DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY
-
Upload
stuart-green -
Category
Documents
-
view
215 -
download
1
Transcript of DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY
![Page 1: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/1.jpg)
DEVELOPMENTS IN ACCELERATOR BASED BORON
NEUTRON CAPTURE THERAPY
STUART GREEN
Department of Medical Physics, University Hospital Birmingham NHS Trust, Edgbaston,Birmingham B15 2TH, UK
AbstractÐThis paper will review the current status of Boron Neutron Capture Therapy (BNCT), frombasic physical mechanisms and clinical indications, to neutron beam development and dosimetry. Forin-hospital facilities, particle accelerators presently provide the favoured option, and this paper concen-trates on this approach to neutron beam production for BNCT. Various accelerator-based approacheswill be reviewed, but discussion will concentrate on the Birmingham programme, particularly the designof a suitable neutron beam delivery system and the experimental validation of Monte Carlo simulationson a mock-up neutron beam moderation system. The use of dose modifying factors to evaluate thelikely clinical utility of an epithermal neutron beam will also be discussed, with illustrations from theBirmingham programme. # 1998 Elsevier Science Ltd. All rights reserved
INTRODUCTION
There has been a resurgence of interest in BNCT inrecent years as a consequence of two separateevents. Firstly, the publication of promising clinical
results from BNCT treatments carried out in Japan(Hatanaka and Nakagawa, 1994) has encouragedclinicians that there might be some future in this
treatment modality. Secondly, a combination ofwork in neutron beam design and in characterisingthe various available compounds which can be used
to carry boronated drugs to the tumour site, hasencouraged scientists that a signi®cant doseenhancement can be achieved.
This paper provides a review of the signi®cantdevelopments in BNCT in recent years, and assessesthe opportunities for this treatment modality todevelop into mainstream radiotherapy practice. The
review will concentrate on future options which useparticle accelerators as the primary source of neu-trons, since these give greatest potential for in-hos-
pital facilities, and will draw speci®c data from thework on accelerator based BNCT in Birmingham.
OVERVIEW
Clinical indications for BNCT
BNCT may be a suitable treatment for a numberof tumour types. If the tumour is located such that
neutrons can be suitably delivered, and it is of atype which takes up a boronated drug, then treat-ment may be possible. In addition, in common with
all other radical radiotherapy treatments, local con-trol of the primary tumour should be the principalclinical problem. There are a number of tumoursfor which these factors apply, but the majority of
interest world-wide has focused on glioblastoma
multiforme and metastatic melanoma.
Glioblastoma multiforme
Two factors were mentioned above as being sig-
ni®cant in the recent resurgence of interest inBNCT. There is also an underlying third factorwhich is the very poor prognosis for the majority ofpatients with the tumour glioblastoma multiforme.
This is a grade IV astrocytoma, and is a tumour ofthe glial tissue which is the structural material ofthe brain. High grade (III or IV) astrocytomas are
around 1% of cancer diagnosis in the UK, andaround 2.5% of cancer deaths (James, 1996).Incidence increases with age with prognosis dete-
riorating with age and performance status. Patientstend to die as a result of uncontrolled local diseaserather than from the e�ects of metastasis. However,
there are a sub-group of the younger, ®tter patientsthat do receive substantial bene®t from convention-al therapy, both surgery and external beam photonradiotherapy. Hence great care must be exercised in
patient selection for BCNT trials which will almostcertainly be sub-therapeutic in their early stages.
BNCT physical mechanisms
BNCT is a form of binary radiotherapy andtherefore involves two key stages. The ®rst is thepreferential accumulation, in tumour cells, of an
isotope with a suitable a�nity for neutrons at a cer-tain energy. This must then be followed by anintense irradiation of these cells with neutrons at an
energy such that their probability for capture ismaximised. In BNCT, this means a preferential ac-cumulation of boron, followed by submerging thetumour and the surrounding healthy brain tissue in
Radiat. Phys. Chem. Vol. 51, No. 4-6, pp. 561±569, 1998# 1998 Elsevier Science Ltd. All rights reserved
Printed in Great Britain0969-806X/98 $19.00+0.00PII: S0969-806X(97)00203-X
561
![Page 2: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/2.jpg)
a bath of thermal neutrons. These thermal neutrons
may either be incident directly onto the tissue, for
super®cial tumours or intra-operative treatment, or
may be produced locally by moderation of a higher
energy neutron source which is incident on to the
patient surface.
The interaction of thermal and epithermal neu-
trons is dominated by scattering. As a result, neu-
trons of these energies cannot be directed in a
beam-like manner. In order to achieve the required
thermal neutron ¯uence at the tumour, the whole
head (in the case of brain tumours) is given a sub-
therapeutic radiation dose, which is therapeutically
e�ective in the tumour cells because of the increased
dose from the boron capture reaction. However, for
the currently available boron carrier compounds,
the degree of selectivity in concentration between
tumour and healthy tissues is still so low that care-
ful treatment planning, combined with a detailed
understanding of the e�ects of the neutron ir-
radiation on the healthy brain, are essential prere-
quisites to a therapy programme.
