ALARM VALVE COMBO UNIT COMBINATION ALARMS 2 in 1 design combines the Area Alarm and Zone Valve Box to compensate for space restrictions True digital, illuminated LED display is readable even in poor lighting conditions High/low alarm set-points are field adjustable for each gas service Repeat alarm is adjustable from 1 to 60 minutes or off Gas-specific sensor with DISS nut and nipple Alarm buzzer in excess of 90 decibels Each module is marked with an approved medical gas identification label Pre-mounted pull-out ring allows for ease of maintenance Individual microprocessor for each display and sensor module; digital sensor is mounted locally Dry contacts for remote monitoring of the high and low alarm ORDERING INFORMATION A20AVEXGXGXGXGXGXGXG NOTE: The power supply will require 1 gang per box. Example: English, 3 valves, 1/2" [1.3 cm] oxygen, 3/4" [1.9 cm] medical air and 1-1/4" [3.18 cm] medvac. Will be in a 4 gang box. AVU- 1O2A4V The 1-1/2" [3.8 cm] valve will have an empty space above the valve. Example: A box with a 1/2" [1.3 cm] oxygen, 3/4" [1.9 cm] medical air and a 1-1/2" [3.8 cm] medvac valves will require a 5 hole box not a 3 hole box. The 2" [5.1 cm] valve will be assembled in a 6-1/2" [16.5 cm] deep box and shall accommodate up to a maximum of 4 gases. The 2" [5.1 cm] valve shall be positioned on the bottom of the assembly. Refer to 2" [5.1 cm] Valve Drawing provided on page 2. The height measurements of the 6-1/2" [16.5 cm] valve box are the same as the 3-7/8" [9.84 cm] valve box. When comparing height measurements, consider the gap above the 2" [5.1 cm] valve. Example: A valve box with 2" [5.1 cm], 1/2" [1.3 cm] and a power supply would be in a 3 gang 6-1/2" [16.5 cm] deep box, which would be the same height as the 4 gang 3-7/8" [9.84 cm] box. DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com G Defines the Type of Gas O Oxygen A Medical Air V Medical Vacuum 2 Nitrous Oxide N Nitrogen C Carbon Dioxide W WAGD (NFPA) E AGSS (ISO) I Instrument Air E – English (CSA/ISO) X Defines the Valve Size 1 ½” (1.3 cm) 05 2 ¾” (1.9 cm) 07 3 1” (2.5 cm) 10 4 1¼” (3.2 cm) 12 5 1½” (3.8 cm) 15 6 2” Full Port (5.1 cm) 20F

Transcript of COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass...

Page 1: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ALARM VALVE

COMBO UNIT

COMBINATION ALARMS • 2 in 1 design combines the Area Alarm and Zone Valve Box to

compensate for space restrictions • True digital, illuminated LED display is readable even in poor

lighting conditions • High/low alarm set-points are field adjustable for each gas service • Repeat alarm is adjustable from 1 to 60 minutes or off • Gas-specific sensor with DISS nut and nipple • Alarm buzzer in excess of 90 decibels • Each module is marked with an approved medical gas

identification label • Pre-mounted pull-out ring allows for ease of maintenance • Individual microprocessor for each display and sensor module;

digital sensor is mounted locally • Dry contacts for remote monitoring of the high and low alarm

ORDERING INFORMATION A20AVEXGXGXGXGXGXGXG

NOTE: The power supply will require 1 gang per box. Example: English, 3 valves, 1/2" [1.3 cm] oxygen, 3/4" [1.9 cm] medical air and 1-1/4" [3.18 cm] medvac. Will be in a 4 gang box. AVU-1O2A4V The 1-1/2" [3.8 cm] valve will have an empty space above the valve. Example: A box with a 1/2" [1.3 cm] oxygen, 3/4" [1.9 cm] medical air and a 1-1/2" [3.8 cm] medvac valves will require a 5 hole box not a 3 hole box. The 2" [5.1 cm] valve will be assembled in a 6-1/2" [16.5 cm] deep box and shall accommodate up to a maximum of 4 gases. The 2" [5.1 cm] valve shall be positioned on the bottom of the assembly. Refer to 2" [5.1 cm] Valve Drawing provided on page 2. The height measurements of the 6-1/2" [16.5 cm] valve box are the same as the 3-7/8" [9.84 cm] valve box. When comparing height measurements, consider the gap above the 2" [5.1 cm] valve. Example: A valve box with 2" [5.1 cm], 1/2" [1.3 cm] and a power supply would be in a 3 gang 6-1/2" [16.5 cm] deep box, which would

be the same height as the 4 gang 3-7/8" [9.84 cm] box.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

G Defines the Type of Gas

O Oxygen A Medical Air V Medical Vacuum 2 Nitrous Oxide N Nitrogen C Carbon Dioxide W WAGD (NFPA) E AGSS (ISO) I Instrument Air

E – English (CSA/ISO)

X Defines the Valve Size

1 ½” (1.3 cm) 05

2 ¾” (1.9 cm) 07

3 1” (2.5 cm) 10

4 1¼” (3.2 cm) 12

5 1½” (3.8 cm) 15

6 2” Full Port (5.1 cm) 20F

Page 2: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GENERAL SPECIFICATIONS

The Alarm Valve Combo Unit shall be an Amico Alert-1 Series.

Each AVC shall consist of the following components: an 18 Gauge steel valve box complete with a baked white enamel finish which can house one to six shut-off ball valves with tube extensions, gas-specific sensor (DISS nut and nipple connection), a hinged gas-specific compact alarm with illuminated LED digital display and an error message for an incorrect connection, an aluminum frame and a pull-out removable window.

Affixed to the opposing sides of the box will be two adjustable steel brackets for the purpose of mounting the box to the structural support. The steel brackets shall accommodate various finished wall thicknesses between 3/8" [9.5 mm] and 1-3/16" [30 mm] and shall be field adjustable. The frame assembly shall be constructed of anodized aluminum and shall be mounted to the back box assembly by standard number #6 x 3/8" tapping screws as provided.

The digital alarm shall read from 0-250 psi [0-1,724 kPa] for pressure and 0-30" Hg [-100-0 HgkPa] for vacuum. The digital read-out shall provide a constant indication of each service being measured. It will indicate a green ”NORMAL” and a red ”HIGH” or ”LOW” alarm condition. If an alarm occurs, the ”RED” alarm light shall flash and the audible alarm (which exceeds 90 decibels) will sound. Pushing the ”ALARM MUTE” button will cancel the audible alarm but the unit will remain in the alarm condition until the problem is rectified. A repeat alarm function shall, when enabled on the compact alarm module, be capable of turning on the buzzer again (after a preset time) if the fault condition has not been rectified.

Each alarm module shall be individually microprocessor based and be field adjustable. The default set point on this alarm shall be +/- 20% variation from normal condition. In the calibration mode the following parameters shall be adjustable: high/low set-points, imperial/metric units, repeat alarm enable/ disable (adjustable from 1 to 60 minutes) as well as psi, kPa or Bar readout (switch selected). Set points shall be adjustable by two on-board push buttons.

