DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

42
Operator’s Manual Fixed/Elevating Float-Top Radiographic Tables Models QT-740 and QT-750 Manual Part No. DC30-006 Revision G

Transcript of DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Page 1: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Operator’s Manual

Fixed/ElevatingFloat-Top

Radiographic Tables

Models QT-740 and QT-750

Manual Part No. DC30-006 Revision G

Page 2: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC. (QMI)

Copyright© 2009 QMI

Made in U.S.A.Quantum Medical Imaging, LLC2002-B Orville Drive NorthRonkonkoma, New York 11779 USAPhone: (631) 567-5800Fax: (631) 567-5074E-mail: [email protected]

Page 3: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Revision History

Models QT-740 & QT-750 - Operator’s Manual Revision GQuantum Medical Imaging, LLC i

REVISION DATE TYPE OF MODIFICATION

A 12/08/00 Initial Release.

B 3/14/01 Incorporated ECO 0245

C 5/17/01 Added UL mark

D 4/30/04 Incorporated ECO 1359

E 5/2/06 Incorporated ECO 1583

F 9/16/09 Incorporate ECO 2123, 2173, 2199, and updated CE mark

G 10/27/09 Incorporate ECO 2011, 2092

PageNumber Rev

PageNumber Rev

PageNumber Rev

i - iv G

1-1 thru 1-6 G

2-1 thru 2-4 G

3-1 thru 3-14 G

4-1 thru 4-4 G

5-1 thru 5-8 G

Page 4: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Revision History

Revision G Models QT-740 & QT-750 - Operator’s Manualii Quantum Medical Imaging, LLC

THIS PAGE INTENTIONALLY LEFT BLANK

Page 5: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Table of Contents

Models QT-740 & QT-750 - Operator’s Manual Revision GQuantum Medical Imaging, LLC iii

Table of ContentsChapter 1, Safety Notices 1-1

Chapter 2, General Information 2-1

Chapter 3, Operation 3-1

Chapter 4, User Maintenance 4-1

Chapter 5, Warranty Information 5-1

Page 6: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Table of Contents

Revision G Models QT-740 & QT-750 - Operator’s Manualiv Quantum Medical Imaging, LLC

THIS PAGE INTENTIONALLY LEFT BLANK

Page 7: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter

1-1

1

SAFETY NOTICES

Page 8: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

1-2

Page 9: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 1 Safety Notices

Models QT-7Quantum M

GENERAL SAFETY INFORMATION

Follow the safety instructions in this manual and all warnings and cautions printed on the warning labels. This manual, along with system components’ manuals, must be read and understood before using the system. Below is a definition of the warning signs used throughout this document:

WARNINGQuantum Medical Imaging, LLC disclaims all responsibility from any injury resulting from improper application of this equipment.

This equipment is sold to be used exclusively under the prescribed direc-tion of a person who is licensed by law to operate equipment of this nature. This equipment must be used in accordance with all safety procedures described in this manual and must not be used for purposes other than those described herein.

Quantum Medical Imaging, LLC cannot assume responsibility for any mal-functioning of this equipment resulting from improper operation, mainte-nance, or repair, or from damage or modification of its components.

X-RAY PROTECTIONX-rays are hazardous to both patient and operator unless established safe exposure factors and operating instructions are observed.

It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the recommendations of the National Council on Radiation Protection and Measurements as published

WARNING Indicates injury or death is possible if the instructions are not obeyed.

CAUTIONIndicates that damage to equipment is possible if the instructions are not obeyed.

NOTE Indicates essential information that should be read to avoid incorrect operation.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 1-3

Page 10: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 1 Safety Notices

R1

in NCRP Reports available from NCRP Publications, 7910 Woodmont Ave-nue, Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of the International Commission on Radiological Protection (www.icrp.org), and take adequate steps to protect against injury.

X-ray equipment may cause injury if used improperly. The instructions con-tained in this manual must be read and followed when operating this unit. Personal radiation monitoring and protective devices are available. You are urged to use them to protect against unnecessary x-ray exposure.

