DOC316.53.01114_Ed7

download DOC316.53.01114_Ed7

of 8

Transcript of DOC316.53.01114_Ed7

  • 8/12/2019 DOC316.53.01114_Ed7

    1/8

    Phosphorus, Reactive

    Page 1 of 8

    Phosphorus, Reactive HR, 8114

    How to use instrument-specific information

    The Instrument-specific informationtable displays requirements that may vary between

    instruments. To use this table, select an instrument then read across to find the corresponding

    information required to perform this test.

    Phosphorus, Reactive DOC316.53.01114Molybdovanadate Rapid Liquid Method1

    1 Adapted from Standard Methods for the Examination of Water and Wastewater.

    Method 8114

    HR (0.3 to 45.0 mg/L PO43) Pour-Thru Cell

    Scope and Application: For treated and natural waters

    Test preparation

    Table 344 Instrument-specific information

    Instrument Pour-Thru Kit Pour-Thru Cell Orientation Adapter

    DR 6000 LQV175.99.20002 Arrow faces right

    DR 5000 LZV479

    DR 3900LQV157.99.10002 Align cell flow arrows with arrows on cell

    compartment

    DR 3800, DR 2800, DR 27005940400 1-inch (round) path aligned with arrow on

    the adapter

    LZV585 (B)

    Before starting the test:

    See the User Manual for Pour-Thru Module installation instructions.

    Clean the Pour-Thru cell and all labware as specified inAnalysis labware treatment.

    Protect the Pour-Thru Cell from contamination when not in use by inverting a small beaker over the top of the glass funnel.

    Make sure the PourThru cell is completely seated in the sample cell compartment.

    Final samples will contain molybdenum. In addition, final samples will have a pH less than 2 and are considered corrosive

    (D002) by the Federal RCRA. Refer to the current MSDS for safe handling and disposal information.

    Description Quantity

    Molybdovanadate Reagent 2 mL

    Water, deionized 25 mL

    Cylinder, graduated, 25-mL, poly 1

    Dispenser, fixed volume, 1.0-mL, w/bottle 1

    Flask, Erlenmeyer, 125-mL, PMP w/cap 2

    Instrument-specific information 1

    http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/
  • 8/12/2019 DOC316.53.01114_Ed7

    2/8

    Phosphorus, Reactive

    Page 2 of 8

    Phosphorus, Reactive

    See Consumables and replacement itemsfor reorder information.

    Molybdovanadate Rapid Liquid method

    1. Select the test.

    Insert an adapter if

    required (see Instrument-

    specific information).

    2. I 3. Rinse a clean plastic

    125-mL Erlenmeyer flask

    and a 25-mL graduatedcylinder with deionized

    water.

    4. Blank Preparation:

    Measure 25 mL of

    deionized water in thegraduated cylinder. Pour

    the water into the flask.

    5. Rinse another clean

    plastic 125-mL Erlenmeyer

    flask with deionized water.

    6. Prepared Sample:

    Measure 25 mL of sample

    in the graduated cylinder.

    Pour the water into the

    flask.

    7. Add 1.0 mL of

    Molybdovanadate

    Reagent to each flask

    using a Repipet Jr.

    Dispenser. Swirl to mix.

    A yellow color will develop

    in the sample if phosphate

    is present. A small amount

    of yellow may be present

    in the blank due to the

    reagent.

    8. Start the instrument

    timer.

    A 7-minute reaction period

    will begin.

    If the sample

    concentration is greater

    than 30 mg/L PO43, read

    at exactly seven minutes

    or make a 1:1 dilution of

    the sample and begin the

    test again.

    Description Quantity

    489 P React. Mo. HR RL

    Stored Programs

    Start

  • 8/12/2019 DOC316.53.01114_Ed7

    3/8

    Phosphorus, Reactive

    Phosphorus, Reactive

    Page 3 of 8

    Interferences

    See the Interfering substancestable for a list of substances, interference levels and type of

    interference. See the Noninterfering substances at low concentrations (less than 1000 mg/L)table

    for a list of substances that do not interfere in concentrations less than 1000 mg/L.

    9. When the timer

    expires, pour the blank

    from the flask into the

    Pour-Thru Cell.

    10. ZEROthe instrument.

    The display will show:

    0.0 mg/L PO43

    11. Pour the prepared

    sample from the flask into

    the Pour-Thru Cell.

    12. READthe results in

    mg/L PO43.

    Flush the Pour-Thru Cell

    with 50 mL of deionized

    water.

    Table 345 Interfering substances

    Interfering substance Interference level

    Arsenate Negative interference. Positive interference if sample is heated.

    Bismuth Negative interference.

