Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC...

16
Stationary Phase Specifications | Phases Compared According to Relative Hydrophobicity | Phases Compared According to Relative Polarity | Categorization of Phases According to Hydrophobicity and Polarity | Comparison of Column Efficiency for a Neutral Compound | Comparison of Column Efficiency for Basic Compounds | Phases Grouped According to Silanol Activity | Comparison of Phases According to Metal Activity Fourth Edition June 2008 Comparison Guide to C18 Reversed Phase HPLC Columns Comparison Data on Commonly Used C18 Phases

Transcript of Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC...

Page 1: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Stationary Phase Specifications | Phases Compared According to Relative Hydrophobicity | Phases Compared According to Relative Polarity | Categorization of Phases According to Hydrophobicity and Polarity | Comparison of Column Efficiency for a Neutral Compound | Comparison of Column Efficiency for Basic Compounds | Phases Grouped According to Silanol Activity | Comparison of

Phases According to Metal Activity

Fourth Edition June 2008

Comparison Guide to C18 Reversed Phase HPLC Columns

Comparison Data on Commonly Used C18 Phases

Page 2: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

MAC-MOD Analytical 103 Commons Court

P.O. Box 587 Chadds Ford, PA 19317

Phone: 1.800.441.7508 Fax: 1.610.358.5993

E-mail: [email protected] Web Site: www.mac-mod.com

Page 3: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Comparison Guide to C18 Reversed Phase HPLC Columns

Introduction

There are so many different C18 columns to choose from, that finding the right column for a particular separation can be very time consuming and expensive. Two apparently similar C18 phases can give very different results. For example, Figure 1 compares the separation of the same sample mixture on a Hypersil HyPurity C18 and a Symmetry C18 column under identical mobile phase conditions. Even though both columns are packed with base deactivated C18 stationary phases, the band spacing (selectivity) between peaks is very different on the two columns. Without more information, it is impossible to predict how the performance of different stationary phases will compare.This Comparison Guide to C18 Reversed Phase HPLC Columns provides basic comparison information on commonly used C18 columns to help you more easily identify similarities and differences before investing time and money in chromatographic testing. Hopefully, this infor-mation will help you find the right column for your application quicker.Only silica based C18 bonded phases are evaluated in this Guide. Other bonded phases, such as C8, CN, Phenyl and polar embedded phases, are excluded.This Guide does not identify an overall “best” column. The column that works best for one application will not necessarily be the column that will work best for other applications. And, there certainly is not a single column that will work best for all applications. However, this Guide can help you identify columns that are likely to perform well so that at least you can narrow the number of columns for chromatographic testing. You may find that this Guide helps you identify several columns that provide good separations and performance. It is always desirable to have more than one column identified for an application, especially if you are running routine assays.Increasingly, chromatographers are seeking to identify alternate brands of HPLC columns suitable for their assays. Having an alternate column choice for a method reduces the risk of “down time” due to column problems such as a change in selectivity from one manufactured lot to another or slow supplier delivery. Finding an alternate or back-up column that will provide acceptable selectivity and performance when substituted into a method can be as expensive and time consuming as finding the right column for developing an initial separation. It is our hope that this Guide will make that job easier by identifying columns with similar chromatographic characteristics.

Figure 1Apparently Similar C18 Phases Can Give Very Different Chromatographic Results

Mobile Phase: 60% CH3CN 40% 50mM KH2PO4, pH 3.2Sample: 1. Uracil 5. N,N-Dimethylaniline 2. Pyridine 6. 4-Butylbenzoic acid 3. Phenol 7. Toluene 4. Dimethyl phthalate

Both Hypersil HyPURITY C18 and Symmetry C18 are base deactivated phases. You would expect them to provide similar performance, and in some cases they do. However, in the example given here you can see significant differences in peak retention times, selectivity and even peak shape.

This Guide provides the following comparison data on commonly used C18 phases:• Stationary Phase Specifications

Specifications provided by column manufacturers• Phases Compared According to Relative Hydrophobicity

Retention data for hydrophobic and neutral compounds• Phases Compared According to Relative Polarity• Categorization of Phases According to Hydrophobicity

and Polarity• Comparison of Column Efficiency for a Neutral

Compound• Comparison of Column Efficiency for Basic Compounds

Also measures peak tailing• Phases Grouped According to Silanol Activity• Phases Compared According to Metal Activity

0� 1� 2� 3� 4� 5� 6� 7� 8� 9� 10� 11Time (min)

1 23 4 5

6

7

0 1 2 3 4 5 6 7 8 9 10 11Time (min)

