Milestone Srl 1
DMA-80 Direct Mercury Analyzer
Milestone Srl
![Page 2: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/2.jpg)
Milestone Srl 2
Introduction• Mercury in the atmosphere can
be from manmade sources (coal-fired power plants, municipal incinerators, industrial boilers) or from natural sources (forest fires, geologic formations, volcanoes)
• Precipitation is the primary mechanism for transporting airborne gaseous or particulate mercury from the atmosphere to surface water and land
![Page 3: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/3.jpg)
Milestone Srl 3
![Page 4: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/4.jpg)
Milestone Srl 4
Mercury Analysis
![Page 5: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/5.jpg)
Milestone Srl 5
Aqueous samplesand extracts
ASV7472
Solid or semi-solid wasteCV-AAS7471A
Liquid wasteCV-AAS7470A
Sample typeTechniqueMethod
![Page 6: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/6.jpg)
Milestone Srl 6
CV-AAS
• Cold Vapor-Atomic Absorption Spectroscopy– Samples are digested or dissolved– Mercury is reduced to free atomic state by using tin
(II) chloride or sodium boro-hydride in a closed reaction vessel
– Mercury is measured by atomic absorption
![Page 7: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/7.jpg)
Milestone Srl 7
CV-AAS
Few minutesAASAnalysis
Mercury reduction1 hour
Interference removalWet chemistry
20 minutes- 1 hourClosed vessel microwave
2 hours- 2 daysHot plateSample acid digestion
TimeDescriptionStep
![Page 8: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/8.jpg)
Milestone Srl 8
CV-AAS and ASV
• Both methods require sample digestion and further wet chemistry
• Sample digestion can be problematic because of Hg volatility
• Sample preparation is the bottleneck for higher analytical productivity
![Page 9: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/9.jpg)
Milestone Srl 9
DMA-80 Direct Mercury Analyzer
![Page 10: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/10.jpg)
Milestone Srl 10
DMA-80 Advantages
• Direct Hg determination at trace level on solid and liquid samples
• No sample digestion step• No wet chemistry pre-treatment step• Fast, approximately 6 minutes per sample• Eliminates waste disposal• Validated results (US EPA method 7473) for
solid and liquid matrices
![Page 11: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/11.jpg)
Milestone Srl 11
DMA-80Principle of operation
• Solid or liquid samples are weighed and introduced in the DMA-80
• The sample is initially dried and then thermally decomposed in a oxygen flow
• Combustion products are carried off and further decomposed in a hot catalyst bed
• Mercury vapors are trapped on a gold amalgamator and subsequently desorbed for quantitation
• The mercury content is determined using atomic absorption spectrophotometry at 254 nm
![Page 12: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/12.jpg)
Milestone Srl 12
![Page 13: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/13.jpg)
Milestone Srl 13
DMA-80
• Major components– Sample dosing system– Thermal process furnaces– Atomic absorption spectrophotometer– System controller
![Page 14: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/14.jpg)
Milestone Srl 14
Sample Dosing System
• Built-in 40 position auto-sampler for high throughput unattended operation, for solid and liquid samples
• Maximum sample weight 500 mg• Maximum sample volume 500 µl• Possibility of multiple sample dosing for Hg pre-
concentration on amalgamator
![Page 15: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/15.jpg)
Milestone Srl 15
Sample Introduction
![Page 16: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/16.jpg)
Milestone Srl 16
Fully Loaded Autosampler
![Page 17: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/17.jpg)
Milestone Srl 17
Pipetting Liquid Samples
![Page 18: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/18.jpg)
Milestone Srl 18
Thermal process
![Page 19: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/19.jpg)
Milestone Srl 19
Thermal Process
12 s0-30 s900°C850-950°CAmalgam
600°C550-650°CCatalysis
180 s0-420 s
850°C800-1.000°CDecomposition
