Presented and adapted by Todd Lang, MD
![Page 2: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/2.jpg)
Improve survivalStabilize patient/relieve symptomsDecrease need for other interventions
By EMSIn EDSpecifically, intubations
Decrease ICU (and transfers out of area)
![Page 3: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/3.jpg)
Reasonably well studied and seems to:Be safe—very rare complications, EMT-B in
MI, WIBe effective—several studies show effects
above both pre- and in-hospitalBoussignac device seems to be best current
option to decrease up front cost for this rare intervention
![Page 4: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/4.jpg)
CPAPNon-invasiveEasily discontinuedEasily adjustedUse by EMT-BMinimal complicationsDoes not require
sedationComfortable
IntubationInvasiveIntubated stays
intubatedRequires highly
trained personnelSignificant
complicationsCan require sedation
or RSIPotential for infection
![Page 5: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/5.jpg)
Conscious patientsTransient interventionIn the “pre-crash” phase of illnessNo risk of complications that RSI bringsNo downstream healthcare implications like
intubation/RSIEasier to do with limited people
![Page 6: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/6.jpg)
Airway pressure maintained at set level throughout inspiration and expiration
Maintains patency of small airways and alveoli
Improves gas exchangeImproves delivery of bronchodilatorsMoves extracellular fluid into vasculatureReduces work of breathing
![Page 7: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/7.jpg)
Adult Respiratory Distress(Age greater than 12)
Routine Medical Assessment
Oxygen2 LPM via Nasal Cannula
Titrate to maintain Pulse ox of >92%
Is Patient a candidate for Mask CPAP?-Respiratory Rate > 25 / min
-Retractions or accessory muscle use-Pulse ox < 94% at any time
See CPAP Protocol
No
Yes
No
No
Is the Patient wheezing and/or doesthe Patient have a history of Asthma/COPD?
Does the Patient have rales and/or does the Patient have a history of congestive heart
failure (CHF)?
YesAdminister Albuterol / Atrovent by Nebulizer
Contact Medical Control, consider
sublingual NTG ifsystolic BP is >100mmHg,
possible furosemide if long transport
Yes
Contact Medical Control
![Page 8: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/8.jpg)
Asses Patient, record vital signsand pulse ox before applying oxygen
Does the Patient meet two or moreInclusion Criteria?
No
Yes
Does the Patient meet anyExclusion Criteria?
Continue standardRespiratory Distress Protocol
Administer CPAP5 cm H2O of pressure AND
Reassess patient, vital signs, andrespiratory distress scale every 5 min.
Patch to Medical ControlRespiratory Distress Protocol
Patient condition is stableor improving
Continue CPAPReassess patient every
5 minutes
Patient condition is deterioratingDecreasing LOC
Decreasing Pulse Ox
Notify Medical Control
Remove CPAPApply BVM Ventilation
Mask CPAP for EMS
CPAP Inclusion Criteria(2 or more of the following)
-Retractions or Accessory muscle use-Respiratory Rate > 25 / minutes
-Pulse Ox < 94% at any time
CPAP Exclusion Criteria-Unable to follow commands
-Apnea-Vomiting or active GI bleed
-Major trauma / pneumothorax
Conditions Indicated for CPAPCongestive Heart Failure
COPD / AsthmaPneumonia
Yes
No
Complete Airway Data Form and submit PHC for each patient placed
on CPAP
![Page 9: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/9.jpg)
CHF Pulmonary Edema
Near Drowning Inhalation Exposure
COPD Asthma Pneumonia
![Page 10: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/10.jpg)
Code/cardiac arrestTrauma to faceUnable to get a seal with maskPneumothoraxVomiting/upper GI bleedingPatient intolerance (despite treatment with
midazolam)
![Page 11: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/11.jpg)
Oxygen supply is rapidly depleted at higher pressure rates
Pressure levelMost studies show 5cm H20 sufficientComplication rate goes up with pressure5cm is probably best place to start EMS CPAP
and titrate if needed from there
![Page 12: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/12.jpg)
CPAP is a non-invasive procedure that is easily applied and can be easily discontinued without untoward patient discomfort
Data supports its use in CHF, pulmonary edema, COPD/Asthma, and pneumonia
Data support EMS CPAP use and its safety and benefits to patients
![Page 13: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/13.jpg)
Boussignac CPAP System
Instructions for Use
![Page 14: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/14.jpg)
Select the appropriate size face mask for your patient:
- Child Mask (Size 3) - Small Adult Mask (Size 4) - Medium Adult Mask (Size 5) - Large Adult Mask (Size 6)
![Page 15: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/15.jpg)
Set Oxygen flow to deliver CPAP in cmH2O of water pressure:
- 15 liters = 5cmH2O - 20 liters = 7.5 cmH2O - 25 liters = 10 cmH2O
![Page 16: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/16.jpg)
Attach the head strap to the hook rings on one side of the Boussignac CPAP mask.
Explain to the patient how the Boussignac CPAP will help their breathing.
Gently hold the mask to the patients face insuring a good face/mask seal.
Turn the flow control device to the desired liters/min, generally 15 l/min, to begin the CPAP.
Gradually adjust the flow to achieve the desired level of CPAP.
![Page 17: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/17.jpg)
Move the head strap around the patient’s head. Insure that the round disk on the head strap is located on the crown of the patient’s head. Attach the head strap on the hook rings.
Check around the mask for any leaks.
Adjust the mask and/or head strap accordingly to insure the patient’s comfort.
Monitor patient’s vital signs.
![Page 18: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/18.jpg)
If the patient requires suctioning of the oral cavity, insert French size suction catheter through the open end of the Boussignac CPAP System.
CPAP pressure will not be affected.
![Page 19: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/19.jpg)
Fill the nebulizer with the prescribed medication.
![Page 20: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/20.jpg)
Insert the 22 mm male end of the nebulizer into the face mask.
Insert the 22 male end (white) of Boussignac CPAP into the nebulizer.
Monitor patient’s vital signs.
![Page 21: Concepts and Use Presented and adapted by Todd Lang, MD.](https://reader036.fdocuments.in/reader036/viewer/2022062407/56649cd65503460f9499d969/html5/thumbnails/21.jpg)
Connect the funnel end of the nebulizer’s gas delivery tube to a gas source.
Turn on gas source to 6 liters/min. to power the nebulizer.
Maintain O2 flow to Boussignac CPAP System at 15 l/min. (5cmH2O).