CT Dose Optimisation

125
RANZCR QUDI MMC July 2010 Multi Multi-detector CT (MDCT) detector CT (MDCT) Technology & Optimization Technology & Optimization John C.P. Heggie, Ph.D. Multi Multi-detector CT (MDCT) detector CT (MDCT) Technology & Optimization Technology & Optimization John C.P. Heggie, Ph.D. Consultant Medical Physicist Consultant Medical Physicist BreastScreen BreastScreen VIC VI

description

Optimizacion de dosis en CT

Transcript of CT Dose Optimisation

  • RANZCR QUDI MMC July 2010

    MultiMulti--detector CT (MDCT) detector CT (MDCT)

    Technology & OptimizationTechnology & Optimization

    John C.P. Heggie, Ph.D.

    MultiMulti--detector CT (MDCT) detector CT (MDCT)

    Technology & OptimizationTechnology & Optimization

    John C.P. Heggie, Ph.D.

    Consultant Medical PhysicistConsultant Medical Physicist

    BreastScreenBreastScreen VICVI

  • RANZCR QUDI MMC July 2010

    The Evolution of CT from The Evolution of CT from

    Head scan circa 1974 Head scan circa 1974 Matrix 80 x 80Matrix 80 x 80EMI Mark I s canner

  • RANZCR QUDI MMC July 2010

    To To

    Siemens Somatom Sensation 64 and beyond

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

    MDCT What about Dose?MDCT What about Dose?

    The potential The potential downside of downside of MDCTMDCT

  • RANZCR QUDI MMC July 2010

    Topics for DiscussionTopics for Discussion

    Dose terminology for risk assessmentDose terminology for risk assessment

    Should we have any concerns about Should we have any concerns about MDCT patient doses?MDCT patient doses?

    What can we do to keep doses under What can we do to keep doses under control in MDCT? control in MDCT? Understanding t he technology o f MDCTUnderstanding t he technology o f MDCT

    OPTIMIZATIONOPTIMIZATION

  • RANZCR QUDI MMC July 2010

    Useful Dose IndicesUseful Dose Indices

    CTDICTDIvolvol Dose length product (DLP)Dose length product (DLP)

    Effective dose (E)Effective dose (E)

    Absorbed dose (D)Absorbed dose (D)

  • RANZCR QUDI MMC July 2010

    Effective Dose (E)Effective Dose (E)

    Measure of Measure of riskrisk of hereditary effects & of hereditary effects & cancer caused by ionizing radiation cancer caused by ionizing radiation

    Accounts f or varying t issue Accounts f or varying t issue

    radiosensitivitiesradiosensitivities

    Accounts for partial body irradiation Accounts for partial body irradiation

    Measured in Measured in SievertSievert ((SvSv), ), mSvmSv or or SvSv 1 1 SvSv = 1,000 = 1,000 mSvmSv = 1,000,000 = 1,000,000 SvSv

  • RANZCR QUDI MMC July 2010

    Conversion from DLP to E(from ICRP 102)

    Conversion from DLP to E(from ICRP 102)

    Example: Example: AbdoAbdo pelvis CT has DLP of 500 mGy.cm. pelvis CT has DLP of 500 mGy.cm. Effective dose is 500 x 0.015 = 7.5 Effective dose is 500 x 0.015 = 7.5 mSvmSv

    Body part k (mSv/mGy.cm)

    Head & neck 0.0031

    Head 0.0021

    Neck 0.0059

    Chest 0.014

    Abdomen & pelvis 0.015

    Trunk 0.015

  • RANZCR QUDI MMC July 2010

    MDCT Doses SVHM Median ValuesMDCT Doses SVHM Median ValuesCT Examination (Adult) Effective Dose (mSv)

    Brain 1.4

    Chest 4.0

    with contrast 7.2

    Chest/abdo/pelvis with contrast 11

    Abdo/pelvis 4.4

    with contrast 6.0

    triple phase 13

  • RANZCR QUDI MMC July 2010

    Comparison with other modalitiesComparison with other modalitiesProcedureProcedure SVHM Ef fective Dose (SVHM Ef fective Dose (mSvmSv))

    Chest (P/A)Chest (P/A) 0.0230.023

    Abdomen (A/P)Abdomen (A/P) 0.350.35

    Lumbar spine seriesLumbar spine series 0.470.47

    PTCAPTCA 3.13.1

    IVP (8 images)IVP (8 images) 3.73.7

    Bone scan (Bone scan (99m99mTc)Tc) 4.24.2

    Brain SPECT (Brain SPECT (NeuroliteNeurolite)) 6.96.9

    Barium enemaBarium enema 7.97.9

    Natural Natural bgbg/annum/annum 2.02.0

  • RANZCR QUDI MMC July 2010

    CT Organ Doses CT Organ Doses Examination Organ Absorbed Dose (mGy)

    Brain lens 45-50brain 38

    Chest - with contrastlung 11breast 7stomach/liver 14

    Abdo/pelvis - triple phasestomach/liver/kidneys 30-35gonads (f) 13SI 22

  • RANZCR QUDI MMC July 2010

    As far as diagnostic radiological procedures are concerned CT is a

    high dose procedure.

    But are these doses a cause for concern?

    CT DosesCT Doses

  • RANZCR QUDI MMC July 2010

    Risks at CT DosesRisks at CT Doses Brenner et al AJR 2001: 176; 289Brenner et al AJR 2001: 176; 289

    Sparked debate about number of possible Sparked debate about number of possible cancer deaths caused by paediatric CT cancer deaths caused by paediatric CT scanningscanning

    Brenner et al NEJM 2007: 357; 2277Brenner et al NEJM 2007: 357; 2277 Discussed increasing CT usage patterns & Discussed increasing CT usage patterns &

    riskrisk

  • RANZCR QUDI MMC July 2010

    Risks at CT DosesRisks at CT Doses

    What do we really know about cancer risks What do we really know about cancer risks at low dose?at low dose? Most information about human exposure Most information about human exposure

    comes from the 1945 Japanese experimentscomes from the 1945 Japanese experiments

  • RANZCR QUDI MMC July 2010

    Risks at CT DosesRisks at CT Doses

    Japanese LSS low dose data (RERF Japanese LSS low dose data (RERF Study)Study)

    440 deaths attributable to radiation out of ~9350440 deaths attributable to radiation out of ~9350 Risk decreases with increasing age at time of Risk decreases with increasing age at time of

    exposureexposure

    Other sourcesOther sources Canadian & Massachusetts TB fluoroscopy Canadian & Massachusetts TB fluoroscopy

    patientspatients Swedish benign breast disease patientsSwedish benign breast disease patients

    IN SHORT - NOT MUCH!