    Fluoride Negative interference.

    Iron, FerrousBlue color is caused by ferrous iron but this does not affect results if the ferrous ironconcentration is less than 100 mg/L.

    Molybdate Negative interference.

    Silica Positive interference if sample is heated.

    Sulfide

    Negative interference. Sulfide interference may be removed by oxidation with Bromine Water

    as follows:

    1. Measure 25 mL of sample into a flask.

    2. Add Bromine Water1drop-wise with constant swirling until permanent yellow color

    develops.

    3. Add Phenol Solution1drop-wise until the yellow color just disappears. Proceed with

    step 7.

    1 See Optional reagents and apparatus.

    Thiocyanate Negative interference.Thiosulfate Negative interference.

    Thorium Negative interference.

    Highly buffered samples or

    extreme sample pHMay exceed the buffering capacity of the reagents and require sample pretreatment.

    Molybdovanadate Rapid Liquid method (continued)

    Zero Read

  • 8/12/2019 DOC316.53.01114_Ed7

    4/8

    Phosphorus, Reactive

    Page 4 of 8

    Phosphorus, Reactive

    Sample collection, preservation and storage

    Collect samples in clean plastic or glass bottles that have been cleaned with 1:1 Hydrochloric

    Acid Solution*and rinsed with deionized water.

    Do not use detergents that contain phosphate for cleaning labware.

    Analyze samples immediately for best results.

    If prompt analysis is not possible, preserve samples by filtering immediately and storing the

    sample at 4 C (39 F) or below for up to 48 hours.

    Warm stored samples to room temperature before analyzing.

    Analysis labware treatment

    1. Clean containers by normal means (do not use detergents containing phosphorus), then rinse

    with deionized water.

    2. Soak for several minutes in a 1:25 dilution of Molybdovanadate Reagent in deionized water.

    3. Rinse well with deionized water. Dedicate these containers for HR PO43analysis.

    4. Fill the Pour-Thru Cell with this same mixture of Molybdovanadate reagent and deionized

    water and let stand for several minutes.

    5. Rinse the Pour-Thru Cell with 50 mL of deionized water.

    Table 346 Noninterfering substances at low concentrations (less than 1000 mg/L)

    Pyrophosphate Tetraborate Benzoate

    Citrate Lactate Formate

    Oxalate Tartrate Salicylate

    Al3+ Selenate Mg2+

    Ca2+ Ba2+ Sr2+

    Li+ Na+ K+

    NH4+ Cd2+ Mn2+

    NO3 NO2

    SO42

    SO32 Pb2+ Hg+

    Hg2+ Sn2+ Cu2+

    Ni2+ Ag+ U

    Zr4+ AsO3 Br

    CO32 ClO4

    CN

    IO3 Fe3+ SiO44

    * See Optional reagents and apparatus.

  • 8/12/2019 DOC316.53.01114_Ed7

    5/8

    Phosphorus, Reactive

    Phosphorus, Reactive

    Page 5 of 8

    How to clean the Pour-Thru Cell

    The Pour-Thru Cell may accumulate a buildup of colored products, especially if the reacted

    solutions are allowed to stand in the cell for long periods after measurement. To clean the Pour-

    Thru Cell:

    1. Remove the color by rinsing with a 1:5 dilution of ammonium hydroxide*

    2. Follow with several deionized water rinses.

    3. Invert a beaker over the glass funnel of the Pour-Thru Cell when not in use.

    Accuracy check

    Standard additions method (sample spike)

    Required for accuracy check:

    Phosphate VoluetteAmpule Standard Solution, 500-mg/L as PO43

    Ampule breaker

    TenSette Pipet and tips

    Mixing cylinders (3)

    1. After reading test results, leave the sample cell (unspiked sample) in the instrument.

    2. Select standard additions from the instrument menu: OPTIONS>MORE>STANDARD

    ADDITIONS.

    3. Accept the default values for standard concentration, sample volume and spike volumes. After

    the values are accepted, the unspiked sample reading will appear in the top row. See the user

    manual for more information.

    4. Open the standard solution ampule.

    5. Use the TenSette Pipet to prepare spiked samples: add 0.1 mL, 0.2 mL and 0.3 mL of

    standard to three 25-mL mixing cylinders and fill to 25 mL with the fresh sample. Mixthoroughly.

    6. Follow the Molybdovanadate Rapid Liquid methodtest procedure for each of the spiked

    samples, starting with the 0.1 mL sample spike. Measure each of the spiked samples in the

    instrument.

    7. Select GRAPHto view the results. Select IDEAL LINE (or best-fit) to compare the standard

    addition results to the theoretical 100% recovery.

