Development of a High Pressure Xenon Imager with Optimal
Energy Resolution
Azriel Goldschmidt on work with David Nygren
Instrumentation Series SeminarLBNL, March 2009
![Page 2: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/2.jpg)
Physics Motivations
• Neutrinoless double beta decay ():– Tests Majorana nature of neutrino – Helps determine absolute neutrino mass– If observed, lepton number NOT conserved– Current situation: controversial (one claim), may
require new and richer approach
• WIMP dark matter:– Direct detection– Current situation: controversial (one claim), may
require new and richer approach This talk focuses on the
![Page 3: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/3.jpg)
• Rare nuclear transition between same mass nuclei– Energetically allowed for even-even nuclei
• (Z,A) → (Z+2,A) + e-1 + 1 + e-
2 + 2
• (Z,A) → (Z+2,A) + e-1 + e-
2
• (Z,A) → (Z+2,A) + e-1 + e-
2 +
![Page 4: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/4.jpg)
Double Beta Decay Spectra
![Page 5: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/5.jpg)
50 meVOr ~ 1027 yr
Normal
Inverted
H-M Claim
0.1
1
10
100
1000
Eff
ecti
ve
M
ass
(meV
)
12 3 4 5 6 7
102 3 4 5 6 7
1002 3 4 5 6 7
1000Minimum Neutrino Mass (meV)
Ue1
= 0.866 m2
sol = 70 meV
2
Ue2
= 0.5 m2
atm = 2000 meV
2
Ue3 = 0
Inverted
Normal
Degenerate
![Page 6: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/6.jpg)
How to look for neutrino-less decay
• Measure the spectrum of the electrons
![Page 7: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/7.jpg)
Experiments• CANDLES 48Ca CaF2 scintillator crystals• COBRA116Cd CdZnTe crystals• CUORE 128Te TeO2 Bolometers• EXO 136Xe Liquid Xenon TPC• GERDA 76Ge Enriched Ge diode• MAJORANA 76Ge Enriched Ge diode• SNO+ 150Nd Nd loaded liquid scintillator• SuperNEMO 82Se Foils in track/calorimeter
![Page 8: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/8.jpg)
Past Results
Elliott & VogelAnnu. Rev. Part. Sci. 2002 52:115
48Ca >1.4x1022 y <(7.2-44.7) eV
76Ge >1.9x1025 y <0.35 eV
76Ge >1.6x1025 y <(0.33-1.35) eV
76Ge =1.2x1025 y = 0.44 eV
82Se >2.1x1023 y <(1.2-3.2) eV
100Mo >5.8x1023 y <(0.6-2.7) eV
116Cd >1.7x1023 y <1.7 eV
128Te >7.7x1024 y <(1.1-1.5) eV
130Te >3.0x1024 y <(0.41-0.98) eV
136Xe >4.5x1023 y <(1.8-5.2) eV
150Nd >1.2x1021 y <3.0 eV
102
103
104
105
20202000198019601940Year
Ge-76Ge-76
Te-128
Te-128
Se-82
Ge-76
Ge-76Ca-48
Nd-150Nd-150
![Page 9: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/9.jpg)
20
15
10
5
205020402030202020102000
Energy (KeV)
20
15
10
5
205020402030202020102000
Energy (KeV)
NIM
A52
2, 3
71 (
200
4)• H-M: Only claimed evidence of detection
with 11 kg of 86% enriched 76Ge for 13 years
•Klapdor-Kleingrothaus et al Phys.Lett.B586:198-212,2004.
T1/2~1.19x1025y<m> ~ 0.44 eV
![Page 10: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/10.jpg)
CUORE• Cryogenic “calorimeters”
• CUORICINO: 40.7kg TeO2 (34% abundant 130Te)
– T ≥ 3.0 × 1024 yr (90% C.L.)
– <m> ≤ 0.19 – 0.68 eV
– Resolution E/E = 2 x 10-3 FWHM at 2.5 MeV
• CUORE ~1000 crystals, 720 kg 60Co (cosmogenic) decays, 2 gammas summed
![Page 11: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/11.jpg)
“Gotthard TPC” Pioneer TPC detector for 0- decay search
– 5 bars, enriched 136Xe (3.3 kg) + 4% CH4
– MWPC readout plane, wires ganged for energy– No scintillation detection no TPC start signal!
