maxfacD E N T A L
www.maxfac.no
/ / 2.2/3.0 3.0/3.3 3.0/3.6 3.0/3.8 3.0/3.6/4.1 3.0/3.6/4.3 3.0/4.1/4.6
Short 2D2230S 2D3033S 2D3036S 2D3038S 2D3041S 2D3043S 2D3046S
Long 2D2230L 2D3033L 2D3036L 2D3038L 2D3041L 2D3043L 2D3046L
Coloring Yellow - Green - Blue - Red
Y-Dim. 0.7mm
5HIOSSEN4 123 KIT
Short Long 123 Drill Dia.
Packing unit : Each Part
The color of 123 Drill handle means the diameter and the main fixture to be used. - Yellow: F3.5, Green: F4.0, Blue: F4.5, Red: F5.0Install drill stop to adjust the drilling depth to intended level.
* 123 Twist Drill has good cutting force but the control of drilling depth may be difficult; therefore, it is highly recommended to usethe drill stop.
Packing unit : Each Part
The length of drill stop means what remains of the actual length when the drill stop is installed on 123 Twist Drill.The lengths are differentiated with colors for convenient identification and for the placement in the KIT.
ODS06 ODS07 ODS08 ODS09 ODS11 ODS12 ODS14 ODS16
L(mm) 6.2 7 8 9.5 11 12.5 14 16
Coloring Purple White Yellow Red Blue Green Black Purple
123 KIT Components
TSII SA / SSII SA / USII SA Fixture123 KIT
7 8.5
L
10 11.5 1315
34
40.4
1
1
123 Twist Drill
123 Drill Stop
3.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 F3.5 Cortical drill 3.5 Fixture
Implant placement
Y-Dim.
4.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 4.0 Fixture
Implant placement
F4.0 Cortical drill
Y-Dim.
4.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 3.0/3.6/4.1 F4.0 Cortical drill 4.5 Fixture
Implant placement
Y-Dim.
5.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6/4.1 3.0/4.1/4.6 5.0 Fixture
Implant placement
F5.0 Cortical drill
Y-Dim.
TS/GS fixture implant depth guide- In case of normal or higher-quality bone, it is recommended to implant deeper than bone level by 1mm or less.- In case of soft bone, it is recommended to implant to meet the bone level for maintenance of anchoring force.In case of HA Fixture, it may crack or separate HA coating layer in case of hard bone and the use is not guided.
Straight implants
7HIOSSEN6 123 KIT
TSIII SA / TSIII HA / SSIII SA / USIII SA Fixture
3.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0
F3.5 Cortical drill- Line
F4.0 Cortical drill- Line
F4.0 Cortical drill- Line
F4.5 Cortical drill- Line
F4.5 Cortical drill- Line
F5.0 Cortical drill- Line
F5.0 Cortical drill- Line
F3.5 Cortical drill- Line 3.5 Fixture
Implant placement
Y-Dim.
4.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 4.0 Fixture
Implant placement
Y-Dim.
4.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 3.0/3.6/4.1 4.5 Fixture
Implant placement
Y-Dim.
5.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6/4.1 3.0/4.1/4.6 5.0 Fixture
Implant placement
Y-Dim.
TS/GS fixture implant depth guide- In case of normal or higher-quality bone, it is recommended to implant deeper than bone level by 1mm or less.- In case of soft bone, it is recommended to implant to meet the bone level for maintenance of anchoring force.In case of HA Fixture, it may crack or separate HA coating layer in case of hard bone and the use is not guided.
Care of 123KIT
123 KIT
During surgical procedure, keep used tools in saline solution or
distilled water.
When the surgery has been completed, soak the used tools in alcohol
for cleaning.
Washing with hydrogen peroxide is prohibited.
Exposure to hydrogen peroxide may discolor the laser
marking and TiN coating.
Wash blood stains and other foreign matter clean with distilled water
or flowing water.
Remove moisture with a dry cloth or a hot air blower.
Place the dried tools in the KIT case.
(Refer to the color coding for placement of tools in the kit case.)
After placement, sterilize the kit in an autoclave at 132 for 15
minutes and store at room temperature.
Caution: After a surgical procedure, separate all the tools used in the surgeryimmediately, and clean and sterilize before storage.It is highly recommended to sterilize the Surgical KIT before and after eachsurgical procedure in an Autoclave (temperature: 132 C, time: 15 min)The warranty period of the Surgical KIT is One Year after first opening thepackage, and the warranty cycles of the Drills and Drivers is 50 cycles.