  • RANZCR QUDI MMC July 2010

    Cancer Mortality Excess Radiation Induced Risk Male 30y

    Cancer Mortality Excess Radiation Induced Risk Male 30y

    0

    0.02

    0.04

    0.06

    0.08

    0.1

    0 25 50 75 100 125 150

    Effective dose (mSv)From uni form whole bo dy e xposure

    E

    x

    c

    e

    s

    s

    R

    e

    l

    a

    t

    i

    v

    e

    R

    i

    s

    k

    Zerorisk

    CT dose range

    Pierce et al: Rad Res 146 (1996) 1Preston et al Rad Res 160 (2003) 381

    Linear model based on risk o ver range 0-2 Sv

  • RANZCR QUDI MMC July 2010

    Cancer Incidence Excess Radiation Induced Risk Male 30y

    Cancer Incidence Excess Radiation Induced Risk Male 30y

    0

    0.05

    0.1

    0.15

    0.2

    0.25

    0.3

    0 50 100 150 200

    E

    x

    c

    e

    s

    s

    R

    e

    l

    a

    t

    i

    v

    e

    R

    i

    s

    k

    Zerorisk

    Pierce & Preston: Rad Res 154 (2000) 178-186

    Effective dose (mSv)From uni form whole bo dy e xposure

    CT dose range

    Linear model based on risk o ver range 0-2 Sv

  • RANZCR QUDI MMC July 2010

    Hypothetical CaseHypothetical Case

    A 30 y old man undergoes two triple phase A 30 y old man undergoes two triple phase CT CT abdoabdo procedures with an effective dose of procedures with an effective dose of ~25 ~25 mSvmSv. What is the hypothetical risk of a . What is the hypothetical risk of a radiation induced cancer in this man?radiation induced cancer in this man?

    Assuming cancer incidence of ~40% in male Assuming cancer incidence of ~40% in male population then:population then:

    AAdditional risk = 0.025 x 40% = 1%dditional risk = 0.025 x 40% = 1% Young children are thought to be more Young children are thought to be more

    radiosensitive radiosensitive risk will be higherrisk will be higher ffood for thought at the very least!

  • RANZCR QUDI MMC July 2010

    Helical CT - Key ParametersHelical CT - Key Parameters

    kVpkVp determines subject contrast determines subject contrast Increasing t he Increasing t he kVpkVp will decrease the subject contrast will decrease the subject contrast Increasing t he Increasing t he kVpkVp will increase dose will increase dose -- all other all other

    parameters being kept constantparameters being kept constant Consider using 80 Consider using 80 kVpkVp or 100 or 100 kVpkVp with children & with children &

    for CT angiography where intrinsic contrast is highfor CT angiography where intrinsic contrast is high

    mAmA & rotation time (& rotation time (mAsmAs)) determines noise & hence low contrast determines noise & hence low contrast detectabilitydetectability Increasing t he Increasing t he mAsmAs increases dose in direct increases dose in direct

    proportionproportion

  • RANZCR QUDI MMC July 2010

    Helical CT - Key ParametersHelical CT - Key Parameters

    Reconstruction algorithm (kernel) Reconstruction algorithm (kernel) High resolution kernel will increase noise

    SD=6.0

    Smoothing filter

    Wire

    SD=32

    Bone filter

    0

    C

    T

    -

    n

    u

    m

    b

    e

    r

    0

    BoneSmoothing

  • RANZCR QUDI MMC July 2010

    Helical CT - Key ParametersHelical CT - Key Parameters

    Reconstruction algorithm (kernel) Reconstruction algorithm (kernel) If high spatial resolution is not crucial then

    reconstruct using a smoothing k ernel

    And/or reconstruct using thicker slices

    To decrease impact of noise in any CT image utilise a smoothing k ernel rather than increase the mAs

  • RANZCR QUDI MMC July 2010

    Helical CT - Key ParametersHelical CT - Key Parameters

    90 mAs Kernel = B80

    Mean = 43 98

    90 mAs Kernel = B41

    Mean = 44 19

    To achieve the same level of noise reduction would require an INCREASE in mAs (dose) by a factor of 25!

  • RANZCR QUDI MMC July 2010

    Helical CT - Key ParametersHelical CT - Key Parameters

    Image slice widthImage slice width impacts on spatial resolution in zimpacts on spatial resolution in z--direction direction

    & image noise & image noise

    Scanned volume Scanned volume Increase the volume and effective dose Increase the volume and effective dose

    increases because more organs includedincreases because more organs included

    PITCH Dictates the total scan time and MAY

    impact on image quality & dose

  • RANZCR QUDI MMC July 2010

    Single slice Helical CT - PitchSingle slice Helical CT - Pitch

    Patient motion

    Distance advanced by table during one rotation of X-ray tube (TF)

    Pitch = table feed per rotation collimated nominal slice width= TF/h

    Pitch values typically 1Pitch values typically 1--22 A pitch = 1 at same mAs A pitch = 1 at same mAs

    has same dose implications has same dose implications as contiguous step and as contiguous step and shoot sequential scanningshoot sequential scanning

    A pitch of >1 at same mAs A pitch of >1 at same mAs implies a dose saving but implies a dose saving but tissues are not scanned tissues are not scanned from all anglesfrom all angles

    A pitch

  • RANZCR QUDI MMC July 2010

    MDCT Helical & PitchMDCT Helical & Pitch A modified definition is required A modified definition is required

    because image slice width & collimated because image slice width & collimated beam width are decoupled:beam width are decoupled:

    Pitch = Table feed per rotationPitch = Table feed per rotationTotalTotal nominal beam widthnominal beam width

    = = TF / (N . h)TF / (N . h)N= Number of detector rows usedN= Number of detector rows used

    With this definition a pitch > 1 implies With this definition a pitch > 1 implies dose reduction as for single slice helical dose reduction as for single slice helical scanningscanning

    Historically we have recommended USING A PITCH > 1

    N . h

    Detector elements

    (4 slices N=4)