Access to the shut-off valves shall be by merely pulling the ring assembly to remove the window from the frame. The window can be reinstalled without the use of tools only after the valve handles have been returned to the open position. The window shall be marked to prohibit unauthorized persons from tampering with the valves with the following silk-screen caution:

”Medical Gas Control Valves with Alarms” ”Close Valves Only in Emergency”

The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm]. Seats shall be reinforced Teflon (PTFE) and seals Viton for 1/2" to 2" [1.3 cm to 5.1 cm]. A blow-out proof stem shall be used and the valve shall have a maximum pressure rating of 600 psi [4,137 kPa].

Valves shall be operated by a lever-type handle requiring only a quarter turn from a fully open position to a fully closed position. All valves shall be equipped with a type ”K” washed and degreased copper pipe stub. Each valve will be identified for gas specification as indicated on the hinged alarm label.

Input power to the Amico Alarm Valve Combo Unit is: 115 VAC to 220 VAC, 50 to 60 Hz.

Amico products comply with NFPA 99 and CSA Z7396.1.

The Amico Alert-1 Alarm complies with the following electromagnetic compatibilty standards: FCC Part 15 Class A

and ICES-003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 3: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 4: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ALERT- 3 LCD ALARM

ALERT-3 SERIES • Highly visible 10" [25.4 cm] LCD Screen • Automatic Gas Detection • Industry’s Smallest 8 Gas Alarm • Clear Display of Alarm Condition • Self Diagnostic and Error Message

Display • Upgrade existing alarms via the simple

“Retro-fit,” with no need to break walls or cut pipes

• Quality and design will bring you years of accuracy and reliability

GENERAL SPECIFICATIONS The Digital LCD Area Alarm system shall be an Amico Alert-3 series, complete with a five-year warranty. The LCD alarm shall be microprocessor based with a 10" [25.4 cm] screen and capable of monitoring up to 8 sensors. Sensors shall be mounted locally (in the rough-in box) by installing the copper pipe provided or mounted remotely. Sensors will be automated for gas specific detection. Each sensor shall be gas specific and an error message shall be displayed for an incorrect connection. Each specific gas service shall have a digital read-out comprising of 0-249 psi [0-1,717 kPa] for pressure and 0-30 inHg [-100-0 kPa] for vacuum. The digital read-out shall provide a continuous indication of each service being measured. A separate indicator shall be provided for each service indicating a green “NORMAL” and a red “HIGH” or “LOW” alarm condition. If an alarm occurs, the green indicator will change to red and a continuous audible alarm will sound. Pushing the (mute button/ push to test button) will cancel the audible alarm, but the unit will remain in the alarm condition until the problem is rectified. The default set-points shall be +/- 20% variation from normal condition. In the calibration mode High/Low set-points shall be adjustable by on board push buttons. To view the set points and audible alarm sound level, press and hold the mute button for twenty (20) seconds. The box shall be fabricated from 18 gauge [1.3 mm] steel with a 3/8" [9.53 mm] O.D. type “K” copper pipe for connection to the service line. The box mounting brackets shall be adjustable to accommodate for different wall thicknesses. LCD Sensor operating pressure range: Mid-Pressure: (0 psi to 99 psi) - OXY, AIR, N₂O, CO₂ High-Pressure: (0 psi to 249 psi) - NIT, IAR Vacuum: (0 inHg to 30 inHg) - VAC, WAGD, AGSS Input power to the Amico Alert-3 alarm is: 115-220 VAC, 50-60 HZ. The Amico Alert-3 LCD Alarm is cETLus listed to UL standard 1069 and CSA standard C22.2 No. 205. Amico products comply with NFPA 99 , CSA Z7396.1 and CSA Z305.1. The Amico Alert-3 LCD Alarm complies with the following electromagnetic compatibility standards: FCC Part 15 Class A and

ICES-003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 5: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20A3AREXXXXXXXX

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

X Defines the Type of Gas

O Oxygen A Medical Air V Medical Vacuum 2 Nitrous Oxide N Nitrogen C Carbon Dioxide W Waste Anesthetic Gas Disposal E AGSS (ISO) I Instrument Air

E – English (CSA/ISO)

R Remote / Local

C Conversion (Retrofit)

Page 6: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ALERT-4 LCD

ETHERNET AREA ALARM

ALERT-4 SERIES

GENERAL SPECIFICATIONS

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 7: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 8: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

BALL VALVE WITH EXTENSIONS

AND DUAL PURGE PORT

ISOLATION VALVE • 3 piece design for ease of maintenance

• 100% Pressure tested

• Cleaned for medical gas service

• Blow-out proof stem

• Available in sizes 1/2" to 4"

• Full-Port design

• Dual port design

• Available in locking or non-locking

GENERAL SPECIFICATIONS

All purge port isolation valves shall be an Amico Alert-1 Series. The Valve shall be 3 piece ball-type design with a bronze body and chrome plated brass ball for sizes 1/2” to 2-1/2”, for sizes 3” - 4” the ball is stainless steel. Seats shall be Teflon (TFE) and seals Viton for 1/2” to 2-1/2”and Teflon (TFE) for 3”- 4” valves. A blow-out proof stem shall be used and the valve shall have a maximum pressure rating of 600 psi [4,137 kPa]. Valves shall be operated by a lever-type handle requiring only a quarter turn from a fully open position to a fully closed position. Ball valves shall be equipped with type “K” copper pipe stub extensions at both the inlet and outlet sides of the valve port to facilitate installation. Valves shall be designed in such a manner that it can be “swung-out” during installation so as to prevent damage due to heat transfer during the brazing operation. Each valve assembly shall be washed and degreased for medical gas service. Pipe stub extensions shall be capped at both ends. The valve shall be supplied in a sealed plastic bag to prevent contamination prior to installation. Amico products comply with NFPA-99, CSA Z7396.1-06 and CSA Z305.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 9: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION

MODEL# SIZE A B C D E F G H BOLTS

A20VVISOG2L05 1/2 (13)

2-1/16 (52)

1-13/16 (46)

1/2 (13)

2-1/32 (52)

3-29/32 (99)

1-1/16 (27)

7-1/4 (184)

11-3/4 (299)

4

A20VVISOG2L07 3/4 (19)

2-15/16 (75)

1-15/16 (49)

3/4 (19)

2-9/32 (58)

4-21/32 (118)

1-7/16 (37)

6-3/4 (172)

11-1/2 (292)

4

A20VVISOG2L10 1 (25)

3-21/32 (93)

2-1/2 (64)

1 (25)

2-1/2 (64)

4-21/32 (118)

1-27/32 (47)

6 (152)

11-1/4 (286)

4

A20VVISOG2L12 1-1/4 (32)

3-29/32 (99)

2-11/16 (68)

1-1/4 (32)

3-1/8 (79)

6-11/16 (170)

1-31/32 (50)

5-5/8 (143)

11-1/4 (286)

4

A20VVISOG2L15 1-1/2 (38)

