Instructions for Use
0x6.001.067_AB 15.11.12 07:00 Seite Cvr1
0x6.001.067_AB 15.11.12 07:00 Seite Cvr2
Dynamic Locking Screw Instructions for Use Synthes 1
Table of Contents
Introduction
Instructions for Use
Product Information
DLS Dynamic Locking Screw 2
Indications 5
Dynamic Locking Screw Technique 6
Compatibility with the LCP Percutaneous 13Instrument System 3.5
Implant Removal 14
Technical Information 15
Implants 16
Instruments for DLS 3.7 mm 18
Instruments for DLS 5.0 mm 20
Image intensifier control
WarningThis description alone does not provide sufficient background for direct use ofthe product. Instruction by a surgeon experienced in handling this product ishighly recommended.
Reprocessing, Care and Maintenance of Synthes InstrumentsFor general guidelines, function control and dismantling of multi-part instruments,please refer to: www.synthes.com/reprocessing
0x6.001.067_AB 15.11.12 07:00 Seite 1
Angular stable screwLocking screws can be placed in eachthreaded hole of the appropriate LCP.
Screws lock tightly in the plate to:– eliminate tension on the bone– eliminate compression between
plate and bone– protect periosteum from potential
damage and retain circulation
Pin-sleeve design of the DLS allows micro-motion within the angular stablesystem.
DLS can be used with all Synthes’ tita-nium or stainless steel locking plates.
Cobalt chromiummolybdenum alloy (CoCrMo)enables DLS to be as strong asstandard locking screws.
CoCrMo encourages less in-growth and facilitates screw removal – similar to stainlesssteel screws.
Laserwelded. For stable fixa-tion between pin and sleeve.
2 Synthes Dynamic Locking Screw Instructions for Use
DLS Dynamic Locking Screw.Combined with LCP LockingCompression Plate.
DLS Dynamic Locking Screw
Stardrive recess. For bettertorque transmission.
0x6.001.067_AB 15.11.12 07:00 Seite 2
New screw tip. Roundedtip and five flute design.
Standardlocking head.Compatible withall Synthes’ locking plates.
Pin-sleeve design.Micro-motion withinthe screw.
Standard thread inthe shaft. For stronganchorage in bone.
DLS allows surgeons to modulate therigidity of the locked plate construct forbetter load distribution.
DLS construct can deliver nearly parallelfracture site motion.
Boosting biological bone healing: Micro-motions on the plate-side of thelocking construct improve circumferen-tial callus formation.
Dynamic Locking Screw Instructions for Use Synthes 3
Principle of Dynamic Locking Screw (DLS)
0x6.001.067_AB 15.11.12 07:00 Seite 3
A
B
4 Synthes Dynamic Locking Screw Instructions for Use
LCP Locking Compression Plate
– Provides angular stable support offragments regardless of bone quality
– Reduces risk of primary and second-ary loss of reduction even underhigh dynamic loading
– Reduces impairment of periostealblood supply due to limited plate-periosteum contact
– Maintains favorable hold in osteo-porotic bone and multiple fragmentfractures
A. Stable plate-screw connection– Locking screws reduce screw loosen-
ing– Excessive torque is not applied to
the cortical bone– The conical screw head facilitates
screw insertion
DLS – Self-tapping dynamic locking screw– Requires precise measurement of
length (diaphysis and metaphysis)– Bicortical use in the shaft– Not necessary to separately tap
thread
Self-tapping locking screw– Requires precise measurement of
length (metaphysis)– Monocortical or bicortical use– Not necessary to separately tap
thread
Standard screw– Dynamic compression is created
by the eccentric insertion of thestandard screw (analogous to LC-DCP)
LCP combi-hole
B. Compatibility– The proven dynamic compression
hole allows for use of all standardscrews
0x6.001.067_AB 15.11.12 07:00 Seite 4
Dynamic Locking Screw Instructions for Use Synthes 5
Indications
Synthes Dynamic Locking Screws (DLS) in combination withSynthes Locking Compression Plates (LCP) are intended forthe fixation of metaphyseal and diaphyseal shaft fractures of long bones (humerus, femur and tibia). The DLS allowsmicro-motion within the angular stable system.
