Transcript of dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1...

Page 1: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

dds digital data systems (pty) ltd

Page 2: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Table Of Contents: Page No: 1. INTRODUCTION....................................................................................... 1

1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ......................... 1 1.2 WHY USE A DDS BOMB CALORIMETER SYSTEM? ..................................... 1 1.3 HISTORY........................................................................................... 2

2. THE CAL2K PRODUCT RANGE................................................................... 3 3. THE CAL2K .............................................................................................. 3

3.1.1 SYSTEM FEATURES ...................................................................... 3 3.1.2 SOFTWARE FEATURES .................................................................. 4 3.1.3 ENHANCED FEATURES .................................................................. 4

3.2 THE CALORIMETER ............................................................................. 4 3.3 THE VESSEL ...................................................................................... 4 3.4 THE FILLING STATION ........................................................................ 4 3.5 THE COOLER...................................................................................... 4 3.6 SYSTEM COMPONENT GUIDE................................................................ 5 3.7 OPTIONAL EXTRAS ............................................................................. 5 3.8 ITEMS TO SOURCE LOCALLY ................................................................ 6 3.9 TECHNICAL SPECIFICATIONS............................................................... 6 3.10 SYSTEM CONFIGURATION.................................................................... 7

3.10.1 MINIMUM SYSTEM FOR MANUAL MASS ENTRY ................................. 7 3.10.2 STANDARD SYSTEM FOR AUTOMATED MASS ENTRY.......................... 7 3.10.3 SYSTEM WITH DOCKING STATION ................................................. 7

4. THE E2K .................................................................................................. 8 4.1 SYSTEM FEATURES............................................................................. 8 4.2 SOFTWARE FEATURES......................................................................... 8 4.3 ENHANCED FEATURES......................................................................... 9 4.4 THE E2K CALORIMETER....................................................................... 9 4.5 TECHNICAL SPECIFICATIONS............................................................... 9 4.6 OPTIONAL EXTRAS ........................................................................... 10 4.7 ITEMS TO SOURCE LOCALLY .............................................................. 10 4.8 SYSTEM OPTIONS............................................................................. 11

5. THE ECO ................................................................................................ 12 5.1 SYSTEM FEATURES........................................................................... 12 5.2 THE ECO CALORIMETER .................................................................... 13 5.3 SYSTEM COMPONENT GUIDE.............................................................. 13 5.4 TECHNICAL SPECIFICATIONS............................................................. 14 5.5 OPTIONAL EXTRAS ........................................................................... 14 5.6 ITEMS TO SOURCE LOCALLY .............................................................. 14

6. OPTIONAL EXTRAS................................................................................ 15 6.1 DOCKING STATION........................................................................... 15

6.1.1 OPERATIONS WITHOUT THE DOCKING STATION ............................ 15 6.1.2 OPERATIONS WITH THE DOCKING STATION.................................. 15 6.1.3 COMPLEX BARCODE FEATURES.................................................... 15 6.1.4 SYSTEM SET UP......................................................................... 16 6.1.5 ADVANTAGES OF VESSEL MASS OPERATION ................................. 16

6.2 E2K AIR COOLER.............................................................................. 16 6.3 HIGH PRESSURE OXYGEN REGULATOR ................................................ 17

7. CONSUMABLES ...................................................................................... 17 7.1 CAL2K INSTALLATION KITS ............................................................... 23

7.1.1 CAL2K-1-KT – CALORIMETER INSTALLATION KIT ........................... 23 7.1.2 CAL2K-2-KT – COOLER INSTALLATION KIT.................................... 23 7.1.3 CAL2K-3-KT – FILLING STATION INSTALLATION KIT....................... 23 7.1.4 CAL2K-4-KT – VESSEL INSTALLATION KIT..................................... 23

7.2 E2K INSTALLATION KITS ................................................................... 24 7.2.1 E2K-KT – CALORIMETER INSTALLATION KIT .................................. 24 7.2.2 CAL2K-2-KT – COOLER INSTALLATION KIT.................................... 24

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com

Page 3: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com

7.2.3 CAL2K-3-KT –FILLING STATION INSTALLATION KIT........................ 24 7.2.4 CAL2K-4-KT –VESSEL INSTALLATION KIT...................................... 24 7.2.5 E2K-ACC-KT –ACCESSORIES INSTALLATION KIT............................ 25

7.3 ECO INSTALLATION KITS................................................................... 25 7.3.1 CAL2K-E-KT – CALORIMETER INSTALLATION KIT ........................... 25 7.3.2 CAL2K-3-KT –FILLING STATION INSTALLATION KIT........................ 25 7.3.3 CAL2K-4-KT –VESSEL INSTALLATION KIT...................................... 25 7.3.4 ECO-ACC-KT –ACCESSORIES INSTALLATION KIT ........................... 26 7.3.5 ECO FREE DEMO KIT .................................................................. 26

8. COMPARING BOMB CALORIMETRY METHODS........................................ 27 9. COMPARISON TABLE DDS BOMB CALORIMETER SYSTEMS..................... 29 10. SOFTWARE COMPARISON/OVERVIEW .................................................. 30

10.1 CAL2K SOFTWARE OVERVIEW............................................................ 30 10.2 E2K SOFTWARE OVERVIEW................................................................ 30 10.3 ECO SOFTWARE OVERVIEW ............................................................... 31

11. INTERNATIONAL STANDARDS............................................................... 32 12. CERTIFICATES....................................................................................... 33 13. WARRANTY ........................................................................................... 34

13.1 PURPOSE OF PRODUCT WARRANTY..................................................... 34 13.2 OWNER’S WARRANTY........................................................................ 34 13.3 EXCLUDED (NON-WARRANTY) PARTS.................................................. 34

14. SHIPPING AND RETURNS POLICIES ...................................................... 35 14.1 SHIPPING POLICY............................................................................. 35 14.2 RETURNS POLICY ............................................................................. 35

15. TECHNICAL SUPPORT & TRAINING ....................................................... 35 16. PRODUCT CATALOGUE DISCLAIMER...................................................... 36 17. CONTACT INFORMATION....................................................................... 36

Page 4: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com

1

1. INTRODUCTION

Digital Data Systems (dds) specializes in the design and manufacture of bomb calorimeter systems for different applications from the coal mining industry to educational institutes. The bomb calorimeter systems are the most advanced bomb calorimeter system available today. The CAL2k, ECO and e2K are "dry" static jacket isothermal calorimeters. The system is used to measure the calorific value of both solids and liquids. dds have three bomb calorimeter systems in their CAL2k range:

CAL2K ECO E2K The CAL2k is a calorimeter aimed at a higher volume market, which requires higher quantities of CV samples per day. A determination will take approximately 7 minutes to run, and depending on the number of vessels used 10 samples can be carried out every hour.

The ECO is an inexpensive calorimeter aimed at a lower volume market, which requires fewer quantities of CV samples per day. A determination will take approximately 50 minutes to run, and 1 sample can be carried out every hour.

The e2K is an inexpensive calorimeter language compatible system suited to both the lower volume and higher volume market. A determination can take 10 minutes using the CAL2k-2 Cooler and 1 vessel (Standard system).

1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM?

