Dose Optimization in Interventional Radiology
Transcript of Dose Optimization in Interventional Radiology
Dose Optimization in
Interventional Radiology
Aida M. Lobriguito, MSc, CSci, MIPEM
Medical Physicist
King Fahad Medical City
National Training Course on Justification and Optimization
of Protection in Diagnostic X-ray Imaging and Interventional
Radiology (18-20 October 2016)
INTRODUCTION
% IR Procedures from US Survey
Source: NCRP Report No. 160 Ionizing Radiation Exposure of the Population of the US
What is Interventional
radiology ? Interventional radiology is concerned with
providing diagnosis and treatment of disease
by a variety of percutaneous procedures
performed under the guidance of radiologic
imaging performed by :
Radiologists
Cardiologists
Neurologists
Other clinicians
Who are at risk in IR
procedures?
Patients
Staff
For skin injuries and eye lens
opacities (cataract)
What is Optimization?
Optimization is the radiation
dose to the patient that is
commensurate with the
medical purpose and
avoidance of radiation that is
clinically unnecessary or
unproductive.
It involves the design,
appropriate selection, and
use of equipment, protection
tools and working
procedures.
Interventional
Radiology
CT
Radiography
Comparison of radiography with IR
and cardiology doses
Procedure Mean Effective Dose
(mSv)
Chest (radiography) 0.03
Abdomen (radiography) 0.6
ERCP (endoscopic retrograde 3.9
choangiopancreatography)
Cardiology:
Coronary angiography (CA) 3.1
PTCA (percutaneous trnasluminal 15.1
coronary angioplasty)
Cardiovascular embolization 19.5
Interventional radiology:
Renal angiography 13.7
Vascular stenting 10.4
TIPS (transjugular intrahepatic 53.6
portosystemic shunt)
PATIENTS: RISKS AND
SAFETY
Deterministic Effects
• Tissue reactions/Skin injuries
• Cataract
Skin injury from
Transjugular
Intrahepatic
Portosystemic Shunt -
TIPS
Stochastic Effects
Procedure Approx. onset 1. Erythema at 2Gy hours 2. Permanent epilation at 7Gy 3 wk 3. Delayed skin necrosis at 12 Gy >1 yr 4. Moist desquamation at 18 Gy 4 wk
5
Acute radiation doses, delivered to tissues
during a single procedure or closely
spaced procedures, will cause:
(a) 6-8 weeks after multiple coronary angiography and angioplasty
procedures.
(b) 16-21 weeks
(c) 18-21 months after the procedures showing tissue necrosis .
(d) Close-up photograph of the lesion shown in (c).
(e) Photograph after skin grafting. (Photographs courtesy of T. Shope & ICRP).
(d) (e)
(a) (c)
(b)
Coronary angioplasty twice in a day followed
by bypass graft because of complication. Dose
20 Gy (ICRP 85)
Deterministic effects caused by cell death:
burns, organ failure, death
IAEA Training Course on Radiation Protection for Doctors (non-radiologists, non-cardiologists) using
Fluoroscopy
At 3 wks At 6.5 mos Surgical flap
Following ablation procedure with arm in beam near port and separator
one removed. About 20 minutes of fluoroscopy.
Reproduced from Wagner – Archer,
Minimizing Risks from Fluoroscopic X Rays,
3rd ed, Houston, TX, R. M. Partnership, 2000
Reports Received by FDA of Skin Injury from Fluoroscopy.
Procedure with
Report of Injury
Number of Injuries
Reported from Procedure
RF cardiac catheter ablation 12 Catheter placement for chemotherapy 1 Transjugular interhepatic portosystemic shunt 3 Coronary angioplasty 4 Renal angioplasty 2 Multiple hepatic/biliary procedures 3 (angioplasty, stent placement, biopsy, etc.) Percutaneous choloangiogram followed 1 by multiple embolizations
Skin Injuries: IR and IC
Why do they occur?
