Transcript of DC30-009 QS-550 Service Manual Rev G
Free-StandingTubestand
Service Manual
™
Manual Part No. DC30-009 Revision G
Model QS-550
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC. (QMI)
This Service Manual is incomplete without Operator’s Manual DC30-034
Copyright© 2010 QMI
Made in U.S.A.Quantum Medical Imaging, LLC2002-B Orville Drive NorthRonkonkoma, New York 11779 USAPhone: (631) 567-5800Fax: (631) 567-5074E-mail: [email protected]
Revision History
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC i
REVISION DATE TYPE OF MODIFICATIONA 12/7/00 Initial Release.
B 2/28/01 Minor modification made to installation of tube-stand on floor track procedure.
C 9/19/01 Incorporated ECO’s 0396, 0402, 0423.
D 6/13/02 Minor revisions to assembly and installation proce-dure. Incorporated ECO 0788.
E 3/7/03 Incorporated ECO’s 0945, 1057.
F 3/15/04 Incorporated ECO’s 1359, 1372.
G 3/22/10 Incorporated ECO 1404, 1691, 1889, added Tech-Vision option
PageNumber Rev
PageNumber Rev
PageNumber Rev
i - iv G
1-1 thru 1-6 G
2-1 thru 2-26 G
3-1 thru 3-6 G
4-1 thru 4-10 G
5-1 thru 5-6 G
Revision History
Revision G X-ray Tube Stand - Service Manualii Quantum Medical Imaging, LLC
This product has been classified as Class I, Type B by Underwriters Laboratories, Inc. Equipment not suitable for use in the presence of a flammable anesthetic mixture of air with oxygen or with nitrous oxide.
MEDICAL ELECTRICAL EQUIPMENTWITH RESPECT TO ELECTRIC SHOCK, FIRE,
MECHANICAL HAZARDS ONLY IN ACCORDANCEWITH UL 60601-1 AND CAN/CSA C22.2 NO. 601.1
98UA
EU Authorized Representative:Medizintechnik Berlin GmbHAltentreptower, Strasse 5912683 Berlin - Germany
Phone: +49-302-82 4726Fax: +49-302-82 6382
E-mail: [email protected]
Table of Contents
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC iii
CHAPTER 1 - SPECIFICATIONSPHYSICAL SPECIFICATIONS .................................................................1-3
SYSTEM SPECIFICATIONS ..............................................................1-3HANDGRIP SPECIFICATIONS ..........................................................1-3PERFORMANCE SPECIFICATIONS ....................................................1-3
ELECTRICAL SPECIFICATIONS ..............................................................1-4
SYSTEM OPTIONS ................................................................................1-4
SYSTEM OPERATING ENVIRONMENT ....................................................1-4
NON-OPERATING ENVIRONMENT .........................................................1-5
COMPATIBILITY STATEMENT ...............................................................1-5
CHAPTER 2 - ASSEMBLY & INSTALLATIONOVERVIEW ..........................................................................................2-3
ROOM CONSTRUCTION ..................................................................2-3INSTALLATION PRECAUTIONS ........................................................2-3CDRH CERTIFICATION ...................................................................2-3REQUIRED TOOLS AND MATERIALS ................................................2-3UNPACKING ...................................................................................2-4INSTALLATION INSTRUCTIONS ......................................................2-4
Floor Track Installation .............................................................2-4Vertical Column/Base Assembly Attachment ...............................2-4
INSTALLING TUBESTAND ON FLOOR TRACK ...................................2-6INSTALLING THE X-RAY TUBE, COLLIMATOR AND HANDGRIPS ...........................................................................2-7
Trunnion Configuration .............................................................2-7Tube Mounting Fork Configuration .......................................... 2-13Leveling the Collimator ........................................................... 2-14TRANSVERSE ARM COVERS INSTALLATION ............................. 2-14
VERTICAL COUNTERWEIGHT BALANCING INSTRUCTIONS ............. 2-16VERTICAL COLUMN ROTATION ADJUSTMENT ................................ 2-17ADJUSTING THE LONGITUDINAL BRAKE (ELECTROMAGNETIC LOCK) ......................................................... 2-17HORIZONTAL SID SWITCH ASSEMBLY INSTALLATION ................... 2-18VERTICAL SID SWITCH ASSEMBLY INSTALLATION ........................ 2-20TUBESTAND INPUT POWER .......................................................... 2-21CONNECTION OF THE HIGH-VOLTAGE CABLES/CABLE ROUTING ............................................................. 2-22HORIZONTAL SID SWITCH ACTUATOR/INDICATOR INSTALLATION ............................................................................ 2-24VERTICAL SID SWITCH ACTUATOR INSTALLATION ....................... 2-24VERTICAL DISTANCE SCALES ....................................................... 2-25
Bucky SID ............................................................................. 2-25Table SID .............................................................................. 2-25
ALIGNMENT OF THE SYSTEM ....................................................... 2-26Collimator Light Field Versus Actual X-Ray Field ........................ 2-26
Table of Contents
Revision G X-ray Tube Stand - Service Manualiv Quantum Medical Imaging, LLC
CHAPTER 3 - THEORY OF OPERATIONOVERVIEW .......................................................................................... 3-3
THEORY OF OPERATION ...................................................................... 3-3VERTICAL MOTION OPERATION ..................................................... 3-3LONGITUDINAL MOTION OPERATION ............................................ 3-3ANGULATION MOTION OPERATION ................................................ 3-3TRANSVERSE MOTION TRAVEL ...................................................... 3-4ALL LOCKS RELEASE OPERATION ................................................... 3-4AUTO-STOP OPERATION ................................................................ 3-4HAND GRIP DISPLAYS ................................................................... 3-5
CHAPTER 4 - SERVICE INSTRUCTIONSOVERVIEW .......................................................................................... 4-3
SERVICE MAINTENANCE ................................................................ 4-3Visual Inspection ..................................................................... 4-3Perform Every Six Months: ....................................................... 4-3Perform Annually: .................................................................... 4-4
TROUBLESHOOTING PROCEDURE .................................................. 4-4REPLACEMENT PARTS AND ORDERING INFORMATION .................... 4-5ORDERING INFORMATION ............................................................. 4-6
CHAPTER 5 - DIAGRAMModel QS-550 Tubestand Sample Configuration Diagram ....................... 5-3
Model QS-550 Tubestand Wiring Diagram - Manual Collimator ................ 5-4
Model QS-550 Tubestand Wiring Diagram - Automatic Collimator (Progeny Linear II) .............................................................................. 5-5
Model QS-550 Tubestand Wiring Diagram - Progeny Linear IV Automatic Collimator ........................................................................... 5-6
Chapter
1-1
SPECIFICATIONS
1
1-2
Chapter 1 Specifications
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 1-3
PHYSICAL SPECIFICATIONSRefer to Figure 5-1 in Chapter 5, DIAGRAMS for Model QS-550 Tubestand dimensions.
