Deriving Nonlinearity Coefficients from Changing Background

27
Deriving Nonlinearity Coefficients from Changing Background Mark Esplin, Deron Scott, Bryce Walker, Ben Esplin Utah State University Space Dynamics Laboratory 1

Transcript of Deriving Nonlinearity Coefficients from Changing Background

Page 1: Deriving Nonlinearity Coefficients from Changing Background

Deriving Nonlinearity Coefficients from Changing Background

Mark Esplin, Deron Scott, Bryce Walker, Ben EsplinUtah State University Space Dynamics Laboratory

1

Page 2: Deriving Nonlinearity Coefficients from Changing Background

Contentsā€¢ Quadric Nonlinearityā€¢ Nonlinearity Model and Sensitivity Studiesā€¢ Application to CrIS Sensorā€¢ Check Method with Stepped Blackbody Dataā€¢ Conclusions

2

Page 3: Deriving Nonlinearity Coefficients from Changing Background

Quadratic Nonlinearity Equations

3

An interferometer with quadratic nonlinearity can be expressed as:š¼š¼ā€² = (š¼š¼ + š‘‰š‘‰) āˆ’ š‘Žš‘Ž2 š¼š¼ + š‘‰š‘‰ 2

where: I = original undistorted interferogramIā€™ = interferogram distorted with quadratic nonlinearityV = offset voltagea2 = quadratic nonlinearity coefficient

āˆ’š‘Žš‘Ž2(š‘†š‘† āˆ— š‘†š‘†) + 1 āˆ’ 2š‘Žš‘Ž2š‘‰š‘‰ š‘†š‘†

)Iā€² = āˆ’š‘Žš‘Ž2š¼š¼2 + 1 āˆ’ 2š‘Žš‘Ž2š‘‰š‘‰ š¼š¼ + (š‘‰š‘‰ āˆ’ š‘Žš‘Ž2š‘‰š‘‰2Rearranging the terms yields:

This term is just a constant and can be removed before taking the Fourier transform. After taking the Fourier transform the result is:

where: S is the Fourier transform of I, and convolution is indicated by āˆ—.If the spectrum is band limited, the first term can be ignored, resulting in:

1 āˆ’ 2š‘Žš‘Ž2š‘‰š‘‰ š‘†š‘†

This is the desired spectrum, S, multiplied by a gain term

Page 4: Deriving Nonlinearity Coefficients from Changing Background

0

40

80

120

0 40 80 120 160

40

80

120

80 120 160

Graphical Representation of Nonlinearity

V -- Voltage Offseta2 -- Quadratic Nonlinearity CoefficientI -- Linear Input InterferogramIā€™ -- Nonlinear Interferogram

š¼š¼ā€² = š¼š¼ + š‘‰š‘‰ āˆ’ š‘Žš‘Ž2 š¼š¼ + š‘‰š‘‰ 2

ā€¢ CrISā€™s photons to voltage relationship is nonlinearā€¢ The magnitude of the nonlinearity has been

exaggerated for clarity

Iā€™

I

V

Page 5: Deriving Nonlinearity Coefficients from Changing Background

Nonlinear Spectrum

ā€¢ When S is band limited as shown, the (š‘†š‘† āˆ— š‘†š‘†) component does not overlap spectrally and can readily be removed

ā€¢ The nonlinearity corrected spectra is then:

5

š‘†š‘†š‘†š‘† āˆ— š‘†š‘†

Sā€™ = āˆ’š‘Žš‘Ž2(š‘†š‘† āˆ— š‘†š‘†) + 1 āˆ’ 2š‘Žš‘Ž2š‘‰š‘‰ š‘†š‘†

V -- Voltage OffsetS -- Correct SpectrumSā€™ -- Nonlinear Spectruma2 -- Quadratic Nonlinearity Coefficient

S = š‘†š‘†ā€²/ 1 āˆ’ 2š‘Žš‘Ž2š‘‰š‘‰

Nonlinear Spectrum

Nor

mal

ized

Resp

onse

Nyquist Frequency

Page 6: Deriving Nonlinearity Coefficients from Changing Background

Standard Nonlinearity Correction Methodā€¢ Measure source with known output radiances

ā€“ Need at least two different radiance levelsā€¢ Derive nonlinearity coefficients that yield known radiancesā€¢ Advantage:

ā€“ Simple intuitive methodā€¢ Disadvantage:

ā€“ Radiance of source needs to be known very accurately at different radiance levels

6

Page 7: Deriving Nonlinearity Coefficients from Changing Background

Alternative Nonlinearity Methodā€¢ Observe external source (ECT) held at a constant output radiance

ā€“ Although required to be stable, the actual value of radiance is unimportant

ā€¢ Make measurements with different sensor temperaturesā€¢ Internal calibration target (ICT) temperature floats with sensor

temperatureā€¢ Reprocess the data with different nonlinearity coefficientsā€¢ Optimize nonlinearity coefficient by minimizing spread in calculated

radiancesā€¢ Why it works:

ā€“ Varying background sensor temperature moves up and down the system gain curve

ā€“ Nonlinearity errors cause calibrated radiance to change even though the radiance from source has not changed