The list of industries is endless; here are a few industries that are currently using our system:

• Coal Production (Collieries) • Coal end user (e.g. Steel, Paper, Sugar, Brick, Cement Production) • Coal Power Station • By-product analysis (Paper, Sugar, Fuel, Fertilizers) • Fuel Analysis (Solid and Liquid) • Waste Product Incineration • Munitions Propellants • Pyrotechnics • Standards Associations • Animal Feed Research (chicken, cattle, sheep) • Maize Production Research • Dairy Production Research • Commercial Analytical Laboratories • Universities, Colleges, Educational Institutions (Chemistry, Metallurgy, Zoology,

Biology, Animal Sciences)

1.2 WHY USE A DDS BOMB CALORIMETER SYSTEM? dds has more than 30 years experience in calorimetry. Our systems have been in the field since 1972 and there are some of our older models, which are still being used in the field today.

Page 5: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

1.3 HISTORY

In 1972, the first unit - AMPC (Automatic Micro Processor Calorimeter) was produced. It was a dual water isothermal unit controlled by a microprocessor. In 1980 a new revolutionary design of vessel, namely the DRY vessel or CP510 was produced (no surrounding water jacket). Determination time was significantly reduced, increasing the unit efficiency by 4 times. With the processing power of the microprocessors available at the time, the CP500 Calorimeter was born with its striking “buttercup yellow” colour. The CP500 proved to be a very fast and reliable system and was exported to the USA, Europe, Scandinavia, Australasia, Africa and India. In early 2002 work began on the CAL2k. The tried and tested DRY system was retained and only the very latest electronic technology was used, including surface mount devices. The CAL2k is now in full production and is sure to be an even bigger success than its predecessor, the CP500. In 2005, dds came to realize the need for a smaller, low sample volume, and inexpensive calorimeter system with the same accuracy and reliability of the CAL2k. The ECO was then created as an alternative system to the CAL2k. In 2007 we introduced the new e2K system which is a language compatible system. The e2K system is suited to both high and low volume markets and is available in a variety of system options. The speed of determination of the e2K is equivalent to the speed of determination of the CAL2k.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 2

Page 6: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 3

2. THE CAL2K PRODUCT RANGE

• The CAL2k Bomb Calorimeter System

• The ECO Bomb Calorimeter System

• The e2K Bomb Calorimeter System

THE CAL2K

he CAL2k system is the most advanced, fully automatic

he CAL2k is easy to use and has special features, which

3.

Tcalorimeter system available today. The result of years of research with dedicated engineers employing the latest state-of-the-art technology and the highest quality materials. Tplaces it in a league of its own. Accuracy is guaranteed with microprocessors that use self-correcting processes,

exceeding the standard requirements of DIN, ASTM and ISO. Its ability to interface with a personal computer, ensure preferred results with displays, data printouts and connectivity across a network.

3.1.1 SYSTEM FEATURES

• Easy to maintain, self-system test for Technicians. • Fully automatic operation – temperature readings and calculations are done for you. • Possible to network up to 7 calorimeters. • Rapid and accurate determinations • 6-8 samples per hour using CAL2k-2 Cooler, 1 vessel • 10 samples per hour using the CAL2k-2 Cooler, 2 vessels. • Isothermal design using a waterless patented vessel (no water bucket, spillage or

measuring). • Large memory for storing more than 2000 determinations, including operating

conditions and user statistics. • Manual or automatic mass entry through the front panel, balance interface or PC. • Automatic correction for firing wire, cotton, spikes, etc. • Fully automatic calibration, with 10 stored calibration curves per vessel for standard

deviation. • Calibrated sensors built into vessel wall. • Vessel is SMART with fault diagnostics and calorific processor. • Adjustable firing limits set per vessel. • High and low mass limits. • Determination cycle adjustment. • Other features include User, Group and Sample ID. • CE Certified. TÜV Certificate. • Precise and reproducible determination of gross calorific values according to ISO

1928, DIN 51900 and BS 1016:105.

Page 7: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

3.1.2 SOFTWARE FEATURES

• User controlled access for vital or routine operations. • Automatic result retrieval for selected and formatted data. • Weigh station operation. • Calorimeter and vessel setup. • Calibration management. • Service and maintenance routines. • Filing, printing and other operator functions, including data export to common

formats. • Real time graphical temperature display. • Network status display. • Grouped sample determinations and analysis. • Colourful general user interface for all windows operating systems

3.1.3 ENHANCED FEATURES

• Computer, balance and calorimeter interface. • Adjustable firing voltages for different firing wire. • Real-time clock with one-week backup. • 1-Watt consumption. • Manufacture and history information stored on each vessel.

3.2 THE CALORIMETER

The CAL2k-1 Calorimeter can operate as either a stand-alone unit or via a PC. When operating as a stand-alone unit it is operated via the keyboard and all relevant information is displayed on the LCD Display. Up to 7 Calorimeters can be connected as a network with or without a PC.

3.3 THE VESSEL

The CAL2k-4 vessel is the first of its kind and is the heart of the CAL2k System. Its sophisticated design allows the temperature to be measured to five decimal places in degrees Celsius. The vessel is an intelligent (SMART) vessel with a microprocessor built into its base. The vessel is capable of: firing counts, identification, memory and reconditioning data.

3.4 THE FILLING STATION The Filling Station is designed to fill the vessel with oxygen to 3Mpa. The filling rate is controlled so as not to disturb the sample in the crucible. The Filling Station is extremely easy to operate and requires minimal adjustments and maintenance.

3.5 THE COOLER

The unit is designed to reduce the temperature of a recently fired vessel, obtained from the calorimeter, to ambient temperature. Solid state cooling is used and the hot junction of the peltier elements is cooled by a continuous trickle from the mains water supply.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 4

Page 8: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 5

3.6 SYSTEM COMPONENT GUIDE OPTIONAL EXTRAS 3.7 OPTIONAL EXTRAS

Centre Electrode Outside Electrode Deflector Plate 5 Firing Wires 10 Top and Bottom O-Rings 5 Lid O-Rings

Filling Station Installation Kit CAL2k-3-KT

Filling Station CAL2k-3

Ambient Cable Mains Cable 10m Water Pipe 3 Water Pipe Hose Clamps

Included in Kit:

Cooler Installation Kit CAL2k-2-KT

Cooler CAL2k-2

Printed Handbook Operating Manual on CD Mains Cable 2 Bottles Benzoic Acid Tablets (50) – Non Certified 5 Crucibles 1000 Firing Cottons Preparation Stand Wire Brush

Balance Cable

Included in Kit:

High Pressure Pipe 1m Oxygen Regulator Connection 2 Flow Adjustor O-Rings 5 Nozzle O-Rings Handling Hook Defiller Cap

Included in Kit:

Vessel Installation Kit CAL2k-4-KT

Vessel CAL2k-4

Included in Kit:

Calorimeter Installation Kit CAL2k-1-KT

Calorimeter CAL2k-1

Description Resolution:

Minimum System

Docking Station CAL2K-5

Description Model

Calorimeter Interconnection Cable CAL2k-CIC

PC Software CAL2k-PCS

PC Cable CAL2k-PCC

REGULATOR High Pressure Regulator – This must be supplied prior to installation, can be sourced locally or supplied by DDS

Page 9: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

3.8 ITEMS TO SOURCE LOCALLY

Note: The recommended balance is Sartorius CP64 but any other equivalent balance can be used. The balance must have the following specifications: • Analytical balance with weighing

