Data Acquisition Protocol CT scanning protocol · Data Acquisition Protocol. CT scanning protocol....

2
Phone: +1 866 596 2022 (toll free) Website: stryker.com Address: 5676 Greenwood Plaza Blvd. Ste. 200 Greenwood Village, CO, 80111, USA Copyright © 2019 Stryker Customer Service: Stryker Sports Medicine Data Acquisition Protocol CT scanning protocol Imaging requirements Pelvis + proximal femur Knee Number of slices below lesser trochanter A scan must include at least the bilateral pelvis and proximal femur. If a knee scan is included femoral torsion will be measured. In such a case the pelvis and knee need to be scanned in the same frame of reference. • Required slice increment / slice thickness: 1 mm. No gap. Recommended field of view (FOV): Include the entire bilateral pelvis and at least 180 mm below the lesser trochanter. Minimally required FOV: Bilateral scan from at least the left and right anterior superior illiac spines (ASIS) to at least 40 mm below the lesser trochanter. • Required slice increment / slice thickness: 5 mm. No gap. Recommended FOV: Include bilateral knees from at least the joint line between the femur and the tibia to 100 mm proximal on the femur. Minimally required FOV: Unilateral knee scan. Slice interval spacing, mm 0.500 0.625 0.750 0.875 1.000 Distance, mm 180 180 180 180 180 Number of slices 360 288 240 206 180 180 mm Pelvis / Femur Knee Recommended FOV Required FOV 40 mm 100 mm ASIS ASIS HipMap Femoroacetabular Impingement Analysis Recommended FOV Required FOV Email: [email protected]

Transcript of Data Acquisition Protocol CT scanning protocol · Data Acquisition Protocol. CT scanning protocol....

Page 1: Data Acquisition Protocol CT scanning protocol · Data Acquisition Protocol. CT scanning protocol. Imaging requirements Pelvis + proximal femur. Knee Number of slices below lesser

Phone:+1 866 596 2022 (toll free)

Website: stryker.com

Address: 5676 Greenwood Plaza Blvd. Ste. 200 Greenwood Village, CO, 80111, USA

Copyright © 2019 StrykerCustomer Service: Stryker Sports Medicine

Data Acquisition Protocol

CT scanning protocolImaging requirements

Pelvis + proximal femur

Knee

Number of slices below lesser trochanter

A scan must include at least the bilateral pelvis and proximal femur. If a knee scan is included femoral torsion will be measured. In such a case the pelvis and knee need to be scanned in the same frame of reference.

• Required slice increment / slicethickness: ≤1 mm.No gap.

• Recommended field of view (FOV):Include the entire bilateral pelvis and atleast 180 mm below the lesser trochanter.

• Minimally required FOV:Bilateral scan from at least the left andright anterior superior illiac spines(ASIS) to at least 40 mm below thelesser trochanter.

• Required slice increment / slicethickness: ≤ 5 mm.No gap.

• Recommended FOV:Include bilateral knees from at least thejoint line between the femur and the tibiato 100 mm proximal on the femur.

• Minimally required FOV:Unilateral knee scan.

Slice interval spacing, mm0.5000.6250.7500.8751.000

Distance, mm180180180180180

Number of slices360288240206180

180 mm

Pelvis / Femur

Knee

Rec

omm

ende

d FO

VR

equi

red

FOV

40 mm

100 mm

ASISASIS

HipMap Femoroacetabular Impingement AnalysisR

ecom

men

ded

FOV

Req

uire

d FO

V

Email:[email protected]

Page 2: Data Acquisition Protocol CT scanning protocol · Data Acquisition Protocol. CT scanning protocol. Imaging requirements Pelvis + proximal femur. Knee Number of slices below lesser

Copyright © 2019 Stryker

Phone:+1 866 596 2022 (toll free)

Website: stryker.com

Data Acquisition Protocol

CT scanning protocolPelvis / Femur

Knee

General scan requirements

General scan recommendations

Imaging acceptance

• Field of view: ≤ 500 mm

• Reconstruction kernel: Bone

• Energy settings: 120-140 kV / 200-250 mA

• Patient position: supine

• Minimize pelvic obliquity:-

-

• Scans that fullfill the requirements andrecommendations will be processed.Scans that deviate from recommendationsare subject to further evaluation and mayneed to be rejected.

1 mm

5 mm

512 pixels

512 pixels

1 / 5 mm

HipMap Femoroacetabular Impingement Analysis

Ensure that the hips and knees are in NEUTRAL alignment: no excessive internal or external rotationEnsure on the scout image that the relative heights of the ASIS are symmetric

• No metal-induced image artifacts

Original axial slices using helical (spiral) scanning

• Data format: DICOM 3 compatible

• No compression

• No annotations embedded in the files

• No motion artifacts

1000902900 Rev. A

Customer Service: Stryker Sports Medicine

Email:[email protected]

Address: 5676 Greenwood Plaza Blvd. Ste. 200 Greenwood Village, CO, 80111, USA