Synthes offers a wide selection of LCP plates, which covers a correspondingly extensive variety of indications. For this reason, this technical guide does not cover specific indicationsor the selection of a plate type for specific clinical situations.For treatment of these subjects, please refer to ”AO Principlesof Fracture Management,” courses offered by the AO(www.aofoundation.org), and the corresponding profes-sional literature.
For general purposes, the following techniques for handlingthe implants and instruments will be demonstrated with astraight 3.5 mm LCP and DLS 3.7 mm.
Important: The DLS 3.7 mm has a 0.2 mm larger shaftthread diameter compared to the standard locking screw 3.5 mm. Therefore dedicated instruments for DLS 3.7 mm(drill bit and drill sleeve) have to be used. The other instru-ments stay the same.
Standard locking screw 5.0 instrumentation can be used forthe DLS 5.0 (see pages 19 and 20).
0x6.001.067_AB 15.11.12 07:00 Seite 5
6 Synthes Dynamic Locking Screw Instructions for Use
Dynamic Locking Screw Technique
1Reduce and fix fracture
Reduce the fracture under image intensification. As needed,provide fixation with a Kirschner wire or reduction forceps.
2Bend plate
Instruments
329.040 Bending Iron for Plates 2.4 to 3.5 length 145 mm (for use with 329.050)
329.050 Bending Iron for Plates 2.4 to 3.5 length 145 mm (for use with 329.040)
329.150 Bending Pliers for Plates 2.4 to 4.0, length 230 mm
329.290 Bending Pliers for Reconstruction Plates 2.7 and 3.5
Adapt the LCP plate to the anatomy using the appropriatebending instruments (as for standard plates), especially in thecase of joint fractures.
Notes– Do not bend the plate back and forth.– The LCP combi-holes are asymmetrical in the plate. In
straight plates, the hole alignment changes in the middleof the plate. This asymmetry enables exertion of uni-directional dynamic compression.
0x6.001.067_AB 15.11.12 07:00 Seite 6
Dynamic Locking Screw Instructions for Use Synthes 7
3Position and fix plate
Position and fix the plate on the bone.
Important: Before setting the first dynamic locking screw,ensure that the plate is provisionally well fixed in place. Otherwise the plate may rotate during screw locking, causingdamage to surrounding soft tissue.
0x6.001.067_AB 15.11.12 07:00 Seite 7
8 Synthes Dynamic Locking Screw Instructions for Use
Dynamic Locking Screw Technique
4Prepare DLS drill sleeve
Instrument
03.213.001 DLS Drill Sleeve 3.7, for Drill Bits � 3.1 mm
Carefully screw the DLS drill sleeve into the desired LCP holeuntil it is gripped completely by the thread. The DLS drillsleeve ensures that the dynamic locking screw is correctlylocked in the plate. Angular stability is reduced if a lockingscrew is inserted obliquely.
Technique tip: To make it easier for the drill sleeve to locatethe start of the thread, it may be useful to slightly rotate thedrill sleeve counterclockwise.
Note: In precontoured plates, the threaded holes might notbe perpendicular to the plate surface due to the anatomy.
Important: Only use the white color-coded DLS drill sleevewith the white color coded drill bit for DLS 3.7 mm. For the DLS 5.0 mm the standard drill sleeve and drill bit as forstandard locking screws can be used.
0x6.001.067_AB 15.11.12 07:00 Seite 8
Dynamic Locking Screw Instructions for Use Synthes 9
5Predrill screw hole and determine screw length
Instrument
03.213.002 DLS Drill Bit � 3.1 mm with Stop, length 165 mm, 2-flute, for Quick Coupling
Carefully drill the screw hole using an appropriate drill bit.
Remove the drill sleeve.