In describing the physical basis for BNCT, it is
necessary to use a number of terms which will be
familiar to clinicians and physicists alike. However,
in the case of BNCT, these terms may not carry
their normal inferences. For example, absorbed
dose is related to energy deposited by the radiation
®eld, and in clinical radiotherapy is directly corre-
lated with tumour control probability. In BNCT
the major dose component, that from the boron
capture reaction 10B(n,a)7Li, is deposited non-uni-
formly on the microscopic scale. The ranges of the
reaction products from this reaction are 5 and 9 mmfor the 7Li and a respectively which are comparable
with typical cellular dimensions of around 10 mm.
Hence, if the boron is located far from the radio-
sensitive structures with in cell nucleus, dose (i.e.
energy) will be deposited during the therapy with
negligible impact on tumour control. This is a cen-
tral problem to the whole of BNCT and has been
investigated by many authors (Gabel et al., 1987;
Kalend et al., 1995).
Since the tumour control is a�ected directly by
the location of the boron atoms within a cell, it will
be a�ected directly by the compound used to carry
boron to the cell. There is therefore a need to
characterise the biological e�ectiveness exhibited by
the boron distribution associated with a particular
compound. For this purpose the term Compound
Biological E�ectiveness (CBE) (Coderre et al., 1993)
has been developed. This is an empirical term which
has utility in characterising a treatment plan. It is
analogous to the conventional term Radio-
Biological E�ectiveness (RBE) since it is a simple
ratio:
CBE �dose of photons required to give a
certain surviving fraction
dose from the neutron capture reactionwhich gives the same surviving fraction
It has tended to be used as a multiplier for physi-cal dose (just as RBE is used) in order to providean overall characterisation of the e�ectiveness of a
particular treatment. However, by agreement of allparties at the recent Seventh Symposium onNeutron capture Therapy, Zurich, September 1996,
papers and patient treatment prescriptions inBNCT will carry full information on physical andbiologically equivalent dose, as well as on the con-version factors (RBE and CBE) which have been
used to derive equivalent doses from absorbed dosemeasurements and calculations.
Clinical BNCT programmes
Clinical experience and results from BNCT treat-ments is still patchy. Following early failures atBrookhaven National Laboratory (BNL) andMassachusetts Institute of Technology (MIT) using
extracted thermal neutron beams from reactors,BNCT was kept alive as a treatment modality bythe work of Hatanaka in Japan. This has resulted
in an extensive body of patient data from intra-op-erative irradiation of tumour residues post debulk-ing. The boronated compound used was BSH.
Hatanaka's work, and his use of BSH, has been avery signi®cant factor in the current European facil-ity at the High Flux Reactor (HFR) in Petten, the
Netherlands. However, this has yet to begin patienttrials. Hatanaka's work has attracted signi®cantcomment and criticism on a number of grounds re-lated to patient selection, tumour classi®cation and
the handling of the BSH compound (Dorn III,1994; Laramore and Spence, 1996). These criticismsnotwithstanding, it remains a very substantial con-
tribution to the ®eld on BNCT and provides sub-stantial encouragement for clinicians dealing withglioblastoma multiforme.
Experience with epithermal neutron beams isbeginning to mount with the on-going programmesat BNL and MIT. The BNL group have establishedthemselves as the focal point for western clinical
interest, and are undertaking a phase I/II clinicalstudy. This involves elements of dose escalation andnormal tissue toxicity, as well as paying obvious
attention to survival and quality of life. The resultscurrently being achieved show very similar meansurvival times to conventional X-ray therapy, with
hopes that further dose escalation may be possiblebefore the limit imposed by toxicity to healthy tis-sue is reached.
The published clinical experience has been sum-marised in Table 1. It should be noted that theongoing clinical studies mean that patient numbersare now increasing rapidly in some centres.
Stuart Green562
![Page 3: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/3.jpg)
Potential agents for neutron capture therapy
Whilst the majority of attention world-wide has
focused on 10B as the neutron capture isotope,there are also signi®cant possibilities from other iso-topes which exhibit high neutron capture cross-sec-
tions for thermal neutrons. These are summarisedin Table 2, with the greatest potential beingattached to 157Gd. This isotope has the highest neu-
tron a�nity of any stable isotope (248 000 barns forthermal neutrons). Gadolinium is widely used toenhance images in MRI, and is therefore widely stu-died for toxicity and chemical manipulation, and it
also has limited potential for imaging. Imaging ofthe boron distribution prior to therapy is proble-matic, although substantial progress is now being
made. It is however commonly assumed that theuse of gadolinium instead of boron would provideexcellent opportunities for imaging on any local
MRI scanner. This is unfortunately untrue sinceconventional MRI images only protons, with gado-linium enhancement producing an e�ect on proton
spin relaxation. Hence, the MRI signal is basicallyrelated to proton concentration rather than gadoli-nium concentration, and will be e�ected by thelocal blood supply to the tumour. Whilst conven-
tional MRI would provide some information, thiswould not be of the quality required for dosimetricutility in radiotherapy.
Neutron sources
The neutron beam characteristics which arerequired for a useful BNCT facility are quite di�-cult to achieve. If we consider only BNCT based on
externally applied beams of epithermal neutrons,then we can consider the design criteria put forwardby the Petten group as a useful guide (Moss et al.,
1992). These criteria, along with the characteristicsof neutron beams in use at BNL, Petten, and thatprojected for Birmingham are shown in Table 3.