Shield • Readability: 0.1mg • Output: RS 232

Water Tap Connection

Balance

PC with Windows Operating Software (Only if running software, PC not required to operate the system

3.9 TECHNICAL SPECIFICATIONS

CAL2k-4 CAL2k-3 CAL2k-2 CAL2k-1

mm 82mm Diameter 150mm Height

220 x 450 x 300

Dimensions: 280 x 290 x 400

280 x 250 x 320 (Width, Height,

Depth)

Weight: kg 9 10 4.5 2

Power: - - 90 – 260 VAC

50/60Hz

90 – 260 VAC

50/60Hz

0 – 60°C 0 – 60°C - - Operating Temperature:

0.1 (%RSD) Repeatability:

0.001 (MJ/kg) Resolution:

0.000001°C Temperature Resolution:

Calibration: Calibration Details per Calibration Curve 10 Stored Calibration and Deviation Curves Automatic Standard Deviation Calculations

Results: 6-8 per Vessel, per Calorimeter per hour

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 6

Page 10: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

3.10 SYSTEM CONFIGURATION

3.10.1 MINIMUM SYSTEM FOR MANUAL MASS ENTRY

Optional: 2 x CAL2k-4 Vessels for Improved Capacity

1 x CAL2k-1 Calorimeter 1 x CAL2k-2 Cooler 1 x CAL2k-4 Vessel 1 x CAL2k-3 Filling Station

3.10.2 STANDARD SYSTEM FOR AUTOMATED MASS ENTRY

NETWORKING Multiple CAL2k-1 Calorimeters (Maximum of 7 Calorimeters)

Networking Systems for Direct Automated Mass Entry

Optional: 3 or 4 x CAL2k-4 Vessels

1 x Sartorius Balance (source locally) 1 x CAL2k-1 Calorimeter 1 x CAL2k-2 Cooler 2 x CAL2k-4 Vessels 1 x CAL2k-3 Filling Station

3.10.3 SYSTEM WITH DOCKING STATION

The purpose of the CAL2K-5 DOCKING STATION is to allow the orderly preparation of bomb vessels in a high production environment with multiple calorimeters, multiple bomb vessels and multiple operators. Connect to: 1. Balance (Source locally) 2. PC Keyboard: a) Keyboard entry b) Barcode scanner c) Both of the above 3. PC via RS232 interface

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 7

Page 11: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

4. THE E2K The e2K is an inexpensive calorimeter language compatible system suited to both the lower volume and higher volume market. The e2K unit is compatible with all languages including Chinese. The LCD display can be read in one standard (English) or alternative language. 4.1 SYSTEM FEATURES

• Intelligent operation support • Very fast operation due to improved software • Easy to maintain, self test for technicians and

operators • Compact size • Operation can be configured to suit every laboratory • Rapid and accurate determinations • Isothermal design using a waterless patented vessel (no bucket, spillage or measuring) • Large memory for storing more than 1000 determinations • Manual or automatic mass entry through the keyboard or balance interface • Automatic correction of firing wire, cotton, spikes etc. • Easy to read LCD display in one standard (English) and a alternative language • Any LCD display language can be loaded in the • Fully automatic calibration in BTU, CAL or MJ/Kg • Calibrated sensors built into vessel wall • Intelligent vessel with fault diagnostics and microprocessor • Adjustable firing limits • High and low mass limits to avoid operator error • Determination cycle adjustment • Interfaces to standard PS2 computer keyboard • Acoustical warning for operator convenience • Auto incrementing 5 digit sample identification • Factory pre-setting parameters can be invoked from the keyboard • Extra 8 digit group identification can be used as a sample id • Can print results directly without PC on standard 80 column serial printer • Real time clock for date/time stamping of all results • Pre-assembled optional cooler interface • Increased MTBF, approximately 10 years • Virtually Maintenance free • Protected card for easy exchange

4.2 SOFTWARE FEATURES

• User controlled access for setting vital parameters once • Automatic result retrieval for selected and formatted data • Filing, printing and other operator functions, including data export to common formats • Real time graphical vessel temperature display • Grouped sample determinations and analysis • Alternative language management • English Windows Operating e2K Software version only • Real time clock setting • Vessel data viewing • High speed communication for fast data retrieval

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 8

Page 12: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

4.3 ENHANCED FEATURES

• Standard balance interface via 25 pin connector • Alternative language switch at rear of unit • Adjustable firing voltages for different firing wire • Real-time clock with one week back up • 1 watt power consumption while display is off • Manufacture and history information stored on each vessel • Floating point calculations for high accuracy • Single control card design • Keyboard function keys for often repeated operations • Complies to ALL International Standards

4.4 THE E2K CALORIMETER

The e2K Calorimeter is a small, desktop apparatus, in a warm buttercup yellow. It performs the temperature analysis of the sample. The operator, balance, vessel and PC interface together. It is insulated against ambient temperature changes via an "air barrier" and polystyrene. The ambient temperature is measured. It "fires" the sample by heating the firing wire inside the vessel. The unit is language compatible, and has an easy to read LCD display in one standard (English) and an alternative language. The system is suited to both high and low volume markets available in a variety of system options.

Note: The e2K uses the same cooler, filling station and vessel as the CAL2k.

4.5 TECHNICAL SPECIFICATIONS

- - 0 – 60°C 0 – 60°C

Operating Temperature:

- - 90 – 260 VAC

50/60Hz

90 – 260 VAC

50/60Hz

Power:

CAL2k-4 CAL2k-3 CAL2k-2 E2K

6 per calorimeter per hour, using one vessel and CAL2k-2 Cooling system

Results:

Calibration Details per Calibration Curve Calibration:

0.000001°C Temperature Resolution:

0.001 (MJ/kg)

0.1 (%RSD) Repeatability:

Resolution:

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 9

Page 13: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

4.6 OPTIONAL EXTRAS Model Description

PC Software E2k-PCS

PC Cable CAL2k-PCC

REGULATOR High Pressure Regulator – This must be supplied prior to installation, can be sourced locally or supplied by DDS

4.7 ITEMS TO SOURCE LOCALLY

Balance

Note: The recommended balance is Sartorius CP64 but any other equivalent balance can be used. The balance must have the following specifications:

• Analytical balance with weighing Shield

• Readability: 0.1mg • Output: RS 232

Water Tap Connection (If cooler is supplied) PC with Windows Operating Software (Only if running the software, PC is not required to operate the system)

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 10

Page 14: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

4.8 SYSTEM OPTIONS

The e2k is available in 3 system options to accommodate different customer budgets. Options currently available: Minimum System Option 1 x E2k Bomb Calorimeter 1 x CAL2k-3 Filling Station 1 x CAL2k-4 Vessel Determination time is 5-7 minutes; however it can take up to 1 hour to cool the CAL2k-4 Vessel naturally without the CAL2k-2 Cooler or the E2K Air Cooler. Making your total

determination time +1 hour. Using the E2k Air Cooler will increase your determination time by at least 30 minutes. Multiple vessels can be used to increase your sample throughput. OPTIONAL: E2K Air Cooler which will reduce determination time by 30 minutes

