CT Made ECasy

download CT Made ECasy

of 47

Transcript of CT Made ECasy

  • 7/29/2019 CT Made ECasy

    1/47

    CT made easy

  • 7/29/2019 CT Made ECasy

    2/47

    Introduction

    The computed tomograpic (CT) scanner isrevolutionary.

    It does not use an ordinary image reseptor, butinstead a well-collimated x-ray beam directed on thepatient, and the attenuated image response istransmitted to a computer.

    The computer reconstructs the image and displays iton a monitor.

    The reconstructions are accomplished withalgorithms adapted for computer processing.

  • 7/29/2019 CT Made ECasy

    3/47

    History of CT

    The first demonstration of the technique was done in1970, but the components to construct a CT scanner

    was avalible 20 years before this. In 1982, Godfrey Houndsfield shared the Nobel prize

    in physics with Alan Cormack. Cormack had earlierdeveloped the mathematics used to reconstruct CTimages.

    No other x-ray equipment development are asimportant as the CT. Some say that MRI and UL areas equally important, but they are, however, not x-ray equipments.

  • 7/29/2019 CT Made ECasy

    4/47

    Principles of use

    CT is a tranaxional/transverese image. Thatmeans that you get axial pictures of the body.

    It is extremly complicated to understand theprecise method on how the CT makes theseimages. You need to have good knowlegdeof physics, engineering and computerscience.

    The basic principles can be demonstrated ifyou consider the simplest CT systems.

  • 7/29/2019 CT Made ECasy

    5/47

    Principles of use

    The x-ray source and and detector are connected sothat they move at the same time.

    When the machine makes one sweep over thepatient the internal structures of the body attenuatethe x-ray beam according to their mass density andeffective atomic number.

    The machine takes several sweeps of the body andcollects this in a computer, this computer thenreconstructs the images of the anatomic structures inthat slice.

  • 7/29/2019 CT Made ECasy

    6/47

    Principles of use

    The detector signal during each sweep is

    registered in increments with values as highas 1000. The value og each increment isrelated to the x-ray attenuation coefficient ofthe total path trough the tissue.

    Through the use of simultaneous equations,a matrix of values is obtained that representsa cross section of anatomy.

  • 7/29/2019 CT Made ECasy

    7/47

    The different generations of CT

    There are four/five generations of CT

    scanners. The fifth is still under development. The first generation: translate-rotate

    configuration, pencil- shaped beam, singledetector, 5-min scan time

    Second generation: translate- rotateconfiguration, fan-shaped beam, detectorarray, 30 s- scan time

  • 7/29/2019 CT Made ECasy

    8/47

    The different generations of CT

    Third generation: rotate- rotate configuration,

    fan-shaped beam, detector array, 1-s scantime, disadvantage: ring artefacts

    Fourth generation: rotate-stationaryconfiguration, fan-shaped beam, detector

    array, 1-s scan time

  • 7/29/2019 CT Made ECasy

    9/47

    Third generation scanner

    In these scanners the x-ray tube and detector arrayare rotated concentrically about the patient.

    They can produce an image in one sec. It uses a curvilinear array containing many detectors

    and a fan beam.

    The curvilinear detector array results in a constant

    source-to-detector path length, which was anadvantage for good image reconstruction.

    This also allows for better x-ray beam collimation toreduce the effect of scatter radiation.

  • 7/29/2019 CT Made ECasy

    10/47

    Third generation scanner

    One disadvantage is the ring artefacts. They

    occur for several reasons. Each detector views a ring of anatomy , so if

    any single detector malfunctions, the aquiredsignal will result in a ring on the

    reconstructed image. Software-corrected image reconstruction

    algorithms minimize such artefacts.

  • 7/29/2019 CT Made ECasy

    11/47

    Fourth generation scanners

    Radiation detection is accomplished through a fixedcircular array, which contains as many as 8000

    individual elements.

    The fixed detector array does not result in a constantbeam path from the source to all the detectors, but itallowes each detector to be calibrated and its signal

    normalized during a scan. They are generally without ring artefacts

    Disadvantage: patient dose and cost of buying

  • 7/29/2019 CT Made ECasy

    12/47

    Fifth generation scanners

    Development of CT is always going on. The

    producers wish to make a CT scanner withimproved image quality at a lesser patientdose.

