Detective-20 0 TM - AMETEK ORTEC · 2020. 8. 5. · 2 Detective-200 TM • Greatly increased...

12
Detective-200 Ruggedized, Ultra-High-Sensitivity, Transportable HPGe Radionuclide Identification System TM “No other radiation instrument has more standoff nuclide identification performance in a small rugged package.”

Transcript of Detective-20 0 TM - AMETEK ORTEC · 2020. 8. 5. · 2 Detective-200 TM • Greatly increased...

  • Detective-200Ruggedized, Ultra-High-Sensitivity, Transportable

    HPGe Radionuclide Identification System

    TM

    “No other radiation instrument has more standoff nuclideidentification performance in a small rugged package.”

  • 2

    Detective-200 TM

    • Greatly increased sensitivity, (3x more sensitive thanDetective and Micro-Detective models).

    • Much faster to identify.• More rugged, compact, IP67 standard stackable

    package.• “Gold Standard” identification performance.• Detective-Remote™ software package; easy

    integration of multiple Detective-200s into highlysensitive systems, for use in a variety of securitysituations:

    • On the ocean (maritime interdiction at sea)• On land (vehicle mounted search)• In the air (aerial search)• Static applications on highways, public events,

    incident triage• The high flexibility of the many possible configurations

    allows matched response to changing security needs.• Wireless Mobile MCB Server software.

    IntroductionThe Detective-200 is a member of the ORTEC Detective productfamily, whose members represent the “Gold Standard” of gamma-ray-emitting radionuclide identifiers, and are deployed to prevent theillicit trafficking of nuclear materials by security authorities globally.With the Detective-200, the range of applicability of theseinstruments is greatly broadened by increased sensitivity andruggedness. Key features and benefits include: • Large Area HPGe Detector (~5675 sq mm) gives definitive

    identification of illicit nuclear materials (SNM) in seconds.• 10 times faster to identify HEU than a "conventional" NaI based

    identifier.• Built-in removable 120° collimator reduces background

    interference in field of view.• IP67 waterproof, dustproof, and shock resistant packaging. It

    actually floats in water.

    Detective-200 in Wide Area Search applications(Land and Air)In these non-maritime applications, Detective-200s are easilyinstalled into unmodified vehicles, cars, vans, and aircraft for mobiledetection applications at higher speeds and larger distances.Additional Detective-200s are easily added into the system forincreased sensitivity. The data can be combined by the Detective-Remote software utility. Multiple Detective-200s are easily stackedvertically.In another class of application the instrument system is stationaryand the potential source is moving, such as traffic choke pointmonitors (toll booths, bridges, tunnels), monitoring entry to a publicevent, monitoring road traffic or ships passing beneath a bridgestructure. The measurement is essentially the same.The configurability of the Detective-200 solution in both mobile andstatic applications is a practical benefit which cannot be over-emphasized as requirementsconstantly change.

    https://www.ortec-online.com/products/nuclear-security-and-safeguards/hand-held-radioisotope-identifiers-riids/detective-100thttps://www.ortec-online.com/products/nuclear-security-and-safeguards/hand-held-radioisotope-identifiers-riids/micro-detectivehttps://www.ortec-online.com/products/nuclear-security-and-safeguards/mobile-detection-systems/detective-remote-softwarehttps://www.ortec-online.com/products/nuclear-security-and-safeguards/mobile-detection-systems/detective-remote-softwarehttps://www.ortec-online.com/products/nuclear-security-and-safeguards/mobile-detection-systems/detective-remote-software

  • Detective-200 in Maritime applications• “Slow Pass” Materials interdictionDetective-200 is very well suited for mounting on naval vessels. TheIP67 packaging is immune to salt spray. The tough fold-flat handlesmake it easy to lift and mount aboard a vessel, or transport fromvessel to vessel. A ship-borne Detective-200 system consisting of one or moreinstruments, controlled through wired or wireless networking viaDetective-Remote, is capable of reliably detecting illicit materials ona nearby vessel in a slow “sail by” measurement. Figure 1 shows anexample of sensitivity for unshielded HEU at representative speedsand distances for multiple units of Detective-200s.

