Compounds related to 3-alkylamine-1H-indolyl acrylate and ...
Development and reproducibility of a novel high-performance liquid-chromatography monolithic column...
-
Upload
kevin-carr -
Category
Documents
-
view
212 -
download
0
Transcript of Development and reproducibility of a novel high-performance liquid-chromatography monolithic column...
![Page 1: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/1.jpg)
Research Article
Received: 27 November 2008, Revised: 10 February 2009, Accepted: 12 February 2009 Published online in Wiley Interscience: 29 April 2009
(www.interscience.wiley.com) DOI 10.1002/bmc.1231
Copyright © 2009 John Wiley & Sons, Ltd. Biomed. Chromatogr. 2009; 23: 1108–1115
11
08
John Wiley & Sons, Ltd.
Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in human urine
Kevin Carr,a Paul Whiteleya and Paul Shattocka,b
ABSTRACT: Elevated levels of trans-indolyl-3-acryloylglycine (IAcrGly) have been reported in the urine of people with variousconditions including pervasive developmental disorders (PDDs) such as autism and Asperger syndrome. Reversed-phasehigh-performance liquid chromatography with ultra-violet detection using traditional particle silica-based columns subse-quent to solid-phase extraction (SPE) has been the preferred assay method; requiring long analytical run times, high flowrates and high solvent usage. Recent developments in monolithic HPLC column technology facilitated the development of anovel analytical method, for the detection and quantification of urinary IAcrGly. The revised method eliminates the require-ment for SPE pre-treatment, reduces sample run-time and decreases solvent volumes. Five urine samples from people diag-nosed with PDD were run in quadruplicate to test the intra- and inter-day reliability of the new method based on retentiontime, peak area and peak height for IAcrGly. Detection was by UV with IAcrGly confirmation by MS/MS-MS. Relative standarddeviations showed significant improvement with the new method for all parameters. The new method represents a majoradvancement in the detection and quantification of IAcrGly by reducing time and cost of analysis whilst improving detectionlimits and reproducibility. Copyright © 2009 John Wiley & Sons, Ltd.
Keywords: high-performance liquid-chromatography; mass spectrometry; monolithic column; urine; indole; tryptophan; autism
Introduction
Trans-indolyl-3-acryloylglycine (IAcrGly) (Fig. 1) is believed to be
a metabolite of the indole amino acid tryptophan and is found
in the urine of most people at differing levels. Concentrations
can be affected not only by existing medical conditions but also
by exposure to sunlight and dietary variables (Marklova, 1999;
Whiteley et al., 1999). Abnormal levels of IAcrGly have been
reported in a number of conditions including the photoder-
matitic condition, polymorphous light eruption (Marklova et al.,1975) and Hartnup disease (Jepson, 1965). Based on differing
symptom phenotypes, various levels of IAcrGly have been
detected in the urine of people with pervasive developmental
disorder (PDD) (Shattock and Whiteley, 2002; Alcorn et al., 2004),
although the precise relationship to the syndrome is unknown.
PDD includes a number of heterogeneous conditions including
autism (Kanner, 1943) and Asperger syndrome (Asperger, 1944).
It is characterized by a triad of cognitive and behavioural impair-
ments including problems with language and reciprocal social
interaction and the presence of a restricted repertoire of activi-
ties (World Health Organization, 1992).
Analytical methods used for the detection and identification of
IAcrGly in biological fluids have principally relied on the use of
reversed-phase high-performance liquid-chromatography (RP-
HPLC) with ultra-violet (UV) detection (Mills et al., 1998; Ander-
son et al., 2002). A gradient RP-HPLC method has also been used
for the detection of small-chain polypeptides and related
material in urine from people with PDD (Solaas et al., 2002).
Whilst urine analysis by RP-HPLC is used by a number of laboratories
Figure 1. Structure of indolyl-3-acryloylglycine (IAcrGly).