  • RANZCR QUDI MMC July 2010

    Effective mAsEffective mAs

    Effective mAs = mA * s / pitchEffective mAs = mA * s / pitch

    = mA * s * beam collimation / table feed= mA * s * beam collimation / table feed

  • RANZCR QUDI MMC July 2010

    Effective mAs & DoseEffective mAs & Dose

    Dose index (Dose index (CTDICTDIvolvol) i s a measure) i s a measure

    of the average absorbed dose to of the average absorbed dose to

    one slice of the patient assuming one slice of the patient assuming

    typical patient dimensionstypical patient dimensions

    CTDICTDIvolvol effective effective mAsmAs= 8.4 = 8.4 mGymGy

    Double the effective Double the effective mAsmAs & & CTDICTDIvolvol doublesdoubles

  • RANZCR QUDI MMC July 2010

    Causes of Dose Increases with MDCT

    Causes of Dose Increases with MDCT

    OverOver--beaming beaming (penumbral effects)(penumbral effects)

    OverOver--scanningscanning (over(over--ranging)ranging)

    MDCT MDCT detector inefficiencies detector inefficiencies

    MDCTMDCT geometrygeometry

    Improved Improved radiation utilisationradiation utilisation

    Increasing use of Increasing use of thin slice reconstructionthin slice reconstruction

  • RANZCR QUDI MMC July 2010

    axis ofrotation

    Single active element detects umbra & penumbra & uses most if

    not all

    Focal spot

    1 x 4 mm single slice CT

    Umbra

    Penumbra4 mm slice includes

    penumbra

    Collimator

    Over beaming (Penumbra)Over beaming (Penumbra)Focal spot

    4 x 1 mm multislice CT

    Each active element detects unequal amounts of the umbra

    Each active element detects the umbra equally

    Collimator opened further

    Focal spot

  • RANZCR QUDI MMC July 2010

    Impact of Total Beam WidthImpact of Total Beam Width

    Unused umbra represents a smaller proportion of beam

    when total beam width (N . h )

    increases

    collimator

    Small beam width N.h Large beam width N.h

  • RANZCR QUDI MMC July 2010

    Over-beamingOver-beaming

    Depends on focal spot size Depends on focal spot size

    OverOver--beaming is increased by having:beaming is increased by having: Large focal spot &Large focal spot &

    Poor radiographic geometry (i.e. short SDD) Poor radiographic geometry (i.e. short SDD)

    OverOver--beaming is common t o both beaming is common t o both sequential (axial) and helical MDCTsequential (axial) and helical MDCT

  • RANZCR QUDI MMC July 2010

    Head scan

    Do the right thing and carefully avoid

    orbits!

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

    Demonstration of Over-beaming (head)

    Demonstration of Over-beaming (head)

    ParameterParameter 12 mm12 mm

    SequentialSequential

    24 mm24 mm

    SequentialSequentialDLP DLP (mGy.cm)(mGy.cm)

    421421 379379

    E (mSv)E (mSv) 1.41.4 1.21.2

    Lens dose Lens dose (mGy)(mGy)

    5.15.1 4.64.6

  • RANZCR QUDI MMC July 2010

    Over-beaming MinimisedOver-beaming Minimised

    Geleijns et al PMB 54 (2009) 3141

    Aquillion One120 kVp

    CTDI normalised to 1.0 for 160 mm collimation

  • RANZCR QUDI MMC July 2010

    Over-beaming MinimisedOver-beaming Minimised

    Note that situation is alleviated somewhat Note that situation is alleviated somewhat with 32 or 64 slice scanners as the with 32 or 64 slice scanners as the unused penumbra b ecomes a smaller unused penumbra b ecomes a smaller proportion of total beam width provided proportion of total beam width provided you use a large you use a large N.hN.h product. product.

    In general recommended that use scan protocols with collimation large (eg. N.h= 64 x 0.5 mm) for all studies except those that require high spatial resolution but there is the then the issue of over-scanning in helical mode!

  • RANZCR QUDI MMC July 2010

    Helical scanners must irradiate a larger Helical scanners must irradiate a larger volume of patient than is ultimately displayed volume of patient than is ultimately displayed in the image data set:in the image data set: Necessary evil to avoid having incomplete data Necessary evil to avoid having incomplete data

    for reconstruction of first and last slices of interest for reconstruction of first and last slices of interest

    Typically an extra rotation is required at Typically an extra rotation is required at beginning a nd end of imaged volumebeginning a nd end of imaged volume

    Common to all scanners used in helical mode Common to all scanners used in helical mode except with a few members of the Siemens except with a few members of the Siemens Definition family of scannersDefinition family of scanners

    Over-scanningOver-scanning

  • RANZCR QUDI MMC July 2010

    Over-scanningOver-scanning

    More significant for MDCTMore significant for MDCT May irradiate several extra cmMay irradiate several extra cm

    Increases patient dose in general but may Increases patient dose in general but may lead to unintended irradiation of key lead to unintended irradiation of key organsorgans Lens of eye in head scanLens of eye in head scan

    Should routinely use sequential scanning unless Should routinely use sequential scanning unless clinically indicated or doing MPR clinically indicated or doing MPR

    Male gonads in abdomen/pelvic scanMale gonads in abdomen/pelvic scan

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

    Outcome of Over-beaming & Over-scanning (head)

    Outcome of Over-beaming & Over-scanning (head)

    ParameterParameter 12 mm12 mm

    SequentialSequential

    24 mm24 mm

    SequentialSequential

    24 mm24 mm

    HelicalHelical

    P =1P =1DLP DLP (mGy.cm)(mGy.cm)

    421421 379379 460460

    E (mSv)E (mSv) 1.41.4 1.21.2 1.51.5

    Lens dose Lens dose (mGy)(mGy)

    5.15.1 4.64.6 35.735.7

  • RANZCR QUDI MMC July 2010

    MDCT Optimization MDCT Optimization

    Perform head scans in sequential mode Perform head scans in sequential mode using broad collimation with angled using broad collimation with angled gantry to avoid orbits gantry to avoid orbits unless o therwise unless o therwise indicated by clinical c onsiderations ( e.g.)indicated by clinical c onsiderations ( e.g.) MPR requiredMPR required

    3D reconstructions required3D reconstructions required

  • RANZCR QUDI MMC July 2010

    Do the right thing and carefully avoid

    testicles!