4-15/32 (117)

3 (76)

1-1/2 (38)

3-11/32 (85)

6-11/16 (170)

2-13/32 (61)

5-1/4 (133)

11-1/8 (283)

4

A20VVISOG2L20 2 (51)

5-25/32 (147)

4 (102)

2 (51)

3-11/16 (94)

6-11/16 (170)

3-3/32 (79)

5 (127)

9-1/8 (232)

4

A20VVISOG2L25 2-1/2 (64)

6-15/16 (176)

5 (127)

2-1/2 (64)

4-11/16 (119)

8 (203)

4 (102)

12 (305)

12 (305)

4

A20VVISOG2L30 3 (76)

8-1/2 (216)

7-11/16 (195)

3 (76)

5-9/16 (141)

18 (457)

5-3/16 (132)

11-7/8 (302)

11-7/8 (302)

6

A20VVISOG2L40 4 (102)

11 (279)

10-1/4 (260)

4 (102)

6-3/4 (172)

18 (457)

6-11/16 (170)

11 (279.4)

11 (279.4)

6

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 10: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

COMBINATION ALARM SYSTEMS

COMPACT/MASTER

COMBINATION ALARMS • Individual Microprocessor for each display and sensor module • Digital sensors can be mounted locally or remotely utilizing a #22 gauge stranded, shielded,

twisted pair cable ONLY • True digital LED read-out with RED alarm and GREEN NORMAL trend indication for each service • Illuminated LED display readable even in poor lighting conditions • Self-diagnostic and error message display for ease of maintenance • HIGH/LOW Alarm set-points are field adjustable for each individual gas service • Alarm buzzer in excess of 90 decibels • A green LED indicator for normal and a RED LED indicator for abnormal conditions • Maintenance mode for ease of troubleshooting • Self-diagnostic circuitry for added reliability • Master modules can be upgraded in the field to interface to a building management system or

slave master alarm • Last alarm flashes, acknowledged alarm shows continuous RED signal • Repeat alarm adjustable from 1, 12, 24 hours or off • Dry contacts are used between master module and source equipment for remote monitoring of

high and low alarms • Sensors housed in a solid, tamper-proof enclosure to act as a barrier against any interference • Gas specific sensor with DISS nut and nipple

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 11: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GENERAL SPECIFICATIONS

Both the Compact and Master Alarm systems shall be an Amico Alert 2 series, complete with a five-year warranty. The Compact Alarm shall be microprocessor based with individual microprocessors on each display, sensor, and master boards. The sensors shall be capable of local (in box) or remote mounting. Each sensor and display unit shall be gas-specific, with an error message display for an incorrect connection. The Compact Alarms shall be of modular construction and shall be field expandable with the addition of extra modules. Up to 12 services can be accommodated per standard box. Each specific service shall be provided with an LED digital read-out comprising of 0-250 psi [1,724 kPa] for pressure and 0-30" Hg [-100-0 kPa] for vacuum. The digital read-out shall provide a constant indication of each service being measured. A bar graph trend indicator shall be provided for each service indicating a GREEN NORMAL and a RED HIGH or LOW alarm condition. If an alarm occurs, the RED alarm led shall flash and the audible alarm will sound. Pushing the (mute button/push to test button) will cancel the audible alarm, but the unit will remain in the alarm condition until the problem is rectified. The default set-points shall be +/-20% variation from normal conditions. In the calibration mode, the following parameters shall be field adjustable: high/low set-points, imperial/metric units, repeat alarm enable/disable. Set-points shall be adjustable by two on board push buttons. Each Master module shall be microprocessor based and field adjustable. Each signal contains a GREEN LED indicator for normal and a RED LED indicator for abnormal conditions. When a fault occurs, the green LED turns off, the RED LED illuminates and the audible alarm sounds. The RED LED flashes until the front panel alarm mute button is pressed. After the alarm mute button is pressed, the RED LED remains illuminated but does not flash. The red indicator automatically turns off and the GREEN LED illuminates when the fault is corrected. A repeat alarm function shall, when enabled, be capable of turning on the buzzer again after a preset time if the fault condition has not been rectified. A maintenance mode shall, when enabled, latch the alarms, requiring a reset after the alarm condition has been rectified. This is to assist in tracking down wiring problems or faulty field devices. The master alarm shall identify the last alarm condition by flashing, while the already acknowledged alarm shows a continuous red signal. Each module shall handle 10 functions and up to 6 modules can be accommodated per standard box for a total of 60 functions. Master alarms shall be modular in construction and shall be capable of adding extra modules in the field. The Alarm system shall be a closed circuit self-monitoring type. A GREEN POWER light shall provide indication that the unit is energized. In addition, TEST and ALARM MUTE buttons shall be easily accessible to operate and test the unit. Every master module shall be field upgradable to allow for interfacing to a building management system with the addition of an add-on circuit board which plugs into the master module. The frame assembly can be added to an existing alarm panel in the field with the addition of an extended mounting frame. The mounting frame is made from 18 gauge (1.3 mm) aluminum that has been painted for aesthetic appeal. Input power to the Amico Alert-2 alarm is: 115 VAC - 220 VAC, 50 - 60 Hz. Amico products comply with NFPA 99, CSA Z7396.1 and CSA Z305.1. The Amico Alert-2 Alarm complies with the following electromagnetic compatabilty standards: FCC Part 15

Class A and ICES-003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 12: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 13: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ORDERING INFORMATION A20A2DREXXXXXXMMM

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

X Defines the Type of Gas

O Oxygen A Medical Air V Medical Vacuum 2 Nitrous Oxide N Nitrogen C Carbon Dioxide W WAGD (NFPA) E AGSS (ISO) I Instrument Air

R Remote/Local Sensors

C Conversion

E – English (CSA/ISO)

Each “M’ Stands for 10 Functions

3 Modules

A20A2DREXXM

4 gang back box

4 Modules

A20A2DREXXMM

4 gang back box

5 Modules

A20A2DREXXXXM

4 gang back box

6 Modules

A20A2DREXXXXMM

7 gang back box

8 Modules

A20A2DREXXXXXXMM

7 gang back box

9 Modules

A20A2DREXXXXXXMMM

7 gang back box

“X” indicates the order of the gases, as shown to the left

Example OAV (position 1 OXY, position 2 AIR, Position 3 VAC)

Page 14: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

COMPACT ALARM

CONVERSION (RETRO FIT)

RETRO-FIT ALARMS • Individual Microprocessor for each display and

sensor module • Digital sensors can be mounted locally or

remotely utilizing a #22 gauge stranded, shielded, twisted pair cable ONLY

• True digital LED read-out with red “alarm” and green “normal” trend indication for each service

• Psi, kPa or Bar read-out (switch selected) • Illuminated LED display readable even in poor

lighting conditions • Self-diagnostic and error message display for ease of maintenance • High / Low Alarm set-points are field adjustable for each individual gas service • Master Alarm modules can be intermixed with Compact Alarm modules • Repeat alarm is adjustable from 1 to 60 minutes or off • Dry contacts for the remote monitoring of high and low alarms • Sensors housed in a solid, tamper-proof enclosure to act as a barrier against any interference • Ability to use existing quick-connect from another alarm panel located in the field • All modules are mounted on a hinged frame for easy accessibility • Alarm buzzer in excess of 90 decibels