Important: Only use the white color-coded DLS drill bit forDLS 3.7 mm to ensure the correctly sized drill hole.
6Option: Determine screw length
Instrument
319.010 Depth Gauge
Determine the screw length with the depth gauge.
Example showing measured length of 24 mm.
Example showing measured length of 24 mm.
0x6.001.067_AB 15.11.12 07:00 Seite 9
10 Synthes Dynamic Locking Screw Instructions for Use
Dynamic Locking Screw Technique
7Insert Dynamic Locking Screw
Instruments
511.770 Torque Limiter, 1.5 Nm
511.773 Torque Limiter, 1.5 Nm, for AO/ASIF Quick Coupling
314.116 Screwdriver Shaft T15, self-holding
397.705 Handle for Torque Limiter Nos. 511.770 and 511.771
311.431 Handle with Quick Coupling
Before insertion of the first dynamic locking screw, anatomi-cal reconstruction must have occurred. After insertion of thedynamic locking screws, additional reduction cannot beachieved without removing the dynamic locking screws. The dynamic locking screws can either be inserted with a powertool (without final locking) or manually and always with atorque limiter.
Important– Always insert the dynamic locking screw with the dedi-
cated torque limiter.– 3 or more dynamic locking screws must be inserted for
each segment / fragment.– Do not combine standard locking screws and/or standard
screws with DLS in the same segment/fragment. This willnullify the dynamization and might overload the standardlocking screw; however, one fragment can be fixed with a standard locking screw and another with DLS.
Note: Do not press the plate onto the bone. This will nullifythe dynamization.
0x6.001.067_AB 15.11.12 07:00 Seite 10
Dynamic Locking Screw Instructions for Use Synthes 11
7aInsertion with power tool
To insert the dynamic locking screw using a power tool, fit atorque limiter to the power tool. Then insert the screwdrivershaft into the torque limiter.
Pick up the dynamic locking screw and insert it into the platehole. To insert the screw, start the power tool slowly, increas-ing the speed and then reducing it again before the dynamiclocking screw is fully tightened. Uncouple the power tooland mount the handle for torque limiter or the handle withquick coupling and manually tighten the screw. After oneclick, the optimum torque is reached.
Notes– To reduce the risk of stripping the screw head, do not lock
dynamic locking screws at full speed. Screw head strip-ping can cause difficulty in removing the implant.
– For long screws and thick cortical bone, ensure sufficientcooling during insertion.
Important: Always insert the dynamic locking screw withthe dedicated torque limiter: DLS 3.7 mm → TLA 1.5 Nm DLS 5.0 mm → TLA 4.0 Nm
0x6.001.067_AB 15.11.12 07:00 Seite 11
12 Synthes Dynamic Locking Screw Instructions for Use
7bManual insertion
To insert the dynamic locking screw manually, attach thetorque limiter handle to the torque limiter and insert a screw-driver shaft. Screw in the dynamic locking screw, and lock itin the plate.
Important– Always insert and lock the dynamic locking screw with the
dedicated torque limiter: DLS 3.7mm → TLA 1.5 Nm DLS 5.0mm → TLA 4.0 Nm
– The dynamic locking screw is only available with Stardriverecess and should only be used with the appropriateStardrive screwdriver. Do not use the hexagonal screw-driver.
Note: Under image intensification you can identify the pin-sleeve design of the DLS.
Dynamic Locking Screw Technique
0x6.001.067_AB 15.11.12 07:00 Seite 12
Dynamic Locking Screw Instructions for Use Synthes 13
Compatibility with the LCP Percutaneous Instrument System3.5 mm
The LCP Percutaneous Instrument Set 3.5 mm consists of acomprehensive series of aiming arms and instrumentation tofacilitate the percutaneous, submuscular insertion of variousplates.
The DLS 3.7 mm is compatible with the LCP Percutaneous Instrument System 3.5 mm.