    Abdomen & pelvis

    scan

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

    Outcome of Over-beaming & Over-scanning (abdo/pelvis)Outcome of Over-beaming & Over-scanning (abdo/pelvis)

    ParameterParameter 12 mm12 mm

    HelicalHelical

    P = 1.0P = 1.0

    24 mm24 mm

    HelicalHelical

    P = 1.0P = 1.0

    24 mm24 mm

    HelicalHelical

    P =0.5P =0.5DLP DLP (mGy.cm)(mGy.cm)

    478478 462462 446446

    E (mSv)E (mSv) 8.48.4 9.29.2 8.28.2

    Testicular Testicular dose (mGy)dose (mGy)

    4.84.8 11.011.0 7.27.2

  • RANZCR QUDI MMC July 2010

    MDCT Optimization MDCT Optimization

    ConsiderConsider using s equential s canning w ith using s equential s canning w ith large collimation when short scan blocks large collimation when short scan blocks intended (e.g. small children)intended (e.g. small children)

    Scan patient anatomy in one large block Scan patient anatomy in one large block rather than two or more smaller blocks:rather than two or more smaller blocks:

    Use pitch < 1.0 !!!!! Use pitch < 1.0 !!!!! PROVIDEDPROVIDED reduce reduce mAsmAs to keep to keep effectiveeffective mAsmAs the samethe same

    There are exceptions to this approach!There are exceptions to this approach!

  • RANZCR QUDI MMC July 2010

    Exceptions to General RuleExceptions to General Rule

    A different level of noise may be tolerated in A different level of noise may be tolerated in one part of anatomy versus anotherone part of anatomy versus another Scan separately with different Scan separately with different mAsmAs

    Combined head/neck scans with some Combined head/neck scans with some 16 slice 16 slice scannerscanner designs would result in a dose penalty designs would result in a dose penalty because head scans ar e usually p erformed with because head scans ar e usually p erformed with less filtration which cannot be increased in less filtration which cannot be increased in middle of scan blockmiddle of scan block dose penalty would more than offset the dose penalty would more than offset the

    advantage of scanning as a single block.advantage of scanning as a single block.

  • RANZCR QUDI MMC July 2010

    MDCT Detector InefficienciesMDCT Detector Inefficiencies The limiting spatial The limiting spatial

    resolution in the zresolution in the z--direction direction is dictated by the width of is dictated by the width of the discrete detector the discrete detector elementselements of necessity there are t iny of necessity there are t iny

    gaps between detector gaps between detector elementselements

    leads to wasted radiation and leads to wasted radiation and a loss of geometric a loss of geometric efficiencyefficiency

    Siemens UFC detector

    GE ceramic detector

  • RANZCR QUDI MMC July 2010

    Single Slice v MDCT GeometrySingle Slice v MDCT Geometry

  • RANZCR QUDI MMC July 2010

    Scan protocols should Scan protocols should NOTNOT be be transferred from one scanner (particularly transferred from one scanner (particularly a single slice CT) to another without due a single slice CT) to another without due consideration to differences in consideration to differences in scanner geometryscanner geometry

    Detector efficiencyDetector efficiency

    filtration.filtration.

    MDCT Detector & GeometryMDCT Detector & Geometry

  • RANZCR QUDI MMC July 2010

    For the CT abdominal example discussed For the CT abdominal example discussed previously, us ing t he same scan protocol previously, us ing t he same scan protocol (effective (effective mAsmAs) would result in a DLP ) would result in a DLP increase of 90% in going to the GE increase of 90% in going to the GE LightspeedLightspeed 16 from the GE CT/16 from the GE CT/ii

    MDCT Detector & GeometryMDCT Detector & Geometry

  • RANZCR QUDI MMC July 2010

    Improved Radiation Utilisation Improved Radiation Utilisation

    A greater proportion of XA greater proportion of X--rays generated rays generated by Xby X--ray tube are utilised with cone ray tube are utilised with cone beam (MDCT) versus fan beam (single beam (MDCT) versus fan beam (single slice). This has advantages from Xslice). This has advantages from X--ray ray tube loading perspective but .. tube loading perspective but ..

  • RANZCR QUDI MMC July 2010

    The DownsideThe Downside

    Volume s canned

    Increase the Increase the scanned scanned volume orvolume or

    Undertake Undertake more phasesmore phases DOSE

    INCREASES Multiphase scan

  • RANZCR QUDI MMC July 2010

    MDCT Optimization MDCT Optimization

    Do Do NOTNOT increase scanned vo lume jus t increase scanned vo lume jus t because you can:because you can: Rather only if clinicallyRather only if clinically justifiedjustified

    MinimiseMinimise the use of multithe use of multi--phase phase examinationsexaminations Do you always need the venous and delayed Do you always need the venous and delayed

    phases?phases?

  • RANZCR QUDI MMC July 2010

    Thin Slice ReconstructionThin Slice Reconstruction

    MDCT allows the same volume to be MDCT allows the same volume to be reconstructed with thin slices without reconstructed with thin slices without increasing scan times increasing scan times However i mage noise increasesHowever i mage noise increases

    Some manufacturers ha ve changed geometry so Some manufacturers ha ve changed geometry so that patient & detector are closer to focus (see that patient & detector are closer to focus (see earlier GE design) and/orearlier GE design) and/or

    Operator may increase tube loading (mAs) to Operator may increase tube loading (mAs) to compensate for increased image noise compensate for increased image noise

    DOSE INCREASES

  • RANZCR QUDI MMC July 2010

    Time Out!Time Out!

  • RANZCR QUDI MMC July 2010

    MDCT OptimizationMDCT Optimization

    Optimization means reducing dose whilst maintaining acceptable image quality

    It is NOT about having either The lowest dose or

    The best image quality

  • RANZCR QUDI MMC July 2010

    MDCT OptimizationMDCT Optimization

    Tailor techniques to clinical Tailor techniques to clinical requirements. Choose:requirements. Choose: scanned vo lume scanned vo lume pitchpitch rotation speed rotation speed and collimation carefullyand collimation carefully

    Adjust Adjust mAsmAs (in absence of AEC) (in absence of AEC) & & kVpkVp to be patient size specificto be patient size specific Paediatric protocolsPaediatric protocols

