Dr Tse, Yee WahDr Koo, Chi Kwan
Dr Tan, Song TuenDr Chan, Yau Wai
Queen Elizabeth Hospital, Hong Kong
Sudden blackout of the Dräger Zeus Anaesthetic Machine
![Page 2: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/2.jpg)
Ventilator failure is not an uncommon incident in operating theatres
Dräger Zeus Anaesthetic Machine failure: not been reported before
Background
![Page 3: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/3.jpg)
Background
TCA(TM) = Target-Controlled Anesthesia automatic control of gas dosagesmanual controltouch screen
Drager Zeus Anaesthetic machine
![Page 4: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/4.jpg)
elective hepatectomy Dräger Zeus Anaestehtic machine failure:
unresponsive touch screen &rebooted subsequently
laid by standby screen shortly after diathermy usage
Incident
![Page 5: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/5.jpg)
Patient disconnected from ventilator, ventilated by manual resuscitator
Dräger Zeus ventilator was replaced by Dräger Narkomed IV
photos were taken to record the setup layout for later investigation
Incident
![Page 6: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/6.jpg)
Preliminary examination by the Dräger Company: no external defects or damages PASS: Self test, gas delivery, ventilatory
test analysis of the device status report:
USB port error of the device computer in the ventilator
Incident
![Page 7: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/7.jpg)
(1) Power cord of electric blanket & (2) data cable
![Page 8: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/8.jpg)
Electrical interference rebooted shortly after the use of diathermyWhere? Between diathermy & electric blanket (the
only device that was plugged to the auxiliary power socket of the ventilator)
data transmission in USB port was affected
Postulation
![Page 9: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/9.jpg)
Simulation
Devices:(1) Dräger Zeus (2) Conmed 6400 (argon beam coagulator) (3) Cusa 200 (ultrasonic surgical aspirator) (4) Force FX (diathermy)(5) KanMed 202W (electric blanket control unit) (6) HP monitorOWON oscilloscopeLEM HEME LH240 ammeter
![Page 10: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/10.jpg)
Simulation
![Page 11: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/11.jpg)
(1) Power cord of electric blanket & (2) data cable
![Page 12: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/12.jpg)
Diathermy use on pork
![Page 13: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/13.jpg)
(1) Oscilloscope & (2) Ammeter connected to power cord
![Page 14: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/14.jpg)
To prove the postulation…
1. Control (no diathermy): no failure2. Diathermy in cutting mode: no failure3. Diathermy in coagulation mode: FAILURE4. without data cable connection: no failure5. with use of optical relay: no failure
![Page 15: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/15.jpg)
Discussion – 2 circuits
![Page 16: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/16.jpg)
Discussion
Capacitive coupling in any eletrosurgery:
high frequency current in a conductor caninduce current in another conductor separated by an insulator
Conductors: diathermy tip & metal wirings in electric blanket
Insulator: gel pad
![Page 17: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/17.jpg)
Discussion – why coagulation mode?
Why diathermy in coagulation mode?
![Page 18: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/18.jpg)
Discussion
Scenarios / Parameters
Voltage (V) Current (A) Frequency (Hz)
Control 8 0.76-1.13 50Cutting mode (90W)
16 0.15-0.42 Out of range
Coagulation mode (90W)
86.4 0.15-0.58 28600
![Page 19: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/19.jpg)
Discussion
without data cable connection: no failure with use of optical relay: no failure
![Page 20: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/20.jpg)
Solutions
Do not use the auxiliary power socket in the Dräger Zeus
Use of an optical signal relay (to interrupt the passage of electric interference via the USB port of the device computer)
![Page 21: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/21.jpg)
Learning points
Electrical safety is an important issue in OT Proper connection of each device requires
an anesthetist’s attention any electrical fault can lead to potential/
actual crisis
![Page 22: Dr Tse, Yee Wah Dr Koo, Chi Kwan Dr Tan, Song Tuen Dr Chan, Yau Wai Queen Elizabeth Hospital, Hong Kong Sudden blackout of the Dräger Zeus Anaesthetic.](https://reader033.fdocuments.in/reader033/viewer/2022061601/56649cf75503460f949c7669/html5/thumbnails/22.jpg)
References Myerson, K R. Potential hazard of East-Radcliffe ventilator. Failure to entrain adequate oxygen. Anaesthesia. 40(3):285-7,
1985 Mar. Williams DJ. Ventilator failure due to non-ISO-standard components. [Case Reports. Letter] Anaesthesia. 56(4):392-3,
2001 Apr. Sosis MB. Drager ventilator failure on changing the respiratory rate setting.[see comment]. [Case Reports. Letter]
Anesthesia & Analgesia. 76(2):453-4, 1993 Feb. Ananthanarayan C. Fisher JA. Drager ventilator failure.[comment]. [Comment. Letter] Anesthesia & Analgesia. 77(3):638,
1993 Sep. Barahal D. Sims C. Ventilator failure during use of a new anesthesia machine. [Letter] Anesthesiology. 95(2):568-9, 2001
Aug. Rafael A. Ortega, M.D.,* Bruce Vrooman, M.D., Rania Hito, M.S. Another Cause for Ventilator Failure. Anesthesiology
2006; 104:1351 Ortega RA. Zambricki ER. Fresh gas decoupling valve failure precludes mechanical ventilation in a Draeger Fabius GS
anesthesia machine. [Case Reports. Letter] Anesthesia & Analgesia. 104(4):1000; discussion 1000-1, 2007 Apr. Photo of Dräger Zeus ventilator from http://medgadget.com/archives/2006/04/zeus_anesthesia.html Photo of electric blanket from User Manual Kanmed Operatherm 202
www.kanmed.se/download/download_files/operaterm/op-202-050_vers5_uk.pdf Haag R. Cuschieri A. Recent advances in high-frequency electrosurgery: development of automated systems. [Review].
Journal of the Royal College of Surgeons of Edinburgh. 38(6):354-64, 1993 Dec. Force FXTM Electrosurgical Generator Technical Specifications
http://www.valleylab.com/product/es/generators/pdf/Lforcefx.pdf (page 4) Tucker RD. Voyles CR. Laparoscopic electrosurgical complications and their prevention. [Review] [16 refs] [Case Reports.
Journal Article. Review] AORN Journal. 62(1):51-3, 55, 58-9 passim; quiz 74-7, 1995 Jul.