DOC316.53.01114_Ed7
-
Upload
ardhito-setiawan -
Category
Documents
-
view
215 -
download
0
Transcript of DOC316.53.01114_Ed7
-
8/12/2019 DOC316.53.01114_Ed7
1/8
Phosphorus, Reactive
Page 1 of 8
Phosphorus, Reactive HR, 8114
How to use instrument-specific information
The Instrument-specific informationtable displays requirements that may vary between
instruments. To use this table, select an instrument then read across to find the corresponding
information required to perform this test.
Phosphorus, Reactive DOC316.53.01114Molybdovanadate Rapid Liquid Method1
1 Adapted from Standard Methods for the Examination of Water and Wastewater.
Method 8114
HR (0.3 to 45.0 mg/L PO43) Pour-Thru Cell
Scope and Application: For treated and natural waters
Test preparation
Table 344 Instrument-specific information
Instrument Pour-Thru Kit Pour-Thru Cell Orientation Adapter
DR 6000 LQV175.99.20002 Arrow faces right
DR 5000 LZV479
DR 3900LQV157.99.10002 Align cell flow arrows with arrows on cell
compartment
DR 3800, DR 2800, DR 27005940400 1-inch (round) path aligned with arrow on
the adapter
LZV585 (B)
Before starting the test:
See the User Manual for Pour-Thru Module installation instructions.
Clean the Pour-Thru cell and all labware as specified inAnalysis labware treatment.
Protect the Pour-Thru Cell from contamination when not in use by inverting a small beaker over the top of the glass funnel.
Make sure the PourThru cell is completely seated in the sample cell compartment.
Final samples will contain molybdenum. In addition, final samples will have a pH less than 2 and are considered corrosive
(D002) by the Federal RCRA. Refer to the current MSDS for safe handling and disposal information.
Description Quantity
Molybdovanadate Reagent 2 mL
Water, deionized 25 mL
Cylinder, graduated, 25-mL, poly 1
Dispenser, fixed volume, 1.0-mL, w/bottle 1
Flask, Erlenmeyer, 125-mL, PMP w/cap 2
Instrument-specific information 1
http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/ -
8/12/2019 DOC316.53.01114_Ed7
2/8
Phosphorus, Reactive
Page 2 of 8
Phosphorus, Reactive
See Consumables and replacement itemsfor reorder information.
Molybdovanadate Rapid Liquid method
1. Select the test.
Insert an adapter if
required (see Instrument-
specific information).
2. I 3. Rinse a clean plastic
125-mL Erlenmeyer flask
and a 25-mL graduatedcylinder with deionized
water.
4. Blank Preparation:
Measure 25 mL of
deionized water in thegraduated cylinder. Pour
the water into the flask.
5. Rinse another clean
plastic 125-mL Erlenmeyer
flask with deionized water.
6. Prepared Sample:
Measure 25 mL of sample
in the graduated cylinder.
Pour the water into the
flask.
7. Add 1.0 mL of
Molybdovanadate
Reagent to each flask
using a Repipet Jr.
Dispenser. Swirl to mix.
A yellow color will develop
in the sample if phosphate
is present. A small amount
of yellow may be present
in the blank due to the
reagent.
8. Start the instrument
timer.
A 7-minute reaction period
will begin.
If the sample
concentration is greater
than 30 mg/L PO43, read
at exactly seven minutes
or make a 1:1 dilution of
the sample and begin the
test again.
Description Quantity
489 P React. Mo. HR RL
Stored Programs
Start
-
8/12/2019 DOC316.53.01114_Ed7
3/8
Phosphorus, Reactive
Phosphorus, Reactive
Page 3 of 8
Interferences
See the Interfering substancestable for a list of substances, interference levels and type of
interference. See the Noninterfering substances at low concentrations (less than 1000 mg/L)table
for a list of substances that do not interfere in concentrations less than 1000 mg/L.
9. When the timer
expires, pour the blank
from the flask into the
Pour-Thru Cell.
10. ZEROthe instrument.
The display will show:
0.0 mg/L PO43
11. Pour the prepared
sample from the flask into
the Pour-Thru Cell.
12. READthe results in
mg/L PO43.
Flush the Pour-Thru Cell
with 50 mL of deionized
water.
Table 345 Interfering substances
Interfering substance Interference level
Arsenate Negative interference. Positive interference if sample is heated.