SYSTEM SPECIFICATIONS
• Column Height: 92.0 inches (2336.8 mm)
• Tube Mount: Fork or Trunnion
• Weight: 600 lbs. (272.0 kg)
• Collimator Compatibility: Manual, Selectable or Automatic, 10-20 lbs. (typi-
cal). Mounting to fork/tube housing requires four 1/4-20 or M6 x 16 bolts spaced on a 3.625" diame-ter circle
• X-ray Tube Compatibility: 3" or 4" rotating anode type, 30 - 40 lbs. (typical)
HANDGRIP SPECIFICATIONS
• Push Button Switches: Vertical lock release, Longitudinal lock release,
Angulation lock release, Transverse lock release, All Locks release, Auto-Stop
• LED Indicators: 40” (100 cm) SID, 72” (180 cm) SID
PERFORMANCE SPECIFICATIONS
• Tubestand Longitudinal Travel: 96.0 inches (1752.6 mm)
• Vertical Travel (x-ray tube): 60.0 inches (1524.0 mm)
• X-ray Tube Height: 72.5 inches (1848.2 mm) maximum to13.75 inches (349.3 mm) minimum
• Transverse Tube Travel: ±5 inches (127 mm) from centerline of table
• Transverse Travel Detent: Center position
• X-ray Tube Angulation: 240° (±120°)
• X-ray Tube Angulation Detents: 0°, 90°(CW), 90°(CCW)
• Column Rotation: 180° (±90°)
• Column Rotation Detents: 0°, 90°(L), 90°(R)
Chapter 1 Specifications
Revision G X-ray Tube Stand - Service Manual1-4 Quantum Medical Imaging, LLC
ELECTRICAL SPECIFICATIONSNote: The following specifications are for the tubestand only and do not apply to the x-ray tube or collimator. Refer to x-ray tube and collimator manufacturer’s documentation (supplied with the tubestand) for applicable specifications.• Locking System
Power: +24 VDC, 1.0 Ampere
• Mode of Operation: Continuous
• Power Supply Type: Double (or Reinforced) Insulated (must meet con-struction/performance criteria per UL 60601-1)
• Electrical Interfacing: Tubestand-mounted terminal strip for interfacing
+24 VDC and, when equipped with automatic colli-mator, tray sensing signals
SYSTEM OPTIONS• Auto-Tracking System, with Motorized Vertical Motion (QS-MAT)
• Fork Mount (replaces Trunnion Mount)
• Automatic collimator interface (QS-ACT-550)
• Floor Grout Kit
• Model QS-550 tubestands equipped with TechVision Hand Grips option:
• TechVision replaces the standard equipment hand grips to provide tube stand position controls and indicators and serves as a generator remote control panel. Refer to the table below for the appropriate documentation for the application:
APPLICATION TYPE OPERATOR MANUAL
SERVICE MANUAL
Quantum(Non-integrated) and DiRex(Integrated) DC30-048 DC30-049
Canon Integration DC30-102 DC30-049
NX-DR Integration DC30-113 DC30-049
Chapter 1 Specifications
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 1-5
SYSTEM OPERATING ENVIRONMENTNote: The tubestand is not designed for use in the presence of explosive or flammable gases as might be found in operating rooms.
Ambient Temperature: +10°C to +40°C
Relative Humidity: 30 to 75%, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa
NON-OPERATING ENVIRONMENT Note: Operation or prolonged storage temperatures below 32 degrees Fahrenheit should be avoided as it will cause lubricants to become more resistive to manually-operated motions.
Ambient Temperature: -18°C to +70°C
Relative Humidity: 20 to 95%, non-condensing
Atmospheric Pressure: 500 hPa to 1060 hPa
COMPATIBILITY STATEMENT
The Quantum Medical Imaging, LLC tubestand is compatible with all Quantum Medi-cal Imaging, LLC. manufactured tables, wall stands, and high-voltage x-ray genera-tors, and with other manufacturer’s equipment having equivalent means for indication of SID and perpendicularity.
Chapter 1 Specifications
Revision G X-ray Tube Stand - Service Manual1-6 Quantum Medical Imaging, LLC
THIS PAGE INTENTIONALLY LEFT BLANK
Chapter
2-1
ASSEMBLY &INSTALLATION
2
2-2
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-3
OVERVIEW
This chapter describes the steps required to assemble and install the Model QS-550 tubestand, including making all required electrical connections, and performing all required alignments and calibrations.
ROOM CONSTRUCTION
A conventional radiographic room construction should be used in the area where the tubestand is to be installed. Consult with your State Health Department or local Building Codes for specific radiation shielding requirements.
INSTALLATION PRECAUTIONS
Install the tubestand before installing the radiographic table.
Check that the space location, conduit arrangements and junction boxes are pre-pared for the permanent installation of the tubestand. Verify that the location and space allowed is sufficient for the full range of tubestand/Transverse Arm travel without hitting any walls or structures in the room (see Figure 5-1 for equipment dimensions). Note that the Transverse Arm in/out adjustment and the Vertical Column position relative to the rear wall are critical. Column rotation in the counterclockwise (right-hand) direction could result in the Tube Housing strik-ing the wall before the 90 degree index position is reached.
CDRH CERTIFICATION
The tubestand is not a CDRH certifiable product. However, it must be compatible and operate in conjunction with other components in the x-ray system so that the x-ray system performs in compliance with CDRH performance standards.
REQUIRED TOOLS AND MATERIALS
The following tools and materials are required to complete the assembly and installation procedures:
• Double-bubble level - two feet minimum length
• Tape Measure - eight feet minimum length
• Standard hand tools
• Alcohol - to clean the high-tension cable ends, tubehead, and high- tension transformer sockets
• Dow-Corning DC4 vapor-proofing compound, silicon insulating grease or equivalent - for packing x-ray tubehead and high-tension cable terminals
• Special film cassette (11" x 14" cassette with scribed centerline)
• Electric drill and drill bits
• Lubricant - white lithium grease
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-4 Quantum Medical Imaging, LLC
UNPACKING
Open crate or carton marked "packing list enclosed" first. Remove packing list and use it as a guide to open the remaining cartons. Do not dispose of packing material until packing list is matched with actual parts received. Should there be a shortage or damage, notify Quantum Medical Imaging, LLC.’s customer service department immediately. The manufacturer is relieved of any responsibility for damage during shipment after unit is picked up by the carrier.
INSTALLATION INSTRUCTIONS
When the tubestand is used with a vertical wall frame or wall stand, care should be taken during room planning and pre-installation to allow for the accurate align-ment of these components. Alignment accuracy must be within 2% of the SID used. After the tubestand Floor Track is aligned with the table receptor, the track cannot be adjusted transversely. Therefore, transverse alignment of the x-ray field to the image receptor in the vertical wall frame or wall stand must be made by adjusting the position of the image receptor within the wall stand or relocation of the entire wall stand.All mounting procedures must be approved before and checked after installation by the hospital or facility engineering department and a registered, professional, structural engineer. It is recommended that a leakage current measurement be taken following installation to verify system meets allowable levels.
To assemble and install the tubestand, proceed as follows:Floor Track InstallationThe area of the floor on which the Floor Track is to be mounted must be level in all directions. Shims may be used, however, grouting is recom-mended.
1. Position the Floor Track in the proper location for installation.
2. Using a bubble-level, check that the Floor Track is level both lon-gitudinally and transversely. If not, shims can be used to level track, however, grout is recommended.
3. Fasten the Floor Track to the floor using appropriate fastening hardware (capable of supporting 100 lb. pull-load) depending on the installation site (see Figure 2-1).
4. Install five (5) Hole Plugs in holes for floor mounting screws.
Vertical Column/Base Assembly Attachment1. Remove the shipping bolt located in the end of the Vertical Col-
umn shaft.