REGULATORY COMPLIANCEThis certified Quantum Medical Imaging, LLC medical device has been designed, manufactured, and calibrated to comply with governing Federal Regulations 21 CFR Subchapter J and the performance standards attendant thereto. Upon installation, all certified products require the filing of Form FD-2579 "Report of Assembly of a Diag-nostic X-ray System" by the assembler (i.e., the installer) with the appropriate agen-cies; the "Installation Quality Assurance Checklist" must also be completed and properly distributed upon installation. A copy of each form (pink copy) is provided to the user. The installation report is completed by the installer and returned to Quan-tum Medical Imaging, LLC.

Those responsible for the planning of x-ray equipment installations must be thor-oughly familiar and comply completely with NCRP Report No. 49, "Structural Shielding Design and Evaluation for Medical Use of X-Rays and Gamma Rays of Energies up to 10 MeV", as revised or replaced in the future. Those authorized to operate, test, par-ticipate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and pro-cedures described in publications such as Subchapter J of Title 21 of the Code of Fed-eral Regulations, "Diagnostic X-Ray Systems and Their Major Components," and NCRP Report No. 102, “Medical X-Ray, Electron Beam and Gamma Ray Protection for Ener-gies Up to 50 MeV—Equipment Design and Use” as revised or replaced in the future.

Scheduled maintenance is essential to the assurance of continued integrity of this equipment with respect to regulatory compliance. The continuance of certified perfor-mance to the regulatory standard is incumbent upon the user's diligent conformance to recommended maintenance instructions.

This product has been classified as Class I, Type B by Underwriters Laboratories, Inc. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or with nitrous oxide.

MEDICAL ELECTRICAL EQUIPMENTWITH RESPECT TO ELECTRIC SHOCK, FIRE,

MECHANICAL HAZARDS ONLYIN ACCORDANCE WITH

UL 60601-1 AND CAN/CSA C22.2 NO. 601.198UA

evision G Models QT-740 & QT-750 - Operator’s Manual-4 Quantum Medical Imaging, LLC

Page 11: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 1 Safety Notices

Models QT-7Quantum M

EU Authorized Representative:Medizintechnik Berlin GmbHAltentreptower, Strasse 5912683 Berlin - Germany

Phone: +49-302-82 4726Fax: +49-302-82 6382

E-mail: [email protected]

The following symbols may be used for marking on this equipment or equipment doc-umentation:

ASSOCIATED DOCUMENTS

This Operator’s Manual is incomplete without QT-740 & QT-750 Fixed/Elevating Float-Top Radiographic Tables Service Manual, document number DC30-007.

Earth (ground) Type B Equipment

Protective Earth (ground) Attention, consult accompanying documents

Dangerous Voltage Enable tabletop float (releases transverse and longitudinal locks)

Enable Tabletop Up Motion Enable Tabletop Down Motion

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 1-5

Page 12: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 1 Safety Notices

R1

THIS PAGE INTENTIONALLY LEFT BLANK

evision G Models QT-740 & QT-750 - Operator’s Manual-6 Quantum Medical Imaging, LLC

Page 13: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter

2-1

GENERALINFORMATION

2

Page 14: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

2-2

Page 15: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 2 General Information

Models QT-7Quantum M

NOTE

The user should read this man-ual in its en-tirety prior to using this equip-ment. It should be kept in a location near the equipment and readily ac-cessible to the those that oper-ate it.

OVERVIEW

This manual provides information for installing, operating, and maintaining Quantum Medical Imaging’s Fixed (Model QT-740) and Elevating (Model QT-750) Radiographic Tables (hereinafter referred to as the radiographic table). It is imperative that all safety procedures described in this manual be strictly adhered to in order to ensure the safety of both patient and user.