    Choke Point MonitoringBecause of the lack of 100% coverage with fixed traffic portalmonitors, it is desirable to be able to quickly setup and configuretemporary or "ad-hoc" portal systems capable of monitoring fornuclear devices, dirty bombs, or contaminated materials in pass-through traffic. One or more Detective-200s can be placedinconspicuously at a roadside, pedestrian entryway, vendor servicesentrance, warehouse dock, or parking garage. The Detective-200internal battery power and “Auto Identification” reporting providessimple deployment and operation.Figure 2 provides estimated performance of multiple units of land-based Detective-200s at 10 MPH and distances, up to 40 meterswith the same alarm conditions as figure 1.

    3

    Figure 1 Marine Bare HEU detection limit (kg) >95% detection possibility, falsealarm rate 95% detection possibility,false alarm rate

  • Standoff DetectionStandoff detection is the problem of correctly identifying a distantstationary or near stationary radioactive source. The standoffdetection limit is essentially a question of signal-to-noise ratio in thedetection system. As distance increases, gamma rays areattenuated by air and a fixed field of view means at further distancesmore background is in the field of view of the source. Both of theseaffects reduce the signal-to-noise ratio. As with other applications inlow resolution systems, background variation is problematic. Thehigh resolution of the Detective-200 and its large area detectormake it an ideal standoff tool either alone or in an array. Figure 3shows the time to identify 2.5 kg of Bare HEU with a >95%probability of detection and

  • Detective-200 Hardware FeaturesSince the first model was introduced in 2004, all ORTEC Detectiveidentifiers have incorporated the same principle technologies andfeatures, namely:• High-resolution, high-purity germanium (HPGe) detector.• Mechanically cooled by miniature Stirling-cycle refrigerator. • Internal high range GM tube provides gamma dose rate.• Fast, Simple and ULTRA-Reliable Classification of NORM,

    Medical, Industrial, SNM and Natural Isotopes, shielded andunshielded.

    • Touch Screen or Remote Software Operation.• ANSI N42.34 and ANSI N42.42 compliant.• Battery/AC/DC powered, highly stable digital electronics.• Wired USB, 802.11 wireless communications.• Local storage of spectra.• Advanced analysis algorithms developed over thousands of real-

    world identification situations:• Low levels of false alarms in the presence of NORM• High-fidelity identification of mixture constituents• Resistance to “spoofing” by masking sources• Immunity to variation in background radiation

    • Easy-to-use through intuitive, touch-sensitive screen GUI withoperational modes configurable according to individual CONOPS,including long count mode for weak/distant sources.

    • Neutron detecting models.

    Detective-200 Operational CapabilitiesSEARCH MODES: Gamma countrates are presented as a timetracking strip chart. MONITOR MODE: Detective collectsone spectrum per second and runsthe ID algorithm against an eightsecond sliding average. Moresensitive to sources which moverelative to the instrument. Monitormode is a valuable search method,but is also useful in ad-hoc portalmonitoring applications.SNM Search Mode: SNM Searchmode is designed to help avoid falsenegatives when determining SNM. Ithelps in finding the point of maximumcount rate which COULD beconsistent with SNM.Key regions of the spectrum aremonitored which are critical to thedetermination of both U-235 (the keyconstituent of HEU) and Pu-239. Thepeak region confidence level isdisplayed in the form of a bar graph.A high and steady reading indicatesthat "something" is present which isworthy of more investigation. Oncethe maximum reading has beenlocated, the "confirm" key initiates thefull identification algorithm.SNM Search mode is an INDICATORof SNM but should always befollowed by the confirmatory ID toavoid false positives. In combination,SNM Search and ID modes minimizeBOTH false negatives and falsepositives.GAMMA DOSE RATE: Gamma Dose Rate is monitored by theHPGe detector and by an internal compensated GM tube. The doserate is displayed at all times. Dose rate units may be chosen asµSv/hr or mR/hr.