2. Remove eight (8) screws securing the left and right side Base Access Panels to the Base Assembly and remove panels.
3. Mount the Base Assembly to the Vertical Column and secure to Vertical Column Shaft using the supplied retaining ring (see Fig-ure 2-2).
NOTE
Examine all cartons and crates at time of delivery. If damage is apparent, have delivery driver write a "Dam-aged Ship-ment Note" on copies of freight bill, sign it, and file appropriate carrier claim. Should you dis-cover con-cealed damage, immediately notify the transporting agent and ask for an "Inspec-tion of Dam-age". Carrier will not accept concealed damage claim if filed after 15 days from date of receipt of merchandise
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-5
Figure 2-1. Floor Track Installation
Figure 2-2. Installing the Vertical Column Retaining Ring
MOUNTINGHOLES (5X)
RETAININGRING
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-6 Quantum Medical Imaging, LLC
INSTALLING TUBESTAND ON FLOOR TRACK
1. Remove the End Cap from one end of the Floor Track.2. At the end of Floor Track without End Cap, slide the Front and Rear Bearing
Bracket Assemblies into the Floor Track.3. Raise the Base and Vertical Column onto the Floor Track.4. Secure the Front and Rear Bearing Brackets to the Base Assembly (see Fig-
ures 2-3 and 2-4).
Figure 2-3. Bearing Bracket Locations
Figure 2-4. Vertical Column Angle Adjustment
FRONTBEARINGBRACKET
BEARING W/ECCENTRIC BUSHING (2X)
REARBEARING
BRACKET (2X)
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-7
CAUTION! When Bearing Brackets are removed,extreme care must be taken not to tip over theBase Assembly and Vertical Column. The BaseAssembly and Vertical Column should not be leftstanding unattended at any time. Re-install Frontand Rear Bearing Brackets immediately once thecolumn is raised.
1. Re-install the End Cap to the end of the Floor Track using four (4) 5/16-18 x 1/2" hex head bolts, lockwashers and flat washers.
2. Using a two-foot (minimum) double-bubble level, placed on the front of the Vertical Column, check that the Vertical Column is plumb. If not, adjustment can be made by rotating the two (2) Vertical Angle Adjustment Bearings with eccentric bushings (see Figure 2-4). Loosen the two (2) eccentric bearings mounted to the rear of the Base Assembly; rotate the bearings to adjust. Re-tighten the bearings after the column angle adjustment is made.
INSTALLING THE X-RAY TUBE, COLLIMATOR AND HANDGRIPS
CAUTION! The tube angulation lock is not amechanical lock; electrical power must be appliedto hold angulation. During installation, the x-raytube and collimator must be held to prevent free-turning. Mount the collimator per instructionssupplied by the collimator manufacturer.
The system is supplied with either a Trunnion type x-ray tube mounting bracket or a Tube Mounting Fork. Refer to the appropriate paragraph that follows corre-sponding with the type of x-ray tube mounting bracket used on your system. On systems equipped with optional TechVision Hand Grips, refer to TechVision Ser-vice Manual for complete hand grips mounting instructions.
Trunnion Configuration
1. Install the Trunnion Assembly onto the Transverse Arm using four (4) 1/4-20 x 2" hex head bolts and lockwashers (see Figures 2-5 and 2-6). There are two sets of four (4) Trunnion mounting holes that can be used as shown in Figure 2-5. The upper set of holes is used when mounting 3-inch x-ray tubes. The lower set of holes is used for mounting 4-inch x-ray tubes.
2. Remove top half of Trunnion Rings.
3. Place x-ray tube onto lower half of Trunnion Rings and re-mount upper half.
4. Install Trunnion Handle on Trunnion Assembly using two (2) 1/4-20 x 1/2" button head socket screws (see Figure 2-7).
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-8 Quantum Medical Imaging, LLC
5. Check trunnion handle locking. If too loose, remove Trunnion Handle and rotate each Trunnion Cam clockwise a few turns. If too tight, remove Trunnion Handle and rotate each Trunnion Cam counterclockwise a few turns. Re-install Trunnion Handle and re-check locking. Repeat until properly adjusted.
Figure 2-5. Trunnion Mounting Holes
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-9
Figure 2-6. Mounting the Tube and Trunnion Assembly
Figure 2-7. Trunnion Handle Installed
1/4-20 X 2" HEX HEAD BOLT AND LOCK-WASHER (4X)
TUBE SUPPORT BRACKET
COLLIMATOR MOUNTING PLATE
TRUNNION ASSEMBLY
1/4-20 X 1/2" BUTTON HEAD SOCKET SCREW (2X)
TRUNNIONHANDLE
TRUNNIONCAM
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-10 Quantum Medical Imaging, LLC
6. Remove the Collimator Mounting Plate from the collimator.
7. Attach the Collimator Mounting Plate to the Tube Support Bracket and x-ray tube. Use required spacer(s) supplied with collimator between x-ray tube and Tube Support Bracket.
8. Mount Tube Support Bracket onto Trunnion using four (4) 1/4-20 hex head bolts.
9. Mount Handgrip Assembly on Tube Support Bracket (in front of the x-ray tube) using four (4) 1/4-20 x 1" hex head bolts (see Figure 2-8).
Figure 2-8. Installing the Handgrips
10. Rotate the Handgrips/tube 120° (i.e., nearly upside down). Install collimator on Collimator Mounting Plate using Collimator Mounting Ring (see Figure 2-9).
11. Affix supplied rubber bumper to Angulation Plate in location where collimator contacts Angulation Plate when in fully rotated position.
1/4-20 X 1" HEX HEAD BOLT (4X)
HANDGRIPASSEMBLY
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-11
12. Affix Trunnion angulation labels (scales and pointers) as follows:
a. Affix pointers on x-ray tube.
b. Place angulation scales on Trunnion Mounting Rings so that a scale can be viewed from either side of the tubestand, and with the pointers (on tube) pointing to 0° on angulation scales.
13. Route Handgrip cable (from terminal block TB1) to D-connector on rear side of the Handgrip Assembly and connect (see note).
Note: When installing Techvision Option, route the Ethernet Cable from the switch or hospital LAN to J2 at rear of TechVision, the Generator Logic Board cable to J3, and the cable from TB1 on the Vertical Column to J5 (see Figure 2-9b). Refer to the TechVision Service Manual (DC30-049) for additional information on install-ing TechVision).
14. Connect collimator cables as required (refer to collimator manual and to Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS).
Figure 2-9. Collimator Installed
COLLIMATOR MOUNTING RING
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-12 Quantum Medical Imaging, LLC
Figure 2-9a. TechVision Option
Figure 2-9b. TechVision Electrical and Communications Connections
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-13
Tube Mounting Fork Configuration
1. Position the Tube Support Bracket on the Tube Arm Extension, with the top two slots centered on mounting holes, and secure using four (4) 1/4-20 x 2" hex head bolts and lockwashers (see Figure 2-10). Remove Collimator Mounting Plate from collimator and install Collimator Mounting Plate (with appropriate spacers, supplied with collimator) and x-ray tube on Tube Fork.