For the use of the DLS 3.7 with LCP Percutaneous InstrumentSystem 3.5 the LCP Small Fragment Percutaneous InstrumentSet 3.5 needs to be upgraded with the following instruments:
03.113.030 Drill Bit � 3.1 mm with Stop, calibrated, length 250/225 mm, for Quick Coupling
03.113.031 Locking Drill Sleeve � 3.1 mm, percutaneous
For the complete instruction and technique details of the useof the LCP Percutaneous Instrument System 3.5 please referto technique guide “LCP Low Bend Medial Distal Tibia PlateAiming Instruments 3.5 mm” 036.001.045 and “LCP Per cutaneous Aiming System 3.5 for PHILOS” 036.001.024.
0x6.001.067_AB 15.11.12 07:00 Seite 13
14 Synthes Dynamic Locking Screw Instructions for Use
To remove the plate, first remove tissue and bone from allscrew heads and drives. Insert a screwdriver in the screw recess and unlock all screws manually. Then completely re-move all screws.
If the screws cannot be removed with the screwdriver, con-sult the separate Synthes publication “Screw Extraction Set:Instruments for removing Synthes screws” (Art. No.036.000.918: pages 27–31 (Situation B) and pages 44 –47),which explains how broken screws can be removed.
These are the specific instruments for the extraction of the dynamic locking screws:
DLS 3.7 mm:
309.035 Hollow Reamer, complete, anticlockwise cutting, for Screws � 3.5 and 4.0 mm
309.038 Spare Reamer Tube, for No. 309.035
309.039 Extraction Bolt, for Screws � 3.5 and 4.0 mm
309.510 Extraction Screw, conical, for Screws � 1.5 and 2.0 mm
309.520 Extraction Screw, conical, for Screws � 2.7, 3.5 and 4.0 mm
DLS 5.0 mm:
309.450 Hollow Reamer, complete, anticlockwise cutting, for Screws � 4.5 mm
309.480 Spare Reamer Tube, for No. 309.450
309.490 Extraction Bolt, for Screws � 4.5 and 5.0 mm
309.520 Extraction Screw, conical, for Screws � 2.7, 3.5 and 4.0 mm
309.521 Extraction Screw for Screws � 3.5 mm
Implant Removal
0x6.001.067_AB 15.11.12 07:00 Seite 14
** * *
* *
Shor
t Pi
n
Shor
t Pi
n
Long
Pin
Long
Pin
26 –
70 m
m
22 a
nd 2
4 m
m
32 –
36 m
m
38 –
90 m
m
Dynamic Locking Screw Instructions for Use Synthes 15
DLS 3.7The two shortest screws (22 and 24 mm) have a shorter pin,and all longer screws (26 – 70 mm) have the same longerlength pin. The result is a comparable elastic behavior acrossall screw lengths.
DLS 5.0The three shortest screws (32 – 36 mm) have a shorter pin,and all longer screws (38 – 90 mm) have the same longerlength pin. The result is a comparable elastic behavior acrossall screw lengths.
Technical Information
Screw and pin lengths
* Maximum deflection 0.2 mm
* Maximum deflection 0.35 mm
0x6.001.067_AB 15.11.12 07:00 Seite 15
16 Synthes Dynamic Locking Screw Instructions for Use
Implants
DLS 3.7 – Dynamic Locking Screw � 3.7 mm, self-tapping– CoCrMo– Sterile
Art no. Length (mm)
09.213.022S 22
09.213.024S 24
09.213.026S 26
09.213.028S 28
09.213.030S 30
09.213.032S 32
09.213.034S 34
09.213.036S 36
09.213.038S 38
09.213.040S 40
09.213.042S 42
09.213.044S 44
09.213.046S 46
09.213.048S 48
09.213.050S 50
09.213.052S 52
09.213.054S 54
09.213.055S 55
09.213.056S 56
09.213.058S 58
09.213.060S 60
09.213.065S 65
09.213.070S 70
Warning: Do not resterilize the Dynamic Locking Screw.