Standard System Option 1 x E2k Bomb Calorimeter 1 x CAL2k-2 Cooler 1 x CAL2k-3 Filling Station 1 x CAL2k-4 Vessel Determination time is 5-7 minutes and using the CAL2k-2 Cooler to cool the vessel will take a further 3-5 minutes, making your total determination time approximately 10 minutes per vessel. Elite System Option 1 x E2k Bomb Calorimeter 1 x CAL2k-2 Cooler 1 x CAL2k-3 Filling Station 2 x CAL2k-4 Vessels Multiple Vessels recommended for high sample volumes. Total determination time 5-7 minutes per vessel. Discounted price offered on quantities of 2 or more vessels. Included in the above options is a standard PS2 computer keyboard and all installation kits required for the calorimeter, cooler, filling station and vessel. Optional Extras: 1 x E2k Windows Operating Software 1 x PC Cable 1 x Oxygen Regulator for Filling Station 1 x E2K Air Cooler (only applicable to the Minimum System Option)

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 11

Page 15: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

5. THE ECO

The new ECO system is to operate together with the existing CAL2k and form part of the CAL2k range. It is aimed at a different type of market and will not be in direct competition to the CAL2k. The ECO is an inexpensive calorimeter aimed at a lower volume market, which requires fewer quantities of CV samples per day. The ECO unlike the CAL2k will use no water; the vessel will cool naturally while standing on a workbench or table. 5.1 SYSTEM FEATURES

• Precise and reproducible determination of gross calorific values according to ISO

1928, DIN 51900 and BS 1016:105 • Ideal for low volume applications up to 1 sample per hour with one vessel. Max. 8

CV samples per day. • 20 minutes per determination and a further 30 minutes to cool the vessel naturally

(no water), total 50 minutes per determination resulting in 1 sample per hour. • No Water required • INEXPENSIVE, Breaks all price/performance norms due to mass production • No operator attention required during analysis, the LCD display prompts the

operator and displays faults and progress • Uses the proven (mass produced) CAL2K-4 “SMART VESSEL” • Small and compact • Uses standard PC keyboard (PS2) • Result retrieval connected to a PC, using the supplied WINDOWS XP software • Factory calibrated, but can be calibrated in the field • Stores 1024 results and other variables • Sample identification is settable and self-incrementing • Vessel temperatures can “stream” to a PC for user spreadsheet analysis • Factory setup for standard operation, however some values can be changed via the

keyboard to suit specific applications • Default setup, which can be invoked from the keyboard • Can use multiple vessels • Can use “SPIKING” for hard to combust samples • Can connect directly to a balance for MASS entry, or the mass can be entered via

the keyboard • Built in user diagnostic, which is ideal for first hand diagnostic • Vessels are pre-programmed for 5000 determination cycles before inspection • Factory guaranteed for 3 years • External power supply for 9 Volts, 100mA is supplied • Works in BTU/lb, CALORIES/g, or MJ/Kg

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 12

Page 16: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 13

5.2 THE ECO CALORIMETER

The CAL2k-ECO Calorimeter is a small, desktop apparatus, in a warm buttercup yellow. It performs the temperature analysis of the sample. The operator, balance, vessel and PC interface together. It is insulated against ambient temperature changes via an "air barrier" and polystyrene. The ambient temperature is measured. It "fires" the sample by heating the firing wire inside the vessel.

Note: The ECO uses the same filling station and vessel as the CAL2k.

5.3 SYSTEM COMPONENT GUIDE

Filling Station

Accessories Required:

Balance Cable Preparation Stand Defiller Cap 2 Benzoic Acid Tables – Non Certified Firing Cotton Wire Brush Handling Hook 5 Crucibles

Vessel CAL2K-4

CAL2K-3

Includes: Printed Handbook Operating Manual on CD CD with Software PC Cable PC Keyboard Power Supply Mains Cable (Local customers)

Includes: Flow Adjustor O-Rings Nozzle O-Rings High Pressure Pipe 1m Oxygen Regulator Connection

Includes: Centre Electrode Outer Electrode Deflector Plate 10 Electrode O-Rings 5 Lid O-Rings 5 Firing Wire

Calorimeter CAL2K-ECO

Description: Model Number:

Minimum System

Page 17: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 14

5.4 TECHNICAL SPECIFICATIONS

5.5 OPTIONAL EXTRAS

5.6 ITEMS TO SOURCE LOCALLY

- - 0 – 60°C Operating Temperature:

- - 90 – 260 VAC 50/60Hz

Power:

2kg 4.5 kg 5 kg Kg Weight:

82mm Diameter 150mm Height

220 x 450 x 300

240 x 240 x 330

mm Dimensions: (Width, Height, Depth)

CAL2K-4 CAL2K-3 CAL2K-ECO

8 per day (absolute maximum is one determination every 40-50 minutes for multiple vessel system)

Results:

Single curve calibration Calibrat on: i

0.000001°C Temperature Resolution:

0.01 (MJ/kg) Resolution:

0.1 (%RSD) Repeatability:

Description Model

REGULATOR High Pressure Regulator – This must be supplied prior to installation, can be sourced locally or supplied by DDS

Note: The recommended balance is Sartorius CP64 but any other equivalent balance can be used. The balance must have the following specifications: • Analytical balance with weighing

Shield • Readability: 0.1mg • Output: RS 232

Balance

PC with Windows Operating Software (Only if running the software, PC is not required to operate the system)

Page 18: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

6. OPTIONAL EXTRAS

6.1 DOCKING STATION

The DOCKING STATION CAL2k-5 is the latest addition to the CAL2k range of equipment. It was designed on request from a customer, but has a wider application. The DOCKING STATION is required when the Mass entry needs to be automated to avoid sample swapping and manual transcription errors. It is suitable in a high production environment with

multiple calorimeters, multiple bomb vessels and multiple operators. The Docking Station can only be used with the CAL2k system. Not required with the ECO or e2K systems.

6.1.1 OPERATIONS WITHOUT THE DOCKING STATION

The CAL2k requires a MASS and a Sample Identification (SID) in order to determine the calorific value and to log the result. Both values can be entered from the CAL2k front panel. If a balance is connected to the CAL2k then the MASS is read directly (no transcription error), but the SID needs to be entered manually, or it is self-incrementing. The CAL2k has no facility to read BARCODE. If the balance is located in a different room from the sample preparation (filling) and the CAL2k, then this can lead to more transcription errors, because log sheets are filled in. It can also lead to Sample Identification (SID) swapping.

6.1.2 OPERATIONS WITH THE DOCKING STATION

The DOCKING STATION is located close to the balance and connects to it. The SID entry can be from 3 sources: a. The Internal numerical keypad b. A optional PC keyboard c. A optional barcode scanner The MASS must come from the balance. Once the sample is weighed and the SID is entered, then this information is entered in the VESSEL. The crucible is placed in the vessel, and the vessel is closed. It can be filled and placed in any CAL2k calorimeter. The CAL2k recognizes the “BOMB MASS”, and reads the MASS, SID, GID from the vessel without any further operator intervention, after the CAL2k lid is closed.

