  • CryoTrapFor Agilent Technologies 7890/6890/6850 GC Systems

    CryoTrap

  • Bad peak shapes and interferences of low boiling compounds are often observed in case of using a sample introduction other than liquid injection, such as headspace, thermal desorption or sampling loop for gas analysis.

    In order to improve peak shape and peak resolution in the chromatogram it is required to refocus high

    volatile components in the front of the column before chromatographic separation. Otherwise incorrect peak

    identification and peak integration may occur.

    The JAS double chamber CryoTrap is a powerful tool of high quality to efficiently trap volatile compounds in

    the front part of a capillary column. It combines the latest advances in CryoTrap technology including low

    thermal mass, efficient insulation and maximum heating power - features that ensure a fast heating and an

    economical cooling. Various appropriate coolants can be used to cool the trap (liquid carbon dioxide, liquid

    nitrogen and pressurized air). An extremely fast but controlled heating of the trapping zone up to 700°C/ min

    allows a rapid re-vaporization of the trapped components for subsequently GC analysis – replacing a

    high speed injection. As a result low boiling compounds occur as sharp peaks with a high resolution in

    the chromatogram. JAS CryoTrap can be actuated either due to control software embedded into Agilent

    ChemStation or by GC keyboard.

    JAS CryoTrap is provided to be fully integrated in

    Agilent 6890 and 7890 GCs. It can be positioned

    inside the GC oven, either at the left hand side wall of

    the GC oven or below the valve cabinet. The devise

    requires one heated zone (150W) and one pneumatic

    inlet slot of the GC. In case of Multi Dimensional GC

    applications, JAS CryoTrap can also be mounted

    outside the GC oven and connected to a second GC

    by means of an optional heated transfer line.

    Features

    Features

    The main advantageous features of JAS CryoTrap system are:

    Flexible In order to meet your analytical needs JAS CryoTrap can be installed at various positions (two locations inside the GC oven and one location outside). Additionally, three different cooling options (liquid CO2, liquid N2 and pressurized air) provide complete flexibility to guarantee the most effective analytical conditions.

    Efficient By means of a double chamber design the JAS CryoTrap features an effective insulation reducing the coolant consumption to a minimum.

    Easy to Use Dedicated and optimized alignment of tuning parameters provides excellent performance at any oven temperature

    Economical JAS CryoTrap is controlled either by Agilent ChemStation software or by GC keyboard. The use of the Trap is easy and may be embedded into every GC method.

    Fig. 1

    JAS CryoTrap

  • Column in Column in

    Column out

    OUT

    Liquid coolant

    IN

    Gaseous coolant

    Column out

    HeatingCatridge

    Cooling Phase Heating Phase

    Operating Mode

    1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00

    1000000

    2000000

    3000000

    4000000

    5000000

    6000000

    7000000

    8000000

    9000000

    1e+07

    1.1e+07

    1.2e+07

    1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00

    5000000

    1e+07

    1.5e+07

    2e+07

    2.5e+07

    3e+07

    3.5e+07

    4e+07

    4.5e+07

    5e+07

    2

    Time (min)

    Abu

    ndan

    ce

    Abu

    ndan

    ce

    Ace

    ticac

    id

    Furfu

    ral 3-

    Met

    hylh

    ydan

    toin

    Without CryoTrap

    GC-Method: 40°C for 6 min, with 20°C/min to 300°C

    Time (min)

    Ace

    ticac

    id Furfu

    ral

    3-M

    ethy

    lhyd

    anto

    in

    CO

    With CryoTrap

    GC-Method: 40°C for 4 min, with 15°C/min to 300°C

    The outer wall of the JAS CryoTrap is made of stainless steel. In order to resist thermal fluctuation a tube of a

    specific material for optimal heat carriage is inserted into the CryoTrap body.

    The analytical column is guided through the CryoTrap chamber. Due to appropriate coolants the trap is cooled

    down to the effective temperature and components transferred from the inlet system to the analytical column

    are trapped before chromatographic separation. Re-vaporization for subsequent GC analysis is caused by

    means of rapid but controlled heating of the trapping zone. Temperature and time of trapping process can be

    optimized depending on the analytes and may easily be adjusted using the control software. A fast heating

    rate is important in order to obtain sharp peaks of volatile components. As a result peak broadering and

    interferences are minimized and quantitative evaluation becomes more precise.