60 s0-300 s300°C20-300°CDrying
Typical timeTime rangeTypical temperature
Temperature rangeStep
![Page 20: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/20.jpg)
Milestone Srl 20
Oxygen
“Research” grade(O2 > 99,95%)
Purity
200 ml/minuteFlow rate
4 bar (60 psig)Inlet pressure
Carrier and decomposition gasFunction
![Page 21: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/21.jpg)
Milestone Srl 21
Atomic Absorption Spectrophotometer
![Page 22: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/22.jpg)
Milestone Srl 22
AAS Technical Data
Si-photodiode sensorDetector
254 nm, 9 nm bandwidthInterference filter
253,65 nmWavelength
Low pressure mercury lampLight source
Single beam spectrophotometer with sequential flow through of
measurement cellsInstrument optics
![Page 23: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/23.jpg)
Milestone Srl 23
Typical Absorbance Profile
![Page 24: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/24.jpg)
Milestone Srl 24
DMA-80Performance Specifications
(1)Sample weight 100 mg(2)Automatic switching when absorbance 0,8
0,20,02Detection limit
(1st) 0,2-350(2nd) 350-6.000
(1st) 0,02-35(2nd) 35-600
Working range(2)
Hg concentration(1)
(µg/kg)Total Hg (ng)Data
![Page 25: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/25.jpg)
Milestone Srl 25
DMA-80Performance Specifications
6 minutesTypical analysis time
< 1,5%Typical reproducibility
![Page 26: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/26.jpg)
Milestone Srl 26
System controller
• Lab TERMINAL– Touch-screen Pentium-based controller, 266 MHz, 64
MB Ram, Windows 98, Color 12,1” TFT SVGA monitor, CD-ROM, FD, Smart Card Reader, 2 COM, 1 USB, LAN ports
• Standard bench-top computer with similar specifications
• Dedicated Terminal 1024
![Page 27: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/27.jpg)
Milestone Srl 27
Terminal 1024
• 12” large high resolution color screen controller
• “Touch control” display• Supplied with keyboard and
mouse• “EasyControl” software for a
complete control of all reaction parameters
• Optional “EasyDoc” software for data export on external computer
![Page 28: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/28.jpg)
Milestone Srl 28
EasyControl Software
• Evolution of previous Windows™ based DMA-80 software
• Selectable calibration algorithm• Virtually unlimited memory for programs and
actual data storage• Built-in service and diagnosis functions• CFR-21 part 11 compliant
![Page 29: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/29.jpg)
Milestone Srl 29
System Status
![Page 30: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/30.jpg)
Milestone Srl 30
1st Cell Linear Calibration
![Page 31: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/31.jpg)
Milestone Srl 31
1st Cell Square Calibration
![Page 32: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/32.jpg)
Milestone Srl 32
Measurement/Graphic
![Page 33: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/33.jpg)
Milestone Srl 33
Balance Setup
![Page 34: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/34.jpg)
Milestone Srl 34
History Trail CFR 21 Part 11
![Page 35: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/35.jpg)
Milestone Srl 35
Mercury Trap
• Optional mercury trap to be located at DMA-80 outlet
• Activated charcoal• Safe operation• Easy and ready to install
![Page 36: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/36.jpg)
Milestone Srl 36
DMA-80Analytical Performance
![Page 37: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/37.jpg)
Milestone Srl 37
Getting Started
• Oxygen cylinder with gas regulator stable at 4 bar (60 psig)
• Power 115-230 V/50-60 Hz• Analytical balance with serial interface• Windows compatible printer• Standard reference materials• Aqueous standards• Precision pipettes
![Page 38: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/38.jpg)
Milestone Srl 38
DMA-80 Calibration
![Page 39: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/39.jpg)
Milestone Srl 39
DMA-80 Calibration
• Liquid standards– Different concentrations, same volume– Same concentration, different volumes– Use fresh standards, stabilized with HNO3 or
HCl/K2Cr2O7