    5

    Detective-200 TM

  • 6

    Detective-200 TM

    GPS Position Information: An internal GPS receiver displays GPScoordinates which may be saved along with spectrum data for futureuse.Storage of Data (spectrum, search data, ID results): To internalRAM and removable SD card.Computer Interfacing: USB connection to laptop. Spectral transferby Microsoft® ActiveSync. Remote control via Microsoft“remotedsp.exe” (supplied). WiFi (802.11) communication. WirelessMobile MCB Server software.Display: Features a large, bright andclear LCD display with touch-sensitive screen. Gamma count rateand gamma dose rate are displayedcontinuously both numerically and inbar graph form. Menu navigation ishighly intuitive. The radionuclidegamma-ray spectrum may bedisplayed and manipulated (e.g.,vertical scale, zoom) like aconventional multichannel analyzer.Y-axis units are now displayed.

    Operating ModesIn response to customer requests, Detective instruments can nowoperate in two modes “Classify” and “ANSI”. Classify Mode: The user presses the Identify button and theinstrument gathers data until manually stopped, without preset.During the acquisition, suspectednuclide classification messagesappear, such as “Suspect Industrial1”, meaning the presence of oneindustrial nuclide is suspected. As thecount continues and confidencelevels increase, this might change to“Found Industrial 1” or disappear asbetter statistics determine thepreviously suspected nuclide wasnot, in fact, present. Clicking on the“Found” or “Suspect” message givesa listing of which nuclides wereactually found (or suspected) by name.

    ANSI Mode: This mode is similar toclassify mode, but dispenses with theclassification, and presents nuclidenames directly, both suspect andfound.Preset Time: This is to allow forCONOPS in which it is required tocount for a preset time, for example60 seconds. At the end of the presetperiod, only what has been found isreported, no suspects are reported.The operator can request a counttime extension, if desired, adding multiples of the original presetperiod. LCX (Low-Confidence Expert) ID Mode: LCX mode is passwordprotected. LCX denotes “Low Confidence-Expert,” and is intendedfor expert users. This identificationmode displays suspected threatalarms and identifications at a lowerconfidence level than the normalmode. This results in more hits onsuspected threat nuclides.Instrument Calibration: The instrument is calibrated prior toshipment from the factory. The energy calibration may be checkedand adjusted with any known source with a clean gamma raybetween 0 and 3 MeV. A higher energy is recommended. Cs-137 isoften used. Calibration can be manual or automatic. Backgroundcollection is now a required part of calibration. By allowing foractivities already in the background, the system will no longer reportnuclides detected in the background. These IDs on former versionssometimes lead to user-confusion. The background must beupdated on a schedule which is chosen by the privileged user.SMART Stabilizer: The “smart stabilizer” stabilizes the gain veryprecisely on the 1460 keV peak of K-40, if present. The smart part isthat if there is no K-40 present or if Eu-152 is detected, which couldinterfere with the K-40 peak, the stabilizer setting is held but notadjusted until “normal” conditions return. Even though the Detectiveis a highly stable instrument, the smart stabilizer allows accuratedetermination of more complicated mixed spectra.

  • 7

    Classify Mode MessagesThe following explains the criteria for selected Classify ID DisplayMode messages.HEU (highly enriched uranium): This message is displayed if themajor lines of uranium are detected and the ratios of the intensitiesof the lines indicates the U-235 content to be above about 70%.Am241 (unshielded) in the “Industrial” category: This message isdisplayed if the 59 keV peak is located. It could mean that anAm-241 source such as a smoke detector is present. Move closerto the source and/or count longer. This will allow the higher-energygamma rays to accumulate in the spectrum, in case plutonium isalso present.Unknown Peak and Unknown/Beta Emitter: This indicates thegamma count rate is higher than can be accounted for based on thepeaks in the library. The implication is that either an unexpectednuclide or a beta emitter is present (beta emitters typicallyproducing counts over a broad range of energies). Move closer tothe source and count longer to determine the nature of the suspectitem. If another ID is found, then the Unknown Peak or theUnknown/Beta Emitter ID are suppressed.“Found Nuclide” Screen MessagesRDD Detected: This message is posted when estimated activity is>100 mCi, whether the activity is from threat or innocent nuclides.The gamma count-rate and dose-rate meters on the Survey Modeand ID Mode screens display a flashing red background andextremely high count and dose rates.Classify Mode Primary ID Message FormatThe form of the primary ID messages is:

    "Found CLASS(#)" or "Suspect CLASS(#)"where “CLASS” is

    MedicalIndustrialNORMBremsstrahlungOtherNuclear UraniumNuclear PlutoniumNuclear Neptunium

    And "#" is the number of nuclides of that class identified.