2. Mount the Handgrip Assembly (in front of the x-ray tube) using four (4) 1/4-20 hex head bolts (see Figure 2-8).
3. Install collimator on Collimator Mounting Plate using Collimator Mounting Ring (see Figure 2-9).
4. Route Handgrip cable (from terminal block TB1) to D-connector on rear side of the Handgrip Assembly and connect.
5. Connect collimator cables as required (refer to collimator manual and to Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS).
Figure 2-10. Mounting the Tube Fork
1/4-20 X 2" HEXHEAD BOLTS (4X)
TUBE FORK
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-14 Quantum Medical Imaging, LLC
Leveling the Collimator
Place a level on top or bottom of the collimator (from front to back). If not level, proceed as follows:
1. Loosen four (4) 1/4-20 x 2" hex head bolts securing Trunnion (or Tube Mounting Fork) to Transverse Arm.
2. Adjust set screws as required to level collimator (see Figure2-11).
3. Re-tighten four (4) 1/4-20 x 2" hex head bolts.
Figure 2-11. Collimator Leveling Adjustment Set Screws
TRANSVERSE ARM COVERS INSTALLATION
Refer to Figures 2-12 and 2-13 and proceed as follows:
1. If not already installed, position the Transverse Arm Inner Cover as shown in Figure 2-12 and secure to Transverse Arm using six (6) 8-32 x 1/4" button head socket screws.
2. Carefully position the Transverse Arm Outer Cover over the Transverse Arm Inner Cover. The holes in the four (4) mounting tabs attached to the inside bottom edge of the Transverse Arm Outer Cover should be aligned with the holes in the four (4) mounting blocks on the underside of the Transverse Arm Assembly.
3. Position the Transverse Arm Lower Cover as shown in Figure 2-13 and secure it to the Transverse Arm using four (4) 6-32 x 3/4" flat head screws.
COLLIMATORLEVELING SET
SCREW (4X)
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-15
Figure 2-12. Installing the Transverse Arm Inner Cover.
Figure 2-13. Installing the Lower Transverse Arm Cover
8-32 X 1/4" BUTTON HEAD SOCKET SCREW (6X)
TRANSVERSE ARM INNER COVER
6-32 X 3/4" FLAT HEAD SCREW (4X)
TRANSVERSE ARM OUTER COVER
TRANSVERSE ARM LOWER COVER
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-16 Quantum Medical Imaging, LLC
VERTICAL COUNTERWEIGHT BALANCING INSTRUCTIONS
1. Verify that the vertical counterweight is balanced properly, so that it provides a smooth up and down motion of the tube head.
2. If additional weights are necessary, add trim weights as required. There are two configurations of the Vertical Column; one has the trim weight access opening located in the front side, the other has it located in both the front and rear sides). To add trim weight to rear side opening, remove access opening cover, add required trim weight(s), and re-install cover. To add trim weight to front side opening, proceed as follows:
a. Bring the tube head down to minimum vertical height position.
b. Remove top two Column Center Cover mounting screws and carefully tilt cover forward enough to gain access to trim weight opening (see Figure 2-14).
c. Bring the tube head up until the top of the counterweight is in level with bottom of opening.
d. Add trim weight as required.
e. Re-install Column Center Cover mounting screws.
Figure 2-14. Trim Weight Access Opening (Front Side of Vertical Column)
NOTE
Be careful not to drop trim weights when adding through opening in Ver-tical Column.
ACCESS OPENING
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-17
VERTICAL COLUMN ROTATION ADJUSTMENT
Rotate the Vertical Column to feel the mechanical indexing. If too loose or too tight, the tension of indexing is adjusted by loosening or tightening two (2) 1/4-20 socket head screws, one on each side of the Vertical Column near the base (see Figure 2-15).
Figure 2-15. Vertical Column Rotation Adjustment Screws
ADJUSTING THE LONGITUDINAL BRAKE (ELECTROMAGNETIC LOCK)
The Longitudinal Brake Assembly is mounted inside the Base Assembly (on the right side when facing the front of column). To adjust height, proceed as follows:
1. Loosen two (2) 1/4-20 x 1” hex head bolts securing the Longitudinal Brake Assembly to the Base Assembly (see Figure 2-15).
2. The space between electromagnetic lock and contacting part of the Floor Track should not exceed 20/1000" (ideally between 0.010” and 0.020”). Adjust electromagnetic lock brake height as required.
3. When electromagnetic lock height is correct, re-tighten two (2) 1/4-20 x 1” hex head bolts.
COLUMN ROTATION ADJUSTMENT SCREW (2X)
LONGITUDINAL BRAKE ASSEMBLY
1/-20 x 1 HEX BOLTSFOR MOUNTING LONGITUDINAL
BRAKE ASSEMBLY
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-18 Quantum Medical Imaging, LLC
HORIZONTAL SID SWITCH ASSEMBLY INSTALLATION
Figure 2-16. Horizontal SID Switch Installation
The Horizontal SID Switch Assembly mounts inside the Base Assembly (on the left side when facing front of column) as shown in Figure 2-16. Proceed as follows:
1. Loosely install Horizontal SID Switch Assembly on rear wall of Base Assembly using two (2) 10-32 x 1/2" phillips pan head screws, flat washers and kep nuts.
2. Using a Switch Actuator placed on the Floor Track as a reference (do not peel off adhesive backing at this time), adjust the height of the switch so that when it contacts the actuator, the switch activates (clicks).
3. When the switch height is correct, tighten the two (2) 10-32 x 1/2" phillips pan head screws.
4. Remove terminal strip cover from back of tubestand.
HORIZONTALSID SWITCHASSEMBLY
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-19
5. Connect wires from Horizontal SID Switch Assembly and Longitudinal Brake Assembly to terminal block TB1 as follows:
6. If tubestand is being interfaced with automatic collimator, connect four (4) wires from Horizontal SID Switch Assembly as follows:
7. Re-install terminal strip cover on back of tubestand.
8. Install left and right Base Access Panels on the Base Assembly.
Table 2-1. Manual Collimator TB1 Wiring
WIRE DESCRIPTION WIRE COLOR
CONNECTION ON TB1
LONGITUDINAL MAGNET RED TB1-6
LONGITUDINAL MAGNET BLACK TB1-2
40" SID WALL SWITCH YELLOW TB1-1
40" SID WALL SWITCH ORANGE TB1-4
72" SID WALL SWITCH GREEN TB1-1
72" SID WALL SWITCH BROWN TB1-5
40" SID TABLE SWITCH RED TB1-12
40" SID TABLE SWITCH BLACK TB1-4
Table 2-2. Automatic Collimator TB1 Wiring
WIRE DESCRIPTION WIRE COLOR
CONNECTION ON TB1
LONGITUDINAL MAGNET RED TB1-6
LONGITUDINAL MAGNET BLACK TB1-2
40" SID WALL SWITCH BLUE TB1-18
40" SID WALL SWITCH PURPLE TB1-19
72" SID WALL SWITCH WHITE TB1-16
72" SID WALL SWITCH GRAY TB1-19
40" SID TABLE SWITCH BROWN TB1-17
40" SID TABLE SWITCH ORANGE TB1-20
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-20 Quantum Medical Imaging, LLC
VERTICAL SID SWITCH ASSEMBLY INSTALLATION
The Vertical SID Switch Assembly mounts to the lower section of the Cable Clamp (on the anode side) as shown in Figure 2-17. Proceed as follows:
Figure 2-17. Vertical SID Switch Installation
1. Install two (2) Cable Clamp lower sections on each side of Tube Arm Exten-sion using two (2) 1/4-20 x 2" socket head cap screws.
