Cr Prima Ir-391cl 02e
Transcript of Cr Prima Ir-391cl 02e
-
7/22/2019 Cr Prima Ir-391cl 02e
1/896
FCR PRIMA Console
CR-IR 391CLService Manual
-
7/22/2019 Cr Prima Ir-391cl 02e
2/896
Export Restrictions Applying to CR-IR 391CL
Japanese export restrictions and foreign exchange regulations apply tothis equipment. Therefore, government approval obtained through dueprocedure is required in order to export or otherwise remove suchequipment from Japan.
Note that the same restrictions apply not only to the main unit but also toafter-sale parts, service manuals and service training (for exporting ortransferring technology) when such items are dispatched or shipped toforeign countries and when service training is given for trainees fromforeign countries.
-
7/22/2019 Cr Prima Ir-391cl 02e
3/896
CR-IR 391CL Service Manual020-221-01E04.28.2009 FM5535 1
1
1. Handling of This Manual
1.1 Notes on the Use of the Service Manual
1. This document is intended to be read by the service personnel of
FUJIFILM Corporation or authorized agent who is commissioned by
FUJIFILM Corporation to perform servicing tasks.
Under these circumstances, no one is allowed to perform the following actionswithout the prior written permission from FUJIFILM Corporation
Duplicating or reproducing this document in whole or in part
Disclosing this document, in whole or in part, to a person other than theauthorized service personnel
Lending or transferring this document, in whole or in part, to a third party
2. Due care must be exercised when handling this document because it
contains product internal structure descriptions and other condential
information.
* When the contents of this document need be changed, the information aboutsuch changes will be conveyed using Engineering Change Notice (ECN).
The Engineering Change Notice (ECN) must also be handled using the samecare as for this document.
3. Due to a product development schedule, the product specications
and available options/variations prevailing at the time of the release of
this document may differ from the descriptions given in this
document. In such an instance, a Service Engineering Information will
be issued as needed to furnish latest information about inapplicable
specications and components.
4. If you have any doubt about the contents of this document or nd
errors, contact your local FUJIFILM representative.
1.2 Warning and Cautionary Information
Precautions during the service work are indicated by the following terms and
symbols.
WARNING
Used when death or serious injuries may occur if this instruction is not
observed.
CAUTION
Used when minor or medium levels of physical injury may be incurred if the
instruction is not observed.
Also used when the machine may suffer serious trouble (such as
unrecoverable or difcult-to-recover trouble).
IMPORTANT
Used when the machine may suffer damage, or any failure or malfunction may
occur, if the instruction is not observed.
NOTE
Used to indicate the matters that need attention during steps of the procedure.
REFERENCE
Used to indicate terminology or supplemental explanations.
-
7/22/2019 Cr Prima Ir-391cl 02e
4/896
CR-IR 391CL Service Manual020-221-01E04.28.2009 FM5535 2
2
1.3 Servicing Instruments and Tools thatRequire Inspection/Calibration
The machine should be installed and serviced by use of servicing instruments and
tools that have been regularly inspected and calibrated as appropriate.
If the machine were serviced using servicing instruments and tools that have not
been inspected and calibrated, proper performance of the machine could not be
guaranteed. Servicing instruments and tools that require inspection/calibration are
as listed below. The inspection/calibration should be planned and performed inaccordance with the specications and instruction manuals of the applicable
servicing instruments and tools.
Instruments and tools that require inspection/calibration
Name Inspection Calibration
Digital multimeter
-
7/22/2019 Cr Prima Ir-391cl 02e
5/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
3
3
020-221-01E
04.28.2009 FM5535
Contents (CR-IR 391CL Service Manual)
Safety Precautions
1. Precautionary Labels ................................................................Safety-1
2. Cautions Relating to Electromagnetic Compatibility ............ Safety-32.1 Electromagnetic Compatibility Instruction .................................. Safety-3
-
7/22/2019 Cr Prima Ir-391cl 02e
6/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
4
4
020-221-01E
04.28.2009 FM5535
Product Specications
1. Equipment Specications ........................................................ Spec-1
1.1 Product Code ............................................................................. Spec-1
1.2 Optional Components ................................................................ Spec-1
1.3 Dimensions/Weight.................................................................... Spec-2
1.4 Environmental Conditions .......................................................... Spec-2
1.5 Electrical Specications............................................................. Spec-2
1.6 Image Storage Performance ...................................................... Spec-2
1.7 Discarding the Equipment .......................................................... Spec-2
-
7/22/2019 Cr Prima Ir-391cl 02e
7/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
5
5
020-221-02E
01.08.2010 FM5563
MD: Machine Description
1. SystemConguration................................................................. MD-1
1.1 System Conguration .................................................................. MD-1
1.2 Operating Environment................................................................ MD-1
1.3 Connectors .................................................................................. MD-3
2. SystemConnectionExample..................................................... MD-5
3. SoftwareFunctionOverview...................................................... MD-6
4. DescriptionsofSoftwareControl .............................................. MD-12
4.1 Data Flow between RU and Console ........................................... MD-12
4.1.1 Data Flow during Routine Processing ..................................... MD-12
4.1.2 Flow of Network Setting Data.................................................. MD-12
4.2 Error Handling.............................................................................. MD-13
4.3 What Is Sleep Mode? .................................................................. MD-134.4 RU Bootup Processing ................................................................ MD-14
4.5 RU Termination Processing......................................................... MD-15
-
7/22/2019 Cr Prima Ir-391cl 02e
8/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
6
6
020-221-02E
01.08.2010 FM5563
MT: Machine Troubleshooting
1. Error Code Table ......................................................................... MT-1
1.1 Referring to Error Messages ........................................................ MT-1
1.2 Precautions .................................................................................. MT-4
1.3 Error Message ............................................................................. MT-4
2. Using the Event Viewer .............................................................. MT-5
2.1 Overview of the Event Viewer ...................................................... MT-5
2.2 Starting the Event Viewer ............................................................ MT-5
2.3 Conguration and How to View the Event Viewer....................... MT-6
2.4 Details of Events .......................................................................... MT-6
2.5 Saving the Event Log ................................................................... MT-7
2.6 Deleting the Event Log ................................................................ MT-8
2.7 Quitting the Event Viewer ............................................................ MT-9
2.8 Collecting Logs for Analysis Purposes ........................................ MT-9
2.9 Customizing the Event Viewer ..................................................... MT-10
3. Information Useful for Error Recovery ...................................... MT-11
3.1 Collecting Logs for Maintenance Purpose ................................... MT-11
3.2 Error for Not Registering the Client Device .................................. MT-11
4. Troubleshooting (for DELL PC).................................................. MT-12
4.1 Analysis Flow for When Your Computer Does Not Start ............. MT-13
4.2 Analysis Flow for When the Monitor Does Not OperateNormally ....................................................................................... MT-13
4.3 Analysis Flow for When the Mouse Does Not OperateNormally ....................................................................................... MT-14
4.4 Analysis Flow for When the Keyboard Does Not OperateNormally ....................................................................................... MT-14
4.5 Checking the System Lights ........................................................ MT-14
4.6 Conrming the Diagnostic Lights ................................................. MT-15
4.7 Conrmation Required When Beeps Sound atComputer Startup ........................................................................ MT-18
4.8 Error Analysis Tool (Dell Diagnostics) ......................................... MT-19
4.8.1 Running the Dell Diagnostics Express Test ............................ MT-19
4.8.2 Memory Checks ....................................................................... MT-22
4.8.3 CPU Fan Checks ..................................................................... MT-22
4.8.4 CPU Checks ............................................................................ MT-23
4.8.5 Motherboard Checks ............................................................... MT-23
4.9 Checking the Hard Disk ............................................................... MT-24
4.10 Tasks to be Performed at Motherboard Replacement................. MT-25
4.10.1 Installation Work ...................................................................... MT-25
4.10.2 Checkup Work ......................................................................... MT-27
Appendix 1 CR-IR 391CL Error Message Table .................. AppxMT 1-1
-
7/22/2019 Cr Prima Ir-391cl 02e
9/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
7
7
020-221-02E
01.08.2010 FM5563
MC: Checks, Replacement and Adjustment of Parts
1. Reinstalling the OS and the Application .................. .................. ......... MC-1
1.1 Installation Flowchart (DELL PC) ................................................. MC-1
1.2 Preparation .................................................................................. MC-2
1.3 Installing and Setting Up the OS .................................................. MC-3
1.3.1 Installing and Setting Windows Vista ....................................... MC-3