  • RANZCR QUDI MMC July 2010

    Dose & Technique FactorsDose & Technique Factors

    Dose increases with kVp for same mAsDose increases with kVp for same mAs

    Dose increases wi th volume scanned Dose increases wi th volume scanned

    use small volumeuse small volume

    Dose increases with mAs Dose increases with mAs

    use low mAsuse low mAs

    Dose decreases with pitchDose decreases with pitchDosemAs effective mAs

    pitch

  • RANZCR QUDI MMC July 2010

    MDCT OptimizationMDCT Optimization

    Check your doses against published Check your doses against published Diagnostic Reference Levels ( DRLs)Diagnostic Reference Levels ( DRLs) If others are using much lower doses you If others are using much lower doses you

    probably need to modify your protocolsprobably need to modify your protocols

    Employ Automatic Exposure Control Employ Automatic Exposure Control AEC (AEC (mAmA modulation) technology modulation) technology whenever possible:whenever possible: Smart Smart mAmA -- GEGE Care DoseCare Dose -- SiemensSiemens DoseRightDoseRight -- PhilipsPhilips SureExposureSureExposure -- ToshibaToshiba

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    The scanner adjusts the XThe scanner adjusts the X--ray tube ray tube mAmAautomatically with changes in anatomy automatically with changes in anatomy both during the scan and from patient to both during the scan and from patient to patientpatient Rotational (xRotational (x--y) current modulationy) current modulation

    Longitudinal (ZLongitudinal (Z--axis) current modulationaxis) current modulation

    Combined rotational & zCombined rotational & z--axis modulationaxis modulation

  • RANZCR QUDI MMC July 2010

    Z-axis Current ModulationZ-axis Current Modulation

    The The mAsmAs is altered from rotation to is altered from rotation to rotation as the patient moves through the rotation as the patient moves through the gantrygantry Uses attenuation data generated from a Uses attenuation data generated from a

    SPR (Scout) to determine required SPR (Scout) to determine required mAsmAs as as a function of distance (z)a function of distance (z) To be fully successful need to ensure the SPR To be fully successful need to ensure the SPR

    is acquired is acquired over the full length of the intended helical acquisitionover the full length of the intended helical acquisition at the same at the same kVpkVp as the helical acquisitionas the helical acquisition

  • RANZCR QUDI MMC July 2010

    Effective mAs versus distanceReference effective mAs = 300

    0

    50

    100

    150

    200

    250

    0 50 100 150 200 250Relative distance (mm)

    E

    f

    f

    e

    c

    t

    i

    v

    e

    m

    A

    s

    Lung apex 160 mAs

    Mid cervical 58 mAs

    Orbits 75 mAs

    SPR Attenuation MeasurementsSPR Attenuation Measurements

    Aortic arch 200 mAs

  • RANZCR QUDI MMC July 2010

    Rotational Current ModulationRotational Current Modulation

    The The mAmA is altered as Xis altered as X--ray tube rotates ray tube rotates about patient about patient

  • RANZCR QUDI MMC July 2010

    Rotational Current ModulationRotational Current Modulation

    The The mAmA is altered as Xis altered as X--ray tube rotates ray tube rotates about patient about patient It may use angleIt may use angle--dependent a ttenuation dependent a ttenuation

    information on the fly information on the fly -- Siemens & PhilipsSiemens & Philips

    OROR AP and Lateral SPRs AP and Lateral SPRs GE & ToshibaGE & Toshiba

    Typically the Typically the mAmA used in the AP & PA used in the AP & PA projections will be significantly less than projections will be significantly less than when the lateral projections are being when the lateral projections are being acquiredacquired

  • RANZCR QUDI MMC July 2010

    Image noise is largely dictated by the projections in which the radiation is most strongly attenuated

    Why Use Rotational Current Modulation?

    Lateral projection

    AP projection

    Intensity IAP = IlatNoise sdAP = sdlat

    AP projection

    Intensity IAP > IlatNoise sdAP < sdlat

  • RANZCR QUDI MMC July 2010

    (b) Constant tube current

    (c) mA modulation - for the same dose level the noise can be significantly reduced

    (d) & (e) or for the same noise level the dose can be reduced, for example by 32%

    Rotational Current Modulation

    sd = 33 sd = 25

    (from Kalendar 2005)

  • RANZCR QUDI MMC July 2010

    (a) Typical mA and noise distributions in scanning with constant mA. Noise (white dots) is highly variable

    AEC Controlled mA Modulation

  • RANZCR QUDI MMC July 2010

    (a) Typical mA & noise distributions in scanning with constant mA. Noise (white dots) is highly variable

    (b) AEC-controlledscanning with the mA adapted to the attenuation & modulated as a function of projection angle. A constant noise level is achieved at significantly reduced dose.

    AEC Controlled mA Modulation

    (from Kalendar 2005)

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlWARNING!

    Automatic Exposure ControlWARNING!

    The use of AEC does NOT in itself The use of AEC does NOT in itself guarantee dose reductions but should guarantee dose reductions but should ensure ensure consitentconsitent image qualityimage quality

    In film/screen radiography the film In film/screen radiography the film blackness (Optical density) controls doseblackness (Optical density) controls dose

    No such physical limitation applies in No such physical limitation applies in CT or any other digital modalityCT or any other digital modality

    REQUIRES HUMAN INTERVENTIONREQUIRES HUMAN INTERVENTION

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    Clinical d ecisions need to be made as Clinical d ecisions need to be made as to what constitutes acceptable image to what constitutes acceptable image quality or noise. Scanners require input quality or noise. Scanners require input of either:of either: a soa so--called predetermined called predetermined referencereference noise noise

    index (NI) or standard deviation (SD) orindex (NI) or standard deviation (SD) or a a reference effective reference effective mAsmAs judged to be judged to be

    suitable for imaging the suitable for imaging the average sizedaverage sized (80 (80 kg) patientkg) patient

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    Once decision is made the AEC should Once decision is made the AEC should ensure consistent image quality and ensure consistent image quality and inevitable r educe doses.inevitable r educe doses.

  • RANZCR QUDI MMC July 2010

    Suppose noise index of about 30 HU is desired for coronary Suppose noise index of about 30 HU is desired for coronary calcium measurements. The calcium measurements. The mAsmAs values can be reduced values can be reduced significantly for smaller patient crosssignificantly for smaller patient cross--sections, shown here for a sections, shown here for a Cardiac CT phantom with 40 and 35 cm lateral diameter, Cardiac CT phantom with 40 and 35 cm lateral diameter, respectively. Dose is reduced by ~ 65% when the respectively. Dose is reduced by ~ 65% when the mAsmAs is lowered is lowered from 200 to 69 under AEC. from 200 to 69 under AEC.