7

Page 8: Deriving Nonlinearity Coefficients from Changing Background

Sensor Simulation Modelā€¢ Built model to study sensitivities of the methodā€¢ Derived nonlinearity a2 coefficients from simulated dataā€¢ Model included:

ā€“ Internal calibration with Deep space (DS) and Internal Calibration Target (ICT)

ā€“ ICT temperature errorsā€“ Time dependent temperature changes for ICT and ECTā€“ Warm instrument background effectsā€“ Noise

ā€¢ Results show low sensitivity to all parameters except for Earth Calibration Target (ECT) radiance changes correlated with sensor/ICT variations

ā€¢ ECT radiance can vary as long as it is not correlated with sensor temperature

8

Page 9: Deriving Nonlinearity Coefficients from Changing Background

Nonlinearity Correction Method Applied to CrIS

ā€¢ Cross-Track Infrared Sounder (CrIS)ā€¢ The CrIS sensor is an infrared

Fourier transform spectrometerā€¢ One CrIS instrument is currently

flying on the Suomi National Polar-orbiting Partnership (NPP) spacecraft

ā€¢ Another CrIS (J1) is being preparing for launch in 2017

http://www.jpss.noaa.gov/images/media-gallery/gallery-jpss1_1.jpg

Page 10: Deriving Nonlinearity Coefficients from Changing Background

ā€¢ 8-second scans ā€¢ 30 Earth locations

ā€“ 9 FOVs per locationā€“ 3 spectral bands per FOV

ā€¢ LWIR 650-1095 cm-1, resolution: 0.625 cm-1

ā€¢ MWIR 1210-1750 cm-1, resolution: 1.25 cm-1

ā€¢ SWIR 2155-2550 cm-1, resolution: 2.5 cm-1

ā€¢ Two calibration views per scanā€“ Internal calibration target (ICT) ā€” Warm (ambient temperature)ā€“ Deep space (DS) ā€” Coldā€“ Separate calibration for two interferometer scan directions

ā€¢ Two redundant electronic sides

CrIS Sensor Includes Self Contained Radiance Calibration

30 Earth Scenes Sampling 9 FOVs

DS(Cold)

ICT(Warm)

8 sec

Page 11: Deriving Nonlinearity Coefficients from Changing Background

Nonlinearity Correction Method Applied to CrISā€¢ CrIS model J1 has completed thermal vacuum (TVAC) testingā€¢ Long Term Repeatability (LTR) data taken during TVACā€¢ Provides opportunity to verify method with real sensor dataā€¢ LTR data consists of 18 collects each 2 hours long taken over a

period of a month ā€¢ The LTR gives a metric for the stability of the sensorā€¢ A small LTR means a stable sensorā€¢ LTR is defined as:

ā€“ Average together the 2 hours of spectra in each setā€“ Convert to a percentage of a 287 K blackbodyā€“ Find the standard deviation at each wavenumber

ā€¢ CrIS J1 meets spec with large margin even without a nonlinearity correction

11

Page 12: Deriving Nonlinearity Coefficients from Changing Background

Time History of LWIR LTR Data

ā€¢ Temperatures for several sensor elements shown on leftā€“ SSM (scene select mirror), ICT (internal calibration target), OMA

(interferometer or opto-mechanical assembly) ā€¢ BT error shown on right

ā€“ Measured brightness temperature minus 287 Kā€¢ BT error tracks sensor temperature changes

12

Sensor Temperatures BT - 287 K

Page 13: Deriving Nonlinearity Coefficients from Changing Background

Derive Nonlinearity a2s from LTR dataā€¢ Vary a2s to minimize LTR using an iterative technique

ā€“ Start with an initial set of a2sā€“ Calculate the LTRā€“ Vary a2s until minimum LTR is found

ā€¢ Derivation of a2s uses no explicit temperature knowledgeā€“ ICT and DS temperatures used by Science Data Record (SDR)

algorithmā€¢ Only consider more central regions of each band

ā€“ LWIR 680 ā€“ 1020 cm-1, MWIR 1220 ā€“ 1600 cm-1, SWIR 2160 ā€“ 2400 cm-1

ā€¢ LTR histories created by concatenated measurements into single time series

ā€¢ Analysis performed for both electronic sides

13

Page 14: Deriving Nonlinearity Coefficients from Changing Background

LWIR LTR Before and After Nonlinearity Optimization

ā€¢ LW FOV5 most nonlinear and shows the largest LTRā€¢ After optimization of a2s, LTR is much lower ā€¢ Excellent agreement between FOVs

14

No Nonlinearity Correction Optimized a2s

Page 15: Deriving Nonlinearity Coefficients from Changing Background

Magnitude of LTR with Different a2s

ā€¢ LTR calculated with a factor times preliminary Exelis a2sā€¢ Calculated at a number of points then spline interpolatedā€¢ Solved for factor that minimizes LTR for each FOV

15

Page 16: Deriving Nonlinearity Coefficients from Changing Background

MWIR LTR Before and After Optimization

ā€¢ Only MW FOV9 has significant nonlinearityā€¢ Only FOV9 a2 optimized, all other MW a2s set to zero

