Transcript of Complying with Wearable Health Device Regulation
Complying with Wearable Health Device Regulation
Which devices does the TGA regulate and how do they do it?
Dr Lee Walsh
Senior Engineer, Biomaterials and Engineering Section
Laboratories Branch
Therapeutic Goods Administration
WT | Wearable Technologies Conference 2017 Australia
7 Dec 2017
Overview of the TGA• Part of the Australian Government Department of
Health
• Administers the Therapeutic Goods Act 1989
• Main offices in Canberra – satellite offices in Sydney,
Melbourne, Adelaide and Brisbane
• Operations cost recovered from industry
• Mission: To safeguard and enhance the health of the
Australian community through the effective and timely
regulation of therapeutic goods
Purpose built science and engineering buildingWT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 1
Overview of the TGA – What we do
• Evaluate therapeutic goods before they are supplied
• Focus on safety, quality and performance
• Monitor products once they are on the market
• Allow for access to unapproved goods in certain
circumstances
• Provide input to therapeutic goods standards
• We do not make decisions based on value for
money or make decisions about which products
receive Government subsidy
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 2
Who does this work?Approximately 750 staff made up of:
Biomedical scientists Engineers Physiotherapists Medical officers
Pharmacists Nurses Toxicologists Lawyers
Nutritionists Dieticians Scientists Administrative staff
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 3
The legislation
The TGA makes decisions based upon:
• Therapeutic Goods Act 1989
• Therapeutic Goods Regulations 1990
• Therapeutic Good (Medical Devices) Regulations 2002
Available on Australian Government: Federal Register of Legislation website
Legislative instruments
• Therapeutics Goods Orders (TGO)
• Excluded Goods Orders
• Medical Device Standards Orders (MDSO)
• Conformity Assessment Standards Orders (CASO)
Available on Therapeutic Goods Administration website
Wearable medical devices
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 4
Is my wearable device regulated by the TGA?41BD What is a medical device
(1) A medical device is:
(a) any instrument, apparatus, appliance, material or other article (whether used
alone or in combination, and including the software necessary for its proper
application) intended, by the person under whose name it is or is to be
supplied, to be used for human beings for the purpose of one or more of the
following:
(i) diagnosis, prevention, monitoring, treatment or alleviation of disease;
(ii) diagnosis, monitoring, treatment, alleviation of or compensation for an
injury or disability;
(iii) investigation, replacement or modification of the anatomy or of a
physiological process;
(iv) control of conception;
and that does not achieve its principal intended action in or on the human body
by pharmacological, immunological or metabolic means, but that may be
assisted in its function by such means; or
(aa) any instrument, apparatus, appliance, material or other article specified under
subsection (2A); or
(ab) any instrument, apparatus, appliance, material or other article that is included
in a class of instruments, apparatus, appliances, materials or other articles
specified under subsection (2B); or
(b) an accessory to an instrument, apparatus, appliance, material or other article
covered by paragraph (a), (aa) or (ab).
(2) For the purposes of paragraph (1)(a), the purpose for which an instrument, apparatus,
appliance, material or other article (the main equipment) is to be used is to be
ascertained from the information supplied, by the person under whose name the main
equipment is or is to be supplied, on or in any one or more of the following:
(a) the labelling on the main equipment;
(b) the instructions for using the main equipment;
(c) any advertising material relating to the main equipment;
(d) technical documentation describing the mechanism of action of the main
equipment.
(2A) The Secretary may, by notice published in the Gazette or on the Department’s website,
specify a particular instrument, apparatus, appliance, material or other article for the
purposes of paragraph (1)(aa). The notice is not a legislative instrument.
(2B) The Secretary may, by legislative instrument, specify a particular class of instruments,
apparatus, appliances, materials or other articles for the purposes of paragraph (1)(ab).
(3) The Secretary may, by order published in the Gazette or on the Department’s website,
declare that a particular instrument, apparatus, appliance, material or other article, or
that a particular class of instruments, apparatus, appliances, materials or other articles,
are not, for the purposes of this Act, medical devices.
Is it a medical device as defined by the TG Act?WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017
5
My tech is a medical device, what now?
Australian regulatory guidelines for medical devices (ARGMD)
• Available on Therapeutic Goods Administration website
• Currently under review
In summary:
• Australian Register of Therapeutic Goods (ARTG)
• Risk based regulation
• Classification
• Essential Principles
• Quality System
• Evidence
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 6
Risk based regulation and classification
Higher risk devices receive higher scrutiny
Class I
The lowest level
Class Is and Class IIa Class IIb
Class Im
Class III and
AIMD
The highest level
Devices are classified with regard to intended purpose
Class 1
The lowest level
Class 2
The highest level
Class 3 Class 4
IVD are classified with regard to the risk to public health
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017
Therapeutic Goods (Medical Device) regulations 2002, Schedule 2
7
Inclusion on the ARTG
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 8
My device is included in the ARTG, am I done?Incident Report Investigation Scheme (IRIS)
• Sponsor must report the details of events associated with their devices that have,
or could have resulted in serious injury or death – TG (MD) Regs 2002
Risk assessment
Investigation Referral
Laboratory testing
Regulatory action
Recall Safety alert Report Cancel or suspend
Other action Product improvement
Data logged for ongoing monitoring
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 9
Other considerationsTGA can approve the supply of unregistered therapeutic goods
• Authorised prescribers
• Special access scheme
• Clinical trials
• Personal importation
• See Therapeutic Goods Administration website
There are penalties for (TG Act 1989, Part 4-11):
• Supplying a medical device that does not comply with the Essential Principles
• Not applying conformity assessment procedures
• Supplying a medical device that is not included in the register
There are advertising and labelling requirements (TG Act 1989, Chapter 5)
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 10
The Essential PrinciplesTherapeutic Goods (Medical Devices) Regulations 2002, Schedule 1
• A legislated requirement for medical devices (TG Act 1989, Section 41CA)
• Used in pre-market evaluations
• Used in post market monitoring
• Standards may or may not be sufficient to demonstrate compliance
TGA assessors will assess your device against the Essential Principles
You should assess your device against the Essential Principles
You should consider the Essential Principles at every stage of you medical device’s life
WT | Wearable Technologies Conference 2017 Australia, 7 Dec 2017 11
The Essential PrinciplesTherapeutic Goods (Medical Devices) Regulations 2002, Schedule 1
General principles1. Use of medical devices not to compromise health
and safety2. Design and construction of medical devices to
conform to safety principles3. Medical devices to be suitable for intended
purpose4. Long-term safety5. Medical devices not to be adversely affected by
transport or storage6. Benefits of medical devices to outweigh any side
effects
Principles about design and construction 7. Chemical, physical and biological properties8. Infection and microbial contamination9. Construction and environmental properties10. Medical devices with a measuring function11. Protection against radiation12. Medical devices connected to or equipped with an energy
source13. Information to be provided with medical devices14. Clinical evidence15. Principles applying to IVD medical devices only
Austrailan Government: Federal Register of Legislation website