1.3.2 Installing the Driver Software ................................................... MC-12
1.3.3 Setting the OS ......................................................................... MC-25
1.4 Setting the Network ..................................................................... MC-35
1.4.1 Installing .NET Framework ...................................................... MC-35
1.4.2 Setting the IP address ............................................................. MC-37
1.4.3 Adding the Windows Components .......................................... MC-40
1.4.4 Setting the Service (Only for Server PC) ................................. MC-42
1.5 RU Settings .................................................................................. MC-43
1.5.1 Installing the RU PC-TOOL ..................................................... MC-43
1.5.2 Setting Up and Checking Firewall Settings ............................. MC-461.5.3 Installing and Registering RU-AP ............................................ MC-52
1.6 Installing the SQL Server ............................................................. MC-58
1.7 Installing the CR-IR 391CL Software ........................................... MC-61
1.7.1 Installing the CR-IR 391CL ...................................................... MC-61
1.7.2 Installing a Settings File ........................................................... MC-63
1.7.3 Installing Menu and Parameter Files ....................................... MC-64
1.7.4 Installing the Standard Kit CD ................................................. MC-65
1.8 SettingsbytheSimpliedSetupWizard...................................... MC-66
1.9 IIS Network Virus Protection ........................................................ MC-67
1.10 Making Checks Using the Service Utility ..................................... MC-68
1.10.1RunningtheSimpliedSetupWizard...................................... MC-68
1.10.2 Checking for Connection with Other Equipment ...................... MC-68
1.10.3 Checking the Date Format ....................................................... MC-68
1.10.4 Installing Adobe Reader and Copying the Reference Guide ... MC-68
1.10.5 Setting the User Account Control (Activate) ............................ MC-69
1.11 RestoringtheCR-IR391CLCongData..................................... MC-70
1.12 OperationCheckandInitializationafterInstallation..................... MC-73
1.13 Restoring the CR-IR 391CL CongurationData(patient information) ..................................................................... MC-74
1.14 Reconstructing the Image Database ........................................... MC-77
2. Service Parts Replacement (for DELL PCs) ............................. MC-81
2.1 Preparation for Parts Replacement ............................................. MC-81
2.2 Hard Disk ..................................................................................... MC-82
2.3 DVD Drive .................................................................................... MC-82
2.4 FD Drive ....................................................................................... MC-84
2.5 CPU ............................................................................................. MC-84
2.6 Memory ........................................................................................ MC-85
2.7 Motherboard ................................................................................ MC-862.8 Power Supply Unit ....................................................................... MC-87
3. Updating the Software ................................................................ MC-88
3.1 Updating the CR-IR 391CL .......................................................... MC-88
3.2 Updating the RU Software ........................................................... MC-88
-
7/22/2019 Cr Prima Ir-391cl 02e
10/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
8
8
020-221-02E
01.08.2010 FM5563
Appendix 1 Setup for Security Enhancement ........................ Appx1-1
Appendix 2 CR-IR 391CL Software Version Upgrade ............ Appx2-1
Appendix 3 Option Key / AP Key Installation ......................... Appx3-1
1. Option Key Installation ........................................................ Appx3-1
2. AP Key Installation .............................................................. Appx3-2
Appendix 4 Setting the RU-Cong .......................................... Appx4-1
Appendix 5 Procedures for Changing the CR-IR 391CL,RU, and Printer IP Addresses .............................. Appx5-1
1. Incorporating into the Network ............................................ Appx5-1
2. Changing the Software Settings ......................................... Appx5-2
2.1 Changing the CR-IR 391RU IP Address .......................... Appx5-2
2.2 Changing the Printer IP Address ..................................... Appx5-2
2.3 Making a setting change usingthe CR-IR 391CLs Service Utility .................................... Appx5-3
2.4 Changing the CR-IR 391RU Setting ................................ Appx5-6
2.5 Changing the Service PC IP Address .............................. Appx5-6
2.6 Changing the CR-IR 391CL IP Address .......................... Appx5-7
2.7 Restart the CR-IR 391CL , RU, and Printer ..................... Appx5-7
2.8 Verifying the connection after IP address changes .......... Appx5-8
Appendix 6 Setting up WakeOnLAN ....................................... Appx6-1
1. Installing the WakeOnLAN Utility ........................................ Appx6-1
2. How to Acquire the RUs MAC Address .................................... Appx6-3
3. How to Acquire the Printers MAC Address ............................... Appx6-4
4. Registering the Other Equipment ....................................... Appx6-5
5. Testing Automatic Startup .................................................. Appx6-7
Appendix 7 Uninstalling APs ................................................... Appx7-1
1. CR-IR 391CL ...................................................................... Appx7-1
2. RU-AP ................................................................................. Appx7-2
3. RU PC-TOOL...................................................................... Appx7-3
Appendix 8 Setting the BIOS ................................................... Appx8-1
1. Setting the Dell OptiPlex 760/780 ......................................... Appx8-1
Appendix 9 Procedure for Changing the Computer Name ... Appx9-1
1. Changing the Computer Name ........................................... Appx9-1
2. Registering with the Service Utility ..................................... Appx9-3
2.1 Starting the Service Utility ................................................ Appx9-3
2.2 Cautions when an Error is Displayed ............................... Appx9-4
3. Checkup Work .................................................................... Appx9-4
-
7/22/2019 Cr Prima Ir-391cl 02e
11/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
9
9
020-221-02E
01.08.2010 FM5563
MU: Maintenance Utility
1. Overview of Service Utility Functions ................................. .......... MU-1
2. Starting, Logging-in to, and Exiting the Service Utility ........... MU-2
3. Top Window ................................................................................. MU-4
4. Regional and Language Options Settings ................................ MU-5
5. SimpliedSetupWizard ............................................................. MU-8
5.1 StartingtheSimpliedSetupWizard ........................................... MU-8
5.2 SettingtheDateandTime ........................................................... MU-9
5.3 System Settings ........................................................................... MU-10
5.4 Server Equipment Settings .......................................................... MU-11
5.5 SettingtheImageReader............................................................ MU-17
5.6 Setting the Film Printer (option) ................................................... MU-18
5.7 Setting for DICOM Storage Device (option) ................................. MU-22
5.8 ExitingtheSimpliedSetupWizard............................................. MU-24
6. Service Settings .......................................................................... MU-25
6.1 System ......................................................................................... MU-25
6.2 ConnectedEquipment ................................................................. MU-50
6.2.1 Server Equipment Settings ...................................................... MU-50
6.2.2 Monitor Settings ....................................................................... MU-52
6.2.3 DICOM Settings ....................................................................... MU-53
6.2.4 Disk Drive Settings .................................................................. MU-56
6.2.5 ImageReaderSettings............................................................ MU-58
6.2.6 Film Printer Settings (option) ................................................... MU-60
6.2.7 Storage Setup for DICOM Storage Device (option) ................. MU-65
6.3 Linkage Function ......................................................................... MU-68
6.4 Film Annotation Characters ......................................................... MU-69
6.4.1 Film Annotation Character Setup(DirectEditingofFilmStrFmt).................................................. MU-74
6.4.2 FilmStrFmt File Description Format ......................................... MU-75
6.4.3 Changing the Film Annotation Characters ............................... MU-77
6.4.4 Format Descriptors .................................................................. MU-806.4.5 Attribute Descriptors ................................................................ MU-84
6.4.6 FilmAnnotationCharacterVericationProcedure .................. MU-88
6.5 Person Name ............................................................................... MU-88
6.6 ModalitySettings ......................................................................... MU-91
7. Maintenance Tool ........................................................................ MU-92
7.1 Database Utility ............................................................................ MU-92
7.1.1 InitializingthePatientDatabase .............................................. MU-93
7.1.2 InitializingtheOutputQueue ................................................... MU-94
7.1.3 InitializingtheImageDatabase ............................................... MU-967.1.4 InitializingthePresetDB ......................................................... MU-98
7.2 AcquiringEDRBackupData........................................................ MU-99
7.3 Connection Check .......................................................................MU-101
7.4 LogDisplay/Edit ...........................................................................MU-102
7.5 Event Log Display ........................................................................MU-103
-