    * Optional reagents and apparatus.

  • 8/12/2019 DOC316.53.01114_Ed7

    6/8

    Phosphorus, Reactive

    Page 6 of 8

    Phosphorus, Reactive

    Standard solution methodNote: Refer to the instrument user manual for specific software navigation instructions.

    Required for accuracy check:

    10.0-mg/L Phosphate Standard

    1. Use the 10.0-mg/L Phosphate Standard solution in place of the sample. Follow the

    Molybdovanadate Rapid Liquid methodtest procedure.

    2. To adjust the calibration curve using the reading obtained with the standard solution, navigate

    to Standard Adjust in the software: OPTIONS>MORE>STANDARD ADJUST.

    3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate

    concentration is used, enter the concentration and adjust the curve to that value.

    Method performance

    Summary of method

    In the molybdovanadate method, orthophosphate reacts with molybdate in an acid medium to

    produce a phosphomolybdate complex. In the presence of vanadium, yellow

    vanadomolybdophosphoric acid is formed. The intensity of the yellow color is proportional to the

    phosphate concentration. Test results are measured at 430 nm.

    Program Standard

    Precision

    95% Confidence Limits of

    Distribution

    Sensitivity

    Concentration change

    per 0.010 Abs change

    489 10.0 mg/L PO43 9.910.1 mg/L PO4

    3 0.34 mg/L PO43

  • 8/12/2019 DOC316.53.01114_Ed7

    7/8

    Phosphorus, Reactive

    Phosphorus, Reactive

    Page 7 of 8

    Consumables and replacement items

    Required reagents

    Description Quantity/Test Unit Catalog number

    Molybdovanadate Reagent 2 mL 500 mL 2076049

    Water, deionized 25 mL 4 L 27256

    Required apparatus

    Description Quantity Unit Catalog number

    Cylinder, graduated, 25-mL, poly 1 each 108140

    Dispenser, adjustable 1 each 2563137

    Flask, Erlenmeyer, 125-mL, PMP w/cap 2 each 2089843

    Recommended standards

    Description Unit Catalog number

    Phosphate Standard Solution, 10-mg/L as PO43 946 mL 1420416

    Phosphate Standard Solution, Voluetteampule, 10-mL, 500-mg/L as PO43 16/pkg 1424210

    Wastewater Influent Standard for NH3N, NO3N, PO4, COD, SO4, TOC 500 mL 2833149

    Optional reagents and apparatus

    Description Unit Catalog number

    Ammonium Hydroxide, ACS 500 mL 10649

    Bromine Water, 30 g/L 29 mL 221120

    Cylinder, graduated mixing, 25 mL each 189640

    Hydrochloric Acid Solution, 6.0 N, 1:1 500 mL 88449

    Phenol Solution, 30 g/L 29 mL 211220

    Pipet, TenSette, Pipet, 0.1 - 1.0 mL each 1970001

    Pipet Tips, for TenSette Pipet 19700011

    1 Other sizes are available

    50/pkg 2185696

    Pipet Tips, for TenSette Pipet 19700011 1000/pkg 2185628

    Filter Paper, folded 12.5 cm 100/pkg 69257

    Filter Funnel, Analytical, 65 mL each 108367

    Thermometer, Non-Mercury, -10 to 225 C each 2635700

    Sampling Bottle with cap, low density polyethylene, 250 mL 12/pkg 2087076

    http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2076049http://www.hach.com/hc/search.product.details.invoker/PackagingCode=27256http://www.hach.com/hc/search.product.details.invoker/PackagingCode=27256http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2076049
  • 8/12/2019 DOC316.53.01114_Ed7

    8/8

    HACH COMPANYWORLD HEADQUARTERSTelephone: (970) 669-3050

    FAX: (970) 669-2932

    FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:Call 800-227-4224

    Contact the HACH office or distributor serving you.

    www.hach.com [email protected]

    In the U.S.A.

    Outside the U.S.A.

    On the Worldwide Web ; E-mail

    toll-free

    Phosphorus, Reactive

    Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7

    Optional standards

    Description Unit Catalog number

    Voluette Ampule breaker, 10 mL each 2196800

    Phosphate, Standard Solution, 3 mg/L 946 mL 2059716

    Phosphate, Standard Solution, 15 mg/L 100 mL 1424342

    Phosphate, Standard Solution, 30 mg/L 946 mL 1436716

    Phosphate, Standard Solution, 50 mg/L, 10 mL Voluette Ampules 16/pkg 17110

    Phosphate, Standard Solution, 100 mg/L 100 mL 1436832

    Phosphate, Standard Solution, 500 mg/L 100 mL 1424232