CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits...

17
NASIR KHAN SBL Meeting 2014, Nov 21 st Measurement of Primary Emission Spectrum of Mixed LS

description

APPARATUS – Agilent Technology Cary Eclipse Spectrophotometer METHODS EMISSION MEASUREMENTS

Transcript of CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits...

Page 1: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

NASIR KHAN

SBL Meeting 2014, Nov 21st

Measurement of Primary Emission Spectrum of Mixed LS

Page 2: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

CONTENTS

• Emission Measurements

• Samples Preparation

• Fluorescence spectra

• Difficulties and limits

• Results and conclusions

Page 3: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

• APPARATUS– Agilent Technology Cary Eclipse Spectrophotometer

• METHODS

EMISSION MEASUREMENTS

Page 4: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

– Diluted PPO in Cyclohexane.

– Diluted bisMSB in Cyclohexane.

– LAB-LS (3g/L PPO + 30mg/L bisMSB)

– Ultima Gold-F (UGF)

– UGF-LS (LAB-LS + UG-F in 9:1 by weight)

SAMPLE PREPARATION

Page 5: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

PPO Back Face samples excited at 300nm• Different concentrations• Low concentration gives the expected primary emission using Back Face geometry • More concentration; more stokes shift• Self absorption

Page 6: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

PPO Front Face

Concentration makes small difference

Page 7: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

PPO Front Face & Back FaceEven concentrated sample gave primary spectrum

Page 8: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

bisMSB Front Face & Back Face

Method makes valuable changes

Page 9: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

LAB-LS Front Face and Back Face We expect Front Face LAB-LS should show spectrum of PPO as primary emission, But it does not show, as the light can be absorbed by bisMSB also.Front Face of a liquid gives detail of its constituents depending upon the excitation range, LAB-LS shows all components.

Page 10: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

Continue.…Using the same Front Face method we tried to find the primary emission of Ultima Gold-F.But unfortunately we found no fruitful results as we don’t know its constituents.

Page 11: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

UGF

Back Face

Front Face

Page 12: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

LS-UGFGreen, Back Face, will reach to detector.

Page 13: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

CONCLUSION Method

– Self-absorption & Re-emission problem in Back Face setup.

– Front Face geometry adopted.

– We expect primary emission of LAB-LS will be more like PPO.

Spectroscopy

– We can not find Primary emission of LAB-LS or bisMSB or UG-F with

photo spectrometer.

– Using X-Ray spectroscopy we may find primary emission of LAB-LS

and may also of Ultima Gold-F.

Page 14: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

Thank you forattention….!

Page 15: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.
Page 16: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

bisMSB Back Face Excited at 340nm

– Low concentration is giving the expected Primary emission.– High concentration shows stoke shift.

Page 17: CONTENTS Emission Measurements Samples Preparation Fluorescence spectra Difficulties and limits Results and conclusions.

Front Face bisMSB Different concentration = same expected Emission.