6.1.3 COMPLEX BARCODE FEATURES

The barcode can be broken up into 3 fields: 1. Group Identification (GID) 2. Sample Identification (SID) 3. Ignored barcode digits (IGN) This allows the structure of the barcode in to ONE long code comprising GID and SID, where the IGN field can ignore any check-digits.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 15

Page 19: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

6.1.4 SYSTEM SET UP

6.1.5 ADVANTAGES OF VESSEL MASS OPERATION

Result includes: Group ID, Sample ID, and Mass from

Vessel

Balance

Barcode Scanner

Optional Keyboard

DOCKING STATION

Vessel CAL2k-4

Calorimeter CAL2k-1

Operation: Vessel Mass Enabled

Vessel Loaded with: Group ID, Sample ID, MASS

1. The vessel can be prepared in a different location away from the calorimeter. 2. The balance can be away from the calorimeter. 3. A bar code scanner can be used for sample identification. 4. The bar code can be “broken up” into Group ID and Sample ID. 5. An optional “PC keyboard” can be used in addition to the scanner (or instead) for Sample

ID. 6. No keyboard operation is required on the calorimeter. 7. The “vessel” can be placed in any calorimeter. 8. Sample ID mix up is avoided. There are no real disadvantages of using “vessel mass” operation, however it is recommend in this type of operation that more than one vessel be used. 6.2 E2K AIR COOLER

This is a new edition to our range of accessories. The E2K Air Cooler is only applicable to the E2K Minimum System Option. The air cooler will reduce your determination time by 30 minutes. It takes approximately 20 minutes for the E2k air cooler to bring the vessel temperature back to ambient temperature making it ready for the next determination cycle. Please note the E2K Air Cooler can be used with the ECO model; however the ECO calorimeter allows for the vessel temperature to be 8 Degrees C above ambient temperature when starting the next determination. With CAL2k and the E2k the vessel temperature needs to be back to ambient

temp before starting the next determination. So therefore the new E2K Air Cooler is more beneficial for the E2K range rather than the ECO. The Air cooler is not applicable to the CAL2k; the CAL2k-2 Cooler must be used.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 16

Page 20: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

6.3 HIGH PRESSURE OXYGEN REGULATOR A supply of oxygen at a pressure of 3.0Mpa (30 bars) (3000Kpa) within 10 meters of the bomb calorimeter system is required. A suitable pressure regulator must be supplied to allow for this high pressure. DDS can supply a suitable regulator at an additional cost or this item should be sourced locally by the agent or customer. However it is important to note that this item is vital and MUST be available before installation of the unit. Without the high pressure regulator you will NOT be able to operate the unit.

7. CONSUMABLES

The CAL2k, ECO and e2K use the same consumables and accessories apart from 4, which are different for the ECO only. See below the part numbers for those consumables, which are different. The other part numbers remain the same when ordering consumables for the ECO, CAL2k or e2K. Please note the Installation kits differ for the CAL2k, ECO and e2K see below. CAL2K-2 COOLER

CAL2K-2-HC – Water Pipe Hose Clamp Hose Clamp

CAL2K-2-AC Ambient Cable

Ambient Cable

CAL2K-2-WP – Water Pipe

10m Water pipe

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 17

Page 21: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

CAL2K-3 FILLING STATION

CAL2K-3-FR – Flow Adjustor O-Ring Flow adjustor O-Ring 4,42 x 2,62 100 per packet

CAL2K-3-NR – Nozzle O-Ring

Nozzle O-Ring 8 x 2,62 100 per packet

CAL2K-3-OR – Oxygen pipe nipple o-ring Oxygen pipe nipple o-ring 3,1 x 1,6 100 per packet

CAL2K-3-PR – Piston O-Ring Piston O-Ring 8 x 2,62 100 per packet

CAL2K-3-OP – High pressure pipe

1m High pressure pipe

CAL2K-3-RC – Regulator Connection

Oxygen Regulator connection

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 18

Page 22: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

CAL2K-4 VESSEL (same for both CAL2k and e2K)

CAL2K-4-CE – Centre Electrode Centre Electrode, top & bottom and sleeve

CAL2K-4-OE – Outside Electrode

Outside Electrode, with sleeve

CAL2K-4-DP – Deflector Plate

Deflector Plate

CAL2K-4-LR – Lid O-Ring

Lid O-Ring 46,0 x 3,53 100 per packet

CAL2K-4-CR – Top & Bottom O-Ring

Top & Bottom O-Ring 3,1 x 1,6 viton 100 per packet

CAL2K-4-PS – Preparation Stand Preparation Stand

CAL2K-4-HH – Handling Hook

Vessel Handling Hook

CAL2K-4-WB – Wire Brush

Wire Brush

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 19

Page 23: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

CAL2K-4-DC – Defiller Cap

Vessel Defiller Cap

CAL2K-4-DA – Defiller Adaptor Vessel Defiller adaptor for C12 samples This adaptor regulates the flow of the gas out the vessel and comes with a 1m oxygen line.

CAL2K-4-BA – Certified Benzoic Acid Tablets

Certified Benzoic Acid Tablets 50 x 0.5gr tablets per bottle

CAL2K-4-CB – Stainless Steel Crucibles

Stainless Steel Crucibles Measurements: D1=30mm D2=25mm H=15mm Sample size 0.5gr 10 per packet

CAL2K-4- Firing Cotton

100% pure cotton 1000 per packet

CAL2K-4-FW – Firing Wire Semi permanent firing wire Pre cut 40mm of 0,4 wire 5 per packet ECO firing wire is not pre-bent. Same wire. Part no: CAL2k-E-FW.

CAL2K-4-FR – Firing Wire (used for Demonstration purposes)

Firing Wire used for firing without oxygen 1000 per packet FOR DEMONSTRATION PURPOSES ONLY

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 20

Page 24: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

ECO PART NUMBERS WHICH DIFFER TO CAL2K and e2K

CAL2K-E-CE – Centre Electrode Centre Electrode and sleeve

CAL2k-E-OE – Outside Electrode Outside Electrode with sleeve

CAL2K-E-CR – Electrode O-Rings

Top and bottom O-rings 3, 1x1, 1.6 Viton

CAL2K-E-FW – Firing Wire

Firing Wire used for firing without oxygen 5 per packet

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 21

Page 25: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

OPTIONAL EXTRAS

CAL2K-BC – Balance Cable Balance Cable

CAL2K-PCS – PC Software

Cal2k Windows Operating Software

CAL2K-PCC – PC Cable

RS232 PC Cable

CAL2K-CIC – Calorimeter Interconnection cable

Calorimeter Interconnection cable

CAL2K-MC – Mains Cable

Mains Cable

REGULATOR – High pressure Oxygen Regulator

GAS-ARC type GA600 Regulator. A high pressure regulator capable of a pressure of 3000Kpa (30 bar) is required before installation. This item can be purchased from DDS or sourced locally by the agent or customer.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 22

Page 26: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

7.1 CAL2K INSTALLATION KITS

When ordering an entire CAL2k system from Digital Data Systems the system is provided with the Calorimeter (CAL2k-1), the Cooler (CAL2k-2), the Filling Station (CAL2k-3) and the vessel (CAL2k-4). Each of these items is supplied with an installation kit, which must be ordered together with the system. The kits are used to assist with installation of the system and to get you started with using the system. The kits are made up as follows:

7.1.1 CAL2K-1-KT – CALORIMETER INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

5 CAL2K-4-CB Crucibles 1 CAL2K-4-PS Preparation stand 1 CAL2K-4-WB Wire Brush 1 CAL2K-HB Printed handbook 1 CAL2K-OM Operating Manual on CD 1 CAL2K-4-FC Firing Cotton (1000 per packet) 2 CAL2K-4-BA Benzoic Acid Tablets – Non certified 1 CAL2K-BC Balance Cable 1 CAL2K-MC Mains Cable