• Solid standards– Certified mercury content– Vary by weight
![Page 40: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/40.jpg)
Milestone Srl 40
0 ng
![Page 41: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/41.jpg)
Milestone Srl 41
10 ng
![Page 42: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/42.jpg)
Milestone Srl 42
20 ng
![Page 43: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/43.jpg)
Milestone Srl 43
30 ng
![Page 44: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/44.jpg)
Milestone Srl 44
100 ng
![Page 45: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/45.jpg)
Milestone Srl 45
200 ng
![Page 46: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/46.jpg)
Milestone Srl 46
500 ng
![Page 47: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/47.jpg)
Milestone Srl 47
Calibration 1st Cell
0,6%0,722330 ng
1,2%0,499620 ng
2,2%0,254910 ng
1,5%0,12725 ng
16,7%0,0012Blank
RSD (n=3)AbsorbanceSolution
![Page 48: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/48.jpg)
Milestone Srl 48
Calibration 2nd Cell
0,9%0,7285500 ng
1,9%0,3117200 ng
2,2%0,1546100 ng
RSD (n=3)AbsorbanceSolution
![Page 49: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/49.jpg)
Milestone Srl 49
Calibration Curves
![Page 50: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/50.jpg)
Milestone Srl 50
DMA-80 Accuracy
![Page 51: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/51.jpg)
Milestone Srl 51
Certified Reference Materials
6.250 ± 190NIST 2711 Soil
1.400 ± 80NIST 2709 Soil
141 ± 19NIST 1633b Fly Ash
93,8 ± 3,7NIST 1630a Coal
34 ± 4NIST 1573a Tomato Leaves
5,8 ± 0,5NIST 1568a Rice Flour
Hg content (µg/kg)Sample
![Page 52: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/52.jpg)
Milestone Srl 52
NIST 1568a Rice Flour
2,03,04,05,06,07,08,0
Hg (µ
g/kg
)
Certified DMA-80
![Page 53: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/53.jpg)
Milestone Srl 53
NIST 1573a Tomato Leaves
24,0
28,0
32,0
36,0
40,0
Hg
( µg/
kg)
Certified DMA-80
![Page 54: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/54.jpg)
Milestone Srl 54
NIST 1630a Coal
80,0
85,0
90,0
95,0
100,0
Hg
( µg/
kg)
Certified DMA-80
![Page 55: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/55.jpg)
Milestone Srl 55
NIST 1633b Fly Ash
100,0
120,0
140,0
160,0
180,0
Hg
( µg/
kg)
Certified DMA-80
![Page 56: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/56.jpg)
Milestone Srl 56
NIST 2709 Soil
1,0
1,2
1,4
1,6
1,8
Hg
(mg/
kg)
Certified DMA-80
![Page 57: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/57.jpg)
Milestone Srl 57
NIST 2711 Soil
5,8
6,0
6,2
6,4
6,6
Hg
(mg/
kg)
Certified DMA-80
![Page 58: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/58.jpg)
Milestone Srl 58
DMA-80 Precision
![Page 59: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/59.jpg)
Milestone Srl 59
Fresh Salmon
![Page 60: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/60.jpg)
Milestone Srl 60
Collecting Sample
![Page 61: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/61.jpg)
Milestone Srl 61
Weighing
![Page 62: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/62.jpg)
Milestone Srl 62
Loading Autosampler
![Page 63: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/63.jpg)
Milestone Srl 63
Loading Furnace
![Page 64: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/64.jpg)
Milestone Srl 64
After Analysis
![Page 65: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/65.jpg)
Milestone Srl 65
Fresh Salmon
56,20,29925
58,00,34044
57,80,23303
54,10,22962
55,40,28491
Hg (µg/kg)Weight (g)Run
![Page 66: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/66.jpg)
Milestone Srl 66
Fresh Salmon
5N
1,6Sd (µg/kg)
56,3Average (µg/kg)
![Page 67: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/67.jpg)
Milestone Srl 67
DMA-80 Detection Limit
![Page 68: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/68.jpg)
Milestone Srl 68
Detection Limit
The detection limit is defined as the concentration which will give an absorbance signal of two (sometimes three) times the magnitude of the baseline noise. The baseline noise may be statistically quantified typically by making ten or more replicate measurements of the baseline absorbance signal observed for an analytical blank, and determining the standard deviation of the measurements. The detection limit is then defined as the concentration which will produce an absorbance signal twice (or three times) the standard deviation