Essentially one is looking to produce a beam withgood penetration, to ensure su�cient dose to thedistal edge of a target volume, with low fast neu-
tron and photon contamination, and of su�cient
intensity to allow a therapeutic dose to be delivered
in an acceptable time (less than 1 hour for
example).
The only external source of neutrons of su�cient
intensity for a practical BNCT treatment that is
currently available is that from a nuclear reactor.
While both boron enhanced fast neutron therapy,
and capture-based brachytherapy are possible,
neither currently available particle accelerators, or
radioisotope neutron sources can provide an exter-
nal beam suitable for BNCT. Nevertheless, the po-
tential utility of accelerators in a real hospital
environment provides such an attractive prop-
osition, that a number of groups around the world
are investing their research e�ort into this topic.
This e�ort gains impetus from the fairly certain
knowledge that new nuclear reactors are very unli-
kely to be located at hospital sites, and that the
potential income which could be generated from
a clinical BNCT programme is unlikely to be
su�cient to support the running costs of a nuclear
reactor.
The use of radioisotope neutron sources for
external `beam' BNCT has also been investigated
(Yanch et al., 1993). These investigations suggest
that a beam of su�cient intensity for a practical
treatment facility is unlikely to be realised by this
route. However, a number of centres are already
working clinically with neutron therapy based on
interstitial placement of neutron sources. The mod-
eration of these, usually ®ssion, neutrons within the
body, might produce a su�ciently high local ther-
mal ¯uence to mean that the introduction of a
boronated capture agent is clinically useful in cer-
tain circumstances (Allen and Ralston, 1996).
The potential for a dose enhancement in fast neu-
tron therapy has been considered at a number of
centres (Maughan et al., 1996). It is signi®cant here
to recall that tumour control probability is a very
strong function of applied dose, so the small poten-
tial for dose enhancement by using tumour speci®c
Table 1. Published BNCT clinical experience world-wide
Country/Lab Compound Condition Neutron beam No. of patients
Japan/Kagawa (Nakagawa et al., 1997) BSH Gliomas (all grades) Thermal 152Japan/JAERI (Matsumura et al., 1997) BSH Gliomas (III and IV) Thermal 4Japan/Kyoto (Ono et al., 1997) BSH/BPA Gliomas (III and IV) Thermal 44USA/BNL (Elowitz et al., 1997) BPA Glioblastoma Epithermal 10USA/MIT (MIT group, 1996) BPA Glioblastoma Epithermal 2USA/MIT (Busse et al., 1997) BPA Melanoma Epithermal 5
Table 2. Characteristics of di�erent isotopes for neutron capture therapy
Isotope Cross-section (barns)aNatural abundance
(%) Reaction Q (MeV) Products
10B 3800 19.6 2.79 a, 7Li, g157Gd 248 000 15.7 7.9 g, eÿ6Li 917 7.4 4.78 a, t235U 566 0.72 200 �, b, g, n
aValues taken from the JEF data library at 0.025 eV (JEF-2.2).
Boron Neutron Capture Therapy 563
![Page 4: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/4.jpg)
capture agents in fast neutron therapy, could have
a signi®cant impact on overall cure-rates. Clinicalenthusiasm for such an approach is constrained bythe fact that the presently available capture agents,
combined with the low capture cross-sections of iso-topes such as 10B at high energy (10B(n,a) cross-sec-tion at 10 MeV is approximately 0.06 barns), means
that the potential for dose enhancement is thoughtto be around 5%. Hence this modality sees unlikelyto have a substantial clinical impact at present.Calculations suggest (Maughan et al., 1996) that
modi®cations to conventional fast neutron beamscould improve this ®gure to in excess of 10%, andthe development of capture agents with improved
speci®city would certainly have a dramatic impacthere.
Accelerator neutron sources
A particle accelerator suitable for BNCT is esti-mated to be of a similar cost to systems for high
energy neutron and proton therapy. That is ap-proximately three to four times the cost of a con-ventional high energy linear accelerator used
routinely in hospitals around the world. It is there-fore substantially cheaper than reactor technology,and would be acceptable for an in-hospital location.From the physics and neutronics point of view
there are also a number of advantages to accelera-tor systems. It is possible, by manipulating suchparameters as incident beam, energy, and modera-
tor thickness, to produce neutron beams of di�erentcharacteristics. There is also the option of usingdi�erent target materials and di�erent incident par-
ticle types but generally attention has focused onlow energy proton accelerators with the choice ofeither lithium or beryllium as the target material.Nevertheless, there are a number of alternative
accelerator con®gurations which also deserve someattention, including moderation of beams from highenergy proton accelerators, and designs for neutron
production at threshold for direct incidence ontothe patient without a moderator. These two alterna-tive approaches will be described ®rst, before con-
centrating on the application of moderated lowenergy proton accelerators as the principal route toaccelerator based BNCT.
High energy accelerators. The role of cyclotronsin the treatment of cancer is now well established.Their contribution to both fast neutron and protontherapy is substantial, and they are now an estab-
lished part of the armoury against cancer around
the world. It is also possible that high energy accel-
erators such as the one at PSI, Switzerland, will be
useful in producing a beam suitable for BNCT.