7/22/2019 Cr Prima Ir-391cl 02e
12/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
10
10
020-221-02E
01.08.2010 FM5563
Appendix 1 Backup Utility ........................................................ Appx1-1
Appendix 2 User Utility ............................................................. Appx2-1
Appendix 3 Creating a User ..................................................... Appx3-1
Appendix 4 Settings for Automatic Output ....................................... Appx4-1
Appendix 5 Setup for Auto-verify ............................................ Appx5-1
Appendix 6 DICOM Tag Mapping Function .................................. ... Appx6-1
Appendix 7 Free Drawing Function Settings ......................... Appx7-1
-
7/22/2019 Cr Prima Ir-391cl 02e
13/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
11
11
020-221-01E
04.28.2009 FM5535
SP: Service Parts
1. Service Parts List ........................................................................ SP-1
1.1 How to Use Service Parts List ..................................................... SP-1
1.2 Service Parts List ......................................................................... SP-3
-
7/22/2019 Cr Prima Ir-391cl 02e
14/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
12
12
020-221-01E
04.28.2009 FM5535
PM: Preventive Maintenance
1. Preventive Maintenance Items ................................................... PM-1
2. Checking for a Fatal Error .......................................................... PM-2
2.1 ConrmingtheEventLog ............................................................ PM-2
3. Safety ........................................................................................... PM-2
3.1 ServiceVoltage............................................................................ PM-2
3.2 PowerPlug/Multi-OutletPowerStrip ........................................... PM-2
4. Hardware Cleaning [reference] .................................................. PM-3
4.1 PCInternalCleaning.................................................................... PM-3
4.2 LCDCleaning .............................................................................. PM-4
5. Executing the Error Analysis Tool (Dell Diagnostics) ............. PM-5
5.1 ConductingtheDellDiagnosticsExpressTest............................ PM-5
6. Functional Checks ...................................................................... PM-5
6.1 ConrmationofImageInputtoStorageontoDVD ...................... PM-5
6.2 ImageOutputtoFilm ................................................................... PM-5
7. BackupofSettingsandVericationofDateandTime ............ PM-6
7.1 BackupofSetupFilesandLogFiles ........................................... PM-6
7.2 VericationofDateandTime ...................................................... PM-12
-
7/22/2019 Cr Prima Ir-391cl 02e
15/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
13
13
020-221-02E
01.08.2010 FM5563
IN: Installation
1. CR-IR 391CL Installation Conditions ......................................... IN-1
1.1 Dimensions and Weight ............................................................... IN-1
1.2 Temperature and Humidity .......................................................... IN-1
1.3 Places of Installation .................................................................... IN-1
1.4 Power Requirements ................................................................... IN-2
1.5 Grounding .................................................................................... IN-2
2. Installation Procedure Overview ............................................... IN-3
3. CR-IR 391CL Hardware Setup .................................................... IN-4
3.1 Unloading and Unpacking ............................................................ IN-4
3.2 Checking the Component Items .................................................. IN-5
3.3 Installing the Accessories and Optional Items ................. .............. IN-6
3.3.1 Keyboard, Mouse, and Power Cable ....................................... IN-6
3.3.2 Monitor ..................................................................................... IN-6
3.3.3 Film Printer Hardware Setup (option) ...................................... IN-6
3.4 Connecting a Cable Between the RU and Other Equipment ....... IN-7
3.4.1 Making a Cable Connection to the RU .................................... IN-7
3.4.2 Film Printer Cable Connection (option) ................................... IN-7
4. Software Setup ............................................................................ IN-8
4.1 Setting the Regional and Language Options ............................... IN-8
4.2 Changing the Displayed Language and Replacingthe Menu Database ..................................................................... IN-12
4.3 Option Key / AP Key Installation .................................................. IN-154.3.1 Option Key Installation ............................................................. IN-15
4.3.2 AP Key Installation .................................................................. IN-16
4.4 SetupbytheSimpliedSetupWizard ......................................... IN-17
4.4.1 Startup of the Service Utility .................................................... IN-17
4.4.2 StartingtheSimpliedSetupWizard....................................... IN-18
4.4.3 Setting the Date and Time ....................................................... IN-19
4.4.4 System Settings ....................................................................... IN-20
4.4.5 Server Equipment Settings ...................................................... IN-21
4.4.6 Setting the Image Reader ........................................................ IN-27
4.4.7 Setting the Film Printer ............................................................ IN-27
4.4.8 Setting the Storage Device ...................................................... IN-27
4.4.9 ExitingtheSimpliedSetupWizard......................................... IN-27
4.5 RU Settings .................................................................................. IN-28
4.5.1 Installing the RU PC-TOOL ..................................................... IN-28
4.5.2 Setting Up and Checking Firewall Settings ............................. IN-33
4.5.3 Installing and Registering RU-AP ............................................ IN-38
4.5.4 Setting the User Account Control (Activate) ............................ IN-42
4.6 Film Printer Software Setup (option) ............................................ IN-43
4.6.1 Installing the Option Key for Application Key ........................... IN-43
4.6.2 Reinstalling the AP Key ........................................................... IN-44
4.6.3 DICOM Print Setup .................................................................. IN-44
4.7 Checking Connections with Other Equipment .............................. IN-45
4.8 Exiting the Service Utility ............................................................. IN-46
5. Checkup Work ............................................................................. IN-47
5.1 CheckingtheStartupandCongurationoftheRU...................... IN-47
5.1.1 Checking the Startup of the RU ............................................... IN-47
5.1.2 Checking the RUs Version ...................................................... IN-50
5.1.3 Checking the RUs Message File ............................................ IN-51
5.1.4 CheckingtheRUConguration............................................... IN-52
5.2 CheckingtheSystemConguration ............................................ IN-53
5.3 Checking Every Step from the Inputting of Images to
the Saving of Images onto DVD .................................................. IN-53
5.4 Checking Output Images (option) ................................................ IN-65
-
7/22/2019 Cr Prima Ir-391cl 02e
16/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
14
14
020-221-02E
01.08.2010 FM5563
6. SetupFileBackupandRUErrorLogVerication/Deletion .... IN-68
6.1 Backing Up the CR-IR 391CL Setup File .......................................... IN-69
6.2 Backing Up the RU Setup File ..................................................... IN-76
6.3 Verifying and Deleting the RU Error Log ..................................... IN-78
6.3.1 Verifying the RU Error Log ...................................................... IN-786.3.2 Deleting the RU Error Log ....................................................... IN-79
6.4 Turning OFF the CR-IR 391CL and RU ....................................... IN-80
7. Links to Individual Option Items ................................................ IN-81
-
7/22/2019 Cr Prima Ir-391cl 02e
17/896
CR-IR 391CL Service Manual
Contents (CR-IR 391CL Service Manual)
15
15
020-221-02E
01.08.2010 FM5563
Appx IN: Installation Appendix
Appendix 1 Settings for Saving Images into DICOM StorageEquipment ............................................................. Appx1-1
1. Hardware Setup .................................................................. Appx1-1
2. Software Setup ................................................................... Appx1-2
2.1 Installing the Option Key .................................................. Appx1-2
2.2 Reinstalling the AP Key ................................................... Appx1-3
2.3 Settings in the Service Utility ........................................... Appx1-3
2.3.1 Settings Regarding This Equipment(DICOM storage SCU settings) .................................. Appx1-3
2.4 Changing the Display Gradation Correction LUT File ...... Appx1-6
3. Connection Check .............................................................. Appx1-9
4. Backing up the Setup File ................................................... Appx1-12
Appendix 2 Settings for Connection of the NAS ................... Appx2-1
1. Hardware Setup .................................................................. Appx2-2
1.1 Specications ................................................................... Appx2-2
1.1.1 LAN ............................................................................. Appx2-2
1.1.2 RAID ........................................................................... Appx2-2
1.1.3 Operation-CertiedModelsandSoftware................... Appx2-2
1.2 Installing hardware ........................................................... Appx2-3
1.2.1 TeraStation ................................................................. Appx2-3
1.2.2 LinkStation .................................................................. Appx2-3
2. Software Setup ................................................................... Appx2-4
2.1 Installation procedures for TeraStation(TS-WX1.0TL/R1 and TS-WX2.0TL/R1) .......................... Appx2-4
2.1.1 Network Setting .......................................................... Appx2-4
2.1.2 TeraStation RAID Setting ........................................... Appx2-9
2.1.3 Shared Folder Setting ................................................. Appx2-13