* Correspondence to: K. Carr, Autism Research Unit, Department of Pharmacy,
Health & Well-being, School of Applied Sciences, University of Sunderland,
Sunderland SR1 3SD, UK. E-mail: [email protected]
a Autism Research Unit, Department of Pharmacy, Health and Well-being, Fac-
ulty of Applied Sciences, University of Sunderland, Sunderland SR1 3SD, UK
b Education and Services for People with Autism (ESPA), 2A Hylton Park Road,
Sunderland SR5 3HD, UK
Abbreviations used: IAcrGly, trans-indolyl-3-acryloylglycine; PDDs, perva-
sive developmental disorders; TFA, trifluoroacetic acid.
Contract/grant sponsor: The Robert Luff Foundation.
![Page 2: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/2.jpg)
Detection and quantification of trans-indolyl-3-acryloylglycine in human urine
Biomed. Chromatogr. 2009; 23: 1108–1115 Copyright © 2009 John Wiley & Sons, Ltd. www.interscience.wiley.com/journal/bmc
11
09
as a research tool for investigations into PDD (Whiteley and
Shattock, 2003; Bull et al., 2003), there remain several issues
regarding reproducibility, length of analysis time, high flow rates
and the volume of solvent used in the existing method.
The present method (Anderson et al., 2002) relies on the use
of solid-phase extraction (SPE) of urine samples prior to HPLC
analysis. SPE removes particulates from the sample which would
block traditional silica particle packed columns. Monolithic col-
umns, by comparison, are composed of a straight rod of highly
porous silica with a bimodal pore structure. The material has a
silica-gel skeleton containing mesopores with diameters of
approximately 13 nm and macropores with diameters of approx-
imately 2 μm. The macropores allow easy solvent flow with low
back pressure whilst the mesopores provide a large surface area
for compound retention (Novakova et al., 2004).
We report on the development of a novel gradient RP-HPLC
method using a monolithic column with direct injection for the
analysis of IAcrGly in urine samples. Intra- and inter-day reliabil-
ity for retention time, peak area and peak height is reported
based on UV detection accompanied by mass spectrometry
(MS) confirmation.
Experimental
Chemicals and Reagents
Deionized water was supplied in-house by an Elga Purelab
Option (Elga, Buckinghamshire, UK) producing water at a con-
ductivity >18 MΩ. Methanol, acetonitrile, isopropyl alcohol
(used for line cleaning), trifluoroacetic acid (TFA) and formic acid
used in this study were all HPLC grade (Sigma, UK). Thymol, used
as an anti-bacterial for urine samples, was also supplied by
Sigma.
Collection of Urine Samples
A fasting morning mid-stream urine sample was collected in a
30 mL universal tube (NELS, UK) containing a small amount of
thymol. Raw urine samples and SPE fractions were stored over-
night at −20°C.
Solid-phase Extraction Sample Preparation for Existing Method
Sample fractions were extracted by vacuum elution using an IST
VacMaster using Varian 5 mL 200 mg C18 Bond Elut LRC SPE car-
tridges (Phenomenex, UK). SPE cartridges were preconditioned
with 2 mL of methanol followed by 2 mL deionized water and
4 mL 0.1% (v/v) aqueous TFA. A 5 mL sample was vacuum
extracted and washed with 2 mL of 0.1% (v/v) aqueous TFA fol-
lowed by 1 mL of a mixture of 10% acetonitrile containing 0.1%
TFA: 90% of 0.1% (v/v) aq. TFA. Final sample collection was made
in aliquots of 1 mL of 40% acetonitrile containing 0.1% TFA: 60%
of 0.1% (v/v) aq. TFA.