    (From Kalendar 2005)(From Kalendar 2005)

  • RANZCR QUDI MMC July 2010

    Toshiba Aquillion 16Toshiba Aquillion 16

  • RANZCR QUDI MMC July 2010

    Toshiba Aquillion 16Toshiba Aquillion 16

    Standard deviation is measure of noise

  • RANZCR QUDI MMC July 2010

    Toshiba Aquillion 16Toshiba Aquillion 16Scanner indicates the

    variation of mA on the SPR

  • RANZCR QUDI MMC July 2010

    GE VCTGE VCT

  • RANZCR QUDI MMC July 2010

    GE VCT ScreenGE VCT Screen

    These arrows adjust NI by 5% and hence dose by 10%

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    Decision about the reference noise index Decision about the reference noise index should be made on a protocol by protocol should be made on a protocol by protocol basis.basis. The nonThe non--contrast phase of contrast phase of abdoabdo/pelvis scan, for /pelvis scan, for

    example, does not need the same low noise example, does not need the same low noise requirement as the contrast phasesrequirement as the contrast phases Set reference noise index higher (or reference Set reference noise index higher (or reference mAsmAs

    level lower)level lower)

    The The mAsmAs actually utilized for a particular actually utilized for a particular patient may be very patient may be very much l owermuch l ower for slightly for slightly built patients than the maximum value the built patients than the maximum value the AEC allowsAEC allows

  • RANZCR QUDI MMC July 2010

    AEC Controlled mA Modulation6 year old child scanned using adult protocol. Reference effective mAs = 165 and mean actual effective mAs = 38

    From McCollough et al Radiographics 26 (2006) p 503 -513

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    The correct choice of noise index (NI) or The correct choice of noise index (NI) or standard deviation (SD) to be used is dictated standard deviation (SD) to be used is dictated by the image slice width of the by the image slice width of the first series in first series in the reconstruction.the reconstruction. A noise index of 12 may seem appropriate for A noise index of 12 may seem appropriate for

    abdominal imaging with 5 mm image slicesabdominal imaging with 5 mm image slices This would be totally inappropriate for This would be totally inappropriate for

    reconstruction with 0.625 mm slices!! If this was reconstruction with 0.625 mm slices!! If this was done the implication would be an increase in dose done the implication would be an increase in dose by a factor of 8!!by a factor of 8!!

    Thin slices should be reconstructed with much Thin slices should be reconstructed with much higher values of the NI.higher values of the NI.

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    A modest change of 10% in NI results in a A modest change of 10% in NI results in a change in dose of change in dose of 20%.20%. For example, a change in the NI from 12 to 14 For example, a change in the NI from 12 to 14

    implies a dose reduction of implies a dose reduction of 36%!!36%!! Such a change is barely perceptible in the Such a change is barely perceptible in the

    imagesimages OPTIMIZATION is not about tinkering at

    the edges large dose savings can be made with little implication for image quality

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    While the avoidance of both extremely low While the avoidance of both extremely low and extremely high tube currentand extremely high tube currenttime pr oduct time pr oduct values is necessary with regard to the xvalues is necessary with regard to the x--ray ray generator, we also found it most appropriate generator, we also found it most appropriate from a patient dose and image quality from a patient dose and image quality perspective. More aggressive dose reduction perspective. More aggressive dose reduction was not acceptable in children, and more was not acceptable in children, and more aggressive dose increase was not necessary in aggressive dose increase was not necessary in obese patients. obese patients. McCollough et alMcCollough et al

  • RANZCR QUDI MMC July 2010

    Automatic Exposure ControlAutomatic Exposure Control

    Translated:Translated: You require You require lessless noise with childrennoise with children

    You can tolerate You can tolerate moremore noise with obese noise with obese patientspatients

    The NOISE INDEX should be changed The NOISE INDEX should be changed when changes in patient size are extreme.when changes in patient size are extreme.

  • RANZCR QUDI MMC July 2010

    Optimization (early days before AEC)Optimization

    (early days before AEC)

    Effective mAs = 380Effective mAs = 380 Effective Effective mAsmAs = 300= 300Equivalent change in NI Equivalent change in NI 12%12%

  • RANZCR QUDI MMC July 2010

    Optimization(with CareDose)Optimization(with CareDose)

    Reference effective mAs = 180Reference effective mAs = 180Effective mAs = 149Effective mAs = 149

    Reference effective Reference effective mAsmAs = 120= 120Effective Effective mAsmAs = 105= 105Equivalent change in NI Equivalent change in NI of of 20%20%

  • RANZCR QUDI MMC July 2010

    OptimizationOptimization

    Fixed effective Fixed effective mAsmAs = 360= 360

    3mm sliceB60s

    L5

    3mm sliceB60s

    L5

    Effective Effective mAsmAs = 232= 232Equivalent change in NI Equivalent change in NI of of 25%25%

  • RANZCR QUDI MMC July 2010

    Optimization - SummaryOptimization - Summary Perform head scans in sequential mode using wide

    collimation with angled gantry to avoid orbits

    Consider using wide collimation to minimise over-beaming and scan patient anatomy in one large block to minimise over-scanning rather than two or more smaller blocks

    Consider using sequential scanning with wide collimation when scanning short scan blocks to avoid over-scanning issues

    Minimise use of multi-phasic examinations

  • RANZCR QUDI MMC July 2010

    Optimization - SummaryOptimization - Summary Do Do NOTNOT transfer scan protocols from one scanner to transfer scan protocols from one scanner to

    another without due consideration to differences in another without due consideration to differences in scanner geometry and filtrationscanner geometry and filtration

    In absence of AEC keep the In absence of AEC keep the effective mAseffective mAs as low as as low as clinically indicatedclinically indicated Effective mAs = mAs/pitch Effective mAs = mAs/pitch

    Use tube current modulation technology (AEC) Carefully select reference effective mAs or noise index

    Do NOT accept manufacturers default scan protocols as being optimized

  • RANZCR QUDI MMC July 2010

    RecommendationsRecommendations Optimization of scan protocols depends upon the Optimization of scan protocols depends upon the

    technical expertise of clinical users (radiographers & technical expertise of clinical users (radiographers & radiologists) and scientific support staff. radiologists) and scientific support staff.

    Thus, before attempting to optimize scan protocols, Thus, before attempting to optimize scan protocols, radiographers, radiologists and support staff need to radiographers, radiologists and support staff need to understand the scanner technology and to appreciate understand the scanner technology and to appreciate clinical imaging requirements.clinical imaging requirements.