2.1.4 ConrmingConnectionbetweenTeraStationandCR-IR 391CL .............................................................. Appx2-14
2.2 Installation Procedures for LinkStation(LS-WH1.0TGL/R1 and LS-WH2.0TGL/R1) .................... Appx2-15
2.2.1 Network Setting .......................................................... Appx2-15
2.2.2 LinkStation RAID Setting ............................................ Appx2-18
2.2.3 Shared Folder Setting ................................................. Appx2-19
2.2.4 ConrmingConnectionbetweenLinkStationandCR-IR 391CL .............................................................. Appx2-20
2.3 Setting in the Service Utility ............................................. Appx2-21
3. Connection Check .............................................................. Appx2-22
3.1 ConrmingConnectionbetweenLinkStationandCR-IR 391CL (Access to the Network Drive)................... Appx2-22
3.2 ConrmingConnectionbetweenTeraStationandCR-IR 391CL ................................................................... Appx2-22
4. Backing up the Setup File ................................................... Appx2-24
5. Turning ON/OFF the NAS................................................... Appx2-25
5.1 Always ON ....................................................................... Appx2-25
5.2 Manual ON/OFF ............................................................... Appx2-25
6. Mirroring Setting for Image Folder ...................................... Appx2-26
6.1 Creating a Save Destination Folder ................................. Appx2-26
6.2 Synchronization Function Setting .................................... Appx2-26
6.3 MirroringbetweenNAS.................................................... Appx2-30
7. Restoring Procedures Using the Automatic Backup File .... Appx2-31
Appendix 3 Setting a DICOM MWM Patient information ....... Appx3-1
1. Hardware Setup .................................................................. Appx3-1
2. Software Setup ................................................................... Appx3-2
2.1 Installing the Option Key .................................................. Appx3-2
2.2 Reinstalling the AP Key ................................................... Appx3-3
2.3 Settings in the Service Utility ........................................... Appx3-3
3. Connection Check .............................................................. Appx3-5
4. Backing up the Setup File ................................................... Appx3-6
-
7/22/2019 Cr Prima Ir-391cl 02e
18/896
CR-IR 391CL Service Manual 16
16
020-221-02E
01.08.2010 FM5563
PC: Performance Check
CR-IR 391CL Performance Checklist ............................................... PC-1
CR-IR 391CL Confg List ................................................................... PC-3
1. Localize Setting ........................................................................... PC-3
2. Setting for Institution .................................................................... PC-33. Setting for Entering Patient Name ............................................... PC-3
4. Setting is Also Possible with the User Utility ................................ PC-3
5. Setting for Image Collection ......................................................... PC-3
6. Setting for Image Browsing .......................................................... PC-3
7. Setting for General-purpose File Output ...................................... PC-4
8. Linkage with the Receipt Computer ............................................. PC-4
9. Setting for Image Link .................................................................. PC-4
10. Image Incidental Information ....................................................... PC-4
11. Print Format ................................................................................. PC-5
12. Film Character Setting ................................................................. PC-6
13. Setting for Output Function .......................................................... PC-6
14. Operation (Low Frequency) Setting ............................................. PC-6
15. Setting for Analysis ...................................................................... PC-6
16. Settings for Endoscopic Image Import ......................................... PC-7
17. Electrocardiographic Data Import Settings .................................. PC-7
18. PDI Output Function Settings ...................................................... PC-7
19. Free Layout Setting ..................................................................... PC-7
20. Setting for DICOM-MWM ............................................................. PC-7
Contents (CR-IR 391CL Service Manual)
-
7/22/2019 Cr Prima Ir-391cl 02e
19/896
CR-IR 391CL Service Manual Safety-I
Safety-I
020-221-02E
01.08.2010 FM5563
Safety Precautions
CR-IR 391CL Service Manual
-
7/22/2019 Cr Prima Ir-391cl 02e
20/896
CR-IR 391CL Service Manual Safety-II
Safety-II
020-221-02E
01.08.2010 FM5563
Control Sheet
Issue date Revision number Reason Pages affected
04/28/2009 01 New release (FM5535) All pages
01/08/2010 02 Revision for software V1.1(B) (FM5563) 3 to 5
Safety Precautions
-
7/22/2019 Cr Prima Ir-391cl 02e
21/896
CR-IR 391CL Service Manual Safety-1
Safety-1
020-221-01E
04.28.2009 FM5535
1. Precautionary Labels
To prevent danger to humans and serious accidents, observe the following.
WARNINGS
Unless indicated otherwise, be sure to turn off the power switch before
maintenance work, and disconnect the power plug from the outlet.
Even if the power switch is OFF, 100 to 240VAC (differs according to
machine specications) will still be supplied to the primary input terminal of
the unit if the power cable is connected, and this may result in possible
electric hazards.
As neckties, scarves, and necklaces may get caught in the unit during work,
remove these items or take precautions against such accidents.
Never modify the unit unless otherwise specied.
Modication of the safety unit (interlock, breaker, temperature fuse, etc.)
will cause serious problems as it is not only dangerous but violates the
related laws and regulations.
CAUTIONS
The grounding wire must be connected.
To avoid damages to the parts/units due to static discharge, use a wrist
strap putting one end on your wrist and the other end on the grounded
section of the equipment, as illustrated below.
When doing so, make sure to unplug the line cord plug to avoid electrical
hazards if using a PC in which the power is energized to the internal power
supply or circuits when it is inserted into an outlet.
00010003.ai
Wrist strap
1. Precautionary Labels
-
7/22/2019 Cr Prima Ir-391cl 02e
22/896
CR-IR 391CL Service Manual Safety-2
Safety-2
020-221-01E
04.28.2009 FM5535
Safety precautions (Control Unit)
Follow the precautions below when installing, using, and servicing the control unit
(PC).
See the Users Guide provided for details on the control unit.
Input voltage : 90V to 135V (60Hz) or 180V to 265V (50Hz)
Electric hazard prevention :
The power cord of the control unit and monitor uses a 3-pin
plug. Connect it to a properly grounded power outlet. When
using an extension cord, use a 3-line type with a grounding pin.
Place of installation :
Avoid places where dust and foreign particles may enter inside
through the computer opening.
Avoid installing near heaters and heat sources.
When installing directly on the oor, ensure that a 5cm or more
distance is secured between the equipments bottom face and
the oor.
Precautions when working on internal parts of the control unit :
(1) Turn off the power of the PC, monitor, and other peripheral devices.
(2) Disconnect the power cord from the outlet to prevent electric hazard.
(3) Open the I/O panel on the rear of the PC, touch the metallic part to fully
discharge body static electricity before accessing.
(4) When removing the board and memory module, etc., hold them without
directly touching the connectors and pins and place them in a static
electricity prevention package.
Safety Precautions (LCD monitor)
WARNING
High voltage is applied inside the LCD monitor.To prevent electric hazards and danger, the LCD monitors internal parts
cannot be accessed.
The LCD monitor does not consist of parts serviceable by the service
engineer.
The unit is also pasted with precautionary labels. Follow the instructions on the
labels when carrying out work.
REFERENCE
PC box:
Labels shown below are attached on the top face of the PC when it is placed
in a vertical position.
Identication Label
Sample year of manufacture
-
7/22/2019 Cr Prima Ir-391cl 02e
23/896
CR-IR 391CL Service Manual Safety-3
Safety-3
020-221-02E
01.08.2010 FM5563
2. Cautions Relating toElectromagnetic Compatibility
2.1 Electromagnetic Compatibility Instruction
This equipment complies with the following electromagnetic compatibility (EMC)
standard by comforting to the requirements of IEC 60601-1-2:2001+A1:2004/EN
60601-1-2:2001+A1:2006.
EMC Standard: CISPR 22:2005/EN 55022:2006
CISPR24:1997+A1:2001+A2:2002/
EN55024:1998+A1:2001+A2:2003
IEC 61000-3-2:2005/EN 61000-3-2:2006
IEC 61000-3-3:1994+A1:2001+A2:2005
However, there is no guarantee that harmful electromagnetic interference will not
occur in any installation environment.
This equipment generates, uses and can radiate radio frequency energy. In a
particular environment, e.g., the equipment is not installed and used in accordance
with the instructions or any device that is not compliant with the EMC Standard is
used in a surrounding area, harmful interference may result, causing the equipmentor surrounding device to malfunction. If this equipment causes harmful interference
to other equipment or receives harmful interference from other equipment, you are
encouraged to try to correct the interference by one or more of the following
measures:
Reorient or relocate the equipment.
Increase the separation between the equipment and the affected equipment.
Connect the equipment into an outlet on a circuit different from that to which the
affected equipment is connected.
If the problem persists, discontinue the use of the equipment and consult the
manufacturer of the affected equipment or our ofcial dealer.
WARNINGKeep the equipment away from sources of strong electromagnetic radiation.
Table 1
Guidance and manufacturer's declaration - electromagnetic emissions
The CR-IR 391CL is intended for use in the electromagnetic environment specied below.The customer or the user of the CR-IR 391CL should assure that it is used in such anenvironment.
Emissions test Compliance
Noise terminal voltage
CISPR22
Class B
Electric eld noise strength
CISPR22
Class B
Harmonic emissions
EN61000-3-2IEC61000-3-2
Class D
Voltage uctuations/
icker emissions
EN61000-3-3IEC61000-3-3
Complies
-
7/22/2019 Cr Prima Ir-391cl 02e
24/896
CR-IR 391CL Service Manual Safety-4
Safety-4
020-221-02E
01.08.2010 FM5563
Table 2
Guidance and manufacturer's declaration - electromagnetic immunity
The CR-IR 391CL is intended for use in the electromagnetic environment specied below.The customer or the user of the CR-IR 391CL should assure that it is used in such an environment.
