Department of Radiation Oncology The University of Michigan
Transcript of Department of Radiation Oncology The University of Michigan
![Page 1: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/1.jpg)
Department of Radiation OncologyUniversity of Michigan
Department of Radiation OncologyDepartment of Radiation OncologyUniversity of MichiganUniversity of Michigan
TheThe
Video and Surface PositioningVideo and Surface Positioning
Scott W. Hadley PhDScott W. Hadley PhD
I for one welcome our AI SUV overlords
![Page 2: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/2.jpg)
AAPM Summer School 2006
ContributorsContributors
• Scott Hadley University of Michigan
• Scott Johnson Varian Medical Systems
• Shidong Li Henry Ford Hospital
• Marco Riboldi TMB Lab @ Politecnico di Milano
• Christoph Bert GSI & MGH
![Page 3: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/3.jpg)
AAPM Summer School 2006
Goal of Patient PositioningGoal of Patient Positioning
• Reproduce patient position as seen in CT simulation– Aligning Skin Marks with Laser
– Matching Boney Anatomy with MV
– Target Positioning with US, markers, MV, kV, CBCT
![Page 4: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/4.jpg)
AAPM Summer School 2006
Goal of Patient PositioningGoal of Patient Positioning
• Frequency of Position Verification
– First Day and weekly
– Adaptive protocol, image for several days, move to average then weekly
– Daily positioning or target localization
![Page 5: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/5.jpg)
AAPM Summer School 2006
Advantages of Video & SurfaceAdvantages of Video & Surface
• No ionizing radiation– Daily use with no extra radiation
– Operators can be at couch side for positioning
• Video cameras see what we see
• Can be "no touch" and "no marks"
![Page 6: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/6.jpg)
AAPM Summer School 2006
Advantages of Video & SurfaceAdvantages of Video & Surface
• Quantitative 3D couch positioning, 6 parameter rigid transformation
• Intrafraction patient monitoring
• 4D acquisition
• Non-rigid positioning possible
![Page 7: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/7.jpg)
AAPM Summer School 2006
Disadvantages of Video & SurfaceDisadvantages of Video & Surface
• Positioning based on skin surface only– Skin surface as a surrogate for target
anatomy
• Can be obstructed by clothes, mask, other objects
![Page 8: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/8.jpg)
AAPM Summer School 2006
History of Video TechniquesHistory of Video Techniques• 1972 Kelsey, C.A., R.G. Lane, and W.G. Connor, Measurement
of patient movement during radiation therapy. Radiology. 103(3): p. 697-8.
• 1975 Connor, W.G., et al., Patient repositioning and motion detection using a video cancellation system. Int J Radiat OncolBiol Phys. 1(1-2): p. 147-53
• 1977 Renner, W.D., et al., The use of photogrammetry in tissue compensator design. Part II: experimental verification of compensator design. Radiology,
• 1982 Andrew, J.W. and J.E. Aldrich, A microcomputer-based system for radiotherapy beam compensator design and patient contour plotting. Med Phys. 9(2): p. 279-83
![Page 9: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/9.jpg)
AAPM Summer School 2006
Three Technologies Come TogetherThree Technologies Come Together
• Photogrammetric Calibration
• 3D Image Registration
• Cheap and good Digital Image Acquisition and Fast Computers
![Page 10: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/10.jpg)
AAPM Summer School 2006
Image Guided SurgeryImage Guided Surgery
Grimson, W.E.L., et al., An automatic registration method for frameless stereotaxy, image guided surgery, and enhanced reality visualization. IEEE Transactions on Medical Imaging, 1996. 15(2): p. 129-140
![Page 11: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/11.jpg)
AAPM Summer School 2006
In Room (Spy) CamerasIn Room (Spy) Cameras
IsocenterIsocenter
AP cameraAP camera Left lateral cameraLeft lateral camera
Right lateral cameraRight lateral camera
Sagittal cameraSagittal camera
![Page 12: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/12.jpg)
AAPM Summer School 2006
Live Video SubtractionLive Video Subtraction
![Page 13: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/13.jpg)
AAPM Summer School 2006
Volume Rendered Ref’ ImagesVolume Rendered Ref’ Images
Volume Rendered Reference image from CT
Live Video of Patient on Table
Video Subtraction for positioning
![Page 14: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/14.jpg)
AAPM Summer School 2006
Patient Surface from CT ScansPatient Surface from CT Scans
Camera 1 Camera 2Camera 1 & 2
![Page 15: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/15.jpg)
AAPM Summer School 2006
3D Surface to 2D Image Matching3D Surface to 2D Image Matching
![Page 16: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/16.jpg)
AAPM Summer School 2006
Latest TechnologyLatest Technology
• 3D Surface Scanners– Video + Laser, interfermometry, light pattern