HPLC Conditions
All HPLC experiments were performed using an Agilent 1100 LC
with a vacuum degasser (G1379A), a quaternary pump (G1311A),
an autosampler (G1313A), a thermostated column compartment
(G1316A) and a diode array detector (DAD) (G1315B). Analysis
was performed using Agilent Chemstation v. 10 (Agilent Tech-
nologies UK Ltd, Berkshire, UK). Mass spectrometry was carried
out on a PE Sciex API 2000 triple quadrupole with ESI TurboIon-
Spray. MS analysis was performed using Analyst v. 1.4.1 (Applied
Biosystems, Warrington, UK). Both the C18 Jupiter column and
the Onyx column had very low levels of use prior to the start of
the comparison test.
Existing reversed-phase gradient high performance liquid
chromatography. The existing method comprised RP-HPLC
carried out on an integrated Agilent 1100 HPLC as described
previously, using a C18 Jupiter (Phenomenex, UK) column
(25 cm × 4.6 mm i.d., 300 Å pore diameter, 5 μm particle diame-
ter). The column temperature was internally regulated at 27°C.
Primary detection of IAcrGly was by UV absorbance (0.1 min
peak width, 4 nm slit) at 326 nm (reference wavelengths
360 ± 100 nm). Mobile phase A was aq. 0.1% TFA and mobile
phase B was 0.1% TFA in acetonitrile. A gradient of 5–50% (v/v)
mobile phase B over 8–40 min at a flow-rate 2 mL/min was used
for analysis followed by a continued gradient of 50–80% (v/v)
mobile phase B (40–55 min) and a post-run time (5 min) back to
original gradient conditions. Sample vials (2 mL; Agilent, UK)
containing 1 mL of total SPE urine sample were injected (10 μL)
by autosampler.
Modified method reversed-phase gradient high perfor-
mance liquid chromatography. The revised method was per-
formed using the same HPLC system with a C18 Onyx monolithic
(Phenomenex, UK) column (10 cm × 3 mm i.d.). The column tem-
perature was regulated at 27°C. Detection was as previously
described. The acid modifier in the mobile phase was changed
from TFA to formic acid to aid MS detection. Mobile phase A was
aqueous 0.1% formic acid and mobile phase B was 0.1% formic
acid in acetonitrile. A gradient method was used as shown in
Table 1. Sample vials containing 1 mL of untreated urine, diluted
1:1 with deionized water (>18 MΩ) were loaded onto the system
and 10 μL of sample injected by autosampler.
Mass Spectrometer Conditions
The output solvent from the DAD was fed directly into the MS
without the need for splitting. The source block temperature
was set at 500°C in positive ion mode with a capillary voltage
5000 eV. Nitrogen was used as a nebulizer gas (45 psi), auxiliary
gas (60 psi), collision gas (7 psi), and curtain gas (50 psi). Mass
parameters for all analysis were (Q1, 100–1000 amu), product
ion (Q3, 10–250 amu), declustering potential (DP, 40 eV),
Table 1. Table showing gradient method used for Onyx
column analysis
Time
(min)
% Water + 0.1%
formic acid
% Acetonitrile + 0.1%
formic acid
Flow
(mL/min)
0.00 99.00 1.00 0.40
1.00 99.00 1.00 0.40
20.00 70.00 30.00 0.40
25.00 20.00 80.00 0.40
25.50 10.00 90.00 1.50
27.50 99.00 1.00 1.00
30.00 99.00 1.00 0.40
31.00 99.00 1.00 0.40
![Page 3: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/3.jpg)
K. Carr et al.
www.interscience.wiley.com/journal/bmc Copyright © 2009 John Wiley & Sons, Ltd. Biomed. Chromatogr. 2009; 23: 1108–1115
11
10
entrance potential (EP, 400 eV), cell exit potential (CXP, 10 eV),
collision energy (CE, 45 eV) and collision gas (CG, 3 psi).
Procedure
Because of the photo-instability of indolyl-3-acryloylglycine all
experiments were performed in a laboratory with no natural
light and reduced levels of artificial light. All urine preparation
was performed as quickly as possible with samples being kept in
the dark wherever possible. Sample vials for the HPLC were
made of brown glass.