0x6.001.067_AB 15.11.12 07:00 Seite 16
Dynamic Locking Screw Instructions for Use Synthes 17
DLS 5.0 – Dynamic Locking Screw � 5.0 mm, self-tapping– CoCrMo– Sterile
Art no. Length (mm)
09.223.032S 32
09.223.034S 34
09.223.036S 36
09.223.038S 38
09.223.040S 40
09.223.042S 42
09.223.044S 44
09.223.046S 46
09.223.048S 48
09.223.050S 50
09.223.055S 55
09.223.060S 60
09.223.065S 65
09.223.070S 70
09.223.075S 75
09.223.080S 80
09.223.085S 85
09.223.090S 90
Warning: Do not resterilize the Dynamic Locking Screw.
0x6.001.067_AB 15.11.12 07:00 Seite 17
18 Synthes Dynamic Locking Screw Instructions for Use
Instruments for DLS 3.7 mm
03.213.001 DLS Drill Sleeve 3.7, for Drill Bits � 3.1 mm
Dedicated instruments for DLS 3.7 mm
03.213.002 DLS Drill Bit � 3.1 mm with Stop, length165 mm, 2-flute, for Quick Coupling
Small fragment instruments
329.040 Bending Iron for Plates 2.4 to 3.5, length 145 mm (for use with 329.050)
329.050 Bending Iron for Plates 2.4 to 3.5, length 145 mm (for use with 329.040)
329.150 Bending Pliers for Plates 2.4 to 4.0, length 230 mm
329.290 Bending Pliers for Reconstruction Plates2.7 and 3.5
319.010 Depth Gauge for Screws � 2.7 to 4.0 mm,measuring range up to 60 mm
511.770 Torque Limiter, 1.5 Nm, for Compact Air Drive and for Power Drive
511.773 Torque Limiter, 1.5 Nm, for AO/ASIF Quick Coupling
0x6.001.067_AB 15.11.12 07:01 Seite 18
Dynamic Locking Screw Instructions for Use Synthes 19
314.116 Screwdriver Shaft Stardrive 3.5, T15, self-holding, for AO/ASIF Quick Coupling
397.705 Handle for Torque Limiter Nos. 511.770and 511.771
311.431 Handle with Quick Coupling
0x6.001.067_AB 15.11.12 07:01 Seite 19
20 Synthes Dynamic Locking Screw Instructions for Use
323.500 LCP Universal Drill Guide 4.5/5.0
314.119 Screwdriver Shaft Stardrive 4.5/5.0, T25,self-holding, for AO/ASIF Quick Coupling
314.164 Screwdriver Stardrive 4.5/5.0, T25, with Groove, length 240 mm
323.460 Universal Drill Guide 4.5/3.2, for neutraland load position
Instruments for DLS 5.0 mm
310.430 LCP Drill Bit � 4.3 mm with Stop, length 221 mm, 2-flute, for Quick Coupling
323.042 LCP Drill Sleeve 5.0, for Drill Bits � 4.3 mm
Large fragment instruments
0x6.001.067_AB 15.11.12 07:01 Seite 20
314.163 Torque-limiting Screwdriver Stardrive, T25, self-holding, for Locking Screws � 5.0 mm
314.281 Holding Sleeve for Screws, for LCP 4.5/5.0
397.705 Handle for Torque Limiter Nos. 511.770and 511.771
511.771 Torque Limiter, 4 Nm, for Compact Air Drive and Power Drive
Dynamic Locking Screw Instructions for Use Synthes 21
0x6.001.067_AB 15.11.12 07:01 Seite Cvr3
0123 036.
001.
067
vers
ion
AB
10
/201
2 30
1012
98©
Syn
thes
, Inc
. or
its a
ffili
ates
Su
bjec
t to
mod
ifica
tions
Sy
nthe
s is
a t
rade
mar
k of
Syn
thes
, Inc
. or
its a
ffili
ates
All technique guides are available as PDF files at www.synthes.com/lit
Ö036.001.067öABNä
0x6.001.067_AB 15.11.12 07:01 Seite Cvr4