    Detective Library V8.5 radionuclides according to their categories in the“Classify” ID mode.

    Detective-200 TM

    MedicalF-18I-125Lu-177Lu-177mMo-99Pd-103Se-75Sm-153Sm-153 (shielded)Tc-99mXe-133Ac-225As-74Ce-139Ce-141Co-58Ga-67Ga-67 (shielded)Ge-68/Ga-68I-123I-123 (shielded)I-124I-131I-131 (shielded)In-111Lu-172Na-24Rb-83Rb-86Ru-106/Rh-106Sc-46Sr-82/Rb-82Sr-89Tl-201Tl-204Tm-170V-48Xe-133mYb-169Zn-62Zn-65Zr-95

    IndustrialAm-241Am-241 (unshielded)Cs-137Ho-166mHo-166m (shielded)Ir-192Ir-192 (shielded)W-187Ac-227Ag-110mAr-41As-72Au-198Ba-133Ba-140

    Be-7Bi-212 (Th-232/U-232daughter)Br-77Ca-47Cd-115Ce-144Cm-242Cm-243Cm-244Co-55Co-57Co-57 (shielded)Co-60Cs-134Hf-181Hg-203I-126I-126 (shielded)I-132I-133I-134I-135Kr-87Kr-88Kr-88 (shielded)Mn-52Mn-56Nb-92mNb-94Nb-95Nb-96Nb-96 (shielded)Nd-147Pa-231Pb-203Pr-144Ra-223Rh-105Ru-103Ru-97Sb-124Sb-124 (shielded)Sb-125Sb-127Sr-85/Kr-85Tc-96Te-132Th-229Th-230Tm-171W-188/Re-188

    NORMLa-138Ra-226Bi-214 (Ra-226 daughter)K-40Lu-176Th-232

    BremsstrahlungBeta emitter

    OtherCr-51Cu-64Eu-152Gd-159La-140Mn-54Neutrons on FeNeutrons on HydrogenUnknown PeakUnknown/Beta emitterXe-131mAt-211Bi-207Br-76Br-76 (heavily shielded)Br-76 (shielded)Cd-109Co-56Co-56 (shielded)Eu-154Eu-155Eu-156Fe-59Ga-64Ga-64 (shielded)Gd-153Ho-166Ir-194 (shielded)Na-22NeutronsOs-194/Ir-194Po-210Sn-113Ta-182Tl-200Tl-202Xe-135Y-88

    Nuclear UraniumEnriched UraniumHEUU-232U-233U-235U-238186 keV peak present2614 keV peak present

    Nuclear PlutoniumPu-239Pu-238375/414 peak presentAm-241 (shielded)

    Nuclear NeptuniumNp-237

  • 8

    Detective-200 TM

    ANSI Mode MessagesThe table is divided according to the threat category used to determine ID background color in Monitor Mode and on the Found and SuspectNuclide reports, e.g., green for innocent IDs, yellow for LCX-mode suspects, and red for threats. NB: if desired and under password protection,the color coding, and therefore the threat classification can be disabled.

    ANSI Mode Messages Table

    Identification Classification Identification Classification Identification Classification Identification ClassificationInnocent

    Ac-225 .....................................MedicalAc-227 .....................................IndustrialAg-110m...................................IndustrialAm-241 (unshielded) ...............IndustrialAr-41 ........................................IndustrialAs-72 .......................................IndustrialAs-74 .......................................MedicalAt-211.......................................MedicalAu-198 .....................................MedicalBa-133 .....................................IndustrialBa-140 .....................................IndustrialBe-7 .........................................IndustrialBeta emitter .............................IndustrialBi-207 ......................................IndustrialBi-212 (Th-232/U-232 daughter) .................................IndustrialBi-214 (Ra-226 daughter) .................................NORMBr-76 ........................................IndustrialBr-76 (heavily shielded) ...........IndustrialBr-76 (shielded) .......................IndustrialBr-77 ........................................IndustrialCa-47 .......................................IndustrialCd-109 .....................................IndustrialCd-115 .....................................IndustrialCe-139 .....................................MedicalCe-141 .....................................MedicalCe-144 .....................................IndustrialCm-242 ....................................IndustrialCm-243 ....................................IndustrialCm-244 ....................................IndustrialCo-55 .......................................IndustrialCo-56 .......................................IndustrialCo-56 (shielded) ......................IndustrialCo-57 .......................................IndustrialCo-57 (shielded) ......................IndustrialCo-58 .......................................MedicalCo-60 .......................................IndustrialCr-51........................................MedicalCs-131 .....................................MedicalCs-134 .....................................IndustrialCs-137 .....................................IndustrialCu-64 .......................................MedicalCu-67/Ga-67 ............................MedicalEu-152 .....................................MedicalEu-154 .....................................MedicalEu-155 .....................................Industrial