2. Loosely mount the Vertical SID Switch Assembly to the lower section of anode-side Cable Clamp using two (2) 10-32 x 1/4" socket head cap screws.
3. Using a switch actuator placed on the Vertical Column as a reference (do not peel off adhesive backing at this time), adjust the height of the switch so that when it contacts the actuator, the switch activates (clicks).
4. When switch height is correct, tighten the two (2) 10-32 x 1/4" socket head cap screws.
5. On SID switch cable (from terminal block TB1), connect red and black wires to NO and COM terminals, respectively, on Vertical SID Switch Assembly switch S7. On automatic collimator systems, connect brown and orange wires to NO and COM terminals, respectively, on Vertical SID Switch Assembly switch S10.
VERTICAL COLUMN
SID SWITCH CABLE
VERTICAL SID SWITCH ASSEMBLY
CABLE CLAMP LOWER HALF
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-21
TUBESTAND INPUT POWER
The tubestand requires a +24 VDC power input connection. Refer to Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS and proceed as follows:
1. Locate 25’ power input cable provided with system.
2. Connect power input cable as follows:
a. Connect end of 25’ power cable with fast-on terminals to tubestand terminal block TB1-1 (+24 VDC) and TB1-2 (24 VDC RTN). Terminal block TB1 is located on the rear side of the Vertical Column.
a. Connect opposite end of cable (end without fast-on terminals) to power source (e.g., x-ray generator) +24 VDC and 24 VDC RTN ter-minals. The input power source(s) must be double (or reinforced) insulated and meet construction/performance criteria per UL 2601-1.
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-22 Quantum Medical Imaging, LLC
CONNECTION OF THE HIGH-VOLTAGE CABLES/CABLE ROUTING
To make the required electrical connections, refer to the appropriate system inter-connection diagram (see Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS) and proceed as follows:
1. Connect the Anode and Cathode high-voltage cables to the x-ray tube hous-ing. Allow sufficient cable drape (see Figure 2-18) in order to angulate the x-ray tube 90° clockwise or counterclockwise.
Figure 2-18. Cable Clamping and Management
2. Insert the high-voltage cables along with the cables from the collimator, Handgrip Assembly, and Vertical SID Switch Assembly into corrugated (slit) tubing provided with system (the tubing will need to be cut to required lengths). Use tie-wraps to secure cable in slit tubing as required.
CABLECLAMP (2X) TERMINAL
STRIP COVER
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-23
3. Install a High Voltage Cable Strain Relief on each high voltage cable attach-ment to x-ray tube horns. (see Figure 2-19).
Figure 2-19. Installing the High Voltage Cable Strain Reliefs
4. Fasten the corrugated tubing into the Cable Clamps (see Figure 18) on each side of the Vertical Trolley by installing upper halves of Cable Clamps to lower halves using two (2) 8-32 x 1” socket head cap screws.
5. From the Terminal Strip Assembly, route the tubestand cables along with the +24 VDC power input cable, to the x-ray generator.
HIGH VOLTAGECABLE
STRAIN RELIEFS(2X)
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-24 Quantum Medical Imaging, LLC
HORIZONTAL SID SWITCH ACTUATOR/INDICATOR INSTALLATION
Install the Horizontal SID switches actuators and indicator decals as follows:
1. Position the tubestand at 40" and 72" SID and mark each position on Floor Track.
2. Clean marked area with alcohol and secure actuator in marked location using double-sided tape. Check switch height to ensure activation at 40" and 72" SID.
3. Affix 40” and 72” (wall) SID indicators on Floor Track at each SID position (see Figure 2-20).
Figure 2-20. Installing SID Indicator Labels
VERTICAL SID SWITCH ACTUATOR INSTALLATION
The Vertical SID Switch actuating arm faces towards the Vertical Column.
1. Attach the Vertical SID Indicator Bracket to the Vertical Trolley. This is done by inserting the end of the bracket with the square cut-out and slot into the opening between the trolley and Vertical Column. The square opening in the bracket is then hung onto the inside of the trolley. Pull down on bracket to ensure it is fully seated in slot.
2. Bring the Vertical Trolley down to the desired 40" SID mark on the Vertical Column.
3. Mark the location for the Vertical SID Switch actuator.
Chapter 2 Assembly & Installation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 2-25
4. Clean marked area with alcohol and secure actuator in marked location using double-sided tape.
5. Check switch height to ensure activation at 40" SID.
VERTICAL DISTANCE SCALES
The Vertical Distance Scales mount on the front side of the Vertical Column (see Figure 2-21). The Vertical Distance Scales, calibrated in either centimeters (met-ric system) or inches (SAE system), indicate the focal spot-to-bucky and focal spot-to-tabletop distances
Bucky SID
1. Set tube position for 40" (100 cm) bucky SID.
2. Affix label on Vertical Column so that the 40" (100 cm) mark lines up with the arrow in the bucky SID window of the SID Indicator Bracket.
Table SID
1. Set tube position for 40" (100 cm) tabletop SID.
2. Affix label on Vertical Column so that the 40" (100 cm) mark lines up with the arrow in the table SID window of the SID Indicator Bracket.
Figure 2-21. Vertical SID Scales Mounted on Vertical Column
NOTE
Bucky and table SID measure-ments and markings are made sepa-rately and inde-pendently of each other.
SID INDICATOR BRACKET
"TABLE" VERTICAL SID INDICATOR
"BUCKY"VERTICAL SID
INDICATOR
Chapter 2 Assembly & Installation
Revision G X-ray Tube Stand - Service Manual2-26 Quantum Medical Imaging, LLC
ALIGNMENT OF THE SYSTEM
Once installed, the tubestand must be aligned with other components of the x-ray system, such as a radiographic table and/or vertical wall stand. If both a radio-graphic table and vertical wall stand are used, align the table first, then the verti-cal wall stand. Alignment instructions are provided in the installation and operation manual provided with each component.
Refer to the collimator manual for additional alignments and performance verifi-cations associated with beam limiting device.
Collimator Light Field Versus Actual X-Ray Field
1. Using any size film with cassette, set the proper technique in the x-ray generator.
2. Make an exposure and develop the film to verify actual x-ray field versus collimator light field alignment is within required specifica-tions.
3. Use collimator horizontal cross light to align the table. Rotate the x-ray tube to check the pitch adjustment as required.
4. Verify vertical movement and alignment with bucky stand or cas-sette holder using collimator light.
NOTE
For this test, it is very impor-tant to use the collimator light for aligning the table and verti-cal wall stand.
Chapter
THEORYOF OPERATION
3-1
3
3-2
Chapter 3 Theory of Operation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 3-3
OVERVIEW
The Model QS-550 tubestand is a free-standing x-ray tube support device (i.e., does not require wall and/or ceiling mounting) that provides a wide range of tube positions through longitudinal, vertical, rotational, and transverse tubestand motions. The Handgrip Assembly is designed for operator convenience with ergonomically posi-tioned push button switches that enable tubestand movement when pressed.
The following paragraphs provide detailed theory of operation of the various tube-stand functions. See Figure 5-2, 5-3, 5-4 in Chapter 5, DIAGRAMS for system sche-matics.