No training in radiation protection
for those performing these studies,
of the:
• Cardiologist
• Urologist
• Gastro-enterologist
• Orthopedic Surgeon
• Vascular Surgeon
• Traumatologist
• Pediatrician
• Anesthesiologist
Skin Injuries are
associated with:
• Complex procedure with lengthy time
over the same skin area
• Prolonged use of high-dose modes of
operation
• Unnecessary body parts in the field of the X
ray beam
• Large patients or steep beam angles requiring
transmission through thick body mass
Skin Injuries are
associated with:
• Patients with health conditions that
predispose them to radiation injury
• Multiple procedures in the same patient over
short periods of time
• Lack of dose monitoring to warn physicians
that high doses are accumulating
• Inexperienced physicians
Radiation Dose-determining Factors
Equipment
Patient Weight
Diagnostic vs. Interventional
Complexity of the Procedure
Constant potential generator
Arc system (X Ray tube down)
High efficiency intensifier/
flat panel detectors
Easy operational controls
Good image saving and retrieving (last hold)
HOW MUST BE AN X Ray SYSTEM BE "SPECIFICALLY DESIGNED" FOR
INTERVENTIONAL RADIOLOGY?
PATIENT DOSE REDUCTION
MEASURES
Performance of the X Ray system used
Old versus new
Use of dedicated units
Well calibrated
With dose reduction features
Knowledge of the operator on the use of
the unit
Fluoroscopy time
Not a good indicator of dose but is good
to evaluate quality of performance
IR/Cardiology procedures
Important factors for dose reduction
Complexity of procedures
Needs training of physician on the procedure
Number of series (Images)
Use cine sparingly.
For pediatric patients at least 60 frames per
sec.
Patient size - affects exposure factors
Available protection tools
Dose monitoring and management
IR/Cardiology procedures
Important factors for dose reduction
THE KNOWLEDGE OF DOSE RATES FOR
DIFFERENT OPERATIONAL MODES AND
FOR DIFFERENT INTENSIFIER INPUT SIZE
IS IMPORTANT
THEN, IT IS POSSIBLE TO HAVE
CRITERIA FOR THE CORRECT
USE OF DIFFERENT OPERATION
MODES
Perform routine and annual QC tests and make
corrections when needed.
Establish image quality criteria.
QA/QC
Equipment
Patient Dose Management
Justification of procedure
Equipment condition – QC performed, well
calibrated
Procedure to be done: simple or complex??
Expected doses known? Verified?
Informed consent
Before the Procedure
During the Procedure
• Area Product (KAP or DAP)
mGy mGy
mGy
KAP or DAP meter - needs proper calibration.
Values can be used to determine equivalent or
effective dose to the whole body.
Measure Patient Doses
Setting of Dose Notification
New machines have alarm system.
Set Rules for patient follow up
Establish Reference levels
“Patient risk”
“Clinical protocol”
“Equipment performance”
Reference levels (indicative of the state of the
practice): an instrument to help operators to conduct
optimized procedures with reference to patient
exposure
Required by international (IAEA) and national
regulations For complex procedures reference levels should
include:
the complexity of the procedures.
more parameters like:
(European Dimond Consortium recommendations)
Procedure: CA PTCA
DAP (Gycm2) 57 94
Fluoroscopy time (min) 6 16
No. of frames 1270 1355
Reference levels in interventional
cardiology
(European proposal 2003)
DIMOND EU project. E.Neofotistou, et al, Preliminary reference levels in interventional
cardiology, J.Eur.Radiol, 2003
Dosimetry Records
• Measure and Record Patient Radiation Dose
• Record Fluoroscopy Time
• Record Available Measures
– DAP (Dose Area Product)
– Cumulative Dose
– Skin Dose
• Use dose conversion factors such as effective
dose per DAP
After the Procedure
Dosimetry Follow-UP
• Develop Methods to Quantify Late Effects
– Design medical records to clearly document the number
and types of interventional procedures received by the
patient.
– Maintain a database of all patients with procedures and
dose information.
– Review dose information to identify patients with high
doses (>3Gy) for follow-up.