Moderator design studies (Teichmann and Craw-
ford, 1996) have been undertaken which indicate
that a suitable epi-thermal beam can be produced
from this facility.
Unmoderated low energy accelerator systems. It
may also be possible to use a low energy accelerator
without a moderation system (Kononov, 1996) In
this case, the target material, usually lithium, is
bombarded with protons with an energy of around
1890 keV, i.e. only slightly above the threshold for
this reaction at 1881 keV. From kinematic consider-
ations, there will be neutron emission in a narrow
forward directed cone, with neutron energies in the
desired epithermal (around 10 keV) region. Since
the neutron production cross-section of the target
material will be changing very rapidly near
threshold, small ¯uctuations in incident particle
energy may lead to dramatic changes in all proper-
ties of the emergent neutron beam, i.e. in its energy,
angular distribution, and intensity. Hence, this
approach, while still requiring the production of
high proton beam currents and targets with high
power dissipation capabilities, also requires very
good voltage stability in the accelerator system.
One very interesting variant of the accelerator
option for BNCT, is that described by Song et al.
(1996) using a miniature accelerator tube, stereotac-
tically inserted into the centre of the tumour
through a bore-hole in the skull. This approach
could, in principle, provide a very substantial
increase in dose-rate around the tumour, but with
some cost in terms of the radiobiological character-
istics of the beam. Further technical work is necess-
ary to determine the heat removal capabilities of
such a miniature accelerator and target system.
Moderated low energy accelerator systems. Low
energy accelerators, generating neutrons up to
around 1.5 MeV, require quite small, compact, and
inexpensive beam moderation systems such as the
one shown in Fig. 1. This kind of accelerator sys-
tem for BNCT is favoured by many laboratories
around the world (Allen and Beynon, 1995; Yanch
et al., 1992; Chu et al., 1996) and appears to o�er
the best possibility for clinical use. The design of
these systems can be tuned to produce neutron
beams with di�erent characteristics, as illustrated in
Table 3. Neutron beam characteristics of a number of BNCT centres
Centre
Useful neutron¯uence-rate(cmÿ2 sÿ1)
Mean neutron kerma perunit neutron ¯uence
(Gn) (Gy cm2)
Mean photon kerma perunit neutron ¯uence
(Gp) (Gy cm2)
Petten Design (Moss et al., 1992) r1� 109 R8.1� 10ÿ13 R2.8� 10ÿ13
BNL I (Liu, 1996) 1.4� 109 4.5� 10ÿ13 1.5�10ÿ13
BNL II (Liu, et al., 1997) 0.84� 109 4.8� 10ÿ13 2.0�10ÿ13
Birmingham (Allen and Beynon, 1995)a 0.82� 109 7.4� 10ÿ13 0.8�10ÿ13
aFor a moderator depth of 20 cm and a proton beam of 5 mA at 2.8 MeV.
Stuart Green564
![Page 5: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/5.jpg)
Fig. 2. Increased thermal components in the beam
(lower Gn in Fig. 2) would be useful for delivering
an enhanced boron-capture dose to more super®cial
region, while beam apertures, decreased moderation
and the use of thermal neutron absorbers such as
lithium in the moderation system, can be used toenhance the beam penetration. In this way it ispossible, in principle, to tune the delivered neutron
beam to suit a range of clinical situations. However,this process must be accompanied by detailed radio-biological studies to assess the impact of each beamadjustment on the response of normal tissues placed
in the beam.
Target cooling
The removal of heat from an accelerator target isone of the principal technological challenges to beovercome by all low energy accelerator systems.
The melting point of lithium metal is around1808C, which means that heat removal must behighly e�cient if the target is to remain solid. The
minimum beam power requirements for a moder-ated low energy accelerator system approximates toa proton beam current of 5 mA at an energy of
2.5 MeV. This is a beam power of 12.5 kW, depos-ited over a few square centimetres in a target layerwhich is only around 150 mm thick. The power den-
sity is therefore approximately 1 kW/cm2 for a uni-form 4 cm diameter beam and is near the limit thatcan be removed by water-based cooling mechanisms(see Table 4).
Fig. 1. The phase I beam moderator which has been experimentally validated in Birmingham.
Fig. 2. The variation in free beam parameters available byaltering the moderator depth in the Birmingham phase Imoderator design. Here Gn and Gp are the mean neutronand photon kermas per unit neutron ¯uence, the protonenergy is 2.8 MeV and the current is 10 mA (reproduced
with permission of IOP Publishing Ltd, Bristol, UK).
Boron Neutron Capture Therapy 565
![Page 6: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/6.jpg)
Of the available options it appears that sub-
merged jet cooling is the only option worth pursu-ing. This involves placing the back surface of thetarget in contact with a ¯uid bath which containsa jet of the same ¯uid, injected via a nozzle onto
the centre of the target. In this way the jet ismaintained at a lower temperature than in forcedconvection cooling and in addition, localised boil-
ing is allowed on the target rear surface. Anyvapour bubbles are blown away by the impact ofthe jet which spreads out radially from the centre.