Immunity test EN/IEC 60601 test level Compliance level
Electrostatic discharge(ESD)
EN61000-4-2
IEC61000-4-2
2kV contact4kV contact2kV air
4kV air8kV air
2kV contact4kV contact2kV air
4kV air8kV air
Electrical fast transient/burst
EN61000-4-4IEC61000-4-4
1kV for power supply lines
0.5kV for input/output lines
1kV for power supply lines
0.5kV for input/output lines
Surge
EN61000-4-5IEC61000-4-5
0.5, 1kV differential mode
0.5, 1.0, 2.0kV common mode
0.5, 1kV differential mode
0.5, 1.0, 2.0kV common mode
Voltage dips, short interruptions and voltage variationson power supply input lines
EN61000-4-11IEC61000-4-11
95% dip in UT)for 0.5 cycle
70% UT(30% dip in UT)for 25 cycles
95% dip in UT)for 250 cycles
95% dip in UT)for 0.5 cycle
70% UT(30% dip in UT)for 25 cycles
95% dip in UT)for 250 cycles
Power frequency (50/60Hz)magnetic eld
EN61000-4-8IEC61000-4-8
1 A/m 1 A/m
Conducted RF
EN61000-4-6IEC61000-4-6
Radiated RFEN61000-4-3IEC61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m80 MHz to 1 GHz
3 Vrms
3 V/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.
-
7/22/2019 Cr Prima Ir-391cl 02e
25/896
CR-IR 391CL Service Manual Safety-5
Safety-5
020-221-02E
01.08.2010 FM5563
BLANK PAGE
-
7/22/2019 Cr Prima Ir-391cl 02e
26/896
CR-IR 391CL Service Manual Safety-6
Safety-6
020-221-01E
04.28.2009 FM5535
BLANK PAGE
-
7/22/2019 Cr Prima Ir-391cl 02e
27/896
CR-IR 391CL Service Manual Spec-I
Spec-I
020-221-02E
01.08.2010 FM5563
Product Specications
CR-IR 391CL Service Manual
S
-
7/22/2019 Cr Prima Ir-391cl 02e
28/896
CR-IR 391CL Service Manual Spec-II
Spec-II
020-221-02E
01.08.2010 FM5563
Control Sheet
Issue date Revision number Reason Pages affected
04/28/2009 01 New release (FM5535) All pages
01/08/2010 02 Revision for software V1.1(B) (FM5563) All pages
Product Specications
S 1
-
7/22/2019 Cr Prima Ir-391cl 02e
29/896
CR-IR 391CL Service Manual Spec-1
Spec-1
020-221-02E
01.08.2010 FM5563
1. EquipmentSpecifcations
1.1 Product Code
Hardware
Product Code Description Qty Remarks
IR 391 CL A1-V1 B EControl Unit
PC control unit(Dell OptiPlex780DT/Core2Duo E7500)
Identication labelCR-IR 391CL Application Software V1.1(B)CR-IR 391CL Standard KitManual Kit for Hospitals
CR-IR 391CL Reference GuideCR-IR 391CL Operation Manual(Software License Agreement is included)CR-IR 391CL Quick Guide
1
11
1
11
1
Software
Product Code Description Qty RemarksIR 391 CL SOFT V1 B E
CR-IR 391CL Application Software V1.1(B)
Identication labelCR-IR 391CL Standard KitManual Kit for Hospitals
CR-IR 391CL Reference GuideCR-IR 391CL Operation Manual(Software License Agreement is included)CR-IR 391CL Quick Guide
111
11
1
1.2 Optional Components
Product Code Description Qty Remarks
IR 391 LCD-17 E 17 inch LCD (DELL E170S) 1
IR 391 LCD-17H E 17 inch LCD (Nanao FlexScan S1721) 1
IR 391 LCD-20W E 20 inch LCD (S2031W-HBK) 1
IR 391 MFP E CR-IR 391CL MFP Key Software 1
IR 391 DICOM PRINT E CR-IR 391CL DICOM Pr int Key Software 1
IR 391 DICOM STORAGE ECR-IR 391CL DICOM Standard CR StorageKey Software
1
IR 391 F-LAYOUT PRINT ECR-IR 391CL Free Layout for DICOMStandard Print
1
IR 391 DICOM ID MWM ECR-IR 391CL DICOM ID MWM KeySoftware
1
S 2
-
7/22/2019 Cr Prima Ir-391cl 02e
30/896
CR-IR 391CL Service Manual Spec-2
Spec-2
020-221-02E
01.08.2010 FM5563
1.3 Dimensions/Weight
Dimensions (width height depth) Weight
PC OptiPlex 780DT 399 114 353 (mm) Approx. 8.3 kg
MonitorS2031W-HBK 427 to 509 469 208.5 (mm)DELL E170S 374.7 380.6 140 (mm)S1721-SHBK 362 280 to 406 200 (mm)
Approx. 9.6 kgApprox. 4.63 kgApprox. 5.6 kg
1.4 Environmental Conditions
Condition Temperature Humidity
Operating 10C - 35C 30% - 80%H (No dew condensation)
Non-operating -20C - 60C 30% - 80%H (No dew condensation)
Increase humidity appropriately if it is too dry to operate the equipment during winter.
1.5 ElectricalSpecifcations
Frequency Input voltage Max. input current
PC 50Hz/60Hz AC115 to 230V 4.0A/2.0A
Monitor
S2031W-HBK 50Hz/60HzAC100V to 120VAC200V to 240V
0.7A
DELL E170S 50Hz/60HzAC100V to 120VAC200V to 240V
1.5A
S1721-SHBK 50Hz/60HzAC100V to 120VAC200V to 240V
0.65A/0.4A
Network cable
2. System Connection Example. of MD: Machine Description. 3.4 Connecting a Cable Between the RU and Other Equipment. of IN:
Installation.
1.6 Image Storage Performance
Presented below are the numbers of images that can be stored on the media
(measurement values). (Note that actual numbers of images that can be stored
may increase/decrease in relation to the values shown in the tables below,
depending on the free space available on the disk or image data capacity.)
Hard disk
Image storage capacity : 240GB
Image Size 14"17" 14"14"
Storable number of images Approx. 15,000 images Approx 17,000 images
DVD+RW/DVD-RW
DVD capacity : 4.7GB, UDF format
Image size ratio : 14"17" : 14"14" : 10"12" : 8"10" = 1 : 6 : 2 : 1
(for standard image/high-density image)
Standard Image High-density Image
UncompressedJPEG lossless-
compressedUncompressed
JPEG lossless-compressed
4.7GB Approx. 650 images Approx. 1,300 images Approx. 200 images Approx. 400 images
1.7 Discarding the Equipment
When discarding the CR-IR 391CL equipment, be sure to dispose of any private
information stored on it completely.
MD I
-
7/22/2019 Cr Prima Ir-391cl 02e
31/896
CR-IR 391CL Service Manual MD-I
MD-I
020-221-02E
01.08.2010 FM5563
MD:MachineDescription
CR-IR 391CL Service Manual
MD IIMD: Machine Description
-
7/22/2019 Cr Prima Ir-391cl 02e
32/896
CR-IR 391CL Service Manual MD-II
MD-II
020-221-02E
01.08.2010 FM5563
Control Sheet
Issue date Revision number Reason Pages affected
04/28/2009 01 New release (FM5535) All pages
01/08/2010 02 Revision for software V1.1(B) (FM5563) 1, 3, 4, 6 to 8, 16 to 18
Changes in pagination (FM5563) 5, 9 to 15
MD: Machine Description
MD 1
-
7/22/2019 Cr Prima Ir-391cl 02e
33/896
CR-IR 391CL Service Manual MD-1
MD-1
020-221-02E
01.08.2010 FM5563
1. SystemConguration
This section shows the system conguration, which includes the CR-IR 391CL
application, and the operating environment for the PC.
The following PC is used.
Dell OptiPlex 760 (desktop)
Dell OptiPlex 780 (desktop)
1.1 SystemConguration
SystemConguration
Video cableMonitor
Mother board
Accessories
(keyboard
and mouse)
Serial
ATA
PC main unit
00000042.ai
Network i/f
System area/image area
(drive C)
USB
Connector
CR-IR 391CL application
CD-ROM
Option key
CD-ROM
DVD-Super
Multi Drive
1.2 OperatingEnvironment
PCMainUnit
Item Description
CPU Intel Core 2 Duo E7300: 2.66GHz or more; secondary cache: 3072KB or more
System memory 2GB or more
HDD 250GB or more
Drive DVD-Super Multi Drive (S-ATA)
Extension slotPCI 1 (Necessary for insertion of a video board when the DVI output is
used.)
Graphic Display resolution: 1280 1024 or more
Number of display colors: 24bit Color or more* DVI 1.0 or VGA output is recommended.
Network100BASE-TX1000BASE-T
OS Windows Vista Business SP1 (32bit)
Port Keyboard connection port: USB or 6-pin mini-DIN PS/2
Mouse connection port: USB or 6-pin mini-DIN PS/2* The USB 2.0 port is recommended.