GENERAL SPECIFICATIONS The Digital Compact Alarm Retro-fit kit shall be an Amico Alert-2 series, complete with a five-year warranty. The alarm shall be microprocessor based with individual microprocessors on each display and sensor board. The sensors shall be capable of local (in box) or remote mounting. Each sensor and display unit shall be gas-specific, with an error message display for an incorrect connection. The compact alarms shall be of modular construction and shall be field expandable with the addition of extra modules. Up to 12 services can be accommodated per standard box. Each specific service shall be provided with an LED digital read-out comprising of 0-250 psi [1,724 kPa] for pressure and 0-30"Hg [-100-0 kPa] for Vacuum. The digital read-out shall provide a constant indication of each service being measured. A bar graph trend indicator shall be provided for each service indicating a green “NORMAL”, and a red “HIGH” or “LOW” alarm condition. If an alarm occurs, the “RED” alarm led shall flash and the audible alarm will sound. Pushing the “MUTE” button will cancel the audible alarm, but the unit will remain in the alarm condition until the problem is rectified. The default set-points shall be +/-20% variation from normal conditions. In the calibration mode, the following parameters shall be field adjustable: high/low set-points, imperial/metric units, repeat alarm enable / disable. Set-points shall be adjustable by two on board push buttons. In addition, the “TEST” & “MUTE” buttons shall be easily accessible to operate and test the unit. The frame assembly can be added to an existing alarm panel in the field with the addition of an extended mounting frame. The mounting frame is made from 18 gauge aluminum that has been painted for aesthetic appeal. Input power to the Amico Alert-2 alarm is: 115 VAC to 220 VAC, 50 to 60 Hz. The Alarm is UL Listed to U.S. and Canadian safety standards. Amico products comply with NFPA 99, CSA Z7396.1 and CSA Z305.1. The Amico Alert-2 Alarm complies with the following electromagnetic compatibilty standards: FCC Part 15 Class A and ICES-003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 15: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20A2DCEXXXXXXXXXXXX

NOTE: Please specify what you are converting from (digital or analog), indicate if a special fitting is required for the sensor. Only conversions from Ohio alarms require a trim plate. Example: 4 gas, English, NFPA, Ohio conversion with Oxygen, Medical Air, Medvac, and Nitrous Oxide (converted from Ohio digital, 4 gang filler box) = A20A2DCEOAV2 2

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

X Defines the Type of Gas

O Oxygen C Carbon Dioxide A Medical Air W WAGD (NFPA) V Medical Vacuum E AGSS (ISO) 2 Nitrous Oxide I Instrument Air N Nitrogen

E – English (CSA/ISO)

C Conversion (Retrofit)

Page 16: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

2 module = A20A2DCEXX 4 gang box

4 modules = A20A2DCEXXXX 4 gang back box

6 modules = A20A2DCE-XXXXXX 4 gang back box

8 modules = A20A2DCEXXXXXXXX 7 gang box

10 modules = A20A2DCEXXXXXXXXXX 7 gang back box

12 modules = A20A2DCE-XXXXXXXXXXXX 7 gang back box

“X’’ indicates the order of gases, as shown above. Example: OAV (position 1 OXY, position 2 AIR, position 3 VAC) IMPORTANT NOTE: Due to unknown factors of the existing equipment, projects requiring Retro-Fit Conversions must be fully investigated by the installer. It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are required.

Page 17: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

COMPACT

DISS WALL OUTLET

WORRY FREE AND DEPENDABLE • Flow rate exceeds the requirements of NFPA and CSA • Accepts only DISS gas specific adaptors • 100% Pressure testing of each outlet and cleaned for

medical gas service • Indexed to eliminate interchangeability of gas services • Up to 3/4" (19 mm) mounting plate thickness adjustment • 360° Swivel Inlet Pipe for easy installation • Universal Rough-In Assembly to accept Quick Disconnects

or DISS front adaptors (can be interchanged at any time) • Outlets are also available for M.R.I. use • Aesthetically pleasing in any patient room, this DISS

connection Amico gas outlet provides a full color fascia for easy gas identification during those critical moments

GENERAL SPECIFICATIONS The medical gas outlet shall be an Amico Alert-1 series. Outlets shall be manufactured with a 7-3/4" (197 mm) length type “K” 1/2" (12.7 mm) outside diameter (3/8" [9.5 mm] nominal) size copper inlet pipe stub which is silver brazed to the outlet body. The body shall be of 1-5/16" (33 mm) diameter, one piece brass construction. For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall be rated at a maximum of 200 psi (1,379 kPa) in the event the primary check valve is removed for maintenance. Outlet bodies shall be gas specific by indexing each gas service to a gas specific pin indexing arrangement on the respective identification module. A color coded front shall be used for ease of gas identification and aesthetic appeal. With the back rough-in mounted, the outlet shall adjust up to 3/4" (19 mm) variation in mounting plate thickness. The latch-valve assembly shall be DISS type and only accept corresponding DISS type gas specific adapters. All outlets shall be cleaned and degreased for medical gas service, factory assembled and tested. The outlets are UL listed. Amico products comply with NFPA 99.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 18: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20OCDISWALEXXX

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

XXX Defines the Gas

OXY Oxygen AIR Medical Air N2O Nitrous Oxide CO2 Carbon Dioxide VAC MedVac WAG Waste Anesthetic Gas Disposal INA Instrument Air NIT Nitrogen

E – English (CSA/ISO)

Page 19: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

DIGITAL VACCHECK

STATE OF THE ART DESIGN, COMPACT, VERSATILE

• Digital VacCheck gauge incorporates digital microprocessor technology – solid state design with no moving parts

• Small enough to carry in a pocket for easy vacuum level accuracy checks wherever required

• Made of rigid ABS shock-resistant plastic case with aluminum vacuum hose barb port connection

• Digital display from 0 to 760 mmHg in 1 mmHg increments

• Anodized aluminum ¼” hose barb connector attaches to standard vacuum tubing

• No user interface required, unit automatically activates when a vacuum source is detected

ITEM DESCRIPTION P70PM760 Digital Vac Check Unit

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

PRODUCT SPECIFICATIONS Dimensions Width: 1.63”, Height: 2.13”, Depth: 1.00” Vacuum Range 0-760 mmHg

Accuracy +/- 1% over full scale range

Connector ¼” hose barb Battery Lithium Ion (8 to 10 year life)