The key features of Model QT-740 and QT-750 radiographic tables are as follows:

• Large Tabletop area (85 inches long x 35.55 inches wide) provides ample examination platform

• Four-way floating tabletop permits 30.5 inches of longitudinal travel and 10 inches of lateral travel for full radiographic coverage

• Foot pedal controls for table float-top motions (foot pedal inhibit switch deactivates foot pedals to prevent inadvertent tabletop motion)

• Float Handle Assembly with FLOAT push button control is provided for operator convenience

• Small patient-to-film plane distance provides improved radiography with minimal magnification

• Accepts all types of image receptors (buckies, grids, and stationary grid cabinets)

• Patient handgrips are standard accessory

• AC line cord, fitted with hospital-grade plug, plugs into a 115VAC, 50/60 Hz receptacle (optionally field wired). Also configurable for 230 VAC, 50/60 Hz power input.

• The Image Receptor Cabinet is equipped with electric locks ensuring precise film receptor alignment

In addition, the Model QT-750 provides:

• Heavy-duty lift mechanism allows for patient loads of up to 650 lbs.* (weight must be centered over table base)

• Variable height tabletop, lowers to 21 inches and raises to 32.5 inches (11.5 inches total vertical travel)

• Foot pedal and hand controls for table up/down motions

• Obstruction sensors disable tabletop down motion in the event an obstruction is detected below tabletop

* 600 lbs. capacity when table is equipped with QT-FCD (Fixed Digital Image Receptor) option; 550 lbs. capacity when equipped when table is equipped with QT-DDS (Digital Drawer Slide) option.

INTENDED USE

The Fixed Float-top (Model QT-740) and Elevating Float-top (Model QT-750) Radiographic Tables are intended for use as a patient support device during the performance of radiographic examinations.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 2-3

Page 16: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 2 General Information

R2

MAIN COMPONENTS

See Figure 2-1. The radiographic table contains:

1 Patient Handgrips

2 Tabletop

3 Image Receptor Lock Release Switch

4 Control Handle: FLOAT Push Button Switch

5 Control Handle: UP/DOWN Rocker Switch (Model QT-750 only)

6 Foot Pedal(s) - Float (Model QT-740)Float, Up and Down (Model QT-750)

7 Table Base

8 Foot Pedal/FLOAT Push Button Inhibit Switch

9 Emergency Shut-off Switch (Model QT-750 only)

Figure 2-1. Radiographic Table Main Components

9

8

7

1 2 3 4

6

5

evision G Models QT-740 & QT-750 - Operator’s Manual-4 Quantum Medical Imaging, LLC

Page 17: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter

OPERATION

3-1

3

Page 18: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

3-2

Page 19: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

OVERVIEW

To provide precise patient positioning, Models QT-740 and QT-750 radiographic tables feature a table top that "floats" in four directions; transversely (forward and backward) and longitudinally (left and right). Model QT-750 has the addi-tional capability of table top elevation (i.e., the table top can be raised and low-ered to facilitate easy transfer of the patient onto and off of the table top).

TABLE CONTROLS

WARNING! All movable assemblies and parts ofthis equipment must be operated with reasonablecare. Manufacturer’s equipment recommenda-tions described in this manual must be observed.

WARNING! To avoid damage to the table, anyload on table top should be distributed as evenlyas possible over the support surface. Do not seatpatient at extreme ends of table top when tabletop is not centered. Patients weighing more thanapproximately 300 lbs. (136 kg) should only betransferred on or off the table from front side - inthe center.

WARNING! Do not allow patient to place his/herhands along the side or below the table rail.Injury to the patient’s hands and/or fingers ispossible.

TABLE TOP FLOAT MOTION (ALL MODELS)

Models QT-740 and QT-750 both feature a "floating" table top. With the "float-ing" table top function activated, both transverse and longitudinal positioning of the patient is enabled. To operate the float-top function, proceed as follows:

WARNING! When transferring a patient on or offthe table top, always make sure the table top islocked. If necessary, the Foot Pedal DisableSwitch may be activated to prevent unintendedactivation of the FLOAT Foot Pedal control.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-3

Page 20: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

R3

1. Position the patient on the table with his or her hands situated along the side of their body, or with the hands grasping the patient handles (see Figure3-1). Be sure that the hands remain above the table top.