16

Page 17: Deriving Nonlinearity Coefficients from Changing Background

MWIR LTR

ā€¢ Only MW FOV9 a2 varied, other a2s set to zeroā€¢ MWIR side 1 data

17

Page 18: Deriving Nonlinearity Coefficients from Changing Background

LTR Before Nonlinearity Optimization

ā€¢ LW FOV5 and MW FOV9 brightness temperature differences on order of 50 mK

ā€¢ Radiance jump at center of test apparent with nonlinear FOVs

18

LWIR MWIR

Page 19: Deriving Nonlinearity Coefficients from Changing Background

LTR After Nonlinearity Optimization

ā€¢ Excellent agreement between LWIR and MWIR time historiesā€¢ ECT gradients rows FOVs 1-2-3 to 7-8-9 clearly visibleā€¢ Remaining structure probably ECT temperature instability

ā€“ Equal magnitude for all bands

19

LWIR MWIR

Page 20: Deriving Nonlinearity Coefficients from Changing Background

LTR After Nonlinearity Optimization

ā€¢ Excellent agreement between LWIR and SWIRā€¢ SWIR has higher noise ā€¢ No nonlinearity correction for SWIR

20

LWIR SWIR

Page 21: Deriving Nonlinearity Coefficients from Changing Background

Derived a2s

ā€¢ Side1: a2s derived from side 1 LTR dataā€¢ Side2: a2s derived from side 2 LTR dataā€¢ Results consistent with a2s generated with conventional method

21

LW 1 LW 2 LW 3 LW 4 LW 5 LW 6 LW 7 LW 8 LW 9 MW 9Preliminary 0.0116 0.0128 0.012 0.0119 0.0386 0.0085 0.0093 0.0121 0.0064 0.0513Flight 0.0141 0.0145 0.0152 0.0159 0.0355 0.011 0.0095 0.0157 0.0085 0.0507S1 0.0131 0.0147 0.0127 0.018 0.0359 0.0113 0.0127 0.0169 0.0086 0.0446S2 0.0139 0.0137 0.0128 0.0158 0.0324 0.0094 0.0076 0.0107 0.0005 0.0321

0

0.01

0.02

0.03

0.04

0.05

0.06

COEF

FICI

ENT

VALU

E

FOV #

a2 Comparisons

Page 22: Deriving Nonlinearity Coefficients from Changing Background

Check of LTR Derived Nonlinearity a2sā€¢ Conventional method of determining nonlinearity uses the ECT

stepped through a range of temperaturesā€¢ Plot stepped blackbody brightness temperatures similar to LTR

time histories ā€¢ Combine data into a time series subtract ECT set point

temperatureā€¢ Again integrate over spectral content

ā€“ LWIR 680 ā€“ 1020 cm-1

ā€“ MWIR 1220 ā€“ 1600 cm-1

ā€“ SWIR 2160 ā€“ 2400 cm-1

ā€¢ No correction for ECT gradients or temperature errorsā€¢ Side 1 a2s consistent with stepped blackbody data

22

Page 23: Deriving Nonlinearity Coefficients from Changing Background

Stepped Blackbody No Nonlinearity Correction

ā€¢ No nonlinearity correctionā€¢ Nonlinearity of LW FOV5 and MW FOV9 clearly visibleā€¢ Mission Nominal Temperature (MN) stepped blackbody, side 1

23

LWIR

200K233K

260K

278K

299K

310K

200K 233K 260K 278K

299K 310K

MWIR

Page 24: Deriving Nonlinearity Coefficients from Changing Background

With Nonlinearity Correction

ā€¢ Note the scale change from the preceding slideā€¢ Good agreement between LW and MWā€¢ a2s derived from side 1 LTR

24

LWIR MWIR

Page 25: Deriving Nonlinearity Coefficients from Changing Background

MWIR Compared to SWIR

ā€¢ Good agreement with SWIRā€¢ High noise for cold temperatures with MWIR and SWIR

25

SWIRMWIR

Page 26: Deriving Nonlinearity Coefficients from Changing Background

Discussionā€¢ Method works well even though sensor temperature difference was

only on the order of 1.5 Kā€¢ Larger change in sensor temperature expected to produce more

accurate a2sā€¢ Variations in observed brightness temperature used to derive a2s

smaller than ECT gradientsā€¢ With quadratic nonlinearity large temperature differences are not

neededā€¢ However, side 2 LTR has less temperature variation and derived

a2s not as consistent with stepped blackbodyā€¢ Method relies on relative differences not absolute differencesā€¢ Method can still work with ECT temperature variation if they are

not correlated with the sensor temperature variations

26

Page 27: Deriving Nonlinearity Coefficients from Changing Background

Conclusionā€¢ a2 nonlinearity coefficients can be derived using alternate method

ā€“ Potential on-orbit application with uniform temperature sceneā€¢ Derived a2s consistent with stepped blackbodyā€¢ Side 1 LTR data showed larger brightness temperature differences

and resulted in better a2sā€¢ Potentially higher accuracy since method can partially compensates

for inaccuracies of other coefficients and modelā€¢ Probably only practical with system like CrIS with self contained

calibration

27