Dialyzers in the 21 Century The Basics - BC Renal. Kiaii... · 2018. 1. 5. · Dr. Mercedeh Kiaii,...

36
Dr. Mercedeh Kiaii, MD, FRCPC St. Paul’s Hospital,Vancouver,BC Dialyzers in the 21 st Century The Basics

Transcript of Dialyzers in the 21 Century The Basics - BC Renal. Kiaii... · 2018. 1. 5. · Dr. Mercedeh Kiaii,...

  • Dr. Mercedeh Kiaii, MD, FRCPCSt. Paul’s Hospital,Vancouver,BC

    Dialyzers in the 21st Century The Basics

  • Overview

    Evolution of Dialyzers

    Basic Components of Dialyzers

    Types of membranes

    Efficiency and Flux

    Sterilization

    Dialyzer reactions

    Local Experience of Significant Thrombocytopenia Related to Dialyzer Use

  • EVOLUTION OF DIALYZERS

  • History of Dialyzers

    1924: Haas Dialyzer

    Tubular Device

    Cellulose trinitrate

    1944: Kolff Dialyzer

    Rotating dialyzer

    Wooden drum

    Celluphane tube

  • History of Dialyzers

    1956: Kolff Dialyzer

    Coil dialyzer

    1960: Kiil Dialyzer

    Plate dialyzer

    Cellulosic flat sheet

  • History of Dialyzers

    1966: Hollow Fiber Dialyzer

    Cellulose Acetate

    1969: First Synthetic Membrane

    Polyacrylonitrile (AN-69)

  • The Hollow Fiber Dialyzer

    Components

    Membrane

    Hollow fibers

    Potting compound

    Header

    Housing

  • BASIC DIALYZER COMPONENTS

  • The Dialyzer Membrane

    Membrane material

    Efficiency

    Pore size (Ultrafiltration coefficient = Flux)

    Sterilization

  • Dialyzer Membrane Material

    Unmodified Cellulose (Cuprophane)

    Polysaccharide based obtained from pressed cotton

    Chains of glucosan rings with abundant free hyroxyl groups

    Substituted cellulose

    Cellulose Acetate: acetate binds to the hydroxyl groups

    Diacetate, triacetate

  • Dialyzer Membrane Material

    Cellulosynthetic

    Addition of a synthetic material such as diethylaminoethyl = Hemophane

    Synthetic membranes

    Polysulfone

    Polyacrylonitrile (PAN)

    Polycarbonate

    Polyamide

    Polymethylmethacrylate (PMMA)

  • Membrane Biocompatibility

    Biocompatible membranes

    Less activation of the immune system and inflammatory response

    Substituted cellulose and the synthetic membranes essentially similar in biocompatibility

    Some patients can have sensitivity to certain membrane material

  • Polysulfone membranes

    High physical strength & chemical resistance

    Heat and radiation sterilization possible

    Often combined with PVP to increase hydrophylicity

    Ability to retain endotoxins

  • Polyethersulfone membranes

    Advanced fiber spinning process

    Creates larger, uniformly sized and densely distributed pores

    Achieve better middle molecule clearance with minimal albumin loss

  • PAN derived membranes

    AN-69 membranes:

    Polymer of PAN and methanllylsulfonate

    Negatively charged surface can activate bradykinins and cause dialyzer reactions

    May be aggravated by use of ACEI

    AN-69 ST membranes:

    Coated with polycationic polymer to decrease reactions

    Heparin (negatively charged) can bind to the membrane

  • Efficiency

    Efficiency related to:

    Membrane size ( Surface Area)

    Porosity

    Thickness

    Internal diameter of the fibers

    Reduces the resistance of the more static blood layers

    Design

    Wavelike or crimped vs straight fibers

  • Flux

    Flux: related to Kuf (Coefficient of ultrafiltration)

    volume of fluid (ml/hr) transferred across the membrane per mmHg of pressure gradient

    High flux dialyzers

    Kuf > 15 ml/mmhg or B2M clearance > 20 ml/min

    Advantages of high flux:

    Improve middle molecule clearance

    Disadvantages of high flux:

    Errors in TMP can cause large UF

    Backfiltration

  • Evidence for Flux

    Cochrane review of RCT’s could not determine overall efficacy of high flux dialysis

    Concluded may reduce CV mortality

    HEMO study

    Decreased CV mortality in pts on HD > 3.7 yrs

    MPO study

    Increased survival in pts with albumin < 40 and in diabetics

  • High Cut-Off Membranes

    Higher molecular weight cut-off ~ 65KDa

    Better B2M clearance compared to high flux dialyzers

    Increase clearance of free serum light chains:

    Potential use in treatment of patients with multiple myeloma

    Increase Albumin clearance

  • Vitamin E coated dialyzers

    Cellulose or PS backbone

    Coated with vitamin E (α- tocopherol)

    Reduce reactive oxygen species

    Decrease inflammatory markers

    No survival benefit yet

  • Sterilization Techniques

    Advances in Sterilization

    Chemical

    Heat

    Radiation

    Gamma-radiation

    Beta-radiation (electron-beam)

  • Microbiological Efficacy

    Effect on Dialyzer Material

    Release of sterilization byproduct

    Cost

    Chemical(ETO)