1

2

3 4

5

6

7

Comparison Guide to C18 Reversed Phase HPLC Columns | 1

Hypersil HyPURITY C18

Symmetry C18

Page 4: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

0 1 2 3 4 5 6 7Time (min)

1

2

3

Figure 3Specifications of C18 Stationary Phases

Particle Pore Surface Carbon High Purity Stationary Phase Size (µm) Size (Å) Area (m2/g) Load (%) Endcapped Silica

ACE C18 5 100 300 15.5 yes yes ACE C18-300 5 300 100 9.0 yes yes ACE C18-HL 5 90 400 20.0 yes yes µBondapak C18 10 125 330 10 yes no Capcell Pak AG C18 5 120 300 15 yes no Capcell Pak UG C18 5 120 300 15 yes yes Develosil ODS-HG 5 140 300 18 yes yes Develosil ODS-MG 5 100 450 15 yes yes Develosil ODS-UG 5 140 300 18 yes yes Exsil ODS 5 100 200 11 yes no Exsil ODS1 5 100 200 11 yes no Exsil ODSB 5 100 200 12 yes no Gemini C18 5 110 375 14 yes yes Hichrom RPB 5 110 340 14 yes yes Hypersil BDS C18 5 130 170 11 yes no Hypersil GOLD 5 180 200 10 yes yes Hypersil HyPurity C18 5 180 200 13 yes yes Hypersil ODS 5 120 170 10 yes no Inertsil ODS 5 100 350 14 yes no Inertsil ODS3 5 100 450 15 yes yes Inertsil ODS2 5 150 320 18.5 yes yes Kromasil C18 5 100 340 19 yes yes LiChrosorb RP-18 10 100 300 17 no no LiChrospher RP-18 5 100 350 21.6 no no Luna 5 C18(2) 5 100 400 17.5 yes yes Novapak C18 4 60 120 7.3 yes no Nucleosil C18 5 100 350 15 yes no Nucleosil C18 HD 5 100 - 20 yes yes Nucleosil C18AB 5 100 350 24 yes no Partisil ODS 10 85 350 5 no no Partisil ODS2 10 85 350 15 yes no Partisil ODS3 10 85 350 10.5 yes no Prodigy ODS2 5 150 310 18.4 yes yes Prodigy ODS3 5 100 450 15.5 yes yes Purospher RP18-e 5 80 500 - yes yes Resolve C18 5 90 200 10 no no SunFire C18 5 100 340 16 yes yes Symmetry C18 5 100 335 19 yes yes TSK ODS-120T 5 120 - 22 yes no TSK ODS-80TM 5 80 - 15 yes no Ultrasphere ODS 5 80 - 12 yes no Vydac 218MS 5 300 70 - yes no Vydac 218TP 5 300 70 8 yes no Vydac Selectapore 300M 5 300 70 - yes yes Vydac Selectapore 300P 5 300 70 - yes yes Vydac Selectapore 90M 5 90 250 - yes yes Waters Spherisorb ODS1 5 80 220 6.2 no no Waters Spherisorb ODS2 5 80 220 11.5 yes no Waters Spherisorb ODSB 5 80 220 11.5 yes no XTerra MS C18 5 125 - 15.5 yes — YMC J’Sphere ODS H80 4 80 510 22 yes no YMC J’Sphere ODS M80 4 80 510 14 yes no YMC ODS A 5 120 300 17 yes no YMC ODS AM 5 120 300 17 yes no YMC Pro C18 5 120 335 16 yes yes Zorbax Extend C18 5 80 180 12.5 yes yes Zorbax ODS 5 70 330 20 yes no Zorbax Rx-C18 5 80 180 12 no yes Zorbax SB-C18 5 80 180 10 no yes Zorbax XDB-C18 5 80 180 10 yes yes

Figure 2High Purity Silicas Provide Better Peak Shape for Basic Compounds

High Purity Silica (ACE C18)

Acidic Silica (Waters Spherisorb ODS2)

Mobile Phase: 60% CH3OH, 40% H2OSample: 1. Uracil 2. Pyridine 3. Phenol

Interaction between cationic compounds and acidic silanol sites on the surface of silica stationary phase supports can contribute to retention and peak tailing. Phases made with high purity silica (less acidic silica) generally can be expected to provide better peak shape for basic compounds.

0 1 2 3 4 5Time (min)

1

2

3

2 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 5: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Figure 4C18 Phases Compared According to Relative Hydrophobicity

Stationary Phase Specifications

Stationary phase specifications provide basic information that can be helpful in deciding which phases to select for evaluation. For example, phases with high surface area and high carbon load will generally retain hydrophobic compounds longer than phases with low surface area and low carbon load. If you are analyzing macromolecules, such as peptides and proteins, a wider pore (200 — 300 Å) phase usually provides better performance than a phase with small pores. New high purity silicas usually provide better peak shape for basic compounds than older, more acidic silicas (see Figure 2). Stationary phase specifications, however, will not give you enough information to accurately predict retention or band spacing (selectivity). This is especially true when separating polar compounds.

