation of the cerebral vasculature .

Three-dimensional (3D) reconstruction of the dataset acquired during rotational DSA represents the latest development in the neurovascular imaging armamentarium.

This technique combines the anatomic resolution of DSA with the 3D visualization abilities previously offered by only CT or MR angiography and provides more detailed information than does DSA alone in the evaluation of neurovascular lesions, such as cerebral aneurysms .

3D-DSA has taken a prominent role in treatment planning by enabling better appreciation of the morphology of complex vascular lesions before endovascular or surgical management.

It is also superior in the performance of sophisticated tasks such as aneurysm volume measurement

On the other hand, the inability of 3D-DSA to simultaneously image osseous and vascular structures is noted as a weakness of this technique compared with CT angiography .

91

Page 92: DSA  DEEPAK GUPTA

04/11/2023 92

THANK

YOU

92