    Eu-156 .....................................IndustrialF-18 .........................................MedicalFe-59 .......................................MedicalGa-64.......................................IndustrialGa-64 (shielded) ......................IndustrialGa-67.......................................MedicalGa-67 (shielded) ......................MedicalGd-153.....................................MedicalGd-159.....................................IndustrialGd-159.....................................IndustrialGe-68/Ga-68............................MedicalHf-181 ......................................IndustrialHg-203 .....................................IndustrialHo-166 .....................................IndustrialHo-166m ..................................IndustrialHo-166m (shielded) .................IndustrialI-123.........................................MedicalI-123 (shielded)........................MedicalI-124.........................................MedicalI-125.........................................MedicalI-126.........................................IndustrialI-126 (shielded)........................IndustrialI-131.........................................MedicalI-131 (shielded)........................MedicalI-132.........................................IndustrialI-133.........................................IndustrialI-134.........................................IndustrialI-135.........................................IndustrialIn-111 .......................................MedicalIr-192 .......................................IndustrialIr-192 (shielded).......................IndustrialIr-194 (shielded).......................IndustrialK-40 .........................................NORMKr-87 ........................................IndustrialKr-87 ........................................IndustrialKr-88 ........................................IndustrialKr-88 (shielded) .......................IndustrialLa-138......................................NORMLa-140......................................IndustrialLu-172......................................IndustrialLu-176......................................NORMLu-177......................................MedicalLu-177m...................................MedicalMn-52.......................................IndustrialMn-54.......................................IndustrialMn-56.......................................IndustrialMo-99.......................................MedicalNa-22 .......................................Medical

    Na-24 .......................................MedicalNb-92m ....................................IndustrialNb-94 .......................................IndustrialNb-95 .......................................IndustrialNb-96 .......................................IndustrialNb-96 (shielded) ......................IndustrialNd-147 .....................................IndustrialOs-194/Ir-194...........................IndustrialPa-231 .....................................IndustrialPb-203 .....................................IndustrialPd-103 .....................................MedicalPo-210 .....................................IndustrialPr-144 ......................................IndustrialRa-223 .....................................IndustrialRa-226 .....................................NORMRb-83 .......................................MedicalRb-86 .......................................MedicalRh-105 .....................................IndustrialRu-103 .....................................IndustrialRu-106/Rh-106 ........................MedicalRu-97 .......................................IndustrialSb-124 .....................................IndustrialSb-124 (shielded) ....................IndustrialSb-125 .....................................IndustrialSb-127 .....................................IndustrialSc-46 .......................................MedicalSe-75 .......................................MedicalSm-153 ....................................MedicalSm-153 (shielded) ...................MedicalSn-113......................................IndustrialSr-82/Rb-82 .............................MedicalSr-85/Kr-85 ..............................IndustrialSr-89 ........................................MedicalTa-182......................................IndustrialTc-96 ........................................IndustrialTc-99m .....................................MedicalTe-132......................................IndustrialTh-229 .....................................IndustrialTh-230 .....................................IndustrialTh-232 .....................................NORMTl-200.......................................IndustrialTl-201.......................................MedicalTl-202.......................................MedicalTl-204.......................................IndustrialTm-170 ....................................MedicalTm-171 ....................................IndustrialUnknown Peak.........................UnknownUnknown/Beta emitter .............Unknown