Inertsil ODS3

LiChrospher RP-18

YMC J’Sphere ODS H80

Nucleosil C18

Kromasil C18

Develosil ODS-MG

Partisil ODS2

LiChrosorb RP-18

Symmetry C18

Inertsil ODS

Develosil ODS-UG

Purospher RP18-e

YMC Pro C18

Prodigy ODS3

Zorbax ODS

Nucleosil C18HD

SunFire C18

YMC ODS AM

YMC ODS A

Inertsil ODS2

Cosmosil C18-AR

Prodigy ODS2

TSK ODS-80TM

ACE C18

Luna C18(2)

Genesis C18

Nucleosil C18AB

Waters Spherisorb ODS2

Zorbax Extend C18

Develosil ODS-HG

Resolve C18

Zorbax RX C18

Zorbax XDB C18

Waters Spherisorb ODSB

Hypersil BDS C18

Hypersil HyPurity C18

Ultrasphere ODS

TSK ODS-120T

Partisil ODS3

Hichrom RPB

Waters Spherisorb ODS1

µBondapak C18

Exsil ODS

Capcell Pak AG C18

YMC J’Sphere ODS M80

Capcell Pak UG C18

Zorbax SB C18

Exsil ODSB

XTerra MS C18

ACE C18-300

Hypersil ODS

Hypersil GOLD

Novapak C18

Exsil ODS1

Vydac SelectaPore 90M

Vydac SelectaPore 300P

Vydac 218TP

Partisil ODS

Vydac 218MS

Vydac SelectaPore 300M

Mobile Phase: 90% CH3OH, 10% H2OElution Order: 1. Dimethyl phthalate 2. Toluene 3. Biphenyl 4. PhenanthreneColumn Temperature: 24°C

Comparison Guide to C18 Reversed Phase HPLC Columns | 3

Page 6: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Inertsil ODS3Purospher RP-18e

Kromasil C18YMC J'Sphere ODS H80

ACE C18-HLDevelosil ODS-MG

Symmetry C18Prodigy ODS3

SunFire C18Zorbax ODS

LiChrospher RP-18Luna C18(2)

Zorbax Extend C18Inertsil ODS2Nucleosil C18

Gemini C18Nucleosil C18 HDZorbax XDB-C18

Prodigy ODS2Develosil ODS-UG

Zorbax Rx C18Inertsil ODS

Waters Spherisorb ODS2Partisil ODS2

ACE C18YMC ODS A

YMC Pro C18YMC ODS AM

XTerra MS C18Capcell Pak AG C18

Ultrasphere ODSResolve C18

Zorbax SB-C18TSK ODS-80TM

Waters Spherisorb ODSBHichrom RPB

LiChrosorb RP-18Develosil ODS-HG

Exsil ODSNucleosil C18AB

Hypersil ODSHypersil BDS

YMC J'Sphere ODS M80TSK ODS-120T

Partisil ODS3Hypersil HyPURITY C18

Waters Spherisorb ODS1Vydac Selectapore 90M

Novapak C18Hypersil GOLD

Exsil ODS1ACE C18-300

µBondapak C18Exsil ODSB

Vydac 218MSVydac 218TP

Vydac Selectapore 300PVydac Selectapore 300M

Partisil ODS

0.0 0.5 1.0k (Toluene)

1.5 2.0 2.5

Figure 5C18 Phases Ranked According to Retention for Toluene

Mobile Phase: 80% CH3OH 20% 25mM KH2PO4, pH 6.0Sample: TolueneTemperature: 24°C

Phases Compared According to Relative Hydrophobicity

Hydrophobicity is measured as the retention of a hydrophobic solute, phenanthrene. Figure 4 gives a comparison of hydrophobicity with the C18 phases listed according to hydrophobicity. Notice, however, that the retention for dimethyl phthalate, the least hydrophobic solute in the mixture, cannot always be predicted from the hydrophobicity ranking. Some low hydrophobicity phases actually have greater retention for dimethyl phthalate than some high hydrophobicity phases. We find that this is not unusual when separating polar compounds. Phases that are significantly more retentive for hydrophobic analytes may show only slightly more retention for polar compounds than low hydrophobicity phases, and sometimes they show less. Alternative Test for Hydrophobicity

Toluene can also be used as a probe to measure hydrophobicity. Notice that the ranking of C18 phases according to retention for toluene (Figure 5) is slightly different from the ranking according to retention for phenanthrene (Figure 4).