Experimental evidence from Blackburn et al.(1996) suggests that submerged jet cooling iscapable of accommodating power densities up to
6 kW/cm2, which gives a factor of 6 safety marginover that required for a uniform beam power, andshould deal adequately with the non-uniform pro-ton beams which are likely to be encountered in
practice.
DOSIMETRY FOR BNCT BEAMS
The dosimetry of the neutron beams used in
BNCT is extremely complex. Neutron interactionsin boron-loaded and normal tissues give rise to arange of secondary particles which deposit the radi-ation dose. The very wide spread in LET of these
reaction products, and by implication the widerange of RBEs involved, means that these dosecomponents must be separately quanti®ed if an
accurate overall assessment of the radiation dose isto be made.Practical beam dosimetry in photon radiotherapy
is exclusively based on ion-chambers, with referenceto Primary standards which may either be based onionisation yield or on more direct measures ofenergy deposited such as calorimetry. Centres in
North America that already have active clinicalBNCT programmes have also chosen to use ion-chambers as their routine dosimetric tool (Rogus
et al., 1994) with reference to a Primary standardthrough a photon calibration of the dosimeters. InEurope, e�orts are in hand to develop an agreed
dosimetry protocol amongst the centres that arepursuing clinical BNCT facilities (Stechet-Rasmussen et al., 1996).
A number of authors have reported on the use ofthe microdosimetric technique for BNCT dosimetry(Wu et al., 1992; Maughan et al., 1992). InBirmingham, we are also attempting to further
understand the radiation absorbed dose depositionby making use of the microdosimetric technique
using specially designed proportional counters. Thefeasibility of this approach has been demonstratedat low ¯uence rates, but needs to be con®rmed in
the full therapy beam (Green et al., 1996).These techniques notwithstanding, macroscopic
dosimetry (i.e. determination of absorbed dose in
cm sized voxels) is not truly possible withoutsome information on the macroscopic distributionof boron in the tissues of the patient. It has
become routine practice in active BNCT centres,to rely on a pre-therapy bio-distribution study tomeasure the boron partitioning between tumour,blood and normal brain tissue, for a particular
infusion protocol. These relative partitions areassumed to be reproduced at the time of therapy,even when this follows a de-bulking procedure
(Chanana, 1996).Whilst these approximations are certainly appro-
priate for early trials, greater understanding of clini-
cal outcomes may be possible if the borondistribution can be measured at the time of thetherapy irradiation. Promising techniques in this
area are based on spectroscopic magnetic resonanceimaging (Flego et al., 1996) where the spin relax-ation of 11B is measured, and also on directmeasurement and possible imaging of the capture
gammas from the 10B capture reaction (Allen et al.,1996).
EXPERIENCE IN BIRMINGHAM
The dynamitron accelerator
The Birmingham BNCT programme is basedaround the 3 MV Dynamitron accelerator in the
School of Physics and Space Research at TheUniversity of Birmingham. In such an accelerator,the terminal voltage is maintained through a
series of Cockroft±Walton type voltage doublingstages, but unlike a conventional Cockroft Waltoncascade accelerator where power is delivered in
series through the chain of doubling stages, in theDynamitron, radio-frequency power is delivered inparallel from large cylindrical plates placed out-side the doubling chain, and coupled to it via cir-
cular metal strips. This coupling of the powerdelivery in parallel means that much higher char-ging rates are possible, giving the Dynamitron a
potential for stable production of high beampowers.In addition, the Dynamitron is built with an in-
ternal electrode geometry which is designed toreduce the path for scattered and divergent par-ticles, preventing them from reaching signi®cant
energies within the accelerator column. Hence theproduction of background X-rays and neutronswithin the accelerator column is minimised. In itspresent con®guration, the accelerator has produced
Table 4. Heat transfer coe�cients from a range of cooling mechan-isms (Blackburn et al., 1996)
Mechanism Coe�cient (W/m2 K)
Free Liquid Convection 50±1000Forced Liquid Convection 50±20 000Convection with a phase change 2500±105
Submerged jet with a phase change >6� 105
Stuart Green566
![Page 7: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/7.jpg)
reliable beams of around 2 kW power (1 mAprotons at 2 MV).
The Birmingham programme is nearing the endof a second design stage. Work on the initial beammoderation system has now been completed. Thisincludes design (Allen and Beynon, 1995), measure-
ment validation (Tattam et al., 1996) and dosimetriccharacterisation (Green et al., 1996). Work isshortly to begin on two key areas, namely the
upgrade of the accelerator ion source to give thepotential for beams up to 10 mA, and constructionof a beam moderation system which will produce
a horizontal epithermal neutron beam from anincident vertical proton beam.
CHARACTERISATION OF THE BIRMINGHAMEPITHERMAL BEAM
Initial design studies for an accelerator basedneutron source for BNCT applications have been
performed using the radiation transport codeMCNP (Briesmeister, 1993). These have been fol-lowed by experimental validation based on the sys-
tem similar to that shown in Fig. 1, but without theouter D2O and Li shield. This is basically the Litarget and moderator system described by Tattam
et al. (1996).
Outline of the experimental programme
The experimental programme comprised the fol-
lowing elements.