Figure 3-1. Patient Positioned on Table Top

2. Press the FLOAT Foot Pedal or the FLOAT push button on the Table Control Handle to enable transverse and longitudinal motion of the table top (see Fig-ures 3-2 through 3-4).

Figure 3-2. Model QT-740 FLOAT Foot Pedal Control

FLOAT Foot Pedal

evision G Models QT-740 & QT-750 - Operator’s Manual-4 Quantum Medical Imaging, LLC

Page 21: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

3. Releasing the FLOAT Foot Pedal or the FLOAT push button (located on the Table Control Handle) immediately activates the table top locks, thereby inhibiting table top "float" motion.

Figure 3-3. Model QT-750 - FLOAT Foot Pedal Controls

Figure 3-4. Table Control Handle FLOAT Push Button Switch

FLOAT Foot Pedals

MODEL QT-740 MODEL QT-750

TABLE TOP FLOAT ENABLE PUSH BUTTON

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-5

Page 22: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Re3-

FOOT PEDAL/CONTROL HANDLE DISABLE SWITCH (ALL MODELS)

To prevent accidental activation of both the Foot Pedal controls (i.e., FLOAT Foot Pedal on Model QT-740 and FLOAT, UP, and DOWN Foot Pedals on ModelQT-750) and the Control Handle push buttons, a Foot Pedal/Control Handle Dis-able Switch is provided. This switch is located below the front table rail on the left end of the table (see Figure 3-5). Pressing the switch until the switch light is illuminated disables all Foot Pedal and Control Handle functions. Pressing the switch until the Foot Pedal/Control Handle Disable Switch light is not illuminated restores normal operation of both the Foot Pedal and Control Handle switches.

.

Figure 3-5. Foot Pedal/Control Handle Disable Switch Location

FOOT PEDAL/CONTROL HANDLE

PUSH BUTTONDISABLE SWITCH

vision G Models QT-740 & QT-750 - Operator’s Manual6 Quantum Medical Imaging, LLC

Page 23: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

NOTE

The Model QT-750 table is equipped with Obstruction Sensors, which detect the presence of an obstruction below the table top when the table top is being driven downward. This feature is only designed to avoid dam-age to the table and is not intended as a safeguard against per-sonal injury.

ELEVATING MOTION (MODEL QT-750 ONLY)

On Model QT-750 radiographic tables, the elevation function allows the table top to be lowered down to as low as 21 inches (533.4 mm) from floor, up to a work-ing height of 32.5 inches (825.5 mm). To operate the table top elevation func-tion, proceed as follows:

WARNING! For safety purposes on ModelQT-750 tables, an Emergency Stop Switch (seeFigure 3-6) when activated immediately removespower from the table top up/down motor. Rotat-ing the Emergency Stop Switch knob 1/4-turncounter-clockwise releases the Emergency StopSwitch and restores up/down motor power.

Figure 3-6. Emergency Stop Switch Location

IMPORTANT! On Model QT-750 tables equippedwith Digital Receptor Drawer Slide Option (QT-DDS): Table top up or down motion is disabledwhen the digital receptor drawer slide isextended out from under the table top. Pushreceptor in fully to enable up or down motion.

EMERGENCY STOPSWITCH

(MODEL QT-750 ONLY)

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-7

Page 24: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

R3

Figure 3-7. Model QT-750 - Up/Down Foot Pedal Controls

Figure 3-8. Model QT-750 - Control Handle Table Top Up/Down Rocker Switch

LOWERING THE TABLE TOP

1. It is recommended that the table top be centered before lowering it (refer to table top Float Motion operation described in preced-ing paragraph).

2. Press the DOWN Foot Pedal (see Figure 3-7) or, on the Control Handle (see Figure 3-8), press the Table Top Rocker Switch toward the down arrow to lower the table top to the desired height.

3. Transfer patient onto table top.

DOWN PEDAL UP PEDAL

PRESS HERE TO ENABLETABLE TOP UP MOTION

PRESS HERE TO ENABLETABLE TOP DOWN MOTION

evision G Models QT-740 & QT-750 - Operator’s Manual-8 Quantum Medical Imaging, LLC

Page 25: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

4. Ensure patient is positioned with his or her hands situated along the side of their body, or with the hands grasping the patient handles (see Figure 3-1). Be sure patient’s hands remain above the table top.