Input device Keyboard and mouse
Monitor
Item Description
Display resolution
17 inch LCD (S1721-SHBK) 1280 1024
17 inch LCD (DELL E170S) 1280 1024
20 inch LCD (S2031W-HBK) 1680 1050
Input format DVI 1.0 (DVI-I 29-pin) or analog RGB (D-Sub 15-pin)
-
7/22/2019 Cr Prima Ir-391cl 02e
34/896
MD 3
-
7/22/2019 Cr Prima Ir-391cl 02e
35/896
CR-IR 391CL Service Manual MD-3
MD-3
020-221-02E
01.08.2010 FM5563
Memorymodules
2 GB memory (DIMM1/DIMM2) is installed and shipped from factory.
Riser card (ATAPI circuit board) in PCI (SLOT2) is installed and shipped from
factory.
DIMM 4
DIMM 3
DIMM 2
DIMM 1
SATA 1
SATA 0
SATA 3
SATA 2
MD000274.AI
PCI Express x16(SLOT1)
PCI(SLOT2)
PCI(SLOT3)
System boardDIMM sockets
(DIMM1/DIMM2 is used)
1.3 Connectors
This section shows locations of the connectors and memory slots of the PCs below.
Dell OptiPlex 760
Dell OptiPlex 780
DellOptiPlex760
Systemboardlocation
System board
System lights (power light and
hard disk light)
Diagnostic lights
MD000263.EPS
Connectorconguration(PCrearpanel)
Parallel port connector NIC connector
USB port
(6-port)Serial port connector
MD000273.AI
Video connector D-Sub
(analog output)
MD-4
-
7/22/2019 Cr Prima Ir-391cl 02e
36/896
CR-IR 391CL Service Manual MD-4
MD-4
020-221-02E
01.08.2010 FM5563
DellOptiPlex780
Systemboardlocation
System board
System lights (power light and
hard disk light)
Diagnostic lights
MD000263.EPS
Connectorconguration(PCrearpanel)
Parallel port connector
NIC connector
USB port (6-port)
Serial port connector
MD000275.eps
Video connector D-Sub
(analog output)
eSATA connector
DisplayProt connector
Video connector DVI
(digital output)
Memorymodules
1 GB or 2 GB memory (DIMM1/DIMM2) is installed and shipped from factory.
DVI adapter card in PCI Express x16 (SLOT1) is installed and shipped from
factory.
Riser card (ATAPI circuit board) in PCI (SLOT2) is installed and shipped from
factory.
DIMM 4
DIMM 3
DIMM 2
DIMM 1
SATA 1
SATA 0
SATA 3
SATA 2
MD000276.eps
PCI Express x16(SLOT1)
PCI(SLOT2)
PCI(SLOT3)
System boardDIMM sockets
(DIMM1/DIMM2 is used)
MD-5
-
7/22/2019 Cr Prima Ir-391cl 02e
37/896
CR-IR 391CL Service Manual MD-5
MD-5
020-221-02E
01.08.2010 FM5563
2. SystemConnectionExample
This section shows an CR-IR 391CL system connection example.
NOTE
To connect the CR-IR 391CL to the RU, use the following network cable.
The UTP type which is compliant with category 5 or more should be used.
IEC950/UL1950-compliant cables should be used.
ForthenetworktypeRUoperation
With 100BASE-TX (straight) cables, connect the RU to the CR-IR 391CL via a hub.
00000028.EPS
HUB RU
Other connected hosts
(Film Printer)
100BASE-TX
cable (straight)
100BASE-TX
cable (straight)
100BASE-TX
cable (straight)
CR-IR 391CL
MD-6
-
7/22/2019 Cr Prima Ir-391cl 02e
38/896
CR-IR 391CL Service Manual MD-6
MD 6
020-221-02E
01.08.2010 FM5563
3. SoftwareFunctionOverview
CRStudy
Function Description
Study registrationfunction
Registers a study to perform image input/verication.
Displaying a scheduled
study Displays a study registered using the study registration function.
Scheduled studymodication function
Modies information of a registered study.
Adding/changing/deleting exposure menus
Replacing images
Inactivating images/determining mis-exposed images andre-exposures
Relocating order of image display
Displaying/editing patient information
Setting a target destination for image output
Performing a studyInputs images, in conjunction with RU, in an already registered studyand veries images thus collected.
Indicator functionIndicates the CR-IR 391CL system status according to the following6 classications: RU, Disk, NAS, Printer, Output Queue condition
and Event, and then informs the user of the result of indication.
PatientList
Function Description
Patient-originated(ID input)
Execute new registration for patient.
Patient-originated(Patient list)
Search for patient.
Patient-originated(Refer to study data)
Refer to and operate the study for selected patient.
Patient-originated(Execute CR study)
Execute a CR study for selected patient.
Patient-originated(Import endoscopic andgeneral-purpose image)
Import a general-purpose image and endoscopic image for selectedpatient. (Endoscope Image Import Function Option/General-PurposeImage Import Function Option are required)
Monitoring Automatic di sp lay of retrieved images from the moda lity.
Starting Utilities Displays the version information
Start up Customize screen
Start up Client Utilities/Start up Study Utilities
Customize Customize Main window behavior to adjust user workow
StudyUtilities
Function Description
Searching for study data
Performs a searching for the study data by way of study of thisequipment and study contained on the disks.Filtering is applied to facilitating selection of a target study and thenstudy data is searched for.
Displaying the study listDisplays the study list of this equipment and the study list containedon the disks.
Study data selection Selects study data on the study list of this equipment and the studylist contained on the disks, and then veries such study data.
Study data operation Operates the selected study data.
NASDrive(NetworkAttachedStorage)
Function Description
Data managementRead/Delete images from NAS.Observe remaining capacity for NAS and study/patient DB.
Automatic ImageTransfer Process in thebackground
When the capacity of PC HDD is full, the oldest image is transferredto NAS and deleted from the PC HDD.
Auto backup of setting/log
Backup the setting/log automatically to NAS at the time of exiting anapplication.
MD-7
-
7/22/2019 Cr Prima Ir-391cl 02e
39/896
CR-IR 391CL Service Manual MD-7
MD 7
020-221-02E
01.08.2010 FM5563
Function Description
RestandardizationModies standardization conditions so that restandardizationprocessing is performed appropriately.
Adding Annotations
Arrows, shapes, text (character strings), and markers can beadded to an image. For the image of an FCR study (notprocessed) as the object, it is also possible to measure theimage and to display the measuring result. Up to 32annotations can be attached to a single image. Shapes (Ellipses, rectangles, polygons)
Markers Measuring result (length, cardiothoracic ratio, angle)
Editing selection ofmeasurement results
Selects, moves and edits already input length and heart-to-ribcage ratio.
Saving measurement results Saves measurement results.
Length Measures dimensions of a tumor or the like.
Heart-to-ribcage ratioMeasures the ratio of the widthwise length of the shadow ofthe heart to that of the ribcage.
Angle measurementMeasures the angle of two line segments, contact or non-contact, for general-purpose use.
Free Drawing
Lines can be drawn freely on the viewer window just as witha pen. The following can be done with this free drawingfunction. Selection of the thickness and the color of the lines to be
drawn Capture of the free drawing image (clipboard copy) Deleting of the drawn lines
Toolbar setting Customizes the toolbar setting.
Study status settingSets Verify or Verication Suspended individually for eachstudy.
SaveSaves image processing parameter information of a studyimage that has been subjected to image adjustmentprocessing.
CR study linkage function
Displaying mis-exposed
imagesDisplays mis-exposed images.
Modify exposure menu Modies exposure menus.
Re-exposure Instructs re-exposure of an image.
Additional exposures Adds exposure menus in a study.
Replacing images Replaces study images.
Modifying patient information Displays/edits patient information.
Setting an output device Species a target destination for image output.
ImageViewer
Function Description
Image display/adjustment
Windows pattern changeChanges display pattern of the study window and the imagedisplay window.
Display position relocation Relocates image display positions on the study window.
Displaying maximized images Displays fully magnied study images on the viewer window.
Feeding display images Alternately changes display images individually by a patient,a study or an image.
Displaying additionalinformation
Displays additional information of the shown image as animage overlay, and also changes between Display and Hidestatuses alternately.
Displaying thumbnailsDisplays thumbnails of the shown study images. Alsochanges between Display and Hide statuses alternately.
Displaying the measuredresult
Displays the measured result, and also changes betweenDisplay and Hide statuses alternately.
Tree view displayDisplays listing of patients/studies in a tree view form, andselects from such list a study you wish to display images.
Displaying comparativeimages
Comparatively displays a currently shown image and thecorresponding past image.
Changing gradation Changes gradation at real- time basis.
Image magnication/imagereduction
Displays magnied/reduced images.
Panning Moves display position of an image to a desired position.
Stack Display Switches the display image by mouse dragging.
Preset Image ProcessingRegisters the selected image processing parameters withpreset, and changes the preset name or deletes the presetimage processing parameters.
Reverse displayPerforms reverse displays of an image in relation to itsdensity.
Reset Resets a display image to a previously stored status.
Rotation Flipping Rotates or ips an image.
