Credentialing Results from the RPC’s Spine
Anthropomorphic Phantom
Andrea Molineu, Paola Alvarez, Nadia Hernandez, Fang-Fang Yin,
and David FollowillAugust 2012
CPR
lung
lung
heart
Spine phantom
CPR
lung
lung
heart
Spine phantom
PTV
bonespinal cord
esophagus
CPR
RTOG 0631
• Phase II/III study of IGRT/SBRT
for localized spine mets
• Facility Questionnaire
• IGRT credentialing
• Phantom credentialing
CPR
Treatment Plan• 6 Gy to at least 90% of PTV
• Min PTV dose at least 5.25 Gy
Normal structure
Volume Volume Max (Gy)
Max Point Dose (Gy)
Spinal Cord <0.35 cc<1.2 cc
3.752.63 5.25
Heart <15 cc 6.00 8.25
Esophagus <5 cc 4.46 6.00
Skin <10 cc 8.63 9.75
Whole Lung 1000 cc 2.78CPR
Representative Treatment Plan
CPR
Criteria
• TLD±7%
• Film (axial and sagittal planes)85% meet 5%/3mm
• Plan DVHmust comply with protocol
CPR
Criteria
• TLD±7%
• Film (axial and sagittal planes)85% meet 5%/3mm
• Plan DVHmust comply with protocol
CPR
Good Agreement
PlanMeasurement
PTV
Cord/boneOAR
CPR
Good Agreement
Gamma
Measurement100%
passing
CPR
Poor AgreementPTV
Cord/boneOAR
Measurement Plan CPR
Poor Agreement
Gamma
Measurement100%
passing26%
passing
CPR
Overall results
• 138 irradiations at 98 institutions
• 44 irradiations failed
• 1 failed TLD criteria only
• 29 failed gamma criteria only
• 14 failed both TLD and gamma
• 64 sites credentialed for IGRT by RTOG
CPR
Spine Results
PTV HeartAxial
GammaSagittal Gamma
mean 0.99 0.94 87.7 88.4
std dev 0.04 0.15 13.8 13.2
range 0.88 – 1.23 0.67 – 1.51 22 - 100 36 - 100
CPR
Results by Machine
Machine Manufacturer Pass Rate (%) Attempts
Criteria Failed
TLD Dose Gamma Dose and Gamma
CyberKnife 64 22 1 3 4
Elekta 56 18 0 3 5
Novalis 69 13 0 3 1
Siemens 67 3 0 1 0
TomoTherapy 79 14 0 3 0
Varian 71 68 0 16 4
Total 138 1 29 14
CPR
Results by TPS
TPS Pass Rate (%) Attempts
Criteria Failed
TLD Dose Gamma Dose and
Gamma
CyberKinfe 64 22 1 3 4
Eclipse 68 53 0 14 3
Hi-Art 79 14 0 3 0
iPlan 71 17 0 3 2
Monaco 100 1 0 0 0
Pinnacle 68 25 0 4 4
XiO 50 6 0 2 1
total 138 1 29 14
CPR
Results by AlgorithmAlgorithm Pass Rate
(%) AttemptsCriteria Failed
Dose Gamma Dose and Gamma
AAA 69 48 0 13 2
adapt conv/cc conv 74 23 0 4 2
Monte Carlo 67 21 0 4 3
pencil beam 74 19 0 3 2
raytracing 40 5 1 0 2
superposition 64 22 0 5 3
total 138 1 27 14CPR
Conclusions
•Challenging credentialing process
•Treatment couch should be considered in these treatments
•Cannot make generalizations with regard to machine, TPS, or algorithm
CPR
ConclusionsConclusions
• Important QA tool• Aids improvements to IMRT delivery• All major linear accelerator and planning
systems have ability to pass