RAISING THE TABLE TOP

1. Press the UP Foot Pedal (see Figure 3-7) or, on the Control Han-dle (see Figure 3-8), press the Table Top Rocker Switch toward the up arrow to raise the table top to the desired height.

2. Proceed with examination.

IMAGE RECEPTOR CABINET LOCK

The Image Receptor Cabinet rides on four linear ball-bearings mounted to adjust-able brackets attached to either side of the cabinet. These bearings travel on the hardened and polished steel shafts of the Receptor Cabinet Carriage. An electro-magnet on the underside of the housing "locks" the cabinet into place until the LOCK RELEASE switch is depressed, thereby releasing the magnet’s hold and allowing free movement of the cabinet (see Figure 3-9).

Figure 3-9. Image Receptor Cabinet LOCK RELEASE Switch

CASSETTE TRAY OPERATION

The radiographic table is equipped with either a Midwest or Poersch cassette tray, depending on the system ordered. The following paragraphs describe the operat-ing instructions for each. Additional information is contained in the cassette tray manufacturer’s documentation, which is provided with the table.

LOCK RELEASESWITCH

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-9

Page 26: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

R3

Loading Cassette Tray (Midwest Type)

To load a film cassette into a "Midwest" type cassette tray, proceed as fol-lows:

1. Float the table top fully back.

2. Pull cassette tray from receptor using tray handle (see Figure3-10).

3. Pull back on front cassette grip to open both grips.

4. Position slide guides ("L" brackets seated in channels) to cassette size using indicators on cassette tray (press center brass button to allow movement).

5. Insert cassette into tray (back end first).

6. Lock cassette into tray.

7. Push front grip against cassette.

8. While pushing front grip against cassette, turn cassette lock han-dle to lock position.

9. Push tray into receptor. The cassette is now in exposure position.

Figure 3-10. Midwest Cassette Film Tray

CASSETTE GRIP

TRAY HANDLE

SLIDE GUIDES

LOCK HANDLE

evision G Models QT-740 & QT-750 - Operator’s Manual-10 Quantum Medical Imaging, LLC

Page 27: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

Loading Cassette Tray (Poersch Type)

To load a film cassette into a "Poersch" type cassette tray, proceed as fol-lows:

1. Float the table top fully back.

2. Pull cassette tray from receptor using tray handle (see Figure3-11).

3. Place thumb under lip of clamp handle. Lift handle to release.

4. Grasp handle and slide clamp back for inserting cassette.

5. Insert cassette into tray (back end first). Center it using center-ing label.

6. Slide clamp forward pushing clamp firmly against cassette.

7. While pushing clamp against cassette, press clamp handle down.

Figure 3-11. Poersch Cassette Film Tray

TRAY HANDLE

CLAMP HANDLE

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-11

Page 28: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

R3

DIGITAL DRAWER SLIDE OPTION (QT-DDS) OPERATION

The following paragraphs describe operating instructions for the QT-DDS Digital Drawer Slide option. For instructions pertaining to operation of the digital imag-ing system, refer to the Digital Receptor manufacturer’s documentation provided with the digital imaging system.

Digital Receptor (Stored)

Figure 3-12. Digital Image Receptor - Stored Position

WARNING! Keep fingers away from area betweentop of receptor and table top when pushing inreceptor; fingers may get caught.

Lift up on the Receptor Lock Handle and push the receptor forward until the receptor is fully withdrawn below the table top as shown in Figure3-12.

RECEPTOR LOCK HANDLE

evision G Models QT-740 & QT-750 - Operator’s Manual-12 Quantum Medical Imaging, LLC

Page 29: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

Models QT-7Quantum M

Digital Receptor (Extended)

Figure 3-13. Digital Image Receptor - Extended Position

WARNING! Do not apply a load of more than 32lbs. (14.5 kg) on the image receptor when it is inthe extended position; the unit may be collapse.