4 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 7: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

YMC Pro C18Develosil ODS-UG

Luna C18(2)Gemini C18Exsil ODSB

Zorbax XDB-C18Develosil ODS-MG

ACE C18Capcell Pak UG C18

Develosil ODS-HGHichrom RPBACE C18-300

SunFire C18Prodigy ODS2

Hypersil HyPurity C18Inertsil ODS3Kromasil C18

Nucleosil C18 HDACE C18-HL

Hypersil GOLDProdigy ODS3

Zorbax Extend C18Inertsil ODS

XTerra MS C18Inertsil ODS2Partisil ODS2

Symmetry C18YMC ODS AM

YMC ODS ANucleosil C18AB

Capcell Pak AG C18Partisil ODS3

Vydac Selectapore 90MNucleosil C18

Ultrasphere ODSVydac Selectapore 300MWaters Spherisorb ODSB

Purospher RP18-eTSK ODS-120T

YMC J'Sphere ODS M80YMC J'Sphere ODS H80

Zorbax Rx-C18Zorbax SB-C18

LiChrosorb RP-18Novapak C18

TSK ODS-80TMµBondapak C18

Vydac 218MSWaters Spherisorb ODS2

Exsil ODSLiChrospher RP-18

Vydac Selectapore 300PVydac 218TP

Hypersil BDS C18Partisil ODSZorbax ODS

Exsil ODS1Waters Spherisorb ODS1

Resolve C18Hypersil ODS

0 1 2k (Pyridine) / k (Phenol)

3 4 5

HighPolarity

LowPolarity

Figure 6C18 Phases Ranked According to Polarity

Phases Ranked According to Relative Polarity

There have been several chromatographic tests suggested for measuring polarity of stationary phases. Although there is no one test that we think provides a definitive measurement, we have chosen to use the ratio of k values for pyridine and phenol as our measure of relative polarity for these C18 phases.Figure 6 ranks stationary phases according to relative polarity using the test conditions given. In this ranking, there is not necessarily a significant difference between consecutive listings. If a different mobile phase condition was used for the test, e g., a lower mobile phase pH, or if different probes were used, the ranking may be somewhat different. However, phases at the high polarity end of the ranking and phases at the low polarity end of the ranking are likely to test that way under most polarity tests conditions. Therefore, this ranking can be used to identify relative differences and similarities in polarity that can affect selectivity for polar compoundsSince silanol activity is a major contributor to phase polarity, the test conditions used here to measure polarity have also been used by some chromatographers as an indication of silanol activity. This seems consistent with the fact that most phases at the high polarity end of the ranking use more acidic silicas as stationary phase supports where phases at the low polarity end of the ranking use less acidic (high purity) silicas. However, there are other factors that contribute to the retention of pyridine and phenol that prevent us from using their relative retention as a reliable measure of silanol activity. For example, Inertsil ODS has moderate polarity but shows significant silanol activity in other tests. Also, we see that Prodigy ODS2 tests with similar polarity as the ACE C18, but in tests for silanol activity, the ACE C18 shows significantly less silanol activity (see Figures 10 and 13).

Mobile Phase: 60% CH3OH 40% H2OPolarity = k Pyridine / k PhenolTemperature: 24°C

Comparison Guide to C18 Reversed Phase HPLC Columns | 5

Page 8: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

High Polarity/ High Polarity/ High Polarity/ Low Hydrophobicity Moderate Hydrophobicity High HydrophobicityExsil ODS1 Exsil ODS LiChrosorb RP-18Hypersil ODS Hypersil BDS C18 LiChrospher RP-18Novapak C18 Resolve C18Partisil ODS TSK ODS-80TMVydac 218MS µBondapak C18Vydac 218TP Waters Spherisorb ODS1 Vydac Selectapore 300P Waters Spherisorb ODS2 Zorbax ODS

Moderate Polarity/ Moderate Polarity/ Moderate Polarity/ Low Hydrophobicity Moderate Hydrophobicity High HydrophobicityVydac Selectapore 300M Capcell Pak AG C18 ACE C18-HLVydac Selectapore 90M Inertsil ODS Nucleosil C18Xterra MS C18 Inertsil ODS2 Partisil ODS2 Nucleosil C18 AB Symmetry C18 Partisil ODS3 YMC J’Sphere ODS H80 Prodigy ODS3 Purospher RP18-e TSK ODS-120T Ultrasphere ODS Waters Spherisorb ODSB YMC J’Sphere ODS M80 YMC ODS A YMC ODS AM Zorbax Extend C18 Zorbax Rx-C18 Zorbax SB-C18