• No measurement of drift distance
E/E ~ 80 x 10-3 FWHM (1592 keV) 66 x 10-3 FWHM (2480 keV)
Reasons for this less-than-optimum resolution are not clear…
Likely: uncorrectable losses to electronegative impurities
Possible: Undetectable losses to quenching (4% CH4)
But: ~30x topological rejection of interactions!
![Page 12: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/12.jpg)
EXO-200 200 kg Enriched 136XeCharge & scintillation
light readout
![Page 13: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/13.jpg)
EXO-200 expected E resolutionAnticorrelation between ionization and scintillation signals in
liquid xenon can be used to improve the energy resolution
Extrapolates to E/E = 33 10-3 FWHM @ Q0
570 keV
![Page 14: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/14.jpg)
Energy resolution in a Xe Dual Phase (XENON)
Aprile, Paris 2008
Extrapolates to E/E = 21 10-3 FWHM @ Q0
![Page 15: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/15.jpg)
What’s needed…
• Long lifetimes (>1025 years) require:– Large Mass of relevant isotope (>100 kg)– Small or No background:
• Clean materials• Underground, away from cosmic rays• Background rejection methods:
– Energy resolution– Event topology– Particle identification– Identification of daughter nucleus
– Years of data-taking
![Page 16: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/16.jpg)
Why use Xe for search
• Only inert gas with a candidate• Long lifetime ~1022-1023 y (not seen yet) • No need to grow crystals• Can be re-purified in place (recirculation)• No long lived Xe isotopes• Noble gas: easier to purify• 136Xe enrichment easier (natural 8.9%):
- noble gas (no chemistry involved)
- centrifuge efficiency ~ Δm (136Xe vs. ~131Xe)
![Page 17: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/17.jpg)
Energy partition in xenon
• When a particle deposits energy in xenon, where does the energy go?– Ionization
– Scintillation: VUV ~170 nm (1, 2 …)
– Heat
• How is the energy partitioned?– Complex responses, different for , , nuclei– Dependence on xenon density , E-field– Processes still not perfectly understood
![Page 18: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/18.jpg)
LXe or HPXe?
With high-pressure xenon (HPXe)
A measurement of ionization alone
is sufficient to obtain
good energy resolution…
![Page 19: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/19.jpg)
Xenon: Strong dependence of energy resolution on density!
For >0.55 g/cm3, energy resolution deteriorates rapidly
Ionization signal only
![Page 20: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/20.jpg)
E/E ~ 1-2 10-3 FWHM
E/E ~ 35 10-3 FWHM
![Page 21: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/21.jpg)
![Page 22: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/22.jpg)
![Page 23: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/23.jpg)
What is this factor “G”?
• In a very real sense:
G is a measure of the precision with which a single electron (from an ionizing track) can be counted.
![Page 24: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/24.jpg)
Electro-Luminescence (EL) (Gas Proportional Scintillation)
– Electrons drift in low electric field region– Electrons then enter a high electric field region– Electrons gain energy, excite xenon, lose energy– Xenon generates UV– Electron starts over, gaining energy again– Linear growth of signal with voltage– Photon generation up to ~1000/e, but no ionization– Early history irrelevant, fluctuations are small– Maybe… G ~ F?
![Page 25: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/25.jpg)
Electroluminescence in 4.5 bar of Xenon
Corresponds to 5 x 10-3 FWHMWhen naively extrapolated toQ of 2.5 MeV(compare with the Fano limited 2.8 x 10-3 FWHM best case)
![Page 26: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/26.jpg)
Fluctuations in Electroluminescence (EL)
EL is a linear gain process
G for EL contains three terms:1. Fluctuations in nuv (UV photons per e):
2. Fluctuations in npe (detected photons/e):
3. Fluctuations in photo-detector single PE response:
G = 2 = 1/(nuv) + (1 + 2pmt)/ npe)
For G = F = 0.15 npe ≥ 10
The more photo-electrons, the better!