    V-48 .........................................MedicalW-187 ......................................IndustrialW-188/Re-188..........................IndustrialXe-127 .....................................IndustrialXe-131m ..................................MedicalXe-133 .....................................MedicalXe-133m ..................................MedicalXe-135 .....................................IndustrialY-88..........................................MedicalY-91..........................................IndustrialYb-169 .....................................MedicalZn-62 .......................................MedicalZn-65 .......................................MedicalZr-95 ........................................Medical

    Suspect (LCX Mode only)186 keV Peak Present .............SNM2614 keV peak present............—375/414 Peak Present .............SNM375/414 Peak Present .............SNM

    ThreatAm-241 ....................................Weapon .................................................IndicatingAm-241 (shielded) ...................Weapon .................................................IndicatingEnriched Uranium ....................SNMHEU .........................................SNMNeutrons ..................................Weapon .................................................IndicatingNeutrons CR ............................Weapon .................................................IndicatingNeutrons on Fe ........................Weapon .................................................IndicatingNeutrons on Hydrogen.............Weapon .................................................IndicatingNp-237 .....................................SNMPu-238 .....................................SNMPu-239 .....................................SNMU-232 .......................................Weapon .................................................IndicatingU-233 .......................................SNMU-235 .......................................SNMU-238 .......................................Weapon .................................................Indicating

  • 9

    OPERATION MODESSEARCH Scanning mode for location of radioactive sources.Speed settings 0.1 to 50 seconds/point.SNM Search Mode™ Nuclide-specific search mode for U-235,Pu-239 and neutron counts. Ba-133 surrogate detection may beturned on for training purposes. Bar graph display of nuclideconfidence level. Aid to Identify mode.MONITOR Mode The instrument collects one spectrum persecond and runs the ID algorithm against an 8 second slidingaverage. This mode is more sensitive to sources which moverelative to the instrument.LCX “Low confidence Expert” Mode.IDENTIFY Gamma Proprietary scheme for identification andclassification of radionuclides. Background subtraction.ANSI Mode: See nuclide list above.Classify Mode:

    Nuclides classified according to:IndustrialMedicalNatural (NORM)Nuclear

    These classifications are based on an internal, fixed libraryaccording to ANSI N42.34. Customized libraries for specificapplications can be supplied by special order.Remote Mode Detective V3 instruments can participate as nodeswithin the Detective Mobile search system.Dose Rate Visual over range indication and continuous audiblealarm, user settable. Over-ride alarm at dose rates>10,000 µSv/hr.

    DETECTORSInternal HPGe DetectorCrystal Nominal Dimensions: 85 mm diameter x 30 mm deep.P-type high-purity germanium. Coaxial construction.Cooler: High reliability, low power Stirling cooler. Cooler design life>5 years continuous running. Dual piston design, 1 W nominal liftat 100 K.Digital Noise Suppression: "LFR Filter," ORTEC Patent Pending.

    Gamma Dose Rate Detector Two detectors determine thegamma dose rate over a wide range from 10000 µSv/h, a dose-rate range of around six decades. For lowdose rates, below ~20 µSv/h, the dose rate is determined from theGe detector spectrum. For dose rates above this value, the internalcompensated GM tube is used. Instrument switches between thetwo automatically.Dose rate uncertainty 10,000 µSv/h (fixed maximum threshold),user settable threshold below this.

    DIGITAL MCA AND DATA PROCESSORDisplay VGA 640 x 480 TFT sunlight readable, touchscreen,operate with finger or stylus.Data Processor Marvel 806 MHz XScale.Data Storage (Spectrum, Search Data, ID Results) To internalRAM and removable SD card.Computer Interfacing USB connection to laptop. Spectraltransfer by Microsoft® ActiveSync. Remote control via Microsoft“remotedsp.exe” (supplied). Wi-fi (802.11) communicationsoftware. Wireless Mobile MCB Server software.GPS Internal NMEA compliant WAAS capable.Digital MCA with Internal Storage of Multiple Spectral Data.“Smart” digital spectrum gain stabilizer.Digital Noise Suppression LFR Filter.Conversion Gain 8k channel.Maximum Number of Stored Spectra Unlimited on removablemedia.

    DISPLAYS AND MENUSMain ScreenGamma Count Rate Bar Graph 20 kcps full scale. Dose Rate Bar Graph 10 mSv/hr full scale, flashes on overrange.Status Lines:WARNING!! High Dose Rate — Displayed when Dose rateexceeds 10 mSv/hr.Detector is Warm — Displayed when crystal temperature is aboveworking limit.Bias Supply Error — Displayed if any power supply is bad.WARNING!! Low Battery.