    High Very low mechanical stressNo thermal stress

    Residual ETO from potting compound

    Low

    Heat High High thermal stress. Not suitable for all membranes

    None High

    Radiation High Intermediate material stress

    Cytotoxic compounds from some dialyzers & from some potting compunds

    High

    Overview of Sterilization Techniques

  • Radiation Sterilization

    Gamma Radiation

    Use radioactive isotope

    Usually cobalt 60

    Very penetrating

    Beta Radiation (electron-beam)

    Stream of high energy electrons accelerated by means of a linear accelerator

    Limited depth of penetration

    More focused and precisely delivered

    Exposure time is shorter

  • Dialyzer Reactions

    Type A

    IgE mediated, hypersensitivity reactions

    Often severe and usually associated with exposure to an antigen (most common ETO)

    Type B

    Complement mediated, less severe

    Can result in changes in count or function of the blood cell lines

    Can result in neutropenia followed by rebound leukocytosis

  • Dialyzer Reactions

    Protein-Membrane Interactions:

    Dialysis membrane and the coagulation proteins which can result in binding of fibrinogen, platelet adhesion and thrombocytopenia

    Platelet count can drop during initial part of dialysis but usually rebound to pre-dialysis levels at the end of the session

    Overall drop of greater than 7-9% is unusual with biocompatible membranes

  • LOCAL CANADIAN EXPERIENCE BRITISH COLUMBIA AND CALGARY, CANADA

  • SPH Dialysis Unit Review

    Switch to different dialyzer

  • BC-Wide Provincial Investigation (broken down by health authority)

    Proportion  of  patients by % of platelet count change from pre‐dialysis  values

    7 7 13 4 5 9

    2012

    17

    12 1515

    2828

    25

    31 2427

    37 47 38 46 52 43

    8 5 7 7 4 6

    A B C D E Total BC

    > ‐25% ‐15% to ‐24% ‐5% to ‐14% ‐4% to +15% > +15%

    Health Authority 

  • Calgary Experience

    Switch to different dialyzer

  • Prevalence of Significant Thrombocytopenia in HD Population

  • Impact of Intervention

    11.4%

    6.7%

    10.8%

    3.6%

    23.4%

    9.4%

    32.0%

    5.7% 7.2%

    2.1%

    7.3%

    1.6%

    0%

    5%

    10%

    15%

    20%

    25%

    30%

    35%

    40%

    0%

    5%

    10%

    15%

    20%

    25%

    30%

    35%

    40%

    E-Beam Non E-Beam E-Beam Non E-Beam E-Beam Non E-Beam E-Beam Non E-Beam E-Beam Non E-Beam E-Beam Non E-Beam

    BC AB BC AB BC AB

    Post-dialysis platelet count 15% Post-dialysis platelet count 15%

    % af

    fect

    ed p

    atie

    nts

    Percentage of patients with significant thrombocytopenia before and after switch to non e-beam sterilized dialyzer

  • CHOOSING A DIALYZER

  • Dialyzer Specifications

    Membrane material

    Polysulfone: Fresenius, Asahi

    Poly(aryl)ethersulfone: Gambro (Revaclear), Baxter (Xenium)

    AN-69-ST: Gambro (Nephral ST)

    Cellulose Triacetate: Baxter (Exeltra)

    Clearance (Qb: 200-400 ml/min and Qd: 500 ml/min)

    Urea, creatinine, vitamin B12, B2M

  • Dialyzer Specifications

    Kuf (Flux): High flux standard, required for HDF

    Sterilization:

    Steam and Gamma radiation preferred

    Other:

    Heparin requirement

    Coating of dialyzer

    Vitamin E, Heparin grafted (Evodial)

    Wet vs Dry

    Cost

  • Summary

    Most dialyzers currently available are high efficiency & have similar clearance performance

    High flux dialyzers have some advantages and required for delivering hemodiafiltration

    Every unit must have options for patients with possible sensitivity to certain dialyzer membrane

    Have choice of two different membrane materialf

    Current preferred modes of sterilization are steam and gamma radiation

  • QUESTIONS / DISCUSSION

    Dialyzers in the 21st Century�The BasicsOverviewEVOLUTION OF DIALYZERSHistory of DialyzersHistory of DialyzersHistory of DialyzersThe Hollow Fiber DialyzerBASIC DIALYZER COMPONENTSThe Dialyzer MembraneDialyzer Membrane MaterialDialyzer Membrane MaterialMembrane BiocompatibilityPolysulfone membranesPolyethersulfone membranesPAN derived membranesEfficiencyFluxEvidence for FluxHigh Cut-Off MembranesVitamin E coated dialyzersSterilization Techniques Overview of Sterilization TechniquesRadiation SterilizationDialyzer ReactionsDialyzer ReactionsLOCAL CANADIAN EXPERIENCE SPH Dialysis Unit ReviewBC-Wide Provincial Investigation�(broken down by health authority)Calgary ExperiencePrevalence of Significant Thrombocytopenia in HD PopulationImpact of InterventionCHOOSING A DIALYZERDialyzer SpecificationsDialyzer SpecificationsSummaryQUESTIONS / DISCUSSION