Low Polarity/ Low Polarity/ Low Polarity/ Low Hydrophobicity Moderate Hydrophobicity High HydrophobicityACE C18-300 ACE C18 Develosil ODS-MGExsil ODSB Capcell Pak UG C18 Inertsil ODS3Hypersil GOLD Develosil ODS-HG Kromasil C18 Develosil ODS-UG Gemini C18 Hichrom RPB Hypersil HyPurity C18 Luna C18(2) Nucleosil C18 HD Prodigy ODS2 SunFire C18 YMC Pro C18 Zorbax XDB-C18

Figure 7C18 Phases Grouped According to Hydrophobicity and PolarityPhases are listed in alphabetical order by category.

Figure 8Chromatographic Comparison of Stationary Phases from Different Categories

ACE C18Low Polarity/Moderate Hydrophobicity

Hypersil BDS C18High Polarity/Moderate Hydrophobicity

Symmetry C18Moderate Polarity/High Hydrophobicity

Mobile Phase: 80% CH3OH, 20% 25mM KH2PO4, pH 6.0Sample: 1. Norephedrine 4. Imipramine2. Nortriptyline 5. Amitriptyline3. Toluene

Categorization of phases according to hydrophobicity and polarity.

The hydrophobicity and polarity data can be used to group phases with similar characteristics into categories. The following criteria was used for the categories:Hydrophobicity

High > 2.0 Moderate 1.30 to 1.99 Low < 1.30

High > 1.00 Moderate 0.50 to 0.99 Low < 0.50

k pyridine k phenol

Polarity

k for phenanthrene

Figure 8 provides an example of how columns from different polarity/hydrophobicity categories will compare. In this separation of antidepressants, Symmetry C18 (high hydrophobicity) is slightly more retentive than Hypersil BDS-C18 (moderate hydrophobicity), and the band spacing of ACE C18 (low polarity) is more similar to Symmetry C18 (moderate polarity) than it is to Hypersil BDS C18 (high polarity).

6 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 9: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

N/Meter (Toluene)

Exsil ODS1ACE C18

YMC Pro C18ACE C18-300

Zorbax SB-C18ACE C18-HL

Nucleosil C18Waters Spherisorb ODS1

YMC ODS AKromasil C18Hichrom RPB

Zorbax XDB-C18Hypersil ODS

Exsil ODSSymmetry C18

Develosil ODS-UGWaters Spherisorb ODSB

SunFire C18Waters Spherisorb ODS2

Hypersil GOLD C18Zorbax Rx-C18

Luna C18(2)Nucleosil C18AB

Zorbax ODSDevelosil ODS-HGNucleosil C18 HD

YMC ODS AMExsil ODSB

Vydac Selectapore 300MZorbax Extend C18LiChrospher RP-18

Vydac Selectapore 90MUltrasphere ODS

Vydac 218MSHypersil BDS C18

Gemini C18LiChrosorb RP-18

Inertsil ODSHypersil HyPURITY C18

Develosil ODS-MGYMC J'Sphere ODS H80

Vydac 218TPProdigy ODS3

TSK ODS-80TMInertsil ODS3Novapak C18

YMC J'Sphere ODS M80Capcell Pak UG C18

TSK ODS-120TVydac Selectapore 300P

Partisil ODS3Capcell Pak AG C18

Prodigy ODS2Partisil ODS

XTerra MS C18Resolve C18

Purospher RP18-ePartisil ODS2

µBondapak C18Inertsil ODS2

0.0 20,000 40,000 60,000 80,000 100,000 120,000

Figure 9Comparison of Column Efficiency for a Neutral Compound Column efficiency reported as Plates per meter (N/Meter)

Comparison of Column Efficiency for a Neutral Compound

Column efficiency is reported as plates per meter (N/Meter). Using a neutral compound (toluene) for the measurement greatly reduces the effects of secondary retention on the measurement of N and allows us to obtain data that is a better indication of the following factors:• Particle size

Smaller average packing particle size = Larger N

• Particle size distribution Broader particle size distribution = Smaller N

• Packing efficiency Better packing procedures = Larger N

Mobile Phase: 90% CH3OH 10% H2OSample: TolueneTemperature: 24°C

Comparison Guide to C18 Reversed Phase HPLC Columns | 7

Page 10: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

N/Meter (Pyridine)

ACE C18Develosil ODS-UG

YMC Pro C18ACE C18-300

SunFire C18Hypersil GOLD

Develosil ODS-MGHichrom RPB

Zorbax XDB-C18Develosil ODS-HG

Capcell Pak UG C18Exsil ODSB

Gemini C18Inertsil ODS3

Waters Spherisorb ODSBInertsil ODSACE C18-HLYMC ODS A

Zorbax Extend C18Luna C18(2)

