Coal Analyzer Performance. Coal analyzer performance. The performance of the coal analyzer is dependent on a number of site specific variables such as: Range of quality that will be seen Belt loading Time period used for analysis Reference system the analyzer is compared to. - PowerPoint PPT Presentation

Transcript of Coal Analyzer Performance

Page 1: Coal Analyzer Performance

The world leader in serving science

Coal Analyzer Performance

Page 2: Coal Analyzer Performance

2

Coal analyzer performance

The performance of the coal analyzer is dependent on a number of site specific variables such as:• Range of quality that will be seen• Belt loading• Time period used for analysis• Reference system the analyzer is compared to

Page 3: Coal Analyzer Performance

3

CQM Performance example

Suflur

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0.0 0.2 0.4 0.6 0.8 1.0

CQM

La

b

CQM Precision: 0.025

Ash

6.0

7.0

8.0

9.0

10.0

11.0

12.0

6.0 7.0 8.0 9.0 10.0 11.0 12.0

CQM

La

b

CQM Precision: 0.204

One-hour Comparisons

Sulfur

As reported data

As reported data

Page 4: Coal Analyzer Performance

4

ECA performance example

As Received Ash

7.00

8.00

9.00

10.00

11.00

12.00

13.00

14.00

15.00

16.00

7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00

ECa

LAB

Standard Error = 0.54

As Received Sulfur

0.60

0.70

0.80

0.90

1.00

1.10

1.20

1.30

1.40

1.50

0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50

ECa

LA

B

Standard Error = 0.050

ECA 1000 on a 42-inch clean coal conveyor belt.

As reported data

As reported data

Two-hour Comparisons

Page 5: Coal Analyzer Performance

5

Analyzer Comparison

1.55 1.6 1.65 1.7 1.75 1.8 1.85 1.9 1.95 2 2.050

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

Control Variable

Pro

ba

bili

ty

CQM: Setpoint = 1.831812C: Setpoint = 1.82ECA: Setpoint = 1.80

Page 6: Coal Analyzer Performance

6

Thermo Coal Analyzer typical guarantees

ASTM CQM Typical ECA typical

Reproducibility* Guarantee Guarantee

(1 sigma) (1 sigma) (1 sigma)

Total Moisture

less than 5% Total Moisture 0.15 0.50** 0.75**

5% or greater Total Moisture 0.25 0.50** 0.75**

As Received Ash

8% M & 8% dry ash 0.273 0.35 0.60

8% M & 16% dry ash 0.483 0.50 0.75

As Received Sulfur

Average sulfur = 1.50% 0.078 0.035 0.07

Average Sulfur = 2.25% 0.111 0.05 0.08

* ASTM reproducibility considers laboratory precision only – does not include sample collection precision.

** Thermo will pass through the moisture guarantee of the Moisture Meter Manufacturer

Range will also effect the guarantee offered. Guarantees above are for typical shipment ranges.