Caution
Taper implants Tlf. 22 20 50 50 • www.osstem.no
maxfacD E N T A L
www.maxfac.no
3HIOSSEN2 123 KIT
123 KIT 123 KIT
What is 123 KIT 123 KIT (H123K)
123 KIT features
123KIT reduces the number of conventional drilling sequence to once in soft bone, to twice in normalbone, and 3 times in hard bone. Both patients and physicians can both have a win-win by shorteningthe time of implant surgery. The 123Kit can also maximize the convenience for the users by integratingdrilling sequences for TS, SS, and US Fixture.
(* Use only 123KIT for SA or HA surface-treated fixture.)
1. Speedy Drilling Sequence: Only three times
(TSII 5.0 Fixture: Hard bone)
2. Simple & Easy Sequence: Unification of drilling sequence
(Separate drilling sequence by system Unified drilling sequence by type.)
3. Design of drill stop
: Drill Stop designed with fixture lengths, the drilling depth can easily be adjusted safely and conveniently.
Guide Drill 1 2 3 4 5 6 7
Conventional
Drilling
Sequence
123
Drilling
Sequence
One sequence
TSII SSII USII
One sequence
TSIII SSIII USIII
Use range (Use )
TSII SA SSII SA USII SA GSII
TSIII SA/HA SSIII SA USIII SA GSIIIUltra-wide MS OS
Guide DrillGD2027L
SideCut DrillOSLMD20L
Parallel PinOPP400: 2eaOPP500: 2ea
Drill ExtensionADE
Depth GaugeODG
Torque WrenchTW30B
Drill StopODS06, ODS07, ODS08, ODS09, ODS11, ODS12,
ODS14, ODS16
123 Cortical Drill2CD35, 2CD40, 2CD45,2CD50, 3CD35, 3CD40,3CD45, 3CD50
Removal Toolfor Mount
HRFRHand DriverAHD12SHAHD12LH
Torque DriverTRHD12STRHD12L
Simple MountDriverASMDS, ASMDL
Simple MountExtensionASMEL
Fixture DriverGSMFDLGSRGDL
NoMount DriverGSNMD32LGSNMD35L
123 Twist Drill2D2230S, 2D2230L,2D3036S, 2D3036L,2D3041S, 2D3041L,2D3046S, 2D3046L
Tlf. 22 20 50 50 • www.osstem.no
maxfacD E N T A L
www.maxfac.no
Tlf. 22 20 50 50 • www.osstem.no
maxfacD E N T A L
www.maxfac.no
7HIOSSEN6 123 KIT
TSIII SA / TSIII HA / SSIII SA / USIII SA Fixture
3.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0
F3.5 Cortical drill- Line
F4.0 Cortical drill- Line
F4.0 Cortical drill- Line
F4.5 Cortical drill- Line
F4.5 Cortical drill- Line
F5.0 Cortical drill- Line
F5.0 Cortical drill- Line
F3.5 Cortical drill- Line 3.5 Fixture
Implant placement
Y-Dim.
4.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 4.0 Fixture
Implant placement
Y-Dim.
4.5 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6 3.0/3.6/4.1 4.5 Fixture
Implant placement
Y-Dim.
5.0 Fixture (Length : 10mm)
Soft
Normal
Hard
10mm
Guide DrillBonequality 2.2/3.0 3.0/3.6/4.1 3.0/4.1/4.6 5.0 Fixture
Implant placement
Y-Dim.
TS/GS fixture implant depth guide- In case of normal or higher-quality bone, it is recommended to implant deeper than bone level by 1mm or less.- In case of soft bone, it is recommended to implant to meet the bone level for maintenance of anchoring force.In case of HA Fixture, it may crack or separate HA coating layer in case of hard bone and the use is not guided.
Care of 123KIT
123 KIT
During surgical procedure, keep used tools in saline solution or
distilled water.
When the surgery has been completed, soak the used tools in alcohol
for cleaning.
Washing with hydrogen peroxide is prohibited.
Exposure to hydrogen peroxide may discolor the laser
marking and TiN coating.
Wash blood stains and other foreign matter clean with distilled water
or flowing water.
Remove moisture with a dry cloth or a hot air blower.
Place the dried tools in the KIT case.
(Refer to the color coding for placement of tools in the kit case.)
After placement, sterilize the kit in an autoclave at 132 for 15
minutes and store at room temperature.
Caution: After a surgical procedure, separate all the tools used in the surgeryimmediately, and clean and sterilize before storage.It is highly recommended to sterilize the Surgical KIT before and after eachsurgical procedure in an Autoclave (temperature: 132 C, time: 15 min)The warranty period of the Surgical KIT is One Year after first opening thepackage, and the warranty cycles of the Drills and Drivers is 50 cycles.
Caution