Five urine samples, from people diagnosed with PDD, were
randomly selected from a sample bank. A proportion of each
sample was prepared by SPE. A separate portion was allowed to
settle and 1 mL of the top fraction was removed via pipette and
diluted 1:1 with deionized water. The samples treated by SPE
were analysed by HPLC with both the Jupiter and Onyx columns,
based on the relevant methods described. The diluted untreated
urine samples were analysed using the Onyx column only. The
five samples were each run in quadruplicate and RSD% calcu-
lated from the results (Tables 2 and 3). The same five samples
(both SPE and untreated) were run on three separate days on
the onyx column; the SPE samples were run for just one day on
the Jupiter column. Both the inter-day and intra-day RSD% were
calculated. Confirmation of the IAcrGly peak was made by MS
(positive parent ion) and MS-MS (relevant daughter ions).
To determine linearity a set of IAcrGly standards (2, 3, 4, 5, 6, 7,
8, 9 and 10 μg/mL) were prepared in deionized water and analysed
with the Onyx column and associated method. The integrated
area under the curve results was plotted. The concentrations of
IAcrGly used for the test were chosen because they represent
the levels of IAcrGly routinely found in urine by our laboratory.
To calculate recovery, the same IAcrGly concentrations used in
the linearity test were prepared in aliquots of a urine sample and
analysed as described previously. After subtracting the amount
of IAcrGly originally contained in the urine, the results were used
to plot a graph of IAcrGly detected against IAcrGly added.
The protocols for the collection and analysis of urine samples
were carried out in accordance with the Declaration of Helsinki
and have been ratified by the University of Sunderland. Parents
Table 2. Table showing inter-day RSD% for Onyx and Jupiter columns for SPE treated and diluted untreated urine samples
SPE samples Diluted neat urine samples
Onyx column RSD% Jupiter column Onyx column RSD%
Day 1 Day 2 Day 3 Day 1 Day 1 Day 2 Day 3
Sample 1Time 0.09 0.05 0.01 0.13 0.11 0.05 0.12
Area 1.67 3.04 2.67 26.58 4.82 0.91 3.15
Height 1.78 3.32 3.22 23.25 3.71 0.3 2.77
% Total area 1.76 2.25 2.65 22.01 2.64 8.52 2.3
% Total height 1.22 1.27 3.92 15.61 4.52 4.76 0.61
Sample 2Time 0.04 0.05 0.03 0.05 0.39 0.04 0.09
Area 0.68 1.89 6.86 25.71 4.23 0.32 3.44
Height 1.9 1.4 7.07 22.08 3.79 0.25 2.42
% Total area 1.39 1.31 3.81 23.96 3.02 0.63 10.63
% Total height 0.15 0.72 1.32 12.57 1.49 3.79 5.51
Sample 3Time 0.09 0.06 0.01 0.02 0.1 0.07 0.13
Area 2.94 3.19 2.33 18.23 3.42 0.07 1.84
Height 3.75 3.77 3.19 19.03 2.97 1.12 0.9
% Total area 4.97 5.06 2.68 23.96 2.66 0.81 5.69
% Total height 3.57 3.46 2.66 23.96 2.07 1.06 1.02
Sample 4Time 0.07 0.02 0.04 0.02 0.14 0.06 0.12
Area 1.65 1.62 5.15 28.77 4.11 0.28 3.68
Height 1.86 1.95 5.33 25.35 4.26 0.11 3.38
% Total area 1 1.11 1.1 23.96 2.81 2.52 2.11
% Total height 1.83 1.58 1.46 23.96 0.52 0.94 1.94
Sample 5Time 0.02 0.05 0.05 0.04 0.03 0.08 0.06
Area 1.16 1.22 1.5 23.18 5.11 3.17 3.54
height 0.8 2.44 1.3 21.99 3.09 1.88 3.41
Height 2.28 4.79 5.8 23.96 2.26 2.15 0.97
% Total height 3.02 4.39 0.81 23.96 3.62 3.42 1.26
![Page 4: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/4.jpg)
Detection and quantification of trans-indolyl-3-acryloylglycine in human urine
Biomed. Chromatogr. 2009; 23: 1108–1115 Copyright © 2009 John Wiley & Sons, Ltd. www.interscience.wiley.com/journal/bmc
11
11
of all participants provided consent for the supply of urine sam-
ples as stipulated by the Ethics Committee of the University of
Sunderland.