THEORY OF OPERATION
VERTICAL MOTION OPERATION
The vertical motion of the x-ray tube is controlled by a fail-safe type "vertical brake"; when power is removed from the system, the brake remains mechanically locked. A +24 VDC input, typically provided from the radiographic system’s x-ray generator, is applied through connector TB1-1 and switch S1-2 to Vertical Lock Switch S2-1. When S2 is pressed, the +24 VDC is applied to electromagnetic Ver-tical Lock Coil L1, releasing the brake (i.e., magnetic lock) and enabling vertical positioning of the x-ray tube. When Vertical Lock Switch S2 is released, +24 VDC is immediately removed from L1 and the vertical brake (magnetic lock) is acti-vated inhibiting x-ray tube vertical motion.
LONGITUDINAL MOTION OPERATION
The longitudinal motion of the x-ray tube (i.e., entire Vertical Column) is con-trolled by a "longitudinal brake" located in the tubestand’s Base Assembly. The +24 VDC input is applied through connector TB1-1 and switch S1 to Longitudinal Lock Switch S3-1. When S3 is pressed, the +24 VDC is removed from electro-magnetic Longitudinal Lock Coil L3, releasing the brake (i.e., magnetic lock) and enabling longitudinal positioning of the x-ray tube. When push button S3 is released, +24 VDC is immediately applied to L3 and the longitudinal brake (mag-netic lock) is activated inhibiting x-ray tube longitudinal motion.
ANGULATION MOTION OPERATION
The angulation motion of the x-ray tube (i.e., clockwise and counter-clockwise rotation of the x-ray tube) is controlled by an "angulation brake" located in the tubestand’s Transverse Arm Assembly. The +24 VDC input is applied through connector TB1-1 and switch S1 to Angulation Lock Switch S4-1. When S4 is pressed, the +24 VDC is removed from electromagnetic Angulation Lock Coil L2, releasing the brake (i.e., magnetic lock) and enabling rotation of the x-ray tube. When push button S4 is released, +24 VDC is immediately applied to L2 and the angulation brake (magnetic lock) is activated inhibiting x-ray tube angulation motion.
Chapter 3 Theory of Operation
Revision G X-ray Tube Stand - Service Manual3-4 Quantum Medical Imaging, LLC
TRANSVERSE MOTION TRAVEL
The transverse motion of the x-ray tube (i.e., cross-table movement of the x-ray tube) is controlled by a "transverse brake" located in the tubestand’s Transverse Arm Assembly. The +24 VDC input is applied through connector TB1-1 and switch S1 to Transverse Lock Switch S5-1. When S5 is pressed, the +24 VDC is removed from electromagnetic Transverse Lock Coil L4, releasing the brake (i.e., magnetic lock) and enabling transverse motion of the x-ray tube. When push button S5 is released, +24 VDC is immediately applied to L4 and the transverse brake (magnetic lock) is activated inhibiting x-ray tube transverse motion.
ALL LOCKS RELEASE OPERATION
The All Locks Release button provides the operator with a single button that, when pressed, enables x-ray tube positioning in the vertical, longitudinal, rota-tional, and transverse directions simultaneously. The +24 VDC input is applied through connector TB1-1 to normally-closed switch S1-2 and S1-5. With the switch closed, the +24 VDC power input is routed through contact S1-1 to the Transverse Lock Switch S5-1, Longitudinal Lock Switch S3-1, and Angulation Lock Switch S4-1 and to contact S1-4, which is not connected. The +24 VDC is then routed through switches S5, S3 and S4 to the Transverse, Longitudinal, and Angulation Lock Coils L4, L3, and L2, respectively, which activate the locks for each of these motions. The fail-safe Vertical lock remains locked since the +24 VDC is not connected in this switch state.
When switch S1 is pressed, the +24 VDC power input is routed to switch contacts S1-3 and S1-4. In this state, +24 VDC input power is removed from the Trans-verse Lock Switch S5-1, Longitudinal Lock Switch S3-1, and Angulation Lock Switch S4-1 disabling the locks (thus enabling transverse, longitudinal, and angu-lation x-ray tube motions). The +24 VDC power input is routed through switch contact S1-6 to the fail-safe Vertical Lock, releasing the brake, thereby enabling vertical motion of the x-ray tube.
AUTO-STOP OPERATION
The Auto-Stop function is provided to enable quick longitudinal positioning of the x-ray tube at the 40" (or 100 cm) and 72 "(or 180 cm) SID positions. This func-tion requires that the operator simultaneously press the Longitudinal Lock Release button (switch S3) and the Auto-Stop button (switch S6). The circuit operation is as follows:
The +24 VDC input is applied through connector TB1-1 to 40" Table SID switch S7, 40" Wall SID switch S8, and 72" Wall SID switch S9. If the tubestand’s Verti-cal Column is moved over a 40" or 72" SID actuator on the floor track, it closes the corresponding SID switch, which routes the +24 VDC to the Longitudinal Lock Coil L3. The longitudinal brake (magnetic lock) is activated, locking the x-ray tube in the current SID position. To re-activate the Auto-Stop function, both the Longitudinal Lock Release and Auto-Stop buttons must be pressed.
Chapter 3 Theory of Operation
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 3-5
HAND GRIP DISPLAYS
The analog-type Hand Grip display includes an analog angle indicator for display-ing the rotation angle (i.e., "angulation") of the x-ray tube, and two SID light-emitting diodes (LEDs). When the Vertical Column is positioned over the 72" SID actuator, located on the floor track, +24 VDC is applied from connector TB1-1 through 72" Wall SID switch S9, to the 72 Inch LED, lighting the LED. Similarly, when the Vertical Column or the Vertical Trolley is positioned over the 40" SID actuator, +24 VDC is applied from connector TB1-1 through 40" Wall SID switch S8 or 40" Table SID switch S7, respectively, to the 40 Inch LED, lighting the LED.
Optionally, TechVision Hand Grips replace the standard analog-type hand grips on systems equipped with this feature. Refer to the TechVision Service Manual for theory of operation.
Chapter 3 Theory of Operation
Revision G X-ray Tube Stand - Service Manual3-6 Quantum Medical Imaging, LLC
THIS PAGE INTENTIONALLY LEFT BLANK
Chapter
4-1
SERVICEINSTRUCTIONS
4
4-2
Chapter 4 Service Instructions
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 4-3
OVERVIEW
This chapter is designed to assist the service technician in maintaining the smooth operation of the tubestand. The tubestand will function reliably when maintained according to the instructions provided in this chapter. Safe equipment performance requires the attention of service personnel who are specifically trained and experi-enced with medical x-ray apparatus. Applicable preventive maintenance or any repair service should be performed by these skilled individuals. Failure to follow manufac-turer’s or service personnel’s recommendations may result in serious injury.
WARNING! Personnel engaged in maintenanceactivities should exercise normal caution and carewhile working with electro-mechanical equip-ment. Before removing or opening any electricalpower panels or covers, verify that the incomingpower supply is turned OFF. In the event mainte-nance procedures require power to be supplied tothe unit, extreme care MUST be exercised toinsure the safety of service or other personnel inthe area. Verify equipment is properly groundedbefore attempting any electrical operation.
SERVICE MAINTENANCE
WARNING! Always disconnect the equipmentfrom the main power supply prior to any cleaning.
Visual Inspection
The following inspection and adjustment procedures are recommended to maintain the system in its original operating condition.