– Establish procedures for follow-up; including skin
examination at 30 days.
After the Procedure
• Remove the grid during procedures on small patients or when the image intensifier cannot be placed close to the patient.
• When the procedure is unexpectedly prolonged, consider positioning the patient or altering the X ray field or other means to alter beam angulation so that the same area of skin is not continuously in the direct X Ray field.
• Patients should be counseled on radiation risks if the procedure carries a significant risk of such injury.
Practical Actions in Controlling
Patient Doses
Practical Actions in Controlling Dose
• Records of exposure should be kept if the estimated maximum cumulative dose to skin is 3Gy or above.
• All patients with estimated skin doses of 3 Gy or above should be followed up 10 to 14 days after exposure.
• The patient’s personal physician should be informed of the possibility of radiation effects.
• If the dose is sufficient to cause observable effects,
• the patient should be counseled after the procedure.
• A system to identify repeated procedures should be set up.
IAEA-KFSHRC Highest MSD and DAP values
for 9 Procedures
Procedure Highest MSD Highest Total DAP
(mGy) (cGy-cm2)
CA 889.2 13,784
PTCA 4,470 33,658
TIPSS 734.7 17,177
Cerebral 713.8 Not available
PTC 983.0 not available
Abdomen 675.6 30,794
Nephrostomy 84.1 1,887
Gastrostomy 349 6,342.7
Embolization 499.9 11, 498.4
Findings & Recommendations
A review of the protocols for TIPSS and PTCA
should be made and standardized. Dose reductions
techniques should be determined.
High DAP values are due to cineradiography as can be
seen from the fluoro time variances. Dose reduction
techniques for should be considered such as no. of
frames.
Image quality criteria and assessment should be set
and standardized.
Counseling of patients especially for PTCA should be
made when doses exceed 2 Gy and dose should be
recorded on the patient’s chart.
Findings & Recommendations
Skills and training needs of operators should be
assessed and structured training program should be
developed.
Introduction of reference dose levels should be made.
Although the use of Mathlab based program showed
to be efficient in dose evaluation, more clinical trials
should be made by intercomparing with other
institutions.
More data should be collected for low frequency
procedures for validation.
Golden Rules
• Radiation dose to the patient should be limited to
that required for the procedure being performed
• Appropriate collimation
• Patient should be positioned as close as reasonably
possible to the image receptor
• Distance between the patient and the X-ray tube
should be maximized to the extent practicable
• Each arm of the patient should be kept outside the
radiation field unless an arm is intentionally imaged
as part of the procedure
Golden Rules
• Lowest dose rate that is clinically
• When electronic magnification is necessary, the lowest
acceptable magnification factor should be used
• Fluoroscopy should be used sparingly and only when
real-time imaging guidance is needed
• Last-image-hold feature or loop replay should be used
whenever possible
• Image acquisition should be activated only when higher
quality image review is essential, and it should be limited
to the frame rate and run duration necessary to
accomplish the immediate task
• In some cases, retrospectively stored fluoroscopy may
reduce the need for image acquisition
STAFF: RISKS AND SAFETY
Deterministic effects in cardiology
studies
Skin injuries on hand
Eye cataract
Staff
Shielding & Positioning Eye Shielding Is Imperative
• Leaded Ceiling Shield 98% reduction.
• Leaded Glasses Shield ~74-91%
• Differences in performance related to operator position, likely representing interplay of design and fit.
• Sports wraparound or side shields important. Rehani. Cataract RASSC Dec
2011
Diligent use personal protective
equipment (PPE).
Lead goggles provide effective
protection but there are issues
of comfort.
Use of suspended shields
protects head and neck.
Use of PPE can reduce eye
doses to about 3/10 of the dose
limit.
What can be done in the light of the new
recommendation for the lens of the eye?
It is important to note that…
Reported eye lens doses:
0.3-11 mGy/study (without use of
protective devices)
0.011-0.33 mGy/study (with protective devices)
Rehani. Cataract RASSC Dec
2011
45
3 As
Awareness
Appropriateness
Audit
Thank you !