Bismuth Negative interference.
Fluoride Negative interference.
Iron, FerrousBlue color is caused by ferrous iron but this does not affect results if the ferrous ironconcentration is less than 100 mg/L.
Molybdate Negative interference.
Silica Positive interference if sample is heated.
Sulfide
Negative interference. Sulfide interference may be removed by oxidation with Bromine Water
as follows:
1. Measure 25 mL of sample into a flask.
2. Add Bromine Water1drop-wise with constant swirling until permanent yellow color
develops.
3. Add Phenol Solution1drop-wise until the yellow color just disappears. Proceed with
step 7.
1 See Optional reagents and apparatus.
Thiocyanate Negative interference.Thiosulfate Negative interference.
Thorium Negative interference.
Highly buffered samples or
extreme sample pHMay exceed the buffering capacity of the reagents and require sample pretreatment.
Molybdovanadate Rapid Liquid method (continued)
Zero Read
-
8/12/2019 DOC316.53.01114_Ed7
4/8
Phosphorus, Reactive
Page 4 of 8
Phosphorus, Reactive
Sample collection, preservation and storage
Collect samples in clean plastic or glass bottles that have been cleaned with 1:1 Hydrochloric
Acid Solution*and rinsed with deionized water.
Do not use detergents that contain phosphate for cleaning labware.
Analyze samples immediately for best results.
If prompt analysis is not possible, preserve samples by filtering immediately and storing the
sample at 4 C (39 F) or below for up to 48 hours.
Warm stored samples to room temperature before analyzing.
Analysis labware treatment
1. Clean containers by normal means (do not use detergents containing phosphorus), then rinse
with deionized water.
2. Soak for several minutes in a 1:25 dilution of Molybdovanadate Reagent in deionized water.
3. Rinse well with deionized water. Dedicate these containers for HR PO43analysis.
4. Fill the Pour-Thru Cell with this same mixture of Molybdovanadate reagent and deionized
water and let stand for several minutes.
5. Rinse the Pour-Thru Cell with 50 mL of deionized water.
Table 346 Noninterfering substances at low concentrations (less than 1000 mg/L)
Pyrophosphate Tetraborate Benzoate
Citrate Lactate Formate
Oxalate Tartrate Salicylate
Al3+ Selenate Mg2+
Ca2+ Ba2+ Sr2+
Li+ Na+ K+
NH4+ Cd2+ Mn2+
NO3 NO2
SO42
SO32 Pb2+ Hg+
Hg2+ Sn2+ Cu2+
Ni2+ Ag+ U
Zr4+ AsO3 Br
CO32 ClO4
CN
IO3 Fe3+ SiO44
* See Optional reagents and apparatus.
-
8/12/2019 DOC316.53.01114_Ed7
5/8
Phosphorus, Reactive
Phosphorus, Reactive
Page 5 of 8
How to clean the Pour-Thru Cell
The Pour-Thru Cell may accumulate a buildup of colored products, especially if the reacted
solutions are allowed to stand in the cell for long periods after measurement. To clean the Pour-
Thru Cell:
1. Remove the color by rinsing with a 1:5 dilution of ammonium hydroxide*
2. Follow with several deionized water rinses.
3. Invert a beaker over the glass funnel of the Pour-Thru Cell when not in use.
Accuracy check
Standard additions method (sample spike)
Required for accuracy check:
Phosphate VoluetteAmpule Standard Solution, 500-mg/L as PO43
Ampule breaker
TenSette Pipet and tips
Mixing cylinders (3)
1. After reading test results, leave the sample cell (unspiked sample) in the instrument.
2. Select standard additions from the instrument menu: OPTIONS>MORE>STANDARD
ADDITIONS.
3. Accept the default values for standard concentration, sample volume and spike volumes. After
the values are accepted, the unspiked sample reading will appear in the top row. See the user
manual for more information.
4. Open the standard solution ampule.
5. Use the TenSette Pipet to prepare spiked samples: add 0.1 mL, 0.2 mL and 0.3 mL of
standard to three 25-mL mixing cylinders and fill to 25 mL with the fresh sample. Mixthoroughly.
6. Follow the Molybdovanadate Rapid Liquid methodtest procedure for each of the spiked
samples, starting with the 0.1 mL sample spike. Measure each of the spiked samples in the
instrument.