WARNING! On Model QT-750 tables equippedwith Digital Receptor Drawer Slide Option(QT-DDS): Table top up or down motion is dis-abled when the digital receptor drawer slide isextended out from under the table top. Pushreceptor in fully to enable up or down motion.

Lift up on the Receptor Lock Handle and pull the receptor forward until the receptor is fully extended as shown in Figure 3-13.

Digital Receptor Lateral (Left-to-Right) Motion

Lift up on the Receptor Lock Handle and slide the digital left or right until the desired lateral receptor position is attained. Release the Receptor Lock Handle to lock the receptor in its current position.

RECEPTOR LOCK HANDLE

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 3-13

Page 30: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 3 Operation

R3

CXDI-50G DIGITAL RECEPTOR CABINET OPTION (TYPE RT-DOK50-T) OPERATION

Quantum offers an optional, motorized, rotating receptor tray for use with a Canon CXDI-50G digital receptor used in QT-740 and QT-750 X-ray table applica-tions. The RT-DOK50-T option provides a rotating tray that allows the X-ray Tech-nician to change receptor orientation without removing the receptor from the tray.

Detailed instructions for operation of the RT-DOK50-T is available in User and Installation manual, Hoorn-Holland document number L05001L manual-01A.wpd (shipped with the X-ray table).

CXDI-50G DIGITAL RECEPTOR CABINET OPTION (TYPES Q-DRT-CT ANDQ-DRT-CT-R)

Quantum offers rotating and non-rotating receptor tray models for Canon CXDI-50G digital receptors used in QT-740 and QT-750 X-ray table applications. The non-rotating version (Q-DRT-CT) provides a tray capable of receiving theCXDI-50G receptor in portrait or landscape orientation (changed manually), while the rotating version (Q-DRT-CT-R) provides a rotating tray that allows the X-ray Technician to change receptor orientation without removing the receptor from the tray.

Detailed instructions for operation of the Q-DRT-CT and Q-DRT-CT-R are available in document DC30-093 (shipped with the X-ray table).

evision G Models QT-740 & QT-750 - Operator’s Manual-14 Quantum Medical Imaging, LLC

Page 31: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter

4-1

USERMAINTENANCE

4

Page 32: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

4-2

Page 33: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 4 User Maintenance

Models QT-7Quantum M

OVERVIEW

This chapter is designed to assist the system user in maintaining the smooth opera-tion of the table. The user is responsible for performing certain routine maintenance and inspection procedures. Aside from routine maintenance, any abnormal noise, vibration, or unusual performance should be investigated by a qualified service repre-sentative. Preventive maintenance or any repair service should be performed only by qualified service personnel.

WARNING! Failure to follow manufacturer’s orservice personnel’s recommendations may resultin serious injury.

USER MAINTENANCE

User maintenance consists of the following activities, which should be performed on a daily basis:

• Visually inspect the table for wear and cleanliness

• Clean the tabletop, cassette tray, and exterior painted surfaces of the table

• Disinfect the table top surface after each use in accordancew with facility requirements.

Cleaning

The system user is responsible for the basic cleanliness of the equipment. On a regular basis, the table surface should be wiped clean. Painted metal sur-faces should be cleaned using a clean cloth slightly moistened in warm soapy water (use mild soap). Wipe with a clean wet cloth, then dry. Never use abrasive polish on this equipment.

WARNING! Always disconnect the equipmentfrom the main power supply prior to any cleaning.On 115 VAC systems, the power supply cord is thepower disconnect device. On 230 VAC systems,the power disconnect is the main disconnect (orpower cable) on systems wired directly to facilitypower.

To remove the cassette tray from the Receptor Cabinet for cleaning, proceed as follows:

Midwest Trays

1. Float the table top fully back.

2. Pull the cassette tray out by the handle until it stops.

3. Grasp both sides of the tray with your right hand placed near the table and your left hand towards the tray front.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 4-3

Page 34: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 4 User Maintenance

Re4-

4. Locate the spring-loaded lever on the right side of the tray bottom. A spring-loaded plate covers this hole. Press the lever with your right index finger, lift the cassette tray up slightly and pull the tray away from the Receptor Cabinet.