7.1.2 CAL2K-2-KT – COOLER INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

3 CAL2K-2-HC Water Pipe Hose Clamp 1 CAL2K-2-AC Ambient Cable 1 CAL2K-2-WP 10m Water Pipe 1 CAL2K-MC Mains Cable

7.1.3 CAL2K-3-KT – FILLING STATION INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

2 CAL2K-3-FR Flow Adjustor O-Rings 5 CAL2K-3-NR Nozzle O-Rings 2 CAL2K-3-OP 1 m High pressure pipe 1 CAL2K-3-RC Oxygen Regulator connection 1 CAL2K-4-DC Defiller Cap 1 CAL2K-4-HH Handling Hook

7.1.4 CAL2K-4-KT – VESSEL INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

1 CAL2K-4-CE Centre Electrode 10 CAL2K-4-CR Top & Bottom O-ring 5 CAL2K-4-LR Lid O-Ring 1 CAL2K-4-OE Outside Electrode 1 CAL2K-4-FW Firing Wire 1 CAL2K-4-DP Deflector Plate

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 23

Page 27: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

7.2 E2K INSTALLATION KITS

When ordering an e2K system note there are different system options. The Minimum System option consists of the Calorimeter (e2K), Filling Station (CAL2k-3) and the Vessel (CAL2k-4). The Standard System option consists of the Calorimeter (e2K), Cooler (CAL2k-2), Filling Station (CAL2k-3) and the Vessel (CAL2k-4). The Elite System option consists of the Calorimeter (e2K), Cooler (CAL2k-2), Filling station (CAL2k-3) and two Vessels (CAL2k-4). Each of these items is supplied with an installation kit. The installation kits are included with the system. The kits are used to assist with installation of the system and to get you started with using the system. The kits are made up as follows:

7.2.1 E2K-KT – CALORIMETER INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

1 CAL2K-1-SM Power Supply 1 CAL2K-KB PC Keyboard 1 CAL2K-MC Mains Cable 1 E2K-HB Printed handbook 1 E2K-OM Operating Manual on CD

7.2.2 CAL2K-2-KT – COOLER INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

3 CAL2K-2-HC Water Pipe Hose Clamp 1 CAL2K-2-AC Ambient Cable 1 CAL2K-2-WP 10m Water Pipe 1 CAL2K-MC Mains Cable

7.2.3 CAL2K-3-KT –FILLING STATION INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

2 CAL2K-3-FR Flow Adjustor O-Rings 5 CAL2K-3-NR Nozzle O-Rings 2 CAL2K-3-OP 1m High Pressure Pipe 1 CAL2K-3-RC Oxygen Regulator Connection

7.2.4 CAL2K-4-KT –VESSEL INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

1 CAL2K-4-CE Centre Electrode 10 CAL2K-4-CR Top & Bottom O-ring 5 CAL2K-4-LR Lid O-Ring 1 CAL2K-4-OE Outside Electrode 1 CAL2K-4-FW Firing Wire 1 CAL2K-4-DP Deflector Plate

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 24

Page 28: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

7.2.5 E2K-ACC-KT –ACCESSORIES INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

1 CAL2K-BC Balance Cable 1 CAL2K-4-PS Preparation Stand 1 CAL2K-4-DC Defiller Cap 2 CAL2K-4-BA Benzoic Acid Tablets – Non Certified 1 CAL2K-4-FC Firing Cotton 1 CAL2K-4-WB Wire Brush 1 CAL2K-4-HH Handling Hook 5 CAL2K-4-CB Crucibles

7.3 ECO INSTALLATION KITS

When ordering an entire ECO system from Digital Data Systems the system is provided with the Calorimeter (CAL2k-ECO), the Filling Station (CAL2k-3) and the vessel (CAL2k-4). Each of these items is supplied with an installation kit. The installation kits are included with the system. The kits are used to assist with installation of the system and to get you started with using the system. The kits are made up as follows:

7.3.1 CAL2K-E-KT – CALORIMETER INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

1 CAL2K-1-SM Power Supply 1 CAL2K-KB PC Keyboard 1 CAL2K-MC Mains Cable 1 CAL2K-E-HB Printed handbook 1 CAL2K-E-OM Operating Manual on CD 1 CAL2K-E-SW Windows Operating Software on CD 1 CAL2K-PCC PC Cable

7.3.2 CAL2K-3-KT –FILLING STATION INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

2 CAL2K-3-FR Flow Adjustor O-Rings 5 CAL2K-3-NR Nozzle O-Rings 2 CAL2K-3-OP 1m High Pressure Pipe 1 CAL2K-3-RC Oxygen Regulator Connection

7.3.3 CAL2K-4-KT –VESSEL INSTALLATION KIT

QTY. PART NUMBER

DESCRIPTION

1 CAL2K-E-CE Centre Electrode 10 CAL2K-E-CR Top & bottom O-Rings 5 CAL2K-4-LR Lid O-Rings 1 CAL2K-E-OE Outside Electrode 1 CAL2K-E-FW Firing Wire 1 CAL2K-4-DP Deflector Plate

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 25

Page 29: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

7.3.4 ECO-ACC-KT –ACCESSORIES INSTALLATION KIT QTY. PART NUMBER

DESCRIPTION

1 CAL2K-BC Balance Cable 1 CAL2K-4-PS Preparation Stand 1 CAL2K-4-DC Defiller Cap 2 CAL2K-4-BA Benzoic Acid Tablets – Non Certified 1 CAL2K-4-FC Firing Cotton 1 CAL2K-4-WB Wire Brush 1 CAL2K-4-HH Handling Hook 5 CAL2K-4-CB Crucibles

7.3.5 ECO FREE DEMO KIT When ordering an ECO demo system, the demo system is supplied with a free demo kit. The purpose of the Demo Kit is so that the ECO unit can be demonstrated without the use of a balance or oxygen. This kit should be used for demonstration purpose ONLY WHEN NO OXYGEN IS AVAILABLE. QTY. PART NUMBER

DESCRIPTION

1 CAL2K-E-SP Sample powder to run 20 determinations 1 CAL2K-4-FR Firing wire 100 per packet

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 26

Page 30: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

8. COMPARING BOMB CALORIMETRY METHODS

The CAL2k, ECO and e2K bomb calorimeter systems are "dry" static jacket isothermal calorimeters.

