Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas...

28
Coriolis two-phase meter Øyvind Nesse NFOGM 22-03-2012 2012-03-14

Transcript of Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas...

Page 1: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Coriolis two-phase meterØyvind Nesse NFOGM 22-03-2012

2012-03-14

Page 2: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Agenda

• Project

• Allocation Metering Concept

• Coriolis Test Results

− Endress & Hauser

• Mass Flow

• Density

− MicroMotion

• Mass Flow

• Density

• Conclusions

2012-03-142

Page 3: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

The Svalin Field PL169 tie-in host Grane

PL028C

PL169

Grane Unit

M

C

PL028C

PL169

Grane Unit

M

C

In-place (MSm3) P90 Mean P10 Reserves (MSm3)

Svalin C 7.3 11.2 15.4 6.4

Svalin M north 6.4 11.9 17.3 6.9

2012-03-143

Page 4: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Svalin M production profiles

01/1

4

01/1

6

01/1

8

01/20

01/2

2

01/2

4

01/26

01/28

01/30

0

50K

100K

150K

200K

250K

300K

Ga

s p

rod

uc

tio

n r

ate

(S

M3

/DA

Y)

0

500

1000

1500

2000

2500

3000

Wa

ter p

rod

uctio

n ra

te (S

M3

/DA

Y)

Gas production rateWater production rate

01/1

4

01/1

6

01/1

8

01/2

0

01/2

2

01/2

4

01/2

6

01/28

01/30

0

1000

2000

3000

4000

5000

Oil

pro

du

cti

on

ra

te (

SM

3/D

AY

)

0

2M

4M

6M

8M

10M

Oil p

rod

uc

tion

tota

l (SM

3)

Oil production rateOil production total

Field rate limitations (design rates) M structure

Oil (Sm3/sd) 5 000

Liquid (Sm3/sd) 5 000

Water (Sm3/sd) 3 000

Gas (Sm3/sd) 400 000

Lift gas (Sm3/sd) 250 000

2012-03-144

Page 5: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Svalin Metering ConceptChallenge: GVF Oil Outlet < 1%

Test separator

Inlet separator

Multiphase meters

Svalin Subsea

Topside Grane

Metering

Metering

Metering

Gas

Gas

Oil

Oil

Water

Choke

Choke

CGas lift

M

Metering

Water

2012-03-145

Page 6: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Deviations from ”Forskrift om måling av petroleum for fiskale formål og for beregning av CO2 –avgift”

• Simplified metering of oil

− NPD: “Salg og allokeringsmåling av olje ± 0,30% av standard volum ”

− Svalin specification: ± 5% of oil mass rate (multiphase meter)

• Justification

− 1st stage separator (test separator)

− Small field

− Cost Benefit Analysis (NPD §4):

• 8 MNOK may be invested for a 1% reduction in measurement uncertainty

for oil

2012-03-146

Page 7: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Concept Study, Test Separator Metering upgrade• Gas

− Double Ultrasonic flow meters in parallel (two metering runs)

− Gas analyzer (double gas chromatograph)

− Redundant P&T

• Oil

− Automatic condensate sampler for daily/monthly samples, and water cut meter

− Double Coriolis flow meters in series for condition based

maintenance/calibration

− Master Coriolis flow meter on normally bypass (double block/bleed valves for

removal during operation)

− Redundant P&T

• Water ???

2012-03-147

Page 8: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Amy Ross, NSFMW 2011 3’’ Coriolis Vertical Oil-Gas Flow

2012-03-148

Page 9: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Fiscal metering test separator

2012-03-149

Page 10: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Technology Qualification Program for Coriolis two-phase meter (Technology Readiness Level 4)

2012-03-1410

Page 11: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Qualification Activity TRL 4

Qualification Activity (QA) Acceptance Criteria (AC)

Carry out horizontal flow tests of Coriolis 4” meters at a recognized laboratory.

Test matrix shall be based on Svalin design parameters.

Identify working range and flow conditions:• Absolute mass flow uncertainty < 2%

Test minimum two different Coriolis designs, i.e. two different vendors.

Verify and rank the performance of the meters

2012-03-1411

Page 12: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Test matrix

Kinematic viscosity 14 cStTemperature 37 oCDensity 833 kg/m3

2012-03-1412

Page 13: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

NEL Oil Flow Test Facility (± 0.08% at k=2)Traceable to National Standards and is accredited to ISO 17025 by UKAS

PDCoriolis

2012-03-1413

Page 14: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ MicroMotion Coriolis Installed at NEL

2012-03-1414

Page 15: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Baseline Calibration of 4’’ Endress & Hauser

2012-03-1415

Page 16: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ E & H: Two Phase Oil & Gas Tests Deviation in Mass Flow Error

2012-03-1416

Page 17: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

-5,00

-4,00

-3,00

-2,00

-1,00

0,00

1,00

2,00

3,00

4,00

5,00

0,00 10,00 20,00 30,00 40,00 50,00 60,00

Mas

s Fl

ow

Err

or

[%]

Ref. Mass Flow [kg/s]

4'' E & H Coriolis, s/n: 980B5302000, Jan 2012 Mass Flow Error

0.0% GVF

0.5% GVF

1.0% GVF

1.5% GVF

4’’ E & H: Two Phase Oil & Gas TestsDeviation in Mass Flow Error

2012-03-1417

Page 18: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ E & H: Two Phase Oil & Gas TestsDeviation in Mass Flow Error

2012-03-1418

Page 19: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ E & H: Two Phase Oil & Gas TestsMixture Density Error

2012-03-1419

Page 20: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ E & H: Two Phase Oil & Gas TestsMixture Density Error

2012-03-1420

Page 21: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Baseline Calibration of 4’’ MicroMotion

2012-03-1421

Page 22: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ MicroMotion: Two Phase Oil & Gas TestsDeviation in Mass Flow Error

2012-03-1422

Page 23: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

-5,00

-4,00

-3,00

-2,00

-1,00

0,00

1,00

2,00

3,00

4,00

5,00

0,00 10,00 20,00 30,00 40,00 50,00 60,00

Ma

ss F

low

Err

or

[%]

Ref. Mass Flow [kg/s]

4'' MicroMotion Coriolis, s/n: 11035721, Jan 2012 Mass Flow Error

0.0% GVF

0.5% GVF

1.0% GVF

1.5% GVF

3.0% GVF

4’’ MicroMotion: Two Phase Oil & Gas TestsDeviation in Mass Flow Error

2012-03-1423

Page 24: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ MicroMotion: Two Phase Oil & Gas TestsDeviation in Mass Flow Error

2012-03-1424

Page 25: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ MicroMotion: Two Phase Oil & Gas TestsMixture Density Error

2012-03-1425

Page 26: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

4’’ MicroMotion: Two Phase Oil & Gas TestsMixture Density Error

2012-03-1426

Page 27: Coriolis two-phase meter · 2019-03-20 · Concept Study, Test Separator Metering upgrade • Gas − Double Ultrasonic flow meters in parallel (two metering runs) − Gas analyzer

Conclusions• New application or product

− Test flow meter at intended environment !!!!

• A robust two-phase flow meter has yet to be found/developed.

• Coriolis:

− The presence of small levels of gas results in significant mis-measurement of

the oil flow.

− At low GVF the mass flow respons improved sligthly with increased oil flow.

− The under-reading of the mass flow and density increases with increasing

GVF (over-reading at low GVF may occur)

• MicroMotion over-reading of 142% at 0.5% GVF at 6kg/s.

• E&H under-reading of 50% at 4% GVF at 50kg/s.

• Simplified metering is a challenge !