– Large FOV
– Multiple Frames per second
• Surface Registration software– 6 parameter rigid transformation
– Direct measurement of positioning error
![Page 17: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/17.jpg)
AAPM Summer School 2006
![Page 18: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/18.jpg)
AAPM Summer School 2006
![Page 19: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/19.jpg)
AAPM Summer School 2006
![Page 20: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/20.jpg)
AAPM Summer School 2006
![Page 21: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/21.jpg)
AAPM Summer School 2006
![Page 22: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/22.jpg)
AAPM Summer School 2006
![Page 23: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/23.jpg)
Surface imaging at MGH
3D surface imaging for PBI setup
Christoph Bert1/2, Katie Metheany2,Karen Doppke2/3, Alphonse Taghian2/3,
Simon N. Powell2/3, George T.Y. Chen2/3
1Gesellschaft für Schwerionenforschung, Darmstadt, Germany2Massachusetts General Hospital, Boston, Ma
3Harvard Medical School, Cambridge, Ma
![Page 24: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/24.jpg)
Surface imaging at MGH
Shape conservation CT-stereovision• “Patient”
contour TPS– Binary cube
fill →marching cubes →decimation
• Any surface image
• Surface registration →distance projection →0.65mm RMS
+2-2mm
![Page 25: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/25.jpg)
Surface imaging at MGH
Comparison to CT surface• Purpose: Check if CT, linac, and 3D
camera coordinate systems match• Methods:
– CT scan phantom, extract surface, DRRs– a) line up according to 3D surface ⇒
b) compare portfilm with DRR (lat.+a.p.)– b) ⇒ a)
![Page 26: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/26.jpg)
Surface imaging at MGH
Surface data + Setup
3D surfaces and originMarkerblock
![Page 27: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/27.jpg)
Surface imaging at MGH
Film samples
a.p. lateral
![Page 28: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/28.jpg)
Surface imaging at MGH
Combined results - alignment
![Page 29: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/29.jpg)
Surface imaging at MGH
Stability test - results
![Page 30: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/30.jpg)
AAPM Summer School 2006
![Page 31: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/31.jpg)
AAPM Summer School 2006
![Page 32: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/32.jpg)
AAPM Summer School 2006
![Page 33: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/33.jpg)
AAPM Summer School 2006
![Page 34: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/34.jpg)
AAPM Summer School 2006
![Page 35: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/35.jpg)
AAPM Summer School 2006
![Page 36: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/36.jpg)
AAPM Summer School 2006
![Page 37: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/37.jpg)
AAPM Summer School 2006
![Page 38: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/38.jpg)
AAPM Summer School 2006
![Page 39: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/39.jpg)
AAPM Summer School 2006
![Page 40: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/40.jpg)
AAPM Summer School 2006
![Page 41: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/41.jpg)
AAPM Summer School 2006
![Page 42: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/42.jpg)
AAPM Summer School 2006
ConclusionsConclusions
• What role is it going to play– Laser replacement?
– Daily Targeting of underlying target?
– Monitor patient during treatment?
– Tracking and Gating?
If You're Considering Video/SurfaceIf You're Considering Video/Surface
![Page 43: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/43.jpg)
AAPM Summer School 2006
ConclusionsConclusions
• What Body Site– Head / Intracranial works well.
– PBI near surface works well.
– PBI near chest wall X-ray chest wall.
– Prostate not so well but may be an aid to the therapists.
If You're Considering Video/SurfaceIf You're Considering Video/Surface
![Page 44: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/44.jpg)
AAPM Summer School 2006
ConclusionsConclusions
• Questions to Ask When Purchasing– How many systems are needed?
– What reference surface can be used?
– Does it work with your planning system?
– How does it calibrate and know isocenter?
– What type of R&V system is included?
If You're Considering Video/SurfaceIf You're Considering Video/Surface
![Page 45: Department of Radiation Oncology The University of Michigan](https://reader036.fdocuments.in/reader036/viewer/2022071600/613d1a97736caf36b7595b12/html5/thumbnails/45.jpg)
AAPM Summer School 2006
ConclusionsConclusions
• Installation, Commissioning and QA– Manufacture’s CAP
– Commissioning• Known phantom and site specific
– QA• Daily isocenter check
• Longer term stability
If You're Considering Video/SurfaceIf You're Considering Video/Surface