Nucleosil C18 HDYMC ODS AM

Vydac Selectapore 300MXTerra MS C18TSK ODS-120T

Kromasil C18Zorbax Rx-C18

Partisil ODS3Prodigy ODS3

µBondapak C18Vydac Selectapore 90M

Purospher RP18-eInertsil ODS2

YMC J'Sphere ODS M80Hypersil HyPURITY C18

Nucleosil C18Zorbax SB-C18

Partisil ODSProdigy ODS2Novapak C18

TSK ODS-80TMVydac Selectapore 300P

LiChrosorb RP-18YMC J'Sphere ODS H80

Vydac 218MSVydac 218TPPartisil ODS2

Nucleosil C18ABCapcell Pak AG C18

Symmetry C18Ultrasphere ODS

Waters Spherisorb ODS2Hypersil BDS C18

Exsil ODS1Exsil ODS

Waters Spherisorb ODS1Zorbax ODS

Hypersil ODSLiChrospher RP-18

Resolve C18

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000 80,000

ACE

Hypersil HyPURITY

Symmetry

Figure 10Comparison of Column Efficiency for a Basic Compound: Pyridine

Figure 11Comparison of Peak Shape

Mobile Phase: 60% CH3OH 40% H2OSample: PyridineTemperature: 24°C

ACE C18 gave the best peak shape and highest column efficiency for pyridine.

Comparison of Column Efficiency for a Basic Compound

Measuring column efficiency using a neutral compound is not very useful in predicting column performance when separating ionic compounds. Interaction between polar solutes and silanol sites on the stationary phase can cause tailing peaks and poor column efficiency.To gain a better understanding of column performance with basic compounds, columns were tested using pyridine and amitriptyline as probes. Although columns are ranked somewhat differently on the two tests, phases at the higher end of the ranking scale can be expected to give better peak shape and higher resolution for basic compounds than phases at the lower end of the scale. Not surprisingly, stationary phases that use high purity silicas exhibit better peak shape and higher column efficiency than stationary phases that use more acidic silicas as their stationary phase supports.

Mobile Phase: 60% CH3OH 40% H2OSample: PyridineTemperature: 24°C

8 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 11: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Figure 12 Comparison of Column Efficiency For a Basic Compound: Pyridine

ACE C18

65,200pl/m

ACE C18-300

52,100pl/m

SunFire C18

46,500pl/m

Zorbax XDB C18

39,400pl/m

Gemini C18

36,100pl/m

Inertsil ODS3V

35,900pl/m

Luna C18(2)

21,900pl/m

XTerra MS C18

18,600pl/m

Zorbax SB-C18

10,000pl/m

Symmetry C18

3,100pl/m

Hypersil BDS C18

600pl/m

Zorbax ODS

100pl/m

Column efficiency measured at 10% pyridine peak height to account for peak tailing effects

Column Dimensions: 250 x 4.6mm, 5µm Sample: 1) uracil 2) pyridine 3) phenolMobile Phase: 60:40 MeOH/H2O Temperature: 24°CFlow: 1.0ml/min

Comparison Guide to C18 Reversed Phase HPLC Columns | 9

Page 12: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

N/Meter (Amitriptyline)

ACE C18-300ACE C18

Hypersil GOLDYMC J'Sphere ODS M80

Gemini C18ACE C18-HL

YMC Pro C18SunFire C18

Develosil ODS-UGLuna C18(2)

Prodigy ODS3Hichrom RPB

Vydac Selectapore 300MNucleosil C18 HD

Inertsil ODS3Develosil ODS-MG

Prodigy ODS2Capcell Pak UG C18

Hypersil HyPURITY C18XTerra MS C18

Hypersil BDS C18Symmetry C18

Develosil ODS-HGPurospher RP18-e

Kromasil C18Nucleosil C18AB

YMC ODS AVydac Selectapore 90M

Zorbax XDB-C18YMC ODS AM

Zorbax Extend C18Vydac Selectapore 300P

Vydac 218MSYMC J'Sphere ODS H80

Exsil ODS1Zorbax SB-C18

Inertsil ODS2Inertsil ODSExsil ODSB

Novapak C18TSK ODS-80TM

Waters Spherisorb ODSBTSK ODS-120T

Vydac 218TPLiChrospher RP-18

Waters Spherisorb ODS2Partisil ODS

Partisil ODS3LiChrosorb RP-18

µBondapak C18Nucleosil C18

Zorbax Rx-C18Ultrasphere ODS

Capcell Pak AG C18Exsil ODS

Hypersil ODSPartisil ODS2Resolve C18

Waters Spherisorb ODS1Zorbax ODS

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000

Figure 13Comparison of Column Efficiency for a Basic Compound: Amitriptyline

Figure 14Top 10 Columns Ranked According to Peak Shape and Efficiency

Sample: AmitriptylineMobile Phase: 80% CH3OH 20% 0.025M Phosphate, pH 6.0

Sample: PyridineMobile Phase: 60% CH3OH 40% H2O

Note: All columns are packed with 5 micron size particles.