Weight 0.15 lb Warranty 1 year

Oxygen Cylinder Connection CGA 870 Post Valve

Page 20: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

DISS WALL OUTLET

WORRY FREE AND DEPENDABLE • Flow rate exceeds the requirements of NFPA and CSA • Accepts only DISS gas specific adaptors • 100% Pressure testing of each outlet and cleaned for

medical gas service • Indexed to eliminate interchangeability of gas services • Up to 1" (25 mm) wall thickness adjustment • 360° Swivel Inlet Pipe for easy installation • Universal Rough-In Assembly to accept Quick Disconnects

or DISS front adaptors (can be interchanged at any time) • Outlets are also available for M.R.I. use • Aesthetically pleasing in any patient room, this DISS

connection Amico gas outlet provides a full color fascia for easy gas identification during those critical moments

• Five Year Warranty

GENERAL SPECIFICATIONS The medical gas outlet shall be an Amico Alert-1 series. Outlets shall be manufactured with a 7-3/4" (197 mm) length type ”K” 1/2" (12.7 mm) outside diameter (3/8" [9.5 mm] nominal) size copper inlet pipe stub, which is silver brazed to the outlet body. The body shall be of 1-5/16" (33 mm) diameter, one piece brass construction. For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall be rated at a maximum of 200 psi (1,379 kPa) in the event the primary check valve is removed for maintenance. Outlet bodies shall be gas specific by indexing each gas service to a gas specific dual pin indexing arrangement on the respective identification module. A large color coded front plate shall be used for ease of gas identification and aesthetic appeal. A one piece chromed fascia plate shall frame the outlet. With the back rough-in mounted, the outlet shall adjust from 3/8" (10 mm) to 1" (25 mm) variation in wall thickness. The outlets shall be of modular design and include a gas specific 1.6 mm (16 ga.) steel mounting plate designed to permit on-site ganging of multiple outlets, in any order, on 5" (127 mm) spacing. The latch-valve assembly shall be D.I.S.S. (Diameter Index Safety System) and shall only accept corresponding gas specific adapters. All M.R.I. outlets are manufactured from nonferrous materials. All outlets shall be cleaned and degreased for medical gas service, factory assembled and tested. The outlets are UL Listed and ETL Listed. Amico products comply with NFPA 99 and CSA Z7396.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 21: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20ODISWALEXXX

MRI Model Number: A20ODISWALMEXXX

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

XXX Defines the Gas

OXY Oxygen AIR Medical Air N2O Nitrous Oxide CO2 Carbon Dioxide VAC MedVac WAG Waste Anesthetic Gas Disposal (NFPA) AGS Anesthetic Gas Scavenging System INA Instrument Air NIT Nitrogen

E – English (CSA/ISO)

Page 22: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GAS CONTROL

PANEL RETRO FIT

RETRO-FIT GAS CONTROL PANELS • Inlet and outlet pressure display

gauges in psi/kPa • Manual shut-off valve • Outlet supply pipe for remote

mounted outlets • 3/8" (9.5 mm) internal tubing for

high flow • High flow capacity regulator • Aluminum front simplifies cleaning • Maximum supply pressure of 250

psi (1,724 kPa) • Maximum delivery pressure:

DISS: AIR CO₂ = 80 psi NIT, IAIR = 200 psi HP (Safety Swing Coupling) = 250 psi

GENERAL SPECIFICATIONS The Gas Control Panel Retro-Fit kit allows you to change from Chemetron or Ohmeda to Amico Alert-1 series without breaking the pipeline or cutting into the existing wall. The control panel shall be supplied with a quarter turn shut-off ball valve, rated at no less than 300 psi (2,069 kPa). The unit shall have an inlet pressure indicator with a 0 to 400 psi (0 to 2,750 kPa) valve to indicate the inlet supply pressure. The pressure regulator shall be adjustable between 0 to 250 psi (0 to 1,724 kPa). The outlet controlled pressure shall be monitored by a pressure indicator with a range of 0 to 400 psi (0 to 2,750 kPa). The DISS outlet shall be a Diameter Index Safety System (D.I.S.S.) for Air or Nitrogen Service Outlet or “Safety Swing Coupling” for pressures above 200 psi. The outlets shall be used for connection to pneumatic surgical tools. The mounting frame is made from 18 gauge carbon steel that has been painted for aesthetic appeal. The unit shall be factory tested for the intended gas service. The Gas Control Panel is UL Listed to U.S. and Canadian safety standards. Amico products comply with NFPA-99 and CSA Z7396.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 23: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION

ITEM DESCRIPTION

A20NCONCERPB Nitrogen Control Panel Retro-Fit - Puritan - Bennett

A20NCONCERCH Nitrogen Control Panel Retro-Fit-Chemetron (English, relieving)

A20NCONCEROH Nitrogen Control Panel Retro-Fit-Ohmeda (English, relieving)

A20NCONCEHPROH Nitrogen Control Panel High Pressure Retro-Fit - Ohmeda - English, relieving (NFPA)

A20NXCONADAB25 High pressure adapter with 1/4" Hose barb (supplied with HP Gas Control Panel)

IMPORTANT NOTE: Due to unknown factors of the existing equipment, projects requiring Retro-Fit conversions must be fully investigated by the installer.

It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are required.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 24: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GAUGE FOR ZONE INDICATOR

PANEL AND ISOLATION VALVE

FEATURES • Size: 2" • Accuracy: +/- 1.5% of full scale. ASME B40.1

Grade B • Range: 0-30" Vac (-100 - 0 kPa)

0 - 100 psi (0 -700 kPa) 0 - 300 psi (0-2000 kPa)

• Temperature: Ambient: -13°F to 140°F • Media: Max. 140°F • Connection: 1/8" Male NPT Center Back • Bourdon Tube: Copper alloy • Movement: Copper alloy • Dial: White plastic with stop pin • Pointer: Black aluminum • Case: Black steel • Window: Snap-in acrylic • The gauges shall be an Amico Alert-1 series • The gauge shall be with dual scale psi and

kPa for pressure, and Hg and kPa for Vacuum

• Amico products comply with NFPA 99 and CSA Z7396.1

ORDERING INFORMATION

ITEM DESCRIPTION A20VXGAUGEVAC Gauge for Zone Indicator Panel and Isolation Valves, Vacuum (0-30” Hg) A20VXGAUGE100 Gauge for Zone Indicator Panel and Isolation Valves, Gas (0-100 psi)

A20VXGAUGE300 Gauge for Zone Indicator Panel and Isolation Valves, Nitrogen (0-300 psi)

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 25: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

LCD RETRO-FIT ALARM

RETRO-FIT ALARMS • 10" [25.4 cm] LCD Screen • LCD alarm panel incorporates the latest

microprocessor based technology • Smallest 8 gas alarm in the industry • Auto detecting gas sensors • Digital sensors can be mounted locally or remotely

utilizing a #22 gauge stranded, shielded, twisted pair cable ONLY

• Sensors housed in a solid, tamper-proof enclosure to act as a barrier against any interference

• LCD brightness and volume is field adjustable • LCD display readable in poor lighting conditions • Self diagnostic and error message display for ease of maintenance • Customizable screen text for displaying gas locations • High / Low alarm set points for each gas are field djustable