    Rotate-nutate scanners: Toshiba has

    produced a novel extension of the fourthgeneration. To maintain the x-ray source atthe same distance from the patient as thedetectors, the detector array nutates, as the

    x-ray source rotates.

  • 7/29/2019 CT Made ECasy

    13/47

    Fifth generation scanners

    Electron-beam CT (EBCT) is a fundamentally

    different way to produce CT images. Imatroncame up with the idea for scanning the heart.

    Currently , EBCT is used to scan all tissues,but especially when ultrafast imaging is

    helpful. EBCT images are produced in 50 ms.

  • 7/29/2019 CT Made ECasy

    14/47

    System components

    The gantry

    The computer The operating console

  • 7/29/2019 CT Made ECasy

    15/47

    The gantry

    Includes the x-ray tube, the detector array,

    the high-voltage generator, the patientsupport couch and the mechanical supportfor each.

    X-ray tube: it has special requirements. The

    power capacity must be high. The anodeheating capacity must be atleast severalmillion heat units (MHU).

  • 7/29/2019 CT Made ECasy

    16/47

    The gantry

    High speed rotors are used in most tubes for thebest heat dissipation.

    Focal-spot size is important. CT scanners designedfor imaging using high spatial resolution incorporatex-ray tubes with small focus-spot.

    Detector assembly: Early scanners had one

    detector. Modern scanners have up to 8000, devidedin to groups; scintillation detectors and gas-filleddetectors.

  • 7/29/2019 CT Made ECasy

    17/47

    The gantry

    Scintillation detectors:

    Containes scintillation crystal-photodiode

    assemblies. They convert light into electronicsignals. They are highly efficient at detecting x-rays,almost 90 % of the x-rays are absorbed andcontribute to the output signal.

    But the space between each detector is big, so theoverall detection efficiency may only be 50 %. Theygive dose to patient but do not contribute to theimage.

  • 7/29/2019 CT Made ECasy

    18/47

    The gantry

    Gas-filled detectors: Contructed of a large metallic chamber with baffles

    spaced with 1 mm intervals. The baffles are like grid stripes and devide the large

    chambers into small ones. Each small chamber is one detector. It is sealed and filled under preassure with an inert

    gas with high atomic number (xenon/xenon-kryptonmixture)

    The overall total detetction efficiency is 45 %, almostthe same as scintillation detectors.

  • 7/29/2019 CT Made ECasy

    19/47

    The gantry

    Collimators: Required for the same reason asconventional x-ray. Correct collimation reduces

    patient dose and improves image quality due to lessscattered radiation.

    In CT there are normally two collimators.

    One is the prepatient collimator; on the x-ray tube

    housing/adjacent to it. It limits the area of the patient that intercepts the

    useful beam and thereby the slice thickness and thepatient dose.

  • 7/29/2019 CT Made ECasy

    20/47

    The gantry

    Improper adjustment of this collimator is the cause ofmost of the un-necessary dose to patient.

    The predetector collimator; located under the patient,over the detector array.

    Reduces scatter radiation improves image quality

    When coupled correctly with the prepatient detector,it defines the slice thickness.

    Has nothing to do with patient dose.

  • 7/29/2019 CT Made ECasy

    21/47

    The gantry

    High-voltage generator;

    All CT scanners operate on three-phase orhigh-frequenzy power.

    Most manufactors built them into the gantryor by mounting on the rotating wheel of the

    gantry. It reduces the amount of spaceneeded, and winding and unwinding a powercable is unnecessary.

  • 7/29/2019 CT Made ECasy

    22/47

    The gantry

    Patient positioning and support couch;

    It has to be made of a material with a lowatomic number (carbon fiber) so that is doesnot interfere with x-ray beam transmissionand patient imaging.

    It should move smoothly for accurate patientpositioning, and is especially important forspiral CT

  • 7/29/2019 CT Made ECasy

    23/47

    Computer

    It is unique for the CT and a must! A ultra-high speeddigital computer is needed for making CT images.