    Practices should work towards development of Practices should work towards development of optimized doseoptimized dose--image q uality relationship(s).image q uality relationship(s).

    When choosing/optimizing protocols take care to When choosing/optimizing protocols take care to ensure that they are tailored to clinical need & patient ensure that they are tailored to clinical need & patient size size -- no universal right dose answersno universal right dose answers..

  • RANZCR QUDI MMC July 2010

  • RANZCR QUDI MMC July 2010

    CT doses are of a magnitude that may represent a risk of cancer in the patient.

    MDCT doses may be higher t han single slice CT doses

    Many factors may contribute to increased patient dose when MDCT is utilised.

    Care must be taken in choosing protocols so that they are tailored to clinical need & patient size.

    SummarySummary

  • RANZCR QUDI MMC July 2010

    ReferencesReferences Heggie JCP et al Importance in optimization of multiHeggie JCP et al Importance in optimization of multi--slice slice

    computed tomography scan protocols Austral Radiol (2006) computed tomography scan protocols Austral Radiol (2006) 50, 27850, 278--285.285.

    Kalra M et al Strategies for CT radiation dose optimization Kalra M et al Strategies for CT radiation dose optimization Radiol (2004) 230, 619Radiol (2004) 230, 619--628.628.

    McCollough CH et al CT Dose Reduction and Dose McCollough CH et al CT Dose Reduction and Dose Management Tools: Overview of Available Options Management Tools: Overview of Available Options RadioGraphics (2006) 26, 503RadioGraphics (2006) 26, 503512.512.

    ICRP Publication 102 (2007). ICRP Publication 102 (2007).

    Heggie JCP Patient doses in multiHeggie JCP Patient doses in multi--slice CT and the slice CT and the importance of optimization Austral Eng Phys Sci Med importance of optimization Austral Eng Phys Sci Med (2005) 28, 86(2005) 28, 86--96.96.

  • RANZCR QUDI MMC July 2010

    In the end there was a beginning!In the end there was a beginning!

    Pregnancy diagnosed by CTPregnancy diagnosed by CT

  • RANZCR QUDI MMC July 2010

    The Public Perception of Radiation!

    The Public Perception of Radiation!

    He has grown a foot since I last saw him

  • RANZCR QUDI MMC July 2010

    Breast Dose & Lifetime Attributable Risk of Cancer

    Induction per 100,000

    Breast Dose & Lifetime Attributable Risk of Cancer

    Induction per 100,000Age at exposure (yr)

    Dose (mGy) 15 30 50

    10 55 25 7

    50 270 126 35

    100 553 253 70

    Derived from BEIR VII Report 2006

  • RANZCR QUDI MMC July 2010

    Achievable MDCT Doses Achievable MDCT Doses

    Examination (Adult) DLP (mGy.cm) E(mSv)

    Brain 610 1.4

    CTPA 160 3.3

    KUB 240 4.2

    Lumbar spine 435 6.9

  • RANZCR QUDI MMC July 2010

    Additional ReadingAdditional Reading

  • RANZCR QUDI MMC July 2010

    The Physicists Compliant Friend:Mr/Mrs Walker

    The Physicists Compliant Friend:Mr/Mrs Walker

    ImPACT

    CT Patient Dosimetry Calculator

    version 0.99x, 20/01/2006

    CT-Expo V 1.5.1 Nov 2005

  • RANZCR QUDI MMC July 2010

    Siemens Sensation 16Siemens Sensation 16ProcedureProcedure Reference Effective mAsReference Effective mAsHeadHead 260260Chest routineChest routine 9090Chest with portal venousChest with portal venous 90/15090/150Chest HRChest HR 100100Chest/Abdo/pelvisChest/Abdo/pelvis 90/15090/150Abdo/Pelvis with contrastAbdo/Pelvis with contrast 150150Abdo/Pelvis triple phaseAbdo/Pelvis triple phase 120/150/150120/150/150Renal (KUB)Renal (KUB) 120120Lumbar s pineLumbar spine 300300

  • RANZCR QUDI MMC July 2010

    Focal Spot WanderingFocal Spot Wandering Thermal/mechanical Thermal/mechanical

    effects cause focal effects cause focal spot position to move spot position to move during scanningduring scanning Some manufacturers Some manufacturers

    increase umbra & increase umbra & signals from two signals from two detector arrays detector arrays outsideoutside imaging imaging volume are used to volume are used to control position of control position of collimatorscollimators

    DOSE INCREASES

    Signals from outside detector elements are used to monitor X-ray

    output

    X-ray beam in red contributes to dose

    but not images

    collimator

  • RANZCR QUDI MMC July 2010

    GEs Dynamic RollersGEs Dynamic Rollers

    Computer controls rollers to adjust position

    of collimator laterally maintaining constant X-

    ray flux over central image detector rows

    Signals from outside rows of detector elements fed to

    computer

  • RANZCR QUDI MMC July 2010

    Rotate - RotateRotate - Rotate 3rd generation3rd generation

    Fan XFan X--ray beamray beam XX--ray tube & detector ray tube & detector

    array rotate as one about array rotate as one about patientpatient

    600600--1000 elements1000 elements Whole body subsecond Whole body subsecond

    single slice scanssingle slice scans PrePre--patient collimation patient collimation

    defines slice width ~1defines slice width ~1--10 10 mmmm

    X-ray tube

    Detector array

    Bow tie filter

    PatientProfile

  • RANZCR QUDI MMC July 2010

    Multi-slice Helical CT (3)Multi-slice Helical CT (3)

  • RANZCR QUDI MMC July 2010

    Multi-slice Helical CT (4)Multi-slice Helical CT (4) Now use Cone Beam Reconstruction:Now use Cone Beam Reconstruction:

    Corrects for finite angle of cone beamCorrects for finite angle of cone beam

    In reconstructing a particular axial image slice uses In reconstructing a particular axial image slice uses only those detector elements that instantaneously view only those detector elements that instantaneously view the slice of interest the slice of interest not ones viewing adjacent tissues.not ones viewing adjacent tissues.

    As table advances and detector rotates different detector As table advances and detector rotates different detector elements will contribute to the image sliceelements will contribute to the image slice

    Improved imaging in sagittal, coronal & oblique Improved imaging in sagittal, coronal & oblique planesplanes

    Improved volume imaging allowing CTImproved volume imaging allowing CT--angiography, ECG gated cardiac studies etc.angiography, ECG gated cardiac studies etc.