Column ranking does differ in the two tests of column effi-ciency for a basic compound (Figures 10 and 13). However, of the 14 columns that ranked in the top 10 on at least one of the tests, 6 ranked in the top 10 on both tests.

Plate count is measured at 10% of peak height to include peak tailing in the calculation. Both tests use mobile phases at neutral pH to encour-age interaction between the basic probes and silanols on the stationary phase.

Mobile Phase: 80% CH3OH 20% 0.025M KH2PO4, pH 6.0Sample: AmitriptylineTemperature: 24°C

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000

ACE C18-300ACE C18

Hypersil GOLDYMC J'Sphere ODS M80

Gemini C18ACE C18-HL

YMC Pro C18SunFire C18

Develosil ODS-UGLuna C18(2)

ACE C18Develosil ODS-UG

YMC Pro C18ACE C18-300SunFire C18

Hypersil GOLDDevelosil ODS-MG

Hichrom RPBZorbax XDB-C18

Develosil ODS-HG

N/Meter (Amitriptyline)

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000

N/Meter (Pyridine)

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000

ACE C18-300ACE C18

Hypersil GOLDYMC J'Sphere ODS M80

Gemini C18ACE C18-HL

YMC Pro C18SunFire C18

Develosil ODS-UGLuna C18(2)

ACE C18Develosil ODS-UG

YMC Pro C18ACE C18-300SunFire C18

Hypersil GOLDDevelosil ODS-MG

Hichrom RPBZorbax XDB-C18

Develosil ODS-HG

N/Meter (Amitriptyline)

0 10,000 20,000 30,000 40,000 50,000 60,000 70,000

N/Meter (Pyridine)

10 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 13: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Figure 15Comparison of Column Efficiency For a Basic Compound: Amitriptyline

ACE C18-300

65,300pl/m

ACE C18

63,300pl/m

Gemini C18

48,500pl/m

SunFire C18

45,000pl/m

Luna C18(2)

43,200pl/m

Inertsil ODS3V

30,500pl/m

Hypersil HyPURITY C18

24,800pl/m

XTerra MS C18

24,400pl/m

Hypersil BDS C18

22,700pl/m

Symmetry C18

21,800pl/m

Zorbax SB-C18

4,900pl/m

Waters Spherisorb ODS2

1,300pl/m

Column efficiency measured at 10% amitriptyline peak height to account for peak tailing effects

Column Dimensions: 250 x 4.6mm, 5µm Sample: 1) norephedrine 2) nortriptyline 3) toluene 4) imipramine 5) amitriptylineMobile Phase: 80:20 MeOH/0.025M KH2PO4, pH 6.0Temperature: 24°CFlow: 1.0ml/min

Comparison Guide to C18 Reversed Phase HPLC Columns | 11

Page 14: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Peak Asymmetry

HigherSilanolActivity

LowerSilanolActivity

1.03

8.998.3

6.503.19

3.073.053.05

2.742.21

2.071.93

1.881.78

1.611.52

1.261.12

ACE C18XTerra MS C18

Inertsil ODS3Luna C18(2)

Hypersil Elite C18Discovery C18

Waters Spherisorb ODS1Inertsil ODS2

Symmetry C18Hypersil BDS-C18

Partisil ODS2Zorbax XDB-C18

Novapak C18Zorbax SB-C18

Nucleosil C18Vydac Selectapore 300M

Cosmosil C18 AR-IIµBondapak C18

1 2 4 6 8 10

Peak Asymmetry

HigherMetalActivity

LowerMetalActivity

1.07

3.072.56

2.422.41

2.271.46

1.431.37

1.341.31

1.271.26

1.241.16

1.121.09

1.07ACE C18

Zorbax XDB-C18Hypersil Elite C18

XTerra MS C18Cosmosil C18 AR-II

Inertsil ODS3Hypersil BDS-C18

Inertsil ODS2Symmetry C18

Vydac Selectapore 300MDiscovery C18

Zorbax SB-C18Luna C18(2)

Waters Spherisorb ODS1Nucleosil C18Novapak C18

µBondapak C18Partisil ODS2

1.0 1.5 2.0 2.5 3.0 3.5

Figure 16Comparison of Metal Activity Using the NIST Test: Asymmetry for Quinizarin