GENERAL SPECIFICATIONS The Digital LCD Area Alarm system shall be an Amico Alert-3 series, complete with a five-year warranty. The LCD alarm shall be microprocessor based with a 10" [25.4 cm] screen and capable of monitoring up to 8 sensors. Sensors will be automated for gas specific detection. Each sensor shall be gas specific and an error message shall be displayed for an incorrect connection. Each specific gas service shall have a digital read-out comprising of 0-249 psi [0-1,717 kPa] for pressure and 0-30 inHg [-100-0 kPa] for vacuum. The digital read-out shall provide a continuous indication of each service being measured. A separate indicator shall be provided for each service indicating a green “NORMAL” and a red “HIGH” or “LOW” alarm condition. If an alarm occurs, the green indicator will change to red and a continuous audible alarm will sound. Pushing the (mute button/push to test button) will cancel the audible alarm, but the unit will remain in the alarm condition until the problem is rectified. The default set-points shall be +/- 20% variation from normal condition. In the calibration mode High/Low set-points shall be adjustable by on board push buttons. To view the set points and audible alarm sound level, press and hold the mute button for twenty (20) seconds. The frame assembly can be added to an existing alarm panel in the field with the addition of a filler frame. LCD Sensor operating pressure range: Mid-Pressure: (0 psi to 99 psi) - OXY, AIR, N₂O, CO₂ Vacuum: (0 inHg to 30 inHg) - VAC, WAGD, AGSS High-Pressure: (0 psi to 249 psi) - NIT, IAR Input power to the Amico Alert-3 alarm is: 115-220 VAC, 50-60 HZ. The Amico Alert-3 LCD Alarm is cETLus listed to UL standard 1069 and CSA standard C22.2 No. 205. Amico products comply with NFPA 99, CSA Z7396.1 and CSAZ305.1. The Amico Alert-3 LCD Alarm complies with the following electromagnetic compatibility standards: FCC Part 15 Class A and ICES-

003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 26: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20A3ACEXXXXXXXX

Example: If the Alarm is converting from an Ohio Alarm consisting of oxygen, medical air and vacuum then the following part number is applicable for the conversion = A20A3ACEOAV Notes: 1. An LCD FIller Frame is required if the existing alarm width is more than 10 inches. 2. Specify if converting from digital or analog. Please also note if there are any special requirements. 3. An Ohio Alarm requires a trim plate to cover the bottom gap.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

X Defines the Type of Gas

O Oxygen C Carbon Dioxide A Medical Air W WAGD (NFPA) V Medical Vacuum E AGSS (ISO) 2 Nitrous Oxide I Instrument Air N Nitrogen

E – English (CSA/ISO)

C Conversion (Retrofit)

IMPORTANT NOTE: Due to unknown factors of the existing equipment, projects requiring Retro-Fit Conversions must be fully investigated by the installer. It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are required. Amico is not responsible for any additional components required.

FILLER FRAME PART NUMBERS A20A3PLCDFILLER4 Alert-3 LCD Conversion Alarm Filler Frame 4-Station A20A3PLCDFILLER5 Alert-3 LCD Conversion Alarm Filler Frame 5-Station A20A3PLCDFILLER7 Alert-3 LCD Conversion Alarm Filler Frame 7-Station

Page 27: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

MASTER ALARM

CONVERSION (RETRO FIT)

RETRO-FIT ALARMS • Microprocessor based with an individual

microprocessor on each module • A green LED indicator for normal and a red LED

indicator for abnormal conditions • Up to 60 functions in a standard configuration • LED Alarm lights utilized for long life • Area Alarm Modules can be intermixed with

Master Alarm Modules • Maintenance mode for ease of troubleshooting • Self-diagnostic circuitry for added reliability • Modules can be upgraded in the field to interface

to a building management system or slave alarm • Alarm buzzer in excess of 90 decibels • All modules to be mounted on a hinged frame for easy accessibility • Last alarm flashes, acknowledged alarm shows continuous RED signal • Repeat alarm adjustable from 1, 12, 24 hours or off • Dry contacts for remote monitoring of high and low alarms and the distance between the master

module and source equipment can be up to 10,000 ft. (3,000 m)

GENERAL SPECIFICATIONS The Master Alarm system shall be an Amico Alert-2 series, complete with a five-year warranty.

Each module shall be microprocessor based and field adjustable. Each signal contains a green LED indicator for normal and a red LED

indicator for abnormal conditions. When a fault occurs, the green LED turns off, the red LED illuminates and the audible alarm sounds.

The red LED flashes until the front panel alarm mute button is pressed. After the alarm mute button is pressed, the red LED remains

illuminated but does not flash. The red indicator automatically turns off and the green LED illuminates when the fault is corrected.

A maintenance mode shall, when enabled, latch the alarms, requiring a reset after the alarm condition has been rectified. This is to

assist in tracking down wiring problems or faulty field devices. The master alarm shall identify the last alarm condition by flashing, while

the already acknowledged alarm shows a continuous red signal.

A repeat alarm function shall, when enabled, be capable of turning on the buzzer again, after a preset time, if the fault condition has

not been rectified.

Each module shall handle 10 functions and up to 6 modules can be accommodated per standard box for a total of 60 functions. Master

alarms shall be modular in construction and shall be capable of adding extra modules in the field.

The Alarm system shall be a closed circuit self monitoring type. A green “POWER” light shall provide indication that the unit is

energized. In addition, “TEST” and “ALARM MUTE” buttons shall be easily accessible to operate and test the unit.

Every module shall be field upgradable to allow for interfacing to a building management system with the addition of an add-on circuit

board which plugs into the master module.

Input power to the Amico Alert-2 alarm is: 115 VAC to 220 VAC, 50 to 60 HZ.

The alarm is UL Listed. Amico products comply with NFPA-99, CSA Z7396.1 and CSA Z305.1.

The Amico Alert-2 Alarm complies with the following electromagnetic compatibilty standards: FCC Part 15 Class A and ICES-003 Class A.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 28: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION A20A2MCEXX

NOTE: Please specify which type of alarm you are converting from (Digital or Analog). Example: Ohio 20 points master alarm converted to Amico = A20A2MCE20.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Available in 10 points through 60 points

E – English (CSA/ISO)

C Conversion (Retrofit)

20 functions = A20A2MCE20 4 gang back box

60 functions = A20A2MCE60 7 gang back box

Page 29: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

MULTIPLE ZONE

INDICATOR PANEL BOX CSA

The zone shut-offs shall be 3 piece, ball-type design with a brass forging body and a chrome plated, brass ball for sizes 1/2" to 2". Ball seats, stem seals and stem washer shall be reinforced Teflon (PTFE), with Viton stem and flange O-rings. A blow-out proof stem shall be used and the zone shut-off shall have a maximum pressure rating of 600 psi (4,137 kPa). All zone shut-offs materials shall be compatible with medical gases or vacuum service to 29" Hg (98.205 kPa).