FCR image processingparameters
Adjusts FCR image processing parameters.
Applying Modality LUT
Applying following Modality LUT according to the modality ofeach image. N/A P-Value gamma 2.2 FCR
MD-8
-
7/22/2019 Cr Prima Ir-391cl 02e
40/896
CR-IR 391CL Service Manual MD-8
MD 8
020-221-02E
01.08.2010 FM5563
Function Description
Image Output
Film print Outputs a lm image to a desired printer.
Work disk output Outputs work media to a desired disk device.
General-purpose leOutputs a study image in any of the following formats:BITMAP, JPEG and DICOM Stream
Using explanatory description
Explanatory description usefunction Displays the explanatory description use dialog.
Indicator function
Indicates the CR-IR 391CL system status according to thefollowing 5 classications: RU, Disk, Printer, Output Queuecondition and Event, and then informs the user of the resultof indication.
Indicator
Function Description
IndicationInforms the user of the device connectivity and error information bymeans of putting the related icon on/off or blinking status.
RU status display Displays status of a connected RU and performs scanner cleaning.
Disk status display
Displays disk status (volume label, application status, usage, writeenabled/disabled).
Displays listing of disks created as Image storage disks*1.
Performs disk settings.
NAS status display
Displays the following NAS status.
Remaining capacity
Host name
Status (Creating write data/Write enable/Temporarily)
Printer status display Displays status of connected printers.
Output status displayDisplays Output Queue status and performs Output Queueoperations, such as deletion, forced output, and other.
Event displayDisplays listing of events that occurred in the server, such as error,
warning, information and other.
*1 Image storage diskIn the CR-IR 391CL system, a disk used for storing study result images as originals isdenominated an Image storage disk.
ServiceUtility
Function Description
Service settings
Simplied setup wizard Interactively performs various kinds of settings successively.
System Performs basic system setups.
Connected equipmentPerforms settings for equipment units to be connected to theCR-IR 391CL system.
Linkage function Performs settings for functions that enable linkage with othersystems.
Film annotationcharacters
Performs settings for character strings to be output on lm andthe formats necessary for output processing.
Person name Performs settings necessary for convert ing an input person name.
Modality sett ing Performs LUT setting corresponding to each modality.
Maintenance tool
Database UtilityDiscards currently used database and restores the stored data,when database les are damaged accidentally or there areinconsistencies between the database and the local images.
Acquiring EDR backupdata
Acquires/stores EDR backup data les of the selected localimages.
Connection checkChecks connection with the equipment determined according to
Connected equipment by Service settings.
Log display/edit Displays or clears off the number of IP exposures.
Event log displayStarts running the Windows XP event viewer for troubleshootingpurposes.
-
7/22/2019 Cr Prima Ir-391cl 02e
41/896
MD-10
-
7/22/2019 Cr Prima Ir-391cl 02e
42/896
CR-IR 391CL Service Manual MD-10020-221-02E01.08.2010 FM5563
Function Description
User information User settings
Explanatory information N/A
Log information
IP counter information
Operation log
Changed history log
Output history log
CR-IR 391CL system event log
Analysis log
Windows event log Registry le
Image database
DDO le
ClientUtilities
Function Description
Edit patient information. Registers, edits and deletes patient information.
Display of IP usefrequency
Displays the IP use count.
User registration
Adds and deletes images to be displayed on the user-selected
window.For the administrator: Registers, edits and deletes users. Operationauthority granted to the user group will be determined.
User editing Edits the user logged in the system.
Registering explanatorydescription
N/A
History display Displays access history and image transfer history.
Disk operation Performs disk settings, disk registration and disk reconstruction.
Database restorationRe-registers studies saved on the Image storage disk
*1to the study
database.
*1 Image storage diskIn the CR-IR 391CL system, a disk used for storing study result images as originals is
denominated an Image storage disk.
DiskCopyTool
Function Description
Window on which copiesare made
Species a source drive to perform copy processing.
Displays process progression status using the progress bar.
Enables canceling of copy processing.
Sets volume label of the copy source as that of the target copydestination, after completion of le copy processing.
ImageOutput
Function Description
DICOM print outputGenerates lm printout to a printer connected to the CR-IR 391CLserver.
DICOM media storageoutput
Generates output to a disk device connected to the CR-IR 391CLserver.
General-purpose leoutput
Manually outputs study images transferred from the viewer, in eitherof the BITMAP, JPEG or DICOM Stream format.
DICOM storage outputStores images on a DICOM storage device that supports the DICOMcommunication protocol (service class: CR Image Storage).
Free Layout DICOM printoutput
On the free layout print screen, free layout of images is possible for
output of multiple images to one piece of lm. The image size alsocan be adjusted.
ImageProcessing
Function Description
Standardization Image standardization processing (conversion to a 10bit format le)
EDR processing Density correction processing (Exposure Data recognizer)
GP processing Gradation Processing
RP processing Response Processing
DRC process ing Dynamic Range Cont rol Process ing
TAS processing Tomographic Art ifacts Suppression processing
MFP process ing Mul ti -object ive Frequency Processing
FNC processing Flex ible Noise Control process ing
GPR processingGrid Pattern Removal processing (removes moire of the grid withoutsacricing the image quality)
MD-11
-
7/22/2019 Cr Prima Ir-391cl 02e
43/896
CR-IR 391CL Service Manual MD-11020-221-02E01.08.2010 FM5563
MediaStorage
Function Description
Image managementMaintains and manages patient/study/image information that issaved on the disk into the system.
Image updateAdds in-study images and updates information for the images storedon the Image storage disk
*1.
Database restoration(Client utilities)
Captures information stored on various disks into the image serverand includes such images as those targeted for management.
Disk reconstruction(Client utilities)
If data contained on disks is illegal, subjects DICOMDIR toreconstruction processing based on contents of the les saved onthe disks so that those disks can be used properly.
Disk storage(Image output)
Stores images onto several disks.
Image search display(Study list)
Uses the Web browser to display listing of images saved on theimage server or disks or images themselves.
Disk alteration check(Client utilities)
Checks if the disks are altered by third parties.
Disk settings(Client utilities)
The user sets whether data is write-enabled or -disabled onto disks.
*1 Image storage disk
In the CR-IR 391CL system, a disk used for storing study result images as originals isdenominated an Image storage disk.
General-purposeImageCapture
Function Description
Image capture
This function is available for capture of images derived fromconnection devices such as digital camera, USB memory, FD/CD/DVD and other, as well as those located in a folder
*1that can be
viewed from My Computer on Windows. A maximum of 50 imagescan be captured for each study.
Image managementManages general-purpose images as study images on the CR-IR391CL.
Image browseEnables browsing studies that include general-purpose imagesselected on the study list on the CR-IR 391CL viewer.
Image outputThis function enables image output onto a DVD for storage/DVD foroperation and general-purpose image output.
*1 To view a folder located on the network, the network drive must be assigned ahead of time.
PDI(PortableDataforImaging)PatientCD
Function Description
Image output
Writes exposed images and captured images from the outside ontoa CD-R and DVD-R.Use this function in a region where patients are required to storeimages compulsorily or when making referral to hospitals.
Conforms to IHE*1
TF Transaction RAD-47 Distribute Imaging
Information on Media.
*1 IHE : Integrating the Healthcare Enterprise
MD-12
-
7/22/2019 Cr Prima Ir-391cl 02e
44/896
CR-IR 391CL Service Manual MD-12020-221-02E01.08.2010 FM5563
4. DescriptionsofSoftwareControl
NOTE
In this chapter, the "RU Application", "RU IP Address", and "RU Confguration" are
referred to shortly as RU-AP, RU IP ADDR, and CONFIGURATION, respectively.
4.1 DataFlowbetweenRUandConsoleThe exchange of data between the RU and Console is described below.
4.1.1 DataFlowduringRoutineProcessing
Flowofimagedata
The image data read from the IP is stored in the SDRAM of the CPU board, and
transferred to the Console via the network.
Flowoflogdata
The log data is written into the FTP server upon RU shutdown processing or
upon occurrence of a FATAL error. The log data is retained in the FTP server until a certain amount of that data is
reached, and, once that amount is exceeded, the data is overwritten on a rst-in,
rst-out basis.
Console RU
FLASH ROMCOMMON
FRRB102002.ai
FTP server
Memory (SDRAM)
Image data
Log data
RU-specific data
Log data
Image data
CPU board
4.1.2 FlowofNetworkSettingData
Settableitemsanddefaultvalues
The settings are made from the RU PC-TOOL. The settable items and their default
values are as follows.
Setupitems Settingmeans Defaultvalues
RU IP address RU PC-TOOL 172.16.1.10
RU subnet mask RU PC-TOOL 255.255.0.0
RU default gateway RU PC-TOOL 172.16.1.0
FTP server IP address RU PC-TOOL 172.16.1.20
RU secure IP address RU PC-TOOL 0.0.0.0
RU secure net address RU PC-TOOL 0.0.0.0
Flowofnetworksettingdata
The network setting data is not written into the CPU board only by clicking [SET]
button on the RU PC-TOOL window. You need click [UPDATE] button to write the
network setting data into the CPU board.