Poersch Trays

1. Float the table top fully back.

2. Pull the cassette tray out by the handle until it stops.

3. Grasp the left side of the tray with your left hand.

4. Locate the square block at the back right side of the cassette tray. A spring-loaded plate is on the front side of the block. While pressing the plate, pull the tray out from the Receptor Cabinet.

CXDI-50G Digital Receptor CabinetOption (Type RT-DOK50-T)

Maintenance instructions for the Hoorn-Holland digital receptor tray (Quan-tum option RT-DOK50-T) is described in detail in document User and Installa-tion manual, Hoorn-Holland document number L05001L manual-01A.wpd (shipped with the X-ray table).

Quantum Q-DRT-CT/Q-DRT-CT-R Trays for Canon CXDI-50G DigitalReceptors

Maintenance instructions for the Q-DRT-CT and Q-DRT-CT-R digital receptor trays are described in detail in document DC30-093 (shipped with the X-ray Table).

vision G Models QT-740 & QT-750 - Operator’s Manual4 Quantum Medical Imaging, LLC

Page 35: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter

5-1

WARRANTYINFORMATION

5

Page 36: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

5-2

Page 37: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

Models QT-7Quantum Me

WARRANTY STATEMENTQuantum Medical Imaging, LLC (herein after known as “QMI”) warrants to the buyer that any new product manufactured by QMI will be free from defects in material and workmanship, and will substantially conform to the applicable specifications in effect on the date of shipment when subjected to normal, proper and its intended use by properly trained personnel. QMI shall be the sole judge in determining whether said equipment or component is defective by reason of manufacture.

All QMI products shall be so warranted for a period of 12 months from the date of original installation, such date to be evidenced by means of a completed Warranty Card returned to QMI within 30 days of installation. In no case shall the warranty extend beyond 15 months from the date of shipment. If the attached warranty card is not so returned to QMI, then the warranty period will be deemed to have com-menced on the date of shipment (the invoice date) and extend for a period of twelve months. The buyer should submit only one such card per system or major compo-nent purchased.

WARRANTY CARD

40 & QT-750 - Operator’s Manual Revision Gdical Imaging, LLC 5-3

Page 38: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

R5

Clip or Copy Warranty Card on reverse side and submit to QMI

See Reverse Side for Warranty Card

evision G Models QT-740 & QT-750 - Operator’s Manual-4 Quantum Medical Imaging, LLC

Page 39: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

Models QT-7Quantum M

WARRANTY STATEMENT (Continued)Promptly complete the warranty card and mail or fax it to:

Quantum Medical Imaging, LLC2002-B Orville Drive NorthRonkonkoma, N.Y. 11779 USA631 567-5074 fax 631 567-5800 voice

Replacement components furnished by QMI to the Buyer/Dealer during the warranty period shall be warranted for the remainder of the original product warranty or 90 days, whichever is longer. This warranty extends only to the original purchaser and is not transferable unless expressly authorized in writing by Quantum Medical Imaging, LLC.

Products manufactured by parties other than QMI, whereby QMI acts solely as distrib-utor or reseller, are warranted exclusively by their manufacturers according to each of their independent warranty terms and conditions.

Warranty consideration can only be given for defective QMI products properly returned to the factory in accordance with the QMI Returned Materials Procedure (refer to Dealer Price Book or contact QMI customer service).

WARRANTY EXCLUSIONSThe foregoing warranties are exclusive and in lieu of all other warranties, whether written, oral, express, implied or statutory. NO IMPLIED WARRANTY OF MERCHANT-ABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Quantum Medical Imaging, LLC Warranty is exclusive of:

1) Failure of the Buyer/Dealer to prepare the site and operating environment in accordance with applicable instructions and recommendations of QMI.2) Failure of Buyer/Dealer to provide the proper incoming power required to support the equipment in accordance with the requirements of QMI.3) Modification of QMI products performed by a party other than QMI.4) Combining products deemed by QMI to be incompatible.5) Improper or extraordinary use of a product, improper maintenance of the prod-uct, or failure to comply with any applicable instructions and recommendations of Quantum Medical Imaging, LLC.6) Misuse, abuse, tampering, or negligent storage or handling of a product by the Buyer, its employees, agents, or contractors.7) Fuses, glassware, high voltage cables and other items deemed by QMI to be expendable.