Perform Every Six Months:
• Check for evidence of loose hardware or loose wires
• Check all bearings and bearing surfaces for cleanliness and corrosion
• Clean Bearing Tracks, Floor Track
• Conduct a general inspection for worn or damaged parts
• Check for proper counterbalancing and smoothness of vertical (up/down) tubestand motion
• Verify that all ground conductors are properly and securely installed and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying
Chapter 4 Service Instructions
Revision G X-ray Tube Stand - Service Manual4-4 Quantum Medical Imaging, LLC
Perform Annually:
• Lubricate tube rotation detents, angulation plate, transverse assembly bearing wheels, column rotation detents using lithium grease
TROUBLESHOOTING PROCEDURE
The following table provides instructions for troubleshooting problems that could be encountered on Model QS-550 tubestand.
Table 4-1. Troubleshooting Information
Problem/Indication Possible Cause(s) Remedial Action
System will not oper-ate
No Power to the tubestand 1. Verify +24 VDC present across terminal block TB1-1 and TB1-2.
40" SID or 72" SID indicators not function-ing
• Loss of +24 VDC power.
• Horizontal Switch (40" WALL SID or 72" WALL SID) failure
• SID lamp failure
1. Check for +24 VDC on hand- grips cable connector A3P1-24.
2. Check Horizontal Switch (40" WALL SID or 72" WALL SID).
3. Replace SID lamp.
Chapter 4 Service Instructions
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 4-5
REPLACEMENT PARTS AND ORDERING INFORMATION
Table 4-2 provides a list of replaceable parts for the tubestand. Figures 4-1 through 4-10 show the locations of the replaceable parts in the system.
Table 4-2. Replaceable Parts
Item Description Part Number
Qty Used on System
1 Magnet, Small (Transverse) EL58-007 1
2 Magnet, Large (Longitudinal, Angula-tion)
EL58-006 2
3 Fail-safe Brake EL80-002 1
4 SID Switch Actuator ME10-218 3 (manual col-limator)
4 (automatic collimator)
5 SID Switch EL50-006 2
6 SID Indicator Light, Blue EL31-003 2
7 Push-button Switch, Transverse EL50-002-1 1
8 Push-button Switch, Longitudinal EL50-002-2 1
9 Push-button Switch, All Locks Release EL50-002-3 1
10 Push-button Switch, Auto-Stop EL50-002-4 1
11 Push-button Switch, Angulation EL50-002-5 1
12 Push-button Switch, Vertical EL50-002-6 1
13 Bearing, R6ZZ ME40-018 4
14 Bearing, 1628-ZZ ME40-017 8
15(not shown)
+24 VDC Power Cable EL75-005 35 ft.
16(not shown)
Horiz. SID Switch, DPDT (automatic collimator systems only)
EL50-019 2
Chapter 4 Service Instructions
Revision G X-ray Tube Stand - Service Manual4-6 Quantum Medical Imaging, LLC
ORDERING INFORMATION
To order replacement parts for the tubestand, contact the Service Department at;
Quantum Medical Imaging, LLC.2002-B Orville Drive NorthRonkonkoma, New York 11779-7661 USAPhone: (631) 567-5800 (x2)Fax: (631) 567-5074e-mail: [email protected]
When ordering replacement parts, supply the following information:
• Model and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in Table 4-2, a complete descrip-tion of the part, including its function and location should be provided with the model number and serial number of the unit. Use only original replacement parts, as supplied or recommended by Quantum Medical Imaging, LLC
Figure 4-1. Model QS-550 Tubestand Parts Location Diagram
21
(ANGULATIONMAGNET)
Chapter 4 Service Instructions
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 4-7
Figure 4-2. Model QS-550 Tubestand Parts Location Diagram
Figure 4-3. Model QS-550 Tubestand Parts Location Diagram
2(LONGITUDINAL
MAGNET)
3
Chapter 4 Service Instructions
Revision G X-ray Tube Stand - Service Manual4-8 Quantum Medical Imaging, LLC
Figure 4-4. Model QS-550 Tubestand Parts Location Diagram
Figure 4-5. Model QS-550 Tubestand Parts Location Diagram
4
5NOTE: Vertical SID switch and actuator are shown in this figure; the Horizontal SID switch, located in the Base Assembly, and longitudinal actuators have the same respective part numbers.
67
8
9
10
11
12
Chapter 4 Service Instructions
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 4-9
Figure 4-6. Model QS-550 Tubestand Parts Location Diagram
13
14
Chapter 4 Service Instructions
Revision G X-ray Tube Stand - Service Manual4-10 Quantum Medical Imaging, LLC
THIS PAGE INTENTIONALLY LEFT BLANK
Chapter
5-1
DIAGRAMS
5
5-2
Chapter 5 Diagrams
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 5-3
Figure 5-1. Model QS-550 TubestandSample Configuration Diagram
DC80-017 Rev. D
9. DIMENSION WITH AUTO-TRACK (Q-TRACK) OPTION. 10. WHEN ACTIVATED, RECEPTOR AND TUBE WILL
AUTOMATICALLY TRACK RECEPTOR TRAVEL. (Q-TRACK OPTION ONLY).