    Depending on the format the computer has to do upto 250 000 equations at the same time! In the computer there is a microprocessor and a

    primary memory. These determine thereconstruction time= the time from end of scanning

    to image appearance. Array processors are becoming more common. They

    are faster than the microprocessor and canreconstruct an image in less than 1 s.

  • 7/29/2019 CT Made ECasy

    24/47

    Operating console

    Many CT scanners have 2 or 3 consoles.

    One for the CT radiologic technologist to operate thescanner.

    One for an other technologist to postprocess images.

    One for the physician to view the image, manipulatecontrast, size and general visual appearance.

    A typical operating console contains controls andmonitors for the various technique factors.

  • 7/29/2019 CT Made ECasy

    25/47

    Image characteristics

    With CT, the x-rays form a stored electronic

    image that is displayed as a matrix ofintensities.

    The CT scan format consists of many cellswith its own number which is shown as a

    brightness level. A matrix of 512 x 512 = 262 144 cells of

    information.

  • 7/29/2019 CT Made ECasy

    26/47

    Image characteristics

    Each cell is a pixel (picture element)

    The numerical information in each pixel is aCT number/ Houndsfield Unit (HU)

    It is a two dimensional representation of acorresponding tissue volume.

    The diameter of image reconstruction iscalled the field of view (FOV)

  • 7/29/2019 CT Made ECasy

    27/47

    Image characteristics

    When the FOV is increased for a fixed matrix ( forexample: from 12 to 20 cm) the size of each pixel is

    increased proportionately. When the matrix size is increased for a fixed FOV

    (for example 512 x 512 to 1024 x1024)the pixel sizegrows smaller.

    Pixel size = FOV/matrix size

    The tissue volume is known as a voxel (volumeelement)

    Voxel size= Pixel size x slice thickness

  • 7/29/2019 CT Made ECasy

    28/47

    CT numbers

    Each pixel is displayed on the video monitoras a level of brightness and on the

    photographic image as a level of opticaldensity.

    The levels correspond to a range of CTnumbers from -1000 up to +1000 for eachpixel.

    -1000 is air, +1000 is dense bone and 0 iswater.

  • 7/29/2019 CT Made ECasy

    29/47

    CT numbers

    The CT number is related to the x-rayattenuation coefficient of the tissue containedin the voxel.

    Remember: the degree of x-ray attenuationis determined by the avarage energy of the

    x-ray beam and the effective atomic numberof the absorber and is expressed by theattenuation coefficient.

  • 7/29/2019 CT Made ECasy

    30/47

    CT numbers

    By the scale of HU there is a range of 2000different gray scales with imformation, butmost of it goes lost.

    The screen only shows 32 grayscales.

  • 7/29/2019 CT Made ECasy

    31/47

    Image reconstruction

    Filtered back projection = all the projectionsduring on CT examination is stored in thecomputers memory, and the reconstructionsare made by these.

    With filter we do not mean a metal filter as in

    the tube of the x-ray, but it is a mathematicalfunction. A difficultone!

  • 7/29/2019 CT Made ECasy

    32/47

    Image reconstruction

    In CT there over 250 000 pixels toreconstruct from, that means that themachine has to solve 250 000 equations tofind the solutions for the images.

  • 7/29/2019 CT Made ECasy

    33/47

    Image quality

    Spatial resolution

    Contrast resolution Noise

    Linearity

    Uniformity

  • 7/29/2019 CT Made ECasy

    34/47

    Spatial resolution

    If you scan a regular geometric structure that has asharp interface,the image at the interface will be

    blurred. The degree of blurring is a measure of spatialresolution of the system and is controlled by severalfactors.

    If you take a scan over an area that has a high

    contrast interface, for example the brain and theskull, the image will be blurred. The system will fix some of the blurring, and

    smoothen the picture.

  • 7/29/2019 CT Made ECasy

    35/47

    Spatial resolution

    This, however, reduses the spatial resolutionbecause of some features of the scanner.

    The larger the pixel size and the lower the subjectcontrast, the poorer the spatial resolution will be.

    The detector size and design of prepatient andpostpatient collimation affect the level of scatterradiation and influence the spatial resolution byaffecting the contrast of the system.

