DEGASSED CATION CONDUCTIVITY DG 105...2017/06/16 · DESCRIPTION The cation conductivity degasser...
Transcript of DEGASSED CATION CONDUCTIVITY DG 105...2017/06/16 · DESCRIPTION The cation conductivity degasser...
DESCRIPTION
The cation conductivity degasser DG-105 is a device designed to obtain the value of the conductivity of a sample, eliminating the contribution to it of possible gases that may be dissolved. After subjecting the sample to boiling, releasing any gas can be dissolved, a cooling is performed at 25-30 ° C, and then the treated sample to a conductivity cell is introduced.
Degassed cationic conductivity is an important parameter for the boiler, most boiler manufacturers have specified that the value of the cation conductivity in the steam must be below a certain limit of conductivity (less than 0.5 microsecond / cm), to avoid damage to it or the turbine.
APPLICATIONS
• Steam-water analysis systems.
FEATURES
• According with ASTM D4519
• Including all necessary components as a complete unit, just piping, plug-in and ready to use.
• Provide with necessary shell protection to prevent operators from hot injury.
• Compact wall mount design.
• Sample cooler ensures CC and DCC probes receive identical sample temperatures, eliminating errors (Optional).
• Flow indication.
• Switch protection No Flow
• CE Marking
DEGASSED CATION CONDUCTIVITY DG 105. SH.1DEGASSED CATION CONDUCTIVITY DG 105. SH.1. DG105r00.2017
DEGASSED CATION CONDUCTIVITY DG 105
www.analyzersmanvia.comPoligono Riaño III, Parcela 43 33920 Riaño, Asturias, SpainPhone:+34 984204549. Email: [email protected]
Set pointA.- ManualB.- Automatic (atmospheric pressure conpensation)
Sample Cooling0.- Self cooled1.- Cooling water
Electrical supplyC.- 120 VAC 60 HzD.- 230 VAC 50 Hz
ORDERING INFORMATION DG-105 -- -- --
DEGASSED CATION CONDUCTIVITY DG 105
SPECIFICATIONS
Max Pressure/Temp Rating:
Sample Flow:
Sample Connection:
Cation Resin:
Cooler Cooling Water Connect. (Only Opt. 1)
Cooling Water Consumption (Only Opt. 1):
Input Power:
Power Consumption:
Plate mounting :
Enclosure Rating:
Dimensions (HxWxD)
4 bar(g) / 50ºC
80 - 250 ml /min.
1/4 ´´ OD inlet / outlet.
600 cc
1/4 ´´ NPT(f)
4 l/min
110-120/220-240 VAC, 50/60 Hz.
1.6 KVA
Stainless steel
IP66 / NEMA 4X
900 x 450 x 300 mm
SPARE PARTSPart #
H1500-120
H-1500-230
RD-105
STH-TXIH-2
MV-PT105
SC-1
Name
Heating Cartidge 120 VAC
Heating Cartidge 230 VAC
Solid State relay 110-230 VAC 50/60 Hz
Turbine
Temperature sensor
Sample cooler
DEGASSED CATION CONDUCTIVITY DG 105. SH.2DEGASSED CATION CONDUCTIVITY DG 105. SH.2 . DG105r00.2017
www.analyzersmanvia.comPoligono Riaño III, Parcela 43 33920 Riaño, Asturias, SpainPhone:+34 984204549. Email: [email protected]
DEGASSED CATION CONDUCTIVITY DG 105
DEGASSED CATION CONDUCTIVITY DG 105. SH.3. DG105r00.2017
MAIN ELEMENTS
1. - Mounting Plate2. -Electrical box3.- Sample Cooler
4.- Turbine5.- Heater & Thermocouple
DRAWINGSin mm
Self Cooled Option Cooling Water Option
*Only in Cooling Water Option