    Detective-200 Detailed Specifications Summary

    Detective-200 TM

  • Search Mode (Gamma/Neutron) Dwell times 0.1–50 secondsper point. Over-range warning.SNM Search Mode™ Nuclide-specific search mode for U-235and Pu-239. Bar graph display of nuclide confidence level. Identify Nuclide ID and classification."Intense" shows the most intense lines list, which is acontinuously updating list of the 12 best peaks currently detected.The nuclides and energies are based on the internal nuclidelibrary. The rank is based on the confidence value for the peak."Save" Saves the spectrum. Format choices: ORTEC “.CHN”,ORTEC “.SPC” and ANSI N42.42."Display" brings up the spectral display. The spectrum may bemanipulated via the arrow keys and various accelerator keys forcursor movement. Energy and channel contents are displayed withthe spectrum.Advanced Setup Password protected.Calibration Check Manual or Automatic Calibration Check.Automatic may be triggered by interval or time of day. Instrument issupplied calibrated from factory.View Data Acquisition Parameters Reports instrument status.

    PHYSICAL SPECIFICATIONSMaximum Overall Dimensions (including Ge detector end capand shock absorbers) 42.72 cm L x 24.13 cm W x 38.7 cm H(16.82" L x 9.5" W x 15.24" H)Weight 47 lb (21.32 kg).Internal Battery Life Approximately 3 hours at 25°C when HPGedetector is cold, depending on battery condition. Battery lifetimemay be extended indefinitely by the use of external battery packswhich are available in "battery belt" formats.Input Power 10 to 17 V DC from battery or DC power supply(universal mains supply included). Battery charger circuit is insideinstrument.

    Power Usage Greatest during cool down:

  • 11

    Detective-200 TM

    Detective-200 Mobile Detection System Rack KitThe rack kit provides a simple, transportable solution for creating “Ad-Hoc” mobile detection systems without the need for permanent vehicleinstallation.

    • Supports up to four Detective-200s in varying configurations.• Portable, light weight, and collapsible.• Easy to install and remove.• Rugged frame for mobile applications.• Can be configured for multiple vehicle types.• Transportable case for rack and two Detective-200s.• Vehicle laptop mounting kit.

  • Specifications subject to change052820

    Detective-200 TM

    Ordering Information

    DETDX-200Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifier (Gamma Only) with AC/DC power adapter charger, automobilepower cable, external battery kit, wheeled transport case, and MAESTRO-PRO software.DETDX-200-2Includes 2 each DETDX-200 Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifiers (Gamma Only) with AC/DCpower adapter charger, automobile power cable, external battery kit, and wheeled transport case, and 1 each Detective-Remote software withLaptop computer, external GPS, and MAESTRO-PRO software.DETDX-200-4Includes 4 each DETDX-200 Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifiers (Gamma Only) with AC/DCpower adapter charger, automobile power cable, external battery kit, and wheeled transport case, and 1 each Detective-Remote software withLaptop computer, external GPS, and MAESTRO-PRO software.DETEX-200Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifier (Gamma and Neutron) with AC/DC power adapter charger,automobile power cable, external battery kit, wheeled transport case, and MAESTRO-PRO software.DETEX-200-2Includes 2 each DETEX-200 Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifiers (Gamma and Neutron) withAC/DC power adapter charger, automobile power cable, external battery kit, and wheeled transport case, and 1 each Detective-Remotesoftware with Laptop computer, external GPS, and MAESTRO-PRO software.DETEX-200-4Includes 4 each DETEX-200 Ultra-High-Sensitivity, Ruggedized, Transportable HPGe Radioisotope Identifiers (Gamma and Neutron) withAC/DC power adapter charger, automobile power cable, external battery kit, and wheeled transport case, and 1 each Detective-Remotesoftware with Laptop computer, external GPS, and MAESTRO-PRO software.DETECTIVE-REMOTE-MOB-SYSDetective-Remote software with Detective software update, Laptop computer, external GPS, and MAESTRO-PRO software.

    For price and delivery, email [email protected].