Equivalent noise: much less than 1 electron rms!
![Page 27: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/27.jpg)
Virtues of an EL readout
• Immune to microphonics
• Absence of positive ion space charge
• Linearity of gain versus pressure, HV
• Isotropic signal dispersion in space
• Trigger, energy, and tracking functions accomplished with optical detectors
![Page 28: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/28.jpg)
Detector Concept
• Use enriched High Pressure Xenon
• TPC to provide image of the decay particles
• Design to also get an energy measurement as close to the intrinsic resolution as possible
![Page 29: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/29.jpg)
High-pressure xenon gas TPC
• Fiducial volume surface:– Single, continuous, fully active, variable,...– 100.00% rejection of charged particles
(surfaces) – TPC with t0 to place event in z coordinate
• Tracking:– Available in gas phase only– Topological rejection of single electron events
![Page 30: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/30.jpg)
Separated Function TPC with Electroluminescence
Readout Plane A
- position
Readout Plane B- energy
Electroluminescent Layer
![Page 31: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/31.jpg)
Electro-Luminescent Readout
For optimal energy resolution, 105 e- * 10 pe/e- = 106 photoelectrons need to be detected!
Energy readout plane is a PMT array
• electron (secondary) drift is very slow: ~1 mm/s• This spreads out the arriving signal in time - up to 100 s
for many events• The signal is spread out over the entire readout cathode-
side, 100’s of PMTs• These two factors greatly reduce the dynamic range
needed for readout of the signals No problem to read out <5 kev to >5000 keV
![Page 32: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/32.jpg)
Single 2.49 MeV e- in 20 atm Xe(background) MC simulation
~5 cm
Using tracking information, separate single electrons (like these) from 2-electron events that should have “blobs” at BOTH ends of the combined track
![Page 33: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/33.jpg)
Backgrounds for the search
NEXT Collaboration
![Page 34: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/34.jpg)
Can one measure Ba++ Directly?
• Extract the ion from the high pressure into a vacuum
• Measure mass and charge directly
• A mass 136, ++ ion is a unique signature of Ba++. (Assumption is Xe++ cannot survive long enough to be a problem)
• This has been done for Ba++ in Ar gas
Sinclair, TPC Workshop Paris 2008
![Page 35: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/35.jpg)
Barium ions are guided towards the exit orifice and focused using an asymmetric field technique. The second chamber is maintained at a pressure of ~10-30 mbUsing a cryopump and is lined with an RF carpet. An RF funnel guides the ionsTowards the RF quadrupole which is at high vacuum. The ion is identified usingTOF and magnetic rigidity
Sinclair, TPC Workshop Paris 2008
![Page 36: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/36.jpg)
Top EL/Scint Detector (Tracking)
Bottom EL/Scint Detector (Energy)
Field Cage
EL Grid
CathodeGrids
BaChannel
Sinclair, TPC Workshop Paris 2008
![Page 37: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/37.jpg)
HPXe and the Dark Matter search
• Liquid Xenon has the lead on this (since energy resolution is not critical), however
• HPXe offers better discrimination between nuclear recoils and electrons
• There are ideas that would enable a lower threshold in the gas phase (useful for testing the DAMA/LIBRA positive result)– Challenge at low recoil energy for both LXe and HPXe
is that the primary scintillation signal (for trigger ing and fiducialization) is Tiny
![Page 38: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/38.jpg)
Big Impact for WIMP Search in LXe
Scintillation (S1) & Ionization (S2) are the
signals used to reject electron recoils: S2/S1
But, in LXe:
S2/S1 fluctuations are anomalously large
Bad news for discrimination power in LXe…
(though may be not critical in the presence of self shielding)
![Page 39: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/39.jpg)
7-PMT 20 Bar Test Cell
anode + fluorescence grid
cathode
J. White, TPC08
![Page 40: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/40.jpg)
7-PMT,20 barTest Cell
1 inch
R7378AJ. White, TPC08
![Page 41: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/41.jpg)
Going from concept to an R&D program at LBNL
• Build a test detector large enough to demonstrate ~ E/E5 * 10-3 at ~2.5 MeV– Gas system (to take out electronegative impurities)– Energy side readout (PMTs most likely)– Enough “tracking-side” sensors to achieve energy resolution– 50 cm size chamber to house ~20 Atm of Xe
• Monte Carlo simulation for detector optimization ongoing (2 students and 0.25*yours-truly)
• Groups in Canada/US (part or EXO) and in Spain (NEXT collaboration) pursuing this line of research as well –both healthy competition and collaboration-
Stay tuned…and thanks for listening!