Acts of God, fires, floods, power failure or electrical power surges. Strikes, sabotage, labor disturbances, war, riots, acts of civil or military authority, or other causes beyond the reasonable control of QMI.

Installation, routine troubleshooting and repair are also excluded from warranty. Technical service and maintenance is the responsibility of the Dealership selling the equipment.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 5-5

Page 40: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

R5

The Manufacturer is hereby relieved of all responsibility for damage during shipment of the product following the freight carrier’s pick-up for transportation to the delivery point.

BUYER’S REMEDIESIf QMI determines that a product fails to meet any specification during the applicable warranty period, QMI shall correct any such failure as follows:A) By repairing, adjusting, or replacing any defective or non-conforming component

or product.B) By making available any necessary repair or replacement parts or assemblies for

exchange.

Quantum Medical Imaging, LLC shall have the option to furnish either new or rebuilt replacement parts or assemblies for exchange. All returned parts shall become the property of Quantum Medical Imaging, LLC upon exchange.

The preceding Paragraphs set forth the Buyer’s sole remedies and QMI’s sole liability for claims based upon failure of the product to meet any warranty, whether the claim is on contract, warranty, tort (including negligence and strict liability) or otherwise, and however instituted.

Upon the expiration of the applicable warranty period, all such liability shall terminate. In no event shall QMI be liable for special or consequential damages arising out of the use of or inability to use its equipment, whatsoever.

The warranties and remedies available to the buyer are conditioned upon claims under this warranty being made in accordance with the aforementioned warranty statement.

WARRANTY RETURN PROCEDUREA fully completed Field Returned Material Evaluation Form must be returned with any defective product or any returned item. All returns must include the Serial Number of the Equipment and the Specific Part Number written on the Field Returned Material Evaluation Form. All freight charges resulting from Warranty Returns are the respon-sibility of the Buyer/Dealer.

EQUIPMENT IN TRANSITQMI assumes no responsibility for equipment damaged in transit to or from QMI. To protect the Buyer/Dealer, the receiver of any equipment should examine all cartons and crates carefully at the time of delivery. If damage is apparent, make a notation on the delivery receipt, request an inspection by the freight carrier, and if applicable, file an appropriate carrier claim. Should concealed damage be detected, immediately notify the carrier and request an inspection. The purchaser (Buyer/Dealer/Customer) is fully responsible for the filing of freight damage claims to the freight carrier.

QMI assumes no responsibility for any loss or damage to products once they have been shipped from our factory. As such, the Buyer/Dealer and Customer remain fully responsible for payment to QMI for all invoices, according to our standard payment terms, regardless of freight damage or processing of an insurance claim, by the dealer or customer.

evision G Models QT-740 & QT-750 - Operator’s Manual-6 Quantum Medical Imaging, LLC

Page 41: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

Models QT-7Quantum M

VOIDING WARRANTYTampering with, or any attempt at installation, maintenance, repair, service, reloca-tion, or alteration of or to a QMI product, when performed by any person or entity other than Quantum Medical Imaging, LLC or its Certified Dealer without the written approval of an Authorized Person at Quantum Medical Imaging, LLC, shall immedi-ately Void and Cancel all warranties with respect to the affected product.

40 & QT-750 - Operator’s Manual Revision Gedical Imaging, LLC 5-7

Page 42: DC30-006 Quantum QT-740 & QT-750 Operator Manual Rev G

Chapter 5 Warranty Information

R5

THIS PAGE INTENTIONALLY LEFT BLANK

evision G Models QT-740 & QT-750 - Operator’s Manual-8 Quantum Medical Imaging, LLC