Results
Limit of detection (LOD) and limit of quantification (LOQ) were
assessed using IAcrGly standards produced in house, as per
Anderson et al. (2002). Standards were dissolved in deionized
water at concentrations of between 50 ng/mL and 2.5 mg/mL
and analysed as per the Onyx procedure (2.4.2) with MS detec-
tion. HPLC gave one major peak at 17.55 ± 0.1 min which corre-
sponded to a peak on the mass spectrometer of 245.10 amu at
an m/z of 17.71 ± 0.1 min (Fig. 2); this matches with an M + 1 of
IAcrGly (244.1 amu). Analysis of results from the standards of
IAcrGly gave an LOD (S/N = 3) of 75 ng/mL and an LOQ (S/
N = 10) of 250 ng/mL, with respect to peak area by diode array
detection.
The analysis was also performed by MS/MS of the main peak
at 245.1 amu. This produced two daughter ions at 170.2 and
142.1 amu and corresponded to expected results, as shown in
Fig. 3.
To ensure assay precision, within-day repeatability (n = 4) and
between-day repeatability (n = 3) were assessed for the 10 sam-
ples (five SPE samples and five untreated samples) using the
Onyx monolithic column. The results are summarized in Tables 2
and 3. Intra-day RSDs ranged from 0.01 to 0.39% for retention
time and from 0.07 to 6.86% for peak area. Inter-day repeatabil-
ity of the SPE samples gave an RSD for retention time of 0.06–
0.15% and an RSD of 6.51–7.95% for peak area. The inter-day
repeatability of the untreated samples gave an RSD for retention
time of 0.09–0.15% and an RSD of 5.46–7.92% for peak area. RSD
was also calculated for peak height and percentage IAcrGly as a
ratio of total chromatogram peak area and height (Tables 2 and
3). Confirmation of IAcrGly was performed by MS/MS with the
same results as per standards.
The five SPE samples were also analysed with the Jupiter
5 μm C18 packed column. This produced a within-day RSD for
retention time of between 0.02 and 0.13% and an RSD of
18.23–28.77% for peak area. The results are summarized in
Table 2. Analysis with the Jupiter column was not performed
between days.
The proposed analytical method was validated with regard to
linearity using a set of nine IAcrGly standards (2–10 μg/mL). The
Table 3. Table showing intra-day RSD% for Onyx column for SPE treated and diluted
untreated urine samples
SPE samples RSD% Diluted neat urine samples RSD%
Sample 1Time 0.15 0.13
Area 6.51 7.92
Height 6.29 6.52
% Total area 7.13 9.15
% Total height 7.29 9.26
Sample 2Time 0.06 0.09
Area 7.54 6.48
Height 6.94 5.34
% Total area 7.29 7.34
% Total height 7.54 6.49
Sample 3Time 0.12 0.15
Area 7.95 5.46
height 6.49 4.69
% Total area 5.96 7.26
% Total height 5.67 3.06
Sample 4Time 0.09 0.1
Area 6.74 5.69
Height 6.59 6.25
% Total area 7.29 7.98
% Total height 7.46 5.45
Sample 5Time 0.06 0.09
Area 6.54 6.94
Height 6.98 5.49
% Total area 7.24 7.59
% Total height 7.29 6.52
![Page 5: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/5.jpg)
K. Carr et al.
www.interscience.wiley.com/journal/bmc Copyright © 2009 John Wiley & Sons, Ltd. Biomed. Chromatogr. 2009; 23: 1108–1115
11
12
calibration curve was constructed and fitted to a linear equa-
tion. Each point of the calibration curve is the average of three
peak–area measurements. The method showed satisfactory lin-
earity with a determination coefficient of 0.9979 (Fig. 5).