7. Select GRAPHto view the results. Select IDEAL LINE (or best-fit) to compare the standard
addition results to the theoretical 100% recovery.
* Optional reagents and apparatus.
-
8/12/2019 DOC316.53.01114_Ed7
6/8
Phosphorus, Reactive
Page 6 of 8
Phosphorus, Reactive
Standard solution methodNote: Refer to the instrument user manual for specific software navigation instructions.
Required for accuracy check:
10.0-mg/L Phosphate Standard
1. Use the 10.0-mg/L Phosphate Standard solution in place of the sample. Follow the
Molybdovanadate Rapid Liquid methodtest procedure.
2. To adjust the calibration curve using the reading obtained with the standard solution, navigate
to Standard Adjust in the software: OPTIONS>MORE>STANDARD ADJUST.
3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate
concentration is used, enter the concentration and adjust the curve to that value.
Method performance
Summary of method
In the molybdovanadate method, orthophosphate reacts with molybdate in an acid medium to
produce a phosphomolybdate complex. In the presence of vanadium, yellow
vanadomolybdophosphoric acid is formed. The intensity of the yellow color is proportional to the
phosphate concentration. Test results are measured at 430 nm.
Program Standard
Precision
95% Confidence Limits of
Distribution
Sensitivity
Concentration change
per 0.010 Abs change
489 10.0 mg/L PO43 9.910.1 mg/L PO4
3 0.34 mg/L PO43
-
8/12/2019 DOC316.53.01114_Ed7
7/8
Phosphorus, Reactive
Phosphorus, Reactive
Page 7 of 8
Consumables and replacement items
Required reagents
Description Quantity/Test Unit Catalog number
Molybdovanadate Reagent 2 mL 500 mL 2076049
Water, deionized 25 mL 4 L 27256
Required apparatus
Description Quantity Unit Catalog number
Cylinder, graduated, 25-mL, poly 1 each 108140
Dispenser, adjustable 1 each 2563137
Flask, Erlenmeyer, 125-mL, PMP w/cap 2 each 2089843
Recommended standards
Description Unit Catalog number
Phosphate Standard Solution, 10-mg/L as PO43 946 mL 1420416
Phosphate Standard Solution, Voluetteampule, 10-mL, 500-mg/L as PO43 16/pkg 1424210
Wastewater Influent Standard for NH3N, NO3N, PO4, COD, SO4, TOC 500 mL 2833149
Optional reagents and apparatus
Description Unit Catalog number
Ammonium Hydroxide, ACS 500 mL 10649
Bromine Water, 30 g/L 29 mL 221120
Cylinder, graduated mixing, 25 mL each 189640
Hydrochloric Acid Solution, 6.0 N, 1:1 500 mL 88449
Phenol Solution, 30 g/L 29 mL 211220
Pipet, TenSette, Pipet, 0.1 - 1.0 mL each 1970001
Pipet Tips, for TenSette Pipet 19700011
1 Other sizes are available
50/pkg 2185696
Pipet Tips, for TenSette Pipet 19700011 1000/pkg 2185628
Filter Paper, folded 12.5 cm 100/pkg 69257
Filter Funnel, Analytical, 65 mL each 108367
Thermometer, Non-Mercury, -10 to 225 C each 2635700
Sampling Bottle with cap, low density polyethylene, 250 mL 12/pkg 2087076
http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2076049http://www.hach.com/hc/search.product.details.invoker/PackagingCode=27256http://www.hach.com/hc/search.product.details.invoker/PackagingCode=27256http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2076049 -
8/12/2019 DOC316.53.01114_Ed7
8/8
HACH COMPANYWORLD HEADQUARTERSTelephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:Call 800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com [email protected]
In the U.S.A.
Outside the U.S.A.
On the Worldwide Web ; E-mail
toll-free
Phosphorus, Reactive
Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7
Optional standards
Description Unit Catalog number
Voluette Ampule breaker, 10 mL each 2196800
Phosphate, Standard Solution, 3 mg/L 946 mL 2059716
Phosphate, Standard Solution, 15 mg/L 100 mL 1424342
Phosphate, Standard Solution, 30 mg/L 946 mL 1436716
Phosphate, Standard Solution, 50 mg/L, 10 mL Voluette Ampules 16/pkg 17110
Phosphate, Standard Solution, 100 mg/L 100 mL 1436832
Phosphate, Standard Solution, 500 mg/L 100 mL 1424232