ORDERING INFORMATION A20ZIEMXXXXXXX

Example: Triple box with gauge; English (CSA/ISO); 1/2", 3/4" and 1-1/4" zone shut-off; Monitoring = A20ZIEM124

NOTES: All zone shut-offs are full port. 1/2" to 1-1/2" shut-offs in a multiple zone indicator panel box will be in a 3-7/8" deep box. Shut-offs 1-1/2" and larger will have an empty space above the shut-off. Please refer to the "Single Zone Indicator Panel Assembly" spec sheet for 2-1/2" and 3" assemblies. Example: A box with a 1/2", 3/4" and 1-1/2" shut-off will require a 4 hole box not a 3 hole box. The 2" shut-off will be assembled in a 6-½" deep box and shall accommodate up to a maximum of 5 gases. The 2" shut-off shall be positioned on the bottom of the assembly. Refer to 2" zone indicator box assembly

drawing on page 3.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Replace X with Size Required

½” = 1 ¾” = 2 1” = 3 1¼” = 4 1½” = 5 2” = 6

E – English (CSA/ISO)

M = Monitoring (with gauge)

Page 30: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GENERAL SPECIFICATIONS

Zone Indicator Panel Boxes shall be an Amico Alert-1 series. Each recessed zone indicator panel box shall consist of the following components: A steel box which can house two to seven zone shut-offs with tube extensions, an aluminum frame and a pull-out removable window. Gauges are included. The indicator panel box shall be constructed of 18 gauge steel complete with a baked white enamel finish. Affixed to the opposite sides of the box will be two adjustable steel brackets for the purpose of mounting the box to the structural support. The steel brackets shall accommodate various finished wall thicknesses between 3/8" (9.5 mm) and 1-3/16" (30 mm) and shall be field adjustable. The frame assembly shall be constructed of anodized aluminum and shall be mounted to the back box assembly by standard #6 * 3/8" tapping screws as provided. The removable front shall consist of a window with a pull-out ring pre-mounted to the center of the window. Access to the zone shut-offs shall be by merely pulling the ring assembly to remove the window from the frame. The window can be reinstalled without the use of tools only after the shut-off handles have been returned to the open position. The window shall be marked to prohibit unauthorized people from tampering with the zone indicator panel with the following silk-screen caution: “MEDICAL GAS CONTROL VALVES CLOSE ONLY IN EMERGENCY” The zone shut-offs shall be 3 piece, ball-type design with a brass forging body and a chrome plated, brass ball for sizes 1/2" to 2". Ball seats, stem seals and stem washer shall be reinforced Teflon (PTFE), with Viton stem and flange O-rings. A blow-out proof stem shall be used and the zone shut-off shall have a maximum pressure rating of 600 psi (4,137 kPa). All zone shut-offs materials shall be compatible with medical gases or vacuum service to 29" Hg (98.205 kPa). Zone shut-offs shall be operated by a lever-type handle requiring only a quarter turn from a fully open position to a fully closed position. All zone shut-offs shall be equipped with type “K” washed and degreased copper pipe stub extensions of sufficient length to protrude beyond the sides of the box. Each shut-off shall be supplied with an identification bracket which shall be riveted to the zone indicator panel box for the purpose of applying an approved medical gas identification label. A package of labels shall be supplied with each Zone Indicator Panel box assembly for application by the installer. Amico products comply with CSA Z7396.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 31: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 32: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

OHMEDA RETRO-FIT

GAS CONTROL PANEL

RETRO-FIT GAS CONTROL PANELS • Inlet and outlet pressure display

gauges in psi/kPa • Manual shut-off valve • Outlet supply pipe for remote

mounted outlets • 3/8" (9.5 mm) internal tubing for

high flow • High flow capacity regulator • Aluminum front simplifies cleaning • Maximum supply pressure of 250

psi (1,724 kPa) • Maximum delivery pressure:

DISS: AIR CO₂ = 80 psi NIT, IAIR = 200 psi HP (Safety Swing Coupling) = 250 psi

GENERAL SPECIFICATIONS The Gas Control Panel Retro-Fit kit allows you to change from Chemetron or Ohmeda to Amico Alert-1 series without breaking the pipeline or cutting into the existing wall. The control panel shall be supplied with a quarter turn shut-off ball valve, rated at no less than 300 psi (2,069 kPa). The unit shall have an inlet pressure indicator with a 0 to 400 psi (0 to 2,750 kPa) valve to indicate the inlet supply pressure. The pressure regulator shall be adjustable between 0 to 250 psi (0 to 1,724 kPa). The outlet controlled pressure shall be monitored by a pressure indicator with a range of 0 to 400 psi (0 to 2,750 kPa). The DISS outlet shall be a Diameter Index Safety System (D.I.S.S.) for Air or Nitrogen Service Outlet or “Safety Swing Coupling” for pressures above 200 psi. The outlets shall be used for connection to pneumatic surgical tools. The mounting frame is made from 18 gauge carbon steel that has been painted for aesthetic appeal. The unit shall be factory tested for the intended gas service. The Gas Control Panel is UL Listed to U.S. and Canadian safety standards. Amico products comply with NFPA-99 and CSA Z7396.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 33: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

ORDERING INFORMATION

ITEM DESCRIPTION

A20NCONCERPB Nitrogen Control Panel Retro-Fit - Puritan - Bennett

A20NCONCERCH Nitrogen Control Panel Retro-Fit-Chemetron (English, relieving)

A20NCONCEROH Nitrogen Control Panel Retro-Fit-Ohmeda (English, relieving)

A20NCONCEHPROH Nitrogen Control Panel High Pressure Retro-Fit - Ohmeda - English, relieving (NFPA)

A20NXCONADAB25 High pressure adapter with 1/4" Hose barb (supplied with HP Gas Control Panel)

IMPORTANT NOTE: Due to unknown factors of the existing equipment, projects requiring Retro-Fit conversions must be fully investigated by the installer.

It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are required.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 34: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

SINGLE ALARM VALVE

COMBO UNIT

COMBINATION ALARMS • 2 in 1 design combines the Area Alarm and Zone Valve Box to compensate for

space restrictions • True digital, illuminated LED display is readable even in poor lighting conditions • High/low alarm set-points are field adjustable for each gas service • Repeat alarm is adjustable from 1 to 60 minutes or off • Gas-specific sensor with DISS nut and nipple • Alarm buzzer in excess of 90 decibels • Each module is marked with an approved medical gas identification label • Pre-mounted pull-out ring allows for ease of maintenance • Individual microprocessor for each display and sensor module; digital sensor is

mounted locally • Dry contacts for remote monitoring of the high and low alarm

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 35: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

GENERAL SPECIFICATIONS

The Alarm Valve Combo Unit shall be an Amico Alert-1 Series.

Each AVC shall consist of the following components: An 18 Gauge steel valve box complete with a baked white enamel finish which can house one shut-off ball valve with tube extensions, gas specific sensor (DISS nut and nipple connection), and a hinged gas specific compact alarm with illuminated LED digital display. It also includes an error message for an incorrect connection, an aluminum frame and a pull-out removable window.