    Also the x-rays focal spot has influence on spatialresolution.

  • 7/29/2019 CT Made ECasy

    36/47

    Contrast resolution

    Contrast resolution = the ability to distinguishone soft tissue from another without regardfor size or shape.

    Contrast resolution is superior in CT,principally because of x-ray beam

    collimation.

  • 7/29/2019 CT Made ECasy

    37/47

    Contrast resolution

    Imagine a scan over abdomen, where youhave spine, liver and fat. The atomic

    numbers are different, but in conventional x-ray it is difficult to seperate them. With CTand the CT numbers it makes it easy! WithHU the CT can amplify these contrast

    differences, and make the contrast high.Then we can cleary see differences betweentissue.

  • 7/29/2019 CT Made ECasy

    38/47

    Noise

    Noise= the precentage of standard deviation of alarge number of pixels obtained with a water-bath

    scan. Noise depends on the following factors:

    1: kilovolt peak filtration

    2: Pixel size

    3: Slice thickness4: Detector efficiency

    5: Patient dose

  • 7/29/2019 CT Made ECasy

    39/47

    Noise

    Example:

    If you scan a homogeneous medium like waterthe pixel value should be zero. But becausethe system is not perfect some pixel valueswill be both higher and lower than zero.These variations in HU will show in the

    image as graininess, and is what we callnoise. The larger the variations in pixel value,the more noise you get in the image.

  • 7/29/2019 CT Made ECasy

    40/47

    Linearity

    The CT must be calibrated frequently so that the HU arecorrect.

    There is a test you can do with a phantom and a water bucket. The result from this test should show a linear line passing

    through the CT number of water (0)

    If the test shows deviation from linearity its a sign ofmalfunction of the CT.

    It may not show on the visual image, but could greatly affectquantitative analysis of tissue, the determination of tissuecomposition based on CT number.

  • 7/29/2019 CT Made ECasy

    41/47

    Uniformity

    When you scan a uniform object (water) thepixel value should be zero (for water!).Butsince the machine is very complicatedmechanically this does not happen. Thevalue may drift from day to day/hour by hour.

    If it is scanned, and the pixel value isconstant in all regions of the reconstructedimage, this is called Spatial uniformity

  • 7/29/2019 CT Made ECasy

    42/47

    Uniformity

    There is also a test for this, where you scan abucket of water and plot the numbers along

    an axis of the image. If this axis is within 2standard deviations of the mean value, thesystem has acceptable spatial uniformity.

    Because of the x-ray beam hardening, there

    may be a decrease of CT numbers, so themiddle of the image is darker than it should.

    This is called cupping artefact.

  • 7/29/2019 CT Made ECasy

    43/47

    Summary

    The collimated x-ray beam is directed to thepatient.

    The attenuated image-forming x-ray beam ismeasured by a detector array.

    The signal from the detector array ismeasured by a computer.

    The image is reconstructed in the computer.

    The image is displayed on a TV monitor.

  • 7/29/2019 CT Made ECasy

    44/47

    Summary

    CT makes transverse images (axial images)

    The internal structures of the body attenuate the x-

    ray beam according to their mass density and atomicnumber.

    All data are processed in digital form.

    The resulting computer image is an electronic matrix

    of intensities.

    Matrix size is generally 512x512 individual cells orpixels.

  • 7/29/2019 CT Made ECasy

    45/47

    Summary

    In each pixel is numerical information called a CTnumber or HU.

    The pixel is a two-dimensional representation of acorrensponding tissue volume.

    The voxel (volume element) is determined bymultiplying the square of the pixel size by the

    thickness of the CT scan slice. HU -1000=air, 0= water, 1000= dense bone

  • 7/29/2019 CT Made ECasy

    46/47

    Summary

    The CT scanner has exellent contrastresolution because of the reduction of scatter

    radiation by the x-ray beam collimators. The ability to scan low-contrast anatomic

    structures is limited by the noise of thesystem.

    System noise is determined by the numbersof x-rays used by the detector array toproduce the image.

  • 7/29/2019 CT Made ECasy

    47/47

    Questions...

    ?