The contents set from the RU PC-TOOL are stored on the CPU board FLASHROM.
Console RU
RU PC-TOOL
FLASH ROM
[SET]
[UPDATE]
CPU board
FRRB102001.ai
Network settingdata
Network settingdata
-
7/22/2019 Cr Prima Ir-391cl 02e
45/896
MD-14
-
7/22/2019 Cr Prima Ir-391cl 02e
46/896
CR-IR 391CL Service Manual MD-14020-221-02E01.08.2010 FM5563
RecoveryfromSleepMode
Recoverrequirements
Press the stand-by switch.
Processinguponrecovery
Upon recovery from the Sleep mode, the RU performs a home-positioning
operation of the suction arm.
In addition, the initial self-diagnosis of the mechanical components takes place, aswell as home-positioning of the RU suction arm, under one of the following
conditions.
The mechanical components are located at other than the home positions.
One of sensors detects an IP.
A cassette is set in the RU.
4.4 RUBootupProcessing
The following self-diagnoses are carried out after the power is turned ON.
Connection verication between the RU and the Console is also carried out
concurrently with the self-diagnoses.
InitializationoftheMechanicalComponents
The initialization of the mechanical components consists of the following ve
stages.
BeforeIPsearchstage
The cassette cover opening mechanism is moved to the reference position, and
the suction arm is moved to the home position.
IPsearchstage
The motor is rotated by an appropriate amount to convey the IP to a position where
it can be conrmed by the sensor, on the assumption that the IP is located at a
position where it cannot be detected by the sensor within the machine.
AfterIPsearchstage
The suction arm is moved to the home position.
RemainingIPejectionstage
When an IP is found within the machine during the IP search stage, its size is
measured and the IP is returned to the cassette. If a cassette is not set in the RU,
the message prompting to set an empty cassette appears.
AfterremainingIPejection
The self-diagnosis of the cassette cover closing mechanism is carried out, and the
cassette hold is released.
InitializationoftheScannerMechanismThe scanner mechanism is initialized.
MD-15
-
7/22/2019 Cr Prima Ir-391cl 02e
47/896
CR-IR 391CL Service Manual MD-15020-221-02E01.08.2010 FM5563
4.5 RUTerminationProcessing
Termination processing of the mechanism and the log data storage take place.
NOTE
If the power is turned OFF without performing RU shutdown processing, log data
will not be saved.
FlowofRUShutdownProcessing
(1)"Shutdownprocessing"command
Two instructions are available for a case where system shutdown of the Console
is carried out and a case where the stand-by switch of the RU is pressed.
(2)Execute"savingoflogdata",andturnOFFthepower.
Saving of log data:
Save HISTORY LOG, TRACE LOG, and ERROR LOG to the FTP server.
MD-16
-
7/22/2019 Cr Prima Ir-391cl 02e
48/896
CR-IR 391CL Service Manual MD-16020-221-02E01.08.2010 FM5563
BLANKPAGE
MD-17
-
7/22/2019 Cr Prima Ir-391cl 02e
49/896
CR-IR 391CL Service Manual MD-17020-221-02E01.08.2010 FM5563
BLANKPAGE
MD-18
-
7/22/2019 Cr Prima Ir-391cl 02e
50/896
CR-IR 391CL Service Manual MD-18020-221-02E01.08.2010 FM5563
BLANKPAGE
-
7/22/2019 Cr Prima Ir-391cl 02e
51/896
MT-IIMT: Machine Troubleshooting
-
7/22/2019 Cr Prima Ir-391cl 02e
52/896
CR-IR 391CL Service Manual MT-II020-221-02E01.08.2010 FM5563
Control Sheet
Issue date Revision number Reason Pages affected
04/28/2008 01 New release (FM5535) All pages
01/08/2010 02 Revision for software V1.1(B) (FM5563) 11, 15 to 17, 19 to 23, 28 to 30,
Appx MT 1-97 to 348
Changes in pagination (FM5563) 18, 24 to 27
MT-1
-
7/22/2019 Cr Prima Ir-391cl 02e
53/896
CR-IR 391CL Service Manual MT-1020-221-01E04.28.2009 FM5535
1. Error Code Table
For errors occurring while the CR-IR 391CL is being used, take appropriate
countermeasures following the error codes (and error messages) listed herein.
See 2. Using the Event Viewer.
1.1 Referring to Error Messages Event (error code) and event source name
When an error occurs, use the event and event source name information to
separate the assumed causes.
The event and event source name are displayed in the Source and Event
columns of the Event Viewer main window shown below.
Source column Event ID column
00000072.ai
Meaning of the event (error code)
A ve-digit code displayed in the Event column implies the following.
A B C C C
A : Error-detected process
1: Module used in common in the system/indicator/user
authentication/Client Utility
2: Main process/routine/study list
3: Input process/CR study4: Image browse/acquisition process
5: Output process/media control
6: Backup Utility/User Utility/Service Utility/disk copy tool
9: RU
B : Error level (See the Table Error Levels.)
CCC : For error level 0 or 1, this section of the error code is provided so
that it is unique in the process.
MT-2
-
7/22/2019 Cr Prima Ir-391cl 02e
54/896
CR-IR 391CL Service Manual MT-2020-221-01E04.28.2009 FM5535
Table Error Levels
Error Level Description
0 : Fatal ErrorAn error that fully disablessystem operation.
A serious, fatal problem that absolutely disables the systemsstable operation.
DB access error
System le access le error
1: WarningAn error that enables thesystems operation to be
continued after userintervention.
Inevitably involves user notication and the system operation canbe continued by the user intervention.Causes of this error type can be assumed as follows.
Data le access error
Disk full
Faulty LP board
No response from the target destination for connection
2: WarningAn error that enables thesystems continuousoperation.
A slight error that does not require user notication.The system operation can be continued with no user intervention.The following causes can be assumed.
Unspecied data entered from networks. (An error implying thatthe system can be continued being substituted by the default).
3 : InformationSimple information
Information that the system development section wishes to keep.
9: Logical ErrorDefective program
An error presumably caused by defective program.Basically, this error does not occur at users site. This error type
has been aborted by the FPD previously. The following causescan be assumed.
Unspecied value returned from the method
Failure in securing memory
Failure in acquiring Windows resources
Error returned from the Windows API
Message
The message is for describing errors and is displayed in the General tab of the
Event Viewers Properties window shown below.
General tab
00000073.ai
MT-3
-
7/22/2019 Cr Prima Ir-391cl 02e
55/896
CR-IR 391CL Service Manual MT-3020-221-01E04.28.2009 FM5535
Identifying the problem cause when two or more errors encounteredhave the same error code
The same error code may appear on the display even when the error cause varies.
If you encounter such a situation, note both the error code and error message to
identify the error cause.
Event Event Source Description Causes
*
13005
FFCustom MsgBox MessageBox DialogShow Error.
A message box could not bedisplayed. The problem isattributable to resource inadequacyor system instability.
13005FFId ConnectionWatch
Recverror(%1).%n[%2]
The connection to the RU wasclosed or an error occurred in theconnection.
13005
FFIdMenuChange
[%1] failed in theliberatingprocessing of thedisplay group dataacquisition
The connection to the display grouptable was not successfully closed.
13005
FFIdPatInfoInput
It failed that [%1]gets the SexMalecaption
The LangStrDB key was not foundin Directory.ini or the ID 30705 wasnot registered in the character stringdata base.
13005FFIdRuMainControl
Recv error[%1] The connection to the RU wasclosed or an error occurred in theconnection.
13005FFInput It failed that [%1]
connects to STRDB
The connection character string wasnot properly registered for theLangStrDB key in Directory.ini.
* There are two or more occurrences of the same error code (event).
To identify the error cause, perform the following procedure:
1. Start the Event Viewer.
For details of the startup procedure, see 2.2 Starting the Event Viewer.
2. Open the Event Properties window.
For details the procedure for opening the above window,see 2.4 Details of Events.
3. See both the error code and error message, and then locate the
associated error description in Appendix 1 CR-IR 391CL Error
Message Table.
For error messages, see Appendix 1 CR-IR 391CL Error MessageTable.
Error Code
Error
Message
00000472.ai
MT-4
-
7/22/2019 Cr Prima Ir-391cl 02e
56/896
CR-IR 391CL Service Manual MT-4020-221-01E04.28.2009 FM5535
1.2 Precautions
An error such as incorrectly entered image processing parameters will not be
handled as an error.
An error code starting with 9 (e.g.: 9XXXX) is that generated by the RU.
The correlation between the error level (the second digit of an error code