  • RANZCR QUDI MMC July 2010

    Detectors & X-ray Beam Width

    Routine abdo scan

    Detectors & X-ray Beam Width

    Routine abdo scan

  • RANZCR QUDI MMC July 2010

    Pitch & Table FeedRoutine abdo scan

    Pitch & Table FeedRoutine abdo scan

    Pitch Pitch = table feed per rotation / beam collimation= table feed per rotation / beam collimation

    = 36 / 24 = 1.5= 36 / 24 = 1.5

  • RANZCR QUDI MMC July 2010

    Effective mAsEffective mAs

    Effective mAs = mA * s / pitchEffective mAs = mA * s / pitch

    = mA * s * beam collimation / table feed= mA * s * beam collimation / table feed

  • RANZCR QUDI MMC July 2010

    Effective mAs & DoseEffective mAs & Dose

    Dose index (CTDIDose index (CTDIvolvol) i s a measure) i s a measure

    of the average absorbed dose to of the average absorbed dose to

    one slice of the patient assuming one slice of the patient assuming

    typical patient dimensionstypical patient dimensions

    CTDICTDIvolvol effective mAseffective mAs= 8.4 mGy= 8.4 mGy

    Double the effective mAs & Double the effective mAs & CTDICTDIvolvol doublesdoubles

  • RANZCR QUDI MMC July 2010

    Care Dose & CareDose 4DCare Dose & CareDose 4D

    This is Siemens version of AEC This is Siemens version of AEC

    Incorporates rotational and ZIncorporates rotational and Z--axis axis

    modulation of the tube currentmodulation of the tube current

    The The Eff mAsEff mAs is the reference is the reference

    mAs that has been chosen from mAs that has been chosen from

    clinical experience to be clinical experience to be

    appropriate for imaging the average appropriate for imaging the average

    adult patient adult patient

  • RANZCR QUDI MMC July 2010

    Multi-slice Helical CT (2)Multi-slice Helical CT (2)

    Cone beamFan beam

  • RANZCR QUDI MMC July 2010

    Care DoseCare Dose

    Constant current

    Attenuation correctionfrom topogram

    Shoulders

    Chest

    Abdo

    Pelvis

    Modulation on the fly

    as it rotates

    Current (mA)

    Distance z-axis

  • RANZCR QUDI MMC July 2010

    Helical ScanningHelical Scanning

    Patient motion

    Path o f continuously rotating X-ray tube relative to patient

    33rdrd generation generation plus: Slip ring technology Slip ring technology

    allows for continuously allows for continuously rotating Xrotating X--ray tube with ray tube with transfer of low voltage transfer of low voltage and data from/to outside and data from/to outside worldworld

    Continuous advance of Continuous advance of patient couch @ 1patient couch @ 1--10 10 mm/smm/s

    Single breath hold Single breath hold imaging (~60 s)imaging (~60 s)

  • RANZCR QUDI MMC July 2010

    MDCTMDCT Helical plus:Helical plus:

    2 to 64 discrete 2 to 64 discrete detector rowsdetector rows

    Reconstructed Reconstructed image slices of image slices of between 0.5 & between 0.5 & 10 mm10 mm

    Improved Improved volume imaging volume imaging AngiographyAngiography

    Calcium Calcium scoring etc.scoring etc.

  • RANZCR QUDI MMC July 2010(from Kalendar 2005)

  • RANZCR QUDI MMC July 2010

    Siemens Sensation 16 (with CareDose 4D)

    Siemens Sensation 16 (with CareDose 4D)

    Cerebrum: Reference Effective mAs = 260

    0

    5

    10

    15

    20

    25

    30

    180 190 200 210 220 230 240 250 260 >260

    Effective mAs

    F

    r

    e

    q

    u

    e

    n

    c

    y

  • RANZCR QUDI MMC July 2010

    Siemens Sensation 16 (with CareDose 4D)

    Siemens Sensation 16 (with CareDose 4D)

    Abdo / Pelvis contrast: Reference Effective mAs = 150

    0

    1

    2

    3

    4

    5

    6

    7

    8

    70 80 90 100 110 120 130 140 150 160 170 >180

    Effective mAs

    F

    r

    e

    q

    u

    e

    n

    c

    y

  • RANZCR QUDI MMC July 2010

    Siemens Sensation 16 (with CareDose)

    Siemens Sensation 16 (with CareDose)

    Abdomen/pelvis with Contrast

    0

    5

    10

    15

    20

    25

    30

    35

    40

    250 300 350 400 450 500 550 600 650 >650

    DLP (mGy.cm)

    F

    r

    e

    q

    u

    e

    n

    c

    y

    Ref effective mAs = 180Ref effective mAs = 120

  • RANZCR QUDI MMC July 2010

    Siemens Sensation 16 (with CareDose)

    Siemens Sensation 16 (with CareDose)

    Lumbar Spine

    0

    5

    10

    15

    20

    25

    300 350 400 450 500 550 600 650 700 >700

    DLP (mGy.cm)

    F

    r

    r

    e

    q

    u

    e

    n

    c

    y

    Fixed effective mAs = 360Ref effective mAs = 300

  • RANZCR QUDI MMC July 2010

    MDCT Doses DLP (mGy.cm) MDCT Doses DLP (mGy.cm) Examination (Adult) CDose1

    Brain 960

    Chest 220

    with contrast 485

    CAP with contrast 765

    AP with contrast 555

    triple phase 1155

    Lumbar spine 560

  • RANZCR QUDI MMC July 2010

    MDCT Doses DLP (mGy.cm) MDCT Doses DLP (mGy.cm) Examination (Adult) CDose1 CDose2

    Brain 960 660

    Chest 220 195

    with contrast 485 370

    CAP with contrast 765 560

    AP with contrast 555 345

    triple phase 1155 725

    Lumbar spine 560 455455

  • RANZCR QUDI MMC July 2010

    MDCT Doses DLP (mGy.cm) MDCT Doses DLP (mGy.cm) Examination (Adult) CDose1 CDose2 CD4D

    Brain 960 660 605

    Chest 220 195 190

    with contrast 485 370 370

    CAP with contrast 765 560 555

    AP with contrast 555 345 325

    triple phase 1155 725 700

    Lumbar spine 560 455455 420420