![Page 42: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/42.jpg)
Other possible uses of HPXe imagers with optimal resolution
• Nuclear safeguardsCheck fuel content of fuel rods
• Homeland security Directional information from
“Compton camera” to identify U and Pu isotopes from a “source”
![Page 43: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/43.jpg)
Summary
![Page 44: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/44.jpg)
Backup Slides
![Page 45: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/45.jpg)
Double beta decay
Only
2-v decays
Rate
Energy Q-value
Only
0-v decays
No backgrounds above Q-value
The ideal result is a spectrum of all events, with a 0- signal present as a narrow peak, well-separated from 2-
0
![Page 46: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/46.jpg)
A scary result: adding a tiny amount of simple molecules (CH4, N2, H2 ) to HPXe quenches both ionization and scintillation for ’s
particle: dE/dx is very high
Gotthard TPC: 4% CH4
Loss(): factor of 6
For particles, what was effect on energy resolution?
Surely small but not known, and needs investigation(~25 bars)
particlesK. N. Pushkin et al, 2004
IEEE Nuclear Science Symposium proceedings
![Page 47: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/47.jpg)
Molecular Chemistry of Xenon
• Scintillation:• Excimer formation: Xe*+ Xe Xe2* h + Xe• Recombination: Xe+ + e– Xe*
• Density-dependent processes also exist: Xe*+ Xe* Xe** Xe++ e- + heat
• Two excimers are consumed!• More likely for both high + high ionization density
– Quenching of both ionization and scintillation can occur!Xe* + M Xe + M* Xe + M + heat (similarly for Xe2*, Xe**, Xe2*+… )
Xe+ + e–(hot) + M Xe+ + e–(cold) + M*
Xe+ + e–(cold) + M + heat e–(cold) + Xe+ Xe*
![Page 48: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/48.jpg)
Energy Resolution Factors in Xenon Gas Detectors
– Intrinsic fluctuations• Fano factor (partition of energy): small for < 0.55 g/cm3
– Loss of signal (primary):• Recombination, quenching by molecular additives (heat)
– Loss of signal (secondary):• Capture by grids or electronegative impurities
– Gain process fluctuations:• Avalanche charge gain fluctuations are large
– Gain process stability:• Positive ion effects, density and mix sensitivity,...
– Long tracks extended signals • Baseline shifts, electronic non-linearities, wall effect,...
![Page 49: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/49.jpg)
Sensitivity Issues
• Target (from oscillations): m ~0.050 eV = 50 meV
– Masses could be higher… ∑m < 0.61 eV – There are ~109 relic neutrinos for each baryon
• the total mass could be ∑all visible matter
• Goal: 100’s to 1000’s kg active mass likely to be necessary
• Rejection level of internal/external backgrounds:– Less than one event per ~1027 atoms/year!
• Energy resolution needed: E/E <<10 x 10-3 FWHM, with gaussian behavior
![Page 50: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/50.jpg)
Xenon: Strong dependence of energy resolution on density!
For >0.55 g/cm3, energy resolution deteriorates rapidly
Ionization signal only
![Page 51: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/51.jpg)
TPC: Signal & Backgrounds
-HV planeReadout plane BReadout plane A
.
ionselectrons
Fiducial volume surface
Signal: event Backgrounds
*
![Page 52: Development of a High Pressure Xenon Imager with Optimal Energy Resolution Azriel Goldschmidt on work with David Nygren Instrumentation Series Seminar.](https://reader034.fdocuments.in/reader034/viewer/2022051401/56649ced5503460f949bafd1/html5/thumbnails/52.jpg)