1. Construction of a rectangular Perspex enclosedwater phantom. Following the computational
work of Gupta et al. (1993) we have chosen aphantom of dimensions 17 cm�14 cm�15 cm,which is the intermediate of the three phantoms
which they investigated.2. Measurements at four positions along the central
axis of this phantom of:
. fast neutron dose from a Tissue Equivalent(TE) microdosimetric detector
. photon dose from a TE microdosimetric detec-
tor. boron dose indirectly using thermal ¯uence®gures derived from gold foil activation
3. Comparison of absolute experimental results (interms of dose rate/mA of proton beam) with anabsolute MCNP simulation of the exact exper-imental con®guration.
4. Assessment of the clinical utility of this beamusing the CBE factors derived for BPA and thefast neutron RBE from the Brookhaven group
(Chanana, 1996). These are:
CBE, Tumour + BPA 3.8CBE Normal Brain + BPA 1.3RBE, photons 1.0RBE, neutrons 3.2Tumour 10B concentration 30 mg gÿ1
Normal Brain 10B concentration 10 mg gÿ1
Validation of the MCNP simulations
The comparisons between experimental andMCNP simulation results are represented in Table 4as ratios between the experimental and simulation
results at each depth. Hence good agreementbetween experiment and calculation would result ina value of 1.0. It is clear from Table 5 that the vali-dation is successful for fast neutron, photon, and
boron dose derived from gold foils.
Dosimetric data
In order to derive estimates of the likely biologi-cally equivalent dose which will be delivered by theBirmingham beam, and to further understand its
clinical utility, the factors listed above for concen-tration, CBE and RBE have been combined withthe physical dose measurements to give informationon the variation with depth of each dose com-
ponent to both tumour and normal tissue. Thesedata are represented in Table 6 which shows bothphysical (Gy) and Gy-equivalent dose data at 5 cm
deep in phantom.Hence, from Table 6, the total dose to normal
brain (in Gy-Eq) at 5 cm deep is only approxi-
mately 26% of the total dose to tumour at thisdepth, whereas the corresponding ®gure in terms ofphysical dose is 57%. It is therefore clear that by
using our understanding of the di�erent degrees ofcell damage that result from the di�erent dose com-ponents in BNCT, a signi®cantly increased relativedose to the tumour is predicted, being greater than
that derived from considerations of physical dosealone.
SUMMARY AND CONCLUSIONS
The ®eld of accelerator based BNCT is rapidlyexpanding, re¯ecting an increased clinical interest
Table 6. Relative doses at 5 cm deep in a water phantom exposedto an epithermal neutron beam
Physical dose (RBE or CBE)�dose
Component TumourNormalbrain Tumour
Normalbrain
10B 64.5 21.5 83.2 9.5Neutron 6.5 6.5 7.0 7.0Photon 29.0 29.0 9.8 9.8Total 100.0 57.0 100.0 26.3
Table 5. Ratio of experiment to MCNP simulation at di�erentdepths in a light water phantom
Depth (cm) Photon Neutron 10B (Au foils)
2.5 0.9920.09 0.9720.07 0.9820.065.0 0.9620.10 1.0020.07 0.9920.067.5 0.9920.10 0.9420.08 0.9120.0512.7 1.1820.13 0.8320.07 1.2320.07
Boron Neutron Capture Therapy 567
![Page 8: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/8.jpg)
following positive results from work in Japan andNorth America. Data presented here indicate that
the enhanced toxicity of the boron capture reactionleads to a dose enhancement which appears favour-able for therapy. The ways in which particle accel-
erators may be used in BNCT are diverse, andcover a possible range of incident particle (almostexclusively proton) energies from around 2 to
600 MeV, with accelerators ranging in size fromthose which might be su�ciently small to allowdirect insertion into the tumour site, to spallation
sources which use synchrotron rings of many metresin diameter.This area has become very a very productive one
for radiation physicists throughout the world, who
function as part of multi-disciplinary teams broughttogether to struggle with the many scienti®c andtechnical challenges that BNCT involves. However,
this is not simply a subject for abstract research.Despite the best e�orts of modern medicine,patients diagnosed with the tumour glioblastoma
multiforme will be dead within 5 years of the initialdiagnosis, the vast majority dying within the ®rst 2years. BNCT o�ers some hope for these patients.
AcknowledgementsÐThis paper has relied heavily on visitsmade in 1996 to Brookhaven National Laboratory, MITand to the Seventh International Symposium on NeutronCapture Therapy, Zurich. Detailed discussion and datasupplied by the following people is gratefully acknowl-edged Jacek Capala, Je� Coderre, Ben Liu, LucianWielopolski and Dr Annan Chanana (all BNL) andRobert Zamenhof, Jacqualine Yanch and Prof. OttoHarling (all MIT). Thanks goes to the RadiologicalResearch Trust for funding for the trip to the Zurich con-ference. For their experimental microdosimetry atBirmingham, thanks also go to Richard Maughan andChandrasekar Kota of Harper Hospital, Detroit, and inaddition the assistance of the many members of theBirmingham group is gratefully acknowledged, particularlyDavid Tattam and Dennis Allen.
REFERENCES
Allen, D. A. and Beynon, T. D. (1995) A design study foran accelerator-based epithermal neutron beam forBNCT. Physics in Medicine & Biology 40, 807±821.