    Chromatograms of a wood sample are illustrated for comparison. Figure 3a shows the bad quality of

    peak resolution for high volatile components, like acidic acid or furfural, in case no Cryo Trap is used. An

    identification and integration of these peaks with reasonable certainty is not possible. Figure 3b demonstrates

    the necessity of using a CryoTrap in order to improve the peak resolution of low boiling compounds.

    Fig. 2

    Functionality of JAS CryoTrap

    Fig. 3a

    Analyses of high volatile compounds in wood samples

    Fig. 3b

  • 2.5e+07

    Benz

    ene

    1.5e+07

    2e+07

    TD-GC/MS: Thermal desorpt ion(200→320°C wi th CryoTrap)

    5000000

    1e+07

    HC

    l

    Hydrochloric acid8000000 Ion 36.00 (35.70 to 36.70):

    Carbon dioxidem/z=44

    m/z=36

    3000000

    Ion 44.00 (43.70 to 44.70):

    Acetic acidm/z=601500000

    Ion 60.00 (59.70 to 60.70)

    /1500000

    Ion 100.00 (99.70 to 100.70)

    m/z=100 2-propenoic acid (methyl ester)

    0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5

    TD-GC/MS: Thermal desorption(200→320°C without Cryo Trap)

    1400000

    Abundance

    Column: UA5 (30m, x 0.25mm x 0.25μm)PyrolyzerFurnace: 200ºC→320ºC (20°C/min)Carrier gas: He 1ml/minSplit ratio: 1:50GC oven: 40→ 20ºC/min→320ºC (1min)Injector: 320ºC

    1100000

    1400000

    These are residual volatiles which can be separated by cryo-trapping during desorption.

    800000

    500000

    1 00 00 10 00

    200000

    1.00 5.00 10.00Time (min)

    Operating Mode

    Figure 4:

    Comparison of thermal desorption analysis (200– 320°C)

    with (4a) and without (4b) Cryo Trap (TD-GC/MS system)

    4a

    4b

    Thermal desorption often results in broad initial

    peaks including numerous individual components.

    While reducing of the initial column temperature or

    using thick film columns will just narrow this initial

    band width, cryo-focussing is the most efficient and

    cost effective way of ensuring peak separation.

    Figure 4a and b illustrate the separation

    improvement of the volatile fraction by means of

    cryo-trapping. Qualitative and quantitative data

    quality will be enhanced tremendously.

  • Specifications

    Specifications of JAS CryoTrap

    Description Double chamber CryoTrap design with combined inlet/exhaust port

    Order Numbers for GC 6890: for GC 7890:

    JAS 65600 JAS 66600 - for air cooling JAS 66601 - for liquid N2 JAS 66602 - for liquid CO2

    Instrument Dimensions Length: 105 mm Width: 20 mm (o.d.) / 1 mm (i.d.)

    Compatibility Agilent GC 7890/6890/6850

    Power Int. Power Supply 24V, Input 150 VA

    Maximum Temperature Agilent GC 6890: 500°C Agilent GC 7890: 450°C

    Cooling Temperature Liquid N2: -160°C Liquid CO2: -60°C

    Pressurized Air: ambient temperature or oven start temperature

    Heating Rate 1-700°C/min

    In case of Multi Dimensional GC application the JAS CryoTrap is positioned outside the oven to trap

    components eluting from the first separation column.

    Using liquid nitrogen as cooling agent a dewar vessel with a dip tube is required (minimum 1 bar head

    pressure for full perfomance mandatory). In case of liquid carbon dioxide as cooling agent a cylinder with

    a dip tube and a variable restriction is required.

  • Carl-Zeiss-Str. 49

    47445 Moers

    Germany

    Phone: +49 2841 9871 100

    Fax: +49 2841 9871 222

    e-Mail: [email protected]

    Internet: www.jas.de

    joint analytical systems GmbH

    About joint analytical systems

    Since 1995 JAS is a major partner and Value Added Reseller of Agilent Technologies. We are an innovative-driven organization that offers customized solutions for GC, GCxGC, µGC, GC-AED, GC-MS, GC-QQQ, LC, LC-MS, LC-QQQ and Q-TOF LC-MS applications.

    JAS serves key industries such as

    Chemical•Environmental•FoodandFlavor•Forensic•MechanicalEngineering•Petrochemical•Pharmaceutical•

    JAS Products for GC

    AtomicEmissionDetector•CryoTrap•CustomizedValvingSystem•EzPrep-PreparativeFractionCollector•Olfactometer•Scotti-RFIDSystems•UNISInlets•

    Copyright © 2013

    joint analytical systems GmbH

    All rights reserved