![Page 69: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/69.jpg)
Milestone Srl 69
Detection Limit
0,0006100,00075
0,000890,00084
0,001080,00063
0,000870,00102
0,001060,00071
AbsorbanceRunAbsorbanceRun
![Page 70: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/70.jpg)
Milestone Srl 70
Detection Limit
0,02 ngDL
0,0004683*Sd
0,000156Sd
0,0008Average
![Page 71: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/71.jpg)
Milestone Srl 71
DMA-80 Memory Effect
![Page 72: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/72.jpg)
Milestone Srl 72
DMA-80 Memory Effect
0,40,0097 (1st)Blank 2
1,50,0374 (1st)Blank 1
500 (set)0,7285 (2nd)Standard 500 ng
Mercury content (ng)AbsorbanceSample
![Page 73: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/73.jpg)
Milestone Srl 73
DMA-80 Memory Effect
![Page 74: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/74.jpg)
Milestone Srl 74
Quartz Boats
• Analysis of liquids with DMA-80 autosampler
• Improved performance – Lower memory effect– Better reproducibility
• Longer life time
![Page 75: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/75.jpg)
Milestone Srl 75
Quartz vs. Metal Boats
0
0,1
0,2
0,3
0,4
1 2 3 4 5 6 7 8 9 10Run
Sig
nal
Metal Quartz
![Page 76: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/76.jpg)
Milestone Srl 76
Memory Effect
0,010,000,0000,1001
0,450,050,00110,1001
1.106,01110,600,14740,1001
Result (µg/kg)Hg (ng)HeightWeight (g)Sample ID
![Page 77: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/77.jpg)
Milestone Srl 77
Reproducibility at Low Concentrations
AVG 8,55SD 0,09 µg/kgRSD 1,01%Statistics
8,500,850,02170,100
8,490,850,02160,100
8,510,850,02170,100
8,530,850,02180,100
8,720,870,02220,100
Results (µg/kg)Hg (ng)HeightWeight (g)
![Page 78: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/78.jpg)
Milestone Srl 78
Reproducibility at High Concentrations
AVG 1.143,7RSD 0,69%SD 7,88 µg/kgStatistics
1.143,4114,340,15230,100
1.158,7115,870,15430,100
1.141,3114,130,15200,100
1.136,2113,620,15140,100
1.138,8113,880,15170,100
Results (µg/kg)Hg (ng)HeightWeight (g)
![Page 79: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/79.jpg)
Milestone Srl 79
Lifetime
![Page 80: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/80.jpg)
Milestone Srl 80
• US EPA method 7473– Mercury in solids and solutions by thermal
decomposition amalgamation and atomic absorption spectrophotometry
![Page 81: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/81.jpg)
Milestone Srl 81
ComparisonDMA-80 vs. CV-AAS
![Page 82: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/82.jpg)
Milestone Srl 82
ComparisonDMA-80 vs. CV-AAS
YesMinimalWaste generation
Yes, to eliminate interferencesNot requiredWet chemistry
ElevatedModerateRunning cost
Limited to low rangeUp to 600 ngWorking range
100 ml solutionUp to 500 mg sampleSample size
AqueousLiquid and solidsSample type
Acid digestionNot requiredSample preparation
CV-AASDMA-80Technique
![Page 83: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/83.jpg)
Milestone Srl 83
Time ComparisonDMA-80 vs. CV-AAS
0
60
120
180
240
DMA-80 CV-AAS
Tim
e (m
inut
es)
Sample preparation Analysis
![Page 84: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/84.jpg)
Milestone Srl 84
DMA-80 Major Applications
• Biological– Blood, urine, hair, tissue, plankton
• Liquids– Waste water, crude oil, heavy oil, detergents, paints
• Solids– Coal, fly ash, soil, sediment, sludge, minerals, food, feed, plastic,
wood, vegetables, leaves, waxes• Medicinal
– Pharmaceuticals, gelatin capsules, lipstick, lotions
![Page 85: DMA-80 Direct Mercury Analyzer - IQ USPallchemy.iq.usp.br/agregando/wpa/Palestra19.pdf · DMA-80 Direct Mercury Analyzer Milestone Srl. ... 4 0,0008 9 0,0008 3 0,0006 8 0,0010 ...](https://reader033.fdocuments.in/reader033/viewer/2022042605/5b448a0c7f8b9a56348b68ea/html5/thumbnails/85.jpg)
Milestone Srl 85
DMA-80 Direct Mercury Analyzer