Allen, D. A., Beynon, T. D. and Perks, J. (1996) In-vivoon-line 10B(n,a) 3D dose measurements using binaryGabor zone plate encoded g-ray holography. ConferenceProceedings, 7th International Symposium on NeutronCapture Therapy. Vol. 1, 229±236.
Allen, B. J. and Ralston, A. (1996) Boron dose enhance-ment for 252Cf brachytherapy. Conference Proceedings,7th International Symposium on Neutron CaptureTherapy. Vol. 1, 271±274.
Blackburn, B. W., Klinkowstein, R. E., Yanch, J. C.,Song, H. and Howard, W. (1996) Development of ahigh-power, water-cooled beryllium target for the pro-duction of neutrons in a high-current tandem accelera-tor. 14th International Conference on the Applicationof Accelerators, Denton, Texas, Nov. 1995.
Briesmeister, J. F. (Ed) (1993) MCNP2ÐA generalMonte Carlo N-particle transport code, version 4A. LosAlamos 12625, M Manual UC705 & UC700.
Busse, P., Zamenhof, R., Madoc-Jones, H., Solares, G.,Kiger, S., Riley, K., Chuang, C., Rogers, G. and
Harling, O. (1997) Clinical follow-up of patients withmelanoma of the extremity treated in a phase I boronneutron capture therapy protocol. ConferenceProceedings, 7th International Symposium on NeutronCapture Therapy. Vol. 1, 60±64.
Chanana, A. D. (1996) Boron neutron-capture therapy ofglioblastoma multiforme at the Brookhaven MedicalResearch Reactor, Protocol #4.
Chu, W. T., Donahue, R. J., Gough, R. A., Kwan, J.,Leung, K.-N., Ludewigt, B. A., Phillips, T. L.,Reginato, L., Staples, J., Wells, R. and Yu, S. (1996)Design of a new BNCT facility based on ESQ accelera-tor. Conference Proceedings, 7th InternationalSymposium on Neutron Capture Therapy. Vol. 1, 533±537.
Coderre, J. A., Maker, M. S., Micca, P. L., Nawrocky, M.N., Liu, H. B., Joel, D. D., Slatkin, D. N. and Amols,H. I. (1993) Derivations of RBE for the high LET reac-ton produced during BNC irradiations of the 9L ratgliosarcoma in vitro and in vivo. Int. J. Radiat. Oncol.Biol. Phys. 27, 1121±1129.
Dorn, R. V. III (1994) Boron neutron capture therapy(BNCT): A radiation oncology perspective. Int.J. Radiat. Oncol. Biol. Phys. 28, 1189±1201.
Elowitz, E. H., Chadha, M., Coderre, J. A., Joel, D.,Slatkin, D. N. and Chanana, A. D. (1996) A phase I/IItrial of BNCT for glioblastoma multiforme using intra-venous boronophenylalanine±fructose complex andepithermal neutrons: Early clinical results. ConferenceProceedings, 7th International Symposium on NeutronCapture Therapy. Vol. 1, 56±59.
Flego, M., Nicolay, K. and Watkins, P. (1996) Three-dimensional back-projection reconstruction methods formagnetic resonance imaging (MRI) of nuclei with shortrelaxation times. Conference Proceedings, 7thInternational Symposium on Neutron Capture Therapy.Vol. 1, 289±293.
Gabel, D., Foster, S. and Fairchild, R. G. (1987) TheMonte Carlo simulation of the biological e�ect of the10B(n,a) 7Li reaction in cells and tissue and its impli-cation for boron neutron capture therapy. RadiationResearch 111, 14±25.
Green, S., Kota, C., James, N., Maughan, R., Tattam, D.A., Beddoe, A. H., Beynon, T. D. and Weaver, D. R.(1996) Dosimetric characteristics of an accelerator basedbeam for boron cancer therapyÐexperimental results.Conference Proceedings, 7th International Symposium onNeutron Capture Therapy. Vol. 1, 217±221.
Gupta, N., Niemkiewicz, J., Blue, T. E., Gahbauer, R. andQu, T. X. (1993) E�ect of head phantom size on 10Band 1H(n,g)2H dose distributions for a broad ®eld accel-erator epithermal neutron source for BNCT. Med. Phys.20, 395±404.
Hatanaka, H. and Nakagawa, Y. (1994) Clinical results oflong-surviving brain tumour patients who underwentboron neutron capture therapy. Int. J. Radiat. Oncol.Biol. Phys. 28, 1061±1066.
James, N. (1996) Presentation at the 1996 Meeting of theAssociation for Radiation Research, Birmingham, UK.
JEF-2.2. Joint Evaluated File version 2.2, Nuclear EnergyAgency Data Bank, 12, Bld des Iles, 92130 Issy-Les-Moulineaux, France.
Kalend, A. M., Bloomer, W. D. and Epperly, M. W.(1995) Dosimetric consequences of 10B(n,a)7Li reactionoccurring at the cellular membrane. Int. J. Radiat.Oncol. Biol. Phys. 31, 171±178.
Kononov, V. N., Regushevskiy, V. I., Soloviev, N. A. andLeipunskiy, A. I. (1996) Accelerator-based and intensedirected neutron source for BNCT. ConferenceProceedings, 7th International Symposium on NeutronCapture Therapy. Vol. 1, 528±532.