Comparing the different bomb calorimetry methods:

Adiabatic System: In the adiabatic system the environment is controlled so that no energy is lost or gained. To achieve this state, the Calorimeter Vessel (bomb) and bucket are surrounded by a water jacket which is strictly temperature controlled by a system of circulators, heaters and coolers, so that it has the same temperature as the inner bucket. Isothermal System: The Isothermal calorimeter measures the effect that the environment has on the bucket before and after the determination and corrects the result accordingly. This implies that the environment is stable during the determination. To achieve stability a large body of water surrounds the vessel and bucket assembly, which is thermally stable. DDS Developed their first calorimeter system in 1972, this system was a full isothermal system with water jacket. Prior to development of the CP500, it was demonstrated by Digital Data Systems that a polystyrene jacket, which has very little thermal mass and negligible heat conductivity, could replace the water jacket. The system was referred to as the "Dry Method". Isoperibol (Isothermal) calorimeter: With an isoperibol calorimeter a water jacket surrounds the bucket in which the vessel sits. The surrounding water jacket is controlled to be at a constant temperature. When a sample is fired, the heat from the vessel is transferred to the water in the bucket and the temperature rise of the bucket water is measured to determine the calorific value of the sample. "Dry" Static Jacket (Isothermal) calorimeter (DDS - CAL2k, ECO and e2K):

The "dry" static jacket isothermal method has no water inside the calorimeter. An aluminium sleeve that is an integral part of the vessel replaces the bucket. The sleeve transfers the heat from the combustion process rapidly and evenly around the outside shell of the vessel. The temperature sensors are located between the combustion vessel and the sleeve. The temperature rise of the vessel is measured to determine the calorific value of the sample. The surrounding water jacket is replaced by polystyrene that is insulated from the outside environment by an aluminium box. The timing of the determination and the quality of the insulation is such that the environment has negligible effect on the determination.

The advantages to the "dry" static jacket isothermal method are:

• No complicated water pipes and valves. • The associated switching circuitry is not required. • This makes the unit virtually maintenance free. • The electronic control circuits are reduced thus making the unit very easy to fault find

in case of failure. • With reduced electronics in the calorimeter, the training of a service technician or

operator is reduced to an absolute minimum. • Precise and reproducible determination of gross calorific values according to ISO 1928

and DIN International Standards.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 27

Page 31: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

By making the vessel removable and by removing the cooling of the vessel from the calorimeter, the speed of each determination is greatly increased.

• Once a result is obtained for a determination, the hot vessel can be removed and cooled in a separate unit.

• This allows for another vessel to be inserted into the calorimeter and another determination to be started even before the previous vessel has been cooled.

• This allows for higher throughput in a single calorimeter

The use of the dry static jacket method allows water to be removed from the vessel surrounds which in turn allows electronics to be built into the vessel.

The advantages to this are:

• The temperature measurements of the vessel are totally independent of the calorimeter and all calibration data is stored inside the vessel.

• Multiple vessels can be used in a single calorimeter. • Any vessel can be used in any calorimeter without re-calibration. • If a vessel needs reconditioning or fails the calorimeter can still be used, using another

vessel.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 28

Page 32: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

9. COMPARISON TABLE DDS BOMB CALORIMETER SYSTEMS

FEATURES CAL2K E2K ECO

Size: Small, compact Small, compact Small, compact Samples: Up to 6 samples per hour

using CAL2k-2 Cooler and 1 vessel Up to 10 samples per hour using CAL2k-2 cooler and 2 vessels

Up to 6 samples per hour using CAL2k-2 Cooler and 1 vessel Up to 10 samples per hour using CAL2k-2 Cooler and 2 vessels Up to 2 samples per hour using E2k Air Cooler and 1 vessel Up to 3 samples per hour using E2k Air Cooler and 2 vessels Up to 1 sample per hour without any cooling unit and 1 vessel

Slower, 1 sample per hour. Only 8 samples per day

Determination time: Approx. 7 minutes per sample – single vessel Approx. 5 minutes per sample – two vessels Approx. 3 minutes per sample – three vessels

Approx. 10 minutes per sample – single vessel Approx. 7 minutes per sample – two vessels

Approx. 40 minutes per sample

CAL2k-2 Cooler: Required Optional, is supplied with Standard and Elite System Options

Not Optional

E2K Air cooler: Not Optional Only optional for the Minimum System

Optional, but will not increase number of samples per day. ECO allows for vessel to be 8°C above ambient Temp.

Water: Required for CAL2k-2 Cooler. A steady tap water flow or a closed circuit water tank with recycled water.

Required for CAL2k-2 Cooler if supplied. A steady tap water flow or a closed water tank with recycled water. The E2k Air cooler does not require water.

None

Vessel: Same for all units

CAL2k-4 Vessel, built in electronics and 8 sensors, results in a faster and more stable temperature reading

CAL2k-4 Vessel, built in electronics and 8 sensors, results in a faster and more stable temperature reading

CAL2k-4 Vessel, built in electronics and 8 sensors, results in a faster and more stable temperature reading

Filling station: Same for all units

Standard CAL2k-3 Filling Station

Standard CAL2k-3 Filling Station

Standard CAL2k3 Filling Station

PC software: Advanced with many features

Limited Limited

Printing of results Only via PC Software Capable of printing results without PC on standard 80 column

Only via PC software

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 29

Page 33: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

printer with RS232 port Drift parameters: Settable Settable via PC Fixed Corrections: 2 2 1 Spiking: Settable Settable Settable Reproducibility: 0.1% RSD 0.1% RSD 0.1% RSD Temperature Resolution:

0.000001°C 0.000001°C 0.000001°C

Resolution: 0.001 (MJ/Kg) 0.001 (MJ/Kg) 0.001 (MJ/Kg) Units of Measure: BTU/Mj/Cal BTU/Mj/Cal BTU/Mj/Cal Mass Entry: Auto or manual Auto or manual Auto or manual Keyboard: Built in PC type (standard PS2) PC type (Standard PS2) Balance Interface: Yes Yes Yes Display: LCD – English Only LCD – English and One

Alternative Language LCD – English Only

CE/TUV: Yes Yes Yes Docking Station Operation:

Yes Not required Not required

Calibration: Single or multiple Single Single Network: Yes No No GID: Yes Yes No SID: Auto increment or

manual Auto increment or manual

Auto increment or manual

SID characters: 8 – Alpha Numeric Extra 8 digit group identification – Alpha Numeric

4 – Numeric

Power supply: External 9V External 9V External 9V Memory: 2000 results +1000 results 1024 results Method used: Isothermal “dry” static

jacket method Isothermal “dry” static jacket method

Isothermal “dry” static jacket method

Multiple vessels Yes Yes No Vessel Determinations 5000 5000 5000 Comply to International Standards

Yes Yes Yes

10. SOFTWARE COMPARISON/OVERVIEW

10.1 CAL2K SOFTWARE OVERVIEW

1. Upload results from CAL2k and management of results. 2. Can view CAL2k set up, can change any parameter and download new setup to the

CAL2k or upload and download to files. 3. Can view vessel data. 4. Can view real time temperature curve for each determination – can view average

temperature and /or each sensor. 5. Can do heap management (see article attached on heap management). 6. Can do calibration curve management. 7. Allows for automatic upload of results from the CAL2k. 8. Allows for complete control of operation of CAL2k from PC.