Mobile Phase: 80% CH3OH 20% 5mM Potassium Phosphate, pH 7.0Sample: QuinizarinFlow Rate: 2 mL/minTemperature: 23°C

Figure 17Comparison of Silanol Activity Using the NIST Test: Asymmetry for Amitriptyline

Mobile Phase: 80% CH3OH 20% 5mM Potassium Phosphate, pH 7.0Sample: AmitriptylineFlow Rate: 2 mL/minTemperature: 23°C

Additional Comparison Tests

There are numerous suggestions from different scientific groups about how to best characterize stationary phases. Most of these tests have merit, but the fact that the ranking of columns will often differ among the different tests shows the difficulty in devising a definitive test that will predict column behavior in all, or even most circumstances. The National Institute of Standards & Technology (NIST) has developed test conditions (Standard Reference Material 870) that do a particularly good job of characterizing stationary phases according to metal activity and silanol activity. The presence of metals on the surface of stationary phases can have a significant effect on chromatographic performance. Even trace levels of metal impurities can contribute to peak tailing of some compounds. In addition, subtle lot-to-lot variations in the amount of trace metals are another cause of poor column reproducibility. The NIST test uses peak asymmetry of quinizarin, a strong metal chelating agent, to measure metal activity. Figure 16 ranks stationary phases according to metal activity using the NIST test.To test silanol activity, the NIST test uses amitriptyline, as does the test used to generate the data in Figure 13. However, the NIST test specifies a mobile phase pH of 7.0 rather than 6.0, and measures peak asymmetry rather than plate count to determine silanol activity. The lower the asymmetry value of the amitriptyline peak (less tailing) the less silanol activity. Figure 17 ranks stationary phases according to silanol activity using the NIST test.

12 | Comparison Guide to C18 Reversed Phase HPLC Columns

Page 15: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

Figure 18Grouping of C18 Columns According to Silanol Activity

Material

ACE C18 ACE C18-300 Develosil ODS-MG Hypersil GOLD Hypersil HyPURITY C18 Inertsil ODS3 Luna C18(2) Nucleosil C18 HD SunFire C18 XTerra MS C18 YMC Pro C18

ACE C18-HL Capcell Pak UG C18 Develosil ODS-HG Develosil ODS-UG Gemini C18 Hichrom RPB Inertsil ODS2 Kromasil C18 Prodigy ODS2 Prodigy ODS3 Purospher RP18-e Symmetry C18 YMC ODS A YMC ODS AM Zorbax Extend C18 Zorbax XDB-C18

Capcell Pak C18 SG Exsil ODSB Hypersil BDS C18 Inertsil ODS Nova-Pak C18 Nucleosil C18AB Partisil ODS3 Synchropak CR101 TSK ODS-120T TSK ODS-80TM µBondapak C18 Vydac 218MS Vydac 218TP Vydac Selectapore 300M\ Vydac Selectapore 300P Vydac Selectapore 90M Waters Spherisorb ODSB YMC J’Sphere ODS H80 YMC J’Sphere ODS M80 Zorbax Rx-C18 Zorbax SB-C18

Capcell Pak C18 AG Exsil ODS Exsil ODS1 Hypersil ODS LiChrosorb RP-18 LiChrospher RP-18 Nucleosil C18 Partisil ODS Partisil ODS2 Resolve C18Ultrasphere ODS Waters Spherisorb ODS1 Waters Spherisorb ODS2 Zorbax ODS

Phases Grouped According to Silanol Activity

Amitriptyline and pyridine are both good test probes to use for measur-ing silanol activity of stationary phases. Even a small amount of silanol exposure by the stationary phase can cause measurable peak broadening and peak asymmetry on one or both of these compounds. Chromato-graphic tests using these two probes are the primary measurements used to group these C18 phases according to silanol activity. In general, phases identified as having “very low” silanol activity will give the highest column efficiency in the pyridine and amitriptyline tests (Figures 10, 13 and 17).

SILA

NO

L A

CTI

VIT

Y

Very

Low

Low

Mod

erat

eH

igh

Comparison Guide to C18 Reversed Phase HPLC Columns | 13

Page 16: Comparison Guide to C18 Reversed Phase HPLC Columns...Comparison Guide to C18 Reversed Phase HPLC Columns Introduction There are so many different C18 columns to choose from, that

MAC-MOD Analytical 103 Commons Court P.O. Box 587 Chadds Ford, PA 19317

Phone: 1.800.441.7508 Fax: 1.610.358.5993 E-mail: [email protected] Web Site: www.mac-mod.com

LC497