Laramore, G. E. and Spence, A. M. (1996) Boron neutroncapture therapy (BNCT) for high grade gliomas of the
Stuart Green568
![Page 9: DEVELOPMENTS IN ACCELERATOR BASED BORON NEUTRON CAPTURE THERAPY](https://reader031.fdocuments.in/reader031/viewer/2022020311/575023481a28ab877ea92555/html5/thumbnails/9.jpg)
brain: A cautionary note. Int. J. Radiat. Oncol. Biol.Phys. 36, 241±246.
Liu, H. B. (1996) Neutron capture therapy facilities at theBrookhaven Medical Research Reactor, private com-munication.
Liu, H. B., Capala, J., Joel, D. D., Brugger, R. M. andRover, D. C. (1997) NCT facilities at the BrookhavenMedical Research Reactor. Advances in Neutron CaptureTherapy. Vol. 1, 311±315, Eds. B. Larsson, J. Crawfordand R. Weinreich, Elsevier, Amsterdam.
Matsumura, A., Shibata, Y., Yamamoto, T., Yamada, T.,Fujimori, H., Nakai, K., Nakagawa, Y., Hayakawa, Y.,Isshiki, M. and Nose, T. (1996) The University ofTsukuba BNCT research group: ®rst clinical experiencesat JAERI. Conference Proceedings, 7th InternationalSymposium on Neutron Capture Therapy. Vol. 1, 46±50.
Maughan, R. L., Kota, C. and Forman, J. D. (1996)Feasibility of boron neutron capture enhancement offast neutron therapy using a superconducting cyclotron.Conference Proceedings, 7th International Symposium onNeutron Capture Therapy.
Maughan, R. L., Kota, C. and Yudelev, M. (1992) Amicrodosimetric study of the dose enhancement in a fastneutron beam due to boron neutron capture. Physics inMedicine & Biology 37, 1957±1961.
MIT group (1996) Oral contribution at 7th InternationalSymposium on Neutron Capture Therapy.
Moss, R. L., Stecher-Rasmussen, F., Ravensberg, K.,Constantine, G. and Watkins, P. (1992) Design con-struction and installation of an epithermal neutronbeam for BNCT at HFR Petten. Progress in NeutronCapture Therapy for Cancer, pp. 63±66. Ed. B. J. Allen,Plenum Press, New York.
Nakagawa, Y., Kyonnghon, P., Kitamura, K., Kageji, T.and Minobe, T. (1996) What were important factors inpatients treated with boron neutron capture therapy inJapan? Conference Proceedings, 7th InternationalSymposium on Neutron Capture Therapy. Vol. 1, 65±70.
Ono, K., Ueda, S., Oda, Y., Nakagawa, Y., Miyatake, S.,Takagaki, M., Osawa, M. and Kobayashi, T. (1996)Boron neutron capture therapy for malignant glioma at
Kyoto University reactor. Conference Proceedings, 7thInternational Symposium on Neutron Capture Therapy.Vol. 1, 39±45.
Rogus, R. D., Harling, O. K. and Yanch, J. C.(1994) Mixed ®eld dosimetry of epithermal neutronbeams for boron neutron capture therapy at the MITR-II research reactor. Med. Phys. 21, 1611±1625.
Song, H., Yanch, J. C. and Klinkowstein, R. E. (1996) Aninterstitial intracavity accelerator-based neutron sourcefor fast neutron brachytherapy. Conference Proceedings,7th International Symposium on Neutron CaptureTherapy. Vol. 1, 543±548.
Stecher-Rasmussen, F., Auterinen, I., Beddoe, A.,Goncalves, I., Jarvinen, H., Kosunen, A., Larsson, B.,Marek, M., Mijnheer, B., Raajmakers, C. P. J.,Savolainen, S., Voorbraak, W. P., Watkins, P. andZsolnay, E. (1996) A code of practice for the dosimetryof BNCT in Europe. Conference Proceedings, 7thInternational Symposium on Neutron Capture Therapy.Vol. 1, 237±240.
Tattam, D. A., Allen, D. A., Beynon, T. D., Constantine,G., Green, S., Scott, M. C. and Weaver, D. R. (1996)Preliminary neutron ¯uence measurements in theBirmingham BCT beam. Conference Proceedings, 7thInternational Symposium on Neutron Capture Therapy.Vol. 1, 472±476.
Teichmann, S. and Crawford, J. F. (1996) Theoreticalstudy of a spallation neutron source for BNCT.Conference Proceedings, 7th International Symposium onNeutron Capture Therapy. Vol. 1.
Wu, C. S., Amols, H. I., Kliauga, P., Reinstein, L. E. andSaraf, S. (1992) Microdosimetry for boron neutron cap-ture therapy. Radiation Research 130, 355±359.
Yanch, J. C., Kim, J. K. and Wilson, M. J. (1993) Designof a californium-based epithermal neutron beam forneutron capture therapy. Physics in Medicine & Biology38, 1145±1155.
Yanch, J. C., Zhou, X.-L., Shefer, R. E. andKlinkowstein, R. E. (1992) Accelerator-based epithermalneutron beam design for neutron capture therapy. Med.Phys. 19, 709±721.
Boron Neutron Capture Therapy 569