Affixed to the opposing sides of the box will be two adjustable steel brackets for the purpose of mounting the box to the structural support. The steel brackets shall accommodate various finished wall thicknesses between 3/8" [9.5 mm] and 1-3/16" [30 mm] and shall be field adjustable. The frame assembly shall be constructed of anodized aluminum and shall be mounted to the back box assembly by standard #6 x 3/8" tapping screws as provided.

The digital alarm shall read from 0-250 psi [0-1,724 kPa] for pressure and 0-30"Hg [-100-0 HgkPa] for vacuum. The digital read-out shall provide a constant indication of each service being measured. It will indicate a green “NORMAL” and a red “HIGH” or “LOW” alarm condition. If an alarm occurs, the “RED” alarm light shall flash and the audible alarm (which exceeds 90 decibels) will sound. Pushing the “ALARM MUTE” button will cancel the audible alarm, but the unit will remain in the alarm condition until the problem is rectified. When alarm is enabled on the compact alarm module, the repeat alarm function shall be capable of turning on the buzzer again (after a present time) if the fault condition has not been rectified.

Each alarm module shall be individually microprocessor based and be field adjustable. The default set point on this alarm shall be +/- 20% variation from the normal range. In the calibration mode the following parameters shall be adjustable: High/Low set-points, Imperial/Metric units, Repeat alarm Enable/Disable (adjustable from 1 to 60 minutes), psi and kPa or Bar readout (switch selected). Set points shall be adjustable by two on-board push buttons.

Access to the shut-off valves shall be by merely pulling the ring assembly to remove the window from the frame. The window can be reinstalled without the use of tools only after the valve handles have been returned to the open position. The window shall be marked to prohibit unauthorized persons from tampering with the valves with the following caution:

“Medical Gas Control Valves with Alarms” “Close Valves Only in Emergency”

The valve shall be of a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 3" [1.3 cm to 7.6 cm]. Ball seats, stem seals and stem washer shall be Reinforced Teflon (PTFE) with Viton stem and flange O-rings. A blow-out proof stem shall be used and the valve shall have a maximum pressure rating of 600 psi [4,137 kPa].

Valves shall be operated by a lever-type handle requiring only a quarter turn from a fully open position to a fully closed position. All valves shall be equipped with type “K” washed and degreased copper pipe stub. Each valve will be identified for gas specification as indicated on the hinged alarm label.

Input power to the Amico Alarm Valve Combo Unit is: 115 VAC to 220 VAC, 50 to 60 HZ.

Amico products comply with NFPA 99 and CSA Z7396.1.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 36: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

TECHNICAL SPECIFICATIONS

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

Page 37: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ORDERING INFORMATION A20AVEXG

NOTE: All valves are full port. The power supply will require 1 gang per box. Example: English, 1 valve, 1/2" [1.3 cm] Oxygen will be in a 2 gang box. AVU-1O. The 1-1/2" [3.8 cm] valve will have an empty space above the valve. Example: A box with 1-1/2" [3.8 cm] Medical Vacuum and a power supply, will require a 3 gang box not 2 gang. The circuit board shall be positioned above the valve. Refer to 1/2" and 1-1/2" [1.3 cm and 3.8 cm] assembly drawing. The 2" [5.1 cm] valve will be assembled in a 6-1/2" [16.5 cm] deep box. The 2" [5.1 cm] valve shall be positioned on the bottom of the assembly, with the power supply above the valve. The 2-1/2" [6.4 cm] and 3" [7.6 cm] valve assembly will be assembled in a 15-1/2" [39.4 cm] wide x 6-1/2" [16.5 cm] deep box. With the valve positioned on the bottom and the power supply on top. Refer to the 2-1/2" [6.4 cm] and 3" [7.6 cm] assembly drawing.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

G Defines the Type of Gas

O Oxygen A Medical Air V Medical Vacuum 2 Nitrous Oxide N Nitrogen C Carbon Dioxide W WAGD (NFPA) E AGSS (ISO) I Instrument Air

E – English (CSA/ISO)

X Defines the Valve Size

1 ½” (1.3 cm) 05

2 ¾” (1.9 cm) 07

3 1” (2.5 cm) 10

4 1¼” (3.2 cm) 12

5 1½” (3.8 cm) 15

6 2” Full Port (5.1 cm) 20F

7 2½” (6.35 cm) 25

8 3” (7.62 cm) 30

Page 38: COMBINATION ALARMS · 2019. 8. 12. · The valve shall be a 3 piece ball-type design with a brass forging body and a chrome-plated brass ball for sizes 1/2" to 2" [1.3 cm to 5.1 cm].

ADVANCED VACUUM CLEANING SOLUTION

Advanced Vacuum Cleaning Solution is designed to remove debris from vacuum system outlets and all subsidiary vacuum components. After time and heavy use, vacuum outlets can become restricted by the buildup of debris. These buildups can easily be removed by inserting a small amount of Advanced Vacuum Cleaning Solution into the outlet. This process may eliminate the need to isolate, disassemble and rebuild the outlet. When used as directed Advanced Vacuum Cleaning Solution can increase the restricted vacuum flow of sluggish medical vacuum system outlets, adapters, hose assemblies, etc. Advanced Vacuum Cleaning Solution is ready-to-use and works on partially plugged outlets and equipment.

ADVANCED VACUUM CLEANING SOLUTION MAY BE USED FOR ALL VACUUM SYSTEM APPLICATIONS, INCLUDING: - Operating rooms

- Recovery Rooms

- Patient rooms

- Labor and delivery rooms

- Vacuum regulators

- Portable vacuum units

- Vacuum adapters

- Vacuum hose assemblies

Instructions for Use: Advanced Vacuum Cleaning Solution is safe, easy, ready-to-use and is designed to work on partially plugged outlets/equipment. 1. Use an appropriate adapter and tubing (small I.D. tube is recommended) 2. Suction approximately 1 cup of cleaning solution from the container and

apply to each outlet. 3. Let stand for approximately 15 minutes to allow the solution to soften

existing debris 4. After the solution has softened debris, suction up approximately 1 cup of

warm water and apply to each outlet, flushing away the cleaning solution. 5. Repeat if necessary 6. If problem persists or the outlet/line is completely plugged, disassembly

and repair may be required.

WARNING Follow the above steps carefully. NEVER apply large amounts of liquid. The liquid can be pulled back into the vacuum pumps. Large amounts of liquid could damage vacuum source equipment

WARNING NEVER use any flammable product such as alcohol to flush vacuum lines or equipment.

Advanced Vacuum Cleaning Solution: Clears sluggish medical vacuum system outlets and lines. Increases restricted vacuum flow. Cleans vacuum system components and accessories. Easy to use. Eliminates costly downtime. Safe - when used as directed - for all vacuum components. When used as instructed Advanced Vacuum Cleaning Solution

is harmless to vacuum pumps, outlets, hose assemblies and other components.

DNR Medical 20448 Walnut Crescent Email: [email protected] Phone: 604-308-1633 Maple Ridge BC V2X 0M1 www.dnrmedical.com

PART NUMBER

A055100

** LOW FOAMING **

** NEW IMPROVED FORMULA **

1 GALLON SIZE