The linear equation constructed from the standards was used
to determine the amount of IAcrGly in relation to the HPLC inte-
grated area under the curve. To calculate the percentage recov-
ery, the same set of nine IAcrGly standards (2–10 μg/mL) were
prepared in aliquots of a urine sample and refrigerated (4°C) for
48 h. These were then analysed by the same method as the stan-
dards. The amount of IAcrGly originally contained in the urine
sample was subtracted from all the results. Using the linear
equation produced from the standards, the amount of IAcrGly
detected was determined for each of the IAcrGly concentrations.
These results were plotted against the concentration of IAcrGly
added.
The graph produced a slope of 0.9950, demonstrating good
correlation between the amount of IAcrGly added and the
amount detected, with an error of less than 1%. Linearity was
good with a determination coefficient of 0.9983 (Fig. 6).
Discussion
Results from our experiments showed that the Onyx column and
revised method performed significantly better than the existing
method (Anderson et al., 2002). RSD for peak area and height
showed on average an improvement by a factor of 10 with the
new method compared with the existing one. Other advantages
include the elimination of the SPE clean-up procedure, reduc–
tion in HPLC analysis time and reduced solvent usage. The new
method can also be applied directly to mass spectrometry fol-
lowing changes to the acid modifier from TFA to formic acid.
Total analysis time has been reduced from 60 to 31 min with the
IAcrGly peak appearing at 17.5 min for the new method com-
pared with 21 min for the existing method (Fig. 4).
Results showed excellent linearity between the concentrations
of 2 and 10 μg/mL for the new Onyx method and a recovery rate
in excess of 99%. Although this method was devised purely for
the analysis of IAcrGly, removal of the SPE clean-up allows for
the analysis of other compounds which may have been lost
by the SPE method and possibly retained by the Jupiter column.
Figure 2. Plot of MS Total Ion Count (TIC) from IAcrGly standard.
![Page 6: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/6.jpg)
Detection and quantification of trans-indolyl-3-acryloylglycine in human urine
Biomed. Chromatogr. 2009; 23: 1108–1115 Copyright © 2009 John Wiley & Sons, Ltd. www.interscience.wiley.com/journal/bmc
11
13
Whilst the new HPLC method shows excellent levels of detec-
tion, quantification of IAcrGly in human urine relating to actual
levels of IAcrGly present in the body is still problematic. Initial
studies of creatinine, the normal reference to determine urine
concentration, not being a reliable standard for quantification
of urinary compounds from samples from people with PDD
(Whiteley et al., 2006), require further confirmation. Studies by our
laboratory are on-going to confirm if creatinine is a viable refer-
ence or if other compounds may produce more accurate results.
It is intended that this new method of HPLC detection and
quantification of IAcrGly will be used in studies to determine lev-
els of IAcrGly in people displaying differing characteristics of
PDD alongside the potential effects that diet, sunlight and
accompanying disease states may have on its production.
Figure 3. MS/MS plot of IAcrGly parent ion (245.1 amu).
Figure 4. Plots showing urine analysis via Jupiter column and Onyx column at 326 nm (IAcrGly highlighted).
![Page 7: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/7.jpg)
K. Carr et al.
www.interscience.wiley.com/journal/bmc Copyright © 2009 John Wiley & Sons, Ltd. Biomed. Chromatogr. 2009; 23: 1108–1115
11
14
Figure 5. Graph plotting IAcrGly standards analysed by HPLC using Onyx column.
Figure 6. Graph plotting amount of IAcrGly detected against amount of IAcrGly added.