10.2 E2K SOFTWARE OVERVIEW

1. Allows for uploading and management of results from E2k. 2. Can upload E2k setup, can change parameters and download setup to the E2k unit. 3. Can view vessel data. 4. Can view real time temperature curve for a determination. 5. Can manage language options.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 30

Page 34: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

Please remember that one can print results directly to an RS232 printer without using the software. This is only applicable to the E2k model. 10.3 ECO SOFTWARE OVERVIEW

1. Upload results from ECO. 2. Can view ECO setup and change only certain parameters and download to the ECO

unit. 3. Can view vessel data. 4. Can view real time temperature curve for each determination. 5. Can view uploaded results. 6. Allows for firmware upgrades.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 31

Page 35: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

11. INTERNATIONAL STANDARDS

ASTM Description Year Complies

D240-02 Heat of Combustion of Liquid Hydrocarbon Fuels by Bomb Calorimeter

2002 Yes

D4809-00 Heat of Combustion of Liquid Hydrocarbon Fuels by Bomb Calorimeter (Precision Method)

2000 Yes

E144-94 Standard Practice for Safe Use of Oxygen Combustion Bombs

1994 Yes

British Description Year Complies

BS 4791:1985 Specification for Calorimeter Bombs 1985 Yes

BS 1016:105:1992 Methods for analysis and testing of coal and coke. Determination of gross calorific value using adiabatic, isothermal or static bomb calorimeter.

1992 Yes

DIN Description Year Complies DIN 51900-2 Determining the Gross Calorific value of solid and liquid

fuels using isoperibol or static jacket calorimeter and calculation of net CV

2003 Yes

ISO Description Year Complies ISO 1928 Solid mineral fuels – Determining Gross calorific value

by bomb calorimetric method and calculation of net CV 1995 Yes

ISO 1928: 1995 Annexure B – Isoperibol and Static Jacket Bomb Calorimeter Below is an extract from this International Standard: Principle The Characteristic feature of isoperibol calorimeters is the isothermal jacket. The temperature of the surrounding thermostat is kept constant throughout the experiment by active control. The static jacket calorimeter has a thermal capacity such that, even without active control, its temperature remains nearly constant during measurements. In both cases there is a flow of heat between the calorimeter itself and the thermostat. Calorimeters surrounded by thermally insulating material behave largely as static jacket calorimeters. DIN 51900-2: 2003-05 Below is an extract from this International Standard: 3 Principle The calorimetric bomb prepared as specified in DIN51900-1 is placed in a calorimeter can, which is filled with water and surrounded by an isoperibol or static jacket insulating jacket. After temperature has equalized and the fore period has elapsed, the fuel sample is fired and the increase in temperature is recorded. Recording is terminated at the end of the after period. The gross calorific value of the sample material is calculated from the corrected temperature rise and the effective heat capacity of the calorimeter. NOTE: The specified method using an isoperibol or static jacket calorimeter covers all procedures in which the temperature of the calorimeter can differ from that of the jacket. As a result of the temperature difference, it is necessary to correct the rise in temperature for the heat exchange (1).

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 32

Page 36: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

12. CERTIFICATES

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 33

Page 37: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

13. WARRANTY

13.1 PURPOSE OF PRODUCT WARRANTY

Any part of a product is subject to ageing and eventual failure. These failures occur during normal use. The purpose of a warranty is to protect the user against a part failure caused by manufacturing defects in workmanship, material or design for a specific time. 13.2 OWNER’S WARRANTY

In the event of the Owner contracting directly with Digital Data Systems (Pty) Ltd “dds”, or in the event of the installation of a unit arising out of a contract between the Owner’s Dealer or any other party on his/her behalf and the said “dds”, the Owner records that not withstanding same, this warranty is the only warranty applicable to “dds” units and is expressly in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular purpose. No amendments or additions to this warranty shall be binding on “dds”, unless recorded in writing and signed by a duly authorized officer of the company. Digital Data Systems (Pty) Ltd (“dds”) agrees to repair and replace with a new or remanufactured part, any part, other than those excluded below, which after inspection, upon its return to the factory in Randburg, Johannesburg by the Owner or the authorized Dealer to which he/she returns the unit, is proven to be defective. Replacement of a defective part will be made only within the applicable warranty period of: CAL2k - 12 (twelve) months E2K - 12 (twelve) months ECO - 12 (twelve) months From date of warranty registration. Not withstanding anything to the contrary herein, any labour involved in repairing or replacing a defective part will be charged at “dds” rates applicable at the time. 13.3 EXCLUDED (NON-WARRANTY) PARTS

Digital Data Systems Pty Ltd (“dds”) shall be entitled to treat this warranty as void unless the relevant warranty form (attached) is sent by registered post to “dds” by the Owner within 10 (ten) days from date of original installation. Regardless of warranty-status, the Owner shall be responsible for labour charges connected with periodic maintenance requirements involving visual and functional inspection of the systems. This warranty expressly covers the unit(s) “dds” supplied only and does not cover any other part, which may, as a consequence of the installation and operation of the unit, become damaged in any way whatsoever. Special attention should be given to the safe operation of the high-pressure vessels. The published rules are guidelines only and enclosed in this leaflet for the convenience of the User. The manufacturer is not responsible for any damages, consequential, indirect or direct, or loss whatsoever, resulting out of electrical shock, vessel explosion, or any other cause.

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 34

Page 38: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

14. SHIPPING AND RETURNS POLICIES 14.1 SHIPPING POLICY

Our shipping prices are not included in the product prices. All shipping costs are included in the invoice to the Dealer before we ship any products. If a Dealer prefers, he may use his own shipping company on his own account. A letter from the Dealer will be required to authorize us to use his shipping company on his account. We are experienced at shipping our products worldwide from very small to large parcels. We also ensure that you get the best rate possible. 14.2 RETURNS POLICY

Any return needs to be authorized in advance. Please contact your dealer directly to arrange authorization and to receive an authorization number. When returning any of our products, please use either of the following methods: For returns via the postal service: Digital Data Systems (Pty) Ltd P O Box 35872 Northcliff 2115 South Africa PLEASE ENSURE THAT YOU INCLUDE A ZERO VALUE AND A DESCRIPTION OF “REPAIR” FOR CUSTOMS PURPOSES. For returns via a shipping company: Digital Data Systems (Pty) Ltd 22 Arbeid Avenue Strijdompark Park Randburg South Africa PLEASE ENSURE THAT YOU INCLUDE A ZERO VALUE AND A DESCRIPTION OF “REPAIR” FOR CUSTOMS PURPOSES. All returns must be sent pre-paid and no collect shipments will be accepted. Please insure the parcel for its full value. Once we receive the returned products they are checked and tested for any manufacturing faults. If we determine that there is a manufacturing fault, we will refund the shipping or postal costs. 15. TECHNICAL SUPPORT & TRAINING

We offer Technical Support via telephone, fax or email and aim to respond within 24 hours during weekdays (Monday to Friday) and 48 hours over a weekend or 36 hours over a public holiday in South Africa. If required, our Technical Support Team at dds is available to travel to your location worldwide, to offer training and technical assistance. Fees for travelling and training can be requested by emailing [email protected]

Digital Data Systems Bomb Calorimeter System Catalogue www.cal2k.com 35

Page 39: dds - Jaagee Limitedjaagee.org/website/wp-content/uploads/2016/05/Bomb-Calorimeter-pdf… · 1.1 WHO WOULD USE A DDS BOMB CALORIMETER SYSTEM? ... DIN 51900 and BS 1016:105. 3.1.2

16. PRODUCT CATALOGUE DISCLAIMER

The material contained in this catalogue is for general information only. We take no responsibility for any loss incurred in reliance on the information contained herein. We reserve the right to alter product specifications and features without prior notice. You should always verify any information before you make any decisions by contacting us at [email protected]

17. CONTACT INFORMATION Visit our comprehensive website at www.cal2k.com for more product information, downloads, technical information and contact details of our local and international Distributors and Dealers.