40"[1016]
92.00[2337]
NOTE 893.38
NOTE 9
72.75 MAX[1848]
10' [3048]
72.75 MAX.[1848]
13.00 MIN.[330]
52.34 [1329] ±5" [127]NOTE 1
55.84±5" NOTE 9
NOTE 2
DELUXE FLOOR MOUNTED
TUBESTANDQS-550
QUIET-LIFT TABLE,REFERENCE
ONLY
FOCALSPOT
13.00 MIN[330]
OF TUBESTANDROTATION
45.50 [1156] ±5" [127]NOTE 1
49.00±5" NOTE 9
5.00 [127]NOTE 4
8.50 MIN. NOTE 9
4" [102]
17.7 [450]C.G. LOCATION
32.5[825]
C.G. NOTE 7
C.G. NOTE 7
6 " [165]1/2
18 " [470]1/2
AUTO-TRACKENABLE SWITCHNOTE 10
AUTO-TRACKENABLE SWITCHNOTE 10
VERTI-QWALL STANDREFERENCE ONLY
7.92
40" F.F.D.[1016] REF
72" F.F.D.[1829] REF
FILMPLANE
(5) .750 [19] DIAMETER MOUNTING HOLES 28" [711] APART
Chapter 5 Diagrams
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 5-4
Figure 5-2. Model QS-550 Tubestand Wiring Diagram- Manual Collimator
ASSEMBLYHANDGRIP
OR
ANG
E
ASSEMBLYTUBESTAND
RED
YEL
LOW
COM
BLACK
GR
EEN
OR
AN
GE
COM
S9
BLA
CK
GR
AY
PURPLE BLU
E
YE
LLO
W
RE
D
GR
EEN
/YE
LLO
W
GROUND
WH
ITE
STUD
BR
OW
N
GROUND STUD
BR
OW
N
BLA
CK
BLA
CK
BRO
WN
BLA
CK
LOCK
RE
D
LOCK
RE
D
LOCK
BLACKRED
OR
ANG
E
LOCK
RE
D
+24 VDC RTN+24 VDC
GENERATOR GROUND
FROMGENERATOR
BLACK
RED
RED
COM
(AY20-020)
P1
J1
PU
R
OR
N
BLU YEL
RE
D
BLK
GR
N
WH
T
BR
N
NO
NC
NO
NC
NO
NC
FROMGENERATOR
24 VAC24 VACGROUND
COLLIMATORWHITEBLACKGREEN
CO
MN
O
CO
MN
C
CO
MN
C
CO
MN
C
COM
COM
NC
NO
NC
NO
GR
Y
1515
12
2020
181919 18
16171617
3
12
3
12
3
12
6
4
3
5
12
21
12
12
131414 13
111212 11 8
810 9910
7 67 6 1
15 45 4
2323
1111
1212
1313
1414
1515 2 4
51010 53
3
88
9
26 4
1
79 1
76
+24V
VERT SW
40" TABLE SID SW
40" WALL SID SW
72" WALL SID SW
+24V
ALL LOCKS REL
AUTOSTOP SW
TRANS SW
S6
1N4005
D5
ANGLE SWLONG SWS3 S4
LED 1 (LEFT)40 INCHES
LED 2 (RIGHT)72 INCHES
TB1
D7
D6
1N4005
1N4005
TRANSVERSE
1N40051N4005
LONGITUDINAL
1N4005
D2
L2
ANGLE
L3
D3
L4
D4
1N4005
L1
D1
VERTICAL
GREEN/YELLOWS7
S8
AY30-023
S2S5
S1
ORN
SC50-002T REV. E
Chapter 5 Diagrams
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 5-5
Figure 5-3. Model QS-550 Tubestand Wiring Diagram- Automatic Collimator (Progeny Linear II)
ASSEMBLYHANDGRIP
OR
AN
GE
NO
ASSEMBLYTUBESTAND
OR
AN
GE
BR
OW
N
RE
D
WH
ITE
InterfaceCollimator
Box
VERTICAL COMLONG. COMMON
40" SID TABLE40" SID WALL
72" SID WALL
BLU
E
YELLOW
PU
RP
LE
COM
COM
40" TABLE SID SWITCHCOM
BLACKCOM
S9
GR
EE
NO
RA
NG
E
COM
COM
BR
OW
N
BLA
CK
GR
AY
PURPLE BLU
E
YE
LLO
W
RED
GR
EEN
/YE
LLO
W
GROUND
WH
ITE
STUD
BR
OW
N
GR
AY
GROUND STUD
BLA
CK
BLA
CK
BR
OW
N
BLA
CK
LOCK
RE
D
LOCK
RE
D
LOCK
BLACKRED
OR
AN
GE
LOCK
RE
D
+24 VDC RTN+24 VDC
GENERATOR GROUND
FROMGENERATOR
RED
RED
BLA
CK
BLA
CK
NC
NC
NO
NC
NO
Automatic
(AY20-020)
P1
J1
PU
R
OR
N
BLU YEL
RE
D
BLK
GR
N
WH
T
BR
N
NC
NO
NC
NO
NC
NO
AY30-023
CO
MN
O
CO
MN
C
CO
M
CO
M
NC NC
COM
COMNC
NO
NC
NO
GR
Y
1515
12
32
1
2020
181919 18
16171617
12
4
3
65
3
12
3
1
6
4
2
5
6
4
3
5
12
21
12
12
131414 13
111212 11 8
810 9910
7 67 6 1
15 45 4
2323
1111
1212
1313
1414
1515 2 4
51010 53
3
88
9
26 4
1
79 1
76
+24V
VERT SWS2
S10
40" WALL SID SW DPDT
S8
40" TABLE SID SWS7
72" WALL SID SW DPDT
+24V
ALL LOCKS REL
S1
AUTOSTOP SW
TRANS SWS5
S6
1N4005
D5
ANGLE SWLONG SWS3 S4
LED 1 (LEFT)40 INCHES
LED 2 (RIGHT)72 INCHES
TB1
D7
D6
1N4005
1N4005
TRANSVERSE
1N40051N4005
LONGITUDINAL
1N4005
D2
L2
ANGLE
L3
D3
L4
D4
1N4005
L1
D1
VERTICAL
GREEN/YELLOW
ORN
SC50-010T REV. E
Chapter 5 Diagrams
X-ray Tube Stand - Service Manual Revision GQuantum Medical Imaging, LLC 5-6
Figure 5-4. Model QS-550 Tubestand Wiring Diagram- Progeny Linear IV Automatic Collimator
SC50-043T REV. D
CO
MN
C
CO
MN
C
CO
MN
C
COM
COM
NC
NO
NC
NO
CO
MN
O
AY30-023
Collimator Interface Box (LINEAR IV)
GENERATOR
INTERFACE BOARD
GN
D
CO
LL R
DY
BLACK/BROWN
RED
ASSEMBLY
72" SID WALLBLACK
HANDGRIP
40" SID TABLE40" SID WALL
VERTICAL COMMON
LONG. COMMONGREENRED
OR
AN
GE
GR
AY
PURPLE
YELLOW
NO
ASSEMBLYTUBESTAND
TB2
PBL CIRCUITQT-750 TABLE
RED
1K
RED
BROWN
BLACK
40" TABLE SID MONITOR
COM
COM
BLU
E
GR
EE
N
PU
RP
LE
WH
ITE
OR
AN
GE
S9GROUND STUD
COM
GR
AY
COM
BR
OW
N
40" TABLE SID SWITCH
COM BLACK
BLU
E
YELL
OW
RE
D
BLA
CK
STUD
GR
EEN
/YEL
LOW
GROUND
WH
ITE
BR
OW
N
LOCK
BLA
CK
BLA
CK
BR
OW
N
BLA
CK
REDR
ED
LOCK LOCK
OR
ANG
E
LOCK
REDBLACK
GROUND
RE
D
GENERATOR
+24 VDC RTN+24 VDC
GENERATORFROM
BLA
CK
BLA
CK
AY20-020
P1
J1
OR
N
BLU YE
L
RE
D
BLK
GR
N
WH
T
BR
N
PU
R
GR
AY
NCCOM
NC
NO
NC
NO
NC
NO
NC
NO
19
1 2 43 5 76
22
2021
5 14
123
54
11
15
1718
16
1314
12
910
8765
34
14
10
76
98
1211
13
15
21
1
1
2
2
12 1011 9 78 6 45 3 12
1
2
3
3
1
12
3
45
6
3
1
6
4
1212
1818
192020 19
14 131617 151517 16 14 13
66
710 9 811
911 10 8 7
3 245 1
35 4 2 1
2
5
2
6
45
3
12
12
12
2 4
553
3
88
9
26 4
1
79 1
76
1010
1111
1212
1313
1414
1515
1N4005
TB2
TS6
TS4
TS5
VERT SW
+24V
S2S5
S6
AUTOSTOP SW
TRANS SW
D5
R1
POT
S7
72" WALL SID SW DPDT40" WALL SID SW DPDT
S8
ALL LOCKS REL
+24V
S1
ANGLE SWLONG SWS3 S4
LED 2 (RIGHT)LED 1 (LEFT)40 INCHES 72 INCHES
VERTICAL
1N4005
TB1
D6
D7
1N4005
1N4005
1N4005 1N4005
LONGITUDINAL
D2
L2
ANGLE
L3
D3
1N4005
TRANSVERSE
L4
D4
L1
D1
RED
RED
GREEN/YELLOW
ORN
P/N: EL20-022