![Page 8: Development and reproducibility of a novel high-performance liquid-chromatography monolithic column method for the detection and quantification of trans-indolyl-3-acryloylglycine in](https://reader031.fdocuments.in/reader031/viewer/2022020508/575002f91a28ab114896983c/html5/thumbnails/8.jpg)
Detection and quantification of trans-indolyl-3-acryloylglycine in human urine
Biomed. Chromatogr. 2009; 23: 1108–1115 Copyright © 2009 John Wiley & Sons, Ltd. www.interscience.wiley.com/journal/bmc
11
15
The new method represents a major advancement in the
detection and quantification of IAcrGly by reducing the time
and cost of analysis whilst improving detection limits and
reproducibility.
Acknowledgements
The authors gratefully acknowledge the financial support pro-
vided by The Robert Luff Foundation.
References
Alcorn A, Berney T, Bretherton K, Mills M, Savery D and Shattock P.Urinary compounds in autism. Journal of Intellect Disability Research2004; 48: 274–278.
Anderson RJ, Bendell DJ, Garnett I, Groundwater PW, Lough WJ, Mills MJ,Savery D and Shattock PE. Identification of indolyl-3-acryloylglycine inthe urine of people with autism. Journal of Pharmacy andPharmacology 2002; 54: 295–298.
Asperger H. Die autistischen Psychopathen. Kindesalter Archiv fürPsychiatrie und Nervenkrankheiten 1944; 117: 76–136.
Bull G, Shattock P, Whiteley P, Anderson R, Groundwater PW, Lough WJand Lees G. Indolyl-3-acryloylglycine (IAG) is a putative diagnosticurinary marker for autism spectrum disorders. Medical ScienceMonitoring 2003; 9: CR2–CR5.
Jepson JB. In The Metabolic Basis of Inherited Disease, 2nd edn, StanburyJB, Wyngaarden JB and Frederickson DS (eds). McGraw-Hill: New York,1965; 1283–1299.
Kanner L. Autistic disturbances of affective contact. Nervous Children1943; 2: 217–250.
Marklova E. Where does indolylacrylic acid come from? Amino Acids1999; 17: 401–413.
Marklova E, Malina L and Hais IM. Urinary excretion of indolyl-3-acryloylglycine in some skin affections. Clinica Chimica Acta 1975; 64:273–280.
Mills MJ, Savery D and Shattock PEG. Rapid analysis of low levels ofindolyl-3-acryloylglycine in human urine by high-performance liquidchromatography. Journal of Chromatography B 1998; 712: 51–58.
Novakova L, Matysova L, Solichova M, Koupparis MA and Soliich P.Comparison of performance of C18 monolithic rod columns andconventional C18 particle-packed columns in liquid chromatographicdetermination of Estrogel and Ketoprofen gel. Journal ofChromatography B 2004; 813: 191–197.
Shattock P and Whiteley P. Biochemical aspects in autism spectrumdisorders: updating the opioid-excess theory and presenting newopportunities for biomedical intervention. Expert Opinion onTherapeutic Targets 2002; 6: 175–183.
Solaas KM, Skjeldal O, Gardner MLG, Kase BF and Reichelt KL. Urinarypeptides in Rett syndrome. Autism 2002; 6: 315–329.
Whiteley P, Rodgers J, Savery D and Shattock P. A gluten-free diet as anintervention for autism and associated spectrum disorders:preliminary findings. Autism 1999; 3: 45–65.
Whiteley P and Shattock P. What makes trans-indolyl-3-acrloylglycine(IAcrGly) identified by high-performance liquid-chromatography(HPLC) relevant to pervasive developmental disorders? Journal ofNutrition and Environmental Medicine 2003; 13: 1–7.
Whiteley P, Waring R, Williams L, Klovrza L, Nolan F, Smith S, Farrow M,Dodou K, Lough WJ and Shattock P. Spot urinary creatinine excretionin pervasive developmental disorders. Pediatrics International 2006;48: 292–297.
World Health Organization. Tenth Revision of the InternationalClassification of Diseases and Related Health Problems. ClinicalDescriptions and Diagnostic Guidelines. WHO: Geneva, 1992.